Milwaukee T75, T76 User Manual

SPECIFICATIONS
Range
Resolution
Accuracy (@20°C) Typical EMC
Deviation Temperature
Compensation TDS factor Calibration
Probe
Environment Battery type Battery life
Dimensions Weight
T75 0 to 1999 ppm (mg/L) T76 0 to 9990 ppm (mg/L) T75 1 ppm (mg/L) T76 10 ppm (mg/L)
±2% F.S. ±2% F.S.
Automatic, 0 to 50°C with β=2%/°C
0.5 Manual, at 1 point
through trimmer
T75 MA73075 (replaceable) T76 MA73076 (replaceable)
0 to 50°C; 100% RH 3 x 1.5V , alkaline approx. 250 hours of
continuous use 165 x 30 x 30 mm 80 g
USER MANUAL
T75 and T76
Waterproof TDS meters
ACCESSORIES
MA73075 Replaceable Conductivity
probe, LR
MA73076 Replaceable Conductivity
probe, HR
M10032B 1382 ppm (mg/L) calibration
solution, 20 mL (25 pcs)
M10038B 6.44 ppt (g/L) calibration
solution, 20 mL (25 pcs)
ISTRT75R1 06/02
This meter is in compliance with the CE directives.
CALIBRATION:
• Immerse the meter in the proper calibration solution: 1382 ppm (M10032) for T75 and 6.44 ppt (M10038) for T76.
• Unscrew the battery compartment cap on the top of the meter.
• Allow the reading to stabilize and with a small screwdriver turn the calibration trimmer inside the bat­tery compartment until the dis­play shows "1382" for T75 or “6440” for T76.
• Replace the battery compartment cap.
The calibration is now complete.
OPERATION:
• Remove the protective cap.
• Turn the meter on by pressing the ON key.
• Immerse the probe in the solu­tion to be tested.
• Stir gently and wait for the read­ing to stabilize.
• After use, rinse the electrode with water to minimize contami­nation.
• Always replace the protective cap after use.
Note: if the reading is over-
range, T75 will display “1” while T76 shows “. . .”.
PROBE REPLACEMENT:
The probe can be easily replaced in the following way:
• Remove the protective cap.
• Unscrew the plastic ring on the top of the probe.
• Pull out the MA73075 (or MA73076) probe and replace it with a new one (aligning the
colored pin towards the LCD
.
side)
• Make sure the gaskets are in place before screwing back the plastic ring.
BATTERY REPLACEMENT:
The meter is supplied with BEPS (Battery Error Preventing System) which avoids any erroneous reading due to low battery level, by switching the meter off. To replace the batteries unscrew the battery compartment cap and replace all three 1.5V batteries while paying attention to their polarity. Make sure the gasket is in place before screwing back the cap. Batteries should only be replaced in a non­hazardous area using the battery type speci­fied in this instruction manual.
WARRANTY:
These instruments are warranted from all defects in materi­als and manufacturing for a period of two years from the date of purchase. The probes are warranted for a period
of one year.
If during this period, the repair or replacement of parts is required, where the damage is not due to negligence or erroneous operation by the user, please return the parts to either dealer or our office and the repair will be effected free of charge.