OPERATOR'S MANUAL
MANUEL de L'UTILISATEUR
MANUAL del OPERADOR
CLAMP METERS
PINCE AFFICHEUR DE COURANT
MEDIDORES DE PINZA
TO REDUCE THE RISK OF INJURY, USER MUST READ AND UNDERSTAND OPERATOR'S
MANUAL.
AFIN DE RÉDUIRE LE RISQUE DE BLESSURES, L'UTILISATEUR DOIT LIRE ET BIEN
COMPRENDRE LE MANUEL DE L'UTILISATEUR.
PARA REDUCIR EL RIESGO DE LESIONES, EL USUARIO DEBE LEER Y ENTENDER EL
MANUAL DEL OPERADOR.
IMPORTANT SAFETY INSTRUCTIONS
Hz
WARNING READ ALL SAFETY WARNINGS AND INSTRUCTIONS.
Failure to follow the warnings and instructions may result in electric shock, fi re and/or serious
injury, as well as instrument damage and/or damage to the equipment being tested.
Save these instructions - This operator’s manual contains important safety and operating
instructions for the MILWAUKEE Clamp Meters. Before using, read this operator’s manual
and all labels on the Clamp Meters.
DANGER
Never make measurement on a circuit in which voltage over AC600V exists. Use only leads
rated 600V or better.
Do not attempt to make measurement in the presence of fl ammable gasses. Otherwise, the
use of the instrument may cause sparking, which can lead to an explosion.
Transformer jaw tips are designed not to short the circuit under test. If equipment under test
has exposed conductive parts, however, extra precaution should be taken to minimize the
possibility of shorting.
Never attempt to use the instrument if its surface or your hand is wet.
Do not exceed the maximum allowable input of any measuring range.
Only test on unenergized circuits unless absolutely necessary.
Check tool functionality on a known circuit fi rst. Never assume tool is working. Assume circuits
are live until they can be proven de-energized.
Do not ground yourself while measuring. Avoid body contact with earthed or grounded surfaces such as pipes, radiators, ranges and refrigerators.
Never open the Battery cover during a measurement.
The instrument is to be used only in its intended applications or conditions. Otherwise, safety
functions equipped with the instrument doesn’t work, and instrument damage or serious
personal injury may be caused.
To reduce the risk of injury from shock and arc blasts, always wear personal protective
equipment where live conductors are exposed.
WARNING
Never attempt to make measurement if any abnormal conditions, such as broken case and
exposed metal parts are found on the instrument.
Do not rotate the Rotary Dial while the test leads are being connected.
Verify proper operation on a known source before use or taking action as a result of the indication of the instrument.
Do not install substitute parts or make any modifi cation to the instrument. For repair or re-
calibration, return the tool to a factory Service/Sales Support Branch or authorized service
station.
Do not try to replace the batteries if the surface of the instrument is wet.
Disconnect all the cords and cables from the object under test and power off the instrument
before opening the Battery Cover for Battery replacement.
This tool is designed to be powered by 2-AA batteries properly inserted into the MILWAUKEE
Clamp Meters. Do not attempt to use with any other voltage or power supply.
Install battery according to polarity (+ and –) diagrams.
Do not leave batteries within the reach of children.
Do not mix new and used batteries. Do not mix brands (or types within brands) of batteries.
Properly dispose of used batteries.
Do not incinerate or dismantle batteries.
Under abusive conditions, liquid may be ejected from the battery, avoid contact. If contact
accidentally occurs, fl ush with water. If liquid contacts eyes, additionally seek medical help.
Liquid ejected from the battery may cause irritation or burns.
2
CAUTION
Set the Rotary Dial to an appropriate position before starting measurement.
Firmly insert the test leads.
Disconnect the test leads from the instrument for current measurement.
Do not expose the instrument to the direct sun, high temperature and humidity or dew fall.
Altitude 2000m or less. Appropriate operating temperature is within -10ºC and 50ºC.
This instrument isn’t dust & water proofed. Keep away from dust and water.
Be sure to power off the instrument after use. When the instrument will not be in use for a long
period, place it in storage after removing the batteries.
