Oakton PCTSTestr 50 Instruction Manual

Operating Instructions
cal esc
cal esc
cal esc
PCTSTestr
Applications
• Agriculture
• Aquaculture
• Aquariums and
fish farms
• Boiler blow-
down
• Car washes
Getting Started
The PCTSTestr 50 Pocket Tester has been factory calibrated and usually works well out of the box. However, after extended periods of non-use, it is best to remove the sensor cap and soak the sensor in warm tap water for 10 minutes or so. Prior to taking the measurements, periodic calibration with certified standards is recommended for best accuracy.
Measurement Parameter Setting
1. Press ON/OFF ( ) to power on the tester.
2. Press MENU/ to select Measure. The display shows pH, Cond, TDS and Salinity.
3. Scroll down by pressing MENU/ Cond, TDS and Salinity. Press HOLD/
4. The display shows the selected parameter with a
menu
Key Info
Scroll
Accept
EXIT
pH Buffer Set Selection
PCTSTestr 50 Pocket Tester features USA (pH 4.01, pH 7.00 and pH 10.01) or NIST standards. Select either one to suit your requirements.
1. Press MENU/ to select Settings. The display shows Buffer, TDS Factor and Backlight.
50 Pocket Tester
• Drinking water
• Ecology
• Electroplating
rinse tanks
• Food sectors
• Hydroponics
• Printing industry
• Swimming pools
• Verication of
reverse osmosis system operation
• Water and waste­water treatment
• Labs
v
v
to toggle between pH,
to select pH.
MENU
HOLD
CAL
Measure Settings
Temp Set Reset
hold
pH Cond TDS Salinity
(pH 4.01, pH 6.86, and pH 9.18)
v
1 2 3
.
hold
pH
2. Press HOLD/ to select Buffer. Display shows USA and NIST.
3. Press HOLD/ MENU/
4. The display shows the selected buffer standard with a
Key Info
menu
Scroll Accept EXIT
MENU HOLD CAL
to select USA or scroll down by pressing
v
to toggle between the two buffer standards.
hold
Measure Settings
Temp Set Reset
Buffer TDS Factor Backlight
hold
USA
NIST
.
hold
USA
TDS Factor Setting
1. Press MENU/ pressing MENU/
2. Press HOLD/ Buffer, TDS Factor and Backlight.
3. Scroll down by pressing MENU/ Buffer, TDS Factor and Backlight. Press HOLD/ the TDS Factor.
4. Press HOLD/ MENU/
5. Press HOLD/ display shows the selected value (TDS factor) with a
menu
Key Info
Scroll Accept EXIT
v
to enter setup window. Scroll down by
v
to select Settings.
to select Settings. The display shows
v
to toggle between the
to select the default TDS factory setting or
v
to adjust the setting.
to confirm the selection of the setting. The
MENU HOLD CAL
hold
Measure Settings
Temp Set Reset
Buffer TDS Factor
Backlight
hold
TDS Factor
Adjust
0.71
to select
.
hold
0.66
Backlight Settings
1. Press MENU/ pressing MENU/
2. Press HOLD/ Buffer, TDS Factor and Backlight.
3. Scroll down by pressing MENU/ Buffer, TDS Factor and Backlight. Press HOLD/ Backlight.
4. The display shows ON and OFF. Scroll down by pressing MENU/ increases readability in low-light conditions.
5. Press HOLD/ display shows the selected backlight option with a
v
to enter setup window. Scroll down by
v
to select Settings.
to select Settings. The display shows
v
to toggle between
to select
v
to toggle between ON and OFF. Backlight ON
to select the desired backlight option. The
.
menu
Key Info
Scroll Accept EXIT
MENU HOLD CAL
hold
Measure Settings
Temp Set Reset
Buffer TDS Factor
Backlight
hold
Backlight ON
OFF
hold
OFF
Temperature Settings
1. Press MENU/ pressing MENU/ to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
2. Press HOLD/ MENU/
3. Press HOLD/ shows the selected temperature setting with a
menu
Key Info
Scroll Accept EXIT
v
to enter setup window. Scroll down by
v
to select Temp Set. Press HOLD/
to select Set °C/°F. Scroll down by pressing
v
to toggle between °C and °F.
to select temperature unit. The display
MENU HOLD CAL
Measure Settings Temp Set Reset
hold
Set °C/°F Temp Cal
Temp Coeff
.
hold
°C
°F
hold
°C
Temperature Calibration
1. Press MENU/ pressing MENU/
2. Press HOLD/ Set °C/°F, Temp Cal and Temp Coeff.
3. Scroll down by pressing MENU/ Set °C/°F, Temp Cal and Temp Coeff. Press HOLD/ to select Temp Cal.
4. The lower display shows the current measured temperature reading based on the last set offset and the upper display shows the current measured temperature reading based on factory default calibration.
