NORTH AMERICAN DRÄGER Oxylog 1000 User manual

Page 1
Oxylog 1000
WARNING! For a full understanding of the
performance characteristics of this equipment, the user should carefully read this manual before use of the device.
MT-974-2000
Emergency ventilator Instructions for Use
Page 2
Contents
For Your Safety and that of Your Patients . . . . . . . . . . . . . . . . .3
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Oxylog 1000 Device Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
IPPV controlled ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Ventilation with PEEP (Special accessory). . . . . . . . . . . . . . . . . . 9
End-expiratory volume measurement (Special accessory) . . . . .10
End of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Care. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Dismantling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Disinfecting/Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Sterilising . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Contents
Preparation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Mounting the ventilation valve . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Installing the Oxylog 1000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Connecting the O2 supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Checking readiness for operation. . . . . . . . . . . . . . . . . . . . . . .18
Testing ventilation function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Testing the »Paw >« alarm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Testing the »Paw <« alarm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Checking the »Psupply« alarm . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
What's what . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Front view. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Side view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Abbreviations and symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Minute volume and O2 concentration as a function of airway pres-
sure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Block circuit diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Order List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
1
Page 3
Contents
This page is intentionally left blank
2
Page 4
For Your Safety and that of Your Patients
For Your Safety and that of Your Patients
Definitions
WARNING!
A WARNING statement provides important information about a potentially hazardous situation which, if
not avoided, could result in death or serious injury.
CAUTION!
A CAUTION statement provides important information about a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or in damage to the equipment or other property.
NOTE:
A NOTE provides additional information intended to avoid inconvenience during operation.
Strictly follow these Instructions for Use
WARNING!
Strictly follow these Instructions for Use. Any use of the product requires full understanding and strict
observation of all portions of these instructions. The device is only to be used for the purpose specified under "Intended Use" on page 4 and in conjunction with
appropriate patient monitoring (see page 6). Observe all WARNING and CAUTION statements throughout this manual and all statements on device labels.
Liability for proper function or damage
The liability for the proper function of the apparatus is irrevocably transferred to the owner or operator to the extent that the apparatus is improperly serviced or repaired by personnel not employed or authorized by DrägerService or if the apparatus is used in a manner not conforming to its intended use.
Dräger cannot be held responsible for damage caused by non­compliance with the recommendations given above. The warranty and liability provisions of the terms of sale and delivery of Dräger are likewise not modified by the recommendations given above.
Dräger Medical b.v., Best, the Netherlands
Maintenance
WARNING!
The device must be inspected and serviced regularly by trained service personnel.
Repair of the device may also only be carried out by trained service personnel.
Dräger recommends that a service contract be obtained with DrägerService and that all repairs also be carried out by them. Dräger recommends that only authentic Dräger repair parts be used for maintenance. Otherwise the correct functioning of the device may be compromised.
See chapter "Maintenance Intervals".
Accessories
Do not use accessory parts other than those in the order list.
3
Page 5
Precautions
Precautions
WARNING!
Standby manual ventilation system If the life-preserving function of the ventilator is no longer
guaranteed due to a fault, the patient must immediately be ventilated with an alternative independent ventilating device, e.g. with a self-filling manual breathing bag.
WARNING!
Do not use this apparatus in explosion hazard areas. Risk of explosion!
WARNING!
Do not use the equipment in conjunction with magnetic resonance imaging (MRI, NMR, NMI).
The apparatus may malfunction, causing danger to the patient.
CAUTION!
Ventilation monitoring During ventilation, the patient must be constantly
monitored by qualified medical personnel.
Intended Use
Oxylog 1000 – a time cycled, volume constant transport and emergency ventilator for patients requiring a minute volume ventilation of at least 3 liters per minute.
General description
Oxylog 1000 is a purely pneumatic-powered transport and emergency ventilator.
With display
— of inspiratory airway pressure P
with audible and visual alarms for:
— airway pressure P — airway pressure Paw high
2 supply pressure Psupply low
—O
aw low
aw,
with applications in:
— mobile use in EMS and primary care of emergency patients — patient transports and transfers by land, sea, or air. — intra-hospital transfers of ventilated patients — secondary transfers between hospitals — the emergency room.
4
Page 6
Operating concept
Operating concept
9 8 7 6 5
3
4 1 1 1 2
1 Three rotary control knobs are located in the middle of the
front panel for setting the upper alarm limit for airway pressure (»Pmax«), the ventilation rate (»Freq.«) and the minute volume (»MV«).
