NORTH AMERICAN DRÄGER Fabius Tiro User manual

RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Technical Service Manual
Fabius Tiro ® Anesthesia System
Emergency Care • OR/ Anesthesia • Critical Care • Home Care
Revision B 5/25/04
6020.002 4118302-002
Because you care
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Copyright by Dräger Medical AG & Co. KGaA, Lübeck, Germany.
No reproduction allowed for commercial purposes.
Read and understand the Instructions for Use/Operator’s Manual.
This Technical Documentation does not replace the Instructions for Use/Operator’s Manual.
The warranty and liability conditions of the general terms and conditions for business transactions of Dräger Medical AG & Co. KGaA are not extended by this Technical Documentation.
Observe all applicable technical laws and regulations.
Insofar as reference is made to laws, regulations or standards, these are based on the legal system of the Federal Republic of Germany. Observe the laws and regulations applicable in your country.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius Tiro Service Manual Table of Contents
What's New in Rev. B
DESCRIPTION
PAGE
General......................................................................................................................................... 1
Recommendations ............................................................................................................... 3
How To Use This Manual ..................................................................................................... 3
General Troubleshooting Guidelines.................................................................................... 3
Copyright.............................................................................................................................. 6
Trademark Notices...............................................................................................................6
Disclaimer ............................................................................................................................ 6
Function Description.................................................................................................................. 7
General Information about the Fabius Tiro .......................................................................... 9
Fabius Tiro Function Diagram............................................................................................ 14
Battery Backup................................................................................................................... 16
Fabius Tiro Piping Diagram................................................................................................ 17
Function Description of Gas Box........................................................................................ 19
SORC (Sensitive Oxygen Ratio Controller) ....................................................................... 20
Compact Breathing System, Cosy II .................................................................................. 22
Manual Ventilation..............................................................................................................25
Spontaneous Breathing...................................................................................................... 28
Volume/Pressure Mode Ventilation .................................................................................... 32
Cosy II Absorber ................................................................................................................37
Ventilator ............................................................................................................................ 37
High Pressure Safety Valve ............................................................................................... 39
Negative Pressure Relief Valve.......................................................................................... 40
Pneumatic System ............................................................................................................. 40
PEEP/Pmax Valve Control ................................................................................................. 41
APL Bypass Valve Control ................................................................................................. 41
Electronic Block Diagram................................................................................................... 43
Control PCB ....................................................................................................................... 44
Flowmeter .......................................................................................................................... 44
Patient Interface................................................................................................................. 44
Ventilator ............................................................................................................................ 44
Front Panel Functions........................................................................................................ 44
Pneumatic Assembly Control............................................................................................. 44
Serial Port Interface ...........................................................................................................44
Battery................................................................................................................................ 44
Control Panel Assembly..................................................................................................... 45
FiO2 Measurement ............................................................................................................ 47
Respiratory Flow Measurement ......................................................................................... 48
Gas Flow Rate Measurement ............................................................................................ 49
I
CONTENTS (continued)
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS TIRO
DESCRIPTION
PAGE
Vaporizer ............................................................................................................................ 50
ELECTROMAGNETIC TESTING AND COMPLIANCE ............................................................. 51
Electromagnetic Testing and Compliance.......................................................................... 53
Guidance and manufacturer’s declaration-electromagnetic emissions.............................. 53
Guidance and manufacturer’s declaration-electromagnetic immunity ............................... 54
Guidance and manufacturer’s declaration-electromagnetic immunity ............................... 55
Recommended separation distances between portable and mobile RF communications
equipment and the Fabius Tiro.................................................................................. 57
TROUBLESHOOTING GUIDE ................................................................................................... 61
Troubleshooting .................................................................................................................63
Power Supply and Voltage Distribution .............................................................................. 63
Battery................................................................................................................................ 68
Troubleshooting Guides..................................................................................................... 68
DIAGNOSTICS ........................................................................................................................... 81
Diagnostics ........................................................................................................................ 83
System Service Screen...................................................................................................... 87
Main Service Screen.......................................................................................................... 89
Service Log ........................................................................................................................ 91
Service Log Entries............................................................................................................ 91
Real Time Values ............................................................................................................... 97
Preventive Maintenance .................................................................................................. 101
Activate Preventive Maintenance Date ............................................................................ 101
General ............................................................................................................................ 103
Serial Number .................................................................................................................. 103
Reset Hours Run ............................................................................................................. 104
