MSI Dyna-Link 2, MSI-7300 Operator's Manual

Presented By
Dallas HoustonAustinFt Worth
NicolScales.com
800.225.8181
Contact Us
Nicol Scales & Measurement is an ISO Accredited Calibration
Company that has provided calibration, repair and sales of all types
of weighing and measurement products since 1931.
MSI-7300
Dyna-Link 2 Tension Dynamometer
Operator’s Manual
PN 152160 Rev C
Technical training seminars are available through Rice Lake Weighing Systems.
Course descriptions and dates can be viewed at www.ricelake.com/training
or obtained by calling 715-234-9171 and asking for the training department.
Contents
1.0 Introduction ................................................................................1
1.1 Safety ..........................................................................................2
1.2 Key Descriptions ........................................................................ 3
1.3 General Information ...................................................................4
1.4 MSI-7300 Annunciators .............................................................6
1.5 Specifications .............................................................................7
1.5.1 Standard Capacities and Resolution ..........................................8
1.6 Features ......................................................................................9
1.7 Options .....................................................................................10
1.8 Unpacking ................................................................................10
1.9 Assembly ..................................................................................10
1.10 Battery Replacement ...............................................................11
2.0 Operation ..................................................................................12
2.1 Power .......................................................................................12
2.2 Zero ..........................................................................................12
3.0 User Key Functions (F1and F2) ................................................13
3.1 OFF ...........................................................................................13
3.2 TEST .........................................................................................13
3.3 TOTAL .......................................................................................13
3.3.1 MANUAL TOTAL .....................................................................13
3.3.2 AUTO TOTAL ..........................................................................14
3.4 Clear Total ................................................................................14
3.5 Net / Gross ...............................................................................14
3.6 TARE .........................................................................................14
3.6.1 To Tare and Display the Net Tension .......................................15
3.6.2 To Clear the Tare and Revert to Gross Tension .......................15
3.6.3 Tare- Rules for Use: ................................................................15
3.7 PEAK HOLD .............................................................................. 15
3.8 2-UNITS/ 5-UNITS ....................................................................16
3.9 HI-RES ......................................................................................17
3.10 PRINT .......................................................................................17
4.0 Dyna-Link Setup .......................................................................18
4.1 Menu Map ................................................................................18
4.2 Function Keys ..........................................................................19
4.3 Auto-Off ....................................................................................20
4.4 Setpoints ..................................................................................21
4.5 Total Mode ............................................................................... 23
4.6 Units ..........................................................................................24
4.7 Filter Setup ............................................................................... 25
5.0 Calibration ................................................................................26
5.1 Calibration Menu ...................................................................... 26
© Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America.
Specifications subject to change without notice.
Rice Lake Weighing Systems is an ISO 9001 registered company.
June 8, 2015
ii MSI-7300 Operator’s Manual
Rice Lake continually offers web-based video training on a growing selection
of product-related topics at no cost. Visit www.ricelake.com/webinars.
5.2 Initial Calibration ......................................................................29
5.3 Guidelines for Capacity and Resolution .................................31
5.4 C-Cal Calibration ...................................................................... 32
5.5 Auto Zero Maintenance (AZM) ................................................ 34
5.6 Service Counters ...................................................................... 35
6.0 Communication Setup .............................................................36
6.1 RF Option ................................................................................. 36
6.2 Printer Setup ............................................................................ 37
6.2.1 Standard Print Strings .............................................................38
6.2.2 Printer Output Setup ...............................................................39
6.3 Comm Port Hardware .............................................................. 41
6.4 802.15.4 RF Network Setup ..................................................... 42
6.5 FCC Statement (For 802.15.4 Option) ..................................... 45
6.6 International RF CERTS (For 802.15.4 Option) .......................45
7.0 Appendix ..................................................................................46
7.1 Troubleshooting .......................................................................46
7.2 Error Codes ..............................................................................48
7.3 Mechanical Dimensions ........................................................... 49
7.4 Firmware Update Procedure ...................................................50
The MSI Limited Warranty ............................................................52
Introduction 1
1.0 Introduction
The MSI-7300 Dyna-Link 2 is a combination of the sound and proven mechanical design of the industry standard Dyna-Link with today’s most advanced electronics to provide a superb feature set unmatched by any dynamometer in its class or price range.
The Dyna-Link 2 is vers
atile, reliable, accurate, and easy to operate. The multi­purpose tension dynamometer is ideal for situations in which headroom is at a minimum. It is designed with safety factors exceeding the industry standard and is fully sealed for outdoor use in any weather.
A remote display option is available to further enhance the saf
ety and usability. The optional RF remote display allows tension monitoring from a distance and adds the ability to print and store data.
If you have any questions or comments, please contact
MSI Scales
Phone (toll free): 1-800-874-4320
MSI SCALES
Authorized distributors and their employees can view or download this manual from the MSI Scales distributor site at:
www.msiscales.com.
2 MSI-7300 Operator’s Manual
1. 1 Sa fe ty
Safety Symbol Definitions:
Important
WARNING
Indicates a potentially hazardous situation that, if not avoided, could result in serious injury or death, and includes hazards that are exposed when guards are removed.
Indicates information about procedures that, if not observed, could result in damage to equipment or corruption to and loss of data.
General Safety
WARNING
Do not operate or work on this equipment unless you have read and understand the instructions and warnings in this manual. Contact any Measurement Systems International dealer for replacement manuals. Proper care is your responsibility.
Failure to heed may result in serious injury of death. DO NOT allow minors (children) or inexperienced persons to operate this unit. DO NOT stand near the load being lifted as it
is a potential falling hazard. Keep a
safe distance. DO NOT use for purposes other then weigh
t taking or dynamic load monitoring.
DO NOT use any load bearing component that is worn beyond 5% of the original dimension.
DO NOT use the dynamometer if any of t
he components of the load train are
cracked, deformed, or show signs of fatigue. DO NOT exceed the rated load limit of the dynamometer, rigging elements, or the
li
fting structure. DO NOT allow multi-point contact with the shackles of the dynamometer unit. DO NOT allow high torque on the dynamometer unless it
is specifically designed
for high torque. DO NOT make alterations or modifications t
o the dynamometer or the shackles.
DO NOT use improperly rated or sized shackles. Use only MSI recommended shackles.
DO NOT remove or obscure warning labels. For guidelines on the safe rigging and l
oading of overhead scales and dynamometers, read the "MSI Crane Scale Safety and Periodic Maintenance Manual" (available at www.msiscales.com).
Keep hands, feet and loose clothing
away from moving parts.
There are no user serviceable parts within the Dyna-Link 2. Any repai
rs are to be
performed by qualified service personnel only.
Function key LEDs
Battery low
Stable
Center of zero
5 Digit 1.22 in (31mm) Sunlight visible LCD Tension Display
Units with maximum capacity of 100000 lb and up come with a 6 digit 1 in (26 mm) display.
Note
Setpoint alarm LEDs
Units & tension mode annunciators
User programmable function keys
Introduction 3
Figure 1-1. MSI-7300 Front Panel
1.2 Key Descriptions
POWER
Power Key - Turns the Dyna-Link 2 on and off. Used as cancel or last menu in the menu mode.
ZERO
Zero Key - Used to zero out residual tension on the link. Used as enter or save in the menu mode
Common uses are units change or peak hold for dynamic testing. F1- Pr
ogrammable to user selectable functions, see Section 3.0. Default – peak hold. Functions as the ENTER/SELECT key in the menu mode.
F2 -
Programmable to user selectable functions, see Section 3.0. Default – display & function test. Functions as the scroll key in the menu mode.
4 MSI-7300 Operator’s Manual
1.3 General Information
1. If the text is discussing a function key, the function key will be displayed as Fx-
YYYYY
with the programmed user key function in italics. F1 and F2 can both be
programmed to all available user functions.
2. If a function key does not work, it is probably because the Dyna-Link 2 is not setup to support the key. For example, if the Function key is set for TOTAL, you must also setup the TOTAL mode in the setup menu.
3. When in setup menus, the ZERO key drops back one menu level. At the root menu level, the ZERO key stores the changes and returns to the tension mode.
4. When in setup menus, the POWER key returns you directly to the tension display without storing the changes.
5000 lbs / 2500 KG
MAXIMUM CAPACITY
7300
lb
Top shackle interface
Cotter pins
Battery access cover
Bottom shackle interface
RS-232 port
Introduction 5
Figure 1-2. Dyna-Link 2
6 MSI-7300 Operator’s Manual
1.4 MSI-7300 Annunciators
The Dyna-Link 2 uses LCD annunciators to indicate tension mode and other information.
Figure 1-3. MSI7300 Annunciators
Stable – indicates the tension force has settled within the motion window (usually ±1d). When standstill is off, the link will not zero, tare, or totalize.
Center-of-Zero – Indicates the ten
sion is within 1/4 d of zero.
BT
Low Battery – Appears when approximately 10% of battery life remains. The BT symbol blinks when automatic shutdown is imminent.
Pk
PEAK- Indicates peak hold mode.
Ttl
TOTAL – Indicates the unit is displaying the total accumulated weight. This is a temporary display lasting less than five seconds.
Net
NET – Indicates the unit is in the net tension mode. A tare weight is subtracted from the gross tension.
M
Metric Ton – In conjunction with the ton annunciator, indicates the unit is displaying metric tons. When used with the total display, it is used for X1000 to allow accumulation of weight beyond the five digit display capacity. It is also used with the service counters when the number of lifts exceeds five digits.
kg
kg – Indicates tension display is in kilograms.
kN
kN – Indicates the tension display is in kilonewtons.
lb
lb – Indicates tension display is in pounds.
Ton
Ton – Illuminated alone, indicates the unit is displaying in US short tons (1 ton = 2000 lb). When illuminated along with the ‘M’ the unit is displaying in metric tons (1 metric ton = 1000 kg)
SETPOINTS – User programmable setpoints for early overload warnings. Setpoints 1 and 2 are high brightness red LEDs
F1 F2 – These blue LEDs are used to indicate various operational features of functions programmed into the F1 and F2 keys. Example: In peak hold mode the associated LED will blink whenever a new peak reading is captured.
0
Introduction 7
1.5 Specifications
Table 1-1. MSI-7300 Specifications
Accuracy
Rated accuracy: ±0.1% of capacity. T
ypical accuracy ±(0.1%+1d) of
reading. ‘d’ equals one displayable increment.
Resolution
Standard displayed resolution: 2000-25
00 ‘d’ . Hi-Res mode increases
resolution to 5000-6250 ‘d’. Internal A/D resolution: 24 bits.
Overload
Power
Display
Operating Te
m p
Operating Tim
e
Load Cell & Enclosure
Load Cell
F1 and F2
Calibration
Auto Zero Maintenance
Auto-Off Mode
Units
Filtering
Totalization
Peak
Setpoints
Service Counte
r
WARNING
The MSI-7300 has a safe mechanical overload of 200% of capacity. Overloads greater than 200% may result in physical damage to the link. The ultimate overload is rated to 500%-700% of capacity (see Table 1-2). At ultimate overload, structural failure and dropped loads may occur. Dropped loads may cause serious personal injury or death.
Safe link mechanical overload, 200% of capacity. Ultimate link overload, >500% of capacity. Typical ultimate overload is 700% or greater. Note that the Dyna-Link is designed to have a greater safety factor than the connecting shackles which have a typical ultimate safety factor of 600%
Battery operated, see Table 1-2 for battery size
and quantity per unit.
Alkaline cells can be replaced with rechargeable NiMH cells. Five digit, 1.22 in (31 mm) numeric digits.
Six digi
t, 1 in (26 mm) on 100000 lb units and higher
- 40°F to +122°F (-40°C to +50°C), Rate
d accuracy range -10°C to +40°C.
>150 hours typical (Alkaline C cells) / >300
hours typical (Alkaline D cells).
NEMA Type 4/IP65 anodized aluminum, 2024-T351 Aircraft Quality.
2000
Bridge
Programmable multifunction buttons for use as TEST, TOTAL, UNIT, PEAK, TARE, NET/GROSS, VIEW TOTAL, PRINT AND High Res mode.
Fully digital calibration from the front panel
or through a computer interface.
Standard, can be disabled internally.
Prolongs battery life by turning the power off after 15, 30, 45, or 60 minutes (oper
ator determined) of no Dyna-Link activity.
kg, lb, Tons (US short ton), Metric Tons, k
ilonewtons (other Units available
with custom calibrations).
Selectable - OFF, Low (LO), Medium (HI-1), High (HI-2).
Standard - Press button or automatic; TOTAL weight up to 99999 X1000 un
its
Uses unfiltered faster reading of A/D, (>22
0 readings per second).
Two internal standard setpoints and two
ultrabright LEDs on indicator panel
Two independent 16 bit registers; registe
r 1 updated each time the force exceeds 25% of capacity; register 2 updated each time the force exceeds overload; when register 1 exceeds 16383 or register 2 exceeds 1023, display reads “LCnt” for load cell counter; test function shows the two readings in order.
8 MSI-7300 Operator’s Manual
1.5.1 Standard Capacities and Resolution
Table 1-2. Standard Capacities and Resolutions
Capacity
Std
‘d’
Std
Counts HiRes ‘d’
HiRes
Counts
Ultimate
Overload Config
1000 lb 500 kg
0.5 Ton
4.9 kN
0.5 lb
0.2 kg
0.0002 T
0.002 kN
2000 2500 2500 2450
0.2 lb
0.1 kg
0.0001 T
