Nsk TI95LA, TI85LA, TI95EX, TI95L, TI85L User Manual

...
0 (0)
Nsk TI95LA, TI85LA, TI95EX, TI95L, TI85L User Manual

1 : 5 Speed Increasing Contra Angle Handpiece FG bur

Ti85L / Ti85LA / Ti85

Ti95L / Ti95LA / Ti95 / Ti95EX

OPERATION MANUAL

: AUTOCLAVABLE

OM-C0179E Rev.B

The EU directive 93/42/EEC was applied in the design and production of this medical device.

Please read this Operation Manual carefully and file for future reference.

Caution

When operating the handpiece always consider the safety of the patient.The handpiece is designed only for dental clinical use.

Check the vibration, noise and overheating outside the patient's oral cavity before use. If any abnormalities are found, stop using and contact your dealer.

If any abnormalities are found in use, stop using immediately and contact your dealer.

Do not allow any impact on the shank. Do not drop the handpiece.

Do not attempt to disassemble the handpiece nor tamper with the mechanism.Do not connect or disconnect the handpiece until the drive motor has

completely stopped.

Do not exceed a speed of 200,000min-1(rpm) at the head.Do not use eccentric, worn or damaged burs.

Use short shank bur for miniature head speed increasing handpiece. (Ti85L,

Ti85LA,Ti85) Using long burs could result in bur spin-out while using, or premature bearing failure.

Do not use a long surgical bur on speed increasing handpiece, which runs to 200,000min-1(rpm). Such long bur as 25mm in length shall not be used. Its shank may warp under load at a high speed and spin out of the handpiece when the load is released.

Before use always confirm that the bur is correctly seated and locked into place.

Depressing the push-button while handpiece bur is in rotation will result in

OVERHEATING of the handpiece head. Special caution must be exercised during use to keep cheek tissue AWAY from the push-button of the handpiece. Contact with cheek tissue may cause the push-button to depress and injury to the patient may occur.

Always operate with spray. Without cooling spray, the handpiece may build up heat.

Supply PANA SPRAY after each use.

Use only the genuine NSK-manufactured oil. (PANA SPRAY)

1. Specifications

Model

Ti85L/Ti85LA

Ti85

Ti95L/Ti95LA

 

Ti95

 

Ti95EX

Gear Ratio

1 : 5 Increasing

 

1 : 5 Increasing

 

Allowable Max. Speed

40,000min-1(rpm)

40,000min-1(rpm)

 

Coolant Delivery

Internal

Internal

Internal

 

Internal

 

External

Spray Type

Single Spray

Quadruple Spray

 

Chuck Type

Ultra Push

 

 

Ultra Push

 

Burs (#)

ø1.59-1.60mm Short Shank Bur

ø1.59-1.60mm Standard Bur

Optics

Glass Rod optic

 

Glass Rod optic

 

 

 

 

#Only use burs capable of 200,000min-1(rpm).

2.Connecting / Disconnecting the motor and the handpiece

To connect the motor and the handpiece, align the positioning marks on the motor and the handpiece, and insert straight. This procedure will also align the light receptacle and allow you to use illumination if your handpiece and motor are equipped for illumination. (Ti85L, Ti85LA, Ti95L, Ti95LA) To disconnect, hold the front part of the motor and the back part of the handpiece, and pull straight out.

Caution

Do not connect or disconnect the handpiece until the motor has completely stopped.

3. Insertion and Removal of the bur

a Bur Insertion

Insert the bur into the chuck until it stops (4-6mm depth).

Depress the push-button and insert further into the chuck until the bur stops. (Fig.1) After the bur is inserted, push and pull the bur to make sure that it is held firmly in place. (Fig. 2)

This action will further increase the bur retention force.

Fig. 1

sBur Removal

Depress the push-button and Pull out the bur. (Fig. 3)

Note : The push-button chuck has a double action. The button stops when it touches the rotor Depress further and pull out the bur.

Fig. 2

Fig. 3