Reset Last Service Date .................................................................................................. 105
Select Market Kit.............................................................................................................. 106
Calibration........................................................................................................................ 109
Fresh Gas Flow Calibration.............................................................................................. 109
Pressure Calibration..........................................................................................................110
O2 Zero Calibration...........................................................................................................111
PEEP Valve Calibration.....................................................................................................112
Configure ..........................................................................................................................115
System Settings (US and Non-US)...................................................................................115
Model Type .......................................................................................................................116
Standard Options ..............................................................................................................117
Flowmeter .........................................................................................................................118
O2 Position (Virtual Flowtubes).........................................................................................119
Gas Selection...................................................................................................................120
Flowtube Resolution ........................................................................................................ 121
O2 Whistle ....................................................................................................................... 122
II
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS TIRO
CONTENTS (continued)
DESCRIPTION PAGE
Alarms.............................................................................................................................. 123
No Fresh Gas................................................................................................................... 124
Fresh Gas Low Alarm ...................................................................................................... 125
Threshold Low Alarm ....................................................................................................... 126
Pressure........................................................................................................................... 127
Ambient Pressure ............................................................................................................ 128
Plateau-Mean Display Screen.......................................................................................... 128
Secure Options ................................................................................................................ 129
Serial Port ........................................................................................................................ 131
Serial Port Parameters..................................................................................................... 131
Pump On/Off .................................................................................................................... 133
REPLACEMENT PROCEDURES ............................................................................................ 135
Replacement Procedures................................................................................................. 137
Core Module Inversion..................................................................................................... 139
Core Module Inversion Disassembly................................................................................ 139
Core Module Inversion Reassembly ................................................................................ 139
Wall Mount Service Access Procedure (If Applicable)..................................................... 143
Cylinder Yokes (Fixed) and Regulators............................................................................ 145
Cylinder Pressure Gauges............................................................................................... 149
Auxiliary Oxygen Flow Meter ........................................................................................... 151
Vaporizers ........................................................................................................................ 153
O2 Flush Valve................................................................................................................. 155
Caster .............................................................................................................................. 157
Ventilator .......................................................................................................................... 159
Gas Inlet Assembly (including O2 supply pressure switch) ............................................. 171
Battery.............................................................................................................................. 173
Power Supply................................................................................................................... 175
Control PCB Assembly..................................................................................................... 177
Pneumatic (PEEP Control) Assembly.............................................................................. 183
Flow Meter Bezel Assembly............................................................................................. 185
Fresh Gas Flow Meter...................................................................................................... 187
Flow Control Valves ......................................................................................................... 189
Fresh Gas Flow Sensors and Filter Assembly................................................................. 191
Pipeline Pressure Gauges ............................................................................................... 193
SORC (Sensitive Oxygen Ratio Controller) ..................................................................... 195
Fresh Gas Display PCB ................................................................................................... 197
Monitor Bezel Assembly................................................................................................... 199
Spirolog Sensor and Cable .............................................................................................. 201
ADJUSTMENT AND CALIBRATION PROCEDURES............................................................. 203
Cylinder Pressure Regulator Adjustment......................................................................... 205
Gas Inlet Regulator Output Adjustment ........................................................................... 207
Oxygen Supply Pressure Alarm Switch Adjustment ........................................................ 209
Sensitive Oxygen Ratio Controller (SORC) Adjustment ...................................................211
III
CONTENTS (continued)
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS TIRO
DESCRIPTION
PAGE
Oxygen Sensor Calibration .............................................................................................. 214
Pressure Calibration......................................................................................................... 216
Fresh Gas Flow Calibration.............................................................................................. 217
PEEP Valve Calibration.................................................................................................... 218
Vacuum Adjustment ......................................................................................................... 219
Vacuum Pressure Pump Assembly (Software Version 2.N)............................................. 221
Pump Calibration - If Applicable....................................................................................... 221
SOFTWARE UPDATE PROCEDURE ...................................................................................... 225
Software Update Procedure............................................................................................. 229
Requirements................................................................................................................... 229
Boot Strap Download Procedure...................................................................................... 229
PMS PROCEDURE ...................................................................................................................241
SPARE AND REPLACEMENT PARTS LIST........................................................................... 307
IV
General
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
1
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS

General Fabius Tiro

1 Recommendations Because of the sophisticated nature of Draeger Medical, Inc. anesthesia

equipment and its critical importance in the operating room setting, it is highly recommended that only appropriately trained and experienced professionals be permitted to service and maintain this equipment. Please contact DrägerService® at (800) 543-5047 for service of this equipment in North America. For service in Europe, call 49 (451) 882-4222. For service in other countries, call (215) 721-5402.
Draeger Medical, Inc. recommends that the Fabius Tiro be serviced at six month intervals. Periodic Manufacturer’s Certification agreements are avail­able for equipment manufactured by Draeger Medical, Inc. Please contact us for further information concerning these agreements.
Draeger Medical, Inc. products/material in need of factory repair shall be sent to:
2 How To Use This
Manual
For North America and other coun­tries except Europe:
DrägerService® 3124 Commerce Drive Telford, PA 18969 U.S.A. (Include RMA Number)
The manual is divided into several sections. The DIAGNOSTICS section describes self-test and service diagnostics for checking the system functions. An understanding of the on-board service capabilities is necessary before any attempt is made to troubleshoot the unit. The TROUBLESHOOTING sec­tion lists error codes and provides troubleshooting guides to assist the Tech­nical Service Representative in locating the source of a problem. The REPLACEMENT PROCEDURES section contains instructions for removal and replacement of assemblies that are considered field-replaceable. The ADJUSTMENT AND CALIBRATION PROCEDURES section contains the field procedures needed to restore original system specifications. The Peri­odic Manufacturer’s Service (PMS) PROCEDURE section outlines the steps required to verify the electrical, mechanical, and pneumatic safety of the unit and also, identifies components requiring periodic replacement. The SPARE PARTS section is provided for use as a reference only to obtain part numbers and descriptions for parts and assemblies for replacement purposes. For items not shown in the Spare Parts section, contact DrägerService®.
For service in Europe:
Dräger Medical AG & Co. KGaA Moislinger Allee 53-55 Reparaturannahme 23542 Lübeck Germany
3 General Trouble-
shooting Guidelines
All rights reserved. Copyright reserved.
K6020002_GeneralUSA.fm 17.06.04
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
Troubleshooting the Fabius Tiro should always begin by communicating with those who observed or experienced a problem with the unit. This may elimi­nate unnecessary troubleshooting steps. Once a general problem is identi­fied, refer to the flow charts in the Troubleshooting Section to determine the proper corrective action to be taken.
After any component is replaced, verify that the unit is operating properly by running the appropriate diagnostic procedure. The PMS PROCEDURE must also be performed after any component is replaced.
3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius Tiro General
There are two possible mounting configurations for the Fabius Tiro, a wall mounted unit and a trolley mounted unit. A general arrangement of both mounting configurations of the Fabius Tiro anesthesia system are shown in
Figure 1 and Figure 2.
Figure 1 Fabius Tiro - Wall Mount Configuration
4
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
All rights reserved. Copyright reserved.
K6020002_GeneralUSA.fm 17.06.04
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
General Fabius Tiro
Figure 2 Fabius Tiro - Trolley Mount Configuration
All rights reserved. Copyright reserved.
K6020002_GeneralUSA.fm 17.06.04
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
5
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius Tiro General

4 Copyright Copyright© 2004 by Draeger Medical, Inc. All rights reserved. No part of this

publication may be reproduced, transmitted, transcribed, or stored in a retrieval system in any form or by any means, electronic or mechanical, including photocopying and recording, without written permission of Draeger Medical, Inc.