0.001 kN
5000 5000 5000 4900
>>700% Alum
2 C-Cells
2500 lb 1250 kg
1.25 ton
12.25 kN
1 lb
0.5 kg
0.0005 T
0.005 kN
2500 2500 2500 2450
0.5 lb
0.2 kg
0.0002 T
0.002 kN
5000 6250 5000 4900
700% Alum
2 C-Cells
5000 lb 2500 kg
2.5 Ton
24.5 kN
2 lb 1 kg
0.001 T
0.01 kN
2500 2500 2500 2450
1 lb
0.5 kg
0.0005 T
0.005 kN
5000 5000 5000 4900
700% Alum
2 C-Cells
10000 lb 5000 kg
5.0 Ton
4.9 kN
5 lb 2 kg
0.002 T
0.02 kN
2000 2500 2500 2450
2 lb 1 kg
0.001 T
0.01 kN
5000 5000 5000 4900
700% Alum
2 C-Cells
25000 lb 12500 kg
12.5 Ton
122.5 kN
10 lb 5 kg
0.005 T
0.05 kN
2500 2500 2500 2450
5 lb 2 kg
0.002 T
0.02 kN
5000 6250 6250 6125
700% Alum
2 D-Cells
50000 lb 25000 kg 25 Ton 245 kN
20 lb 10 kg
0.01 T
0.1 kN
2500 2500 2500 2450
10 lb 5 kg .005 T
0.05 kN
5000 5000 5000 4900
600% Alum
2 D-Cells
100000 lb 50000 kg 50 Ton 490 kN
50 lb 20 kg
0.02 T
0.2 kN
2000 2500 2500 2450
20 lb 10 kg
0.01 T
0.1 kN
5000 5000 5000 4900
550% Alum
2 D-Cells
120000 lb 60000 kg 60 Ton 588 kN
50 lb 20 kg
0.02 T
0.2 kN
2400 3000 3000 2940
20 lb 10 kg
0.01 T
0.1 kN
6000 6000 6000 5880
500% Steel
2 D-Cells
180000 lb 90000 kg 90 Ton 882 kN
100 lb 50 kg
0.05 T
0.5 kN
1800 1800 1800 1764
50 lb 20 kg
0.02 T
0.2 kN
3600 3600 4500 4410
500% Steel
2 D-Cells
260000 lb 130000 kg 130 Ton 1275 kN
100 lb 50 kg
0.05 T
0.5 kN
2600 2600 2600 2550
50 lb 20 kg
0.02 T
0.2 kN
5200 6500 6500 6375
500% Steel
2 D-Cells
380000 lb 190000 kg 190 Ton 1863 kN
200 lb 100 kg
0.1 T 1 kN
1900 1900 1900 1863
100 lb 50 kg
0.05 T
0.5 kN
3800 3800 3800 3726
500% Steel
2 D-Cells
550000 lb 225000 kg 225 Ton 2206 kN
200 lb 100 kg
0. 1 T 1 kN
2750 2250 2500 2206
100 lb 50 kg
0.05 T
0.5 kN
5500 4500 5000 4412
440% Steel
3 D-Cells
Note
Short ton and metric ton resolutions are the same.
Introduction 9
1.6 Features
• Designed to meet or exceed all U.S. and International safety and environmental standards.
• Greater than 150 hours operation with two standard Alkaline ‘C’ cells. Greater than 300 hours with two standard Alkaline ‘D’ Cells (25000 lb/12500 kg capacities and above). Also works with off the shelf NiMH rechargeable cells.
• Automatic power off conserves battery life by sensing no activity after 15,30,45 or 60 minutes, determined by operator, and turns power off.
• Rugged construction throughout. IP65/NEMA Type 4 for outdoor use.
• Designed for use with USA made Crosby shackles (optional).
• Shackle holes reinforced with steel sleeves (25000 lb/12500 kg capacities and above) to reduce wear.
• Shackle stops ensure ease of mounting. The stops prevent the shackles from falling to the side of the unit and are held in position for easy rigging.
• MSI’s ScaleCore technology provides precision, high resolution (2500 division standard and up to 10,000 possible) 24 bit A/D conversion coupled with an advanced RISC microcontroller.
• Five large, 1.22 inch (31 mm) LCD digits for clear tension readings from a distance. Six digits, i in (26 mm) on units 100,000 lb and over.
• Easy to maintain: Full digital calibration assures reliable, repeatable measurements. Can be calibrated without test weights using MSI C-Cal technology.
• Selectable kg/lb/tons (US Short)/metric tons/kilonewtons.
• Automatic or manual weight totalization for loading operations.
• Easily customized for special applications.
• Hi speed PEAK mode for stress and drop test analysis.
• Two setpoints can be set for any in-range tension/weight value for operator alerts or process control. Optional audible alarm output.
• ScaleCore technology provides quick and easy firmware updates and calibration/ setup backup.
• Two Service counters ensure load train safety by warning the user to perform safety checks when the lift count gets high or the Dyna-Link has been overloaded repeatedly. Counter 1 (LFCnt) records the number of lifts above 25% of capacity. Counter 2 (OLCnt) records the number of times the Dyna-Link overloads.
10 MSI-7300 Operator’s Manual
1.7 Options
Options which you may have ordered with your Dyna-Link 2 may include the following:
• Audible alarm (triggered by setpoint 1)
• Top and bottom shackles
• Portable carry case
• Serial I/O cable (RS-232)
• RF remote display (will also operate hard-wired)
. See RF remote display for
ScaleCore user guide.
• Hardwired cable for remote display.
• RF remote modem, RS-232, for direct connection to computers, scoreboards, or serial printers. See
RF modem for ScaleCore user guide.
• RF remote modem, USB, for direct connection to computers. See RF modem for ScaleCore user
guide.
• RF remote gateway for direct connection to an Ethernet LAN and for use with MSI’
s SCCMP program. See RF modem for ScaleCore user guide.
• RF or hardwired scoreboard display.
1.8 Unpacking
When unpacking the Dyna-Link 2 from the shipping container, ensure that all assembly parts are accounted for.
Check for any visible damage and immediately report any damage to your shipper. Retain the original shipping container for future shipping or transporting.
1.9 Assembly
1. Identify and locate the following:
• Batteries, two ‘C’ cells or
‘D’ cells depending on capacity
• Top shackle and pin (option or customer supplied)
• Bottom shackle and pin (option or customer supplied)
• Two cotter pins
2. Slide top shackle over load cell and insert the pin.
3. Screw the shackle nut onto the pin.
WARNING
Note
Note
It is not necessary or desirable to tighten the nut too tight. Make sure the nut is down far enough to expose the cotter pin hole.
4. Lock the shackle pin in place with the supplied cotter pin. Bend the cotter pin.
5. Repeat steps 1-3 for the bottom shackle.
6. Remove the battery access port
cover with a coin or a large screwdriver.
7. Insert the two batteries, positive end first, into the battery shaft.
8. Reinstall the battery access port cover. The Dyn
a-Link 2 is now ready for use.
The Dyna-Link2 will automatically start when the batteries are installed.
The Dyna-Link 2 load train will be unsafe for use i
f the shackle pins
are not properly secured with cotter pins.
Introduction 11
1.10 Battery Replacement
Disposable Batteries
The BT annunciator will display when the is battery is beginning to get low. When the BT annunciator starts to blink, the batteries are close to being completely drained. For maximum life, use the batteries until the system shuts off.
Rechargeable Batteries
When using Nickle-Metal-Hydrid
e (NiMH) Cells, it is recommended that the cells are recharged immediately after the BT annunciator starts to blink. Do not allow the batteries to discharge completely, it may compromise the recharge life of the battery.
NiMH Cells in C and D sizes have a lower
capacity then the Alkaline C and D sizes. MSI recommends having two sets of NiMH batteries, so one set can be charging while the other is in use.
Note
NiMH ‘D’ cells are often repackaged ‘C’ cells so you don’t get an increase in battery life for Dyna-Links large enough for ‘D’ cells.
The use of NiCad batteries is not
recommended.
If the Dyna-Link 2 will not be used for an extended period, the batteries should be removed. A small current is used when powered off which will discharge the batteries in about six months.
12 MSI-7300 Operator’s Manual
2.0 Operation
2.1 Power
To turn on the Power
1. Press
POWER
.
• The LCD will show all segments for a display test.
• The software version number will display.
The D
yna-Link 2 is ready for use.
2.2 Zero
Takes out small deviations in zero when the Dyna-Link 2 is unloaded. See Section 3.6 for zeroing (Taring) package or pallet weights.
Note
Note
The tension reading must be stable within the motion window for the zero function to work.
1. Press
ZERO
. The display reads 0 (or 0.0 or 0.00, etc).
The backup memory stores the zero reading, and can restore it even if power fails.
Zero - Rules for Use:
• Works in GROSS mode or NET mode. Zeroing while in NET mode will zero the gross tension causing the display to show the negative tare value.
• The unit must be stable within the motion window, it will not zero until the stable ann
unciator is on.
• The unit will “remember” that it has a zero
request for two seconds. If motion
clears in that time, it will zero.
• The unit will accept a zero setting over the full range of the Dyna-Link. Zero setti
ngs above 4% of full Dyna-Link will subtract from the overall capacity. For example, if you zero out 100 lb on a 1000 lb Dyna-Link, the overall capacity will reduce to 900 lb plus the allowed over-range amount.
User Key Functions (F1and F2) 13
3.0 User Key Functions (F1and F2)
There are optional functions that can be programmed for the function keys (F1 and
F2) on the front panel, as well as on the RF remote display. See Section 4.2 for setup
instructions.
Note
The functions PRINT (F3), and TARE are available full time on the RF remote display.
3.1 OFF
No USER key function assigned. The F-Key is disabled.
3.2 TEST
The TEST function provides an LCD test that lights all LCD segments and the LED at once and then counts from 00000 to 99999. Other internal tests are performed and if any test fails, an error code will display. See Section 7.2 for a description of all error codes.
3.3 TOTAL
Note
The Total Mode must be programmed from the setup menus before the USER key will function. See Section 4.5.
For accumulation of multiple weighments. The accumulator always uses the displayed weight, so GROSS and NET readings can be added into the same
TOTAL.
There are four modes of totalizing: manual and
three auto modes.
The manual mode requires the
TOTAL key be pressed with the tension on the unit. The
tension will then be added to the previously accumulated value. This assures that a weight on the scale is only added to the total once. Both the manual and three auto total modes require that the tension on the Dyna-Link return below 0.5% (relative to full scale) of GROSS ZERO or NET ZERO before the next weighment can be added. Applied weight must be 1% of full scale above GROSS ZERO or NET ZERO before it can be totaled.
3.3.1 MANUAL TOTAL
The Fx-TOTAL key under the MANUAL TOTAL mode functions in this manner:
1. With the tension at more than 1% of capacity, push the
Fx-TOTAL key to add the
current tension to the TOTAL weight. The
Fx LED blinks to indicate the tension
value was accepted. The TOTAL LCD annunciator and the total weight is displayed for about five seconds, then the number of samples is displayed for about two seconds.
2. Remove the weight, when the tension is les
s than 1% of capacity, place next
load to be accumulated on the unit.
3. Repeat until all loads to be accumulate
d have been added to the total.
4. To view the total weight accumulated, push the
Fx-TOTAL key with the tension
at less than 1% of capacity (no weight on unit), the total weight will display for five seconds (view total) without changing the total value. The TOTAL LCD annunciator and the total weight is displayed for about five seconds and then the number of samples is displayed for about two seconds.
14 MSI-7300 Operator’s Manual
3.3.2 AUTO TOTAL
The Fx-TOTAL key under the AUTO TOTAL mode functions as auto total on/auto total off:
The auto mode has three variations which are programmed in the setup menu:
AutoLoad – Any settled tension above the ‘rise above’ threshold will be automatically totaled. Then the Dyna-Link must fall below the ‘drop below’ threshold before another total is allowed.
AutoNorm – This mode takes the last settled weight to auto total with. The total occurs only once the scale goes below the threshold. This allows the load to be adjusted without a total occurring. Once the load is removed, the scale uses the last settled reading for total.
AutoHigh – Similar to the AutoNorm mode except the scale uses the highest settled reading. Useful for loads that can’t be removed all at once.
3.4 Clear Total
The Fx-VIEW TOTAL key activates the total weight display followed by the number of samples. While the display is showing the total, total is cleared by pressing ZERO.
3.5 Net / Gross
Switches the display between net and gross modes. Net tension is defined as gross tension minus a tare weight.
To switch between net mode and gross mode:
1. Press the Fx-NetGross key (setup to the net/gross function).
2. The Fx-NetGross key will only function if a tare value has been established.
3. Switching back to gross mode from net mode will not clear the tare value. This
allows the operator to use the gross mode temporarily without having to reestablish the tare value. Only clearing the tare or setting a new tare will change the tare value held before switching into gross mode.
3.6 TARE
In force measurement applications, tare is useful as a way to display differential force. By taring out a known force, only positive and negative deviations from the tared force are displayed. This can also increase accuracy as any initial error is removed leaving only slope error. In scale applications, tare is typically used to zero out a known weight such as rigging, a packing container, or pallet and display the load in
NET tension/weight. To use tare, one of the two function keys must be configured to
the
TARE function. A tare value is entered by pressing the Fx-TARE key. The TARE
function in the Dyna-Link 2 is defined as a tare-in/tare-out operation. The first press of the
Fx-TARE key stores the current tension/weight as a tare value and then the
Dyna-Link subtracts the tare value from the gross tension and changes the display to
NET mode. The next press of the Fx-TARE key will clear the tare value and revert the
display to
GROSS mode. The optional RF remote display has a TA R E key
permanently available.
User Key Functions (F1and F2) 15
3.6.1 To Tare and Display the Net Tension
1. Press (programmed as TA RE ).
2. The tension reading must be stable within the motion window for the tare
fu
nction to work.
3. The digits display 0 and the tension mode changes to
NET.
4. The backup memory in the Dyn
a-Link 2 stores the tare reading, and can restore
it even if power fails.
3.6.2 To Clear the Tare and Revert to Gross Tension
1. Press (programmed as TA RE ).
2. The
NET annunciator will turn off.
3. Absence of the
NET annunciator is the only indication that you are in GROSS
tension mode.
Note
To view the
GROSS
tension without clearing the tare value, program
the remaining function key to the function NET/GROSS.
3.6.3 Tare- Rules for Use:
1. Only positive gross tension readings can be tared.
2. The stable annunciator must be on. The tension/force reading must be stable.
3. Setting or changing the tare has no effect on the gross zero setting.
4. Taring will reduce the apparent over range
of the Dyna-Link. For example, taring 100 pounds of rigging on a 1000 lb Dyna-Link, the Dyna-Link will overload at a net tension of 900 lb (1000-100) plus any additional allowed overload (usually ~4% or 9d).
5. The Dyna-Link stores the tare value in non-volatile memory is restored when power is
cycled.
3.7 PEAK HOLD
Peak hold will only update the display when a higher peak tension reading is established. The peak hold function uses a high speed mode of the A/D converter allowing it to capture transient tensions at a far higher rate than typical dynamometers.
Peak hold is cleared and re enabled with the Fx-Pea
k Hold key. When a new peak is detected, the Fx LED will flash three times. The accuracy of the system in peak hold mode is slightly reduced to .2% of capacity ± 5d. The filter setting is turned off while in peak hold mode to ensure the fastest acquisition rate.
Example Peak Hold Application
The peak hold function is useful in materials and ‘Fall’ tests. Common tests for fiber ro
pe include overall breaking strain (OB€), breaking force, and cycled breaking strain. The Dyna-Link 2 combined with a force test stand, meets the speed and accuracy requirements to properly conduct these tests.
1) Program a function key to Peak Hold (P-Hld)
In this example we’ll use F1 for Peak Hold.
2) Prepare the test stand and test sample.
3) Press ZERO to zero out any residual strain on the link.
4) Press F1 PkHold . Confirm that the “Pk” annunciator is on.
A small jump in the reading may occur depending on the stability of your test stand.
5) Apply the test force. The F1 LED will blink three times for every new peak it detects.
6) When the test force is removed, the Peak value can be recorded.
7) To run a new test, press F1 PkHold to clear the peak value. Confirm the “Pk” annunciator is off. Then repeat steps 3-6.
Peak Hold
kg
lb
M
kN
Ton

