Check: Identify whether error cause(s) is (are) main board,
sub-board and/or DC+5V/±12V switching power supply.
Check that actual output voltage is ≧4.8V (nominal: 5VDC)
when measured at test points on main and sub-board.
Action:
When ≧4.8V is not outputted to the main board and sub-board:
・ Replace sub-boardwith new oneor the switching power
supply with new one.
When it is outputted to sub-board, while it is not outputted to
main board:
・ Replace main board with new one.
When it is outputted to the main board and sub-board:
・ Replace main board with new one.
E02 T1 TEMPERATURE SENSOR ERROR
Detection T1 temperature exceeding measurable limit
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Rese
Press [RESET] SW. Note
t
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Continuance
↑
↑
↑
↑
Dialysate line stop
↑
↑
↑
Condition or cause Check and action
Indication error of temperature and
other analog system
Incorrect [T1] terminal connection
Internal disconnection or short circuit
of [T1]
Introduction of high temperature water
from water supply source
Incorrect indicated temperature
calibration of main board
Dialysate line function BP/SP stop CLV function Reset
――○
Check: Check whether indications of all temperature sensors and other
sensors such as dialysate pressure sensor are incorrect.
Check stable output of voltage of ≧4.9V (nominal: 5Vref) at test
point on main board.
Action: Replace main board with new one.
Check: Check no incorrect wiring, loosening, corrosion and/or
disconnection at TM1 T1 terminal of main board.
Action: If loosened, tighten securely.
If incorrectly wired, correct.
If corroded or disconnected, replace [T1] with new one.
Check: Remove TM1 T1 terminal from main board and measure
resistance between Red and White to find short circuit,
disconnection and fluctuation.
Action: Replace [T1] with new one.
Action: Overhaul dialysate line to correct potential damage.
Action: Calibrate temperatures indicated by [T1], [T2] and [T3] circuits
of main board using temperature calibrator.
If the calibration is impossible, replace main board with new
one.
E03 T2 TEMPERATURE SENSOR ERROR
Detection T2 temperature exceeding measurable limit
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Rese
Press [RESET] SW. Note
t
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Continuance
↑
↑
↑
↑
Dialysate line stop
↑
↑
↑
Condition or cause Check and action
Indication error of temperature and
other analog system
Incorrect [T2] terminal connection
Internal disconnection or short circuit
of [T2]
Incorrect indicated temperature
calibration of main board
Dialysate line function BP/SP stop CLV function Reset
――○
After Ver 3, measurable extent of upper limit while
rinse isn’t detectable.
Check: Check whether indications of all temperature sensors and other
sensors such as dialysate pressure sensor are incorrect.
Check stable output of voltage of ≧4.9V (nominal: 5Vref) at test
point on main board.
Action: Replace main board with new one.
Check: Check no incorrect wiring, loosening, corrosion and/or
disconnection at TM1 T2 terminal of main board.
Action: If loosened, tighten securely.
If incorrectly wired, correct.
If corroded or disconnected, replace [T2] with new one
Check: Remove TM1 T2 terminal from main board and measure
resistance between Red and White to find short circuit,
disconnection and fluctuation.
Action: Replace [T2] with new one.
Action: Calibrate temperatures indicated by [T1], [T2] and [T3] circuits
of main board using temperature calibrator.
If the calibration is impossible, replace main board with new
one.
E04 T3 TEMPERATURE SENSOR ERROR
Detection T3 temperature exceeding measurable limit
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
Continuance
↑
↑
↑
↑
Dialysate line stop
↑
↑
↑
Condition or cause Check and action
Check: Check whether indications of all temperature sensors and other
sensors such as dialysate pressure sensor are incorrect.
Indication error of temperature and
other analog system
Check stable output of voltage of ≧4.9V (nominal: 5Vref) at test
point on main board.
Action: Replace main board with new one.
Check: Check no incorrect wiring, loosening, corrosion and/or
disconnection at TM1 T3 terminal of main board.
Incorrect [T3] terminal connection
Action: If loosened, tighten securely.
If incorrectly wired, correct.
If corroded or disconnected, replace [T2] with new one
Check: Remove TM1 T3 terminal from main board and measure
Internal disconnection or short circuit
of [T3]
resistance between Red and White to find short circuit,
disconnection and fluctuation.
Action: Replace [T3] with new one.
Action: Calibrate temperatures indicated by [T1], [T2] and [T3] circuits
Incorrect indicated temperature
calibration of main board
of main board using temperature calibrator.
If the calibration is impossible, replace main board with new
one.
――○
E05 DIALYSATE TEMPERATURE MAX ALARM
Detection Dialysate temperature T3 of ≧upper alarm limit during heater ON
Detection
process
Preparation
Prep. end
Dialysis
Recovery
Press [RESET] SW to reset.
Reset
Auto reset is caused after release from alarm
condition for 10 seconds.
Audible
alarm
○ (Continuous) Auto bypass
Heater OFF
Condition or cause Check and action
Actually higher dialysate temperature
○Failure of main board
○Incorrect calibration of indicated
temperature of main board
○Higher supply water temperature
○Lower priming flow rate
Dialysate line function BP/SP stop CLV function Reset
――○
Note
Check: Check whether actual dialysate temperature is higher than
upper alarm limit by touching dialysate supply line with hand
when the line is connected to dialyzer or by measurement of
temperature of dialysate discharged from coupler outlet using
thermometer when patient is not connected to the machine.
Action: Replace main board with new one.
Action: Calibrate temperatures indicated by [T1], [T2] and [T3] circuits
of main board using temperature calibrator.
If the calibration is impossible, replace main board with new
one.
Action: Adjust supply water temperature as specified.
・When heat exchanger is used : 5 to 30℃
・When heat exchanger is not used:17 to 30℃
Action: Adjust priming flow rate to 700 to 750mL/min.
Actually acceptable dialysate
temperature
○Failure of T3 temperature sensor
○Incorrect calibration of main board
Check: Check whether actual dialysate temperature is acceptable by
touching dialysate supply line with hand when the line is
connected to dialyzer or by measurement of temperature of
dialysate discharged from coupler outlet using thermometer
when patient is not connected to the machine.
Check: Check whether indicated temperature is fluctuated when
harness of T3 temperature sensor is shaken.
Check also whether indicated T3 temperature is deviated when
temperature calibrator is connected.
Action: If check result is unacceptable, replace T3 temperature sensor
with new one.
Check: Check whether indicated T3 temperature is deviated when
temperature calibrator is connected.
Action: Calibrate [T3] circuit of main board using temperature
calibrator.
If the calibration is impossible, replace main board with new
one.
E06 DIALYSATE TEMPERATURE MIN ALARM
Detection Dialysate temperature T3 of ≦lower alarm limit
Detection
process
Dialysis
Reset
Audible
alarm
○ (3 times)
Press [RESET] SW to reset.
Auto reset is caused after release from alarm
condition for 10 seconds.
Dialysate line function BP/SP stop CLV function Reset
Auto bypass
Note
――○
Condition or cause Check and action
Actually lower dialysate temperature
Check: Check whether indicated T3 temperature is deviated when
○ Failure of main board
○ Incorrect calibration of indicated
temperature of main board
○ Lower supply water temperature
○Failure of heater
○ Higher priming flow rate
Actually acceptable dialysate
temperature
○ Failure of T3 temperature sensor
○ Incorrect calibration of main board
Check: Check whether actual dialysate temperature is lower than lower
alarm limit by touching dialysate supply line with hand when the
line is connected to dialyzer or by measurement of temperature
of dialysate discharged from coupler outlet using thermometer
when patient is not connected to the machine.
Action: Replace main board with new one.
Action: Calibrate temperatures indicated by [T1], [T2] and [T3] circuits
of main board using temperature calibrator.
If the calibration is impossible, replace main board with new
one.
Action: Adjust supply water temperature as specified.
・When heat exchanger is used : 5 to 30℃
・When heat exchanger is not used:17 to 30℃
Check: Measure resistance between both ends of heater at room
temperature after power OFF.
Action: Replace heater if the resistance is out of the below ranges:
230V spec. :38.3 to 44.1Ω
110V spec. :8.7 to 10.0Ω
Action: Adjust priming flow rate to 700 to 750ml/min.
Check: Check whether actual dialysate temperature is acceptable by
touching dialysate supply line with hand when the line is
connected to dialyzer or by measurement of temperature of
dialysate discharged from coupler outlet using thermometer
when patient is not connected to the machine.
Check: Check whether indicated temperature is fluctuated when
harness of T3 temperature sensor is shaken.
Check also whether indicated T3 temperature is deviated when
temperature calibrator is connected.
Action: If check result is unacceptable, replace T3 temperature sensor
with new one.
temperature calibrator is connected.
Action: Calibrate [T3] circuit of main board using temperature
calibrator.
If the calibration is impossible, replace main board with new
one.
E07 DIALYSATE PRESSURE MAX LIMIT
Detection Dialysate pressure of ≧+400mmHg
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Auto reset of BP, SP and CLV
Reset
Press [RESET] SW to reset UF stop.
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Condition or cause Check and action
Pressure rise due to failure of [PR1]
Dialysate line function BP/SP stop CLV function Reset
UF stop
ON of V4 and V10 for 2 sec.
Note
2 sec. 2 sec. ○
Action: Adjustment of [PR1]
In adjustment is impossible, repair or replace [PR1].
Leak from [V10] Action: Repair or replacement of [V10]
Error occurrence at gas purge
○Occlusion of [R2]
○Excess tightening of [PC2]
○Impossible opening of [V4]
Action: Change [R2] so that dialysate pressure becomes 0 to
100mmHg during gas purge.
Action: Adjust [PC2] so that dialysate pressure becomes 0 to
100mmHg during gas purge.
Check: Check whether [V4] functions when it is turned on or off in
maintenance mode 4 screen.
If not functions, remove connector from [V4] and check whether
24V is outputted to the connector when [V4] is ON in
maintenance mode 4 screen.
Action:・If 24V is not outputted, replace main board with mew one.
・If it is outputted, replace [V4] with mew one.
Pressure rise due to folding of blood
Action: Correct folding.
line
Check: Disconnect hose from the sensor Assy to expose to
atmosphere for checking whether indicated pressure is nearly
0mmHg.
Failure of dialysate pressure sensor
board or the sensor
Action: If the indicated pressure is not nearly 0mmHg, calibrate the
sensor board.
If calibration is impossible, replace the board or sensor with new
one.
E07 DIALYSATE PRESSURE MAX LIMIT
Condition or cause Check and action
Disconnection of harness from the
sensor Assy or insufficient contact of
connector of it
Check: Pull lead wire (harness) lightly to check whether it is separated
or loosened.
Action: Repair
Drain tube is blocked.
○No folding/kinking of Drain tube
○Drain tube tip is placed too higher
position.
○Drain tube is too long.
Check:Remove drain tube from drain port and check drain liquid is
drained from port correctly or not.
Action:Remove the trouble cause.
Action:Drain tube tip must be placed above the machine floor by 0 to
60 cm.
Action:Use tube of an inter diameter of 8 mm and a length of 2 m or
shorter to connect to the port.
E08 DIALYSATE PRESSURE MIN LIMIT
Detection Dialysate pressure of ≦-400mmHg
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Auto reset of BP, SP and CLV
Reset
Press [RESET] SW to reset UF stop.
