The procedures in this section shall be performed in their entirety each time a
component is removed, replaced, calibrated, adjusted and during all scheduled
Periodic Manufacturer's Certification (PMC) visits. A PMC Checklist form, P/N
S010211 is available from Draeger Medical, Inc. and shall be completed by the
Technical Service Representative each time a PMC is performed. Steps in the
procedure marked with (
checklist form.
Space is also provided on the PMC checklist form to record the results of a vapor
concentration test. Refer to the current Anesthesia Equipment & Monitoring System
Service Information CD-ROM Service Procedures section for vapor concentration
verification procedures.
NOTE:Test equipment listed below with an asterisk (*) requires calibration at a
maximum interval of one year. Verify the dates on test equipment calibration
labels. DO NOT USE any test equipment having an expired calibration date.
Notify your supervisor immediately if any equipment is found to be out of
calibration. In the space provided at the bottom of the PMC checklist form,
record the Model and ID number of all calibrated test equipment used.
In the space provided at the bottom of the PMC checklist form, record the Model and
ID number of all calibrated test equipment used. Also record the calibration due
dates. Examples are: multimeter, digital pressure meter, Riken gas analyzer, safety
analyzer, volumeter, trace gas analyzer, simulators.
ü) require a response at the corresponding line on the
Test Equipment Required:
•*Electrical Safety Analyzer (Biotek 501 Pro or equivalent)
•*Pressure Gauge with DISS Adapters (P/N 4114807 or equivalent)
•*Flowmeter 0-250 ml min. (P/N S000081 or equivalent)
•*Volume Meter (P/N 2212300 or equivalent)
•*Digital Pressure Manometer (SenSym PDM 200CD or Equivalent)
•*Riken Gas Indicator (Model 18H, or 1802D or equivalent)
•Stop Watch
•Test Lung (P/N 4115128)
Materials Required:
•Breathing Bag 3 liter (P/N 9995330 or equivalent)
•Patient Circuit: Y-piece, elbow, 2x 32” x 22mm hoses
•Hose 22 mm x 32” (P/N 9995132)
•Fresh Gas Outlet Volume Test Device (P/N S010158 or equivalent)
•Fresh Gas Leak Test Adapter (P/N 4115041 or equivalent)
•Volumeter/Fresh Gas Adapter (P/N 4115042)
6-1
Page 4
PMC PROCEDURE (continued)
Materials Required (continued):
•Test Terminal 2x (P/N 4104389 or equivalent)
•Breathing System Leak Test Device (P/N S010159 or equivalent)
•PDM/Suction Adapter (P/N 4115038)
•Scavenger Adapter (P/N 4108114)
•Pressure Monitor Test Adapter (P/N 4115043 or equivalent)
Key test equipment and materials illustrations are shown on following pages.
NM MOBILE/MILITARY
6-2
Page 5
NM MOBILE/MILITARY
TEST EQUIPMENT AND FIXTURE ILLUSTRATIONS
4115041
FRESH GAS
LEAK TEST DEVICE
PMC PROCEDURE (continued)
TEST TERMINAL
S010158
FRESH GAS OUTLET VOLUME TEST DEVICE
TEST TERMINAL
S010159
BREATHING SYSTEM LEAK TEST DEVICE
4115042
VOLUMETER/
FRESH GAS HOSE
ADAPTER
SP18201
6-3
Page 6
PMC PROCEDURE (continued)
TEST EQUIPMENT AND FIXTURE ILLUSTRATIONS - Continued
4114807 PRESSURE TEST ASSEMBLY , WITH ADAPTERS
NM MOBILE/MILITARY
SP18202
4114830-0024114830-001
4114830-0044114830-003
4114830-0064114830-005
6-4
Page 7
NM MOBILE/MILITARY
TEST EQUIPMENT AND FIXTURE ILLUSTRATIONS - Continued
PMC PROCEDURE (continued)
SP18203
TEST
VOLUME
METER
4115087 33mm x 22mm ADAPTER
6-5
Page 8
PMC PROCEDURE (continued)
TEST EQUIPMENT AND FIXTURE ILLUSTRATIONS - Continued
PDM TO PATIENT SUCTION ADAPTER
NM MOBILE/MILITARY
4115038
AVENGER
4108114
SCAVENGER ADAPTER
4115043
PDM TO
MONITOR ADAPTER
SP18204
6-6
Page 9
NM MOBILE/MILITARY
TEST EQUIPMENT AND FIXTURE ILLUSTRATIONS - Continued
900
PMC PROCEDURE (continued)
15
Liter
100
S000081
FLOW METER
TEST STAND
800
700
vol/min
600
10
500
5
200
ml
300
400
4104389
TEST TERMINAL
ADAPTER
2212300
MINUTE
VOLUMETER
4115128
SIEMENS TEST LUNG
Rev. D
TEST TERMINAL
SP18205
6-7
Page 10
PMC PROCEDURE (continued)
NM MOBILE/MILITARY
Periodic Manufacturer’s Certification General Instructions
The purpose of these procedures is to provide detailed instructions for performing a Periodic
Manufacturer’s Certification (PMC) inspection on the Narkomed Mobile/Military anesthesia
machine.
A PMC consists of a complete Periodic Manufacturer ’s Service procedure and a certification level
inspection based on Draeger Medical, Inc. Recommendations and equipment performance.
Additional inspections are also performed to ensure proper product labeling.
Several additional documents have been created to assist the technician through the process.
Following is a brief description of the purpose of each document.
Field Service Procedure:
Periodic Manufacturer’s Certification Forms - Part Number SP00175.This procedure illustrates
the sample checklists with typical periodic maintenance items filled in, including vapor
concentration verification tests, parts replaced, general comments and certification levels. Also
included are sample PMC labels marked to show several levels of certifications. An excerpt from
Draeger Medical, Inc.’s Anesthesia System Risk Analysis and Risk Reduction is included, and also
a sample of an Executive Summary to be furnished to the hospital’s Risk Manager or Chief of
Anesthesia.
Field Service Procedure:
DMI Recommendation Guidelines Index Anesthesia Systems - Part Number S010250.This
Guideline was created to provide an assessment of each machine’s certification. It contains
various comprehensive overviews of possible equipment conditions and their associated
certification levels.
The first list in the Recommendation Guidelines is a reference chart for machine certification
based on equipment status. The second is an abbreviated summary of all DMI Recommendations
and Failure Codes including the Condition Number, Equipment Condition, Recommended
Corrections, Certification Code, and Tests Affected when applicable.
There is also a matrix classified as “Failure Codes” which identifies the correct manner in which
to document equipment tests that fail, or were unable to be performed due to circumstances
beyond the control of the service technician performing the inspection. (Ex: Air cylinder supply is
unavailable to perform an Air High Pressure Leak test.) The Failure Codes section also indicates
suggested resolution of the situation. Failure Code numbers begin at 34 and use the same
certification levels strategy, and carry the same weight as NAD Recommendation equipment
condition codes.
The final matrix is the most comprehensive index sorted by machine model and includes
Equipment Condition, Certification Code, and DMI Recommendations. It also specifies any
suggested upgrade path including ordering information that should be taken such as installing a
Bellows with Pressure Limit Control 4109664-S01 Kit, after market modification kit to a machine
not equipped with pressure limit control.
