11.3 Replacing the rechargeable batteries .................................................................................105
IV
K5133001IECIVZ.fm 03.05.06
All rights reserved. Copyright reserved.
Contents
Schematics and Diagrams
1Schematics and diagrams 109
Annex
Technical Documentation acc. to EMV standard IEC/EN 60601-1-2: 2001
Test List
Spare Parts Catalogue
K5133001IECIVZ.fm 03.05.06
All rights reserved. Copyright reserved.
V
Contents
VI
K5133001IECIVZ.fm 03.05.06
All rights reserved. Copyright reserved.
General
1
2
ZeusGeneral
1Symbols and Defini-
tions
WARNING
A WARNING statement provides important information about a potentially hazardous situation which, if not avoided, could result in death
or serious injury.
CAUTION
A CAUTION statement provides important information about a potentially
hazardous situation which, if not avoided, may result in minor or moderate
injury to the user or patient or in damage to the equipment or other property.
NOTE
A NOTE provides additional information intended to avoid inconvenience
during operation.
Definitions according to German standard DIN 31051:
Inspection= examination of actual condition
Maintenance= measures to maintain specified condition
Repair= measures to restore specified condition
Servicing= inspection, maintenance, and repair
2Notes
This Technical Documentation conforms to the IEC 60601-1 standard.
Read each step in every procedure thoroughly before beginning any test.
Always use the proper tools and specified test equipment. If you deviate from
the instructions and/or recommendations in this Technical Documentation,
the equipment may operate improperly or unsafely, or the equipment could
be damaged.
It is our recommendation to use only Dräger parts and supplies.
The maintenance procedures described in this Technical Documentation may
be performed by qualified service personnel only. These maintenance procedures do not replace inspections and servicing by the manufacturer.
The information in this Technical Documentation is confidential and may not
be disclosed to third parties without the prior written consent of the manufacturer.
This Technical Documentation is for the purpose of information only. Product
descriptions found in this Technical Documentation are in no way a substitute
for reading and studying the Instructions for Use/Operating Manual enclosed
with the product at the time of delivery.
All rights reserved. Copyright reserved.
Version 3.0_ Released_Printed on_03.05.06_K5133001_General_Technical_Documentation.fm
5133.001
3
GeneralZeus
Know-how contained in this Technical Documentation is subject to ongoing
change through research and development and Dräger Medical reserves the
right to make changes to this Technical Documentation without notice.
NOTE
Unless otherwise stated, reference is made to laws, regulations or standards (as amended) applicable in the Federal Republic of Germany for
equipment used or serviced in Germany. Users or technicians in all other
countries must verify compliance with local laws or applicable international
standards.
Version 3.0_ Released_Printed on_03.05.06_K5133001_General_Technical_Documentation.fm
All rights reserved. Copyright reserved.
4
5133.001
ZeusFunction Description
3Abbreviations and
Definitions
A-BoxAnaesthetic Gas Box
APLAdjustable Pressure Limit
CANController Area Network
CPCICompact Peripheral Component Interconnect
DAGMARDigital Advanced Gas Mixer for Anesthesia Require-
ments
DIANADIgita ANaesthestic controller in A-Box
DIVADigital Injection of Volatile Agent
DUMAFlowmeter for volatile anaesthetic
DVIDigital Video Interface
EEPROMElectrically Eraseable Programmable Read Only
Memory
Flash-ROMFlash Read Only Memory
GMZGas Measuring module Zeus
HERMESDisplay and operating unit of Zeus
ILCAInfrared Low Cost Analyzer
IRIAInfrared Rapidly Identifying Analyzer
MIBManagement Information Base
MIRMid Infrared Range
MISOMaster Input Slave Output
MOSIMaster Output Slave Input
NIRNear Infrared Range
PLDProgrammable Logic Device
PWMPulse Width Modulation
RAMRandom Access Memory
SDRAMSynchronous Dynamic RAM
SNMPSimple Network Management Protocol
SPISerial Peripheral Interface
SRAMStatic RAM
TFTThin Film Transistor
TIVATotal IntraVenous Anesthesia
UARTUniversal Asynchronous Receiver/Transmitter
USBUniversal Serial Bus
UPSUninterruptible Power Supply
ZeusName of anaesthetic workstation
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_03.05.06_F5133001_abbreviations.fm
5133.001
5
Function DescriptionZeus
Version 1.0_ Released_Printed on_03.05.06_F5133001_abbreviations.fm
All rights reserved. Copyright reserved.
