Replacement of Advantech Hardware with
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Replacement of Advantech Hardware with 4SYS
Medical Grade Workstation
Replacement Procedure
This procedure includes instructions for the replacement of Advantech Hardware with 4SYS Medical Grade Workstation.
Installing the 4SYS Medical Grade Hardware
1. Remove AC power from the Advantech Hardware.
WARNING: Before servicing, ensure that
AC power is removed from the WPU. Failure to observe this precaution may cause injury by electrical shock.
2. Unplug all cables from the rear of the Advantech Hardware. Take note of removed locations for future reference.
3. Utilizing a #2 Phillips Head screwdriver, remove and retain 3 screws (1) attaching the tilt clamp assembly (2) to the patient arm (3). See Figure 1.
4. Remove and retain 4 Phillips Head screws (4) attaching tilt clamp assembly (5) to L bracket (6). See Figure 2.
5. Remove 3 Phillips Head screws (7) attaching the Patient Arm attachment plate (8) to the tilt clamp assembly (5). Discard items (7) and (8). See Figure 2.
6. Utilizing a T-30 Torx Wrench, loosen (do not remove) two screws (9) located on the back of the tilt clamp. See Figure 3.
7. Flip the tilt clamp (10) from vesa bracket (11) to reposition vesa bracket, ensuring remouting of bracket assembly will allow the open slots (12) on the vesa bracket to be located at the top of assembly. See Figure 3 for correct installation positioning of tilt clamp and vesa bracket.
8. Using 3 screws (13) previously removed in step 3, install tilt clamp assembly (14) onto Patient Arm (15).
NOTE: When mounting the tilt clamp
assembly to the patient arm, ensure the open slots (12) on the vesa bracket portion of the tilt clamp assembly are located at the top.
CAUTION: The 4SYS Medical Grade
Workstation is extremely delicate. Use extreme care when handling the Workstation.
9. Using the 4 screws removed in step 4, (16) install the 4SYS Medical grade workstation (17) onto the tilt clamp assembly (18). See Figure 4.
10. Disconnect all cables routed from the Isolation Transformer and the Uninterrupted Power Supply located n the right side of the machine.
1