Milwaukee SM801, SM802 User Manual

Milwaukee
WARRANTY
These instruments are warranted from all defects in materials and manufacturing for a period of two years from the date of purchase.
The probes are warranted for a period of six months. If during this period the repair or replacement of parts
Note: We reserve the right to modify the design,
construction and appearance of our products without advance notice.
Milwaukee
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment.
If noticeable damage is found, notify your Dealer. Each meter is supplied complete with:
MA850 combination, amplified, double-junction, gel pH-electrode with incorporated EC/TDS probe and built-in temperature sensor; 1 m cable
• pH7.01, 1413 µS/cm & 1382 ppm (for SM801) or 1500 ppm (for SM802) solutions, 20 mL each
• 1x9V battery
• Instruction Manual.
OPERATION
Connect the probe to the meter securely by aligning the pins with the socket and pushing the plug in.
Always remove the electrode pro­tective cap before use. Make sure that the meter has been calibrated before taking any mea­surements.
Turn the instrument on by pressing the ON/OFF key.
pH MEASUREMENTS
If the probe has been left dry, soak the tip in a pH7 (M10007) buffer solution for a few minutes to reacti­vate it.
Submerge the tip into the sample to be measured, select the pH mode and allow the reading to stabilize before taking any measurement.
EC/TDS MEASUREMENTS
Immerse the tip of the probe into the sample to be tested.
Select the desired measurement range (EC or TDS) and wait for the temperature sensor to reach thermal equilibrium with the sample before taking any measure­ment.
AFTER MEASUREMENTS, switch the instrument off, clean the probe and store it with a few drops of storage solution in the protective cap.
CALIBRATION PROCEDURE
Remove the protective cap from the probe. Switch the meter on. Select the desired range (pH, EC, TDS) by pressing the appropriate key.
pH Calibration
The instrument can be manually calibrated at one point for pH.
A) PREPARATION
If the samples to be measured are neutral or close to neutral pH, to obtain accurate readings use pH
7.01 buffer solution (M10007). The pH 4.01 buffer (M10004) is suit­able for acidic measurements, while the pH 10.01 (M10010) buffer should be used for alkaline samples.
B) PROCEDURE
Rinse the tip of the probe and immerse it in the buffer chosen for calibration. Wait a couple of minutes for the reading to stabilize.
With an accurate thermometer measure the temperature of the buffer solution.
Adjust the calibration knob until the LCD shows the pH value at the measured temperature (see the “pH versus temperature” chart for reference).
The calibration is now complete and the meter is ready for use.
USER MANUAL
PORTABLE pH/EC/TDS METER
MODEL: SM801. SM802
Smart Combined Meter
Milwaukee
SPECIFICATIONS
SM801 SM802
RANGE 0.0 to 14.0 pH 0.00 to 14.00 pH
0 to 1990 ppm 0 to 4000 ppm 0 to 1990 µS/cm 0.00 to 6.00 mS/cm
RESOLUTION 0.1 pH 0.10 pH
10 ppm 10 ppm 10 µS/cm 0.01 mS/cm
ACCURACY ± 0.2 pH ±0.20 pH (@ 20ºC) ±2% Full scale
±2% Full scale
CAL.SOLUTIONS M10007 (pH 7.01) M10031 (1413 µS/cm)
M10032 (1382 ppm) M10442 (1500 ppm)
CONVERSION 0.5 0.68 FACTOR
CALIBRATION Manual, at 1 point
TEMPERATURE COMPENSA TION
Automatic, from 0 to 50ºC
PROBE MA850 combination pH/EC/TDS probe ENVIRONMENT 0 to 50ºC, 95%RH max
BATTERY TYPE 1x9V alkaline BATTERY LIFE approx. 150 hours of use
DIMENSIONS 185 x 82 x 45 mm WEIGHT 165 g
EC/TDS calibration
The instrument can be manually calibrated at one point for conductivity range.
A) PREPARATION
For performing TDS calibration, use 1382 ppm (M10032) solu­tion for SM801 and 1500 ppm (M10442) solution for SM802.
For performing EC calibration, use 1413 µS/cm (M10031) so­lution for both meters.
B) PROCEDURE Immerse the tip of the probe in the calibration solu-
tion. Wait for the reading to stabi-
lize, and then turn the EC/TDS calibration knob until the dis­play shows the EC or TDS reading at 25°C.
The calibration is now complete and the meter is ready for use.
Note: Since the conversion between EC and TDS is
made by a built-in circuit, it is request to calibrate the meter only in EC or TDS range, and the other range will be automatically cali­brated.
The instrument should be recalibrated at least once a month, or whenever the probe or battery is changed.
BATTERY REPLACEMENT
The meters are powered by a 9V battery that is located on the rear of the instrument.
When the battery becomes weak, the instrument auto­matically switches off.
Battery replacement must only take place in a non­hazardous area using an alkaline 9V battery.
Slide off the battery compartment cover and replace the rundown battery with a new one. Make sure that the battery contacts are tight and secure before in­serting the cover.
pH VERSUS TEMPERATURE CHART
ACCESSORIES
M10004B pH 4.01 buffer solution, 20
mL sachet (25 pcs)
M10007B pH 7.01 buffer solution, 20
mL sachet (25 pcs)
M10010B pH 10.01 buffer solution,
20 mL sachet (25 pcs)
M10031B 1413 µS/cm cal. solution,
20 mL sachet (25 pcs)
M10032B 1382 ppm cal. solution, 20
mL sachet (25 pcs)
M10442B 1500 ppm cal. solution, 20
mL sachet (25 pcs)
MA9015 Electrode storage solution,
220 mL bottle
MA850 pH/EC/TDS spare probe
with 1 m cable
ISMIL801 04/03
TEMP pH VALUES
ºC ºF
0 32 4.01 7.13 10.32 5 41 4.00 7.10 10.24 10 50 4.00 7.07 10.18 15 59 4.00 7.04 10.12 20 68 4.00 7.03 10.06 25 77 4.01 7.01 10.01 30 86 4.02 7.00 9.96 35 95 4.03 6.99 9.92 40 104 4.04 6.98 9.85 45 113 4.05 6.98 9.85 50 122 4.06 6.98 9.82 55 131 4.07 6.98 9.79 60 140 4.09 6.98 9.77 65 149 4.11 6.99 9.76 70 158 4.12 6.99 9.75
M10004B M10007B M100 10B