11. RETURN OF ITEMS ......................................................... 18
1
1. INTRODUCTION
Thank you for purchasing our pH 700 / Ion 700 series benchtop
meter. These microprocessor-based meters are economical and
simple to use. The design incorporates a large LCD for clear
viewing, yet offers a small footprint to conserve space.
The pH 700 measures pH, mV (ORP), and temperature (ºC or ºF).
In addition to these parameters, the Ion 700 features direct ion
concentration measurement of various ions (mono and divalent).
Each meter includes a convenient slide-out card for
quick reference.
Some configurations include an electrode arm and metal bracket
which can be easily attached to the left or right side of the meter
according to your preference.
The 700 series benchtop meters replace our popular 510 series
meters that were introduced in 2000.
We take great pride in every instrument we manufacture and
hope this one serves you well.
2
2. GETTING STARTED
2.1 Keypad Functions
Powers the meter on and off. Upon power on, the meter
automatically begins in the mode that was last used.
Calibration and memory values are retained even if meter
is unplugged.
Toggle between available measurement modes;
pH/Temp, mV/Temp, or ppm/mV (Ion 700 only).
Also used to switch from pH to Temp during pH
calibration mode.
Press and hold for 5 seconds to enter SETUP mode.
Toggles between measurement and calibration modes.
In SETUP mode, returns user to the measurement mode.
MI (Memory Insert) stores values into memory.
Increase value. Scroll up in SETUP mode.
MR (Memory Recall) recalls values from memory
Decrease value. Scroll down in SETUP mode.
Freezes measured reading. Press again to resume
live reading.
Confirms calibration values in CAL mode.
Confirms selections in SETUP mode.
View recalled values in memory mode.
3
2.2 LCD Annunciators
2.3 Meter Connections
pH
REF
ATC
DC
BNC connection for pH, ORP (Redox), or Ion Selective
Electrodes (ISE).
Pin connection for half cell reference electrodes.
Requires separate half cell BNC electrode.
Note: REF is not commonly used and is not required.
For Automatic Temperature Compensation probe.
Power supply.
4
3. CALIBRATION
3.1 pH Calibration
For best results, periodic calibration with known accurate standards is
recommended. Calibrate with standards that bracket your intended
measuring range while including a neutral standard (pH 7.00 or 6.86).
For example, if you expect to measure samples from pH 6.2 to pH 9.5,
calibration with 4.01, 7.00, and 10.01 will work well.
The 700 series meters can be calibrated with up to 5 buffers. The
non-volatile memory retains all calibration values upon meter shut down.
The following calibration standards are automatically recognized;
To eliminate temperature errors associated with the pH electrode, attach
the automatic temperature compensation (ATC) probe for best accuracy.
Without temperature compensation, pH accuracy will worsen as samples
deviate from 25ºC and pH 7.
USA buffer group 1.68, 4.01, 7.00, 10.01, 12.45
NIST buffer group 1.68, 4.01, 6.86, 9.18, 12.45
See Section 5.6 to change the buffer group
1) Press
2) Dip the pH and ATC electrodes into pH buffer and press
The secondary display will lock on the appropriate buffer value. Provide
stirring for best results. When the READY indicator appears, press
to accept. The primary reading will flash briefly before the secondary
display begins scrolling the remaining available buffers.
3) Rinse the pH and ATC electrodes then dip into the next pH buffer. The
secondary display will lock on the appropriate buffer value. When the
READY indicator appears, press
will flash briefly then display the percent efficiency (slope) before the
secondary display begins scrolling the remaining available buffers.
as needed to select pH.
to accept. The primary reading
5
4) To calibrate another buffer repeat step 3) or press to return to the
measurement mode. Note: The meter will automatically return to
measurement mode upon successful completion of the number of
specified calibration points. To specify a different number of pH
calibration points see Section 5.6.
Additional notes:
A single point (offset) calibration is only allowed with pH 7.00 or
pH 6.86 buffers.
When the first calibration value is accepted during a new calibration,
all prior calibration values are erased.
Press
mode.
3.2 Temperature Calibration
The thermistor sensor used for automatic temperature compensation and
measurement is both accurate and stable, so require frequent calibration
isn’t required. Temperature calibration is recommended upon electrode
replacement, whenever the temperature reading is suspect, or if matching
against a certified thermometer is desired.
at any time to abort calibration and return to measurement
1) Connect the temperature probe to the meter and place into a solution
with a known accurate temperature such as a constant temperature
bath. Note: To adjust the manual temperature compensation (MTC)
value, do not connect the temperature probe.
2) Press
3) Press
measured temperature while the secondary display shows the factory
default temperature.
4) Adjust the temperature using
to cancel. The meter allows an adjustable maximum value of ± 5ºC
(or ± 9ºF) from the factory default temperature.
as needed to select pH or mV/R.mV.
followed by . The primary display shows the
or . Press to accept or
Loading...
+ 16 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.