5 Trademark Notices Datagrip, DrägerService, ORM, Quality Service For Life, Respitone, Narko-

med, Vigilance Audit, Vitalert, Vitalink, Narkomed GS, Fabius, Fabius Tiro, and Fabius GS are registered trademarks of Draeger Medical, Inc. All other products or name brands are trademarks of their respective owners.
6Disclaimer The content of this manual is furnished for informational use only and is sub-
ject to change without notice. Draeger Medical, Inc. assumes no responsibil­ity or liability for any errors or inaccuracies that may appear in this manual.
6
All rights reserved. Copyright reserved.
K6020002_GeneralUSA.fm 17.06.04
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Function Description
RETURN TO CD-ROM TABLE OF CONTENTS
7
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
8
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS

Function Description Fabius Tiro

1 General Information
about the Fabius Tiro
The Fabius Tiro comprises the following assemblies:
Bezel assembly: Display and Control Panel
Flowmeter assembly
Gas Box: Gas Inlet Assembly and related items
Breathing system
Pneumatic Assembly
Ventilator
Anesthetic Vaporizer(s)
Trolley Mount and Wall Mount
Figure 1 Front View of Fabius Tiro Anesthesia System - Wall Mount Configuration
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
9
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius Tiro Function Description
Figure 2 Front View of Fabius Tiro Anesthesia System - Trolley Mount Configuration
10
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Function Description Fabius Tiro
Figure 3 Back View - Interface Panel and Power Entry
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
11
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius Tiro Function Description
Figure 4 Side View - Gas Pipeline Inlets and Tethered Yoke - Wall Mount Configuration
12
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Function Description Fabius Tiro
Figure 5 Back/Side View - Gas Pipeline and Cylinder Hose Connections - Trolley Mount Configuration
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
13
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius Tiro Function Description
2 Fabius Tiro Function
Diagram
Figure 6 Functional Diagram of Fabius Tiro - Typical for Breathing System P/N 4116398 or 4117529
14
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Function Description Fabius Tiro
Figure 7 Function Diagram of Fabius Tiro - Typical for Breathing System P/N 4118378 or 4118379
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
15
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius Tiro Function Description