6
5133.001
Function Description
7
8
ZeusGeneral
1General information
about Zeus
Figure 1View of the Zeus anesthesia workstation
1.1Intended use (summary from the
Instructions for Use
manual)
1.2Product classifica-
Zeus is an integrated anesthesia workstation for inhalation and intravenous
anesthesia. Zeus is used
–in operating rooms as well as in induction and recovery rooms
–in adults, children and neonates
–with anesthetic agents Isoflurane, Sevoflurane and Desflurane
–with gas mixtures O2/AIR, O2/N2O
–in partial rebreathing to complete rebreathing mode
–for operation with external fresh-gas outlet - non-rebreathing systems
–for inhalation anesthesia, balanced and intravenous anesthesia
Class II b according to the Directive 93/42 EEC, Annex IX.
tion
1.3Protection classesClass I according to EN 60601-1.
All rights reserved. Copyright reserved.
Version 2.0_ Released_Printed on_03.05.06_F5133001_ZEUS_General_about.fm
5133.001
9
GeneralZeus
1.4Brief description of
the system
1.4.1ConfigurationThe Zeus anesthesia workstation can be subdivided into the display and con-
The integrated anesthetic workstation Zeus includes the following components:
–Windows NT computer with system display unit for operation of therapy
control and monitoring
–electronic gas and anesthetic agent flow control with closed-loop control
system
–electronically controlled and driven blower with rebreathing system
–airway monitoring
–hemodynamics monitoring module
The following extensions are available as optional features:
–Dräger syringe pumps for intravenous anesthesia
trol unit HERMES and the anesthetic gas box (A-Box).
HERMES is a PC based system for display and control of the A-Box or other
connected front-end units, such as the parameter box for measurement of
hemodynamic values. In addition, HERMES provides all external interfaces
of the Zeus anesthesia workstation, e.g. printer and serial ports.
The anesthetic gas box (Figure 2) contains the actuators required for the system, the mixed-gas flow control, the anesthetic-agent flow-control, the breathing system, the lung ventilator and the monitoring system. The anesthetic gas
box also comprises slots and interfaces for future optional features.
1.4.2The most important
external characteristics
The anesthetic workstation Zeus has an excellent ergonomic design. This
includes, for example:
–A system display unit mounted on a hinged arm that allows a 180° rota-
tion. A second flat display (optional feature) can be arranged on top of the
system display unit.
–A housing shape that is suitable both for left-hand/right-hand operation.
This includes, besides the rotatable system display unit, a push-through
work top and a push-through drawer. The absorber can be viewed easily
from both sides.
–Concealed, left-hand and right-hand rails for mounting of accessories.
–O2 flush buttons that can be operated on left-hand side and right-hand
side.
–A central locking brake for trolley castors that can be operated from both
sides.
10
5133.001
Version 2.0_ Released_Printed on_03.05.06_F5133001_ZEUS_General_about.fm
All rights reserved. Copyright reserved.
ZeusGeneral
Figure 2System overview
All rights reserved. Copyright reserved.
Version 2.0_ Released_Printed on_03.05.06_F5133001_ZEUS_General_about.fm
5133.001
11
GeneralZeus
12
5133.001
Version 2.0_ Released_Printed on_03.05.06_F5133001_ZEUS_General_about.fm
All rights reserved. Copyright reserved.
ZeusFunction Description
2Anaesthetic Gas
Box
2.1General introduction
The anaesthetic gas box comprises the trolley and the housing of the Zeus.
Apart from the monitor, it includes all the components required to operate the
functions of the Zeus. They include:
–Power supply unit
–Hermes computer
–PCB box
–CS module (CS = Central Supply)
–DIVA (digital anaesthetic metering unit)
–GMZ (Zeus gas measuring module)
–Blower (ventilator)
–Breathing system
Version 1.0_ Released_Printed on_03.05.06_F5133001_anaesthesiagasbox_General_about.fm
All rights reserved. Copyright reserved.
