Digital Stat-Arc 2 IS for Hazardous Locations
model 282IS
Digital Stat-Arc 3
Description
The Digital Stat-Arc 2, model 282IS and Digital
Stat-Arc 3, model 282A-1 are pocket-sized, noncontacting static meters which produce consistently
accurate readings, and are easy to use.
Features:
••
• Pulsing-beam range finder for always ob-
••
taining the correct reading
••
• Exclusive Auto-Zero circuit
••
••
• HOLD button captures transient readings
••
••
• Recorder output and 40-hour battery
••
••
• Drift-free accuracy even in ionized envi-
••
ronments
••
• Charged plate adapter available
••
••
• Model 282IS - Intrinsically safe (IS) meter
••
UL Certification Class I Groups A, B, C, D
and Class II Groups E, F, G
The right answers anywhere -
easier than ever
With the Model 282, accuracy is unquestioned.
Simply hold the instrument so the range finder LED
beams form a non-pulsing circle indicating you’re at
exactly the right distance from the target for readings
up to 20 kV . Additionally , Monroe’s chopper-stabilized
circuitry is immune to ionization, unlike ordinary static
locators. Zeroing is push-button simple.
Designed for optimal performance as well as low
cost you can’t afford not to keep the Model 282 handy
for:
• Monitoring “static-free” electronic
workstations
• Measuring static buildup on webs in
converting, laminating, and printing operations
• Checking grounds and bonding in dry
particle/powder transport systems
• Testing ionizer performance using optional
charged plate adapter Model 282CPS
model 282A-1
Simple to use:
1. Turn the instrument ON.
2. Discharge your body by touching a grounded
metal object.
3. Point the aperture toward a grounded object
and press the ZERO button.
4. Aim the aperture toward the target surface at a
distance of 1 inch. Adjust the distance until
the flashing beams of the LEDs in the instrument converge. Read the voltage and polarity
of the charged surface on the meter display .
For additional measurements repeat step four only .
T o freeze the display, press the HOLD button. T o
read the voltages, start at a greater distance to the
target as given under specifications.
Maintenance
The battery should be replaced annually , whenever
you plan an extended period of unattended monitoring,
or whenever “BAT” appears in the display for more than
an instant. T o obtain accurate and drift free readings
the sensor plate and especially the area around the
aperture must be kept absolutely clean at all times.
Never touch the aperture with anything – not even a
cotton swab.
Digital Stat-Arc 2 IS model 282 IS for Hazardous Locations
Digital Stat-Arc 3 model 282A-1
Specifications:
Display: LCD, 3½-digit with auto polarity
readout, with HOLD and LOW BA TT
indicators
Range: 0 to ± 19.99kV at 1 inch.
Voltages over 20kV and higher may
be read by increasing the distance to
the target
kV Distance reading by
0-40 4.0 inches 2
0-80 8.5 inches 4
Accuracy: ±5% of reading, + zero offset, ±2 Isd
Analog output 1 V signal denotes 10kV reading
Amplitude: at 1 inch for high impedance loads
Response Typ. 80 - 100 msec 10 - 90%
Time:
Jack Type: Accepts standard 3/32 inch
(2.5mm) monaural phone plug
Battery: 9V NEDA #1604 or equivalent,
Life: 40 hours of normal use, with
alkaline battery
Physical: (L x W x H): 2.4 x 4.2 x 0.9 inches
(6.1 x 10.7 x 2.3 cm) Weight: 5 oz
(0.14kg) with battery
Operating 0-50°C, o 85% RH
Environment: (non-condensing),
unaffected by ionized equipment
Multiply
Calibration:
Monroe Electronics instruments are factorycalibrated prior to shipment. Recalibration should be
performed annually , or more frequently if specified by
contract or company policy . Your instrument should
also be recalibrated any time it has been repaired or
tampered with. We are happy to recalibrate your
instrument for you at a reasonable cost, or provide
information and procedures on calibration upon request.
Warranty:
Monroe Electronics, Inc., warrants that each
instrument and sub-assembly manufactured by them
shall be free from defects in material and workmanship
for a period of two years after shipment from the factory .
This warranty is applicable to the original purchaser
only.
The finest ESD instrumentation and
support:
For more than 40 years - ever since we invented
the feedback--nulled electrostatic voltmeter, Monroe has
been the technology and quality leader in electrostatic
detection and measurement instrumentation. T oday we
offer the world’s most complete array of fieldmeters,
voltmeters, and resistivity meters. Our customers
include the leading makers of photocopiers and laser
printers, converters and microelectronics worldwide.
We know you need quality support as well as
quality products. We pride ourselves on providing our
customers with the most knowledgeable applications
and installation support — as well as superior customer
service.
282IS CE mark approval
Industrial UL Certification Class I Groups A, B,
Approval and C, D, and Class II Groups E, F, G,
certification:
when powered by 9-volt carbon zinc
or zinc chloride battery NEDA 1604
or NEDA 1604D.
UL Certification for Class I Groups C
and D only when powered by 9-volt
alkaline NEDA 1604A.
How can we help?
Contact your Monroe Electronics representive
for price and delivery information on this and other ME
products, to schedule a no-obligation demonstration at
your convenience. For the name of your nearest dealer,
or for technical or applications assistance, contact
Monroe Electronics directly at the address and numbers
below.
For UL approved, Intrinsically Safe meters, customer
must specifically request model 282IS when placing
order.
www.monroe-electronics.com
100 Housel Ave. | Lyndonville | NY | 14098
800-821-6001 | 585-765-2254 | fax 585-765-9330
Specifications subject to change without notice. | Printed in USA | Copyright© 2002 | Monroe Electronics Inc. | 282IS+A ds 022403.p65