The letters A, B, C, D and the Roman Numerals I, II are used as codes in the individual matrix for
each model of anesthesia machine. The letters A, B, C, and D are used in descending order to
indicate the certification level of the equipment. They are as follows:
A = Certified
B = Certified with Recommendations
C = Conditionally Certified
D = No Certification
6-8
Page 11
NM MOBILE/MILITARY
Roman Numerals I and II do not affect the certification level but rather are provided to give
further instructions to the end user as follows:
I = The system in its present configuration shall only be used with a CO2 monitor
incorporating an apnea warning. The operator of the system is advised to frequently scan
the CO2 readings and alarm thresholds.
II =The present configuration of equipment requires that the unit operate at all times
with an oxygen analyzer that includes a low oxygen warning. The operator of the system
is advised to frequently scan the oxygen readings and alarm limits.
Following is an explanation of machine certification levels:
Certified- No DMI Recommendations or Failure Codes apply to machine being inspected. (Only
item number 33 - "No Recommendations" shall apply for this certification level.)
Certified with Recommendations- A numbered DMI Recommendation or Failure Code with a
code of B applies to the machine being examined.
Conditionally Certified- A numbered DMI Recommendation or Failure Code with a code of BCI
or BCII applies to the machine being examined.
No Certification- A numbered DMI Recommendation or Failure Code with a code of D applies to
the machine being examined.
PMC PROCEDURE (continued)
When multiple recommendations apply, "No Certification" would take precedence over
"Conditionally Certified" and "Certified with Recommendations". "Conditionally Certified" would
take precedence over "Certified with Recommendations".
For example:
A Narkomed 2B could have DMI Recommendation number 21 and Failure Code 61.1 that apply.
21 - No ventilator pressure limit control. Code is B. 61.1 - Enflurane agent is unavailable to test.
Code is BC. Correct certification for this machine is BC, which means CONDITIONALLY
CERTIFIED WITH RECOMMENDATIONS.
A Narkomed 4 could have DMI Recommendation numbers 14 and 21 apply.14 - CO2/Agent
monitor exhaust port is not properly connected to the waste gas scavenger. Code B.
21 - No ventilator pressure limit control. Code B.The correct certification for this machine is B,
which means “CERTIFIED WITH RECOMMENDATIONS”.
A Narkomed 2B, 2C or GS could have DMI Recommendation 30 apply.
30 - Anesthesia machine is equipped with inhalation anesthesia vaporizers without an agent
analyzer in the breathing system. Code B. The correct certification for this machine is B, which
means “CERTIFIED WITH RECOMMENDATIONS”.
A Narkomed 6000 could have no DMI Recommendations or Failure Codes apply. The correct
certification level for this machine is Code A, “CERTIFIED”.
The correct certification for this machine is A, which means "CERTIFIED".
Code, D also means "NO CERTIFICATION", also means the machine shall not receive a
Periodic Manufacturer's Certification label. The machine shall receive a "WARNING This System Is Not Certified" label, P/N 4114857. This label shall be placed at a prominent location on the right side of the machine after all other previous PM and "Vigi-
®
lance Audit
Validation" labels have been removed.
6-9
Page 12
PMC PROCEDURE (continued)
NM MOBILE/MILITARY
PM Certification Procedure for Narkomed Mobile/Military Anesthesia System
1. Use the PM Certification form for Narkomed Mobile/Military Anesthesia Systems (P/
N 4115597).
2. Completely fill in the header information.
3. All Narkomed Mobile/Military machines are equipped with Humphrey valves. No
“MJV-2 LUBRICATION” is needed to be performed for this type of anesthesia
machine.
4. Replace the VENTILATOR RELIEF VALVE DIAPHRAGM every 12 months in
accordance with SP00075. Place a check mark and the replacement date at “VENT
VALVE REPLACEMENT” line on the Periodic Manufacturer’s Certification form.
5. If machine is equipped with a HALOTHANE Dräger Vapor 19 or 19.1 vaporizer,
determine if vaporizer must be inspected for soil condition one. Check the serial
number plate located on the rear of the vaporizer for a plus (+) preceding the serial
number. A HALOTHANE vaporizer serial number not preceded with a (+) must be
tested for soil in accordance with SP00073. If vaporizer does not need to be inspected,
indicate so with a plus (+) next to the “Vapor Inspection (H)” line on the Vigilance
Audit form. If vaporizer is soil condition 0, indicate so with “SOIL 0” written next to
the “Vapor Inspection (H)” line on the Vigilance Audit form. If vaporizer is soil
condition one, indicate so with “SOIL 1” written next to the “Vapor Inspection (H)” line
on the Vigilance Audit form. Place a “CAUTION DO NOT USE” label (part # 4114327)
on the vaporizer, and issue a departmental alert. The TSR shall also seek permission
from the equipment operator to remove the failed vaporizer from the machine and
apply a replacement vaporizer or an adapter block onto the mount. All “SOIL 1”
vaporizers must be removed from service for machine to receive certification.
6. Perform the vapor concentration test on all Dräger vapor vaporizers in accordance
with SP00073 at a six month maximum interval. Perform the vaporizer concentration
test on all Desflurane vaporizers in accordance with SP00091 for fixed mount
vaporizers and SP00189 for user removable D-tec vaporizers at a six month maximum
interval. For every vaporizer tested, fill out a "VAPOR VAPORIZER CALIBRATION
CHECK" label (part # S010016). Information on this label shall include your
signature, type of agent, date tested, a No Agent To Test or the test results @ 1%, 2.5%,
4% for H, E, I, or S vaporizers, or @ 4%, 10%, 12%, 16% for Desflurane vaporizers, and
a PASS or FAIL indication. This label shall be attached to the upper right side of the
vaporizer. If vaporizer fails the concentration verification, internal leak, or exclusion
system tests, check "NO" in the "RECOMMENDED FOR USE" section on the PM
Certification form. Place a "CAUTION DO NOT USE" label (part # 4114327) on the
vaporizer, and issue a departmental alert. The TSR shall also seek permission from
the customer to remove the failed vaporizer from the machine and install a
replacement vaporizer or an adapter block onto the mount. All nonfunctional
vaporizers must be removed from service for machine to receive certification.
6-10
Page 13
NM MOBILE/MILITARY
7. Proceed with PM Certification procedure. If any tests fail refer to the "Failure Codes"
listing in DMI Recommendations Guidelines Index (P/N S010250) to determine
correct certification level starting point. Failure codes shall be documented on the
"RECOMMENDATIONS / GENERAL COMMENTS" section of the PM Certification
form and on the Executive Summary. If a test fails that has not been identified by the
"Failure Codes" list, consult with Draeger Medical, Inc. to assess the proper
certification level.
8. Based on the "EQUIPMENT CONDITION" inspect the machine for any "DMI
RECOMMENDATIONS" that would apply. Use the Narkomed Mobile/Military section
of the "RECOMMENDATION GUIDELINES INDEX" (P/N S010250). Note all
applicable DMI recommendations on the Executive Summary. NOTE: If using a
carbon form, indicate the Equipment Condition number and to see reverse side under
the "RECOMMENDATIONS / GENERAL COMMENTS" section of the form.