Uniform colour codes are used to identify different scale ranges of the »Freq.« and »MV« knobs to help rapid presetting: with these colour-coded scales, the initial parameters are adapted to the relevant patient group: infants (green) / children (blue) / adults (brown).
2 The main switch 0/I for switching the ventilator on and off is
in the bottom right-hand corner of the front panel.
3 With the »Air Mix/No Air Mix« switch, the user can choose
between approx. 60 Vol.% O
4 The pressure gauge shows the inspiratory airway pressure.
NOTE:
The color coded ranges and patient identifications are offered solely for the convenience of the user for selecting initial settings. It is ultimately the responsibility of the user to select the correct settings for each patient.
2 and 100 Vol.% O2.
The flag indicators for the alarms are located above the rotary control knobs for »Freq.«, »MV« and »Pmax«:
5 »Psupply« flag indicator.
Green if the supply pressure is sufficient, turns red if the supply pressure is insufficient.
6 Indicator for the upper alarm limit »Paw
Red if the airway pressure exceeds the maximum limit.
7 Indicator for lower alarm limit »Paw
alarm limit is not reached.
8 »
g«button to mute the audio alarm for up to 2 minutes.
9 Indicator »
g« turns yellow when the alarm tone is muted.
>« .
<« turns red if lower
MT-214-99A
5
Page 7
Operation
Operation
Oxylog 1000 Device Check
The device check must be carried out before each use. Any operation of the device requires thorough knowledge of the Instructions for Use.
Type
Serial No.
Check the following points before starting up the device:
O2 pressure supply connected
Cylinder pressure at least 100 bar or ventilator
connected to the central O
Ventilation valve and ventilation hose connected
2 supply
Testing correct operation
Fit the test lung to the ventilation valve.
Set the device when using a reusable hose:
»MV« approx. 10 L/min »Freq.« approx. 10 bpm »
Pmax« approx. 55 mbar Main switch I (ON) Switch »
Set the device when using a disposable hose to the settings
No Air Mix«
as described in the accompanying leaflet of the disposable hose.
Oxylog 1000 must ventilate the test lung After 5 ventilation strokes ventilation is
constant, no alarms should occur.
Squeeze the test lung, so that the airway pressure is approx.
60 mbar.
The »Paw >« indicator turns red, and the audible alarm is sounded.
Remove the test lung:
The »Paw <« indicator turns red, and the audible alarm is sounded.
6
Page 8
Operation
In the event of deviations, see "Troubleshooting", page 21.
Device check completed
Name
Date
Use a prepared, operable and disinfected device. Care, page 11.
Preparation, page 14. Checking device ready for operation, page 18.
IPPV controlled ventilation
For ventilation frequencies of 4 to 54 breaths per minute.
For rapid presetting, uniform colour-coded scale ranges can be used for the rotary control knobs for the ventilation rate Freq. and minute volume MV.
1 Presetting the »Freq.« and »MV« control knobs:
Patient group Freq.
1/min
Green range
28 to 54 3 to 5
MV
L/min
for infants
Blue range
20 to 28 5 to 9
for children
Brown range
4 to 20 9 to 20
for adults
NOTE:
The color coded ranges and patient identifications are offered solely for the convenience of the user for selecting initial settings. It is ultimately the responsibility of the user to select the correct settings for each patient.
D
Oxylog 1000
1
1
002
7
Page 9
Operation
NOTE:
The tidal volume will be equal to the minute volume setting divided by the rate setting:
Minute volume Ventilator Rate (1/min) (L/min) 5 10 25 35 54 (max
3 0.60 0.30 0.12 0.09 0.06
7 1.40 0.70 0.28 0.20 0.13
10 2.00 1.00 0.40 0.29 0.13
15 3.00 1.50 0.60 0.43 0.28
20 4.00 2.00 0.80 0.57 0.37
Tidal volume VT (L)
The I/E-ratio is fixed to approximately 1:1.5 and changes slightly depending on the setting. The I/E-ratio is within the range considered reasonable for most patients during emergency care and transport.
)
2 Set the desired O2 concentration with the switch:
Air Mix approx. 60 % O
2 by volume
or No Air Mix = 100 % O
WARNING!
2 by volume
In Air Mix mode, the applied tidal volume VT is reduced at high airway pressures due to the physical characteristics of the injector used for the mixing and the O
2 concentration
increases due to the smaller amount of air intake. (See also page 28 in the appendix).
WARNING!
In toxic surroundings: — The patient must be ventilated in No Air mix mode in order
to ensure that toxic constituents are not entrained into the breathing gas.
— The patient must immediately be transferred to a
breathable atmosphere in order to prevent inhalation of toxic air when spontaneous breathing resumes.
