NORTH AMERICAN DRÄGER MicroVent User manual

Dräger Medizintechnik
D
Contents
1 General Information About the Dräger Service Mode 3
1.1 Definition of the Original Status ...................................................................................3
1.3 Reaction in the Event of a Technical Error (TE) ........................................................4
1.4 Error Messages During a Specific Test ......................................................................5
1.4.1 Power Supply Low .........................................................................................5
1.4.2 Pressure Low ..................................................................................................5
1.4.3 Pressure High .................................................................................................5
1.4.4 Mode Switch Position (IPPV/CPAP) .........................................................5
1.4.5 Mode Switch Position (IPPV/CPAP) .........................................................5
1.5 Tests in the Dräger Service Mode ...............................................................................6
2 Starting / Exiting the Customer Service Mode (CSM) 7
2.1 Test 001: Adjusting the Language ...............................................................................8
2.2 Test 002: Testing the Gas Supply ............................................................................. 10
2.3 Test 003: Position of the Pneumatic Switch (IPPV/CPAP) ..................................11
2.4 Test 004: Setting the Minute Volume Limits ............................................................12
2.5 Test 005: Buzzer Test, LED Test, and LCD Test .................................................... 13
2.6 Test 006: Adjusting and Reading Out the
Potentiometers f, VT, Tinsp, and Pmax 14
2.7 Test 007: Tachypnea Monitoring Configuration ..................................................... 15
3 Starting / Exiting the Extended Service Mode (ESM) 16
3.1 Test 011: Calibrating the Pressure Sensors P-Supply and P-LA ........................17
3.2 Test 012: Calibrating the Minute Volume Valve ...................................................... 18
3.3 Test 013: Adjusting and Displaying the Flow ......................................................... 20
3.4 Test 014: Reading in the Analog Channels ............................................................ 22
3.5 Test 015: Testing Basic Flow ..................................................................................... 26
For internal use only. Copyright reserved.
GBR5673100SM1IVZ.fm
5673.100 MicroVent 09/97 Service Mode Page I
Dräger Medizintechnik
Contents
3.6 Test 016: Testing the Autozero Solenoid Valve ...................................................... 28
3.7 Test 017: Displaying and Calibrating the Autozero Offset .................................. 29
3.8 Test 018: Testing the Pressure Sensor Paw and
Triggering the Minute Volume Valve 30
3.9 Test 019: Testing the Expiratory Solenoid Valve .................................................... 33
3.10 Test 020: Testing the ASB Solenoid Valve ............................................................. 34
3.11 Test 021: Testing the Supply Pressure .................................................................... 35
3.12 Test 022: Testing the Minute Volume Valve (Slow) ............................................... 36
3.13 Test 023: Testing the Minute Volume Valve (Fast) ................................................. 37
D
3.14 Test 024: Testing the Differential Pressure Sensor P and
Testing the Flow Measurement 38
3.15 Test 025: Testing the Pressure Switch PEEP-ASB .............................................. 40
4 Technical Error (TE) List 41
For internal use only. Copyright reserved.
GBR5673100SM1IVZ.fm
5673.100 MicroVent 09/97 Service Mode Page II

1 General Information About the Dräger Service Mode

You can operate the Dräger Service Mode with the Info key and the Reset key. These keys allow you to change the set values, to scroll through tests, or to confirm execution of a specific test. The Dräger Service Mode is subdivided into the Customer Service Mode (CSM) and the Extended Service Mode (ESM). The working language used throughout the Dräger Service Mode is English.
The Dräger Service Mode in the MicroVent includes 20 tests:
• The first six tests (Test 001 through Test 006) are specifically designed for the customer (CSM), but can also be used by Dräger service technicians.
• The other fourteen tests (Test 011 through Test 024) are part of the Extended Service Mode (ESM) which is used to carry out calibrations. These tests are only accessible to Dräger service technicians. To do so, the Dräger service technician must enter a numeric code.

1.1 Definition of the Original Status

Each function in the Dräger Service Mode can be carried out separately. At the beginning of the Dräger Service Mode and after completion of a specific test, the MicroVent enters one of the following original status.
Table 1: Original Status of the Micro Vent in the Dräger Service Mode
I:E solenoid valve Off (expiration) PN solenoid valve Off (PEEP is inactive) Autozero solenoid valve Operated automatically Exp. solenoid valve Off (normal position = expiration) ASB solenoid valve Off (ASB is inactive) MV valve Closed Alarm LED Off Buzzer Off LCD backlight On Paw alarm Deactivate
For internal use only. Copyright reserved.
GBR5673100SM1.fm
R 5673.100 MicroVent 09/97
Service Mode
Page 3

1.2 Calibration Functions

The following table lists the four calibration functions available in the Dräger Service Mode. The data is stored in the EEPROM:
Table 2: Calibration Functions in the Dräger Service Mode
No. Function Mode Test No. 1 Adjusting the Language CSM/ESM 001 2 Calibrating the Pressure Sensors (Pvers, PLA) ESM 011 3 Calibrating the Minute Volume Valve ESM 012 4 Calibrating the Autozero Offset ESM 017
Caution:
After installing a new Control PCB, perform the calibration functions (Nos. 1 through 4) specified in the table above.
After installing a new Sensor PCB, perform Test 011 “Calibrating the Pressure Sensors P­Supply and P-LA”.