9. Determine the correct certification level of the machine based on the combined lowest
common denominator of "Equipment Conditions" and "Failure Codes". If the machine
is at least conditionally certified fill out the "PM CERTIFICATION" label. Check the
box(s) on the validation label where appropriate. Write the month and year, (three
months from date of PM Certification) next to "NEXT VISIT DUE:". If certification
level is "D", machine shall not receive a "PM CERTIFICATION" label. Any machine
not receiving a PM Certification label shall receive a "WARNING NOT CERTIFIED"
label, P/N 4114857. This label shall be placed at a prominent location on the left side
of the machine after all other previous PMS and Vigilance Audit Validation labels
have been removed.
PMC PROCEDURE (continued)
10.In the "CERTIFICATION LEVEL" section of the PM Certification form, record the
last visit certification level, the current certification level and the next visit due month
and year, (three months from date of PM Certification) in the spaces provided.
11.If applicable, remove the previous PM CERTIFICATION VALIDATION label and
attach the new label (P/N S010006 w/phone #, or S010007 w/o phone #) in a prominent
location on the rear of the anesthesia machine.
12.Check the appropriate boxes on the "PM CERTIFICATION NOTICE" label, (part #
S010011). If the machine is not certified, the last box of this notice label shall be
marked. Attach this notice near the flowmeter shield of the anesthesia machine.
13.Have the customer sign each PM Certification form or the Executive Summary, and
review any Failure Codes equipment conditions and DMI Recommendations with the
customer.
14.Return top copy to Draeger Medical, Inc. Service Department, keep middle copy for
service organization records, give bottom copy to customer.
6-11
Page 14
PMC PROCEDURE (continued)
(
ü)6.1 SELF-DIAGNOSTICS
6.1.1Turn the System Power switch to ON and verify the "ON" LED is
lighted?
6.1.2Verify all LED's on the keypad and ventilator displays are lit.
6.1.3Verify that the following is displayed on the Display:
NARKOMED
COPYRIGHT 1997-1998 NAD. INC.
VERSION X.XX NARKOMED
SOFTWARE ID: XXXX
ü)6.1.4Record the machine software version on the header of the checklist form.
(
6.2 ELECTRICAL SAFETY- One Year Service Interval; Due Date ______
(
ü)6.2.1Battery Check & Ground Continuity
6.2.1.1Unplug the AC power cord for all devices mounted to the
machine that may provide an alternate path to earth ground,
such as a Desflurane vaporizer.
6.2.1.2Unplug the machine’s AC power cord and plug the power cord of
the safety analyzer into this AC receptacle.
NOTE: Do not plug the safety analyzer power cord into a line isolation
monitor circuit, as inaccurate readings may occur.
NOTE: The BIOTECH 501 PRO will automatically test the source
outlet for open ground (or ground resistance of 31 Ohms or
higher), reverse polarity, open neutral and open line. (The latter
two conditions will prevent the analyzer from powering up.)
6.2.1.3Is the green "AC POWER" LED off? __ (Y)
6.2.1.4Does the "AC PWR FAIL" message appear in the Advisory
display? __ (Y)
6-12
Page 15
NM MOBILE/MILITARY
T
SPIROMED
6.2.1.5Press and hold the "BATTERY TEST" button.
6.2.1.6Is green Battery Test LED lighted? __ (Y)
6.2.1.7Release the "BATTERY TEST" button.
6.2.1.8Turn on the safety analyzer and configure it to measure
6.2.1.9Plug the machine’s AC power cord into the safety analyzer.
6.2.1.10 Apply the analyzer’s test lead to a cylinder yoke bolt.
6.2.1.11 What is the value displayed on the safety analyzer? ___ ohm (0-
PMC PROCEDURE (continued)
“Resistance” according to the manufacturers documentation.
Attach a test lead to the “SINGLE LEAD” connector of the
analyzer. Connect the other end of the test lead to the AC
receptacle ground socket on the safety analyzer. Verify a
displayed resistance of 0.00 ohms or, if necessary, calibrate the
unit according to manufacturers documentation.
0.1)
ü)6.2.2Circuit Isolation
(
6.2.2.1Disconnect the respiratory volume sensor cord from the
interface panel.
6.2.2.2Using a multimeter set to its
highest resistance range apply the
test leads between the yoke bolt
and circuit common at the volume
interface test pin. Use the position
indicated on the illustration for
the ultrasonic flow sensor
connector. Use the 12 o’clock
position for the Spiromed
connector. There shall be no continuity between these two
points. Refer to illustrations for proper pin locations.
6.2.2.3Reconnect the respiratory volume
sensor cord to the interface panel.
6.2.3Chassis Leakage Current
6.2.3.1Apply the analyzer test lead to a
cylinder yoke bolt.
6.2.3.2Configure the safety analyzer to
the measure “CHASSIS LEAKAGE CURRENT” according to
the manufacturers documentation.
EY
P15005A
KEY
SP15005
ULTRASONIC
TES
PIN
TEST
PIN
6-13
Page 16
PMC PROCEDURE (continued)
6.2.3.3Record the total leakage current with the Polarity and Ground
switches set as follows:
NM MOBILE/MILITARY
Ground
(ü)6.2.3.3.1NormalNormal
ü)6.2.3.3.2OpenNormal
(
(
ü)6.2.3.3.3OpenReversed
(
ü)6.2.3.3.4NormalReversed
6.2.3.4Verify that the leakage current is 100* microamps or less in
each of the switch positions (110 microamps or less for the 220/
240 volt power supply option).
6.2.3.5300 microamps if external monitors are plugged into
convenience receptacles.
6.2.3.6Shut off and unplug the safety analyzer. Remove the anesthesia
machine plug from the analyzer and plug it back into the
original AC receptacle.
ü)6.3 CONFIGURATION
(
6.3.1Press the CONFIG key.
6.3.2The CONFIGURE screen is displayed.
Polarity
6.3.3Verify the correct Time and Date.
6.3.4Press the exit key to exit the function.
6.4 SERVICE DATA
6.4.1Press and hold the Oxygen High Limit key and the Volume Low Limit
key, and then press the UP ARROW key.
6.4.2The Main Service Screen shall appear.
6.4.3Select and enter the Service Log.
6.4.4Verify any pertinent information from the Service Log. Contact the
Draeger Medical, Inc. Technical Service Department if necessary.
6.4.5Press EXIT to return to the Main Service screen.
6.4.6Using the arrow keys, select the SRVC Service Code.
6.4.7Press the SELECT key and enter your Technical Service Rep. I.D.
number.
6-14
Page 17
NM MOBILE/MILITARY
ü)6.4.8Press the RESET key. This resets the last service date to the current
(
PMC PROCEDURE (continued)
date and resets the hours run since last service to zero.
6.4.9Press the PMS SCHED key.
(
ü)6.4.10Select and enter the month of the next service due date.
NOTE: The internal clock of the machine limits the amount of date advance to a
maximum of six months from the current service date.
6.4.11Press the EXIT key to return the main service screen. If not performing
monitor calibrations, press the EXIT key again to return to normal
operation mode.
6.5 CALIBRATIONS - One Year Service Interval; Due Date ______
6.5.1To bring up the Oxygen Monitor Service Screen, press the Mon Cal key.
6.5.2Remove the Oxygen sensor from the Breathing Pressure Hose Assembly,
and remove the Oxygen Sensor capsule from the Oxygen Sensor
Housing.