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
UF stop
ON of V4 and V10 for 2 sec.
Note
2 ses. 2 sec. ○
Condition or cause Check and action
Check: Check whether the line is folded or clamped by forceps
Occlusion of dialysate line to dialyzer
visually.
Action: Correct folding or clamp.
Excess occlusion in dialyzer Action: Replace dialyzer with new one.
Check: Check stop of leak by clamping [V4] line with forceps.
Leak from [V4]
Action: Clean [V4] by rinsing after disassembling or replace it with new
one.
Opening of [V4] due to malfunction of
[FSW]
Action: Clean air-elimination tank [AS2] by rinsing after disassembling
or replace it with new one.
Action: Adjust pressure in closed line.
Occlusion of [RV2]
If adjustment is impossible, Clean [RV2] by rinsing after
disassembling or replace it with new one.
Error occurrence during gas purge
○Failure of [V10]
Check: Check whether [V10] functions when it is turned on or off in
maintenance mode 4 screen.
If not functions, disconnect connector from [V10] and turn on it
in maintenance mode 4 screen to check whether 24V is
outputted to the connector.
Action: If 24V is not outputted, replace main board with new one.
If it is outputted, replace [V10] with new one.
Check: Disconnect hose from the sensor Assy to expose to
atmosphere for checking whether indicated pressure is nearly
0mmHg.
Failure of dialysate pressure sensor
board or the sensor
Disconnection of harness from the
sensor Assy or insufficient contact of
connector of it
Low water supply flow Action:Increase water supply flow rate more than 750 mL/min.
No water supply Check/Action:See E26 SHORT WATER SUPPLY.
Low water supply pressure Check/Action:See E26 SHORT WATER SUPPLY.
Alarm while gaspurge Check/Action:See E26 SHORT WATER SUPPLY.
Action: If the indicated pressure is not nearly 0mmHg, calibrate the
sensor board.
If calibration is impossible, replace the board or sensor with new
one.
Check: Pull lead wire (harness) lightly to check whether it is separated
or loosened.
Action: Repair
On the acid rince without an acid rince
option.
Check/Action:See E26 SHORT WATER SUPPLY.
E09 DIALYSATE PRESSURE MAX ALARM
Detection Dialysate pressure of ≧set upper alarm limit
Detection
process
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Auto reset of BP, SP and CLV
Reset
Press [RESET] SW to reset UF stop.
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (Continuous) UF stop
ON of V4 and V10 for 2 sec.
2 sec. 2 sec. ○
By the alarm actuation, upper/lower alarm limits of
Note
venous pressure become error free.
Condition or cause Check and action
Simultaneous actuation of this alarm
with venous pressure max alarm
Check: See E13 VENOUS PRESSURE MAX ALARM.
Action: See E13 VENOUS PRESSURE MAX ALARM.
Dialysate pressure rise due to UF stop Action: Remove cause of UF stop.
Dialysate pressure rise due to failure of
[PR1]
Action: Adjust [PR1].
If adjustment is impossible, repair [PR1] or replace it with new
one.
Leak from [V10] Action: Repair [V10] or replace it with new one.
Leak from [V5] and [V7] Action: Repair [V5] and [V7] or replace them with new ones.
E10 DIALYSATE PRESSURE MIN ALARM
Detection Dialysate pressure of ≦set lower alarm limit
Detection
process
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Auto reset of BP, SP and CLV
Reset
Press [RESET] SW to reset UF stop.
Audible
alarm
○ (Continuous) UF stop
Condition or cause Check and action
Simultaneous actuation of this alarm
with venous pressure min alarm
Occlusion of [RV2]
Leak from [V4]
Dialysate line function BP/SP stop CLV function Reset
2 sec. 2 sec. ○
ON of V4 and V10 for 2 sec.
By the alarm actuation, upper/lower alarm limits of
Note
venous pressure become error free.
Check: See E14 Venous pressure min alarm.
Action: See E14 Venous pressure min alarm.
Action: Adjust pressure in closed line.
If adjustment is impossible, clean [RV2] by rinsing after
disassembling or replace it with new one.
Check: Check no dialysate pressure change by clamping [V4] line with
forceps.
Action: In the case of no change, clean [V4] by rinsing after
disassembling or replace it with new one.
Opening of [V4] due to failure of [FSW]
The opening may be simultaneously
caused with E44 AIR ELIMINATION
TANK ERROR
Check: See E44 AIR ELIMINATION TANK ERROR
Action: See E44 AIR ELIMINATION TANK ERROR
Leak from [V6] and [V8] Action: Repair [V6] and [V8] or replace them with new ones.
Excess occlusion in dialyzer Action: Replace dialyzer with new one if required.
E11 VENOUS PRESSURE MAX LIMIT
Detection Venous pressure of ≧+400mmHg
Detection
process
Dialysis wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW to recover. Note
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
UF stop ○―○
Condition or cause Check and action
Folding or occlusion of blood line
between Venous Drip Chamber and
patient or occlusion of the line with
Action: Remove the trouble cause.
Robert clamp
Thrombus formed at mesh of venous
drip chamber or at shunt connector
Error occurrence during single needle
dialysis
Action: Remove thrombus or replace blood line with new one.
Action: Set upper pressure change limit of SN correctly.
Check: Disconnect blood line from venous pressure inlet port to
expose to atmosphere and check indicated venous pressure is
Failure of venous pressure sensor
board
nearly 0mmHg.
Action: If nearly 0mmHg is not indicated, calibrate the board.
If calibration is impossible, replace the board with new one.
E12 VENOUS PRESSURE MIN LIMIT
Detection Venous pressure of ≦-400mmHg
Detection
process
Dialysis wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW to recover. Note
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
UF stop
Condition or cause Check and action
Simultaneous occurrence of this error
with dialysate related error
Simultaneous occurrence of this error
with arterial pressure related error
Action: Remove cause of dialysate related error.
Action: Remove cause of arterial pressure related error
Press of DRAIN SW without removing
coupler of dialyzer under occluding
Note: DRAIN SW operation-miss
venous line with Robert clamp
Check: Disconnect blood line from venous pressure inlet port to
expose to atmosphere and check indicated venous pressure is
Failure of venous pressure sensor
board
nearly 0mmHg.
Action: If nearly 0mmHg is not indicated, calibrate the board.
If calibration is impossible, replace the board with new one.
○―○
E13 VENOUS PRESSURE MAX ALARM
Detection Venous pressure of ≧set upper alarm limit
Detection
process
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW to recover. Note
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (Continuous) UF stop ○ ― ○
Condition or cause Check and action
Check: Check shunt occlusion or blood line folding.
Occlusion of venous line.
Action: Remove the trouble.
Check: Check shunt occlusion or blood line folding.
Occlusion of arterial line
Action: Remove the trouble.
By the alarm actuation, upper/lower alarm limits of
dialysate pressure become error free.
Venous pressure change due to patient
movement
Significant discharge pressure change
of blood pump
Action: If the change is significant, raise somewhat upper alarm limit of
venous pressure at discretion of doctor.
Check: Check whether discharge flow rate, pressure applied to rolling
tube and rotor rotation of blood pump are normal.
Action: Adjust the occlusion or the pump rpm correctly.
If adjustment is impossible, replace the pump or sub-board
with new one.
E14 VENOUS PRESSURE MIN ALARM
Detection Venous pressure of ≦set lower alarm limit
Detection
process
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW to recover. Note
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (Continuous) UF stop ○ ― ○
Condition or cause Check and action
Blood pressure drop of patient
Action: Stop UF and monitor patient condition.
Correct if set lower alarm limit of venous pressure is incorrect.
Check: Check no introduction of blood in venous pressure sensor line.
Dislodgment of venous line from
venous pressure inlet port
Action: When blood is not introduced in the line, connect venous line
to the port. If introduced, replace the sensor line with new one.
By the alarm actuation, upper/lower alarm limits of
dialysate pressure become error free.
Occlusion of venous line
Occlusion of arterial line.
Venous pressure change due to patient
movement
Significant discharge pressure change
of blood pump
Check: Check shunt occlusion or venous line folding.
Action: Remove the trouble.
Check: Check shunt occlusion or arterial line folding.
Action: Remove the trouble.
Action: If the change is significant, lower somewhat lower alarm limit
of venous pressure at discretion of doctor.
Check: Check whether discharge flow rate, pressure applied to rolling
tube and rotor rotation of blood pump are normal.
Action: Adjust the occlusion or the pump rpm correctly.
If adjustment is impossible, replace the pump or sub-board
with new one.
E15 ARTERIAL PRESSURE MAX LIMIT
Detection Arterial pressure of ≧+400mmHg
Detection
process
Dialysis wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW to recover. Note
Audible
alarm
○ (7times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
UF stop ○―○
Condition or cause Check and action
Simultaneous actuation of this error
with dialysate pressure related error
Simultaneous actuation of this error
with venous pressure related error
Action: Remove cause of dialysate pressure related error.
Action: Remove cause of venous pressure related error.
Folding or occlusion of blood line
between arterial Drip chamber and
dialyzer or occlusion of the line with
Action: Remove the trouble cause.
Robert clamp
Folding or occlusion of blood line
between venous Drip chamber and
patient or occlusion of the line with
Action: Remove the trouble cause.
Robert clamp
Thrombus formed at mesh of arterial
drip chamber or
Action: Remove thrombus or replace blood line with new one.
Check: Disconnect blood line from arterial pressure inlet port to
expose to atmosphere and check indicated arterial pressure is
Failure of arterial pressure sensor
board
nearly 0mmHg.
Action: If nearly 0mmHg is not indicated, calibrate the board.
If calibration is impossible, replace the board with new one.
E16 ARTERIAL PRESSURE MIN LIMIT
Detection Arterial pressure of ≦-400mmHg
Detection
process
Dialysis wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW to recover. Note
Audible
alarm
○ (7times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
UF stop ○―○
Condition or cause Check and action
Simultaneous actuation of this error
with dialysate pressure related error
Thrombus formed at arterial shunt
connector
Action: Remove cause of dialysate pressure related error.
Action: Remove thrombus or replace blood line with new one.
Folding or occlusion of blood line
between arterial drip chamber and
patient or occlusion of the line with
Action: Remove the trouble cause.
Robert clamp
Check: Disconnect blood line from arterial pressure inlet port to
expose to atmosphere and check indicated arterial pressure is
Failure of arterial pressure sensor
board
nearly 0mmHg.
Action: If nearly 0mmHg is not indicated, calibrate the board.
If calibration is impossible, replace the board with new one.
E17 ARTERIAL PRESSURE MAX ALARM
Detection Arterial pressure of ≧set upper alarm limit
Detection
process
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW to recover. Note
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (Continuous) UF stop ○ ― ○
Condition or cause Check and action
Check: Check shunt occlusion and blood line folding.
Occlusion of venous line.
Action: Correct the trouble.
Check: Check shunt occlusion and blood line folding.
Occlusion of arterial line.
Action: Correct the trouble.
By the alarm actuation, set upper/lower alarm limits
of dialysate pressure become error free.
Arterial pressure change due to patient
movement
Significant discharge pressure change
of blood pump
Action: If the change is significant, raise set upper alarm limit of
arterial pressure somewhat at discretion of doctor.