kg
lb
M
kN
Ton
Pk
kg
lb
M
kN
Ton

kg
lb
M
kN
Ton
Pk
Example Peak Captured Value
Peak Hold
kg
lb
M
kN
Ton

kg
lb
M
kN
Ton
kg
lb
M
kN
Ton

kg
lb
M
kN
Ton
Peak Cleared Value
F1
F1
0
ZERO
See Section 4.2 for Function Key setup
16 MSI-7300 Operator’s Manual
Figure 3-1. Capture Peak Force
3.8 2-UNITS/ 5-UNITS
The Fx-2.Unit key will switch the force units between pounds force and kilograms force. Selecting the Fx-5.Unit setting will scroll through all available units: lb, kg, tons (U.S. short), metric tons, and kilonewtons.
User Key Functions (F1and F2) 17
3.9 HI-RES
Set a function key (Fx) to Hi-Res, see Section 4.0. Press the Fx key programmed to display the high resolution mode (see Table 1-2 on
page 8). The display will stay in high resolution mode until the selected Fx key is pressed again, or power is cycled. While in the hi-res mode the appropriate Fx LED will blink continuously at a slow rate.
Hi-res mode does not increase the accuracy, but allows for smaller weight increments to be displayed.
Use Tare or the ZERO key to zero out any initial error. Hi-res mode will make the Dyna-Link 2 more sensitive to motion and movement resulting in a less stable display. When hi-res is on, the filter is automatically set to the Hi-1 setting (if Hi-2 is already set, then the filter is not changed). This will have a small effect on settling time. When hi-res is turned off, the filter setting resets to the previous filter setting.
3.10 PRINT
Set a function key (Fx) to Print, see Section 4.0. Pressing the Fx key outputs a configured text string to the RS-232 port on the base of
the unit. If an F-Key is programmed as print and the print setup is configured as continuous,
then the
Fx key is used for start print/stop print. See Section 6.2 for more details on
data output. The print function is always available on the optional RF remote display, so it is not
necessary to program an F-Key to “print” if you intend to trigger print outputs from the remote.
However, if you program F1 or F2 to “print” then pushing F1 or F2 on the Dyna-Link will cause the comm port on the remote to output the selected data string.
If the RF remote display option is installed, the Dyna-Link cannot use its built in comm port except for hard-wire connections to the RF remote display or firmware updates.
18 MSI-7300 Operator’s Manual
4.0 Dyna-Link Setup
4.1
2II WWO2Q $/R$' $/$6W
$+ L*K
)YQF IYQF DII 6W3W 6W3W WRWDO ILOWU
81L7
2II
  