ü)6.5.3When the CURRENT CELL A and CURRENT CELL B readings have
(
stabilized, press the ZERO key and verify that the new offset values are
stored.
NOTE: The higher the offset, the higher the calculated O
2 concentration appears
at high concentrations.
6.5.4Put the Oxygen Sensor capsule into the Oxygen Sensor Housing.
6.5.5Press the PRES MON key.
6.5.6Remove Breathing Pressure Hose Assembly from Inspiratory Valve dome
to expose the Breathing Pressure Sensor line to ambient pressure.
6.5.7Let the Current Pressure Value stabilize and press the ZERO key to
store the value.
6.5.8Remove Breathing Pressure Sensor line from the machine.
6.5.9Connect a Pressure Monitor Adapter, (P/N 4115043) and calibrated
manometer to the Breathing Pressure quik connect on the machine
monitor interface.
NOTE: Zero manometer prior to connecting to monitor interface. Slight pressure
will be noticed on manometer when connected to monitor interface, DO
NOT re-zero manometer.
6-15
Page 18
PMC PROCEDURE (continued)
NM MOBILE/MILITARY
6.5.10Pressurize the circuit to 50 cm H
2O and allow the [Current Value] to
stabilize.
ü)6.5.11Press the SPAN key and verify that the new span values are stored.
(
6.5.12Release the pressure, disconnect the manometer and test fixture,
reconnect the Breathing Pressure Sensor line to the quik connect on the
machine monitor interface and insert the Breathing Pressure Hose
Assembly into Inspiratory Valve dome.
6.5.13Press EXIT to return to the Main Service screen.
6.5.14Press EXIT to return to normal operation.
(ü)6.6 ABSORBER INSPECTION
Absorber System Inspection and Maintenance
6.6.1Remove the O
2 Sensor Housing and/or Oxygen Sensor Plug and
Breathing Pressure Hose Assembly from the inspiratory valve dome and
examine the O-rings on each assembly. Replace O-rings as necessary.
6.6.2Remove the Inspiratory and the Expiratory valve domes.
6.6.3Are any pins on the valve crater damaged? Inspiratory ___ (N)
Expiratory ___ (N) Replace complete valves as necessary.
6.6.4Are any pins on the valve domes damaged? Inspiratory ___ (N)
Expiratory ___ (N) Replace complete valves as necessary.
6.6.5Is the valve disc in good condition? Inspiratory ___ (Y) Expiratory ___ (Y)
Replace as necessary.
6.6.6Are the valve dome washers in good condition? ___ (Y) Replace washers
as necessary.
6.6.7Reinstall the Inspiratory and Expiratory valve domes.
ü)6.6.8Ultrasonic Flow Sensor (Mobile Only)
(
6.6.8.1Remove the Ultrasonic Flow Sensor connector hose.
6.6.8.2Are the connector hose, connector, and O-ring components in
good condition? ___ (Y) Replace components as necessary.
6.6.8.3Remove the Expiratory valve.
6.6.8.4Is the washer under the valve in good condition? ___ (Y) Replace
washer as necessary.
6-16
Page 19
NM MOBILE/MILITARY
6.6.8.5Remove the PEEP valve.
6.6.8.6Is the washer under the PEEP valve in good condition? ____(Y)
6.6.8.7Reattach the PEEP valve.
6.6.8.8Reattach the Expiratory valve.
6.6.8.9Remove the Ultrasonic Flow Sensor from the mounting bracket
6.6.8.10 Remove the flow housing/transducer assembly from the
6.6.8.11 Remove both transducers from the flow housing; examine each
6.6.8.12 Reattach the Ultrasonic Flow Sensor to the mounting bracket.
PMC PROCEDURE (continued)
Replace washer as necessary.
electronics housing.
O-ring and condition of all components, replace O-rings as
necessary, then reassemble the Ultrasonic Flow Sensor.
6.6.8.13 Reattach the connector hose between the Ultrasonic Flow
Sensor and Expiratory valve.
(
ü)6.6.9Spiromed Sensor (Military Only)
6.6.9.1Remove the expiratory valve.
6.6.9.2Is the washer under the valve in good condition? ___ (Y) Replace
washer as necessary.
6.6.9.3Remove the Spiromed Sensor.
6.6.9.4Is the washer under the sensor in good condition? ___ (Y) Replace
washer as necessary.
6.6.9.5Remove the PEEP valve.
6.6.9.6Is the washer under the PEEP valve in good condition? ____(Y) Replace
washer as necessary.
6.6.9.7Reattach the PEEP valve.
6.6.9.8Locate the four lateral holes at the sides of the Spiromed Sensor
marked by arrows.
Caution: Use only the Sensor Lubrication Kit (P/N 2218180) for the following
procedure.
6-17
Page 20
PMC PROCEDURE (continued)
6.6.9.9Dip the tip of the pipette into the lubricant and draw lubricant into the
pipette by pulling the pin backwards.
6.6.9.10Insert the pipette into one of the four holes as far as it will go. Push the
pin forward to its stop to inject lubricant into the hole.
6.6.9.11Repeat the previous 2 steps for lubrication of the three remaining
holes.
6.6.9.12Wipe any lubricant residue from the exterior of the sensor.
6.6.9.13Reattach the Spiromed Sensor to the PEEP valve.
6.6.9.14Reattach the Expiratory valve to Spiromed Sensor.
6.6.9.15Remove the Inspiratory valve assembly.
6.6.9.16Is the washer under the valve in good condition? ___ (Y) Replace
washer as necessary.
NM MOBILE/MILITARY
6.6.9.17Reinstall the Inspiratory valve.
6.6.9.18Is there a wing nut on each absorber rod? ___ (Y)
6.6.9.19Inspect the following: canister, canister gaskets, dust cup and O-ring,
and soda lime.
6.6.9.20Are the canister, canister gaskets, dust cup and O-ring, and soda lime
in good condition? ___ (Y) Replace components as necessary.
6.6.9.21Verify the cm H2O gauge is at zero (0) and readjust/replace as
necessary.
NOTE:The small slotted screw is the zero adjustment.
6.6.9.22Reinstall the Breathing Pressure Hose Assembly into the Inspiratory
valve dome.
6.6.9.23Reinstall the O2 Sensor plug into the Breathing Pressure Hose
Assembly.
6.6.9.24Remove the Freshgas Hose Assembly from the FRESHGAS OUTLET.
6.6.9.25Is the Freshgas Hose Assembly and Freshgas Outlet in good condition?
___ (Y)
6.6.9.26Reconnect the Freshgas Hose Assembly to the FRESHGAS OUTLET.
6-18
Page 21
NM MOBILE/MILITARY
ü)6.7 HIGH PRESSURE LEAK
(
PMC PROCEDURE (continued)
NOTE: Minimum cylinder pressures required for High Pressure Leak tests are:
N2O: 600 psi;
O
2, Air: 1000 psi.
6.7.1Turn the machine to Standby.
6.7.2Verify the Auxiliary Oxygen flow control valve is closed
6.7.3Disconnect all pipeline supply hoses at the wall outlets.
6.7.4Open then close and remove each cylinder. And if applicable remove the
yoke plug from each additional yoke assembly.
6.7.5Note the reading on each of the cylinder pressure gauges and start a stop
watch.