1.3 Reaction in the Event of a Technical Error (TE)

If the MicroVent detects a technical error, it will show the error number on the display. The technical error and the name of the current test are displayed alternately at 2-second intervals. In the event of a technical error, all tests can be carried out normally. The message on the display disappears as soon as the technical error has been corrected.
A new Control PCB will generate the error No. 16 (invalid check sum in the EEPROM) before initial calibration. This technical error will disappear as soon as all data has been calibrated.
For internal use only. Copyright reserved.
GBR5673100SM1.fm
R 5673.100 MicroVent 09/97
Service Mode
Page 4

1.4 Error Messages During a Specific Test

If the system detects and displays an error, you must confirm this error by pressing the Info key and the Reset key simultaneously. Once the error has been removed, the specific test can be selected and carried out again.
All potential errors and their cause are listed below:

1.4.1 Power Supply Low

Problem Accu/Bat. empty

1.4.2 Pressure Low

Problem Pressure low

1.4.3 Pressure High

Problem Pressure high
No external power supply connected. The internal power supply is no longer sufficient for safe operation.
Gas supply pressure is too low.
A gas supply source is connected although no pressure is required for this test.

1.4.4 Mode Switch Position (IPPV/CPAP)

Problem
The mode switch is not in the correct position.
Pos. IPPV/SIPPV

1.4.5 Mode Switch Position (IPPV/CPAP)

Problem
The mode switch is not in the correct position.
Pos. SIMV/CPAP
For internal use only. Copyright reserved.
GBR5673100SM1.fm
R 5673.100 MicroVent 09/97
Service Mode
Page 5

1.5 Tests in the Dräger Service Mode

Table 3: Service Mode Tests
Test number Test Name Mode 001 Adjust language CSM 002 Supply pressure CSM 003 Mode switch CSM 004 Adjust MV limits CSM 005 Buzzer & LED test CSM 006 Adjust f, V
, Ti, Pmax CSM
T
007 TACHYPNEA ALARM CSM 008 to 010 Not assigned 011 Calibration PSWI ESM 012 Calibration MV valve ESM 013 Flow set & measure ESM 014 Analog inputs ESM 015 Test basic flow ESM 016 Test autozero valve ESM 017 Autozero meas & cal. ESM 018 Set MV valve, Paw ESM 019 Test EXP valve ESM 020 Test ASB valve ESM 021 Test supply pressure ESM 022 Test MV valve slow ESM 023 Test MV valve fast ESM 024 Flow measurement ESM 025 Test pressure switch ESM 026-255 Not assigned
For internal use only. Copyright reserved.
GBR5673100SM1.fm
R 5673.100 MicroVent 09/97
Service Mode
Page 6

2 Starting / Exiting the Customer Service Mode (CSM)

Important:
In the Customer Service Mode (CSM), only Test 001 through Test 006 are accessible to the customer or Dräger service technician. The other tests are deactivated. Test 007 is only available in US software version.
• Turn the potentiometers for frequency f and for tidal volume VT clockwise as far as they will go.
• Press and hold the Info key and the Reset key simultaneously and switch on the MicroVent.
Display:
C u s t o m e r
Customer Service Mode (CSM)
Se r v i c e - M o d e
• Release the two keys.
Display:
A d j u s t l a n g u a g e
< – 0 0 1 + >
• Press the Reset key or the Info key to select Test 001 through Test 007.
• Press the Reset key and the Info key simultaneously to activate the respective test.
• Press the Reset key and the Info key simultaneously again to return to the Customer Service Mode (CSM).
Exiting the Customer Service Mode (CSM)
• Switch the MicroVent off and on again to return to the MicroVent normal operating mode.
For internal use only. Copyright reserved.
GBR5673100SM1.fm
R 5673.100 MicroVent 09/97
Service Mode
Page 7