Use a cloth dipped in water or neutral detergent for cleaning the instrument. Do not use
abrasives or solvents.
Functions
Dial PositionRangeResolutionAccuracy
Voltage DC60/600V0.01V/0.1V±1.0%rdg±3dgt
Voltage AC0.0 to 600.0V0.1V
0 to 600.0A
Peak 1500A
CF=2.5@600A
CF=3.0@500A
0-600.0A0.1A±1.5%rdg±5dgt
0.01nF - 4000F
Auto-ranging
Cont Buzzer
0-600.0
0-600.0uA0.1uA ±1.5%rdg±5dgt
-40C - 400C
-40F - 32.0F
32.1F - 752F
ACA: 40-400Hz,
ACV: 1Hz-10kHz
0.1A
0.01nF/0.1nF/
0.001F/0.01F/
0.1F/1F
0.1C
0.1F
0.001/0.01/0.1Hz
0.001/0.01kHz
DC uA (2236-20)
Hertz (2237-20)
Hz
Current AC
Current DC
(2237-20)
Resistance600/6k0.1/0.001k±1.0%rdg±5dgt
Capacitance
Continuity
uA Current
Temperature
(2236-20)
* These instruments are True-RMS sensing. All voltage and current readings are True-RMS values.
3
±1.5%rdg±4dgt (50/60Hz)
±3.5%rdg±5dgt (40~400Hz)
±2.0%rdg±5dgt(50/60Hz)
±3.5%rdg±5dgt(40 ~ 400Hz)
* Add 2% at CF>2
* < 5A, add 3dgt
0.01nF-39.99nF ±2.5%±2nF
40.00nF-1000uF ±2.5%±20dgt
1000uF> ±5%±20dgt
Buzzer sounds at 35 ±10 or
25 or less
±1.0%C ±2.0C
-40.0F to 20.0F; ±1.5%F ±4F
20.1F to 752F; ±1.5%F ±2F
ACA: 40-400Hz ±0.5%rdg±5dgt
ACV:1Hz-10kHz ±0.5%rdg±5dgt
* Minimum Measurable inputs
are: 70 Vrms for VAC or
50 Arms for ACA
General Specifi cationsFUNCTIONAL DESCRIPTION
Accuracy is specifi ed for 1 year after calibration, at
operating temperatures of 18°C to 28°C (64°F to
82°F), with relative humidity at 0 % to 85 %.
Maximum voltage between any terminal and
earth ground .... 600 V
Jaw Opening (maximum conductor size) ....
approx. 1.3” (33 mm)
Temperature ....
Operating: -10°C to 50°C (14°F to 122°F)
Storage: -40°C to 60 °C (-40°F to 140°F)
Temperature Coefficient .... 0.1 x (specified
accuracy)/°C (<18°C or >28°C)
Operating Altitude.... 2,000 meters
Drop Test .... 1 Meter
Battery .... 2 AA, NEDA 15 A,IEC LR6
Battery Life .... Approx. 30 Hours all lights on.
Safety Compliances .... EN61010-1,
UL 61010-1, EN61010-031 (Probes),
IEC 61010-2-32 (Clamp Assemblies),
IEC/EN 61010-1 2nd Edition for measurement
Category III, 600 V,
Pollution Degree 2, EMC EN61326-1
Certifi cations .... cULus, CE
Symbology
Read Operator’s Manual
Double insulation
Risk of electric shock
Indicates that this instrument can
clamp on bare conductors when measuring a voltage corresponding to the
applicable measurement category,
which is marked next to this symbol.
Earth
Danger, Warning, or Caution
Battery compartment
European Conformity Mark
Underwriters Laboratories, Inc.,
United States and Canada
Classifi cation of transient overvolt-
Cat III
ages, based on nominal line voltage
to earth.
1
Cat. No.
2236-20
2
3
4
5
6
7
8
1. Current sensing jaws
2. NCVD indicator
3. Jaw opening trigger
4. Rotary Dial
5. Display
6. Zero button
9
7. °F/°C button (2236-20)
Hz button (2237-20)
8. Terminal inputs
9. Min/Max button
10.Worklight LED
11.Hold button
Cat. No.