6.7.6Are there two (2) yoke index pins installed securely in each yoke? ___(Y)
6.7.7Is the proper gas I.D. label affixed to each yoke? ___ (Y)
6.7.8After two (2) minutes, is the pressure loss equal to or less than 50 PSI?
___(Y)
6.7.9Verify the presence of only one (1) cylinder washer, then reattach and
secure the cylinders to each yoke assembly, then open each cylinder
valve.
6.8 BREATHING SYSTEM
6.8.1Absorber & Freshgas Leak/ Vaporizer Leak
6.8.1.1Close all flow control valves.
6.8.1.2Set the AUTO/BAG selector to BAG.
6.8.1.3Close the APL valve.
6.8.1.4Interconnect the inspiratory valve and expiratory port (on the
Ultrasonic Flow Sensor) or the expiratory valve, as applicable,
with a 22 mm hose (P/N 9995132).
6.8.1.5Attach a Freshgas Leak Test Adapter (P/N 4115041), Test
Terminal Adapter (P/N 4104389), and Digital Pressure
Manometer to the bag mount.
6.8.1.6Apply 50 cm H
2O test pressure to the absorber system and start
a stop watch.
6-19
Page 22
PMC PROCEDURE (continued)
NM MOBILE/MILITARY
6.8.1.7Is the pressure on the manometer within 49 to 51 cm H
(Y)
6.8.1.8Is the pressure displayed on the Breathing Pressure Gauge +/- 3
2O of the value displayed on the test manometer?___(Y)
cm H
(
ü)6.8.1.9After thirty (30) seconds, is the pressure displayed on the
manometer equal to or greater than 40 cm H
2O? ___(Y)
6.8.1.10 If applicable, turn on the vaporizer to the first graduated
marking. Reapply 50 cm H
2O of pressure to the system and
start a stopwatch.
ü)6.8.1.11 After thirty (30) seconds, is the vaporizer test pressure
(
displayed on the manometer equal to or greater than 40 cm
2O? ___(Y). Turn off the vaporizer.
H
(
ü)6.8.2APL Valve
6.8.2.1Open the APL valve to its stop.
6.8.2.2Verify Scavenger is active.
2O? ___
6.8.2.3Turn the System Power switch to ON.
6.8.2.4Set the Oxygen flow to 8 l/min.
6.8.2.5Is the test pressure within 0 to 3 cm H
6.8.2.6Remove Freshgas Leak Test Adapter (P/N 4115041), Test
ü)6.8.3O2 Flush
(
6.8.3.1Attach a 33 mm x 22 Female Adapter (P/N 4115087) to the top
6.8.3.2Remove the ventilator hose from the absorber and bellows, and
6.8.3.3Connect test volumeter to the end of ventilator hose.
6.8.3.4Close the APL valve.
6.8.3.5Press and hold the O
2O? ___ (Y)
Terminal Adapter (P/N 4104389), and Digital Pressure
Manometer (P/N PDM 200CD) from the bag mount.
port of the test volumeter.
connect to the Bag Mount terminal.
2 FLUSH button for 6 seconds; multiply
the value obtained by 10.
6-20
Page 23
NM MOBILE/MILITARY
6.8.3.6Is the calculated Oxygen flush flow rate within 45 to 65 l/min.?
___ (Y)
6.8.3.7After releasing the flush, does the flow of Oxygen stop
immediately? ___ (Y)
6.8.3.8Reattach ventilator hose between absorber and bellows
ventilator terminals.
6.8.3.9Remove all test equipment. Remove 22mm hose from
inspiratory valve and expiratory port on the Ultrasonic Flow
Sensor or the expiratory valve, as applicable.
ü)6.8.4Expiratory Valve Leak
(
6.8.4.1Connect a 22 mm hose (P/N 9995132) between the Inspiratory
valve and the bag mount.
6.8.4.2Connect a test terminal (P/N 4104389) to the expiratory port on
the Ultrasonic Flow Sensor.
PMC PROCEDURE (continued)
6.8.4.3Connect a 0-250 ml/min. flowmeter (P/N S000081) to the test
terminal.
6.8.4.4Turn up the Oxygen flow until the system pressurizes to 30 cm
2O. Adjust the APL valve as necessary to maintain 30 cm
H
2O.
H
6.8.4.5Is the value indicated on the flowmeter within 0 to 60 ml/min.?
___ (Y)
6.8.4.6Turn off O
2 flow. Fully close APL valve.
6.8.4.7Remove all test equipment.
(
ü)6.8.5Inspiratory Valve Leak
6.8.5.1Turn the System Power switch to Standby.
6.8.5.2Connect a test terminal (P/N 4104389) to the Inspiratory valve.
6.8.5.3Connect a Fresh Gas Leak Adapter (P/N 4115041) and
manometer to the test terminal on the Inspiratory valve.
6.8.5.4Connect another test terminal to the bag mount.
6.8.5.5Connect a 0-250 ml/min. flowmeter (S000081) to the test
terminal on the bag mount.
6-21
Page 24
PMC PROCEDURE (continued)
NM MOBILE/MILITARY
6.8.5.6Pressurize the test circuit to 30 cm H
6.8.5.7Is the value indicated on the flowmeter within 0 to 60 ml/min.?
___ (Y)
6.8.5.8Turn the system power switch to ON.
6.8.5.9Remove all test equipment.
(
ü)6.8.6Absorber PEEP Valve
6.8.6.1Open the APL valve.
6.8.6.2Disconnect the pressure line going to the Breathing Pressure
gauge and connect to a test manometer.
6.8.6.3Connect a Fresh Gas Outlet Volume Test Device (S010158) to
the Fresh gas outlet.
6.8.6.4Install a test breathing circuit or disposable breathing circuit
onto the Inspiratory and Expiratory valves of the absorber.
6.8.6.5Connect the test terminal, of the Fresh Gas Outlet Volume Test
Device, to the Y-piece of the breathing circuit.
2O.
6.8.6.6Set the O
6.8.6.7Adjust the absorber PEEP valve clockwise to the maximum
position.
6.8.6.8Does the PEEP valve adjust smoothly? ___ (Y)
6.8.6.9Is the maximum PEEP indicated on the test gauge greater than
15 cm H
6.8.6.10 Adjust the absorber PEEP valve counterclockwise to its
minimum position.
6.8.6.11 Is the PEEP indicated on the test gauge less than or equal to 3
cm H
6.8.6.12 Close the O
6.8.6.13 Remove the test equipment and reconnect the pressure line to
the Breathing Pressure gauge.
6.9 OXYGEN ANALYZER
2 flow to 5 l/min.
2O? ___ (Y)
2O? ___(Y)
2 flow control valve.
6.9.1Press the Cal key.
6-22
Page 25
NM MOBILE/MILITARY
PMC PROCEDURE (continued)
NOTE: Ensure that the sensor has stabilized in ambient air for several
minutes.
ü)6.9.2After calibration is completed, is the O2 concentration 21%? ___ (Y)
(
6.9.3The warning message INSP O
2 LOW shall appear on the central alarm
display, and an audible alarm shall sound.
6.9.4Press the ALARM SILENCE key and verify the audio alarm is silenced.
6.9.5Remove the Breathing Pressure Hose Assembly from the Inspiratory
valve dome.
6.9.6Place the Oxygen Sensor housing directly into the Inspiratory valve
dome.