II
/R +L +L
2))
*UHDW
/(66
Number Entry
Off
Test Display
Total
View Total
Net/Gross
Tare
Peak Hold
2 Units
5 Units
Hi Res
Print
F1 & F2 KEY
FUNCTIONS
TOTAL MODE
Total Off
(default)
Total On Manual
Autototal on Load
Autototal Last
Autototal on High Load
F2 Key F1
F2 Key F2
Auto Off Time
Set Point 1
Set Point 2
Total Mode
Weight Filter
Weight Units
SETUP MENU
Disabled (default)
15 minutes
30 minutes
45 minutes
60 minutes
AUTO OFF TIME
OFF (disabled)
Low
(default)
High Filter
Very High Filter
SOFTWARE FILTER
With the power on, press the F2 key and power key simultaneously
Set Point Off
(default)
Greater Than
Less Than
Totals on first stable load
Totals the last stable load before zero return
Totals the highest stable load
Totals by pressing an assigned F-key
Autototal Last and Autototal High total when the scale is unloaded.
*URVV
QH W*U
WRWDO
WFQW
SP WEIGHT MODE
Gross
(default)
Net or Gross
Total
Total Count (n)
(F1 default)
lb
kg
Ton
MTon
kN
Press Simultaneously
0
ZERO
POWER
F2F1
F2
Pounds kilograms Short Tons Metric Tons kiloNewtons
2))
WHVW WRWDO YWWO
1( W*U
WDUH 3+/G 8QLW 81LW +L5HV 3ULQW
Menu Map
Figure 4-1 MSI-7300 Menu Map
Dyna-Link Setup 19
4.2 Function Keys
The Dyna-Link 2 has two user definable function keys on the front panel that can be programmed to several different functions. F1 is defaulted to peak hold, and F2 is defaulted to test. This procedure also assigns the F1 & F2 keys on the optional RF remote display.
)YQF
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
ENTER/SELECT
F1
ENTER/SELECT
F1
WRWD/
blinking
QHW U
blinking
3+OG
blinking
WHVW
blinking
2II
blinking
1) With the 7300 on, press the F2 key and the POWER key simultaneously.
2)
The first item of the setup menu is Func1.
To setup the F2 Key scroll to the next menu item by pressing F2.
3) To set up the F1 key press F1. The current F1 key function is displayed.
4)
Select the F1 key function by scrolling through the choices with the F2 key. See the list of available functions on the setup menu map.
This procedure scrolls through all available choices for illustration purposes only.
In this example, we’ll set F1 to the TEST function.
5)
When the desired F1 Key function is displayed, press F1
.
The next item in the setup menu appears.
6)
Either press ZERO to exit setup and store all changes, or continue to another setup menu item using the F2 Key.
)YQF
Next Setup Menu Item
Store and return to
weight display
6WRUH
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
7HVW
blinking

blinking
XQLW
blinking
WDUH
blinking
YWWO
blinking
0
EXIT/SAVE
ZERO
Press Simultaneously
POWER
F2
SCROLL
F2
XQLW
blinking
SCROLL
F2
+LUHV
blinking
3ULQW
blinking
2II
blinking
0
ZERO
F1
Figure 4-2 Function Keys
20 MSI-7300 Operator’s Manual
4.3 Auto-Off
When enabled, the Auto-Off prolongs the battery life of the unit by turning the power off when it is not in use.
Any key press or detected tension in motion exceeding 10d, resets the time limit and the un
it remains on.
When disabled, the unit will only turn off by pressing the POWER key (or if the batt
ery is depleted).
1)
With the 7300 on, press the
F2
key and the POWER key simultaneously.
2)
The first item of the Setup Menu is Func1. Scroll to A-OFF with theF2 key.
3)
To set up the A-Off timing, press
F1. The current auto-off time is
displayed.
4)
Select the auto off time by scrolling through the choices with the F2 key.
In this example, we’ll set 60 minutes as the auto-off time.
5)
When the desired time is displayed, press F1. The next item in the setup menu appears.
6)
Either press ZERO to exit setup and store all changes, or continue to another setup menu item using the F2 key.
)YQF $2))
Press Simultaneously
POWER
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
ENTER/SELECT
F1
0
EXIT/SAVE
ZERO
F2
SCROLL
ENTER/SELECT
F1
2II
blinking

blinking

blinking
6W3W
Next Setup Menu Item
Store and return to
weight display
6WRUH

blinking

blinking
0
ZERO
F1
Figure 4-3 Auto-Off Setup
Dyna-Link Setup 21
4.4 Setpoints
The Dyna-Link 2 supports two setpoints. Common uses of set points are for warnings or process control. The MSI-7300 Dyna-Link 2 comes standard with two high brightness red LED outputs for a triggered set point. The MSI-7300 Dyna-Link 2 has an audible output option that is triggered by setpoint 1. Contact MSI for other setpoint output options.
1)
With the 7300 on, press the F2 and POWER keys simultaneously.
2)
Scroll through the setup menu choices by pressing the F2 key. Stop when the LED displays StPt1 or StPt2.
3)
When the desired setpoint is displayed, press F1. The display blinks OFF, or if previously programmed, the last set mode.
4)
Select the setpoint mode by scrolling through the choices with the F2 key. GrEAt (greater than) indicates the setpoint will trigger when the tension exceeds the value. LESS (less than) will trigger the setpoint when the tension is less than the value.
This example scrolled through all available choices for illustration purposes only.
5)
When the desired setpoint mode is displayed, press F1.
6)
Next select the type of tension or weight value the set point is assigned to. Use the F2 key to scroll through the choices.
This example scrolled through all available choices for illustration purposes only.In this example, we’ll enter Gross as the tension mode because we are going to use the set point as a safety warning.
7)
When the desired weight mode is shown, push F1. Next the current setpoint value is displayed.
If there was a previous value, it is displayed. If no value has been entered, a zero will appear. To keep the displayed value, press ZERO .
VWSW
UHDW
2II
blinking
/(66
UHDW
Current Value (if any)
Set Point responds to Gross tension, regardless of the display mode.
Set Point responds to the tension on the display, Net or Gross Weight.
Set Point responds to the Totaled Weight.
Set Point responds to the Total Count (number of samples).
Press Simultaneously
POWER
F2
F2
SCROLL
ENTER/SELECT
F1
ENTER/SELECT
F1
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
F2
SCROLL
Set Point responds to Gross tension, regardless of the display mode.
URV6
blinking
QHW U
blinking
WRWDO
blinking
WFQW
blinking
URV6
blinking
0
ZERO
F1
Figure 4-4 Setpoint Setup
8)
Press the F2 key. The first digit blinks at zero. Use the F2 key to scroll through the numbers. When the desired number is shown, push F1.
In this example,
we’ll enter 240 as a setpoint value.
To enter a decimal point, push
POWER while the digit is blinking.
Error Correction: If you input a wrong value, press ZERO to step back one digit and change the digit with the F2 key.
blinking
ENTER/SELECT
F1
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
 
 
fixed
blinking blinking
blinking
 
blinking blinking
9)
When the desired number is shown, push F1 a second time to set the value. The next setup menu item is displayed.
10)
Either press ZERO to exit setup and store all changes, or continue to another setup menu item using the F2 key.
 
xed blinking
 
blinking blinking
 6W3W
fixed next setup menu item
Store and return to weight display
6WRUH
SCROLL
F2
SCROLL
F2
SCROLL
F2
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
0
EXIT/SAVE
ZERO
22 MSI-7300 Operator’s Manual
Figure 4-4 Setpoint Setup (continued)
Dyna-Link Setup 23
4.5 Total Mode
The MSI-7300 Dyna-Link 2 can keep track of all weighments using the total feature. Either manual total, which totals by pushing a configured
USER key on the front
panel or the optional RF remote display, or auto-total which can be used to automatically add up each weighment. See the total mode descriptions for details on the various total modes. To use manual total, you must also program a user key. Auto total modes do not need a user key, but if a user key is setup for total, then it will function as a total on/total off.
1) With the 7300 on, press the F2 key and the POWER key simultaneously.
2) The first item of the setup menu is Func1. Scroll to totaL with the F2 key.
3) To setup the total mode, press
F1. The current total mode
setting is displayed.
4) Select the total mode by scrolling through the choices with the F2 key.
In this example, we’ll set the total mode to the auto-high mode. The auto high mode uses the highest stable reading as the total value, and totals when the load is removed.
5) When the desired total mode setting is displayed, press F1. The next item in the setup menu appears.
6) Either press ZERO to exit setup and store all changes, or continue to another setup menu item using the F2 key.
)YQF
)LOWU
Next Setup Menu Item
$/$67
Total last stable load
WRWD/
WWO2Q
Total with pushbutton
Store and return to
weight display
6WRUH
Press Simultaneously
POWER
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
ENTER/SELECT
F1
ENTER/SELECT
F1
0
EXIT/SAVE
ZERO
F2
SCROLL
$/RDG
Total on stable load
$+L +
Total highest stable Load
0
ZERO
F1
Figure 4-5 Total Mode Setup
24 MSI-7300 Operator’s Manual
4.6 Units
Units can be changed in two ways:
• Program a user function key to two unit or five unit
• Change the units with the setup menu using the following procedure
To set the accessible units available by a fun
ction key, set the F key either as 2Unit
(lb/kg) or 5Unit (lb/kg/short tons/metric tons/kilonewton).
1)With the 7300 on, press the F2 key and the POWER key simultaneously.
2)The first item of the setup menu is Func1. Scroll to Unit with the F2 key.
3)To set up the weight units, press
F1. The display will blink Unit.
4)Change the weight units by pressing the F2 key until the desired unit is displayed. The selected unit is indicated by the annunciators.
Not all units shown
are available on every capacity.
5)When the desired unit setting is annunciated, press F1. The next item in the setup menu appears.
6)Either press ZERO to exit setup and store all changes, or continue to another setup menu item using the F2 key.
)YQF
)YQF
Next Setup Menu Item
8QLW
Blinking
8QLW
Store and return to
weight display
6WRUH
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
Press Simultaneously
POWER
F2
ENTER/SELECT
F1
ENTER/SELECT
F1
SCROLL
F2
F2
SCROLL
0
EXIT/SAVE
ZERO
F2
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
0
ZERO
F1
Note
If the Dyna-Link Calibration was originally in tons or metric tons, the “2Unit” setting will switch from tons to metric tons instead of lb/kg.
Figure 4-6 Units Select Menu
Dyna-Link Setup 25
4.7 Filter Setup
Changing the filter settings allows the Dyna-Link to adjust to situations where there is a lot of movement in the lift or the crane structure. If the reading is not stable, it can often be improved by increasing the filter setting. Settling time will be longer as the filter setting is increased. However, the MSI-7300 Dyna-Link 2 employs algorithms that speed up large tension changes while still controlling vibration even with higher filter settings.
1) With the 7300 off, press and hold the F2 key, then press the
POWER key.
...or while the 7300 is on, press F2 and
POWER simultaneously.
2) The first item of the setup menu is Func1. Scroll to Filtr with the F2 key.
3) To set up the filter, press F1. The display will blink the current filter setting.
4) Change the filter setting by pressing the F2 key.
There are four available filter settings. Not all choices are shown in this example.
5) When the desired filter setting is annunciated, press F1. The next item in the setup menu appears.
6) Either press ZERO to exit setup and store all changes, or continue to another setup menu item using the F2 key.
)YQF
+L
High Filter 1
2))
Blinking
)LOWU
2))
Blinking
/2
Low Filter, Blinking
8QLW
Next Setup Menu Item
Store and return to
weight display
6WRUH
Press Simultaneously
POWER
F2
0
EXIT/SAVE
ZERO
ENTER/SELECT
F1
ENTER/SELECT
F1
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
F2
SCROLL
0
ZERO
F1
Figure 4-7 Filter Setup Menu
26 MSI-7300 Operator’s Manual
5.0 Calibration
The Dyna-Link 2 is calibrated using standard precision test weights. It is required that the weight used is at least 10% of full capacity in order to achieve rated accuracy. For example, use at least a 500kg test weight to calibrate a 5000kg capacity unit. The Dyna-Link 2 supports load cell linearization with up to four span points that can be calibrated in any order. Usually only one cal span point is necessary and is sufficient to reach rated accuracy.
When adequate test weights are not available,
the Dyna-Link 2 can be calibrated using a constant calibration which is referred to as C-Cal. To use C-Cal, a factory generated C-Cal number must be known. MSI supplies original and replacement load cells for the Dyna-Link 2 with the C-Cal value stamped on the serial number label.
There are three aspects of calibration:
• Standard calibration - used for maintenance and routine calibration.
• Initial calibration - used to setup both the capacity and resolution (d) of the Dyna­Link
. It differs from standard calibration only in the initial steps. Initial calibration is performed after a calibration reset which completely erases the calibration and setup memory.
• C-Cal - If C-Cal values are known, the Dyna-Link can be calibrated without weig
hts.
5.1 Calibration Menu
The calibration menu contains three items: Cal, C-Cal, and auto zero maintenance. The following procedures start with entering into the Cal menu, or for an initial calibration, resetting the Dyna-Link 2 and then going to the Cal menu.
&$/ &&D/ 6HWXS
Calibration Setup Menu
Calibrate with Weights
Calibrate with
Calibration number
Calibrate Menu
Figure 5-2 - Standard Calibration or
Figure 5-3 - Initial Calibration (After Reset)
Figure 5-4
Figure 5-5
Figure 5-1 Calibration Menu
Calibration 27
Procedure for the Routine Calibration of the MSI-7300
XQ/G