5. Dip the tester into a solution of known temperature and allow time for the built-in temperature sensor to stabilize.
6. Press MENU/ the HOLD/ temperature value of the solution.
Note: To exit this program without confirming the calibration, press CAL/ESC.
Key Info
menu
Scroll Accept EXIT
v
to enter setup window. Scroll down by
v
to select Temp Set.
to select Temp Set. The display shows
v
to toggle between
v
to adjust the temperature value or press
to confirm the calibrated value as the new
MENU HOLD CAL
hold
Measure Settings Temp Set Reset
Set °C/°F Temp Cal Temp Coeff
hold
Temp Cal 23 4 °C 23 4 °C
hold
2.00
Temperature Coefficient
Key Info Scroll Accept EXIT
v
to enter setup window. Scroll down by
MENU/
v
to select Temp Set.
v
to toggle between
to select Temp Coeff or MENU/v to
to confirm the Temp Coeff value. The
MENU HOLD CAL
Measure Settings Temp Set Reset
hold
Set °C/°F Temp Cal Temp Coeff
hold
Adjust
Temp Coeff
2.00
.
hold
2.00
1. Pres
MENU/
pressing
2. Press HOLD/to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
3. Scroll down by pressing MENU/ Set °C/°F, Temp Cal and Temp Coeff.
4. Press HOLD/ adjust the Temp Coeff.
5. Press HOLD/ new value is automatically confirmed with a
menu
pH Calibration
Calibration should be done regularly, recommended once a week. Calibrate up to three points using either the USA or the NIST buffer set standards.
1. Press ON/OFF (
2. Dip electrode about 2 cm to 3 cm into the pH standard buffer solution.
3. Stir gently and press CAL/ESC to enter calibration mode. The CAL indicator will be displayed. The upper display will show the measured reading based on the last calibration while the lower display will indicate the pH standard buffer solution.
Note: To abort calibration, press CAL/ESC to escape.
4. Allow about 2 minutes for the tester reading to stabilize. The timer icon blinks during this time. Once the reading is stabilized, the timer stops blinking. Automatic confirmation happens when the buffer is found and the display returned to measurement window with reading calibrated to pH standard buffer solution.
5. Repeat with other buffers if necessary. Rinse electrode before dipping into next buffer.
Note: The calibration mode allows you to perform up to three calibration points. Calibration is automatically con­firmed with the buffer identification. No user interaction is required after starting the calibration by pressing CAL/ESC.
cal
CAL
esc
Place the tester in the buffer to start or press to escape.
) to power on the tester if needed.
CAL pH
pH
7.45
7.01
pH
7.01
25.5°C
Calibration for Conductivity, TDS, or Salinity
For best results, periodic calibration with an accurate standard is recommended prior to measurement. Use the calibration standard value that is close to your intended sample value. The tester will retain one calibration value in each mode (conductivity, TDS, salinity) when the instrument is powered off. The conductivity value can be calibrated automatically or manually, while the TDS & salinity values require manual calibration. The tester will begin in the measurement mode that was used when it was powered off. See “Measurement Parameter Setting” to change the desired parameter.
Automatic Calibration for Conductivity
1. Remove the cap and press ON/OFF ( ) to power on.
2. Dip the sensor in at least 30 mm of calibration standard.
3. Stir gently and press CAL/ESC to begin the calibration.
4. The display will show CAL followed by the default value. CAL is indicated on the display during calibration mode.
5. If the reading is within the calibration range of the automati­cally recognized standards; 80 (84 μS/cm), 1410 (1413 μS/cm), or 12.90 (12.88 mS/cm), the
icon is displayed when the
automatic calibration standard value has been detected.
6. Press HOLD/
to accept the auto conductivity standard
and finish the calibration.
7. Display returns to Measurement window.
cal
CAL
esc
Place the tester in the calibration STD to start or press to escape.
CAL COND
µS/cm
1413
844
µS/cm
844
23.5°C
Manual Calibration
When the conductivity reading is outside calibration range of the automatic conductivity standards or when TDS or salinity is used, the tester will require manual adjustment.
1. Repeat steps 1 to 4 from “Automatic Calibration for Conductivity”.
2. Press MENU/ desired reading.
Note:
The adjustment will decrease only, however the adjustment will eventually cycle to the highest available value after decreasing by 40% of the initial value.
