4.Use retaining ring pliers to remove the top plate (6)
and retaining ring (7). Pull cap (8) and o-ring (9) from
body.
5.Unscrew socket head screw (1), then remove button
(2), spring (3), knob (4), tumblers (5), coupling (10),
and o-ring (11).
6. Remove, as an assembled unit, the adjusting screw
(12) with upper and lower piston assemblies (items
13 thru 19). Remove spring (20).
7. Unscrew the adjusting screw (12), then separate
upper and lower piston assemblies (items 13 thru
19). The adjusting screw has left hand threads.
Make sure o-ring (18) remains attached to the lower
piston (17). If not, retrieve it from inside the upper
piston (13).
8. Unscrew bottom plug (21) to gain access to the
parts (22 thru 33) located in the lower portion of the
body.
CLEANING
1.Clean parts with warm water and soap.
2.Rinse and dry parts. Blow out internal passages in
body (34, 35) with clean, dry compressed air.
3.Inspect parts. Replace those found to be damaged.
ASSEMBLY
1. Lubricate o-rings, o-ring sealing surfaces, and the
adjusting screw threads (12) with a small amount of
good quality o-ring grease.
2. Assemble the lower piston (16*, 17, 18,19), the
regulating spring (15), and upper piston (13, 14),
then turn the adjusting screw (12) into the assembly
until the tip of the screw is flush with the bottom of
the lower piston (17). The adjusting screw has left
hand threads. Place spring (20) in position in the
body, then place the upper and lower piston
assembly (items 12 thru 19) in the body.
3. Insert coupling (10) and o-ring (11) into cap (8).
Place tumblers (5), knob (4), spring (3), and button
(2) in position and secure with screw (1). Make
sure button (2) is installed with the large end
(marked A) up. Place o-ring (9) in position on the
cap (8), then place the cap and knob assembly
(items 1 thru 5, and 8 thru 11) in body, making sure
coupling (10) engages the adjusting screw (12).
4. Use retaining ring pliers to install retaining ring (7),
then install and top plate (6).
WARNING
Make sure retaining ring (7) is fully seated in the
groove in the body.
5. Install the lower section parts (items 21 thru 33)
into body (34 or 35).
6. Torque Table
Item Inch-Pounds (Nm)
21 (Bottom plug) 20 to 30 (2.3 to 3.4)
1 (Screw
)
3 to 4 (0.3 to 0.5). Screw head
must not extend more than
0.050" (1.3 mm) above top of
button (2).
7. Set the R24 for normal operation as described in
paragraph 1-2, page 2 before it is returned to
service.
* Item 16, washer, is used only with 1/4", 3/8", and
1/2" ported R24 regulators manufactured from
September, 1993 to March, 1994. The design of the
piston (17) was changed in early 1994, and washer
(16) was eliminated. The washer support ribs on the
early pistons (Detail A of exploded view) are
recessed approximately 0.188" (5mm) to
accommodate washer (16). The support ribs on
current pistons are recessed approximately 0.063
(2mm). Discard washer (16) when replacing the
early piston with a current piston.
IM-414.210.01
(2/02)
Replaces NIP-243
© Norgren 2002
TECHNICAL DATA
Fluid: Compressed air
Maximum inlet pressure: 300 psig (20 bar)
Operating temperature: 0° to +150°F (-20° to +80°C)
Air supply must be dry enough to avoid ice
formation at temperatures below +35°F (+2°C).
Typical flow at 150 psig (10 bar) inlet pressure, 90 psig
(6.3 bar) set pressure, and 15 psig (1 bar) droop
from set:
1/2" ports: 200 scfm (94 dm
3
/s)
1-1/4" ports: 700 scfm (330 dm
3
/s)
Maximum bleed rate at 50 psig (3.5 bar) outlet
pressure: 0.031 scfm (0.015 dm3/s). Maximum bleed
rate occurs under dead-end (no flow) conditions.
Port sizes:
Main Gauge
1/4" 1/4"
3/8" 3/8"
1/2", 3/4", 1", 1-1/4" 1/2"
Thread type: PTF, ISO G, or ISO Rc
Materials:
Body, top cap: Zinc
Main valve, adjusting screw: Brass
Pilot valve, relief valve, bottom plug: Acetal
Elastomers: Nitrile
REPLACEMENT ITEMS
Service kit - 1/4", 3/8" 1/2" ported regulators (Items 9,
11, 12, 14, 18, 19, 22 thru 26, 28, 29, 33).....5292-52
Service kit - 3/4", 1", 1-1/4" ported regulators (Items 9,
11, 12, 14, 18, 19, 23 thru 26, 30 thru 33) ....5292-53
PANEL MOUNTING DIMENSIONS
Panel mounting hole diameter: 1.26" (32 mm)
Maximum panel thickness: 0.12" (3 mm)
INSTALLATION
1.Shut off air pressure. Install regulator in air line -
● with air flow in direction of arrow on body,
● upstream of lubricators and cycling valves,
● as close as possible to the device being serviced.