Check: Check whether discharge flow rate, pressure applied to rolling
tube and rotor rotation of blood pump are normal.
Action: Adjust the occlusion or the pump rpm correctly.
If adjustment is impossible, replace the pump or sub-board
with new one.
E18 ARTERIAL PRESSURE MIN ALARM
Detection Arterial pressure of ≦set lower alarm limit
Detection
process
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW to recover. Note
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (Continuous) UF stop ○ ― ○
Condition or cause Check and action
Blood pressure drop of patient
Action: Stop UF and monitor patient condition.
Correct if set lower alarm limit of venous pressure is incorrect.
Check: Check no introduction of blood in arterial pressure sensor line.
Dislodgment of blood line form arterial
pressure inlet port
Action: When blood is not introduced in the line, connect arterial line
to the port. If introduced, replace the arterial line with new one.
By the alarm actuation, set upper/lower alarm limits
of dialysate pressure become error free.
Occlusion of venous line.
Occlusion of arterial line.
Arterial pressure change due to patient
movement
Significant discharge pressure change
of blood pump
Check: Check shunt occlusion and blood line folding.
Action: Correct the trouble.
Check: Check shunt occlusion and blood line folding.
Action: Correct the trouble.
Action: If the change is significant, raise set upper alarm limit of
arterial pressure somewhat at discretion of doctor.
Check: Check whether discharge flow rate, pressure applied to rolling
tube and rotor rotation of blood pump are normal.
Action: Adjust the occlusion or the pump rpm correctly.
If adjustment is impossible, replace the pump or sub-board
with new one
E19 TMP ALARM
Detection TMP (VP - DP) of ≧set alarm limit
Detection
process
Dialysis
Recovery
Reset Press [RESET] SW to recover. Note
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (Continuous) UF stop ○ ○ ○
Condition or cause Check and action
Excess occlusion in dialyzer Action: Replace dialyzer with new one at discretion of doctor.
Lower setting of TMP alarm limit Action: Set TMP to appropriate value.
E09 DIALYSATE PRESSURE MAX
ALARM
E10 DIALYSATE PRESSURE MIN
ALARM
E13 VENOUS PRESSURE MAX ALARM
Simultaneous actuation of this alarm
Action: See E09 DIALYSATE PRESSURE MAX ALARM
See E10 DIALYSATE PRESSURE MIN ALARM
See E13 VENOUS PRESSURE MAX ALARM
See E14 VENOUS PRESSURE MIN ALARM
with E14 VENOUS PRESSURE MIN
ALARM
E07 DIALYSATE PRESSURE MAX LIMIT
E08 DIALYSATE PRESSURE MIN LIMIT
E11 VENOUS PRESSURE MAX LIMIT
Simultaneous actuation of this alarm
with E12 VENOUS PRESSURE MIN
Action: See E07 DIALYSATE PRESSURE MAX LIMIT.
See E08 DIALYSATE PRESSURE MIN LIMIT.
See E11 VENOUS PRESSURE MAX LIMIT.
See E12 VENOUS PRESSURE MIN LIMIT.
LIMIT
E20 DIALYSATE CONDUCTIVITY MAX ALARM
Detection Dialysate conductivity of ≧upper alarm limit
Detection
process
Dialysis ○ (Continuous) Auto bypass ― ― ―
Press [RESET]SW.
Reset
Abnormal actual conductivity
○No adjustment of conductivity
○Failure of main board
○Failure of [V13] or [V14]
○Failure of [V2] or [V3]
○Higher or lower priming flow rate
○Failure of [C2]
○ Folding, occlusion or collapsing of
concentrate line tube
Auto reset after release from alarm condition for 10
sec.
Condition or cause Check and action
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Note
Check: Take dialysate sample from [SL] to measure actual
conductivity by specified method.
Action: Set conductivity correctly and calibrate indicated conductivity
in maintenance mode 5 screen.
Check: See E33 VISCOUS PUMP DRIVING ERROR(VPa) and
E34 VISCOUS PUMP DRIVING ERROR(VPb).
Action: See E33 VISCOUS PUMP DRIVING ERROR(VPa) and
E34 VISCOUS PUMP DRIVING ERROR(VPb).
Action: Repair [V13] or [V14] (V15)) or replace it with new one.
Action Repair [V2] or [V3] or replace it with new one.
Action: Adjust the flow rate to 700-750mL/min.
Action: Replace [C2] with new one.
Action: Correct the trouble.
E20 DIALYSATE CONDUCTIVITY MAX ALARM
Condition or cause Check and action
Normal actual conductivity
○[CD] electrode contaminated with dirt
such as CaCO3 or leak from
conductivity sensor
○Incorrect [CD] terminal connection
○Internal disconnection or short circuit
of [CD] harness
○Failure of main board
Check: Take dialysate sample from [SL] to measure conductivity by
specified method.
Action: Rinse the electrode with acetic acid followed by water rinsing
sufficiently.
If the dirt can not be removed, use specified detergent such
as CARON or SYUNMA to remove it after disassembling
conductivity sensor.
Check: Check no miss-wiring, loosening, corrosion and
disconnection of TM1 CD terminal of main board.
Action: If loosened, tighten the connection securely.
If miss-wired, correct.
If corroded or disconnected, replace harness with mew one.
Check: After disconnecting TM1 CD terminal from main board,
measure resistance between [CD] and [GND] to check no short
circuit, disconnection and fluctuation of measured resistance.
Action: Replace [CD] harness with new one.
Check: Change ambient temperature of main board to check whether
conductivity and temperature indications are changed.
Action: If incorrectly indicated, replace main board with new one.
When Na infusion unit (option) is used:
○Higher or lower concentration of Na
concentrate
○ Use of other chemical added
concentrate
○Use of other concentrate
Action: Use 10% Na concentrate.
Action: Set proper upper alarm limit of conductivity by considering
electrolyte concentration.
Action: Use 10% Na concentrate.
E21 DIALYSATE CONDUCTIVITY MIN ALARM
Detection Dialysate conductivity of ≦lower alarm limit
Detection
process
Dialysis ○ (Continuous) Auto bypass ―――
Press [RESET]SW.
Reset
Auto reset after release from alarm condition for 10
sec.
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Note
Condition or cause Check and action
See E20 DIALYSATE CONDUCTIVITY
See E20 DIALYSATE CONDUCTIVITY MAX ALARM.
MAX ALARM.
E22 BLOOD LEAK SENSOR ERROR
Absolute blood leak REAL value of ≧201 in PREP. or PREP. END process
TYPEⅠ(RK-GR1)
Detection
Detection
process
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET]SW. Note
Absolute blood leak REAL value of <201 in DIALYSIS or RECOVERY process
Signal of RINSE alarm
Audible
alarm
○ (Continuous)
Dialysate line function BP/SP stop CLV function Reset
Line function stop
E22 BLOOD LEAK SENSOR ERROR
Absolute blood leak REAL(BLD AUTO Z ERO) value of ≧201 in PREP. or PREP. END process
Detection
Detection
process
Signal of RINSE alarm
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
――○
BLD SENS VALUE :No7 in maintenance mode 1
Cancellation of ditection alarm is avairable with
ROM Ver3.30 at maintenance mode 2.
TYPEⅡ(RK-GR2)
Preparation
Prep. end
Dialysis
Recovery
Reset
○ (Continuous)
Press [RESET]SW
Line function stop
E22 BLOOD LEAK SENSOR ERROR
Absolute blood leak REAL(BLD AUTO Z ERO) value of ≧201 in PREP. or PREP. END process
Detection
Detection
process
Preparation
Prep. end
Press [RESET]SW
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (Continuous)Line function stop― ― ○
― ― ○
BLD AUTO ZERO:№6 in maintenance mode1
Note
Cancellation of ditection alarm is avairable with
ROM Ver3.30 at maintenance mode 2.
TYPEⅡ(BLD-01)
BLD AUTO ZERO:№6 in maintenance mode1
Reset
Note
Cancellation of ditection alarm is avairable with
ROM Ver3.30 at maintenance mode 2.
Condition or cause Check and action
Excess leak from dialyzer
Excess dirt of blood leak sensor window
(RK-GR1,RK-GR2,BLD-01)
Inclusion of large amount of air in blood
leak sensor
Incorrect connection of blood leak
sensor connector
Action: Replace dialyzer with new one.
CAUTION
After setting to error free condition, error is continued when leaked
blood is retained in drain side of viscous chamber.
In such case, release the alarm to operate the machine until the
blood is discharged. If required, clean leak sensor window by wiping
with swab or the like after opening the sensor cover.
Action: Clean the window by wiping with swab or the like after
opening the sensor cover.
Check: Take dialysate sample from drain port of the machine or
dialysate outlet of dialyzer to check blood in the sample
visually or by use of specified tester.
Check: Check visually whether fine air bubbles impossible to be
removed by [AS1] and [AS2] remain in them and whether
negative pressure line (example: pump inlet) in the machine is
leaked.
Action: Repair damaged part or replace damaged component with
new one.
Check: Pull harness lightly to find disconnection and loose
connection and check defect of connector.
Action: Replace harness with new one.
Deterioration or failure of LED of blood
leak sensor
Leak of silicone oil from chamber
Action: Readjust the sensor. If adjustment is impossible, replace the
sensor Assy with new one.
Action. Replace chamber Assy with new one.
Clean chamber inside with detergent to remove dirt formed by
contact with the oil from there.
Replace silicone oil line and/or connectors contaminated by the
oil leak with new one(s) when required.
E23 BLOOD LEAK ALARM
Detection
Detection
process
Dialysis ○ (Continuous) UF stop
Reset Press [RESET]SW. Note
BLD REAL VALUE of ≧set alarm limit
(BLD REAL VALUE: No.5 of maintenance mode 1, Alarm limit: No.12 of maintenance mode 2)
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Auto bypass
Condition or cause Check and action
Check: Take dialysate sample from drain port of the machine or
dialysate outlet of dialyzer to check blood in the sample
visually or by use of specified tester.
Blood leak
Action: If leaked from dialyzer:
After shifting to RECOVERY process, press BYPASS SW and
replace dialyzer with new one followed by priming it.
○ ― ○
No blood leak
○Air inclusion
○Dirt of blood leak sensor window
(Deposition of protein and/or CaCO3)
○ Deterioration or failure of LED of
blood leak sensor
○Failure of main board
○Leak of silicone oil
Check: Take dialysate sample from drain port of the machine or
dialysate outlet of dialyzer to check blood in the sample
visually or by use of specified tester.
Check: Check visually whether fine air bubbles impossible to be
removed by [AS1] and [AS2] remain in them and whether
negative pressure line (example: pump inlet) in the machine is
leaked.
Action: Repair damaged part or replace damaged component with
new one.
Action: Clean the window.
Action: Readjust the sensor. If adjustment is impossible, replace the
sensor Assy with new one.
Action: Replace the sensor with new one and check correct function
of it..
If the replacement is ineffective, replace main board with new
one.
Check: Remove cover from the sensor and check whether dialysate is
turbid or is mixed with the oil.
Action. Replace chamber Assy with new one.
Clean chamber inside with detergent to remove dirt formed by
contact with the oil from there.
