15.2 Required spare parts .......................................................................................................... 190
15.3 Service Equipment Required .............................................................................................. 191
15.4 Removing the pressure regulator ....................................................................................... 192
15.5 Replacing the “Major Overhaul” spare parts set ................................................................. 194
All rights reserved. Copyright reserved.
K5132300IECIVZ.fm 02.06.06
VII
Contents
Schematics and Diagrams
1Primus pneumatic components diagram 203
2Schematics and Diagrams 209
Annex
Parts catalog
Test List
Technical Information
VIII
K5132300IECIVZ.fm 02.06.06
All rights reserved. Copyright reserved.
General
1
2
PrimusGeneral
1Symbols and Defini-
tions
WARNING
A WARNING statement provides important information about a potentially hazardous situation which, if not avoided, could result in death
or serious injury.
CAUTION
A CAUTION statement provides important information about a potentially
hazardous situation which, if not avoided, may result in minor or moderate
injury to the user or patient or in damage to the equipment or other property.
NOTE
A NOTE provides additional information intended to avoid inconvenience
during operation.
Definitions according to German standard DIN 31051:
Inspection= examination of actual condition
Maintenance= measures to maintain specified condition
Repair= measures to restore specified condition
Servicing= inspection, maintenance, and repair
2Notes
This Technical Documentation conforms to the IEC 60601-1 standard.
Read each step in every procedure thoroughly before beginning any test.
Always use the proper tools and specified test equipment. If you deviate from
the instructions and/or recommendations in this Technical Documentation,
the equipment may operate improperly or unsafely, or the equipment could
be damaged.
It is our recommendation to use only Dräger parts and supplies.
The maintenance procedures described in this Technical Documentation may
be performed by qualified service personnel only. These maintenance procedures do not replace inspections and servicing by the manufacturer.
The information in this Technical Documentation is confidential and may not
be disclosed to third parties without the prior written consent of the manufacturer.
This Technical Documentation is for the purpose of information only. Product
descriptions found in this Technical Documentation are in no way a substitute
for reading and studying the Instructions for Use/Operating Manual enclosed
with the product at the time of delivery.
All rights reserved. Copyright reserved.
Version 3.0_ Released_Printed on_02.06.06_General_Technical_Documentation.fm
5132.300
3
GeneralPrimus
Know-how contained in this Technical Documentation is subject to ongoing
change through research and development and Dräger Medical reserves the
right to make changes to this Technical Documentation without notice.
NOTE
Unless otherwise stated, reference is made to laws, regulations or standards (as amended) applicable in the Federal Republic of Germany for
equipment used or serviced in Germany. Users or technicians in all other
countries must verify compliance with local laws or applicable international
standards.
Version 3.0_ Released_Printed on_02.06.06_General_Technical_Documentation.fm
All rights reserved. Copyright reserved.
4
5132.300
Function Description
5
6
PrimusFunction Description
1General
1.1Medical purpose
Primus is an anaesthetic workstation for automatic and manual ventilation
and spontaneous breathing, usable for adults, children and infants.
Application:
–Inhalation anesthesia in rebreathing systems.
–Inhalation anesthesia in virtually closed systems for low-flow and mini-
mum-flow applications.
–Inhalation anesthesia in non-rebreathing systems with separate fresh gas
output for connection of “Bain” or “Magill” system, for example, with a
fresh gas flow of 0.2 L/min to 18 L/min.
Changed ventilation modes (software 2.n or higher):
–Volume-controlled ventilation “Vol u m e Mode”. Switchable functions:
Version 2.0_ Released_Printed on_02.06.06_F5132300T01_Allgemeines.fm
All rights reserved. Copyright reserved.
PrimusFunction Description
1.4.4Monitor control panel
Figure 3Monitor control panel
The parameters for fresh-gas flow control, ventilation, and gas monitoring are
displayed on a 12-inch color screen.
The following parameters are monitored:
–Airway pressure.
–Inspiratory and expiratory flow.
–Circle system leakage.
–Inspiratory and expiratory O2 concentration.
–CO2 measurement and anesthetic gas measurement.
–Anesthetic gas detection.
–Quantitative measurement of mixed-gas values and MAC calculation
(age-relevant).
A data view, a trend view (graphical) and a log view can be selected.
As from software release 2.n the following settings are possible:
–Free configuration of 3 real-time curves and different numerical values.