/R$'
lb

Blinks Capacity
1) With the power on, initiate calibration by holding down the F2 and ZERO keys until the display reads CAL.
The CAL setup menu appears.
2) Press F1 to start the calibration procedure.
3) The display reads UnLd (unload) indicating you should remove all weight from the link.
You can choose to leave bottom fittings on the link as long as they are always part of the load train.
4) Press F1. The 7300 sets the zero calibration point.
5) If the zero is in range, the scale will display PASS. Then LoAd1 is displayed.
6) Load the link with a precision test weight.
For highest accuracy, a test weight of 10% of capacity or more is recommended.
7) Press F1. The 7300 flashes the capacity and calibration units. If you are loading the scale with the capacity weight, skip to step 10.
8) To enter a calibration weight other than capacity, press F2. The displays far left digit will flash zero indicating that a number should be entered.
&$/

blinking
3$66
No weight on link
Flashes Zero
Lift Test Weight
example capacity
Error Correction: If you input a wrong value, press ZERO to step back one digit and reenter.
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
SCROLL
F2
Press
Simultaneously
F2
0
ZERO
F1
Figure 5-2 Standard Calibration Procedure
 
 




 
5 times
fixed
blinking blinking
blinking
blinking
blinking
blinking
fixed
fixed
fixed
3DVV
9) Press the F2 key to scroll the number and the F1 key to enter each digit of the calibration weight.
In this example, we’ll enter 2500 kg on a 5000 kg capacity scale. Do not push the F1 key two times in a row.
To add a decimal point, push the
POWER key while the number is
blinking.
10) When the entire value of the test weight is displayed and the weight and link are stable, press F1 to finish off the weight entry. If the resultant cal value is within limits, the display will read
PASS.
11) The display now reads LoAd2. The Dyna-Link allows multi-point calibration. If more cal points are desired (up to three additional) press
F2
and repeat steps 8-10. If finished with span points, go to step 12.
12) Press ZERO to store the calibration constants. The LCD will read CAL’d to indicate the calibration is acceptable.
13) Skip to step 14, or press F1 to observe the CAL number. The C-CAL number appears. Make a note of the value.
14) Press ZERO to exit calibration. The next menu item Setup appears.
15) Press ZERO to store the calibration and return to standard link operation, or press F1 if you want to adjust additional cal setup parameters
(Standard, AZM, et c.)
You can cancel calibration by pressing Power and the scale will reset to the previous calibration constants.
To add span
points
Go to step 8
To complete
calibration
63DQ
63DQ
/RDG
6HW8S
Next Cal menu item
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
F2
SCROLL
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
0
EXIT/SAVE
ZERO
0
EXIT/SAVE
ZERO
ENTER/SELECT
F1
Optional:
to display
Cal number
&&D/
ENTER/SELECT
F1

blinking
Example Cal#
VWRUH
0
EXIT/SAVE
ZERO
&$/ '
Cal saved to memory
F1
28 MSI-7300 Operator’s Manual
Figure 5-2 Standard Calibration Procedure (continued)
Calibration 29
5.2 Initial Calibration
Use this procedure only if the capacity and count-by (d) needs to be modified. The initial steps of this procedure will totally erase user setups as well as any previous calibration.
1) Turn the 7300 on.
2) Press the ZERO switch and
the F1 switch simultaneously.
3) The display blinks rESEt.
4) To reset all calibration
constants and setup parameters, press F1.
5) The 7300 requests a
confirmation by displaying Sure?.
To cancel the Reset press
the POWER or the ZERO key.
6) To complete the reset, press
F1. The Calibrate menu
appears. You must now recalibrate the system.
7) Press F1 to start the initial calibration procedure. The display shows Unit. You select the force units you wish to calibrate in.
8) Press F1 to select the calibration unit.
9) Use the F2 key to scroll through the available calibration units.
10) When the desired unit is shown, press F1.
11) Next, set the capacity in the selected units. Capacity must be set no higher than the load cell rated capacity.
12) Press F1 to enter the capacity setting screen. A capacity of 10000 is the initial value.
If 10000 is the desired capacity in the selected Calibration Unit, press F1 and skip to step 16.
13) To change the capacity, press
F2.
Resetting Capacity and Countby (d)
UHVHW
blinking
6YUH"
&$/
8QLW
8QLW
blinking
&$3

blinking

blinking
Error correction: if you input a wrong value, press ZERO to step back one digit and reenter.
To enter a decimal point, push
POWER while the digit is blinking.
For very large capacity Dyna-Links Enter the capacity in weight X1000 and push the POWER key twice (X1000 mode is indicated by the M annunciator during data entry).
e.g. For a 100,000 lb 7300, enter 100, then press the POWER key twice so the M indicator is on.
ENTER/SELECT
F1
ENTER/SELECT
F1
EXIT/CANCEL
POWER
To cancel
Reset
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
SCROLL
F2
ENTER/SELECT
F1
ENTER/SELECT
F1
SCROLL
F2
kg
lb
M
kN
Ton
Press &Hold
0
ZERO
F2F1
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
pounds Tons (US Short)
Metric Tons (kg X 1000)
kiloNewtons
kilograms
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
Figure 5-3 Initial Calibration
The first digit blinks. Use the
F2 key to scroll through the
numbers. When the desired number is shown, push F1
.
In this example, we’ll enter 2500 as a capacity.
Continue inputing the desired capacity using the F2 key for scrolling the number and the
F1 key to store the number.
 
 
fixed
blinking blinking
blinking
SCROLL
F2
SCROLL
F2
SCROLL
F2
ENTER/SELECT
F1




 
5 times
blinking
blinking
blinking
fixed
fixed
fixed
15) Finalize the capacity value by pressing the F1
key on an unblinking display. In our example, once the number 2500 is fixed on the display, press F1
to store the capacity
value.
16) Next the scale division size d is set. Press F1 to begin.
In this example we’ll set the d to
2.
17) Use the F2 key to scroll through the recommended scale divisions. The first d
offered is the standard division for the given capacity. Setting a ‘d’ size that results in total resolution higher than 1:5000 is not recommended for stability reasons.
18)When the desired scale d is displayed, press F1. The UnLd display appears and the scale is ready for calibration. Follow the standard calibration procedure starting at step 3.
G

blinking
XQ/G
Capacity is set.
d is set.
Proceed to Standard Calibration starting at Step 3.
F2
SCROLL
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
SCROLL
F2
SCROLL
F2
ENTER/SELECT
F1
Next Cal Screen

blinking
SCROLL
F2
ENTER/SELECT
F1
14)
30 MSI-7300 Operator’s Manual
Figure 5-3 Initial Calibration (continued)
Calibration 31
5.3 Guidelines for Capacity and Resolution
Dyna-Links are subject to forces that static scales do not experience. Many bridge cranes, hoist cranes, and mobile cranes lack rigidity and tend to bounce or swing when loads are lifted. For this reason, MSI recommends that resolution is kept in the 1:2000 to 1:3000 range. Some improvement in stability can be achieved by increasing the filtering. However, you should never program resolution that is far greater than you need. If the MSI-7300 Dyna-Link 2 display is never stable, it is recommended that the resolution is reduced as well as filtering increased. In any circumstance, the resolution should never be set higher than 1:15000 due to temperature and noise considerations common to all strain gage based load cells.
The tension must be stable for certain features to work: ZERO tensi
on must be stable to be zeroed. TARE tension must be stable to be tared. TOTAL tension must be stable to be added to the total registers. One way to improve the stability is to increase the filtering, at the risk of increasing settling time. The other is to increase the ‘d’ (reduce resolution). The third way is to increase the motion window. The MSI-7300 Dyna- Link 2 defaults to ±1d as a motion window. It can be changed at MSI to a higher value if desired. Often ±3d is chosen for bridge cranes as these tend to have a lot of bounce to them. This of course carries an accuracy penalty adding ±3 d to the total accuracy of the Dyna-Link if the zero or tare operation happens to capture the tension in a valley or peak.
Setting capacity is dictated primarily by the
capability of the load cell. MSI supplies the MSI-7300 Dyna-Link 2 in many capacities. Never set the capacity of the Dyna- Link higher than the rating of the load cell. Due to the excellent linearity of the MSI Link load cell, it is acceptable to set lower capacities to better match the crane the MSI-7300 Dyna-Link 2 is used on. For example, if the hoist is rated for 9000lb, you can use a 10000 lb capacity Dyna-Link and reset the capacity to 9000 lb so that the Dyna-Link will indicate overload at 9000 lb instead of 100000 lb. Derating as much as 50% of the capacity is usually acceptable, but the Dyna-Link may be less stable if the ‘d’ is decreased (resolution increased).
Note
The capacity of all the MSI-7300 systems is rated approximately 20% higher than the rated capacity in pounds. This is to allow the kg capacity to be exactly 1/2 the number of the pounds capacity.
32 MSI-7300 Operator’s Manual
5.4 C-Cal Calibration
When adequate test weights are not available, the MSI-7300 Dyna-Link 2 can be calibrated using a cal number calibration which is referred to as C-Cal. To use C-Cal, a factory generated C-Cal number must be known. MSI supplies replacement load cells for the MSI-7300 Dyna-Link 2 with the C-Cal value stamped on the serial number label. When a calibration is preformed with test weights, a new C-Cal is generated.
Important
The C-Cal number must be known prior to starting this procedure. For a MSI-7300 with its original load cell, MSI prints this number on the Calibration Record, the serial number tag. C­Calibration reduces slightly the absolute accuracy of the system and is intended for non-critical use only. For highest accuracy, calibrate the MSI-7300 with precision test weights.
XQ/G

1) With the power on, initiate calibration by holding down the F2 and ZERO keys until the display reads CAL.
The CAL setup menu appears.
2) Press F2 to scroll to the C–Cal selection.
3) Press F1 to start the C-Cal calibration procedure.
3) The display reads UnLd (unload) indicating you should remove all weight from the link.
You can choose to leave bottom fittings on the Link as long as they are always part of the load train.
4) Press F1. The 7300 sets the zero calibration point.
5) If the zero is in range, the scale will display PASS. Then C-CAL is displayed.
6) Press F1. The last known C–Cal value is displayed.
If the offered C-Cal value is correct, push F1 and jump to step 10.
In this example we’ll enter 15200 as an C-Cal value. You must use your known C-Cal value, not 15200.
&$/
3$66
No weight on link
Flashes Zero
ENTER/SELECT
F1
ENTER/SELECT
F1
SCROLL
F2
&&$/
&&D/
lb

example C-Cal Value
ENTER/SELECT
F1
Press
Simultaneously
F2
0
ZERO
F1
Figure 5-4 C-Cal Calibration Procedure
7) To input the C-Cal value, press F2 to start the number entry process.
8) Use F2 to change the number, and F1 to enter the number. Add the next digit by pushing F2 and scrolling as required. Repeat this sequence until the entire C-Cal number is entered.
 