Set »Pmax«
When the patient is connected: 1 Check the »MV« setting and adjust according to the patient. 2 Check the airway pressure on the pressure gauge. 3 Set desired upper alarm limit »Pmax«. When the desired upper alarm limit set »Pmax« is reached, the
machine limits the airway pressure increase by blowing off part of the inspiratory flow. Inspiration is continued by the machine.
WARNING!
Watch the pressure gauge and take note of alarm conditions in order to recognise incorrect ventilation at an early stage and prevent danger to the patient.
D
D
2
003
Oxylog 1000
2
3
1
004
8
Page 10
Operation
For heart-lung resuscitation
For resuscitation of adults using the "two helper method": 4 Set the »Freq.« knob to the heart symbol H, 5 Set the »Pmax« knob to the heart symbol
H,
approx. 55 mbar.
NOTE:
Airway pressure limiting is now activated. When the peak pressure limit is reached, tidal volume may not be fully applied under certain conditions.
Ventilation with a mask
Connect mask to patient connection on breathing valve.
Position mask over the face to cover the bridge of the nose
and the chin, to ensure a tight fit.
D
Oxylog 1000
5
4
006
Ventilation with PEEP (Special accessory)
1 Set the PEEP valve to 0 mbar = turn the knob anti-clockwise
as far as it will go and fit it to the expiration connector of the ventilation valve. Set PEEP = turn knob. The end-expiratory pressure is increased by the set PEEP value.
NOTE:
The PEEP pressure is not displayed on the pressure gauge!
D
028028
Oxylog 1000
1
9
Page 11
Operation
Only for stationary use!
End-expiratory volume measurement (Special accessory)
To measure the end-expiratory tidal volume and end-expiratory minute volume.
Cannot be combined with the PEEP valve.
NOTE:
Strictly follow the Instructions for Use of the Volumeter 3000.
2 Clamp the Volumeter 3000 holder to the wall rail. 3 Screw the Volumeter 3000 to the holder. 4 Screw the elbow connector to the Volumeter 3000. 5 Fit adapter to expiration connector of the ventilation valve. 6 Connect the ventilation valve and Volumeter 3000 with
1.5 m long ventilation hose.
D
Oxylog 1000
5
6
4
3
End of operation
After disconnecting the patient: 1 Switch the main switch to 0.
If the Oxylog is supplied by O
Fully close the cylinder valve.
If the Oxylog is supplied by the central gas supply:
Remove the gas supply connector.
2 cylinder:
D
2
009
Oxylog 1000
1
010
10
Page 12
Care
Care
After each operational use of the ventilator, dismantle and
disinfect the reusable ventilation valve and ventilation hose, the volume metering parts and the reusable PEEP valve.
The ventilator and compressed gas hoses must also be
cleaned/disinfected if severely soiled.
Use the single-use system, consisting of the ventilation
valve, the ventilation hose and the single-use PEEP valve, only once. These parts are marked: For single use only!
Dispose of the single-use components in accordance with
local regulations.
Dismantling
Remove Oxylog 1000 from its holder.
1 Unscrew the O2 compressed gas hose from the
Oxylog 1000.
2 Remove the ventilation hose from the connector on the side
of the ventilator.
CAUTION!
To remove the ventilation hose, always grip it by the sleeve and not the spiral stiffening ridges. Otherwise the spiral ridges may be torn, e.g. at the sleeve, or the hose may be ripped out of the sleeve.
1
2
011
Remove the ventilation valve from the ventilation hose.
Pull the PEEP valve out of the ventilation valve,
or
remove and dismantle the volume metering parts.
012
11
Page 13
Care
Dismantling the ventilation valve
Turn the cover about 45o anti-clockwise = unlock, and remove the cover.
3 Remove the diaphragm. Do not dismantle any further.
WARNING!
Protect the diaphragm from damage.
WARNING!
Do not allow any foreign matter to enter the housing of the ventilation valve.
WARNING!
The red non return valve in the diaphragm should not be removed, damaged or bent.
Otherwise, impaired valve operation will occur, putting the patient at risk.
1
2
013
Disinfecting/Cleaning
Use only preparations classified as »surface disinfectants« for disinfecting. For material compatibility, we recommend preparations based on
—aldehydes, — alcohols, — quaternary ammonium compounds.