● at any angle.
2.Connect piping to proper ports using pipe thread
sealant on male threads only. Do not allow sealant to
enter interior of regulator.
3.
Install a pressure gauge or plug the gauge ports.
Gauge ports can also be used as additional outlets for
regulated air.
4.Install a Norgren general purpose filter upstream of
the regulator.
ADJUSTMENT
1.Before applying inlet pressure to regulator, turn
adjustment (4) counterclockwise until knob stops.
2.Apply inlet pressure, then turn adjustment (4)
clockwise to increase and counterclockwise to
decrease pressure setting.
3.Always approach the desired pressure from a lower
pressure. When reducing from a higher to a lower
setting, first reduce to some pressure less than that
desired, then bring up to the desired pressure.
4. If desired, a maximum or minimum outlet pressure
adjustment limit may be set. See page 2 for
instructions.
DISASSEMBLY
1.Regulator can be disassembled without removal from
air line.
2.Shut off inlet pressure. Reduce pressure in inlet and
outlet lines to zero.
3.Turn adjustment (4) fully counterclockwise.
WARNING
Prior to performing step 4, make certain air
pressure upstream and downstream of the R24 is
zero psi.
R24
Installation & Maintenance
Instructions
* Outlet pressure can be adjusted to pressures in excess of, and less than, those specified. Do not use these units to control pressures outside of the specified ranges.
Special Purpose Regulator
Adjustable Pressure Limits - High Relief Flow
R24 - ★★★ - ★★★★
Thread Form
A....PTF
B....ISO Rc taper
G....ISO G parallel
Port
2....1/4" 6 ....3/4"
3....3/8 8 ....1"
4....1/2" A....1-1/4"
Spring (Outlet Pressure Range) *
C....5 to 30 psig (0.3 to 2 bar)
F ....5 to 60 psig (0.3 to 4 bar)
L....10 to 125 psig (0.7 to 8 bar)
S....10 to 250 psig (0.7 to 17 bar)
Type
R....Relieving piston
Gauge
G....With
N....Without
Option
0....Not applicable
Adjustment Type
0....Knob
a subsidiary of IMI plc
1
2
3
4
5
34
36
29
28
29
16
17
24
25
26
27
0.188"
(5mm)
A
6
7
8
9
11
10
12
14
13
15
18
17
19
20
35
37
38
31
30
32
33
22
23
21
R24
Installation & Maintenance
Instructions
1-3. How to Set Outlet Pressure Limit
when the Pressure Set Button is in the
Tamper Resistant Position (Fig 1).
To Set a Maximum Pressure Limit:
1. With primary pressure applied to the R24, turn
adjusting knob clockwise until the desired maximum
outlet pressure is reached.
2. Hold adjusting knob, then use a 5/64" hex wrench
to loosen button screw 4 to 5 turns.
3. Depress and hold button screw, then turn adjusting
knob clockwise until it stops. Release button
screw.
4. Hold adjustment knob and tighten button screw. Do
not depress screw, as the adjustment limit will
be affected.
5. Turn adjusting knob fully counterclockwise.
To Set a Minimum Pressure Limit:
1. With primary pressure applied to the R24, turn
adjusting knob clockwise until the desired minimum
outlet pressure is reached.
2. Hold adjusting knob, then use a 5/64" hex wrench
to loosen button screw 4 to 5 turns.
3. Depress and hold button screw, then turn the
adjusting knob counterclockwise until it stops.
Release button screw.
4. Hold adjustment knob and tighten screw. Do not
depress screw, as the adjustment limit will be
affected.
WARNING
When the minimum adjustment limit is set
above zero, the upper limit of adjustment
increases correspondingly, and could possibly
approach inlet pressure.
1-2. Regulators Shipped from Factory:
● allow pressure adjustment throughout the entire
outlet pressure range as shown on the regulator
label and as listed in the model number description
at the top of page 1.
● have the red pressure set button (Figure 1) in the
tamper resistant position, preventing the accidental
setting of a pressure limit. When the button is in
the tamper resistant position, the letters TR are
visible on the button; when the button is in the
adjustable position, the letter A is visible.
1-2. Set the R24 for Normal Operation
(Fig 1).
Any time the R24 is disassembled and reassembled,
the adjusting knob must be reset for normal operation
(i.e., set to operate thru the entire outlet pressure
range as listed in the model number description at the
top of page 1).