 

5 times
fixed
blinking blinking
blinking
blinking
SCROLL
F2
SCROLL
F2
SCROLL
F2
F2
SCROLL
ENTER/SELECT
F1
Error Correction: If you input a wrong value, press ZERO to step back one digit and reenter.
To input a decimal point, push POWER while a digit is blinking.
F1
9) Once the entire C-Cal value is entered, press
F1 to finalize the number.
The 7300 modifies its span factor registers to adjust the calibration to the value of the C-Cal. If the C-Cal input was in the acceptable range, the LCD will read PASS.
Multiple C-Cal span points are possible, but are only accessible using MSI’s optional SCCMP program.
10) When the final C-Cal number is finished, press
ZERO to exit and save
the new C-Cal calibration. The LCD reads STORE to indicate a successful calibration.
11) Press ZERO again to exit the CAL menu and start standard link tension operation.



 
blinking
blinking
fixed
fixed
fixed
3DVV
SCROLL
F2
SCROLL
F2
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1

blinking
SCROLL
F2

SCROLL
F2
blinking
0
EXIT/SAVE
ZERO
0
EXIT/SAVE
ZERO
6WRUH

SCROLL
F2
blinking
&$/ '
6HWYS
Calibration 33
Figure 5-4 C-Cal Calibration Procedure (continued)
34 MSI-7300 Operator’s Manual
5.5 Auto Zero Maintenance (AZM)
The Dyna-Link 2 employs an auto zeroing maintenance mechanism to adjust the zero reading to the center-of-zero (COZ). COZ is defined as the tension reading is within ¼’d’ of zero. AZM continuously adjusts zero to maintain COZ. It is recommended that AZM is on to maintain the highest accuracy. However, there are circumstances when it should be turned off. This can happen when minor variations of tension occur while picking up Dyna-Link attachments and the variations fall within the AZM capture window. The AZM capture window (usually 1 ‘d’) and capture time (usually eight seconds) can be adjusted by MSI to meet custom requirements.
1) With the power on, initiate calibration by holding down the
F2 and ZERO
keys until
the display reads CAL.
The CAL setup menu appears.
2) Press F2 to go the the SetUP screen.
3) Press F1. The StAnd menu appears.
4) Press F2 to scroll to the Auto0 screen.
5) Press F1. The AutoZero mode shows On or OFF.
In this example, we will turn off AZM (not recommended for typical operations)
6) Press F2 to change to AZM between on and off.
7) When the desired mode is displayed, press F1.
8) The next cal setup menu item appears.
9) Press ZERO twice to exit cal setup and store all changes, or continue to another cal setup menu item using the F2 Key.
&$/
ENTER/SELECT
F1
F2
SCROLL
6HW8S
SCROLL
F2
ENTER/SELECT
F1
0
EXIT/SAVE
ZERO
6WDQG
2II
blinking
Press
Simultaneously
F2
0
ZERO
F1
ENTER/SELECT
F1
2Q
blinking
F2
SCROLL
$XWR
6WDQG
Next Cal Setup Menu Item
0
EXIT/SAVE
ZERO
&$/
Next Cal Menu Item
6WRUH
Changes saved to
Backup Memory
Figure 5-5 Auto Zero Maintenance Setup
Calibration 35
5.6 Service Counters
The MSI-7300 maintains two service counters for safety.
• The first counter counts the number of times the scale has been overloaded.
• The second counter counts lifts above 25% of capacity.
These counters serve to warn the user to in
spect the load train after a number of overloads, also when there is a chance of fatigue failure. The power up routine will be interrupted when the lift counter exceeds 16383 lifts or the overload counter exceeds 1023 overloads. If the screen displays LFCnt when unit is powered on :
1. Push TARE to display the 25% lift counter.
2. Push TARE again to see the overload lift counter.
3. Push the ZERO key to acknowledge the warning and return to standard scale operatio
n.
Note
The power up warning message won't appear again for another 16383 lifts ( or 1023 overloads).
1) Program a user function key to be TEST (see function key setup) For this example, F1is programmed as TEST.
2) Press F1- TEST .
3) Within two seconds of pressing the
F1– TEST key, press F1 again
(must be F1 regardless of which key is
programmed as TEST).
The test will sequence through steps 4 to 7 automatically unless you stop it by
pressing F2 immediately, then using
F1 and F2 each parameter can be
observed statically.
4) The display flashes LFCnt (for lift counter) followed by the number of times the weight has exceeded 25% of capacity.
5) Next, the display flashes OLCnt (for overload counter) followed by the number of times the weight has exceeded capacity.
6) Next,the display flashes the C­Cal value.
7) The Dyna-Link returns to standard weighing mode. If you interrupted the auto sequence,
press ZERO to return to tension link mode.
/)&QW

2/&QW
Return to standard weight display

ENTER/SELECT
F1
Number of Lifts above 25% of Capacity

Number of Lifts that exceeded Capacity
example value
example value
Start of Test Sequence
All segments on
&&D/
blinking
blinking
example C-Cal value
Fx
Figure 5-6 Access the Service Counters
WARNING
Only a MSI factory representative can reset the service counters, as these are important safety warning features. Depending on the circumstances, a thorough load train inspection might be necessary to ensure user safety.
Reference MSI’s “Crane Scale Safety and Periodic Maintenance manual” (Pub. 243-08-94D) for proper loadi
ng techniques to improve the safety and longevity of
your
MSI-7300
crane scale. This publication is available at www.msiscales.com
and is included in the CD shipped with your crane scale.
36 MSI-7300 Operator’s Manual
6.0 Communication Setup
The Dyna-Link 2 can communicate with peripheral devices using RS-232 or 802.15.4 wireless. Only one communications type can exist at a time. The RS-232 port located on the bottom side of the Dyna-Link 2 is useful for setup and calibration using a computer and MSI’s SCCMP Software (SCCMP operation is detailed in the SCCMP User Guide).
6.1 RF Option
Since the RF options are easily connected they are commonly used for gathering weight data after the initial setup of the unit. For RF operation, the Dyna-Link 2 uses an 802.15.4 transceiver to communicate with the MSI-8000 RF Remote Display.
802.15.4 wireless:
• operates in the 2.4GHz ISM band and does not require the end user to obtain a licens
e.
• can coexist with other 2.4GHz systems if caution is
taken to isolate antennas at
least 10 feet or 3 meters from the Crane Scales and MSI-8000 equipment.
MSI-8000
based RF systems are peer to peer. However, for multiple scale
connections, the MSI-8000 acts as the network coordinator.
Number entry screen
0
TAR E
F2
Exit and Cancel Changes
Exit and Save/Back 1 Level
Enter / Select
Scroll
In Menu Key Functions
PRINT STRINGS
6WUQ*
&QWUO
5DWH
2II 8VHU /RDG &RQW
Off (default)
F key
On Stable Load
Continuous
PRINT CONTROL
Print String Output Rate Number entry screen (0-65535 seconds)
Print String Format
1= Current Wt
(Wt-Unit-Mode )
2= Net Wt (Wt-Unit-Net ) 3= Gross Wt (Wt-Unit-Grs ) 4= Tare Wt (Wt-Unit-Tare ) 5= Total Wt (Wt-Unit-Total ) 6= Total Count (#Samples-TCNT ) 7= Current Wt (no units or mode
)
8=
(Reserved)
9= CR-LF (
)
Affects cont­inuous mode only
3ULQW
U)
Print
RF
COMM SETUP MENU
2Q2II
FLG &KQ/
QHWLG
Enable RF
ScaleCore ID
RF Channel
Network ID
RF
SETUP
Setup Strings
Print Control
Output Rate
Network ID
2Q
2II
Range 1-254, (20-30)
Range 12-23
Range 0 to 99999
RF Network Number entry screens
F2
Press Simultaneously
F1
F2
6WU(Q
Figure 6-1 COMM Setup Menu
Communication Setup 37
6.2 Printer Setup
The RS-232 comm port is capable of outputing tension data. All the weight modes the Dyna-Link can measure are available in user formatted form. The control mode program is what causes the Dyna-Link 2 to print.
• USER – the assigned F-Key is pressed, then one transmission of the selected stri
ng type is output.
• On Load – when the tension on the link is stable, one transmission will output, then
the tension must return to zero to enable another print to output.
• Continuous – program the interval in seconds up to 65,535 seconds.
Setting the interval to 0 will set an interval as fast as the system can go. To disable pri
nting, simply don’t program an F-Key to print and set the control to “USER” or
turn the control mode to “OFF.”
PRINT STRINGS
6WUQ*
&QWUO
5DWH
2II 8VHU /RDG &RQW
Off
(default)
F key
On Stable Load
Continuous
PRINT CONTROL
Print String Output Rate
Number entry screen
(0-65535 seconds)
3ULQW
ENTER/SELECT
F1
Print String Format
Number entry screen
1= Current Wt
(Wt-Unit-Mode ©)
2= Net Wt
(Wt-Unit-Net
)
3= Gross Wt (Wt-Unit-Grs
)
4= Tare Wt (Wt-Unit-Tare
)
5= Total Wt (Wt-Unit-Total
)
6= Total Count (#Samples-TCNT
)
7= Current Wt (no units or mode
)
8=
(Reserved)
9= CR-LF (
)
0
ZERO
F1 F2
Affects continuous mode only
Press Simultaneously
Figure 6-2 Print Output Setup Menu
38 MSI-7300 Operator’s Manual
6.2.1 Standard Print Strings
1 Current
Te nsio n
Fixed output length: 16. Leading zer
os suppressed except for the
LSD.
TTTTTTTspUUspMMMMMcrlf
TTTTTTT – tension data with -sign if necessary. UU – units MMMMM – tension mode which for “1” is either NET or GROSS
2 Net Tension Fixed output length:16. Leading ze
ros suppressed except for the
LSD.
TTTTTTTspUUspNETspspcrlf
3 Gross
Te nsio n
Fixed output length: 16. Leading zeros suppressed except for the LSD.
TTTTTTTspUUspGROSScrlf
4 Tare We i g h t Fixed output length:16. Leading zeros suppressed except for the
LSD.
TTTTTTTspUUspTAR Ecrlf
5 To ta l
We
ight
Fixed output length: 16. Leading zeros suppressed except for the LSD.
TTTTTTTTTspUUspTTLcrlf
6 Number of
Samples To
ta led
Fixed output length: 16. Leading zeros suppressed except for the LSD.
spspspspspsp
SSSSSSSspT- C N Tspcrlf
7 Current
Weight Mode
Net, Gross, Peak, etc spMMMMMcrlf
8/9 Carriage
Retur
n/
Line Feed
Used to add a space between print records. crlf
Table 6-1. Standard Print Strings
In the string number entry screen, you can enter combinations of the standard print strings. For example,
to get a NET GROSS TARE printout with a space between
records, simply enter “2349.” Using SCCMP application (ScaleCore configuration management program),
custom
output strings are possible. See the ScCMP programming guide for details.
Note
ScCMP programming can be found on the CD included with the product or can be downloaded from the MSI website.
The serial output is configured as 9600 baud, Xon/Xoff handshaking, no hardware handshaking, 1 stop bit, no parity. Other baud rates are possible by special order only.
Communication Setup 39
6.2.2
1) With the 7300 on, press the F1 and F2 key simultaneously.
2) The LCD shows Print. Press
F1.
3) The submenu item StrnG appears. Press F1.
4) The current print mode format number is displayed.
5)
Set up a print format with one or more digits representing the data type required for the print output.
In this example, we’ll set the print format for a net, gross, tare output with a carriage return/line feed between each print output. The number entry required will be 2349. The 2 is for net weight, the 3 for gross weight, the 4 for tare weight, and the 9 inserts a space before the next print output.
6) Using the F2 Key, scroll through the digits until the desired digit is shown, then press F1 to enter the digit value. Repeat for the remaining digits.
7) When the entire number is displayed press F1. The next item in the print menu appears, Cntrl.
8) Press F1 to enter the print control menu. The last set control mode will appear.
9) To change the print control mode, press F2.
In this example, we’ll set the print control mode to continuous.
10) Press F2 key until the desired print control mode is shown.
11) When the desired print mode is shown, push F1 to save. The next print setup item, rAte appears. If you have set continuous (cont) as your print control, proceed to step 12). For any other control mode jump to step 15
0
ZERO
F1 F2
Press and Hold




