Oakton pHTestr 30 Instruction Manual

Operating Instructions
®
pHTestr
Large Screen Waterproof pH/Temperature Tester Double-Junction Sensor
Applications
• Aquaculture
• Aquariums
• Boilers
• Car washes
• Ecology studies
Before You Begin
Your instrument has been factory calibrated and usually works well out of the box. However, after extended periods of non-use, it’s best to soak the sensor in electrode storage solution, pH buffer for 10 minutes or so. A brief rinse with deionized (DI) water is OK, but avoid soaking or storing in deionized water as this will shorten the pH electrode life. Prior to taking measurements, periodic calibration with certified standards is recommended for best accuracy.
pH Buffer Set Selection
pHTestr features USA (pH 4.01, pH 7.00 and pH 10.01) or NIST (pH 4.01, pH 6.86, and pH 9.18) standards. Select either one to suit your requirements.
1. While pressing the HOLD/ENT button, switch on the unit by pressing the ON/OFF button.
2. Release the HOLD/ENT button. The display will flash either USA or NIST.
3. Press CAL button to toggle between the two buffer set standards.
4. Press the HOLD/ENT button to confirm the selection of the buffer set.
With meter powered off, press HOLD/ENT and ON/OFF at the same time. First release ON/OFF button, then HOLD/ENT button.
30 Pocket Tester
• Food sectors
• Hydroponices
• Labs
• Pools
• Sanitation plants
Figure 1: Buffer Selection Sequence
• Spas
• Steam generators
• Water & wastewater
treatment
Water quality testing
• More!
pH Calibration
Calibration should be done regularly, preferably once a week. You can calibrate up to three points using either the USA or the NIST buffer set standards.
1. Press ON/OFF button to switch unit on.
2. Dip electrode about 2 to 3 cm into the pH standard buffer solution.
3. Press the CAL button to enter calibration mode. The ‘CAL’ indicator will be shown. The upper display will show the measured reading based on the last calibration while the lower display will indicate the pH standard buffer solution.
Note: To abort calibration, press the CAL button.
4. Allow about 2 minutes for the reading to stabilize before pressing the HOLD/ENT button to confirm the first calibration point. The upper display will be calibrated to the pH standard buffer solution and the lower display will then search for the next pH standard buffer solution.
5. Repeat with other buffers if necessary. Rinse electrode before dipping into next buffer.
Note: The calibration mode allows you to perform up to three calibration points before returning to the measurement mode automatically. However, if you opted to have only one or two calibration points, simply skip the remaining calibration points by exiting to the measurement mode by pressing the CAL button.
To do a 1 point calibration only, press CAL button at this point to exit to the measurement mode. Otherwise, proceed to second buffer for a second buffer for a second point calibration.
To do a 2 point calibration only, press CAL button at this point to exit to the measurement mode. Otherwise, proceed to third buffer for a third buffer for a third point calibration.
After the third point calibration, the meter will automatically return to the measurement mode. At any point, an error message ‘Er. 1’ will be displayed momentarily if the confirmed pH value is not within the pH calibration window.
Figure 2: Example of pH Calibration Sequence
pH Measurement
1. Press the ON/OFF button to switch the unit on.
2. Dip the electrode about 2 to 3 cm into the test solution. Stir and let the reading stabilize. For semi-solid samples using pHSpear, rotation is ideal for sample contact.
CAUTION: Testing dry samples is not accurate and can lead to sensor damage or breakage. Soils must be wet and free of particulates that may scratch the glass sensor. Excessive force into dry samples can cause glass breakage.
3. Note the pH value or press HOLD/ENT button to freeze the reading. To release the reading, press HOLD/ENT again.
4. Press ON/OFF to turn off unit. If you do not press a button for 8.5 minutes, the unit will automatically shut off to conserve batteries.
HOLD Function
This feature lets you freeze the display for a delayed observation.