Due to possible damage to materials, the following preparations are unsuitable:
— compounds containing alkylamine — compounds containing phenol — halogen-releasing compounds — strong organic acids — oxygen-releasing compounds. For users in the Federal Republic of Germany, we recommend
the use of disinfectants listed in the current DGHM list (DGHM = German Society for Hygiene and Microbiology). The DGHM list (published by mhp Verlag GmbH, Wiesbaden) also specifies the active basis of each disinfectant. For countries where the DGHM list is unavailable, we recommend products based on the above active bases.
12
Page 14
Care
Wipe disinfecting
Ventilator, O2 compressed gas hose:
Wipe disinfect with e.g. Buraton 10 F or Terralin. Strictly
follow the Instructions for Use of the disinfectant manufacturer. Coarse impurities must first be removed with a disposable cloth.
CAUTION!
Do not allow any liquids to penetrate inside the ventilator or O
2 compressed gas hose.
Liquid inside the system can impair ventilation.
Immersion disinfecting
Dismantled components of the ventilation valve, ventilation hose, volume metering parts, not the Volumeter 3000:
Disinfect in disinfectant bath with e.g. Gigasept FF =
formaldehyde-free. Strictly follow the Instructions for Use of the disinfectant manufacturer. Thoroughly stir the parts in the solution. Do not clean with hard brushes. Do not allow any foreign object to enter the inside of the ventilation valve.
Rinse thoroughly with Aquadest.
CAUTION!
Allow the parts to dry fully. If any water remains in the ventilation valve, the ventilation function may be impaired.
Sterilising
Sterilise as follows if necessary: Disassembled parts of the breathing valve, ventilation hose
and mask:
Sterilise in hot steam at 134
accordance with EN 285: at least 3 minutes.
NOTE:
Do not expose the plastic parts to the hot steam for more than 10 minutes, since this can accelerate the ageing of the materials.
PEEP valve and Volumeter 3000
treat as specified in their specific Instructions for use.
After care
Preparation, page 14.
Connect the O2 supply, page 16.
Betriebsbereitschaft prüfen, page 18.
o
C. Exposure time in
13
Page 15
Preparation
Preparation
The reusable ventilation hose and the ventilation valve are supplied with the ventilator. Alternatively, the optional, pre-assembled singe-use system, consisting of a ventilation hose and ventilation valve, can be used. These parts are marked: For single use only!
Mounting the ventilation valve
1 Check that the red non-return valve is securely seated in the
diaphragm and lies flat on the diaphragm.
WARNING!
The red non return valve in the diaphragm should not be removed, damaged or bent.
Otherwise, impaired valve operation will occur, putting the patient at risk.
1
2
2 Insert the diaphragm in the valve housing, with the red non-
return valve facing the housing.
3 The bead of the diaphragm must lie flat and flush on the
edge of the housing.
4 Fit the cover as shown, press down and turn 45° clockwise
to lock. The inspiration connector and the expiration connector should be facing one another. The diaphragm should fit into the housing without creasing.
5 Fit the ventilation hose to the inspiration connector of the
ventilation valve.
4
3
014
5
6 Fit the ventilation hose to the connector on the ventilator.
14
015
6
016
Page 16
Preparation
Mounting the PEEP valve (option)
Both a reusable PEEP valve and a single-use PEEP valve are available. The single-use PEEP valve is marked: For single use only!
Mount the PEEP valve on the expiration connector of the
ventilation valve.
Measurement of the expiratory volume is not possible when the PEEP valve is used.
029017
Installing the Oxylog 1000
For stationary use
Place firmly on a level, non-slip surface, ensuring unit
cannot fall, or:
Suspend at the head of the bed,
or:
Suspend on a wall rail, as illustrated.
For mobile use in vehicles
Use the vehicle mount
1 Suspend Oxylog 1000 by its bracket on the rod of the
holder.
2 Swivel the ventilator up into the holder until it clicks into
place.
To pull out the ventilator: 3 Press the release catch from underneath.
D
Oxylog 1000
1
3
2
018
15
Page 17
Preparation
Connecting the O2 supply
Take care when handling oxygen.
WARNING!
O2 is highly flammable and intensively propagates any source of fire.
Do not smoke or allow any naked flame in the vicinity of the O
2 supply.
Protect O to intense heat.
WARNING!
Do not oil or grease O2 fittings, such as cylinder valves and pressure-reducing valves, and do not handle these parts with greasy hands. Fire risk.
WARNING!
Only open/close cylinder valves by hand, turning evenly. Never use a tool to open/close the valves.
2 cylinders against falling, and do not expose them
For supply by O
WARNING!
2 cylinder
Use only compressed gas cylinders that conform to national safety standards and have been duly approved.
Use full cylinders with 200 bar cylinder pressure.