blinking
blinking
fixed
fixed
fixed
&QWUO
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
Next Print Menu Item
3ULQW
6WUQ
ENTER/SELECT
F1
ENTER/SELECT
F1
blinking
SCROLL
F2
SCROLL
F2

ENTER/SELECT
F1
ENTER/SELECT
F1
xed

blinking
SCROLL
F2
SCROLL
F2
Example Value

blinking
SCROLL
F2

blinking
ENTER/SELECT
F1
2))
SCROLL
F2
2))
SCROLL
F2
8V(U
Function Key PrintsPrint is disabled
SCROLL
F2
/R$G
SCROLL
F2
&RQW
Function Key PrintsPrint is disabled
UDWH
Next Print Menu Item
ENTER/SELECT
F1
1= Current Wt
(Wt-Unit-Mode
)
2= Net Wt (Wt-Unit-Net ) 3= Gross Wt (Wt-Unit-Grs ) 4= Tare Wt (Wt-Unit-Tare ) 5= Total Wt (Wt-Unit-Total ) 6= Total Count
(#Samples-TCNT
)
7= Current Wt
(no units or mode
)
8=
(Reserved)
9= CR-LF ( )
Printer Output Setup
Figure 6-3 Printer Output Setup
12) Press F1 to enter the print rate number entry screen. The current print rate appears on the LCD.
In this example, we’ll set the print rate to an output rate of once every two seconds.
13) Press F2 to change the print rate. Use F2 to scroll the number, enter each number with the F1 key.
14) When the entire number is displayed, press F1 again to finalize the seconds entry. The next print menu Item appears, String.
15) Exit the print setup menu by pressing ZERO twice. The message
Store appears briefly
then normal link operation starts.
ENTER/SELECT
F1

blinking
Example Value

blinking
SCROLL
F2

blinking
fixed
ENTER/SELECT
F1
SCROLL
F2
ENTER/SELECT
F1
6WUQ
0
EXIT/SAVE
ZERO
0
EXIT/SAVE
ZERO
U) 6WRUH
The number represents the print interval rate in seconds. Use 0 for the fastest the Dyna-Link can go (~5 times per second). To set a once a minute transmission, set the rate to
60. To set once per hour, set the rate to 3600, etc. Largest possible entry is 65535 which is >18 hours.
40 MSI-7300 Operator’s Manual
Figure 6-3 Print Output Setup (continued)
Communication Setup 41
6.3 Comm Port Hardware
The Dyna-Link 2 RS-232 comm port is used for software updates, connecting to a remote display, and for connecting to any RS-232 device.
Table 6-2. Comm Port Hardware
Connector M12 industrial IP67 rated. An adapter cable (P/N 503363) is
required to connect the Dyna-Link 2 to a computer. This adapter cable converts the
Dyna-Link 2 connector to a
standard D9 serial connector. Data Configuration The data output is fixed at 8-1-N. Baud Rate Programmable for 300 to 230.4k baud in eight steps. The
bootloader for updating softwar
e is always 38.4k baud.
Handshaking No hardware handshaking is supported XON/XOFF softwar
e
handshaking is always on.
It may be necessary to disconnect the shield drain wire at the D-9 connector frame to prevent ground loops. Ground loops can cause unstable readings. In extreme cases it may be necessary to use an opto-isolated RS-232 interface.
Note
Figure 6-4 Serial Cable Schematic, DCE Configuration for
Connecting to a Computer
Figure 6-5 Serial Cable Schematic,
DTE Configuration for Connecting Directly to a DCE Printer
42 MSI-7300 Operator’s Manual
6.4 802.15.4 RF Network Setup
When equipped with the 802.15.4 option, the Dyna-Link 2 can connect with an MSI-8000 Remote Display or
an 802.15.4 modem. The unit uses three numbers to
connect to an 802.15.4 piconet:
1. ScaleCore ID – uniquely identifies each ScaleCore device in a
piconet. It has a
range of 0-254 and must not be duplicated within the same RF channel. For the
MSI-8000 as network coordinator, MSI recommends a number for the Dyna­Link 2 from 0-3 if multiple units will be connected to the MSI-8000. If a single Dyna-Link 2 is all that’s needed than any number up to 254 is acceptable.
2. RF Channel – establishes the base network, all in
terconnected devices must
match. This number must be in the range of 12-23.
3. Network ID – this is a 64-bit number that all interconnected devices must match. The Dyna-Link 2 li
mits this number to a max of 5 digits for a range of 0 - 99999. Do not use a small number here to help avoid other 802.15.4 networks that default to a network ID of 0.
4. RF Strength – Transmission strength can be set from 0 to 4, default is 1. The settin
gs effect the transmission range with zero is lowest power level and four is the highest. Power 4 will use the battery life quicker, so use the lowest number possible for reliable transmission. If maximum range is needed set the strength to four.
3ULQW
U)
Serial Out Setup
rF Modem
2Q2II
6&LG &KQO
QHWLG
RF On/Off
ScaleCore ID
RF Channel
Network ID
2Q
2II
RF On
RF Off
RF SETUP
Range 0-254
Range 12-23
Range 1 to 999999
0
ZERO
F1 F2
Press Simultaneously
6WU(Q
Transmission Strength
Range 0 to 4
Note
For all devices that interconnect, the RF channel and network ID must match. The ScaleCore ID must be unique. The Dyna-Link 2 or other MSI RF equipment that is a weight data source should be set to a ScaleCore ID of 0, then if other slave devices are added, they can be added in sequence.
Figure 6-6 RF Menu
2Q2II
2Q
3ULQW
RF Setup Procedure
VFLG
U)
example SCID Value
 

blinking blinking
Error Correction: If you input a wrong value, press ZERO to step back one digit and reenter.
&KQ/
blinking
1) With the power on, open the COMM setup menu by pressing the F1 and F2 keys simultaneously.
The COMM setup menu
choice is Print.
2) Press F2 to scroll to the RF setup menus.
3) Press F1 to enter the RF setup menu.
4) The display reads OnOff
by pressing F1. If it is off, use the F2 key to change it to on. Then push F1.
Off is only used when the 8000 is hardwired to a Dyna-Link.
5) Press F1. Next the Scid (ScaleCore ID) screen is shown
6) Press F1. The current SCID number is shown.
If the offered SCID value is correct, push F1 and jump to step 10.
In this example we’ll enter 3 as an SCID number. Any value from 1-254 is acceptable. However, MSI recommends a value from 20 – 30.
7) To input the SCID value, press F2 to start the number entry process.
8) Use F2 to change the number, and F1 to enter the number. Add the next digit by pushing F2 and scrolling as required. Repeat this sequence until the entire SCID number is entered.
9) Once the entire SCID value is entered, press F1
10) The next RF setup
parameter is “ChnL” (Channel). Press F1 to start the number entry process.
11) The current RF channel
number is displayed.
If the offered RF Channel value is correct, push F1 and jump to step14.
SCROLL
F2
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
F2
F2
SCROLL
F2
ENTER/SELECT
F1
ENTER/SELECT
F1
Press Simultaneously
F1 F2
Communication Setup 43
Figure 6-7 RF Setup Procedure
12) Press F2 and using the
previous process, enter the RF channel number.
Recommended RF Channel numbers range from 12 to 23. In this example we’ll use 21 as the RF Channel.
13) When the desired RF
channel number is shown and fixed, press
F1.
14) The next RF setup menu
item is the nEtid (Network ID). Press F1 to see the current net ID number.
If the offered network ID value is correct, push F1 and jump to step 17.
15) Using the numeric entry
process as before, input a network ID number. Press F2 to start a new number.
Allowed network ID numbers range from 0 to 99999. MSI recommends a random number of at least four digits to ensure that the 8000 system won’t conflict with any other 802.11.4 networks (Zigbee).
16) When the final number is
shown and fixed, press
F1 to store the network
ID.
17) The RF setup menu goes to the first menu, OnOff. Press the ZERO key to exit out of the RF setup menus.
18) Press ZERO again to exit the COMM setup menu and store the new RF network numbers.
QHWLG
 