5133.001
13
Function DescriptionZeus
14
5133.001
Version 1.0_ Released_Printed on_03.05.06_F5133001_anaesthesiagasbox_General_about.fm
All rights reserved. Copyright reserved.
ZeusFunction Description
3PCB BoxThe PCB box is a housing system for PCBs in the Zeus. It principally com-
prises the Transfer PCB, which performs the function of a motherboard. The
Transfer PCB accommodates additional PCBs: the standard Zeus PCBs as
well as PCBs for later options.
At present the PCB box holds the following PCBs:
–DIANA PCB
–VERONA PCB
–Vent Power PCB
–External Flow PCB
3.1DIANA PCB
3.1.1PurposeThe DIANA PCB handles the communication of all system processors with
the HERMES via an Ethernet link. By way of the Transfer PCB the relevant
signals of various modules and the 24 V supply are routed to the DIANA
PCB. The internal 3.3 and 5 V operating voltages are generated on the
DIANA PCB by means of DC/DC-converters.
3.1.2FunctionThe central module of the DIANA PCB is the processor MPC 850 (Figure 3).
Figure 3Block diagram of the DIANA PCB
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_04.05.06_F5133001_PCB_Box.fm
5133.001
15
Function DescriptionZeus
The operating program is located in a 4 or 8 MB Flash-ROM which is loaded
into a 4 or 8 MB SDRAM as appropriate during start-up. Key software components can be stored in a 256 kB SRAM which is battery-buffered to protect
against failure of the operating voltage. Key parameters can additionally be
saved to an EEPROM via an I2C-port.
In addition to the memory modules, a CAN-controller and a four-way-UART
are also connected to the system bus. The modules are selected by way of a
PLD. The CAN-controller handles most of the communication with the other
processors in the overall system with the aid of the installed SABUS-protocol.
The four-way UART handles the communication with the ILCA module and
with any subsequent add-ons. An Ethernet controller handles the communication with the HERMES system. The IRIA module and a serial port for Service are connected via an electrical isolator directly to the processor. The
power supply unit generates a Power Fail signal indicating failure of its primary voltage (240 V mains voltage). The 24 V is also monitored in order to
utilize time reserves prior to total failure of the supply voltage for regulation
shutdown of the processor. An additional integrated module monitors the
3.3 V supply (Brown-Out monitor).
3.2VERONA PCB
3.2.1PurposeThe VERONA PCB is required in the ZEUS to control the ventilator and to
interface to the CAN bus. For this, a microcontroller system based on the
PPC555 is used.
3.2.2FunctionThe circuit is operated with a 24 V supply (Figure 4).
16
5133.001
Version 1.0_ Released_Printed on_04.05.06_F5133001_PCB_Box.fm
All rights reserved. Copyright reserved.
ZeusFunction Description
Figure 4Block diagram of the VERONA PCB
The voltage supply is fed in via an EMC-filter and monitored by the controller.
The controller PPC555 has an internal RAM which can be optionally batterybuffered. The following operating voltages are generated on the PCB:
–3.3 V for microcontroller
–5 V for temperature sensors
–5 V for digital modules
–12 V for pressure sensors
The processor system has two serial ports (max 115 kbaud). Two CAN-interfaces actively support the Full CAN V2.0A and V2.0B-protocol at up to 1Mb/s.
The CAN-buses are each isolated from the rest of the system and from each
other by an isolating voltage of at least 500 V AC. Optocouplers are used to
isolate the signals. To activate the blower motor the board has a dedicated
interface. The power electronics for the blower are located on the Vent Power
PCB. For speed evaluation the three TTL-signals Hall-A, Hall-B and Hall-C
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_04.05.06_F5133001_PCB_Box.fm
5133.001
17
Function DescriptionZeus
are delivered by the motor electronics. The motor temperature is evaluated,
as are the other temperature sensors. 6 LEDs are provided for output of status signals. The following LEDs are interesting for service work:
– 12 V supply voltage = LED V405
– 5 V supply voltage = LED V402
If only LED V302 of the 4 LEDs between the SUB-D connectors is lit the hardware is in test mode.
3.3Vent Power PCB
3.3.1PurposeThe Vent Power PCB contains the power electronics to activate the blower
and the external valves. It provides the link between the valve control and the
pneumatic system.
3.3.2FunctionThe respective power stages can be disabled by an external signal (Figure
5).