Screw a pressure-reducing valve onto the O2 cylinder (for
supply pressure 2.7 to 6.0 bar).
WARNING!
Use only pressure-reducing valves with an outlet exhaust valve that limits the outlet pressure to approx. 5 bar in the event of failure.
Connect the Oxylog 1000 to the pressure-reducing valve
with the compressed gas hose.
Open the cylinder valve by turning slowly as far as it will go.
CAUTION!
Do not insert metering valves or flow meters in the O
2 supply of the Oxylog 1000.
Risk of impaired function of the ventilator with consequent danger to the patient.
O2
019
16
Page 18
Preparation
Determining the gas content of the O2 cylinder
Example: Cylinder pressure measured at the pressure gauge of the
pressure-reducing valve: 200 bar
2 cylinder capacity: 2.5 L
O Compressed gas content: 2.5 L x 200 bar = approx. 500 L
Estimated operating time of the Oxylog 1000
Example: Freq. 10 bpm, V
T = 1 L, MV = 10 L/min
Operating time =
Operating time = = approx. 45 min
Compressed gas content [L] (MV +1*) [L/min]
500 11
If Oxylog 1000 is switched to »Air Mix«, the gas consumption is reduced by about 50 %, thereby increasing the operating time to about 90 minutes.
For supply from a central O
Screw the compressed gas hose to the Oxylog 1000 and
insert the gas connector in the O
2 supply
2 supply socket.
In exceptional cases, the Oxylog 1000 may also be supplied with compressed air, in which case the O
2 concentration is
always 21 % by volume.
For use with the Dräger Oxator
Screw the O2 compressed gas hose to the Oxylog 1000.
Insert the gas connector securely into one of the two O2
ports until it clicks into place.
NOTE:
Strictly follow the Instructions for Use of the Oxator.
_________
* Intrinsic consumption of the device: approx. 1 L/min
020
021
17
Page 19
Checking readiness for operation
Checking readiness for operation
— after changing the ventilation valve, — after each care and reassembly procedure, — at the latest every 6 months.
ocument testing in the medical products book.
Connecting the test lung
The test lung consists of the mask elbow for connecting to the ventilation valve, the Ø7 catheter connector for simulating the resistance of the airways and a 2 L breathing bag to simulate lung compliance.
Insert the mask elbow of the test lung onto the patient
connector of the ventilation valve.
Switch on the O2 supply
Open the O2 cylinder supply by turning slowly as far as it
will go or
insert the O2 compressed gas connector in the supply
socket until it clicks into place = tapping position.
Testing ventilation function
Set the ventilator when using a reusable hose:
1 »MV« knob approx. 10 L/min 2 »Freq.« knob approx. 10 bpm 3 »Pmax« knob approx. 55 mbar 4 Main switch I (On) 5 Switch »No Air Mix«
Set the device when using a disposable hose to the settings
as described in the accompanying leaflet of the disposable hose.
D Oxylog 1000
5
4
123
022023
Oxylog 1000 ventilates the test lung.
After 5 ventilation strokes ventilation is constant, no alarms should occur.
18
Page 20
Checking readiness for operation
Testing the »Paw >« alarm
With the existing setting 1 Keep test lung completely deflated and observe pressure
gauge:
Oxylog 1000 limits the airway pressure to approx. 55 mbar.
2 The »Paw >« indicator turns red, and the audible alarm is
sounded.
1 Release the test lung. 2 The »Paw >« indicator and audible alarm are deactivated.
D Oxylog 1000
2
Testing the »Paw <« alarm
With the existing setting 3 Remove the test lung from the ventilation valve: 4 The red »Paw
alarm is sounded. 5 Press the » 4 The »Paw
muted for up to 2 minutes. 6 The indicator »
tone has been muted.
3 Refit the test lung: 4 The »Paw <« indicator and audible alarm are deactivated.
<« indicator is displayed, and the audible
g« button,
<« indicator remains red; the audible alarm is
g« turns yellow to indicate that the alarm
1
D
024
456
Oxylog 1000
3
19
025
Page 21
Checking readiness for operation
Checking the »Psupply« alarm
Switch off the gas supply.
Close the cylinder valve or remove the compressed gas connector.
1 The colour of the »P
red, but there is no audio alarm in this case.
Restore the gas supply:
1 The colour of the »P
green.
Remove test lung from ventilation valve.
supply« indicator changes from green to
supply« indicator changes from red to
D
1
Oxylog 1000
After all tests have been successfully completed, the ventilator is ready for operation.
026
20
Page 22
Troubleshooting
Troubleshooting
This table is intended to assist the user to detect the underlying cause in the event of an alarm and to remedy the situation rapidly.
