Replacement of Advantech Hardware with
4SYS Medical Grade Workstation

RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS

RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Replacement of Advantech Hardware with 4SYS
Medical Grade Workstation
Replacement Procedure
This procedure includes instructions for the replacement of Advantech Hardware with 4SYS
Medical Grade Workstation.
Installing the 4SYS Medical Grade
Hardware
1. Remove AC power from the Advantech
Hardware.
WARNING: Before servicing, ensure that
AC power is removed from the
WPU. Failure to observe this
precaution may cause injury
by electrical shock.
2. Unplug all cables from the rear of the
Advantech Hardware. Take note of
removed locations for future reference.
3. Utilizing a #2 Phillips Head
screwdriver, remove and retain 3 screws
(1) attaching the tilt clamp assembly (2)
to the patient arm (3). See Figure 1.
4. Remove and retain 4 Phillips Head
screws (4) attaching tilt clamp assembly
(5) to L bracket (6). See Figure 2.
5. Remove 3 Phillips Head screws (7)
attaching the Patient Arm attachment
plate (8) to the tilt clamp assembly (5).
Discard items (7) and (8). See Figure 2.
6. Utilizing a T-30 Torx Wrench, loosen (do
not remove) two screws (9) located on
the back of the tilt clamp. See Figure 3.
7. Flip the tilt clamp (10) from vesa
bracket (11) to reposition vesa bracket,
ensuring remouting of bracket assembly
will allow the open slots (12) on the vesa
bracket to be located at the top of
assembly. See Figure 3 for correct
installation positioning of tilt clamp and
vesa bracket.
8. Using 3 screws (13) previously removed
in step 3, install tilt clamp assembly
(14) onto Patient Arm (15).
NOTE: When mounting the tilt clamp
assembly to the patient arm,
ensure the open slots (12) on
the vesa bracket portion of the
tilt clamp assembly are located
at the top.
CAUTION: The 4SYS Medical Grade
Workstation is extremely
delicate. Use extreme care
when handling the
Workstation.
9. Using the 4 screws removed in step 4,
(16) install the 4SYS Medical grade
workstation (17) onto the tilt clamp
assembly (18). See Figure 4.
10. Disconnect all cables routed from the
Isolation Transformer and the
Uninterrupted Power Supply located n
the right side of the machine.
1

Replacement Procedure
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
1
3
Figure 1. Tilt Clamp Assembly Removal
2
2

Replacement Procedure
8
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
4
7
5
6
Figure 2. Tilt Clamp Assembly Modification
3

Replacement Procedure
9
15
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
12
11
14
10
13
Figure 3. Tilt Clamp Assembly and Vesa Bracket Installation
4

Replacement Procedure
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
17
16
18
Figure 4. 4SYS Medical Grade Workstation Installation
5

Replacement Procedure
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
11. Remove and discard the Isolation
Transformer and the Uniterrupted
Power Supply by sliding them off the
top of the GCX rail.
KEYBOARD AND MONITOR
SIDE VIEW
12. Reconnect all cables to the rear of the
4SYS Medical Grade Workstation. Refer
to Figure 5 for labeled connection
diagram.
13. Proceed with application install. For
additional support, contact Saturn
Support.
KEYBOARD
NETWORK
CONNECTION
AC POWER
POWER
BUTTON
BACK OF WORKSTATION
Figure 5. 4SYS Medical Grade Workstation Cable Connections
COM 1
COM 2
COM 3
COM 4
SU00073
6

RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS

RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DrägerService is a division of
Draeger Medical, Inc.
3122 Commerce Drive
Telford, PA 18969
Tel: (215) 721-5402
(800) 543-5047
Fax:(215) 721-5784
Web:
www.draegermedical.com
Printed in the U.S.A.