NORTH AMERICAN DRÄGER Oxylog 2000 User manual

Dräger Medical, Inc.
Oxylog 2000 Technical Service Bulletin # 6
Re: Device mount for vehicles P/N 84 12 069 (Accessory for
Oxylog 2000)
Update: May 22, 2000 Reference Doc: IDM nos. 6 / 02.99 Reason: It has been reported that both retaining pins in the device mount for
vehicles P/N 84 12 069 may become loose. If the retaining pins loosen the Oxylog 2000 can fall down in worst case.
Solution: The two retaining pins (P/N 84 12 245) shall be secured with
Loctite 221. Mark secured pins with screw stop varnish 3000 ISO red.
Devices affected: All device mounts for vehicles P/N 84 12 069 shipped prior February
1999.
When: Next Service Call or Preventive Maintenance. Additional Info: N/A Cost: Covered by Customer, post warranty. Ordering Info: Loctite 221, screw locking agent, low P/N 79 01 966
Screw stop varnish 3000 ISO red P/N 79 01 616
5503.160 Oxylog 2000 TSB No. 6 May 2000 1/2
Dräger Medical, Inc.
Distribution: Dräger Service Personnel and Authorized Service Organizations for
CCS products.
If you have any questions, please contact Technical Support by phone at 1-800-543-5047 or by fax at 1-215-721-5789
Dräger Medical, Inc. Technical Product Manager
5503.160 Oxylog 2000 TSB No. 6 May 2000 2/2