Fault Cause Remedy
Ventilator does not build up any airway pressure
»Psupply« indicator red
Ventilator remains on »Inspiration« Supply pressure at ventilator inlet too
Patient cannot exhale, or can only exhale with difficulty
»
Paw >« indicator red
Audible alarm is sounded The minute volume is not fully applied
»
Paw <« indicator red
Audible alarm is sounded
Gas capacity of O
Supply pressure at ventilator inlet too low; no central gas supply or O empty.
low.
Oxylog 1000 faulty. Call Dräger Service.
Ventilation hose kinked. Route the ventilation hose without
Red non-return valve in the diaphragm is faulty or »creased«.
Stenosis in the airways. Free the airways.
Ventilation hose kinked. Route the ventilation hose without
Lung compliance reduced. Set the »Pmax« higher.
Patient is "fighting the ventilator" Change the ventilation pattern or sedate
Disconnection/leakage at the patient connection, ventilation valve or ventilation hose.
Diaphragm of the ventilation valve incorrectly fitted or damaged.
Leaky cuff. Inflate the cuff and check for leaks.
2 cylinder fully spent. Immediately connect the ventilator to a
full O
Establish sufficient supply pressure:
2 cylinder
2.7 to 6 bar.
Establish sufficient supply pressure:
2.7 to 6 bar.
kinking it.
Open the ventilation valve and assemble correctly, see page 14.
kinking it.
the patient.
Ensure that all connections are leakproof.
Mount the diaphragm correctly, or replace diaphragm if faulty – see page 14. Replace the single-use system.
2 cylinder.
21
Page 23
Maintenance intervals
Maintenance intervals
Clean and disinfect the ventilator or parts before each repair, even if returning components to the factory for repair.
Inspection and servicing of the ventilator
Pressure-reducing valve Major overhaul in accordance
Disposing of the apparatus
— At the end of the period of use. — Dispose of Oxylog 1000 correctly, after consulting the
appropriate disposal contractors. Relevant legal provisions should be complied with.
Every 2 years by trained and qualified personnel.
with the operating instructions for the pressure-reducing valve by trained and qualified personnel.
22
Page 24
What's what
Front view
What's what
6 7 8 9 10
11
5 4 3 2 1
1 Main switch 0/I 2 »MV« rotary control knob for minute volume 3 »Freq.« rotary control knob for ventilation frequency 4 Rotary control knob for upper alarm limit »Pmax« 5 Pressure gauge for inspiratory airway pressure 6 Indicator »
g« key for muting the audible alarm tone
7 » 8 Indicator »Paw <« (Paw lower alarm limit) 9 Indicator »Paw 10 Indicator »P 11 »Air Mix / No Air Mix« switch
g« to mute audible alarm tone
>« (Paw upper alarm limit)
supply« for supply pressure
MT-214-99A
23
Page 25
What's what
Side view
1 Connection for ventilation hose and ventilation valve 2 Connection for O
2 compressed gas hose
1 2
MT-975-2000
24
Page 26
Technical data
Technical data
Environmental conditions
Operation
Temperature –18 to 50
Atmospheric pressure 700 to 1100 hPa
Humidity 15 % to 95 % r. h.
Storage
Temperature –18 to 70
Atmospheric pressure 700 to 1100 hPa
Humidity 15 % to 95 % r. h.
o
C
o
C
Performance characteristics
Operating principle Flow chopper
Control Timer-controlled, constant volume
Ventilation frequency 4 to 54 bpm ±15 %, at least ± 2 bpm
Breathing time ratio permanently set 1:1.5; tolerance: 1:1.2 to 1:2
*
*
Minute volume 3 to 20 L/min ±20 %, but at least ±1 L/min
O
2 concentration of the ventilation
gas when supplied with O
2
Switch set to »Air Mix« 60 % O2 by vol. ±10 %
for MV greater than 7 L/min for MV less than 7 L/min:
the O
2 concentration increases up to 100 % O2 by vol.
Switch set to »No Air Mix« 100 % O
Safety valve 80 ±10 mbar
2 by vol.
**
Pressure gauge –10 to 80 mbar
±1.6 % of upper range limit
Main switch I – 0
Patient system consisting of 1.5 m hose with ventilation valve
Compliance approx. 1 mL/mbar
Expiration resistance <6 mbar/L/s
Clearance space approx. 12 mL
Connection for PEEP valve 30 mm cone in accordance with EN 1281-1
* The tolerances indicated apply to the table on page 7 for frequency and minute volume relative to the reference condition NTPD (20 oC,
1013 hPa, dry gas). The tolerances are greater, for physical reasons, if maximum minute volume and maximum frequency are set at the same
time. ** 1 mbar = 0.1 kPa
25
Page 27
Technical data
Alarms
P
supply low Alarm triggered when the supply pressure drops below 2.7 bar
In the case of a slow pressure drop, an audible alarm is also activated.