3. Press HOLD/ to accept and finish the calibration when
the desired value is selected.
Note: To abort calibration, press CAL/ESC to escape.
v
to manually adjust the value to the
4. Once the calibration is finished and user has accepted the changes, measurement window will now show the calibrated reading.
Note: The auto conductivity standards are 84 μS/cm, 1413 μS/cm & 12.88 mS/cm.
cal
CAL
esc
Place the tester in the calibration STD to start or press to escape.
CAL CAL
µS/cm
844
1413
COND
µS/cm
844 820
µS/cm
1413
23.5°C
Measurement
1. Press ON/OFF ( ) to power on the tester if needed.
2. Dip the electrode in about 2 cm to 3 cm into the test solution. Stir and let the reading stabilize. The timer icon will blink during this time. Once the reading is stabilized, the timer stops blinking and
will appear to indicate the
stability of the reading.
CAUTION: Testing dry samples is not accurate
and can lead to sensor damage or breakage. Soils must be wet and free of particulates that may scratch
pH
pH
4.01
25.5°C
the glass sensor. Excessive force into dry samples can cause glass breakage.
3. Note the value or press HOLD/ To release the reading, press HOLD/
4. Press ON/OFF (
) for 3 seconds to turn off tester. If you
to freeze the reading.
again.
do not press a button for 8.5 minutes, the tester will automatically shut off to conserve batteries.
User Reset
Reset to the user’s default settings by using the User Reset function. Buffer selection and user temperature calibration are not affected by the user reset function.
MENU HOLD CAL
v
to enter setup window. Scroll down by
v
to select Reset. Press HOLD/ to select
to select User Reset.
v
to toggle between No and Yes.
to confirm either No or Yes. The display
.
hold
Measure Settings Temp Set Reset
User Reset
Factory Reset
hold
No
Yes
hold
1. Press MENU/ pressing MENU/ Reset. The display shows User Reset and Factory Reset.
2. Press HOLD/
3. The display automatically shows No and Yes. Scroll down by pressing MENU/
4. Press HOLD/ shows the User Reset option with a
Key Info
menu
Scroll Accept EXIT
4 5 6
1065O100_MAN_35634-35.indd 1 9/5/2017 3:09:46 PM
Factory Reset
Reset to the Factory Default Settings by using the Factory Reset function.
1. Press MENU/ pressing MENU/
v
to enter setup window. Scroll down by
v
to select Reset. Press HOLD/ to select
Reset. The display shows User Reset and Factory Reset.
2. Scroll down by pressing MENU/ resets. Press HOLD/
to select Factory Reset.
v
to toggle between the
3. The display automatically shows No and Yes. Scroll down by pressing MENU/
4. Press HOLD/ shows the Factory Reset option with a
menu
Key Info
Scroll Accept EXIT
MENU HOLD CAL
v
to toggle between No and Yes.
to confirm either No or Yes. The display
.
Measure Settings Temp Set Reset
hold
Factory Reset
User Reset
hold
No
Yes
hold
HOLD Function
This feature lets you freeze the display for a delayed observation.
1. Press HOLD/button to freeze the measurement.
COND
µs/cm
820
23.5°C
COND
hold hold
µs/cm
820
Press Hold to release
23.5°C
2. Press HOLD/again to release the measurement.
COND
µs/cm
820
Press Hold to release
23.5°C
COND
µs/cm
820
23.5°C
Sensor Replacement
You can replace the sensor module at a fraction of the cost
of a new tester. When the tester fails to calibrate or gives
fluctuating readings in calibration standards, you need to change the electrode.
1. With dry hands, grip the ring with sensor facing you. Twist the ring counterclockwise. Save the ring for later use.
2. Pull the old sensor module away from the tester.
3. Align the four tabs on the new module so that they match the four slots on the tester.
4. Gently push the module onto the slots to sit it in position. Push the smaller O-ring fully onto the new sensor module. Push the other O-ring over the module and thread it into place by firmly twisting clockwise.
Note: It is necessary that you recalibrate your tester prior to measurement after a sensor replacement.
Replacing the Batteries
The PCTSTestr 50 Pocket Tester uses four AAA 1.5 V batteries.
1. To remove the battery cover, see Figure 1. Clear the front catch and then the back catch, before sliding the cover off.
2. To remove the battery plate, push the center tab towards the front
Fig. 1: Removing battery cover
Fig. 2: Push to unlock
of the tester as shown in Figure 2. Once unlocked, remove the plate to access the batteries.