Replace silicone oil line and/or connectors contaminated by
the oil leak with new one(s) when required.
E24 AIR BUBBLES ARE DETECTED
Detection Air bubbles when BUBBLE [ON/OFF] SW is ON
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Press [RESET] SW to reset UF stop and CLV.
Reset
Reset BP/SP by SW-ON.
Audible
alarm
○ (7times)
↑
↑
↑
○
(Continuous)
↑
↑
↑
↑
Condition or cause Check and action
Air bubbles are detected:
○Air introduction from blood line
Dialysate line function BP/SP stop CLV function Reset
UF stop ○
SW-OFF
Detection of air bubbles only when BUBBLE
Note
[ON/OFF] SW is ON.
○○
Action: Correct air introduction site and collect air in drip chamber.
Air bubbles are not detected:
○Accidental dislodgment of blood line
from air bubble sensor
○Malfunction due to external noise
○Disconnection of the sensor harness
○Failure of air bubble sensor board
Action: Return the line to the sensor.
Action: Remove source of noise or place the source at distant
position.
Action: Replace the sensor with new one.
Action: Adjust the board.
If adjustment is impossible, replace the board or the sensor
with new one.
E25 HEATER OVER HEAT ALARM
Detection Functioning of [LS1] placed at outer wall of heater
Detection
process
Preparation
Prep. End
Dialysis
Recovery
Reset Press [RESET] SW. Note
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (Continuous) Heater OFF ―――
Condition or cause Check and action
Heater overheat
Check: Measure voltage at heater side of TERMINAL BLOCK 3 using
tester to find whether 230V (or 110V) is not applied always.
○Failure of [T1], [T2] or [T3]
○Failure of sub-board
○Failure of SSR
○Failure of RELAY
No heater overheat
○Failure of [LS1]
○ Disconnection of lead wires
connected to [LS1] or incorrect
connector connection
○Failure of main board
DO NOT touch heater with hand because heater temperature
exceeds 60℃ when heater SW is ON.
Action: See E02 T1 TEMPERATURE ERROR;
E03 T2 TEMPERATURE ERROR; and
E04 T3 TEMPERATURE ERROR
Action: Replace sub-board with new one.
Action: Replace SSR with new one.
Action: Replace RELAY with new one.
Check: Check closed contact of [LS1] at room temperature. If
possible, check opening of the contact when heater surface
temperature is about 60℃.
Action: Replace [LS1] with new one if malfunction of [LS1] is
detected.
Check: Pull lightly lead wires of CN19 and those of connector of
heater side to check disconnection or loosening of the wires.
Action: Repair.
Action, If [LS1] is normal, replace main board with new one.
CAUTION
E26 SHORT WATER SUPPLY
Detect alarm depends on the time of Water priming complete.
Disabled detection of PS (Pressure switch) within 160 sec after starts the Preparation.
Detection
Detection
process
Rinse
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Disabled detection of PS within 60 sec after suction of B liquid on Flesh Dialysate mixing cycle. (After
suction of A liquid in case of Acetate Dialysis)
Disabled detection of PS within 65 sec after complete the UF leak check procedure.
Audible
alarm
○ (7times)
○
(Continuous)
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
Line function stop
Condition or cause Check and action
No water supply
○Occlusion of water supply filter
○Folding or occlusion of water supply
line
○Leak from water supply system
○Water supply stop
Check: Remove water supply port to check normal water flow.
Action: Clean the filter or replace it with new one.
Action: Correct the trouble.
Action: Repair leaked site.
Action: Operate water supply equipment.
― ― ○
Low water supply pressure
○Low water supply pressure
○Incorrect adjustment or failure of [PR]
○ Incorrect adjustment or failure of
[RV1]
○Incorrect adjustment or failure of [PS]
○Failure of main board
○Alarm while gaspurge
○On the acid rince without an acid rince
option.
Check: Remove water supply port to check normal water flow.
Action: Raise the pressure to ≧0.049MPa(0.5kgf/cm2.)
Action: Adjust [PR] correctly.
If adjustment is impossible, repair [PR].
Action: Adjust priming completion pressure correctly by operating
[RV1].
If adjustment is impossible, repair [RV1].
Action: Adjust [PS] correctly.
If adjustment is impossible, repair [PS].
Check: Remove connector from [PS] and check ON and OFF of [PS]
by short circuiting and opening connector pins respectively in
maintenance mode 4 screen.
Action: If ON and OFF are not executed, replace main board with new
one
Check: Check the capacity of a RO supply against number of Dialysis
machine.
Action: Adjust the ratio of RO supply capacity to a number of
Dialysis machine.
Check: Check a mode of acid rince option .
Action: Set a mode of acid rince option to off.
E27 SHORT DIALYSATE FLOW
Detection No functioning of [FS] after 85 seconds following chamber changeover
Detection
process
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Audible
alarm
○
(Continuous)
Dialysate line function BP/SP stop CLV function Reset
― ― ― ―
Condition or cause Check and action
Normal dialysate flow (at set rate)
○Low dialysate flow rate
○ Disconnection of lead wires
connected to [FS] or incorrect
connector contact
○Failure of [FS]
○Failure of main board
○Adjust the flow rate to 300 to 600ml/min.
Check: Pull lightly lead wires of CN15 placed on main board and
those of connector of [FS] side to check disconnection or
loosening of the wires.
Action: Repair.
Check: Measure 24V power supply and output voltages of [FS] and
pin voltage of flow sensor.
・Power supply voltage between + and -: about 24V
・Output voltage between OUT and -:
About 24V when float is UP condition
About 0V when float is DOWN condition
Action: Adjust positions of [FS] and [FM].
If adjustment is impossible, replace [FS] or flow meter Assy
with new one.
Check: Measure 24V power supply and output voltages of [FS] and
pin voltage of flow sensor
Action: If voltage between + and - is not about 24V, replace main
board with new one.
Check: If [FS] is normal, check indication of [FS] in maintenance
mode 4 screen or FS function indicating lamp window.
・Float UP condition: OFF in maintenance mode 4 screen
OFF of FS function indicating lamp window
・Float DOWN condition: ON in maintenance mode 4 screen
ON of FS function indicating lamp window
Action: Replace main board with new one.
E27 SHORT DIALYSATE FLOW
Condition or cause Check and action
No dialysate flow
○Occlusion in closed circuit
○Failure of [P2]
○Failure of one or more of [V5] to [V9]
Check: Check no dialysate flow by [FM] placed at front side of the
machine.
Action: Repair occluded component or replace it with new one.
Action: Repair [P2] or replace it with new one.
Action: Repair failed valve or replace it with new one.
E28 DIALYSATE FLOW STOP
Detection
Detection process Audible
Rinse (water
rinse)
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Functioning of [FS] within 20 seconds after chamber changeover
Functioning of [FS] for 170 seconds in DIALYSIS process excluding SQHD
Alarm
○ (7 times)
○ (Continuous)
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
Line function stop ― ― ―
Condition or cause Check and action
Normal dialysate flow
○High dialysate flow rate
○Low dialysate flow rate
○This alarm occur just after gas purge
function.
○ Disconnection of lead wires
connected to [FS] or incorrect
connector contact
○Failure of [FS]
○Failure of main board
Action: Adjust the flow rate to 300 to 600mL/min.
Check: Float of Flow meter cover the FS (Flow sensor) or not by
checking maintenance mode 4.
Action: Adjust the flow rate to 300 to 600mL/min.
Check: Detect the FS by decreasing dialysate flow to less than 300
mL/min by gas purge function under low dialysate flow condition.
Action: Adjust the flow rate to 300 to 600mL/min.
Check: Pull lightly lead wires of CN15 placed on main board and
those of connector of [FS] side to check disconnection or
loosening of the wires.
Action: Repair.
Check: Measure 24V power supply and output voltages of [FS] and
pin voltage of flow sensor.
・Power supply voltage between + and -: about 24V
・Output voltage between OUT and -:
About 24V when float is UP condition
About 0V when float is DOWN condition
Action: Replace [FS] or flow meter Assy.
Check: Measure 24V power supply and output voltages of [FS] and
pin voltage of flow sensor
Action: If voltage between + and - is not about 24V, replace main
board with new one.
Check: If [FS] is normal, check indication of [FS] in maintenance
mode 4 screen or FS function indicating lamp window.
・Float UP condition: OFF in maintenance mode 4 screen
OFF of FS function indicating lamp window
・Float DOWN condition: ON in maintenance mode 4 screen
ON of FS function indicating lamp window
Action: Replace main board with new one.
E28 DIALYSATE FLOW STOP
Condition or cause Check and action
Connector is connected to Dialyzer on
preparation process then Dialysate
flow is auto-bypassed.
After connector connected to
Dialyzer ,Turn BP on and
unbypass.
Too much of flow at [V10]
Wrong dialyzer in use
No dialysate flow
○Reduced chamber capacity
○Inclusion of large amount of air in
closed circuit
○Low priming flow rate
○Incorrect adjustment of [PS]
○Occlusion in closed circuit
○Failure of [P2]
○Failure of one or more of [V5] to [V9]
Action: Connect the connector to the Dialysate on preparation end
process.
Action: BP should be turned off while priming a Dialysis.
Check: There is [R3] at [V10]out line tube or not.
Action: Put [R3] and change the ROM later than Ver.3
Then input D.PRESS offsetting.
Action: Change to the correct dialyzer.
Check: Check no dialysate flow by [FM] placed at front side of the
machine.
Check: Measure chamber capacity.
Action: If the capacity is not 345 to 375mL, add silicone oil or
replace the chamber Assy with new one.
Check: Check air inclusion in the line visually.
Action: Repair air included component or replace it with new one.
Action: Adjust the flow rate to 700 to 750mL/min.
Action: Adjust [PS].
If adjustment is impossible, replace [PS] with new one.
Action: Repair occluded component or replace it with new one.
Action: Repair [P2] or replace it with new one.
Action: Repair failed valve or replace it with new one.
E29 SN SWITCH TIME ALARM
Detection No arrival at SN switching pressure after set alerting time
Detection
process
Dialysis
Reset Press [RESET] SW. Note Enabled alerting only when SN SW is ON
Audible
alarm
○
(Continuous)
Dialysate line function BP/SP stop CLV function Reset
UF stop ○○○
Condition or cause Check and action
Slow venous pressure rise
○ Large gap between rollers and
housing of blood pump
○Folding or fatigue of rolling tube
○Large clamp gap
○Incorrect connection of [CN18] clamp
connector
Check: Check whether the pressure rise is normal by operating blood
pump.
Action: Adjust the gap so that rolling tube is occluded properly.
Action: Replace blood line including rolling tube
Action: Adjust clamp gap.
Check: Remove connector to check corrosion and disconnection.
Action: Replace harness with new one.
No functioning of clamp
Check: See E32 CLV POSITION ERROR.
Action: See E32 CLV POSITION ERROR.
Normal venous pressure rise/lowering
○ Incorrect blood flow rate, SN
operation range or set SN alerting
time
Check: Observe venous pressure change during SN operation.
Check: Measure SN switching time.
Action: Set properly.
E30 BLOOD PUMP ERROR
Overload
Detection
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Reset
Abnormal rotating direction
Abnormal rpm
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Press [RESET] SW.
Turn on [BP/SP] SW.