–Body-weight-relevant ventilator presettings (Vt and frequency) and age-
relevant calculation of minimum alveolar concentration (MAC ) according
to “Mapleson” as well as age-relevant scaling of volumeter and influence
on ventilation monitoring.
All rights reserved. Copyright reserved.
Version 2.0_ Released_Printed on_02.06.06_F5132300T01_Allgemeines.fm
5132.300
11
Function DescriptionPrimus
The safety concept incorporates the following tests:
–Automatic self-test with mixer test, ventilator test and test of the breathing
system.
–Test and automatic calibration of all sensors.
1.4.5OptionsPrimus is prepared for future upgrading with the following options:
–Integrated SpO2 measurement.
–Consumption-free O2 measurement (with software 2.n or higher).
–PAW preview – display of expected airway pressure curve when chang-
ing parameters.
–P/V and flow/V loops (with software 2.n or higher).
–Econometer function (with software 2.n or higher).
–Additional languages available for display texts.
1.5Primus component
structure
Figure 4Primus component structure
In the “Block diagrams” section, you will find a detailed block diagram of the
Primus.
1.5.1NEUTRAL POINT PCBThe NEUTRAL POINT PCB connects the components of the Primus to the
power supply, additional signal wires and the CAN bus.
More details are given in the following section on the NEUTRAL POINT PCB.
12
5132.300
Version 2.0_ Released_Printed on_02.06.06_F5132300T01_Allgemeines.fm
All rights reserved. Copyright reserved.
PrimusFunction Description
1.5.2Graphical User Interface
(GUI)
The GUI has the following components:
–On the Monitor Control Panel (MoBi) the ventilation mode is displayed.
Limit and target values are specified and the ventilation and anesthesia
parameters are displayed.
–S-Box (Interface box). PC interfaces and optional measuring functions
such as SpO2 and BIS
TM
.
–Patient Gas Module (PGM) for measurement of O2, CO2 and anesthetic
gas.
For more details refer to the section headed GUI.
1.5.3MixerThe mixer comprises the following function units:
–Electronically controlled and monitored mixer.
–Vapor plug-in system for one or two conventional vaporizer types.
–External fresh gas outlet, A-cone (optional).
–Pressure monitoring for CS and compressed gas cylinders.
More details are given in the following section on the mixer.
1.5.4VGC (Ventilation and
Gas Controller)
The VGC comprises the following function units:
–Electronically controlled and driven ventilator.
–Integrated breathing system for “low flow” and “minimum flow” applica-
tions.
For more details refer to the section headed VGC.
1.5.5Power packThe power pack comprises:
–Switched-mode power supply unit.
–Uninterruptible power supply (UPS) with one battery pack consisting of
two 12 V lead-gel batteries.
–Battery charging management.
For more details refer to the section headed Power pack.
1.5.6Cylinder pressure regulator
The cylinder pressure regulators reduce the pressure of the optional compressed gas cylinders.
For more details refer to the section headed Cylinder pressure regulators.
The function description relating to the NEUTRAL POINT PCB follows.
All rights reserved. Copyright reserved.
Version 2.0_ Released_Printed on_02.06.06_F5132300T01_Allgemeines.fm
5132.300
13
Function DescriptionPrimus
14
5132.300
Version 2.0_ Released_Printed on_02.06.06_F5132300T01_Allgemeines.fm
All rights reserved. Copyright reserved.
PrimusFunction Description
2NEUTRAL POINT
PCB
The NEUTRAL POINT PCB is the central signal and voltage distributor.
Figure 5 Location of the Neutral Point PCB
All rights reserved. Copyright reserved.
Version 2.0_ Released_Printed on_02.06.06_F5132300T02_LP_Sternpunkt.fm
5132.300
15
Function DescriptionPrimus
Figure 6Component mounting diagram, NEUTRAL POINT PCB, for leg-
end see Table 1
Table 1Legend to Figure 6
ItemConnector
1Monitor Control Panel (MoBi).
2Monitor Control Panel (MoBi).
3Ventilation and Gas Controller (VGC).
4Mixer B.
5Mixer A.
6Safety O2 flow valve (microswitch).
7Power switch (main switch).
8Halogen lamp
9Jack plug (production tests).
10PGM.
11Power pack A.
12Power pack B.
13Not assigned.