3. Invert the tester upside down to remove the batteries.
Fig. 3: Align tabs
Fig. 4: Push down to lock
Each side uses two AAA batteries. Orient each battery with positive terminal facing downward.
4. To lock the battery place, align the small
Fig. 5: Battery plate unlocked Fig. 6: Battery plate locked
tabs (Figure 3) into the guide ribs on the hous­ing and then press down. See Figure 4.
Warranty
This instrument is supplied with a warranty against manufacturing defects for a period of one year from the date of purchase.
Sensor Maintenance
1. Always keep the sensor electrodes clean. Rinse the electrodes with de-ionized water and wipe them dry with clean cloth before storing with its protective cap. For cup type electrodes, remove the white plastic cup and insert to thoroughly clean viscous solutions. Note: Never scratch electrodes with a hard substance.
2. For better performance, soak the electrode in alcohol for 10 to 15 minutes and rinse with de-ionized water before starting any measurement process. This is to remove dirt and oil stains on the electrode, which may affect the accuracy of the measurements.
7 8 9
Self-Diagnostic Messages
Batteries are weak and need replacement soon. Appears when calibration is attempted but the reading
stable error
is not yet stable. Wait for the reading to stabilize or manually confirm the calibration by pressing enter.
buffer error The buffer is outside of the calibration range. slope error
The 2nd and 3rd calibration point is not within 80% to
120% slope range. over range The reading is above the measuring range of tester. under range The reading is below the measuring range of tester.
Specifications
Specifications PCTSTestr 50 pH
pH range –1.00 to 15.00 pH Resolution 0.01 pH Relative accuracy ±0.01 pH Calibration points Up to 3 points Buffer set standard
selection Calibration window ±1.00 pH Calibration type Point to point
Conductivity
Conductivity range Resolution 0.1 μS, 1 μS, 0.01 mS
Relative accuracy ±1% full scale Normalization temperature 25.0ºC (77ºF) Temperature co-efficient 0.0% to 10.0% Calibration points Up to 3 points
TDS
TDS range Resolution 0.1 ppm, 1 ppm, 0.01 ppt
Relative accuracy ±1% full scale Calibration points Up to 3 points TDS factor 0.40 to 1.00 (selectable)
Salinity
Salinity range 0.00 to 10.00 ppt Resolution 0.10 ppt Relative accuracy ±1% full scale Calibration points 1
USA: 4.01/7.00/10.01 NIST: 4.01/6.86/9.18
0.0 to 200.0 μS, 200 to 2000 μS,
2.00 to 20.00 mS
0.0 to 100.0 ppm, 100 to 1000 ppm,
0.10 to 10.00 ppt (TDS factor 0.5)
Specifications (cont.) PCTSTestr 50 Temperature
Temperature range 0 to 60°C (32.0 to 140.0°F) Temperature resolution 0.1°C / 0.1°F
Temperature accuracy Temperature
compensation
From 0 to 50ºC (±0.5°C / ±0.9°F + 1 LSD); from 50 to 60°C (±1.0ºC / ±1.8°F + 1 LSD)
Yes (Automatic Temperature Compensation)
General
Display Graphics, dot matrix 80 x 100 pixel Backlight
Yes, selectable
(30 sec from last key press) Auto off 8.5 minutes (from last key press) Reset User / Factory Power requirement Four AAA 1.5 V batteries Battery life >150 hours Water proofing IP67 Regulatory
certifications
CE, FCC
Environmental Operating Conditions
Ambient operating temperature
5 to 45°C / 41 to 113°F Relative humidity 5% to 85% noncondensing
Storage temperature –20 to 60°C / –4 to 140°F Storage humidity 5% to 85% noncondensing
Return of Items
Authorization must be obtained from your distributor
before returning items for any reason. When applying for
authorization, please include information regarding the reason the item(s) are to be returned.
We reserve the right to make improvements in design, construction and appearance of products without notice. Prices are subject to change without notice.
Accessories
Ordering Code
35634-35 35634-37 Replacement sensor module for PCTSTestr 50
35634-09 Replacement sensor cap 09376-00 17101-45 NIST-traceable calibration with data for pocket testers
Product Description
PCTSTestr 50 pocket tester with case, lanyard, and batteries
Replacement alkaline batteries; AAA, 1.5 V. Pack of 12
www.4oakton.com
1065O100_MAN_35634-35 August, 2017
10 11
1065O100_MAN_35634-35.indd 2 9/5/2017 3:09:48 PM
Loading...