NORTH AMERICAN DRÄGER TEC 6 Desflurane NAD Variant User manual

RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
DrägerService
RETURN TO CD-ROM TABLE OF CONTENTS
®
Field
Service
Procedure
Part Number: SP00091
Date: 11 July 2001
Rev: H
KIT, TEC 6 DESFLURANE
VAPORIZER INSTALLATION, NAD VARIANT
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
KIT, TEC 6 DESFLURANE
VAPORIZER INSTALLATION, NAD VARIANT
INSTALLATION PROCEDURE
NOTE: The following procedure is performed after all host NARKOMED Anesthesia Systems
have been upgraded, if applicable, in accordance with TSB #103.
NOTE: The desflurane vaporizer shall be installed in the extreme right position (as viewed
from the front) on the machine.
NOTE: If the host machineisaNARKOMED4, its gas analyzer must be configured to include
desflurane. Refer to Service Procedure SP00117 for details. If the host machine is a NARKOMED 3, its gas analyzer must be configured to include
desflurane. Refer to Service Procedure SP00121 for details.
NOTE: Proceed to Step 5 if you are not removing an existing vaporizer.
1. Before removing an existing vaporizer or bypass block from the machine, perform a low pressure leak test at the fresh gas outlet to verify gas circuit integrity.
2. If there is an existing vaporizer in place, it must be drained and dried as follows:
CAUTION
The following steps must be performed in the sequence given.
2.3 Open the APL valve by turning it fully counter-clockwise.
2.4 Short circuit the inspiratory and expiratory valves with a 22 mm hose.
2.5 Ensure that the fresh gas hose is connected to the 15 mm connector on the machine.
2.6 Turn the System Power switch to ON.
2.7 Set all vaporizer handwheels to their Zero or OFF position.
WARNING
2.1 Activate the waste gas scavenger system.
2.2 Attach a breathing bag to the bag mount connector.
Do not inhale anesthetic vapors as this could result in personal injury.
1
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTS
2.8 Open the filler and drain plugs, and drain the vaporizer into a suitable container. Dispose of the residual agent in an approved manner.
2.9 Close the filler and drain plugs.
2.10 Turn the vaporizer handwheel to themaximumconcentrationsetting.
2.11 Set the oxygen flow to 10 l/min. for at least 20 minutes.
2.12 Turn the vaporizer handwheel to 0 (zero), turn the oxygen flow off and set the System Power switch to STANDBY.
3. While holding the vaporizer, remove the mounting screws and carefully separate the vaporizer from the machine.
4. Place the vaporizer in a suitable container for transport or storage.
5. Remove the desflurane vaporizer from its packing container and inspect the vaporizer for damage.
6. Verify that the dial (handwheel) is set to Standby.
7. Place an O-ring (supplied by the vaporizer manufacturer) in each recess of the mounting surface of the vaporizer.
WARNING
DO NOT use the Dräger O-rings when installing the desflurane vaporizer. The Dräger O-rings may cause occlusion of the gas passages when used with desflurane.
8. Secure the vaporizer to the machine with two 35 mm long screws (Ohmeda supplied by the vaporizer manufacturer). See Figure 1.
WARNING
If the Desflurane vaporizer is later removed from the machine and a Dräger vaporizer installed in its place, DO NOT use the Ohmeda screws or the Ohmeda O-rings to mount the Dräger vaporizer. The longer Ohmeda screws will bottom out and will not allow a gas tight seal between the Dräger vaporizer and the machine.
2
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTS
SP91001
VAPORIZER
MOUNTING SCREWS
(USE ONLY OHMEDA SCREWS -
SEE TEXT)
O-RINGS
(USE ONLY OHMEDA O-RINGS
SEE TEXT)
Figure 1: DESFLURANE VAPORIZER INSTALLATION
3
L
U
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTS
9. Test the long pivot arm of the vaporizer interlock assembly with a magnet. If the pivot arm is magnetic, skip to Step 14.
NOTE: Unused pivot arms may be
returned to:
North American Dräger Technical Service Department 3122 Commerce Drive Telford, PA 18969
Attn: Desflurane Coordinator
10. Remove the upper retaining clip securing the center pivot pin to the vaporizer interlock assembly. See Figure 2.
11. Slide the pin out from the assembly (do not misplace the two spacers) and remove the long pivot arm.
12. Transfer the two locking hex nuts and the two set screws from the existing pivot arm to the new pivot arm included in the Desflurane Installation Kit.
13. Assemble the new pivot arm onto the interlock assembly as follows: Slide the pivot pin up into the lower support bar, place a spacer on the pin, place the new pivot arm onto the pin, slide the pin up through the pivot arm, place another spacer on the pin and continuesliding the pin through the upper support bar. Secure the pin to the assembly with the retaining clip
14. Attachthedesflurane warning label (included in the Desflurane Installation Kit) to the right side of sloping front vapor box panel. (Choose another location on the sloping front panel if the right side is already occupied by another label.)
VAPORIZER
SP91006
INTERLOCK MECHANISM REAR VIEW
RETAINING CLIP
PPER
SUPPORT BAR
LONG PIVOT
ARM
OWER
SUPPORT BAR
SPACERS
CENTER PIVOT PIN
Figure 2: INTERLOCK MECHANISM - LONG PIVOT ARM INSTALLATION
4
Loading...
+ 12 hidden pages