3 Battery Backup Fabius Tiro backup power is provided by two series-connected 12 V

rechargeable batteries. These batteries remain on charge as long as the machine is plugged into an active AC outlet. Should power to the main con­troller PCB assembly fail while the machine is in operation, the batteries will allow the machine to continue operating for a minimum of 45 minutes, pro­vided the batteries are fully charged.
The batteries are located within the controller housing and are accessible by opening the ventilator compartment. The 3.15A battery fuse is located on the power entry panel at the back of the machine.
Figure 8 Battery Backup Arrangement
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
16
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Function Description Fabius Tiro
4 Fabius Tiro Piping
Diagram
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
17
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius Tiro Function Description
Figure 9 Fabius Tiro 3-Gas Piping Diagram
18
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Function Description Fabius Tiro
5 Function Descrip-
tion of Gas Box
The supply gases flow through the filters and non-return valves in the gas inlet assembly. Pipeline supply pressures are indicated on gauges located on the flowmeter assembly. Cylinder pressure gauges are located on the base machine assembly. The pressures of O2 and N2O delivered to the flowmeter assembly are set by regulators on the gas inlet assembly.
Should the O2 supply fail or its pressure decrease below a certain limit, the O2 supply pressure alarm switch signals an alarm.
Figure 10 Gas Box Functional Diagram (Part 1 of 2)
If the O2 flush button is pressed, oxygen is delivered to the fresh gas outlet. The fresh-gas ejector prevents the fresh gas from flowing back into the anes­thetic vaporizer. This avoids an increase in the anesthetic gas concentration.
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
19
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius Tiro Function Description
Figure 11 Gas Box Functional Diagram (Part 2 of 2)
6 SORC (Sensitive
Oxygen Ratio Con­troller)
20
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
The SORC is a control element that functions like an N2O shut-off device and ensures a vital O2 concentration in the fresh gas. In the event of an O2 short­age, the SORC limits the N2O flow such that the O2 concentration in the fresh gas does not decrease below 23 vol.%.
If the O2 flow control valve is closed or if the O2 flow is lower than or equal to 200 mL/min, the SORC interrupts the N2O flow.
N2O can be added when the O2 flow is approx. 300 mL/min. In this case, the SORC also prevents O2 concentrations below 23 vol.%.
The SORC bypass allows O2 to bypass the restrictor in the SORC when O2 flows above 10 L/min. are needed.
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Function Description Fabius Tiro
Figure 12 SORC Functional Diagram (Part 1 of 2)
The flow control valves are used to adjust the O2 and N2O flows.
Restrictors located at the outlets of the SORC generate back-pressures. These back-pressures exert a force on the control diaphragms of the SORC. The O2 back-pressure opens the SORC. The N2O back-pressure closes the SORC. The pressure ratio at the control diaphragm affects the N2O flow.
The restrictors and the spring tension are dimensioned such that a minimum concentration of 23 vol.% O2 is always ensured. The maximum O2 flow is approx. 12 L/min.
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
21
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius Tiro Function Description
7 Compact Breathing
System, Cosy II
Figure 13SORC Functional Diagram (Part 2 of 2)
The Cosy II compact breathing system allows various modes of patient venti­lation: manual and spontaneous breathing, volume controlled, and pressure controlled.
In the "MAN" position, the compact breathing system is closed to atmo­sphere. This position is used for manual ventilation of the patient. The APL valve opening pressure can be adjusted from 5 to 70 cmH2O.
In the "SPONT" position the APL valve is open to atmosphere. This position is used for spontaneous patient breathing.
The pressure limit (Pmax) can also be adjusted (through front panel interface) during volume control from 15 cmH2O to 70 cmH2O using the control box and the PEEP/Pmax valve.
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
22
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Function Description Fabius Tiro
Figure 14 Compact Breathing System, Cosy II - Typical for P/N 4116398 or 4117529
Table 1 Key, Figure 14
1 PEEP/Pmax valve
2 Flow sensor (Spirolog)
3 MAN/SPONT-APL Valve
4 Fresh-gas decoupling valve
5 Breathing bag hook
6 Inspiratory port
7 Inspiratory valve and O2 sensor port
8 Breathing bag terminal and Y-piece plug
9 Expiratory valve
10 Expiratory port
11 APL Bypass valve
12 Fresh-gas port
13 Ventilator port
14 Anesthetic gas scavenging port
15 Absorber
16 Pressure sensor connector
17 Exhaust port, anesthetic monitor return (non-U.S. systems only)
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released
23
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Fabius Tiro Function Description
Figure 15 Compact Breathing System, Cosy II - Typical for P/N 4118378 or 4118379
Table 2 Key, Figure 15
1 PEEP/Pmax valve
2 Flow sensor (Spirolog)
3 MAN/SPONT-APL Valve
4 Fresh-gas decoupling valve
5 Breathing bag hook
6 Inspiratory port
7 Inspiratory valve and O2 sensor port
8 Breathing bag terminal and Y-piece plug
9 Expiratory valve
10 Expiratory port
11 APL Bypass valve
12 Fresh-gas port
13 Ventilator port
14 Anesthetic gas scavenging port
15 Absorber
16 Pressure sensor connector
17 Exhaust port, anesthetic monitor return (non-U.S. systems only)
All rights reserved. Copyright reserved.
F6020002_Function_Description_USA.fm 17.06.04
24
Dräger Medical AG & Co. KGaA 6020.002 Revision B Released