Paw high > The visual and audible alarm are activated when the set value is
exceeded. Range: 25 to 60 mbar ±10 %, but at least ±4 mbar
Paw low
< The visual and audible alarm are triggered if a pressure of
10 mbar ±3 is not exceeded during inspiration.
Audio alarm muting The audible alarm can be muted for up to 2 minutes.
Audible alarm volume >75 dB (A) at a distance of 1 m
*
.
Gas supply
Supply gas Medical O
2,
In exceptional cases: compressed air
Conditioning of supply gas Dry, oil-free, dust-free
Supply From a central O
2 supply system
or from compressed gas cylinders
Gas pressure 2.7 to 6.0 bar at 60 L/min
O
2 cylinders and pressure-reducing valves Must conform to national safety standards and be duly certified
Pressure-reducing valves Must have an exhaust valve on the outlet side to limit the outlet
pressure to about 5 bar in the event of a fault
O
2 connection NIST O2, NIST AIR/O2, DISS or AFNOR
Gas consumption
Internal control approx. 1.0 L/min
Patient
with »Air Mix« approx. 50 % of the effective minute volume
with »No Air Mix« 100 % of the effective minute volume
Typical operating time for a minute volume of 10 L/min
with 11 L O
2 cylinder / 200 bar approx. 200 minutes without mixing (No Air Mix)
approx. 400 minutes with mixing (Air Mix)
with 2,5 L O
2 cylinder / 200 bar approx. 45 minutes without mixing (No Air Mix)
approx. 90 minutes with mixing (Air Mix)
* 1 bar = 1 kPa x 100
26
Page 28
Technical data
Dimensions (W x H x D) mm 215 x 90 x 215 (without handle)
Weight
Oxylog 1000 (basic unit) 3.15 kg
Ventilation hose, reusable, 1.5 m 0.35 kg
Ventilation valve, reusable 0.06 kg
Single-use hose system
(consisting of ventilation hose and ventilation valve)
Materials used
Ventilator housing Impact-resistant acrylonitrile butadiene styrene (ABS)
Ventilation hose, reusable Silicone rubber
Ventilation valve housing, reusable Polysulphone (PSU), silicone rubber
Single-use hose system Silcone rubber, polycarbonate (PC), polypropylene (PP),
0.11 kg
ethylene vinylacetate (EVA)
Classification according to Directive 93/42/EC, Appendix IX
UMDNS-Code Universal Medical Device Nomenclature System
Type of protection IPX4
Class IIb
18 – 098
Abbreviations and symbols
Air Mix Mixing of O
IPPV Intermittent Positive Pressure Ventilation
Freq. Ventilation frequency, strokes per minute
MV Minute volume, L/min
No Air Mix No mixing with ambient air, 100 % O2 by volume
Paw Airway pressure Paw > Upper alarm limit of airway pressure Paw < Lower alarm limit of airway pressure
PEEP Positive End Exspiratory Pressure
Pmax Set point for upper alarm limit Paw
Psupply Supply pressure
H
g
Setting symbol for ventilation rate and airway pressure during heart-lung resuscitation
Audio alarm muting
2 with ambient air, approx. 60 % O2 by vol.
27
Page 29
Appendix
Appendix
Minute volume and O2 concentration as a function of airway pressure
Air Mix function
Oxylog 1000 supplies the air/O2 mixture (Air Mix) by means of an injector that draws in additional air from the atmosphere to generate an air/O 60 % by volume. Due to the laws of physics, the suction effect of injectors decreases with increasing back pressure. With high airway pressures, the minute volume MV set may be reduced during the Air Mix function, thereby increasing the O concentration. At airway pressures up to 20 mbar, the set and applied minute volume (MV) are the same. An additional increase of airway pressure by 10 mbar leads to a reduction of the volume applied by about 10 %. The set minute volume MV should possibly be increased appropriately. Since less air is drawn from the atmosphere when there is an increase in airway pressure, the O At a low minute volume MV of under 7 L/min, concentrations of up to 100 % by volume can be attained.
2 mix with an O2 concentration of about
2 concentration increases.