The function description relating to the GUI follows.
16
5132.300
Version 2.0_ Released_Printed on_02.06.06_F5132300T02_LP_Sternpunkt.fm
All rights reserved. Copyright reserved.
PrimusFunction Description
3GUI The following section describes the user interface (“GUI = Graphical User
Interface”).
Figure 7GUI block diagram
3.1Monitor Control
Panel (MoBi)
In the “Block diagrams” section, in the block diagram of the Primus, you will
find a block diagram of the GUI.
The GUI has the following components:
–MoBi (monitor control panel).
–S-Box (Interface Box).
–Patient Gas Module (PGM). The function description relating to the
Patienten Gas Module (PGM) follows.
Version 5.0_ Released_Printed on_02.06.06_F5132300T03_GUI.fm
All rights reserved. Copyright reserved.
5132.300
Figure 8 Position of Monitor Control Panel (MoBi)
In the “Block diagrams” section, you will find a block diagram of the MoBi.
17
Function DescriptionPrimus
The user and Primus communicate via the MoBi. The MoBi display presents
system and patient information. It is here that the user sets the parameters
and ventilation modes.
The Patient Gas Module (PGM) is connected to the GUI.
Figure 9Exploded view of MoBi, for legend see Ta b le 2
Table 2Legend to Figure 9
ItemComponents
1Front panel with membrane keypad. Includes keypad membrane
covering with design imprint, keys, LEDs (e.g. for CS gases), the
carrier plate and the shielding, anti-glare glass screen.
212 inch color display (TFT, resolution: 800 x 600).
3MONITOR CONTROL PANEL PCB (motherboard).
4Backlight converter (display backlighting).
5LCD800 PCB (adapter PCB for connection of different makes of
display).
6Loudspeaker.
7Rotary transducer.
8Control knob (central operator control element).
18
5132.300
Version 5.0_ Released_Printed on_02.06.06_F5132300T03_GUI.fm
All rights reserved. Copyright reserved.
PrimusFunction Description
3.1.1MONITOR CONTROL
PANEL PCB
The following software is installed on the PCB:
–GUI software.
–Monitoring and evaluation software for the PGM.
–Software for Medibus connections and SpO2.
A 2-processor system is in operation on the PCB. It comprises a Display Master (DiMa) and a Communication Master (CoMa).
The powerful DiMa processor incorporates the following components:
–Motorola processor (MPC823) with 48 MHz clock frequency and 32-bit
address and data buses.
–Flash-PROM (program memory).
–RAM (data memory).
–CAN controller.
–RS232 interface for in-house development purposes.
–Serial communication channel for Ethernet.
The LCD controller is a programmable logic device (“PLD”). A “DRAM” serves
as the video memory.
The CoMa processor system primarily controls communication with the
other Primus components.
The CoMa incorporates the following components:
–Motorola processor (M68332) with 16.7 MHz clock frequency, internal 32-
bit bus and external 16-bit data bus.
–Flash-PROM (program memory).
–RAM (data memory).
–RS232 interface for communication with the SPO2, PGM and Medibus 1 -
3 modules.
–Real-time clock (RTC).
–Keyboard and rotary knob scan, LED actuation and sound output.
–CAN interface.
Both processor systems communicate by way of a Dual-Port RAM (DPR).
This memory device is battery-buffered. The buffering is provided primarily by
the UPS batteries of the Primus. If they fail, the lithium battery on the MONITOR CONTROL PANEL PCB ensures data is retained.
The operating voltage is provided by an unstabilized voltage of 20 V to 30 V
(Vcc). DC converters on the MONITOR CONTROL PANEL PCB generate all
other voltages on the PCB.
The MoBi is interconnected over the CAN bus with the other components of
the Primus (power pack, mixer and VGC).
Version 5.0_ Released_Printed on_02.06.06_F5132300T03_GUI.fm
All rights reserved. Copyright reserved.
5132.300
19
Function DescriptionPrimus
3.2S-Box (interface
box)
Figure 10 Position of S-Box
In the “Block diagrams” section, you will find a block diagram of the S-Box.
20
Figure 11 Layout of the S-Box, for legend see Tab l e 3
5132.300
Version 5.0_ Released_Printed on_02.06.06_F5132300T03_GUI.fm
All rights reserved. Copyright reserved.
Loading...
+ 306 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.