1. Press HOLD/ENT button to freeze the measurement. A ‘HOLD’ indicator will be displayed and the measurement will be frozen.
2. Press HOLD/ENT again to release the measurement. The ‘HOLD’ indicator will not be displayed anymore indicating the held measurement is released.
Figure 4: Example of HOLD Function
User Reset
You can reset the pH calibration to the factory default by using the user reset function. Buffer set selection and temperature user calibration are not affected by the user reset function.
1. Switch off the unit.
2. While pressing the CAL button, press and release the ON/OFF button to enter the ‘User Reset’ selection menu. The screen will display ‘rSt’ on the bottom display with a flashing ‘nO’ selection.
3. Use the CAL button to toggle between ‘nO’ and ‘YES’ selection.
• nO deactivates reset selection
• YES activates the reset selection
4. Press the HOLD/ENT button to confirm the selection made.
5. If you have selected ‘YES’, the unit will show ‘CO’ momentarily and proceed to the measurement mode with the calibration reset back to factory default value.
6. If ‘nO’ is selected, the unit will proceed to the measurement mode without any calibration reset performed.
With meter powered off, press CAL and ON/OFF at the same time. First release ON/OFF button, then CAL button.
Figure 5: User Reset Sequence
After reset, whole LCD segment lits up and goes back to measurement mode.
ATC - Automatic Temperature Compensation
Through its built-in temperature sensor, the measurement error due to the changes in electrode sensitivity due to changes in temperature is compensated to give the actual pH reading of the sample measured.
Note: To exit this program without confirming the calibration, press the CAL button before the automatic confirmation takes place.
1065O100_MAN_35634-30.indd 1 8/25/2017 7:15:07 AM
Temperature Calibration
From the measurement mode,
1. Press the HOLD/ENT button to bring the unit to the ‘HOLD’ mode.
2. Press the CAL button for 3 seconds to switch to the °C or °F mode setting selection screen. Pressing the CAL button continuously for 3 seconds allows you to toggle in between the °C and °F mode setting selection screen.
3. Release the CAL button to confirm your mode selection and the display will go to the temperature calibration mode with the upper display flashing. The upper display shows the current measured temperature reading based on the last set offset and the lower display shows the current measured temperature reading based on factory default calibration.
4. Dip the unit into a solution of known temperature and allow time for the built-in temperature sensor to stabilize.
5. Press the HOLD/ENT button to set the upper display to the temperature value of the solution.
6. Once the new temperature setting is reached, the new value is automatically confirmed and returns to the measurement mode if no button is pressed after 5 seconds.
Note: To exit this program without confirming the calibration, press the CAL button before the automatic confirmation takes place.
Continuous pressing of CAL button (about 3 seconds) to toggle between temperature setting of degrees Celsius and Farenheit
Release CAL button to start entering the new temperature
During calibration, upper display will flash. Press HOLD/ENT to increase offset up to +5 °C (+9 °F) from default value. It will roll over to -5 °C (-9 °F) from default value and start increasing again.
Figure 6: Temperature Calibration Sequence
When the temperature offset value is reached, the new value is automatically confirmed if no key is pressed for 5 seconds.
Electrode Maintenance
1. Rinse the electrode with electrode storage solution after each measurement. Care has to be taken not to damage the sensor’s glass electrode.
2. In aggressive chemicals, dirty or viscous solutions, and solutions with heavy metals or proteins, take readings quickly and rinse electrode immediately
afterward. Mild detergent can be used to wash the electrode clean.
3. If possible, keep a small piece of paper or sponge in the electrode cap— moistened with clean water or electrode storage solution (NOT de-ionized water)—and close the cap over the electrode.