6.9.7Verify the AUTO/BAG selector is set to BAG.
6.9.8Close the APL valve.
6.9.9Attach a 22 mm hose (P/N 9995132) to the Inspiratory valve.
6.9.10Attach a Breathing System Leak Test Device (P/N S010159) to the bag
mount.
6.9.11Press the O
6.9.12After 20 seconds, is the O
2 Flush.
2 concentration within 90 to 100 %? ___ (Y)
6.9.13Release the O
2 Flush, does the flow cease immediately? ___ (Y)
6.9.14Set the Oxygen flow to 8 l/min.
(
ü)6.9.15After 1 minute, is the O2 concentration within 97 to 100 %? ___ (Y)
6.10 FLOWMETERS & CONCENTRATIONS
ü)6.10.1Oxygen Flowmeter
(
6.10.1.1 Is it possible to adjust the flow of Oxygen over the full range of
the flowmeters? ___ (Y)
6.10.1.2 Set the Oxygen flow to 4 l/min.
6.10.1.3 Is the correct flow control knob and label attached to the
Oxygen flow control valve? ___ (Y)
6-23
Page 26
PMC PROCEDURE (continued)
(
ü)6.10.2Air Flowmeter
6.10.2.1 If not configured with an Air Cylinder yoke, attach the Air
Pipeline hose.
6.10.2.2 Is it possible to adjust the flow of the Air over the full range of
the flowmeter? ___ (Y)
6.10.2.3 Set the Air flow to 2 l/min.
NM MOBILE/MILITARY
6.10.2.4 After the value stabilizes, is the O
77 %? ___ (Y)
6.10.2.5 Close the Air flow control valve.
6.10.2.6 Is the correct flow control knob and label attached to the Air
flow control valve? ___ (Y)
6.10.3Nitrous Oxide Flowmeter
6.10.3.1 Set the Nitrous Oxide flow to 2 l/min.
6.10.3.2 After the value stabilizes, is the O
70 %? ___ (Y)
6.10.3.3 Is the correct flow control knob and label attached to the
Nitrous Oxide flow control valve? ___ (Y)
6.10.3.4 Is it possible to adjust the flow of nitrous oxide over the full
range of the flowmeter? ___ (Y)
ü)6.10.4Oxygen Ratio Control
(
2 concentration within 71 to
2 concentration within 64 to
6.10.4.1 Open the Nitrous Oxide flow control valve to the stop position.
6.10.4.2 After the value stabilizes, is the O
2 concentration within 21 to
29 %? ___ (Y)
6.10.4.3 Set the Oxygen flow to 2 l/min.
6.10.4.4 After the value stabilizes, is the O
2 concentration within 21 to
29 %? ___ (Y)
6.10.4.5 Set the Oxygen flow to 1 l/min.
6.10.4.6 After the value stabilizes, is the O
2 concentration within 21 to
29 %? ___ (Y)
6.10.4.7 Close the Oxygen flow control valve.
6-24
Page 27
NM MOBILE/MILITARY
6.10.4.8 Close the Nitrous Oxide flow control valve.
ü)6.10.5Auxiliary Oxygen Flowmeter
(
6.10.5.1 Connect a test manometer to the Auxiliary Oxygen outlet using
a PDM/Suction adapter (P/N 4115038).
PMC PROCEDURE (continued)
6.10.5.2 Slowly open flowmeter valve until a pressure of 50 cm H
2O is
achieved.
6.10.5.3 Close flowmeter valve.
6.10.5.4 After 10 seconds, is the pressure within 40 to 60 cm H
2O? ___
(Y)
6.10.5.5 Remove the test manometer and adapter.
6.10.5.6 Is it possible to adjust the Auxiliary Oxygen flowmeter over the
full range of the flowmeter? ___(Y)
6.10.5.7 Set the Auxiliary O
2 flow rate to 5 L/min.
6.10.5.8 Hold the Oxygen sensor at the Auxiliary Oxygen flowmeter
outlet.
6.10.5.9 After the value stabilizes, is the O
2 concentration within 80 to
100 %? ___ (Y)
6.10.5.10Reinstall the Breathing Pressure Hose Assembly.
6.10.5.11Insert the Oxygen sensor into the Breathing Pressure Hose
Assembly.
6.10.5.12Close the Auxiliary Oxygen flow control valve.
6.11 HIGH PRESSURE REGULATORS – Six Month Service Interval; Due Date ______
ü)6.11.1N2O Regulator- if applicable
(
NOTE: Minimum cylinder pressure for N2O regulator test is 600 psi.
6.11.1.1 Configure the test gauge (P/N 4114807) using a N2O nut/stem
DISS connector (P/N 4114830-004) on the hose, and N2O DISS
body connector (P/N 4114830-003) on the valve body side. If the
machine is configured with CSA style fittings reverse the
position of the connectors.
6.11.1.2 Connect the test fixture hose to the machine’s Nitrous Oxide
pipeline inlet.
6-25
Page 28
PMC PROCEDURE (continued)
6.11.1.3 Connect the Nitrous Oxide pipeline supply hose to the test
fixture.
6.11.1.4 Open the Nitrous Oxide and the Oxygen cylinder valves.
6.11.1.5 Set the Oxygen and Nitrous Oxide flows to 4 l/min.
6.11.1.6 Depress the push button on the test device.
6.11.1.7 Release the push button. After the pressure decay stabilizes, is
the regulator output pressure within 40 to 49 psi? ___ (Y)
6.11.1.8 Remove the test fixture.
NOTE: If a pressure decrease does not occur, either the hospital’s
supply pressure is too low or the regulator pressure is set too
high.
(
ü)6.11.2O2 Regulator
NM MOBILE/MILITARY
NOTE: Minimum cylinder pressure for O
6.11.2.1 Configure a test gauge (P/N 4114807) using an O
DISS connector (P/N 4114830-006) on the hose and an O
2 regulator test is 1000 psi.
2 nut/stem
2 DISS
body connector (P/N 4114830-005) on the valve body side. If the
machine is configured with CSA style fittings reverse the
position of the connectors.
6.11.2.2 Connect the test fixture hose to the machine’s Oxygen pipeline
inlet.
6.11.2.3 Connect the Oxygen pipeline supply hose to the test fixture.
6.11.2.4 Set the Oxygen flow to 4 l/min.
6.11.2.5 Depress the push button on the test device.
6-26
Page 29
NM MOBILE/MILITARY
6.11.2.6 Release the push button. After the pressure decay stabilizes, is
NOTE: If a pressure decrease does not occur, either the hospital’s
PMC PROCEDURE (continued)
the regulator output pressure within tolerance given in the
following table? ___ (Y)
supply pressure is too low or the regulator pressure is set too
high.
Cylinder Pressure
psi
200038 to 4441 to 47
180039 to 4542 to 48
160040 to 46 43 to 49
140041 to 4744 to 50
120042 to 4845 to 51
100043 to 4946 to 52
ü)6.12 LOW O2 SUPPLY – Six Month Service Interval; Due Date ______
(
USA Compensated Regulator
output tolerances
CSA Compensated Regula-
tor output tolerances
6.12.0.1 Close the Oxygen Cylinder valve and drain Oxygen cylinder
pressure.