2237-20
ASSEMBLY
WARNING
To avoid an electrical hazard, turn the Rotary
Dial to OFF and disconnect the test leads
before replacing batteries.
Loading/Changing the Batteries
Replace batteries when the Low Battery indicator
is displayed.
1. Turn Rotary Dial to OFF
and disconnect the test
leads.
2. Unscrew and remove
battery door.
3. Insert two (2) AA batteries, according to the
polarity marked in the
battery compartment
4. Close the battery door and tighten screw securely.
4
11
10
OPERATION
WARNING
4
Only use MILWAUKEE test leads with
the MILWAUKEE Clamp Meters.
Inspect test leads before each use. Use clamp
meter to run a continuity test.
Before Use
Confi rm the Rotary Dial is set to the correct position,
the instrument is set to the correct measurement
mode, and the Data hold function is disabled. Otherwise, desired measurement cannot be made.
LCD Backlight
The LCD backlight will turn off after about 3 minutes
of inactivity. Push any button or turn the rotary dial
to turn the backlight on.
Making a Measurement
AC Current
DANGER To avoid electrical shock:
Never make measurement on a circuit
in which voltage over AC600V exists. Clamp
tips are designed not to short the circuit under
test. If equipment under test has exposed
conductive parts, however, extra precaution
should be taken to minimize the possibility of
shorting.
Do not use with the Battery Cover removed.
Disconnect the test leads from the instrument
for current measurement.
1. Set the Rotary Dial to position.
AC mark is displayed.
2. Press the jaw opening trigger to
open the jaws and clamp them
onto the conductor under test.
The reading is displayed.
Cat. No. 2237-20 only: Pressing
the “Hz” Key toggles the reading
between AC Current and Hz.
NOTE: Hz Function requires 50A
or more.
NOTE: Do not clamp over 2 or
more wires at the same time.
Irregular results will occur.
CAUTION Maximum conductor size
is about 1.3" diameter. During measurement, keep the jaws fully closed to ensure
accurate measurements.
DC Current (Cat. No. 2237-20 only)
1. Set the Rotary Dial to position.
DC mark is displayed.
2. With the jaws closed and without
clamping them around a conductor, press the ZERO key to zero
adjust the display.
3. Press the jaw opening trigger to
open the jaws and clamp them
around the conductor under test.
The reading is displayed.
NOTE: Do not clamp over 2 or
more wires at the same time.
Irregular results will occur.
4. Set the Rotary Dial to an appropriate position according to current under test.
5. Press the ZERO key again to release the ZERO
function.
CAUTION When current fl ows from
the display side to the underside of
the meter, the polarity is positive; fl ow from
underside to display side, the polarity is
negative.
AC Voltage
DANGER To avoid electrical shock:
Never make measurement on a circuit
in which voltage over AC600V exists.
Do not use with the Battery Cover removed.
Keep fi ngers away from jaws during mea-
surements.
1. Set the Rotary Dial to
position.
2. Connect the red test lead
to the V terminal and the
black test lead to the COM
terminal.
3. Connect the test leads to
the circuit under test. The
reading is displayed.
Cat. No. 2237-20 only:
Pressing the Hz Key toggles
the reading between AC
Voltage and Hz. NOTE: Hz
Function requires 70V or
more.
AC
CAUTION
Readings of frequency may fl uctuate or be
infl uenced in noisy environments.
DANGER To avoid electrical shock:
Never make measurement on a circuit
in which voltage over AC600V exists.
Do not use with the Battery Cover removed.
5
DC Voltage
DANGER To avoid electrical shock:
Never make measurement on a circuit in
which voltage over DC600V exists.
Do not use with the Battery Cover removed.
Keep fi ngers away from jaws during mea-
surements.
1. Set the Rotary Dial to position.
2. Connect the red
test lead to the V
terminal and the
black test lead to
the COM terminal.
3. Connect the red
test lead to the
positive (+) side
and black test
leads to the negative (-) side of the
circuit under test.