1. Make sure the red pressure set button is in the
adjustable position (end marked A up). If the end
marked TR is up, reverse the position of the button
by performing steps 1 and 2 in paragraph 1-5,
Tamper Resistant Procedure, but reinstall the button
with the end marked A up.
2. Install the regulator in a compressed air test line
with gages installed for observing inlet and outlet
pressures. Inlet pressure should not exceed 250
psig (17.2 bar).
3.Turn adjusting knob fully counterclockwise (ccw),
then slowly apply inlet pressure while observing the
outlet pressure gage.
If outlet pressure is indicated with the knob fully
ccw:
a. Depress and hold the pressure set button, then
turn adjusting knob clockwise (cw) until it stops
and release button.
b. Turn adjusting knob ccw until outlet pressure
reaches zero. If outlet pressure does not reach
zero when knob is turned fully ccw, repeat steps (a)
and (b) until it does.
c. When outlet pressure reaches zero, turn knob
an additional 1/4 turn ccw. Depress pressure set
button, then turn knob fully ccw and release
button.
d. Turn knob slowly cw and verify that outlet pressure
starts after knob is turned 1/4 turn. Fine
adjustment can be made by depressing the
pressure set button and turning knob cw or ccw
as needed.
If outlet pressure is not indicated with the knob
fully ccw:
a. Turn adjusting knob clockwise (cw) until outlet
pressure is indicated.
b. If outlet pressure is not indicated when the knob is
turned fully cw, depress pressure set button, then
turn adjusting knob ccw until it stops and release
button. Repeat steps (a) and (b) until outlet
pressure is indicated when the knob is turned cw.
c. When outlet pressure starts, turn knob ccw 1/4
turn. Depress pressure set button, then turn knob
fully ccw and release button.
d. Turn knob slowly cw and verify that outlet pressure
starts after knob is turned 1/4 turn. Fine
adjustment can be made by depressing the
pressure set button and turning knob cw or ccw as
needed.
After the regulator is set for normal operation, the
adjusting knob may be set to limit the maximum or the
minimum adjustable pressure setting of the regulator.
For example, a regulator with a 10 to 125 psig outlet
pressure range could be set to limit the maximum
outlet pressure to 90 psig, or it could be set to limit the
minimum outlet pressure to 25 psig.
IM-414.210.02
(2/02)
Setting Outlet Pressure Adjustment Limits
On The R24
© Norgren 2002
1-4. How to Set Outlet Pressure Limit
when the Pressure Set Button is in the
Adjustable Position (Fig 1).
To Set a Maximum Pressure Limit:
1. With primary pressure applied to the R24, turn
adjusting knob clockwise until the desired
maximum outlet pressure is reached.
2. Depress and hold button screw, then turn the
adjusting knob clockwise until it stops. Release
button screw.
3. Turn adjusting knob fully counterclockwise.
To Set a Minimum Pressure Limit:
1. With primary pressure applied to the R24, turn
adjusting knob clockwise until the desired
minimum outlet pressure is reached.
2. Depress and hold button screw, then turn the
adjusting knob counterclockwise until it stops.
Release button screw.
WARNING
When the minimum adjustment limit is set
above zero, the upper limit of adjustment
increases correspondingly, and could possibly
approach inlet pressure.
1-5. Tamper Resistant Procedure (Fig 1).
Adjustment limits can be made tamper resistant by
reversing the position of the red pressure set button
so the end marked TR is up. To reverse the button:
1. Apply minimum 25 psig (1.7 bar) to the regulator.
Turn adjusting knob fully clockwise, then hold
adjusting knob and and remove button screw
using a 5/64" hex wrench. Do not depress
screw, as the adjustment limit will be affected.
Remove button, then reinstall with the end
marked TR up.
2. Install button screw, taking care not to push screw
in when tightening. Turn adjusting knob fully
counterclockwise.
1-6. Warning.
These products are intended for use in industrial
compressed air systems only. Do not use these
products where pressures and temperatures can
exceed those listed under Technical Data.
If outlet pressure in excess of the regulator
pressure setting could cause downstream equipment
to rupture or malfunction, install a pressure relief
device downstream of the regulator. The relief
pressure and flow capacity of the relief device must
satisfy system requirements.
The accuracy of the indication of pressure gauges
can change, both during shipment (despite care in
packaging) and during the service life. If a pressure
gauge is to be used with these products and if
inaccurate indications may be hazardous to
personnel or property, the gauge should be
calibrated before initial installation and at regular
intervals during use.
Before using these products with fluids other than
air, for non industrial applications, or for life-support
systems consult Norgren.
Figure 1. R24 Pressure Set Button
RED PRESSURE SET BUTTON
TR = Tamper Resistant Position
A = Adjustable Position
ADJUSTING
KNOB
BUTTON
SCREW
a subsidiary of IMI plc