blinking
blinking
fixed

blinking
blinking
fixed

current Net ID blinking
blinking

example value only
6WU(Q
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
SCROLL
F2
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
Repeat number entry procedure
3ULQW
6WRUH
2Q2II
ENTER/SELECT
F1
0
EXIT/SAVE
ZERO
0
EXIT/SAVE
ZERO
blinking
ENTER/SELECT
F1
17)
The menu displays StrEn. Press the F1 to enter.
Press F2, using the numeric entry process, input a transmission strength number.
16) When the desired number
is displayed, press F1 to store the number.
18)
When setting
StrEn
, use the
lowest setting possible to achieve the transmission required. Both the Scale/Dyna-Link and the 8000 should be set at the same transmission strength setting.
Note
44 MSI-7300 Operator’s Manual
RF Setup Procedure (continued)
Communication Setup 45
6.5 FCC Statement (For 802.15.4 Option)
Contains FCC ID: OUR-XBEEPRO The enclosed device complies with part 15 of the FCC rules. Operation is subject to
the following two conditions: (i.) this device may not cause harmful interference and (ii.) this device must accept any interference received, including interference that may cause undesired operation.
6.6 International RF CERTS (For 802.15.4 Option)
Canada Radio Cert. No.: IC: 4214A-XBEEPRO Australia & New Zealand: AS4268:3000 Japan: Certificate of Radio Equipment in Japan No.: 08215111/AA/02 Europe and much of Asia:
This product is compliant with the following standards and/or other normative documents:
Safety (article 3.1A) EN60950-1:2001
EMC (article 3.1b) ETSI EN 301 489-1 v1.7.1 (2007-04) In accordance with the
specific requirements of ETSI EN 301 489-17 v1.2.1 (2002-08)
Spectrum (article 3.2) ETSI EN 300 328 v1.7.1 (2006-10)
46 MSI-7300 Operator’s Manual
7.0 Appendix
7.1 Troubleshooting
Problem Possible Cause Solution
Display is blank when POWER key is pressed
Discharged battery Replace cells, or if using NiMH,
recharge Defective battery Replace Corroded battery or battery
contacts
Clean contacts
Defective switch or circuit board
Requires authorized service
Display does not function properly or front panel keys do not function normally or Dyna-Link will not turn off
Improperly loaded software Reinstall software Faulty circuit board Requires authorized service Loose connectors Requires authorized service
Dyna-Link does not respond to tension changes
Out of calibration Calibrate Faulty load cell Replace Load cell connector Check connector and wires
Display over ranges below 100% of capacity
Tared tension is added to load to determine overload point
Return to gross tension mode
Zero requires adjustment Rezero the Dyna-Link Too much tension/load has
been zeroed
Rezero the Dyna-Link
Display drifts AZM (Auto0) is turned off Tu r n A Z M o n
Rapid temperature changes such as moving the Dyna­Link from indoors to outdoors
Wait until the Dyna-Link
temperature has stabalized
Displayed tension shows larger error
Dyna-Link not zeroed before load is lifted
Zero the Dyna-Link with no load
attached lb/kg units causing confusion Select proper units Requires recalibration Recalibrate
Display reading not stable Excessive vibration in crane
system
Increase filtering or increase ‘d’
in Cal Excessive side loading Improve load train symmetry Load cell faulty Check LC connections
Display toggles between “Error” and “Load”
Tension exceeds capacity Immediately reduce tension Faulty load cell or wiring Check LC and LC wiring.
Table 7-1. Troubleshooting
Appendix 47
Display toggles between “Error” and “buttn”
A key is stuck or is being held down
Check switches for damage
Ensure that a remote is not
continuously transmitting
Optional RF Remote display does not work
Units not mated See “Setting the Transmitter
and Receiver Address”
procedures.
Lo Batt is blinking Battery is low Replace (alkaline) or recharge
batteries
Unit turns on, then immediately off
Battery is low Replace (alkaline) or recharge
batteries
Tension will not zero System not stable Wait for stable symbol to turn
on
Increase filtering for more
stability Zero out of range Zero range might be limited.
Reduce the tension or use tare
instead
Tension will not tare or total
System is not stable Wait for stable symbol to turn
on, or if in a mechanically noisy
crane, increase the filtering or
reduce the size of the Dyna-
Link increment “d.” It is also
possible to increase the motion
window. Contact MSI if you
have a problem getting the
MSI-7300 to zero, tare, or total
due to stability issues.
Setpoint lights blink Setpoint is enabled and the
trigger point has been reached
Disable setpoints if they are not
needed
Manual total does not work
A function key is not set to “Total”
Set up Func1 or Func2 for
“Total” Tension must be stable Increase filtering for more
stability
Auto total does not work Tension must be stable Wait for stable symbol to turn
on, or increase filtering for more
stability Tension thresholds not
reached
You must exceed 1% of
capacity for autototal to work.
Then you must drop below
0.5% of capacity for additional
weighments to register.
Problem Possible Cause Solution
Table 7-1. Troubleshooting
48 MSI-7300 Operator’s Manual
7.2 Error Codes
The ScaleCore processor that is the heart of the MSI-7300 Dyna-Link2 detects errors and generates error codes to aid in troubleshooting.
Table 7-2. Error Codes
Error Code Definition Comment
LcOFF LC Disabled A load cell was not enabled
2CAL In Cal The system is in calibration mode.
Do not send commands unrelated to calibration.
unCAL Not Calibrated System has not been calibrated.
Error
Load
Overload Tension/Weight exceeds set
capacity +9d
Load Cell damaged or disconnected
Error
UnLd
Underload Tension/Weight is more than 20%
negative
Load cell damaged or disconnected
Appendix 49
7. 3
5000 lbs / 2500 KG
MAXIMUM CAPACITY
7300
lb
B
C = DIA.
E = LOAD CELL BODY WIDTH AT SHACKLE INTERFACE.
A
D
G
F
Mechanical Dimensions
Figure 7-1. Mechanical Dimensions
Table 7-3. Mechanical Dimensions
Capacity A B C D E F G
Approx
Shipping
Wt
Shackle
1000 lb 8.0 in 13.53 in .75 in 5 in .99 in 1.69 in 1.75 in 4.4 lb
G-2130 3.25T
500 kg 203 mm 344 mm 19 mm 127 mm 25 mm 43 mm 44.4 mm 2.0 kg
2500 lb 8.5 in 14.03 in .75 in 5 in .99 in 1.69 in 1.75 in 4.9 lb
G-2130 3.25T
1250 kg 216 mm 356 mm 19 mm 127 mm 25 mm 43 mm 44.4 mm 2.2 kg
5000 lb 8.5 in 14.03 in .75 in 5 in .99 in 1.69 in 1.75 in 4.9 lb
G-2130 3.25T
2500 kg 216 mm 356 mm 19 mm 127 mm 25 mm 43 mm 44.4 mm 2.2 kg
10,000 lb 8.5 in 16.14 in 1 in 5.50 in 1.35 in 2.28 in 2.34 in 7.3 lb
G-2130 6.5T
5000 kg 216 mm 410 mm 25 mm 140 mm 34 mm 58 mm 59.4 mm 3.3 kg
25,000 lb 9.5 in 22.66 in 1.63 in 6.38 in 2.24 in 3.88 in 4.69 in 13 lb
G-2130 17T
12,500 kg 241 mm 576 mm 41 mm 162 mm 57 mm 99 mm 119 mm 5.9 kg
50,000 lb 9.63 in 25.67 in 2 in 7.50 in 2.74 in 5.00 in 5.75 in 23 lb
G-2130 25T
25,000 kg 245 mm 652 mm 51 mm 191 mm 70 mm 127 mm 146 mm 10 kg
100,000 lb 12 in 29.75 in 2.25 in 8.13 in 3.11 in 5.75 in 4.81 in 53 lb
G-2140 55T
50,000 kg 305 mm 756 mm 57 mm 207 mm 79 mm 146 mm 122 mm 24 kg
50 MSI-7300 Operator’s Manual
7.4 Firmware Update Procedure
Updating firmware in the MSI-7300 requires the following: a DCE serial cable (MSI 503363-001, or build per DCE cable schematic on page 22), a PC with a terminal program (“Teraterm Pro” recommended), and if the PC does not have standard RS-232 serial ports, then a USB to serial converter. Make sure the driver for the USB converter is properly installed, and that the terminal program is set up for the proper comm port.
The latest firmware code is available from the MSI service department and can be emailed on request. Your firmware version is displayed when the MSI-7300 is turned on in the form of “01-04” (your version will vary). Most firmware updates do not require a recallibration. Consult the version release notes for confirmation.
1. Setup the terminal serial port to 8 data bits, no parity, 1 stop bit, 9600 BAUD, XON/ XOFF (flow control).
2. Connect to the Dyna-Link serial port using the DCE cable. Connect the D9 connector to your PC or USB adapter.
3. (Optional) Test that you have a connection by typing {00FF01?}. If the connection is good the Dyna-Link will respond with {000001r2;0;01E02;2011-07-08;11:05} or something similar.
4. On the terminal keyboard, type {ff0009=0199}
5. Change the terminal serial port to 38400 BAUD. Hit the ‘q’ key to refresh the display. Cycle Power on MSI-7300 by removing and reinstalling batteries. The following menu should appear
MSI-7300 SCALECORE 1 BOOT LOADER Ver. 02-04(c) 2011-09-02 17:06
(u) Download and program application code (q) query app code info (g) execute app code (r) refresh
Your bootloader version may vary.
6. Type u
Terminal should display:
Send File NOW, or press ^ to abort:
7. Send the .prg file using the file send feature of your terminal program. The character “#” will tick away as the ScaleCore programs.
Send File NOW, or press ^to abort:##################################### ################################################################## ################################################################## ################################################################## ###########################################################
Completed
Note
Appendix 51
8. After the file is received, terminal should display “Completed.” Then the boot menu appears again.
MSI8000 SCALECORE2 BOOT LOADER Ver. 02-04 (c) 2011-09-02 16:06
(u) Download and program application code (q) query app code info (g) execute app code (r) refresh
9. Optional step: send q to check the program. The ScaleCore will respond with a message that details the 32b checksum, the product ID and version, and the
application code version number in the following form:
Computed Signature BOB742D Received Signature BOB742D Product ID 07 Dyna-Link product family Product Version ID 00 Optional features code App Code Version 01-04 Firmware version number
If the CRC Signature does not match, go back to step 4 and try again.
10.Send an “r” to restore the boot menu.
MSI8000 SCALECORE2 BOOT LOADER Ver. 00-05 (c) 2012-05-02 10:55
(u) Download and program application code
(your bootloader version may vary) (q) query app code info (g) execute app code (r) refresh
11.Send a “g.” The MSI-7300 should start.
32b CRC must match
52 MSI-7300 Operator’s Manual
The MSI Limited Warranty
MEASUREMENT SYSTEMS INTERNATIONAL, INC., WARRANTS load sensing elements and meters against defects in workmanship and materials for a period of one year from date of purchase and warrants electrical cables and batteries against the same defects for a period of ninety (90) days from date of purchase.
Any device which proves defective during the warranty period will be replaced or repaired at no charge; provided that the defective device is returned to the Company freight pre­paid.
In no event shall the Company be liable for the cost of any repairs or alterations made by others except those repairs or alterations made with its specific written consent, nor shall the Company be liable for any damages or delays whether caused by defective workmanship, materials or otherwise.
The Company shall not be liable for any personal injury or property damage resulting from the handling, possession or use of the equipment by the customer.
The warranty set forth herein is exclusive and is expressly in lieu of all other warranties, express or implied, including without limitation any implied warranties of merchantability or fitness, or of any other obligations or liability on the part of the Company.
The liability of the Company under this warranty is limited solely to repairing or replacing its products during the warranty periods; and the final judgment and disposition of all claims will be made by MEASUREMENT SYSTEMS INTERNATIONAL, INC.
Appendix 53
Notes
54 MSI-7300 Operator’s Manual
Notes
A RICE LAKE WEIGHING SYSTEMS COMPANY
Measurement Systems
International
TM
14240 Interurban Avenue South Suite 200 Ɣ Seattle, WA 98168-4661 Ɣ USA
Phone: 206-433-0199 Ɣ Fax: 206-244-8470
www.ricelake.com www.ricelake.mx www.ricelake.eu www.ricelake.co.in m.ricelake.com
© Rice Lake Weighing Systems 06/2015 PN 152160 Rev C
Loading...