Dialysate line function BP/SP stop CLV function Reset
―○
SW-OFF
Note
――
Condition or cause Check and action
Motor shaft lock
○Catch of foreign matters by rotor of
blood pump
○Damage of motor gear head
○Failure of sub-board
No motor rotation
○ Disconnection of lead wires
connected to [BP] or incorrect
connector contact
○failure of blood pump motor
○Failure of sub-board
Action: Turn off power to remove foreign matters.
Action: Replace blood pump Assy with new one.
Action: Replace sub-board with new one.
Check: Pull lightly lead wires of CN415 placed on sub-board and
those of connector of [BP] side to check disconnection and loosening
of the wires.
Action: Connect or tighten securely.
Action: Replace blood pump Assy with new one.
Action: Replace sub-board with new one.
E31 BLOOD PUMP STOP ERROR
Opening of blood pump cover
Detection
Detection
process
Dialysis
Reset
OFF of [ON/OFF] SW of blood pump
0 setting of flow rate setting volume of blood pump.
Audible
alarm
○
(Continuous)
Press [RESET] SW.
Auto reset by release from error
Dialysate line function BP/SP stop CLV function Reset
UF stop SP stop
Condition or cause Check and action
Opening of blood pump cover
○Opening of the cover
○ Incorrect setting of rolling tube to
blood pump
○ Impossible closing of blood pump
cover due to deformation of hinge
shaft.
○No press of SW by pin of blood pump
cover due to loose holding of hinge
shaft
Closing of blood pump cover
○Failure of [BPC]
○ Disconnection of lead wires
connected to [BPC] or incorrect
connector contact
○Failure of sub-board
― ―
Note
Action: Close the cover.
Action: Set properly.
Action: Correct the shaft deformation or replace the shaft with new
one.
Action: Tighten screws used to hold hinge shaft securely.
Check: Disconnect CN403 from sub-board to measure resistance
between harness pins.
・When the cover is closed: short circuit
・When the cover is opened: open circuit
Check: Check whether resistance between harness pins is open
condition when the cover is closed.
Action: Adjust position of [BPC].
If adjustment is impossible, replace harness with new one.
Check: Pull lightly lead wires of CN403 placed on sub-board to check
disconnection and loosening of the wires.
Action: Replace harness with new one.
Action: Replace sub-board with new one.
E32 CLV POSITION ERROR
Detection Abnormal clamp positioning
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Prep. end
Dialysis
Recovery
Reset
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Press [RESET] SW.
Turn on [BP/SP] SW.
Condition or cause Check and action
Clamped condition
○Failure of main board
○Failure of solenoid
Audible
alarm
Dialysate line function BP/SP stop CLV function Reset
―○
SW-OFF
Note
――
Check: Check whether [CLV] is OFF in maintenance mode 4 screen
and also check clamp head positioning visually.
Check: Measure output voltage between 1 and 2 of CN18 placed on
main board.
Action: If about 24V is outputted, replace main board with new one.
Action: If about 24V is not outputted, replace solenoid with new one.
Non-clamped condition
○Incorrect clamp positioning
○Incorrect positioning of [LS5] or screw
for ON of [LS5]
○Failure of [LS5]
○Failure of main board
Check: Check whether [CLV] is ON in maintenance mode 4 screen
and also check clamp head positioning visually.
Check: Check clamp position visually.
Action: Correct the position.
Check: Check whether [LS5] is turned on by screw for ON visually.
Action: Adjust position of screw for ON.
Check: Disconnect CN18 from main board to measure resistance
between pins 5 and 6.
Turn on and off [LS5] manually.
・When [LS5] is ON : short circuit
・When [LS5] is OFF : open circuit
Action: Replace harness with new one.
Action: Replace main board with new one.
E33 VISCOUS PUMP DRIVING ERROR(VPa)
Disable to change OFF to ON of PH2 (Rotation pulse signal) during a specified time (depend on UFR) at
the UF circle.
Detection
Detection
process
Rinse
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Disable to change OFF to ON of PH1 (Original position signal) during 20 sec at the procedure of Original
position.
Detect opposite rotation during VPa working.
Audible
Alarm
○ (7 times)
○
(Continuous)
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
Line function stop
Condition or cause Check and action
― ― ○
Normal rotation of viscous pump with
stop at error detection
○Failure of rotation detector of sensor
circuit board
○ Failure of original point detector of
sensor circuit board
○ Disconnection of [PHA] harness or
incorrect connector contact
○ Failure of main board
○ Occur with E35 VP DRAWING
VOLUME IS AT MAX LIM.
Check: Disconnect connector from viscous pump and check whether
lower photo-sensor responds to manual slow rotation of
position detection plate under monitoring PH2 of maintenance
mode 4 screen.
If the sensor does not respond, disconnect connector from the
board and measure power supply voltage between pins 1 and
4 to find whether it is about 12V.
Connect the connector and measure voltage between test
points V and G of the board. When the voltage is about 12V,
check whether Hi - Lo cycle is repeated during the pump
rotation by the sensor output between test points OUT2 and G.
Action: If voltage between test points V and G of the board is not
about 12V or that between test points OUT2 and G of the board
is abnormal, replace the board with new one.
Check: Check whether upper photo-sensor responds to manual slow
rotation of position detection plate under monitoring PH1 of
maintenance mode 4 screen.
If the sensor does not respond, disconnect connector from the
board and measure power supply voltage between pins 1 and
4 to find whether it is about 12V.
Connect the connector and measure voltage between test
points V and G of the board. If the voltage is about 12V, check
whether the sensor output between test points OUT1 and G is
Lo at detection of original point and Hi at no detection of the
point.
Action: If voltage between test points V and G of the board is not
about 12V or the sensor output between test points OUT1 and
G of the board is abnormal, replace the board with new one.
Check: Pull lightly lead wires of CN7 placed on main board to check
disconnection and loosening of the wires.
Action: Replace the harness with new one.
Action: If the sensor circuit board is normal, replace main board with
new one.
Action: Refer to E35. then push the reset button two or three times.
E33 VISCOUS PUMP DRIVING ERROR(VPa)
Condition or cause Check and action
Disabled motor rotation or uneven
rotation
○Insufficient motor torque
○High water supply pressure
○High secondary pressure
○High priming pressure
○ Disconnection of lead wired
connected to [Vpa] or incorrect
connector contact
○Failure of sub-board
○Failure of VPa
○Sensor Guide Plate crosses over PH1
by some causes.
Check: Check whether motor torque is sufficient.
Press position detection plate lightly by hand during motor
rotation or at motor start to find whether motor stop is caused.
Action: Replace chamber Assy.
Action: Set the pressure to 0.5 to 7.5MPa(kgf/cm2).
Action: Adjust [PR].
If adjustment is impossible, repair [PR] or replace it with mew
one.
Action: Adjust [RV1].
If adjustment is impossible, replace [RV1] with new one.
Check: Pull lightly lead wires of CN409 placed on sub-board and
those of connector of Vpa side to check disconnection and
loosening of the wires.
Action: Correct the disconnection or loosening.
Action: Replace sub-board with new one.
Action: Replace [Vpa] or chamber Assy with new one.
Action: Disconnect connector from Viscous pump and rotate Position
Detection Plate manually lower than detect position of PH1.
E34 VISCOUS PUMP DRIVING ERROR(VPb)
Disable to change OFF to ON of PH4 (Rotation pulse signal) during a specified time (depend on UFR) at
the UF circle.
Detection
Detection
process
Rinse
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Disable to change OFF to ON of PH3 (Original position signal) during 20 sec at the procedure of Original
position.
Detect opposite rotation during VPb working.
Audible
alarm
○ (7 times)
○ (Continuous)
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
Line function stop
Condition or cause Check and action
― ― ○
Normal rotation of viscous pump with
stop at error detection
○ Failure of rotation detector of sensor
circuit board
○ Failure of original point detector of
sensor circuit board
○ Disconnection of [PHS] harness or
incorrect connector contact
○ Failure of main board
○ Occur with E35 VP DRAWIN
VOLUME IS AT MAX LIM
Check: Disconnect connector from viscous pump and check whether
lower photo-sensor responds to manual slow rotation of
position detection plate under monitoring PH4 of maintenance
mode 4 screen.
If the sensor does not respond, disconnect connector from the
board and measure power supply voltage between pins 1 and
4 to find whether it is about 12V.
Connect the connector and measure voltage between test
points V and G of the board. When the voltage is about 12V,
check whether Hi - Lo cycle is repeated during the pump
rotation by the sensor output between test points OUT2 and G.
Action: If voltage between test points V and G of the board is not
about 12V or that between test points OUT2 and G of the board
is abnormal, replace the board with new one.
Check: Check whether upper photo-sensor responds to manual slow
rotation of position detection plate under monitoring PH3 of
maintenance mode 4 screen.
If the sensor does not respond, disconnect connector from the
board and measure power supply voltage between pins 1 and
4 to find whether it is about 12V.
Connect the connector and measure voltage between test
points V and G of the board. If the voltage is about 12V, check
whether the sensor output between test points OUT1 and G is
Lo at detection of original point and Hi at no detection of the
point.
Action: If voltage between test points V and G of the board is not
about 12V or the sensor output between test points OUT1 and
G of the board is abnormal, replace the board with new one.
Check: Pull lightly lead wires of CN8 placed on main board to check
disconnection and loosening of the wires.
Action: Replace the harness with new one.
Action: If the sensor circuit board is normal, replace main board with
new one.
Action: Refer to E35. then push the reset button two or three times.
E34 VISCOUS PUMP DRIVING ERROR(VPb)
Condition or cause Check and action
Disabled motor rotation or uneven
rotation
○Insufficient motor torque
○High water supply pressure
○High secondary pressure
○High priming pressure
○ Disconnection of lead wired
connected to [Vpb] or incorrect
connector contact
○Failure of sub-board
○Failure of VPb
○Sensor Guide Plate crosses over PH3
by some causes
Check: Check whether motor torque is sufficient.
Press position detection plate lightly by hand during motor
rotation or at motor start to find whether motor stop is caused.
Action: Replace chamber Assy.
Action: Set the pressure to 0.5 to 7.5MPa(kgf/cm2).
Action: Adjust [PR].
If adjustment is impossible, repair [PR] or replace it with mew
one.
Action: Adjust [RV1].
If adjustment is impossible, replace [RV1] with new one.
Check: Pull lightly lead wires of CN409 placed on sub-board and
those of connector of Vpb side to check disconnection and
loosening of the wires.
Action: Correct the disconnection or loosening.
Action: Replace sub-board with new one.
Action: Replace [Vpb] or chamber Assy with new one.
Action: Disconnect connector from Viscous pump and rotate Position
Detection Plate manually lower than detect position of PH3.
E35 VP DRAWING VOLUME IS AT MAX LIM.
Detection Out of operable range of viscous pump
Detection
process
Rinse
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Audible
alarm
○ (7times)
○ (Continuous)
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
Line function stop ――○
Condition or cause Check and action
Check: Measure chamber switching time using stopwatch under
monitoring flow meter indication.
Execution of high UF at low dialysate
flow rate
Action:ROM Ver 3.60 and older
Adjust UFR or dialysate flow rate so that [UFR (L/h)]×[Chamber
Switching Time] becomes <364.
