Test points ..............................................................................................................................................7
“CONTACT SERVICE” on display ........................................................................................................8
This document is intend to be a guide for localizing and repairing electrical faults in the
RH-29 device. First there is a brief guide for fault localizing. Then fault repairing is
divided into troubleshooting paths.
Before any service operation you must be familiar with the RH-29 product and module
level architecture. You have to also be familiar with the RH-29 specified service tools
such as the Phoenix service software, flashing tools and software.
General guidelines for RH-29 troubleshooting
Tools needed for troubleshooting
•Service tools (as listed at service tools chapter in service manual)
•Laboratory power supply with current indicator
•Oscilloscope
•Digital multimeter
General guidelines
If the device cannot be turned on by any means, see “dead device” troubleshooting
Current consumption (missing consumption) gives an idea whether the device is able to
start up.
Dropping supply voltage or very large current consumption indicates a short circuit
Check whether the connection with Phoenix works and what can be discovered with
Phoenix (ADC-readings, baseband self-test, BB-calibrations etc.)
Check baseband self-tests with Phoenix if “CONTACT SERVICE” is shown on the display.
Check visually display and rocker faults.
Force phone to LOCAL mode and make keyboard test by phoenix.
Check that board-to-board connector is OK, and connectors make good contacts.
If liquid damage, stop repairing!
Flash the phone before disassembling it, if the fault is not obvious and the Phoenix connection is OK.
RH-29Company Confidential
6b - Baseband TroubleshootingNokia Customer Care
Check failed module visually:
Mechanical damages?
Solder joints OK?
Continue with specific troubleshooting procedure for the module:
If there is an obvious fault, repair it before reflashing the device.
Flash first if a fault is not obvious.
If flashing is not working go to flashing troubleshooting.
Due to CSP packages short circuits or broken solder joints are not easily seen. If the
examined signal seems to be continuously in low or high level, then measure for possible
short circuit to ground (signal low) or to supply voltage (signal high) Note that if a problem is not found from any visible contact/component it can be under CSPs where the signal is connected.
Care must be taken when assembling and disassembling the transceiver. Failure to do
this may result in unnecessary damage to device.
Nominal current consumption
NOTE: Service tools need some amount of current to work.
The following current consumption values are measured from a complete RH-29.
Vbatt = 3.6V
Measured nominal currents are drawn from the main battery.
Measurements have been made with a current probe connected to an oscilloscope.
Operating modeCurrent consumption
Idle 90 mA
2w audio call (backlights off) 250 mA
RH-29Company Confidential
6b - Baseband TroubleshootingNokia Customer Care
Partially damaged device
If the device is working, but some functionality is missing try to localize where the problems is and see relevant part of this manual. E.g audio is not working see “Audio Troubleshooting”, if charging is not working see chapter Charging troubleshooting etc.
Most common symptoms reported by customer
In this chapter is described most common symptoms reported by customers when the
device is brought in for service. Some tips where the trouble can be found are given also.
When troubleshooting use these tips and follow the given troubleshooting path.
Most common symptoms for audio problems
“Earpiece sound is missing”
”Handsfree sound is missing”
”Headset is not recognized”
”Microphone is not working”
”Volume cannot be adjusted”
” Ringing tones does not work”
”Audio volume too low”
If symptom is something like above see audio troubleshooting.
Symptoms related to energy management
“Phone does not stay on”
”Charging is not working”
”Time is lost during battery change”
”Charging takes too long”
”Operating time is very short”
These symptoms lead to relevant part of energy management troubleshooting
Problems related to UI module:
“UI-module keypad is not working”
”Backlight is dim”
”Backlight not even”
”Backlight is blinking”
”Keypad or display backlight is not working”
”Display related problems”
See UI- module troubleshooting.
Most common RF related symptoms:
“Call cannot be made”
”Phone does not find signal”
”Call is often dropped”
See RF troubleshooting
RH-29Company Confidential
6b - Baseband TroubleshootingNokia Customer Care
“CONTACT SERVICE” on display
CONTACT SERVICE on display (Self-tests by Phoenix)
Display information: “Contact Service”
This fault means that software is able to run and thus the watchdog of UEM can be
served.
Self-test functions are executed when the phone is powered on and if one or more selftest functions fail, the message “Contact Service” is shown on the display.
MCU self-test cases can be split into two categories: The ones that are executed during
power up and the ones that are executed only with a PC connected. These test and the
items included are as follows:
If some self-test failed, see relevant chapter in this troubleshooting document.
RH-29 has three memory components installed on the main PWB. The best indication of
which one is causing problems can be obtained by flashing the device. It has to be kept
in mind that all three flashes are interfaced with UPP WD2 asic that might it self have
some problems. The necessary steps are described below. Phoenix error messages during
flashing greatly help on defining what is wrong. To be able to flash the device, most
device BB area components must function properly.
RH-29Company Confidential
6b - Baseband TroubleshootingNokia Customer Care
Energy management troubleshooting
Device does not stay on
If the device is switched off without any visible reason, there may be problems in the following areas:
•UEM watchdog problem (WD is not updated by SW)
•BSI line problem (BSI line is floating => contact failure)
•Battery line problem
•Soldering problem
The most likely reason is UEM WD (watchdog), which turns the device off after about 32
seconds if SW is jammed.
This may caused by SW problem, UPP_WD2 problem (Not server by SW), UEM or memory
malfunctions.
The following tests are recommended:
•General power checking
•Clocks
•Memory testing
•Serial Interface
If there is something wrong in BSI line, the device seems to be dead after the power key
is pressed. However the regulators of the device are on a few seconds before the powerdown.
This mode can easily be detected from the current consumption of the device. After a
few seconds the current consumption drops almost to 0 mA.
In this case check component or soldering
•Battery connector X382
EMI-filter R381
UEM D190 (pin number C2)
If phone boots to TEST or LOCAL mode with normal battery, BSI is short circuited to
ground. Check EMI-filter and filtering capacitors, which are located to BSI.