The reading is displayed. A reversed connection
is indicated as a negative value.
Resistance/Continuity/Capacitance
Measurements
DANGER
To reduce the risk of electric shock for
Resistance, Continuity, and Capacitance
measurements, never use the meter on an
energized circuit. Make sure a capacitor is
fully discharged before touching or attempting to make a measurement.
Do not use with the Battery Cover removed.
Resistance
1. Set the Rotary Dial to
position.
2. Connect the red test
lead to the V terminal and the black
test lead to the COM
terminal.
Confi rm “OL” is indi-
cated on the display,
and then short-circuit
the tips of test leads
to make the indication zero.
3. Connect the test
leads to both ends of the resistor under test.
4. The reading is displayed.
CAUTION
After shorting the test leads, the displayed
value may not be zero due to the resistance
of test leads themselves.
Continuity
1. Set the Rotary Dial to
position.
2. Connect the red test lead
to the V terminal and the
black test lead to the COM
terminal.
Confi rm “OL” is indicated on
the display, and then shortcircuit the tips of test leads
to make the indication zero.
A buzzer will sound.
3. Connect the test leads to
both ends of the conductor
under test. If the resistance under test is 35 ±10
or 25 or less, the buzzer will sound.
Capacitance
1. Set the Rotary Dial to position.
2. Connect the red test lead to the V terminal and
the black test lead to the COM terminal.
3. Discharge capacitor.
4. Connect the test leads to both ends of the capacitor under test.
5. The reading is displayed.
DC Current (Cat. No. 2236-20 only)
DANGER To avoid electrical shock:
Never make measurement on a circuit in
which voltage over AC600V exists.
Do not use with the Battery Cover removed.
1. Set the Rotary Dial to
position. DC mark
is displayed.
2. Connect the red test
lead to the V terminal and the black
test lead to the COM
terminal. Contact the
red test lead to the
fl ame sensor probe
and the black test
lead to the control
module.
3. Turn on the heating unit. The reading is displayed.
6
Flame
sensor
probe
Control
Module
Temperature (Cat. No. 2236-20 only)
WARNING
Never connect the Temperature Probe to an
energized circuit.
1. Set the Rotary Dial to
position.
2. Connect the K-type Temperature Probe to the input
terminal. The positive (+)
side of Probe should be
connected to V.
3. Place the probe sensor in
the desired location.
4. The reading is displayed.
Non-Contact Voltage Detection (NCVD)
DANGER
The LED may not be displayed due to installation condition of electrical circuit or equipment. Never touch the circuit under test to
avoid possible danger even if the LED for
NCVD is not displayed.
Check the functionality of LED on a wellknown power supply prior to measurement.
When the LED doesn’t light up, do not make
measurement.
NCVD indication is affected by external voltage, and how the meter is held or placed.
When the meter is on in any
function, the non-contact voltage detector will indicate with a
Red LED on the display when
an electric fi eld exceeding 90V
is detected. Place the edge of
the jaw labeled “Voltage Detector” near the electric fi eld.
CAUTION
When the Rotary Dial is set to , the room
temperature should be displayed. If anything
else is displayed, something may be wrong
with the meter. Stop using the meter immediately.
Worklight LED ON/OFF
To turn the light on and off, press the button.
HOLD Key
Data Hold Function - Freezes the value on the
display. Press the “HOLD” button to freeze the
reading. The reading will be held regardless of
subsequent variation in input. HOLD is displayed
with the reading. To exit Data Hold mode, press the
HOLD button again.
CAUTION
The Data Hold readings are released when the
meter enters Sleep Mode.
Sleep Mode
The clamp meter is automatically powered off in
about 20 min after the last Rotary Dial or button
operation. To reset, rotate the Rotary Dial to OFF.
If the display is still blank when a new Rotary Dial
setting is selected, replace the batteries.
The sleep mode is disabled when the MIN/MAX
function is selected.
The clamp meter does use battery power in sleep
mode. Be sure to switch the tool to OFF to conserve
battery power.
Over-fl ow indication
Any time the input exceeds the
measuring range “OL” or “-OL” is displayed.