ROM Ver 3.70 and later
Adjust UFR or dialysate flow rate so that [UFR (L/h)]×[Chamber
Switching Time] becomes <260.
Significant deviation of input values of
viscous pump flow rates
Check: Check whether values of [11.VPa FLOW] and [12.VPb FLOW]
are correctly inputted in maintenance mode 6 screen.
Significant deviation of input values of
chamber capacities
No chamber switching for long time due
to failure of [FS]
Refer
E33 VISCOUS PUMP DRIVING
ERROR(VPa)
E34 VISCOUS PUMP DRIVING
ERROR(VPb)
Check: Check whether values of [9.VCa VOLUME] and [10.VCb
VOLUME] are correctly inputted in maintenance mode 6 screen.
Check: See E27 SHORT DIALYSATE FLOW.
Action: See E27 SHORT DIALYSATE FLOW.
Check: See E33 VISCOUS PUMP DRIVING ERROR(VPa) and
E34 VISCOUS PUMP DRIVING ERROR(VPb).
Action: See E33 VISCOUS PUMP DRIVING ERROR(VPa) and
E34 VISCOUS PUMP DRIVING ERROR(VPb).
After removing the above error causes, press [RESET] SW
twice or thrice.
E36 VCa CLOSED CIRCUIT LEAK ERROR
Detection
Detection
process
Rinse
(Water rinse)
Reset Press [RESET] SW. Note
Leak from closed circuit
Leak from NC side of [V5] or [V6] Action: Repair leaked [V5] or [V6] or replace it with new one.
Leak from NO side of [V7] or [V8] Action: Repair leaked [V5] or [V6] or replace it with new one.
Insufficient negative pressure in A chamber
No arrival at set pressure within set time
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (7 times) ―――○
Enabled release from this error in maintenance
mode 2 .
Condition or cause Check and action
Check: Check leak from closed circuit visually or by contact the
circuit with finger.
Action: Replace leaked component with new one.
Detection of leak error in both A and B
chambers
○Leak from [V3] or [V4]
○Leak from coupler
○ Insufficient deaeration performance
of water supply
○Gas purge pressure setting is wrong .
Action: Identify leaked site(s).
Operate the machine in RINSE or DIALYSIS PREP. process to
apply negative pressure to C2 line.
Shift to DIALYSIS WAIT process.
Turn on and off each of valves [V5] to [V8] manually in
maintenance mode 4 screen and operate viscous pump
manually after removing connector from the pump.
Clamp IN or OUT side of dialysate pressure sensor with forceps
to find whether leak is caused from V5,7 or V6,8 side.
Repeat clamp position change to find other leaked sites.
Action: Repair leaked [V3] or [V4] or replace it with new one.
Action: Replace O-ring of coupler or coupler including O-ring with
new one.
Check: Measure dissolved oxygen % in water.
Action: Improve deaeration performance of the means or change set
closed circuit leak detection value properly.
Action: After input D.PRESS offset appropriately, adjust [PC2]to make
gas purge pressure 0 ~ 100 mmHg .
E37 VCa CLOSED CIRCUIT LEAK ERROR
Detection
Detection
process
Rinse
(Water rinse)
Reset Press [RESET] SW. Note
Negative pressure change in A chamber
Pressure change out of set range during set time
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (7 times) ――○
Condition or cause Check and action
See E36 VCa CLOSED CIRCUIT LEAK
ERROR
See E36 Vca CLOSED CIRCUIT LEAK ERROR.
Enabled release from this error in maintenance
mode2.
E38 VCb CLOSED CIRCUIT LEAK ERROR
Detection
Detection
process
Rinse
(Water rinse)
Reset Press [RESET] SW. Note
Insufficient negative pressure in B chamber
No arrival at set pressure within set time
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (7 times) ―――○
Enabled release from this error in maintenance
mode 2 .
Condition or cause Check and action
Check: Check leak from closed circuit visually or by contact the
Leak from closed circuit
circuit with finger.
Action: Replace leaked component with new one.
Leak from NC side of [V7] or [V8] Action: Repair leaked [V7] or [V8] or replace it with new one.
Leak from NO side of [V5 or [V6] Action: Repair leaked [V5] or [V6] or replace it with new one.
Detection of leak error in both A and B
chambers
○Leak from [V3] or [V4]
○Leak from coupler
○Insufficient deaeration performance of
water supply
Action: Identify leaked site(s).
Operate the machine in RINSE or DIALYSIS PREP. process to
apply negative pressure to C2 line.
Shift to DIALYSIS WAIT process.
Turn on and off each of valves [V5] to [V8] manually in
maintenance mode 4 screen and operate viscous pump
manually after removing connector from the pump.
Clamp IN or OUT side of dialysate pressure sensor with forceps
to find whether leak is caused from V5,7 or V6,8 side.
Repeat clamp position change to find other leaked sites.
Action: Repair leaked [V3] or [V4] or replace it with new one.
Action: Replace O-ring of coupler or coupler including O-ring with
new one.
Check: Measure dissolved oxygen % in water.
Action: Improve deaeration performance of the means or change set
closed circuit leak detection value properly.
E39 VCb CLOSED CIRCUIT LEAK ERROR
Detection
Detection
process
Rinse
(Water rinse)
Reset Press [RESET] SW. Note
Negative pressure change in B chamber
Pressure change out of set range during set time
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (7 times) ―――○
Condition or cause Check and action
See E38 VCb CLOSED CIRCUIT LEAK
ERROR.
See E38 VCb CLOSED CIRCUIT LEAK ERROR.
Enabled release from this error in maintenance
mode 2 .
E40 VCa CLOSED CIRCUIT LEAK ERROR
Detection
Detection
process
Rinse
(Water rinse)
Reset Press [RESET] SW. Note
Insufficient positive pressure in A chamber
No arrival at set pressure within set time
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (7times) ―――○
Condition or cause Check and action
Occlusion of water supply filter
See E36 VCa CLOSED CIRCUIT LEAK
ERROR.
Action: Wash up filter or change to a new filter
See E36 Vca CLOSED CIRCUIT LEAK ERROR.
Enabled release from this error in maintenance
mode 2 .
E41 VCa CLOSED CIRCUIT LEAK ERROR
Detection
Detection
process
Rinse
(Water rinse)
Reset Press [RESET] SW. Note
Positive pressure change in A chamber
Pressure change out of set range during set time
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (7 times) ―――○
Condition or cause Check and action
See E36 VCa CLOSED CIRCUIT LEAK
ERROR.
See E36 Vca CLOSED CIRCUIT LEAK ERROR.
Enabled release from this error in maintenance
mode 2 .
E42 VCb CLOSED CIRCUIT LEAK ERROR
Detection
Detection
process
Rinse
(Water rinse)
Reset Press [RESET] SW. Note
Insufficient positive pressure in B chamber
No arrival at set pressure within set time
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (7 times) ―――○
Condition or cause Check and action
Occlusion of water supply filter
See E38 VCb CLOSED CIRCUIT LEAK
ERROR.
Action: Wash up filter or change to a new filter
See E38 VCb CLOSED CIRCUIT LEAK ERROR.
Enabled release from this error in maintenance
mode 2 .
E43 VCb CLOSED CIRCUIT LEAK ERROR
Detection
Detection
process
Rinse
(Water rinse)
Reset Press [RESET] SW. Note
Positive pressure change in B chamber
Pressure change out of set range during set time
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (7 times) ―――○
Condition or cause Check and action
See E38 VCb CLOSED CIRCUIT LEAK
ERROR.
See E38 VCb CLOSED CIRCUIT LEAK ERROR.
Enabled release from this error in maintenance
mode 2 .
E44 AIR ELIMINATION TANK ERROR
Detection
Detection
process
Rinse
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Functioning of air elimination tank [FSW] for ≧15 seconds in DIALYSIS process or for ≧30 seconds in any
process other than DIALYSIS process
Audible
alarm
○ (7times)
○ (Continuous)
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
―
↑
↑
Line function stop
―
Condition or cause Check and action
Lowered float positioning
○Increased float density due to water
absorption
○ Lowered float positioning due to
attachment of foreign matters to
float and float shaft
○No air discharge due to failure of [V4]
○Occlusion of [R2] by foreign matters
○ No air discharge due to failure of
electric circuit to drive [V4]
Check: Check function of [FSW] in maintenance mode 4 screen.
When ○ is indicated, float is lowered position.
Check: Measure float weight.
Float weight exceeding 8g is unacceptable.
Action: Replace float with new one.
Action: Clean float and float shaft to remove foreign matters.
Action: Repair [V4] or replace it with new one.
Check: Measure gas purge pressure which shall be 0 to 100mmHg.
Action: Clean [R2] by rinsing or replace it with new one.
Check: Turn on [V4] manually to check whether 24V is outputted to
[V4] in maintenance mode 4 screen.
Action: If 24V is not outputted, replace main board with new one.
― ― ○
○Wrong dialyzer in use
Raised float positioning
○Failure of float SW due to short circuit
○Failure of input circuit of float SW due
to short circuit
Change to the correct dialyzer.
Check: Check function of [FSW] in maintenance mode 4 screen.
Failure of electric system is indicated by continuous indication
of ○.
Check: Measure resistance of float side after removing connector
from float.
Action: If short circuited when float is raised position, replace float
SW with new one.
Action: If float SW is normal, replace main board with new one.
E45 BLOWN-OUT FUSE DETECTION
Detection Blown out of pump fuse
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
Line function stop ―――
Condition or cause Check and action
No functioning of [P1] or [P2]
○Blown out of pump fuse
○Incorrect contact of pump driving line
○Failure of sub-board
Check: Observe [FUSE 402] (P) and [FUSE 404] (P2) placed on
sub-board visually or check using tester to find whether fuses
are blown out.
Action: If blown out, replace blown out fuse with new one.
Check: Pull lightly each of harnesses connecting POWER PLUG-
SWITCH GEAR- TERMINAL BLOCK2- SUB-BOARD CN404-
SUB-BOARD CN401-PUMP in series to check disconnection
and loosening of the harnesses.
Action: Tighten harness connection securely if loosening is detected.
Replace harness with new one if incorrect contact or
disconnection is detected
Action: Replace sub-board with new one.
Functioning of [P1] and [P2]
○Failure of sub-board
○Failure of main board
○Effect of noise
Action: Replace sub-board with new one.
Action: Replace main board with new one.
Check: Identify noise source such as hyperthermia device,
electro-cardiograph, laser knife or clean tower
Action: Place the machine at a position distant from the source.
Remove noise by any means such as use of insulating
transformer.
E46 WATCH-DOG ERROR
Detection Reset of system due to system error detected by watchdog
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note Initialization of RAM after error detection
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
――――
Condition or cause Check and action
Reset of system by the machine when
system error is detected by watchdog
Incorrect function of software by effect
of external noise.
Check: Identify noise source such as hyperthermia device,
electro-cardiograph, laser knife or clean tower
Action: Place the machine at a position distant from the source.
Remove noise by any means such as use of insulating
transformer.
Failure of main board Action: Replace main board with new one.
E47 RAM ERROR
Detection RAM error
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Audible
alarm
○ (7times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
――――
Error detection at OFF→ON of POWER SW or return
to normal after power failure
Initialization of RAM after error detection
Condition or cause Check and action
Detection of RAM error by self diagnosis
after operation start of the machine
Action: Replace main board with new one.