6.12.0.2 Depress the push button on the test device.
6.12.0.3 Adjust the Oxygen flow to 500 ml/min.
6.12.0.4 Release the test device push button.
6.12.0.5 Is the pressure on the test gauge when the LO O
2 SUPPLY
message appears on the monitor within 34 to 40 psi? ___ (Y)
6.12.0.6 Close O
2 Flow control valve.
6.12.0.7 Remove the test equipment.
6.13 OXYGEN SUPPLY FAILURE PROTECTION
6.13.1Connect all pipeline supplies.
(
ü)6.13.2*Is the flow of Oxygen 150 to 200 ml/min.? ___ (Y)
6.13.3Open the Nitrous Oxide flow control valve.
(
ü)6.13.4*Is the flow of Nitrous Oxide within 375-750 ml/min.? ___ (Y)
6-27
Page 30
PMC PROCEDURE (continued)
NM MOBILE/MILITARY
6.13.5Adjust the Oxygen, Nitrous Oxide and AIR gases to 4 l/min.
6.13.6Disconnect the Oxygen pipeline supply and close the Oxygen cylinder
valve.
(
ü)6.13.7Do all fresh gas flows cease when the Oxygen pressure is depleted?
___(Y)
6.13.8Reconnect the Oxygen pipeline supply.
6.13.9Close all cylinder valves and then disconnect the Nitrous Oxide pipeline
supply.
6.13.10 Drain the cylinder contents then reconnect the pipeline supplies.
6.13.11 Close all flow control valves.
* Nitrous Oxide Bypass flow and Minimum Oxygen flow specifications
are given @ 50 psi. Pipeline pressure deviations may affect these tests.
6.14 PRESSURE MONITOR
6.14.1Disconnect the Breathing Pressure Sensor line from the Breathing
Pressure interface on the machine.
6.14.2Connect a PDM to Monitor Adapter (P/N 4115043) and test manometer
to the Breathing Pressure interface on the machine.
6.14.3Adjust the test pressure to 0 cm H
2O.
6.14.4Simultaneously turn “ON” the ventilator and start a stopwatch.
(
ü)6.14.5Does the APNEA PRESSURE alarm appear on the alarm display as a
CAUTION within 13 to 17 seconds? ___ (Y)
ü)6.14.6Increase the test pressure slowly. Does the APNEA PRESSURE alarm
(
deactivate within 10 to 14 cm H
2O? ___ (Y)
6.14.7First decrease the pressure then increase the test pressure above the
threshold line shown on the display, and begin timing with a stopwatch.
(
ü)6.14.8Does the CONTINUOUS PRES alarm appear as a warning within 13 to
17 seconds? ___ (Y)
ü)6.14.9Decrease the pressure slowly. Does the CONTINUOUS PRES alarm
(
deactivate within 10 to 14 cm H
(
ü)6.14.10 Increase the test pressure slowly. Does a VENT PRESS HI alarm
activate as a warning alarm within 47 to 53 cm H
2O? ___ (Y)
2O? ___ (Y)
6-28
Page 31
NM MOBILE/MILITARY
ü)6.14.11 Increase the sub-atmospheric test pressure slowly. Does the SUB ATM
(
PRES warning alarm activate within –7 to –13 cm H
PMC PROCEDURE (continued)
2O? ___ (Y)
6.14.12 Turn the ventilator “OFF”.
6.14.13 Open APL valve.
6.14.14 Remove the test equipment and reconnect the Breathing Pressure
Sensor line to the Breathing Pressure interface on the machine.
6.15 VENTILATOR
NOTE: Readjustment of inspiratory flow to limit the inspiratory plateau may be
required to reduce erratic tidal volumes and breath rates caused by
artifact volumes.
6.15.1Remove the bellows vent hose and the scavenger hose at the ventilator
relief valve. Remove the bellows sub-assembly and remove bellows.
6.15.2Visually inspect the bellows for deterioration particularly at it’s seems
and corrugations. Replace as necessary.
6.15.3Verify the presence and condition of it’s sealing O-ring. Replace as
necessary. Reassemble all components.
6.15.4Set O
2/AIR mode switch to “O2”.
6.15.5Turn the ventilator “ON”.
6.15.6Set the FREQUENCY to 10 BPM.
(
ü)6.15.7Press and hold the EXTENDED RANGE switch and set the I:E ratio to
2:1. Using a stopwatch, time the extended I:E ratio. Is the inspiratory
time within 3.6 to 4.4 seconds and the expiratory time within 1.8 to 2.2
seconds? ___ (Y)
ü)6.15.8Set the I:E RATIO to 1:2. Using a stopwatch, time the I:E ratio. Is the
(
inspiratory time within 1.8 to 2.2 seconds and the expiratory time within
3.6 to 4.4 seconds? ___ (Y)
6.15.9Adjust the Oxygen flow to 500 ml/min.
6.15.10 Set the Tidal Volume to 1200.
6.15.11 Attach a patient circuit to the absorber system.
6.15.12 Set the pressure limit control to MAX.
6.15.13 Adjust the Inspiratory Flow to the bottom of the LOW zone.
6-29
Page 32
PMC PROCEDURE (continued)
6.15.14 Occlude the Y-piece.
NM MOBILE/MILITARY
6.15.15 Press the O
6.15.16 Adjust the Inspiratory Flow until a peak pressure of 80 cm H
2 Flush momentarily to inflate the bellows.
2O is
achieved.
ü)6.15.17 Set the Pressure Limit Control to within the 30 range. Readjust within
(
the band as necessary to achieve proper value, is the peak pressure at
the 30 range within 27 to 33 cm H
(
ü)6.15.18 Set the Pressure Limit Control to the MIN position. Is the peak pressure
at the MIN range within 0 to 15 cm H
2O? ___ (Y)
2O? ___ (Y) Return the Pressure
Limit control to MAX.
6.15.19 On the Military machine, loosen the expiratory valve dome. On the
Mobile machine, disconnect the hose between the expiratory valve and
the ultrasonic flow sensor and blow into it.
ü)6.15.20 Does the Reverse Flow message appear on the display? ___ (Y)
(
6.15.21 Reconnect the hose between the expiratory valve and the flow sensor.
6.15.22 On the Military machine, insert a test minute volumeter between the
Spiromed and the absorber dome. On the Mobile machine, insert a test
minute volumeter between the exhalation valve and PEEP valve.
6.15.23 Open the Y-piece.
6.15.24 Turn the ventilator off and press the ALL STBY key to clear alarms.
6.15.25 Turn the ventilator “ON” and start a stop watch.
6.15.26 Does APNEA-VOLUME appear as a Caution within 13 to 17 seconds?
___ (Y)
6.15.27 Attach a 3 liter breathing bag to the Y-piece.
NOTE: Bag should be placed on a flat horizontal surface to reduce artifact
volume.
6.15.28 Press the O
2 Flush momentarily to inflate the bellows.
6.15.29 Set the Inspiratory Flow to the MED and readjust as necessary to fully
collapse the bellows.
(
ü)6.15.30 Observe the operation of each unidirectional valve disc at eye level and
make sure the inspiratory valve disc raises only during the inspiration
phase, and the expiratory valve raises only during the exhalation phase.