Self-Diagnostic Messages
3 Bars indicates battery is full (100%)
Low Battery Indicator
2 Bars indicates 50% of the battery life is left
1 Bar indicates 25% of the battery life is left
Blinking battery casing indicates the need to replace batteries with fresh ones as specified by manufacturer
Electrode is not in contact with solution or electrode is failing
Over Range/ Under Range Signal
Error
Message
Or / Ur
(Still)
ATC / Or / Ur
(Blinking)
Er.0
Er.1
Replacement sensor is not connected properly to the unit during sensor replacement
Measured pH value or temperature value exceeds its specified maximum or minimum value
Blinking ‘ATC’, ‘Or’ or ‘Ur’ indicates that there is a short or open circuit at the built in temperature sensor
Temperature calibration error of attempting to calibrate unit to a value which is out of range or under range
pH calibration error of attempting to confirm a calibration value which is not within the specified calibration window
Changing Batteries
1. Open battery compartment lid (with attached lanyard loop).
2. Remove old batteries; replace with fresh ones. Note polarity.
Electrode Replacement
You can replace the electrode module at the fraction of the cost of a new tester. When the tester fails to calibrate or gives fluctuating readings in calibration standards, you need to change the electrode.
1. With dry hands, grip the ribbed collar with electrode facing you. Twist the collar counterclockwise (see picture A). Save the ribbed collar and O-ring for later use.
2. Pull the old electrode module away from the unit.
3. Align the four tabs on the new module so that they match the four slots on
the unit (see picture B).
4. Gently push the module onto the slots to sit it in position. Push the smaller O-ring fully onto the new electrode module. Push the collar over the module and thread it into place by firmly twisting clockwise.
Note: It is necessary that you recalibrate the tester prior to measurement after an electrode replacement.
Rotate collar away from you
Rlbbed collar
Figure 7: Removal of collar
from tester
Insert Electrode Module
Electrode Module
Small O-ring
Figure 8: Example of electrode
Small Tab Large Tab
module fitting alignment
Large O-ring
Warranty
The waterproof pHTestr is warranted to be free from manufacturing defects for 1 year and electrode module for 6 months, unless otherwise specified. If repair, adjustment or replacement is necessary and has not been the result of abuse or misuse within the time period specified, please contact your local distributor for assistance.
Return Of Items
Authorization must be obtained from your distributor before returning items for any reason. When applying for authorization, please include information regarding the reason the item(s) are to be returned.
Note: We reserve the right to make improvements in design, construction and appearance of products without notice. Prices are subject to change without notice.
Accessories
Item Order Code
pHTestr 30 pocket tester 35634-30 Replacement sensor for pHTestr 30 35624-38 Replacement button cell batteries, 1.5 V. Pack of 6 09377-16 NIST-traceable calibration with data for pH pocket testers 17101-45
Specifications
Description pHTestr 30
pH range –1.00 to 15.00 pH Resolution 0.01 pH Relative accuracy ±0.01 pH Calibration points Up to 3 points
Buffer set standard selection Calibration window
(USA buffer set standard) Calibration window
(NIST buffer set standard) Temperature 0 to 50.0°C / 32.0 to 122.0°F Temperature compensation ATC Temperature resolution 0.1°C / 0.1°F Temperature accuracy 0.5°C / 0.9°F Temperature calibration window Auto off After 8.5 minutes from last key press User reset Yes Nonvolatile memory backup Yes LCD Dual Power requirement 4 x 1.5 V “A 76” micro alkaline batteries Battery life More than 500 hrs Operating temperature 0 to 50°C / 32 to 122°F Dimensions 6.5"L x 1.5"dia. (165 x 38 mm) Weight 3.25 oz (90 gm)
USA: 4.01/7.00/10.01
NIST: 4.01/6.86/9.18
±1.00 pH (pH 4.01 & pH 10.01),
±1.50 pH (pH 7.00)
±1.00 pH (pH 4.01 & pH 9.18),
±1.25 pH (pH 6.86)
±(5°C / 9°F) from default value
www.4oakton.com
1065O100_MAN_35634-30 August, 2017
1065O100_MAN_35634-30.indd 2 8/25/2017 7:15:08 AM
Loading...