Operating Manual
EcoTestr™ TDS1 Pocket Tester
Keypad Functions
Short press = <2 seconds
Long press = >2 seconds
1. Short press to turn on; long press to
turn off.
MEAS
CAL
<
2. When turned off, long press to enter
setup mode.
3. In mode setting, short press to change
parameter.
1. When turned on, long press to enter
calibration mode.
2. In calibration mode, short press to
confirm calibration.
3. In mode setting, short press to confirm
parameter selection.
Calibration
1. Rinse the probe in distilled water and gently dry it
with a clean tissue. Short press
the tester.
2. Long press CAL/ to enter calibration mode;
short press
3. Dip the probe into 1413 µS/cm (1003 ppm)
calibration solution. Stir gently, leave it to
stand. Wait for the measurement stability
icon (
) to appear and stay on the display
(see Fig 1); then short press CAL/ to
complete the 1st calibration. Tester returns
to measurement mode, and calibration
icon “M” appears on bottom left side
of display.
4. Rinse probe in distilled water and dry it. Repeat steps
#2 and #3 to complete 2nd calibration in 12.88 mS/cm
(9.14 ppt) calibration solution. Tester returns to measurement mode, and calibration icons “M” and “H”
/MEAS to exit.
/MEAS to turn on
Cond
1433
Figure 1
1413
will appear on bottom left side of display.
Measurement
1. Short press /MEAS to turn on the tester. Rinse probe
in distilled water and dry it gently with a clean tissue to
remove excess water.
2. Stir the probe in the sample solution gently, leave it to
stand. Wait for the stability icon (
take a reading.
3. Rinse off the tester thoroughly in distilled water after
each test.
Notes:
• If this is the rst-time use or the tester hasn’t been used
for a long time, we recommend soaking the probe in
12.88 mS/cm solution for 15 to 30 minutes to restore its
sensitivity and speed up the response time.
• The tester adopts 1413 µS/cm and 12.88 mS/cm standard
calibration solutions. User can use 1 and 2 point calibra-
tions as needed. For most circumstances, calibrating in
1413 µS/cm to complete the 1st point calibration will
meet testing requirements.
• The tester has already been calibrated after manufac-
µ
S
ture. User can use the tester immediately or can test it in
the calibration solutions to test its accuracy. If error is
large, calibrate the tester before using.
• We recommend replacing the calibration solution after
5 to 10 calibrations to maintain accuracy.
) to remain on, then
Setting the Parameters
When tester is off, long press /MEAS to enter setup mode.
Short press
press CAL/ and parameter will flash, then short press
/MEAS to choose desired parameter. Short press
CAL/ to confirm parameter selection. Long press
to return to measurement mode.
Symbol Menu setting Selection
P1 Set TDS factor 0.40 to 1.00 0.71
P2
P3
Notes:
• Users can adjust TDS factor (P1) by experimental data
or experience. The following chart lists some commonly
used TDS factors for reference:
Conductivity of the solution TDS factor (P1)
• Select “Yes” for P3 to restore the calibration to the
theoretical values and parameter setting to original
values. When tester’s calibration or measurement performs abnormally, this function can be adopted so the
tester returns to factory default setting and then users
can conduct calibration or take measurements again.
/MEAS to switch from P1 to P2 to P3. Short
Select temperature
unit
Restore to factory
default
0 to 100 µS/cm 0.60
100 to 1000 µS/cm 0.71
1 to 10 mS/cm 0.81
10 to 100 mS/cm 0.94
°F – °C °C
No – Yes No
/MEAS
Factory
default
1065O101_MAN_TDS1_35634-08.indd 1 03/22/2018 3:59:12 PM
Self-Diagnostic Messages
Symbol
Er 1
Er 2
Self-diagnosis
information
Wrong calibration
solution, or measured
value is not within the
range of the tester.
CAL/ is pressed
before measurement
is stable ( appears).
How to fix
• Check if calibration
solution is correct.
• Check if probe is
damaged.
Wait for the
measurement
stability icon ()
to appear and stay,
then press CAL/ .
Specifications
Range
Resolution
TDS
Accuracy ±1% full-scale
Calibration
points
Automatic
temperature
compensation
Range 32 to 122°F (0 to 50°C)
Temperature
Resolution 0.1°F/°C
Accuracy ±0.9°F (0.5°C)
Power: four AAA batteries (included); >400 hours of
continuous operation
Low-voltage warning:
battery status icon flashes
Auto power-off: tester automatically turns off after 8 minutes
of nonuse
IP rating: IP67 (waterproof), oats on water when sensor cap
is on
Dimensions (L x W x H): 7" x 1.5" x 1.25" (17.8 x 4 x 3.1 cm)
Weight: 3.8 oz (107 g)
0 to 100.0 ppm,
0 to 1000 ppm,
0 to 10.00 ppt
0.1 ppm, 1 ppm,
0.01 ppt
1 or 2 points, auto
standard recognition
32 to 122°F
(0 to 50°C)
Ordering Information
Model Product description Catalog number
TDS1
EcoTestr pocket total
dissolved solids tester
35634-08
Battery Installation
The tester uses four AAA batteries. Please install batteries
according to the following steps. Note the correct direction
of battery installation: the positive side (+) of every single
battery must face up. Incorrect installation of batteries will
cause damage to the tester and create a potential hazard.
4 x AAA Batteries
2
1
3
+
+
_
_
+
+
_
_
1. Pull the battery cap up.
2. Slide the battery cap along the direction of arrow.
3. Open the battery cap.
4. Insert the batteries (ALL POSITIVE SIDES FACING UP).
5. Close the battery cap.
6. Slide and lock the battery cap along the direction of
arrow.
7. Fit the tester’s cap while making sure to push all the
way down. The tester’s waterproof design may be
compromised if the cap is not fitted correctly.
5
7
4
6
Warranty
We warrant this instrument to be free from defects in
material and workmanship and agrees to repair or replace
free of charge, at option of Oakton Instruments, any
malfunctioned or damaged product attributable to
responsibility of Oakton Instruments, for a period of
two years from the delivery (a six-month limited warranty
applies to probes). This warranty does not apply to defects
resulting from actions such as misuse (violation of the
instructions in this manual or operations in the manner
not specified in this manual), improper maintenance, and
unauthorized repairs. Warranty period is the time limit to
provide free service for the products purchased by
customers, not the service life of the tester or probe.
Oakton Instruments reserves the right to update the
information in this manual without giving notice in
advance.
www.4oakton.com
1065O101_MAN_TDS1_35634-08 March 2018
1065O101_MAN_TDS1_35634-08.indd 2 03/22/2018 3:59:13 PM