E48 EEPROM ERROR
Detection EEPROM error
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Audible
alarm
○ (7times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
――――
Condition or cause Check and action
Detection of EEPROM error by self
Action: Replace main board with new one.
diagnosis after operation start of the
machine
Error detection at OFF→ON of POWER SW or return
to normal after power failure
E49 EPROM ERROR
Detection EPROM error
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
――――
Error detection at OFF→ON of POWER SW or return
to normal after power failure
Initialization of RAM after error detection
Condition or cause Check and action
Detection of EPROM error by self
diagnosis after operation start of the
machine
Action: Replace main board with new one.
E50 RAM INITIALIZE
Detection
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset
Initialization of RAM
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Press [RESET] SW.
Dialysate line function BP/SP stop CLV function Reset
― ― ― ―
Condition or cause Check and action
Detection of different version of ROM by
check of RAM version
Detection of error as E46 WATCHDOG
ERROR.
Error detection at OFF→ON of POWER SW or return
to normal after power failure
Note
Automatic shift to DIALYSIS WAIT process after
error detection
Normal
Normal
Detection of error as E47 RAM ERROR Normal
Detection of error as E49 EPROM
ERROR
Normal
Check: Check if you have not turned on power breaker of the machine
for long term (about 7 to 10 days).
Discharged memory back up battery on
the Main board.
Turn on power breaker when you install the machine.
Action: Set date, time and display of SETTING MODE 1 because it was
initialized.
Turn on power breaker of the machine to full charge battery for
22 hours
Failure of main board Action: Replace main board with new one
Failure of sub-board Action: Replace sub-board with new one.
E51 EEPROM INITIALIZE
Detection Initialization of EEPROM
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
――――
Error detection at OFF→ON of POWER SW or return
to normal after power failure
Automatic shift to DIALYSIS WAIT process after
error detection
Condition or cause Check and action
Detection of different version of
EEPROM by check of EEPROM version
Normal.
Failure of main board Action: Replace main board with new one.
E52 SWITCH ERROR
Detection ON of any SW on operation panel for 20 sec continuously at POWER SW ON
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
― ― ― ―
●ROM Ver~3.93
Error detection at OFF → ON of POWER SW or
return to normal after power failure
●ROM Ver 4.01~
Error detection at anytime
Condition or cause Check and action
Press of any SW on operation panel
No press of any SW on operation panel
○Failure of operation panel
○Failure of operation panel board
○Failure of main board
Check: Press SW’s on operation panel one by one to find whether any
SW remains to be pressed.
Action: If any SW remains to be pressed, replace operation panel with
new one.
Action: Replace operation panel with new one.
Action: Replace operation panel board with new one.
Action: Replace main board with new one.
E53 I/O PORT ERROR
Detection Error of I/O port on board
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset Turn on [POWER] SW. Note
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
Line function stop ○――
Condition or cause Check and action
Latch up of IC (71054) on main board
Check: Check whether the machine function becomes normal by
Disabled all machine and operation panel SW
functions after error detection
○Application of abnormal voltage to IC
○Effect of noise
Failure of main board
turning on power breaker.
Check: Check whether abnormal voltage is applied to IC accidentally
during repair or maintenance of the machine.
Action: Turn on power breaker.
Check: Identify noise source such as hyperthermia device, electro-
cardiograph, laser knife or clean tower
Action: Place the machine at a position distant from the source.
Remove noise by any means such as use of insulating
transformer.
Check: Check whether error is alerted at each time when power
breaker is turned on frequently.
Action: Replace main board with new one.
E54 CONNECTION ERROR SBY0151-SBY0156
Detection Disconnection of harness used to connect main board to operation panel board
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset Turn on [POWER] SW. Note
Audible
alarm
○(7 tomes)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
Line function stop ○――
Condition or cause Check and action
Disconnection of [CN2] from operation
panel board and that of [CN10] from
main board
Connection of [CN2] to operation panel
board and that of [CN10] to main board
○ Disconnection of [OPERATION 2]
harness or incorrect connector
contact .
○Failure of operation panel board
○Failure of main board
Check: Check whether [CN2] and [CN10] are disconnected from
respective boards.
Action: If disconnected, connect correctly.
Check: Pull lightly lead wires of [CN2] of operation panel board and
those of [CN10] of main board to check disconnection and
loosening of the wires.
Action: Replace harness with new one.
Action: Replace operation panel board with new one.
Action: Replace main board with new one.
Disabled all machine and operation panel SW
functions after error detection
E55 BP BATTERY ERROR
Detection Voltage drop of blood pump backup battery to ≦8.5V
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
―――○
Condition or cause Check and action
Discharged battery
○Long time battery operation of blood
pump or insufficient battery charge
Check: Turn off power breaker of the machine and operate blood
time
pump by battery to check whether the pump is operated
normally. If the operation is abnormal, the cause is discharge or
○Revere connector connection
damage of battery.
○Battery life
Action: Turn on power breaker of the machine to full charge battery
for 48 hours.
○Failure of battery charge circuit
Action: Correct.
Action: Replace battery with new one.
Action: Replace sub-board with new one.
Enabled release from error in maintenance mode 7
screen
Check: Measure battery voltage using tester under loaded condition.
Charged battery
Action: If voltage is ≧ 8.5V, replace main board or sub-board.
E56 COUPLER SW ON ALARM(DIALYSIS)
Detection ON of [LS2]
Detection
process
Dialysis ○ (Continuous) Line function stop ―――
Reset Press [RESET] SW. Note
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Condition or cause Check and action
Functioning of [LS2]
○ON of [LS2] in dialysis process
Check: Operation miss
Action: No
No functioning of [LS2]
○Failure of [LS2]
○Disconnection of harness connected
to [LS2] or incorrect connector
contact
Check: Turn on and off [LS2] manually and check its function in
maintenance mode 4 screen.
Check: Disconnect connector from [LS2] side and measure resistance
between pins 1 and 2.
・When [LS2] is ON: the pins shall be short circuit condition.
・When [LS2] is OFF: the pins shall be open circuit condition.
Action: If malfunction of [LS2] is detected, replace it with new one.
Check: Pull lightly lead wires of [CN15] placed on main board and
those of connector of [LS2] side to check disconnection or
loosening of the wires.
Action: Replace harness with new one.
Action: Replace main board with new one.
○Failure of main board
E57 COUPLER SW ON ALARM(PREP.)
Detection Press of [UF DIALYSE] SW under [LS2] being ON
Detection
process
Prep. end
Recovery
Press [RESET] SW.
Reset
Auto reset by release from error
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (3 times) ――――
Condition or cause Check and action
「 Functioning of [LS2]
○ Press of [UF DIALYSE] SW under
setting of coupler to its receiver
No functioning of [LS2]
○Failure of [LS2]
○Disconnection of harness connected
to [LS2] or incorrect connector
contact
○Failure of main board
Note
Check: Operation miss
Action: Set coupler to dialyzer.
Check: Turn on and off [LS2] manually to check its function in
maintenance mode 4 screen.
Check: Disconnect connector from [LS2] side and measure resistance
between pins 1 and 2.
・When [LS2] is ON: the pins shall be short circuit condition.
・When [LS2] is OFF: the pins shall be open circuit condition.
Action: If malfunction of [LS2] is detected, replace it with new one.
Check: Pull lightly lead wires of [CN15] placed on main board and
those of connector of [LS2] side to check disconnection or
loosening of the wires.
Action: Replace harness with new one.
Action: Replace main board with new one.
E58 COUPLER SW OFF ALARM
Detection OFF of [LS2]
Detection
process
Rinse ○ (3 times) Line function stop ―――
Press [RESET] SW.
Reset
Auto reset by release from error
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Note
Condition or cause Check and action
No setting of coupler to its receiver
○Shift to RINSE process without setting
coupler to its receiver or dislodgment
of coupler from its receiver during
Check: Operation miss
Action: Set coupler correctly.
RINSE process
Setting of coupler to its receiver
○Damage of buttons of [LS2]
○ Deviated [LS2] position due to
loosening of set screws
○ Deviated coupler receiver position
due to loosening of set screws
○Failure of [LS2]
○ Disconnection of lead wires
connected to [LS2] or incorrect
connector contact
Check: Turn on and off [LS2] manually and check its function in
maintenance mode 4 screen.
Action: Replace [LS2] with new one.
Action: Correct the position and set [LS2] securely.
Action: Correct the position and set the receiver securely.
Check: Disconnect connector from [LS2] side and measure resistance
between pins 1 and 2.
・When [LS2] is ON: the pins shall be short circuit condition.
・When [LS2] is OFF: the pins shall be open circuit condition.
Action: If malfunction of [LS2] is detected, replace it with new one.
Check: Pull lightly lead wires of [CN15] placed on main board and
those of connector of [LS2] side to check disconnection or
loosening of the wires.
Action: Replace harness with new one.
Action: Replace main board with new one.
○Failure of main baord
E59 UF MAX ALARM
Detection Non-conformance with requirement of [UFR]≦[Blood Flow Rate]×60×[Set Upper UF Alarm Limit]
Detection
process
Dialysis ○ (Continuous) UF stop ―――
Press [RESET] SW.
Reset
Auto reset by release from error
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Note
Condition or cause Check and action
Non-conformance with requirement of
[UFR] ≦ [Blood Flow Rate] × 60 × [Set
Upper UF Alarm Limit]
○Decrease of blood pump flow rate
○Increase of UFR
[UF Alarm Limit] improperly set in
Action: Increase blood pump flow rate.
Action: Decrease UFR.
Action: Set [UF Alarm Limit]properly.
maintenance mode 2 screen
E60 SYRINGE INFUSION COMPLETION
Detection Output of syringe pump overload signal [LS3] during ON of SP SW
Detection
process
Rinse wait
Rinse
Rinse end
Air purge
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Press [RESET] SW.
Reset
Auto reset by release from alerting
Audible
alarm
○ (7 times)
↑
↑
↑
○ (Continuous)
↑
↑
↑
↑
Dialysate line function BP/SP stop CLV function Reset
―
Note Alerting during ON of SP SW
SP stop
――
Condition or cause Check and action
Completion of fluid drug infusion Check: Observe no fluid drug in syringe visually.
Action: Turn off [ON/OFF] SW of SP.
Catch of syringe by any object Action: Turn off power to release the catch.
Continuous indication of ○ (overload)
at [LS3] in maintenance mode 4 screen
○ Incorrect positioning of [LS3] or
screw for ON of it
○failure of [LS3]
○ Disconnection of [LS3] harness or
incorrect connector contact
○Failure of main board
Check: Observe whether screw for ON is positioned to allow ON of
[LS3] visually.
Action: Adjust the screw position correctly.
Check: Disconnect [CN16] from main board and measure resistance
between pins 1 and 3.
Turn on and off [LS3] manually.
・When [LS3] is ON (normal): the pins shall be short circuit
condition.
・When [LS3] is OFF (overload): the pins shall be open circuit
condition.
Action: Replace harness with new one.
Check: Pull lightly lead wires of [CN16] placed on main board to check
disconnection and loosening of the wires.
Action: Correct the disconnection and/or loosening.
