MyWeigh DuraScale User Manual

DuraScale Manual & Advanced Programming
"I tried using my scale for the first time yesterday. The display would only show the A.ch.Er error. Pulling
up on the tray as described in the user manual's A.ch.Er User Repair section provided no change. I was
able to locate the "For advanced user repair of Durascale 100 V2 ONLY" section on your website. I was
unsure of my scale's version but decided to give these instructions a try. Everything appears to function
properly since. I simply wanted to thank you for providing the necessary information on your website. I certainly appreciate it. David B. "
Low Batteries & bad battery connections are the #1 cause of scale malfunction and inaccuracy! We test all of our scale returns from consumers. Fully 60% of consumer returns are battery related problems. This sounds silly but it's true! A scale will perform slowly, or read inaccurately when it has low batteries.Please replace the batteries often (and only use good qualitybatteries). We include good quality batteries with all of our scalesbut batteries can run low in storage. If your scale simply won't turn on while on Battery power, this is often caused by loose battery connections. Battery prongs (terminals) are made of metal. They must be making good contact with your batteries in order for the scale to power on. You can use a paperclip to slightly bend the battery prongs to make them have a better connection. Also some poorly designed batteries have recessed or partially obstructed battery terminals. This may cause your prongs to be touching the plastic housing of the battery instead of the metal battery terminal.
Thank you for purchasing The DuraScale 100 or 50. With normal care and proper treatment it will provide years of reliable service. Please read all operating instructions carefully. Although this manual was current at the time of printing, we often update our user manuals and provide operation tips and guidance on our website.
* Description of error codes that may show on the display:
‘50.00 g’ ----- max processing weight is 50g; ‘0
‘100.00 g’ ----- max processing weight is 100g; ‘0
————
----- zero-point is too large;
————
----- zero-point is too large;
‘A.oFF.x’----- auto-off time is ‘x’ minutes; ‘CAL-0 ’ ----- zero-point calibration;
‘CAL-F ’ ----- full range calibration; ‘A.ch.E r’ ----- there is an error in analog channel; Please try
recalibration, if this does not work please see at the bottom of this page for more instructions.
—————
------ the weight is over 50.09g/100.09g; ‘______’------ the weight is below -0.09g.
‘EEP.Er’ ----- there is an error in accessing eeprom;
‘Hld.on’/‘Hld.oF’ ----- display data holding function is available or disabled
‘Cnt.Er’ ----- the unit weight (the weight of one part) is less 1d; (in counting function)
‘Cnt.xx’ ----- calculate unit weight using xx’s sample parts weight
* NOTE: errors "‘0
————
—————
’, ‘
’,‘______’ can usually be fixed by recalibration
KEEP THE FOLLOWING POINTS IN MIND:
* The tray of the scale can also be used as a weighing cup or bowl. Simply turn it upside down and place it on the weighing platform.
* Avoid lengthy exposure to extreme heat or cold, your scale works best when operated at normal room temperature. Always allow the unit to acclimate to a normal room temperature for at least one hour before use.
* Allow sufficient warm up time. Turn the scale On and wait several seconds to give the internal components a chance to stabilize before weighing.
* The cleaner the environment the better. Dust, dirt, moisture, vibration, air currents and proximity to other electronic equipment can all cause an adverse effect on the reliability and accuracy of your scale. Only use on a stable, vibration free surface
* Handle with care. Gently apply all items to be weighed onto tray top. Although this scale is designed to be quite durable, try to avoid rough treatment as this may permanently damage the internal sensor and void your warranty.
* Avoid shaking, dropping or otherwise shocking the scale. This is a precision instrument and MUST BE HANDLED WITH EXTREME CARE.
*IMPORTANT ADVICE: Place the item to be weighed on the platform, after the stable weight is
displayed remove the item immediately. This will prolong the longevity and accuracy of this weighing instrument.
.** These electronic scale are precision instruments. Do not operate near an in-use cell phone, cordless phone, radio, computer or other electronic device. These devices emit RF and can cause unstable scale readings. If your scale ever performs poorly, try moving the scale to a different room or location. This is a very precise scale - the display may seem to wander or jump when weighing. This is due to air currents or vibrations. Stable weighing is achieved when the display remains fixed for 3 seconds.
FEATURE_______________________________________________
* Power Up Segment Test & Stabilization
When first turning the unit on, the segments of the display will appear as shown below. This display will remain for approximately 5 seconds and then reset to 0. OPERATION____________________________________________________
When all segments flash on the display, this indicates the batteries are low
Battery Replacement:
DO NOT USE EXCESSIVE FORCE & DO NOT PRESS ON THE TRAY!!!
1) Four "AAA" size alkaline batteries are required.
2) To install batteries:
a) Release both battery covers by holding the tab and pulling out-wards.
b) Place batteries into battery compartment aligned correctly.
c) Replace battery cover.
3) The scale is now ready for battery operation.
(Advanced users only) CALIBRATION
NOTE: These instructions are for the 2000-2005 models only.
If you have a 2006 or later Durascale DO NOT FOLLOW THESE INSTRUCTIONS BELOW - Please see the 2006/7 Instructions at the bottom of this page
Only use the instructions that came with your scale. For a copy please contact us.
IMPORTANT: This scale was professionally factory calibrated before shipment and does NOT need to be recalibrated by the end user. However, if you wish to recalibrate your scale we provide these instructions for calibration: Repeat calibration if the scale performs inaccurately. Incorrect calibration can occur if you do not follow the steps exactly. You will need a 50gram weight or set of weights to calibrate this scale (included in the box).
DURASCALE 50 Easy Calibration:
There are two models of the Durascale. The older Durascales need to begin calibration with the scale turned OFF. The newer Durascales need to begin calibration with the scale turned ON
*Start with the scale powered off or on as per above and nothing on the weighing platform.
1) press and hold both the ‘ON/OFF’ and ‘TARE’ keys until ‘CAL-0’ shows on the display
Loading...
+ 6 hidden pages