2.1 Test 001: Adjusting the Language

This test allows you to adjust the language to be used during the MicroVent operation.
• Press the Reset key to select the desired language.
Display:
d e u t s c h
or
e n g l i s h
or
f r a n c a i s
or
e s p a n o l
or
s u o m e k s i
or
p o r t u g u e s e
For internal use only. Copyright reserved.
GBR5673100SM1.fm
R 5673.100 MicroVent 09/97
Service Mode
Page 8
or
i t a l i a n o
or
a m e r i c a n e n g l i s h
or
f r a n c a i s f r a n c e
or
n e d e r l a n d s
• Press the Info key and the Reset key simultaneously to adjust the desired language.
The MicroVent returns automatically to the Customer Service Mode (CSM).
Exiting the Customer Service Mode (CSM)
• Switch the MicroVent off and on again to return to the MicroVent normal operating mode.
For internal use only. Copyright reserved.
GBR5673100SM1.fm
R 5673.100 MicroVent 09/97
Service Mode
Page 9

2.2 Test 002: Testing the Gas Supply

The pressure sensors must have been calibrated (test 011 in Extended Service Mode) befre selecting test 002.
• Press the Reset key to select Test 002.
• Press the Info key and the Reset key simultaneously to activate Test 002.
Display:
P r e s s u r e h i g h
The gas supply pressure is sufficient.
or
P r e s s u r e l o w
The gas supply pressure is too low or the device is not connected to a gas supply source.
• Press the Info key and the Reset key simultaneously to return to the Customer Service Mode (CSM).
Exiting the Customer Service Mode (CSM)
• Switch the MicroVent off and on again to return to the MicroVent normal operating mode.
For internal use only. Copyright reserved.
GBR5673100SM1.fm
R 5673.100 MicroVent 09/97
Service Mode
Page 10

2.3 Test 003: Position of the Pneumatic Switch (IPPV/CPAP)

This test allows you to check position of the IPPV/CPAP mode switch is checked. The respective position is shown on the MicroVent display. The pressure sensors must have been calibrated (test 011 in Extended Service Mode) befre selecting test 003.
If the pressure in the device is too low, or if the device is not connected to a gas supply source, the warning “Problem pressure low” will appear. The problem must then be rectified (establish gas supply) and subsequently confirmed with the Info and Reset key before test 003 can be called up again.
• Press the Reset key to select Test 003.
• Press the Info key and the Reset key simultaneously to activate Test 003.
Display:
P o s. I P P V / S I P P V
or
P o s. S I M V / C P A P
• Press the Info key and the Reset key simultaneously to return to the Customer Service Mode (CSM).
Exiting the Customer Service Mode (CSM)
• Switch the MicroVent off and on again to return to the MicroVent normal operating mode.
For internal use only. Copyright reserved.
GBR5673100SM1.fm
R 5673.100 MicroVent 09/97
Service Mode
Page 11

2.4 Test 004: Setting the Minute Volume Limits

This test allows you to check the settings of the minute-volume double-potentiometer (plausibility test).
• Press the Reset key to select Test 004.
• Press the Info key and the Reset key simultaneously to activate Test 004.
Display:
M V – l o w = X X . X X
M V – h i g h = X X . X X
• Adjust the values listed in the table below using the potentiometer and check the values shown on the display.
Table 4: Minute Volume Alarm Limit Settings
Values Adjusted with the Potentiometer Values shown on the Display
MV - high 2 L/min 10 L/min 20 L/min
MV - low 2 L/min 10 L/min 20 L/min
1.9 to 2.1 L/min
9.5 to 10.5 L/min 19 to 21 L/min
1.9 to 2.1 L/min
9.5 to 10.5 L/min 19 to 21 L/min
• Press the Info key and the Reset key simultaneously to return to the Customer Service Mode (CSM).
Exiting the Customer Service Mode (CSM)
• Switch the MicroVent off and on again to return to the MicroVent normal operating mode.
For internal use only. Copyright reserved.
GBR5673100SM1.fm
R 5673.100 MicroVent 09/97
Service Mode
Page 12

2.5 Test 005: Buzzer Test, LED Test, and LCD Test

This test allows you to check the audible and visible signals. The LEDs of the internal/external power supply, the alarm LED and the backlight of the display come on. The buzzer is activated and the pixels of the display are triggered.
• Press the Reset key to select Test 005.
• Press the Info key and the Reset key simultaneously to activate Test 005.
Display:
I N F O T E S T O N
R E S E T T E S T O F F
• Press the Info key to start Test 005.
• Press the Reset key to stop Test 005.
• Press the Info key and the Reset key simultaneously to return to the Customer Service Mode (CSM).
Exiting the Customer Service Mode (CSM)
• Switch the MicroVent off and on again to return to the MicroVent normal operating mode.
For internal use only. Copyright reserved.
GBR5673100SM1.fm
R 5673.100 MicroVent 09/97
Service Mode
Page 13
Loading...
+ 29 hidden pages