CAUTION
Clamp the jaws around the conductor under
test and press the MIN/MAX button after an
appropriate range is selected by Auto-ranging function. ZERO and Hz keys are disabled
while MIN/MAX Function is active.
MIN/MAX Function
The MIN/MAX function can be
used during measurements of AC
or DC current, AC or DC voltage,
Temperature (2236-20 only), uA DC
(2236-20 only), and Resistance.
The MIN/MAX function does not
work in the Capacitance or Continuity measurements.
To measure the minimum or maximum of the function, set the dial to
the appropriate dial position and
then press the MIN/MAX button
to capture the maximum reading.
Press the MIN/MAX button again to
capture the minimum reading.
Press the MIN/MAX button to toggle between
minimum or maximum. The minimum or maximum
reading is displayed and held until the MIN/MAX is
turned off by holding in the MIN/MAX button for 2
seconds or changing the Rotary Dial position.
7
CAUTION
Auto-Ranging and MIN/MAX button are disabled when ZERO function is in use. ZERO
function operates only in AC Current , DC
Current and Capacitance .
ZERO Function
AC Current
/ DC Current : When the LCD
doesn’t read ZERO while the Jaws are closed,
press the ZERO Button to indicate ZERO before
starting a measurement. In this case, the “ZERO”
mark appears on the LCD
to indicate the ZERO Function is activated.
Capacitance : Effective
at 40nF Range only. Press
the ZERO Button with the
Test leads opened before
measuring capacitances to
indicate ZERO before starting a measurement.
ACCESSORIES
WARNING Always remove batteries
before changing or removing accessories. Only
use accessories specifi cally recommended for
this tool. Others may be hazardous.
For a complete listing of accessories refer to your
MILWAUKEE Electric Tool catalog or go online
to www.milwaukeetool.com. To obtain a catalog,
contact your local distributor or service center.
MAINTENANCE
WARNING To reduce the risk of injury ,
always remove the batteries from the tool
before performing any maintenance. Never
disassemble the tool. Contact a MILWAUKEE
service facility for ALL repairs.
Maintaining Tool
Keep your tool in good repair by adopting a regular
maintenance program. After one year, it is recommended to return the tool to a MILWAUKEE service
facility for calibration.
If the tool does not start or operate at full power
with fully charged batteries, clean the contacts
on the battery door. If the tool still does not work
properly, return the tool to a MILWAUKEE service
facility for repair.
WARNING To reduce the risk of per-
sonal injury and damage, never immerse
your tool in liquid or allow a liquid to fl ow
inside them.
Cleaning
Clean dust and debris from the tool. Keep tool
handles clean, dry and free of oil or grease. Use
only mild soap and a damp cloth to clean the tool
since certain cleaning agents and solvents are
harmful to plastics and other insulated parts. Some
of these include gasoline, turpentine, lacquer thinner, paint thinner, chlorinated cleaning solvents,
ammonia and household detergents containing
ammonia. Never use fl ammable or combustible
solvents around tools.
Repairs
For repairs, return the tool to the nearest service
center listed on the back cover of this operator's
manual.
FIVE YEAR TOOL LIMITED WARRANTY
MILWAUKEE T est & Measurement Product (including bare tool, battery pack(s) and battery charger) is warranted to the original
purchaser only to be free from defects in material and workmanship. Subject to certain exceptions, MILWAUKEE will repair
or replace any part on this product which, after examination, is determined by MILWAUKEE to be defective in material or
workmanship for a period of fi ve (5) years* after the date of purchase. Return of the Test & Measurement tool to the nearest MILWAUKEE Electric Tool Corporation - factory Service Center, freight prepaid and insured is required. A copy of the proof of
purchase should be included with the return product. This warranty does not apply to damage that MILWAUKEE determines
to be from repairs made or attempted by anyone other than MILWAUKEE authorized personnel, misuse, alterations, abuse,
normal wear and tear, lack of maintenance, or accidents.