Figure 5 Vent Power PCB block diagram
This signal is generated by an independent monitoring computer. In addition
to the power stages, the sender information from the TurboVent and the status signals of the various output stages are processed and passed on for
evaluation to the TurboVent control computer. The supply voltage of the TurboVent is 48 V DC and that of the valves is normally 24 V. Because of the different supply voltages and the substantial peak currents in the TurboVent
activation circuit these blocks are electrically isolated by optocoupler.
18
5133.001
Version 1.0_ Released_Printed on_04.05.06_F5133001_PCB_Box.fm
All rights reserved. Copyright reserved.
ZeusFunction Description
The supply voltage for the TurboVent is externally stabilised in the power
supply unit to 48 V and has a current limit of 3 A. An external 68000 uF
capacitor is included to deal with the high peak currents of the motor.
The function of the blower power stage is to provide phase-adjusted activation of the motor coil with the aid of the Hall sensors built into the motor and to
control the speed of the motor. The Hall signals are formatted by an inverter
and passed to the Hall decoder. This module decodes the Hall signals of the
motor and codes the activation signals of the MOSFET driver from them. The
power MOSFETs are activated by a MOSFET driver. The output stage comprises 3 half-bridges each with two identical n-channel MOSFETs. At a TurboVent electronics supply voltage below 38 V a protective circuit disables the
MOSFET activation. As the motor is run in a highly dynamic mode at high
speeds, the power stage is rated for a peak power output of approximately
480 W. The mean load in operation is approximately 100 W.
The fan delivers one pulse per fan wheel revolution. This pulse is filtered by
an RC element and routed to a retriggerable monoflop. If the pulses are not
received, the monoflop drops out and disables the power electronics for the
TurboVent. This ensures that the TurboVent is only activated when the fan is
running. This prevents harmful overheating or excessive oxygen concentration build-up at the TurboVent motor.
The Vent Power PCB also contains the activation electronics for the valves in
the breathing system and their interface to the TurboVent control computer.
Additionally, the Ventilator-OFF signal is evaluated by the DIANA PCB to cut
the power to the driver stages.
The A-cone valve is a single-turn bipolar valve. The turn must be activated in
bipolar mode by a voltage pulse. This valve driver is independent of the Ventilator-OFF signal.
The power electronics for the flow valves are located directly at the pneumatic interface underneath the valves, and not on this PCB. However, the
valves are shut off by the Ventilator-OFF signal by way of a HIGH-side FET
switch on this module. The switch cuts the supply voltage to the flow valves.
The valves, seal, fresh gas, switch-over and BAG are controlled by a low-side
MOSFET switch. This module has a monitoring output which is routed to the
TurboVent control computer. The monitoring covers short-circuit and open
load.
3.4External Flow PCB
3.4.1PurposeThe “Alveon” flow sensor is designed for adult patient-local directional flow
measurement and is built into the patient Y-piece. The External Flow PCB
digitises the sensor signals and supplies the data via SPI.
3.4.2External Flow PCB function
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_04.05.06_F5133001_PCB_Box.fm
5133.001
The External Flow PCB circuitry is divided into 3 areas (Figure 6):
–Supply voltage with electrical isolation from the overall system
–Analog measurement processing
–Digital measurement processing with data interface and electrical isola-
tion from VERONA PCB via SPI
19
Function DescriptionZeus
3.4.3Sensor measurement
principle
20
Figure 6 External Flow PCB block diagram
The “Alveon” flow sensor is based on the principle of a filament anemometer
(Figure 7). In conjunction with the Zeus anaesthetic workstation, the measur-
ing wire (2) is used for compensation of the breathing gas, the measuring
wire (3) for “flow measurement” and the measuring wire (5) with the shading
(4) for detecting the direction of flow.
5133.001
Version 1.0_ Released_Printed on_04.05.06_F5133001_PCB_Box.fm
All rights reserved. Copyright reserved.
ZeusFunction Description
Figure 7“Alveon” flow sensor layout
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_04.05.06_F5133001_PCB_Box.fm
5133.001
21
Function DescriptionZeus
22
5133.001
Version 1.0_ Released_Printed on_04.05.06_F5133001_PCB_Box.fm
All rights reserved. Copyright reserved.
Loading...
+ 149 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.