Check: Monitor indication of [LS3] in maintenance mode 4 screen.
・Live condition : no indication
・Overload condition: indication of ○
Action: Replace main board with new one.
E61 COMPLETION OF SYRINGE.P INFUSION
Detection Arrival at set syringe infusion time
Detection
process
Dialysis ○ (3 times) ―SP stop ――
Press [RESET] SW.
Reset
Auto reset by release from alerting
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Note
Condition or cause Check and action
Check: Not required (Normal)
Arrival of set syringe infusion time
Action: To re-infuse, set infusion rate by adjusting syringe pump flow
rate setting SW.
If not re-infused, turn off [ON/OFF] SW of SP.
Check: Not required(Normal)
Setting of syringe stop tine to left
dialysis time in maintenance mode 2
screen
Action: To re-infuse, set infusion rate by adjusting syringe pump flow
rate setting SW.
If not re-infused, turn off [ON/OFF] SW of SP.
Check: Not required(Normal)
Change of UFR and UF goal and setting
of dialysis left time to syringe stop time
Action: To re-infuse, set infusion rate by adjusting syringe pump flow
rate setting SW.
If not re-infused, turn off [ON/OFF] SW of SP.
E62 PERFORMED PRIMING
Detection Completion of priming of target volume
Detection
process
Priming ○ (3 times) ―BP stop
Press [RESET] SW.
Reset
Auto reset by release from alerting
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
SW-OFF
Note
Condition or cause Check and action
Check: Not required
Completion of priming
Action: To re-prime, set target priming volume by operating [F・1] SW.
If not to re-prime, terminate priming function by operating [F・4]
SW.
Check: Not required
Reduction of target priming volume to
currently primed volume
Action: To re-prime, set target priming volume by operating [F・1] SW.
If not to re-prime, terminate priming function by operating [F・4]
SW.
○―
E63 PERFORMED TARGET UF.
Detection Completion of UF of target volume
Detection
process
Dialysis ○ (Continuous) UF stop ―――
Press [RESET] SW.
Reset
Auto reset by release from alerting
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Becoming error free of set upper alarm limits of
Note
venous and dialysate pressures by alerting
Condition or cause Check and action
Check: Not required.
Completion of UF of target volume
Action: To perform UF again, increase target UF volume.
If not to perform UF, terminate dialysis.
Check: Not required.
Setting of target UF volume lower than
total UF volume
Action: To perform UF again, increase target UF volume.
If not to perform UF, terminate dialysis.
E64 NO WATER RINSING
Detection
Detection
process
Dialysis wait ○ (3 times) ― ― ― ―
Reset
Shift to DIALYSIS WAIT process without performing water rinsing for a set time after acid rinsing, disinfection,
hotwater rinsing, and hotdisinfection acid rinsing
Audible
alarm
Press [RESET] SW.
Reset by performing water rinsing for a set time
Dialysate line function BP/SP stop CLV function Reset
Hotwater rinsing, and hot disinfection
Note
are not avairable with ROM Ver.2.40
and older
Condition or cause Check and action
No completion of water rinsing after
acid rinsing, disinfection, hotwater
rinsing, and hotdisinfection.
Action: Perform water rinsing for a set time.
E65 PREPARE DIALYSATE CONC.
Detection No connection to concentrate tank due to no detection of pressure rise in A or B concentrate line by [PS]
Detection
process
Preparation ○ (Continuous) Line function stop ―――
Reset Press [RESET] SW. Note
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Condition or cause Check and action
No dislodgment of concentrate
connector from rinse port during
PREPARATION process
No rise of priming end pressure
○No water supply
○Incorrect adjustment or failure of [PR]
○ Incorrect adjustment or failure of
[RV1]
○Incorrect adjustment or failure of
[RV3]
○Failure of [P1]
○Failure of [C2]
○Leak from rinse port
Action: Check the both A and B lines if connect to correct port or not
Dislodge concentrate connector from rinse port.
Action: Supply water.
Action: Adjust [PR].
If adjustment is impossible, repair [PR] or replace it with mew
one.
Action: Adjust priming end pressure by operating [RV1].
If adjustment is impossible, repair [RV1] or replace it with mew
one.
Action: Repair [RV3] or replace it with new one.
Action: Repair [P1] or replace it with new one.
Action: Replace [C2] with new one.
Check: Observe visually whether leak is caused from rinse port.
Action: Replace rinse port with new one.
Rise of priming end pressure
○Incorrect adjustment or failure of [PS]
○Failure of main board
Check: Measure priming end pressure.
Action: Adjust [PS].
If adjustment is impossible, repair [PS] or replace it with mew
one.
Check: Disconnect connector from [PS] and check ON and OFF of
[PS] in maintenance mode 4 screen by short circuiting and
opening the connector pins respectively.
Action: If ON/OFF function of [PS] is incorrect, replace main board
with new one.
E66 CONC CONNECTOR TO RINSING PORT.
Detection No connection to rinse port due to detection of pressure rise in A or B concentrate line by [PS]
Detection
process
Rinse ○ (3 times) Line function stop ―――
Reset Press [RESET] SW. Note
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Condition or cause Check and action
Acetate dialysis
No connection of concentrate
connector to rinse port while on
Check:Check if the undiluted solution is correct or not
Action:Connect line A to the undiluted solution.
Connect line B to the B rinse port
Action: Connect concentrate connector to rinse port.
preparation process.
Connection of concentrate connector to
rinse port
○Incorrect adjustment or failure of [PR]
○ Incorrect adjustment or failure of
[RV1]
○ Incorrect adjustment or failure of
[RV3]
○Incorrect adjustment or failure of [PS]
○Failure of main board
○ Failure of [V11], [V13], [V14], [V5],
[V6], [V7] and/or [V8]. (Failure of [V15]
when Na option is used.)
○Failure of main board
Action: Adjust [PR].
If adjustment is impossible, repair [PR] or replace it with new
one.
Action: Adjust priming end pressure by operating [RV1].
If adjustment is impossible, repair [RV1] or replace it with new
one.
Action: Repair [RV3] or replace it with new one.
Action: Adjust [PS].
If adjustment is impossible, repair [PS] or replace it with new
one.
Check: Disconnect connector from [PS] and check ON and OFF of
[PS] in maintenance mode 4 screen by short circuiting and
opening the connector pins respectively.
Action: If ON/OFF function of [PS] is incorrect, replace main board
with new one.
Check: Check ON and OFF of each valve in maintenance mode 4
screen.
If ON/OFF function of each valve is incorrect, check whether
output voltage is 24V.
Action: If 24V is outputted, replace relevant valve with new one.
If 24V is not outputted, replace main board with new one.
E67 RESET SW
Detection No press of [RESET] SW for a set time after alerting
Detection
process
Dialysis
wait
Preparation
Prep. end
Dialysis
Recovery
Reset Press [RESET] SW. Note
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
○ (Continuous) ―――○
Condition or cause Check and action
Check: Not required.
No press of [RESET] SW
Action: Press [RESET] SW.
Enabled release from alerting in maintenance mode
2 screen
E68 UF DIALYSE SW
Detection No press of [UF DIALYSE] SW for a set time after alerting
Detection
process
Prep. end ○ (3 times) ―――○
Reset Press [RESET] SW. Note
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Enabled release from alerting in maintenance mode
2 screen
Condition or cause Check and action
Check: Not required
No press of [UF DIALYSE] SW
Action: Press [UF DIALYSE] SW to shift to DIALYSIS process.
E69 BYPASS SW
Detection No release from manual ON condition of [BYPASS] SW for a set time
Detection
process
Dialysis ○ (3 times) ―――○
Reset Press [RESET] SW. Note
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Enabled release from alerting in maintenance mode
2 screen
Condition or cause Check and action
Check: Not required.
No press of [BYPASS] SW
Action: Press [BYPASS] SW to release from bypass condition.
NO INDICATION IN LCD
Detection
Detection
process
Reset
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Note
Condition or cause Check and action
Simultaneous alerting of this error with
E54 CONNECTION ERROR
SBY0151-SBY0156
Screen disappearance during RINSE
WAIT process
○Functioning of auto OFF mode
Check: See E54 CONNECTION ERROR SBY0151-SBY0156.
Action: See E54 CONNECTION ERROR SBY0151-SBY0156.
Check: Not required.
Note: Back-light is OFF after no SW operation for ≧30 minutes during
“indication of rinse execution screen and Pre-Rinse Wait
process” or “indication of rinse end screen”.
Action: Press any SW on operation panel to turn on back-light.
OFF of LCD
○Incorrect contrast adjustment
○Disconnection of [OPERATION PART
1] harness or incorrect connector
contact
○Disconnection of LCD harness or
incorrect connector contact
○Failure of LCD
○Failure of main board
Action: Adjust contrast by turning volume placed on operation panel
board.
Check: Pull lightly lead wires of [CN1] of operation panel board and
[CN10] of main board to check disconnection and loosening of
the wires.
Action: Replace harness with new one followed by turning on power
SW.
Check: Pull lightly lead wires of [CN4] and [CN5] of operation panel
board to check disconnection and loosening of the wires.
Action: Replace lead wires (harness) with new ones.
Action: Replace LCD with new one.
Action: Replace main board with new one.
NO INDICATION IN LCD
Condition or cause Check and action
Deviated indication
○Indication deviated by external noise
○Disconnection of [OPERATION PART
1] harness or incorrect connector
contact
○Disconnection of LCD harness or
incorrect connector contact
Check: Recover by turn on power SW again.
Identify noise source such as hyperthermia device,
electro-cardiograph, laser knife or clean tower.
Action: Move the source to a position distant from the machine.
Use noise reducing means such as insulating transformer.
Check: Pull lightly lead wires of [CN1] of operation panel board and
[CN10] of main board to check disconnection and loosening of
the wires.
Action: Replace harness with new one followed by turning on power
SW.
Check: Pull lightly lead wires of [CN4] and [CN5] of operation panel
board to check disconnection and loosening of the wires.
○Failure of LCD
○Failure of main board
Action: Replace lead wires (harness) with new ones.
Action: Replace LCD with new one.
Action: Replace main board with new one.
DISABLED SW FUNCTION
Detection
Detection
process
Reset
Audible
Dialysate line function BP/SP stop CLV function Reset
alarm
Note
Condition or cause Check and action
Simultaneous alerting of this error with
E54 CONNECTION ERROR
SBY0151-SBY0156
Disabled SW function; or
Enabled SW function with error
(function of any SW different from that
of pressed SW)
Check: See E54 CONNECTION ERROR SBY0151-SBY0156.
Action: See E54 CONNECTION ERROR SBY0151-SBY0156.
Check: Pull lightly flexible cable of [CN3] placed on operation panel
○Disconnection of operation panel
sheet cable or incorrect connector
contact
○Disconnection of [OPERATION PART
1] harness or incorrect connector
contact
○Failure of operation panel sheet
○Failure of operation panel board
○Failure of main board
board to check disconnection and loosening of the cable.
Action: Replace flexible cable with new one.
Check: Pull lightly lead wires of [CN1] of operation panel board and
[CN10] of main board to check disconnection and loosening of
the wires.
Action: Replace harness with new one.
Action: Replace operation panel sheet with new one.
Action: Replace operation panel board with new one.
Action: Replace main board with new one.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.