2
No Air Mix function
The applied minute volume MV is not a function of the airway pressure. The O
Minute volume as a function of ambient air pressure
The applied minute volume MV is also dependent on the atmospheric pressure of the surrounding area (reference value 1013 hPa). If the atmospheric pressure is reduced by 100 hPa, the supplied minute volume MV is increased by about 10 %.
2 concentration is always 100 %.
28
Page 30
Block circuit diagram
Appendix
1413
1
10
12
11
Function description of the Oxylog 1000 with reference to the block circuit diagram
The supply gas, O filter 1, the on/off switch 2 to the ventilation block 3, which can optionally be connected to the "Air Mix" module
The minute volume is set by the "MV" valve 5 and controls the ventilation block 3.
The frequency is set by the "Freq." valve 6 and acts on the frequency control system 8.
The upper alarm limit "Pmax" is set with the "Pmax" rotary knob 9 and opens a valve in the ventilation block 3, so that the airway pressure is limited during inspiration.
The alarm logic control" 10 by push-button » pressed, the yellow indicator » audible alarm is suppressed for up to 2 minutes.
Additional relief valve 13 is permanently set and opens at a specified pressure of around 80þmbar.
The emergency air valve 14 permits spontaneous breathing by the patient in the event of ventilator failure.
2 (or compressed air), is channelled via the
7, which is coupled to the alarm logic
8 is connected to the "audible alarm muting
g« 11. When button »g« 11 is
g« 12 is displayed, and the
2
8
7
9
Pmax Freq. MV
4.
34
Air Mix
56
MV
04329033
29
Page 31
Order List
Order List
Name and description Part No.
Basic unit
Oxylog 1000 Time-cycled, constant volume emergency ventilator with alarms for high and low airway pressure and low supply pressure
Optional accessories
Caddy 57 03 300
Carrying system 1000 2M 86 001
Wall Holder for Carrying System 1000 2M 86 103
Spare bag for Carrying System 1000 AB 41 047
Vehicle bracket 84 12 069
Test lung 84 03 201
Ventilation hose E ISO, reusable, 1.5 m 2M 86 511
Ventilation hose E ISO, reusable, 3.0 m 21 12 760
Ventilation valve, reusable 2M 86 800
Reusable PEEP valve (0 to 10 mbar) 84 07 475
Single-use hose system (25 pieces) consisting of ventilation hose with ventilation valve
Single-use PEEP valve (5 to 20 mbar) 2M 86 832
2M 86 105
2M 86 837
Connecting hoses
*
-connecting hoses
CG
Gas Supply System 57 04 500
* Central piped gas supply
30
Page 32
Index
Index
A
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Accessories Applications
B
Before using for the first time . . . . . . . . . . . . . . . . . . . . . . . . . 4
C
Central O2 supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Checking readiness for operation Connection O2 supply
D
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Device Check Disinfecting/Cleaning Dismantling Disposing of the apparatus Dräger Oxator, use of
E
End of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
End-expiratory volume measurement Environmental conditions
F
Front view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
H
Heart-lung resuscitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
I
Immersion disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Installing the Oxylog 1000 Intended use IPPV
M
Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
O
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Order List
P
PEEP valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Performance characteristics Pmax Precautions
S
Side view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Sterilising Symbols
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
. . . . . . . . . . . . . . . . . . . 18
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . 22
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
. . . . . . . . . . . . . . . . . 10
. . . . . . . . . . . . . . . . . . . . . . . . . . . 25
. . . . . . . . . . . . . . . . . . . . . . . . . . 15
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
. . . . . . . . . . . . . . . . . . . . . . . . 25
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
T
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Test lung, connecting Testing the »Paw Testing the »Paw >« alarm Troubleshooting
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
. . . . . . . . . . . . . . . . . . . . . . . . . . 19
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
V
Ventilation valve, Dismantling . . . . . . . . . . . . . . . . . . . . . . . . 12
Ventilation with a mask Ventilation with PEEP
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
W
What's what . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Wipe disinfecting
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
31
Page 33
Index
This page is intentionally left blank
32
Page 34
These Instructions for Use apply only to Oxylog 1000 with Serial No.:
If no Serial No. has been filled in by Dräger these Instructions for Use are provided for general information only and are not intended for use with any specific machine or device.
0344
Directive 93/42/EEC concerning Medical Devices
Dräger Medical b.v.
Kanaaldijk 29
z
5683 CR BEST The Netherlands
+31 499 331 331
y
FAX +31 499 331 333
medical.best@draeger.com
!
www.draeger-medical.com
90 37 394 - 06 en - GA 5503.410 © Dräger Medical b.v. 5th edition - January 2007 Subject to alteration
Loading...