6-30
Page 33
NM MOBILE/MILITARY
ü)6.15.31 Is the tidal volume on the volume monitor and on the test volumeter
(
PMC PROCEDURE (continued)
within 20 % of each other? ___ (Y)
6.15.32 Does the volume monitor display 10 BPM? ___ (Y)
6.15.33 Does the display correctly track the Breathing Pressure waveform? ___
(Y)
6.15.34 Verify AIR pipeline hose is connected.
(
ü)6.15.35 Set O2/AIR mode switch to “AIR”. Does ventilator continue to cycle
properly. ___(Y)
6.15.36 Set O
2/AIR mode switch back to “O2”.
6.16 BELLOWS ADULT
(
ü)6.16.1What is the tidal volume indicated on the test volumeter within 960 to
1440 ml? ___ (Y)
ü)6.16.2Does the bellows remain fully inflated during the expiratory pause
(
phase? ___ (Y)
6.16.3Remove the ventilator hose from the bellows hose terminal.
6.16.4Attach a test terminal to the bellows hose terminal.
6.16.5Connect a 0-250 ml/min. flowmeter (P/N S000081) to the test terminal.
6.16.6Set the FREQUENCY to 1 BPM.
(
ü)6.16.7Is the drive gas leakage indicated during the inspiratory phase within 0
to 50 ml/min.? ___ (Y)
6.16.8Remove the test equipment from the ventilator hose terminal and
reconnect the ventilator hose to the bellows ventilator hose terminal.
6.16.9Set the FREQUENCY to 10 BPM.
6.16.10 Adjust the O
2 flow to 8 l/min.
6.16.11 Adjust the Tidal Volume to maximum.
6.16.12 Press the O
2 Flush momentarily to inflate the bellows.
6.16.13 Adjust the INSPIRATORY FLOW to fully compress the bellows.
(
ü)6.16.14 Is the Tidal Volume on the test volumeter greater than or equal to 1400
ml? ___(Y)
6-31
Page 34
PMC PROCEDURE (continued)
(
ü)6.16.15 Is the PEEP value displayed on the monitor within 0 to 3 cm H2O? ___
NM MOBILE/MILITARY
(Y)
6.16.16 Remove the breathing bag from the Y-piece and replace it with a test
lung.
6.16.17 Adjust the O
2 flow to 300 ml/min.
6.16.18 Adjust the Tidal Volume to 200 ml.
6.16.19 Does the bellows stop adjust smoothly and engage properly? ___ (Y)
6.16.20 Adjust the INSPIRATORY FLOW to fully compress the bellows.
(
ü)6.16.21 Is the Tidal Volume on the test volumeter within 125 to 250 ml? ___ (Y)
6.16.22 Close the Oxygen flow control valve.
6.16.23 Remove the test lung and set the AUTO/BAG selector to BAG.
6.16.24 Press the Alarms All Standby key.
6.17 SCAVENGER- Six Month Service Interval; Due Date ______
NOTE: If the ambient air in the local environment contains a significant amount of
dust and lint, the cleaning frequency must be increased to compensate for
these conditions.
(
ü)6.17.1Scavenger - Passive Mode
6.17.1.1 Remove the scavenger hoses and drain all accumulated
moisture. Inspect all scavenger hoses for deterioration and
replace as necessary. Reinstall hoses and bag.
6.17.1.2 Ensure that the suction needle valve (not used in passive mode)
is closed.
6.17.1.3 Positive Pressure Test:
6.17.1.3.1 Connect a 19mm scavenger hose between the Vent
Relief valve and the top port of the scavenger.
Connect another 19 mm scavenger hose between
APL valve and the second scavenger port.
6.17.1.3.2 Interconnect the inspiratory valve and expiratory
valve or expiratory port on the ultrasonic flow sensor,
if applicable with a 22 mm hose (P/N 9995132).
6.17.1.3.3 Set Man/Auto valve to “Auto”.
6-32
Page 35
PMC PROCEDURE (continued)
6.17.1.3.4 Verify PEEP valve is at minimum (counter-
NM MOBILE/MILITARY
clockwise).
6.17.1.3.5 Set O
2 Flow to 8 l/min.
6.17.1.3.6 Occlude the bottom 19 mm connector (EXHAUST).
6.17.1.3.7 After the bellows inflates, the flow of oxygen will exit
the system through the positive pressure safety relief
valve.
6.17.1.3.8 Is the pressure displayed on absorber cm H
is less than or equal to 5 cm H
(
ü)6.17.2Scavenger - Suction Mode
6.17.2.1 Remove the scavenger hoses and drain all accumulated
moisture. Inspect all scavenger hoses and scavenger bag for
deterioration and replace as necessary. Reinstall hoses and bag.
6.17.2.2 Remove the filter from the negative pressure relief valve. Clean
and reinstall the filter.
6.17.2.3 Negative Pressure Test:
6.17.2.3.1 Interconnect the inspiratory valve and expiratory
valve or expiratory port on the ultrasonic flow sensor,
if applicable with a 22 mm hose (P/N 9995132).
2O gauge
2O?___(Y)
6.17.2.3.2 Set Man/Auto valve to “BAG” position
6.17.2.3.3 Fully open APL valve.
6.17.2.3.4 Attach a Breathing System Leak Test Device (P/N
S010159) to the bag mount (occlude bag mount).
6.17.2.3.5 Verify Suction system is active. Adjust needle valve
to allow typical suction through scavenger.
6.17.2.3.6 Verify all flow control valves are closed.
6.17.2.3.7 Install scavenger adapter (P/N 4108114) between
either of the 19 mm scavenger connectors and
scavenger hose. Connect a manometer to the hose
barb on the scavenger adapter and observe reading
on manometer. The manometer shall indicate a
pressure less than or equal to 0.5 cm H
2O.
6-33
Page 36
PMC PROCEDURE (continued)
6.17.2.4 Positive Pressure Test:
6.17.2.4.1 Fully close scavenger needle valve.
NM MOBILE/MILITARY
6.17.2.4.2 Push the O
2 flush to inflate scavenger bag. Open
oxygen flow control valve to 8 l/min.
6.17.2.4.3 Verify reading on manometer is less than or equal to
5 cm H
2O.
6.17.2.4.4 Remove all test equipment. Adjust scavenger needle
valve to allow typical suction through scavenger.
6.18 FINAL TESTS
(
ü)6.18.1Is the machine's Operator's Instruction manual in close proximity of the
machine? ___ (Y)
6.18.2Verify all cylinder pressure gauges indicate zero.
6.18.3Verify the pipeline hoses are connected to the hospital pipeline.
6.18.4Verify the APL valve knob is turned completely counterclockwise (fully
open).
6.18.5Place the AUTO/BAG selector in the BAG position.
6.18.6Verify the ventilator hose is connected between the AUTO/BAG valve
and bellows ventilator hose terminal.
6.18.7Verify the Breathing Pressure Sensor line is connected to the Breathing
Pressure interface on the machine.
6.18.8Verify the Oxygen sensor is removed from the inspiratory valve dome.
6.18.9Verify that the Inspiratory valve dome is plugged.
6.18.10 Verify that the machine is plugged into a live A/C receptacle.
6-34
Page 37
Page 38
DrägerService is a division of
Draeger Medical, Inc.
3122 Commerce Drive
Telford, PA 18969
Tel: (215) 721-5402
(800) 543-5047
Fax: (215) 721-5784
Web: www.draegermedical.com
Printed in the U.S.A.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.