*The warranty period for the LITHIUM-ION battery pack that ships with Test & Measurement Product is two (2) years from
the date of purchase. *Alkaline battery that ships with Test & Measurement Product is separately warranted by the battery
manufacture.
*The warranty period for a NON-CONTACT VOLTAGE DETECTOR – 2201 20 or Plumb 2-Beam Plumb Laser – 2230 20 is
one (1) year from the date of purchase.
Warranty Registration is not necessary to obtain the applicable warranty on MILWAUKEE product. The manufacturing date
of the product will be used to determine the warranty period if no proof of purchase is provided at the time warranty service
is requested.
ACCEPTANCE OF THE EXCLUSIVE REPAIR AND REPLACEMENT REMEDIES DESCRIBED HEREIN IS A CONDITION
OF THE CONTRACT FOR THE PURCHASE OF EVERY MILWAUKEE PRODUCT. IF YOU DO NOT AGREE TO THIS
CONDITION, YOU SHOULD NOT PURCHASE THE PRODUCT. IN NO EVENT SHALL MILWAUKEE BE LIABLE FOR ANY
INCIDENTAL, SPECIAL, CONSEQUENTIAL OR PUNITIVE DAMAGES, OR FOR ANY COSTS, A TTORNEY FEES, EXPENSES,
LOSSES OR DELAYS ALLEGED TO BE AS A CONSEQUENCE OF ANY DAMAGE TO, FAILURE OF, OR DEFECT IN ANY
PRODUCT INCLUDING, BUT NOT LIMITED TO, ANY CLAIMS FOR LOSS OF PROFITS. SOME STATES DO NOT ALLOW
THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES, SO THE ABOVE LIMITATION OR
EXCLUSION MAY NOT APPLY TO YOU. THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES,
WRITTEN OR ORAL. TO THE EXTENT PERMITTED BY LAW, MILWAUKEE DISCLAIMS ANY IMPLIED WARRANTIES,
INCLUDING WITHOUT LIMITA TION ANY IMPLIED W ARRANTY OF MERCHANT ABILITY OR FITNESS FOR A P ARTICULAR
USE OR PURPOSE; TO THE EXTENT SUCH DISCLAIMER IS NOT PERMITTED BY LAW, SUCH IMPLIED WARRANTIES
ARE LIMITED TO THE DURA TION OF THE APPLICABLE EXPRESS WARRANTY AS DESCRIBED ABOVE. SOME ST A TES
DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE LIMITATION MAY NOT
APPLY T O YOU, THIS WARRANTY GIVES YOU SPECIFIC LEGAL RIGHTS, AND YOU MAY ALSO HAVE OTHER RIGHTS
WHICH VARY FROM STATE TO STATE.
This warranty applies to product sold in the U.S.A. and Canada only.
LIMITED WARRANTY - MEXICO, CENTRAL AMERICA AND CARIBBEAN
TECHTRONIC INDUSTRIES' warranty is for 5 year since the original purchase date.
This warranty card covers any defect in material and workmanship on this Power Tool.
To make this warranty valid, present this warranty card, sealed/stamped by the distributor or store where you purchased the product, to the
Authorized Service Center (ASC). Or, if this card has not been sealed/stamped, present the original proof of purchase to the ASC.
Call toll-free 1 800 832 1949 to fi nd the nearest ASC, for service, parts, accessories or components.
Procedure to make this warranty valid
Take the product to the ASC, along with the warranty card sealed/stamped by the distributor or store where you purchased the product, and
there any faulty piece or component will be replaced without cost for you. We will cover all freight costs relative with this warranty process.
Exceptions
This warranty is not valid in the following situations:
a) When the product is used in a different manners from the end-user guide or instruction manual.
b) When the conditions of use are not normal.
c) When the product was modifi ed or repaired by people not authorized by TECHTRONIC INDUSTRIES.
Note: If cord set is damaged, it should be replaced by an Authorized Service Center to avoid electric risks.
Model:
Date of Purchase:
Distributor or Store Stamp:
SERVICE AND ATTENTION CENTER
Rafael Buelna No.1.
Col. Tezozomoc Mexico, Azcapotzalco D.F.
Ph. 01 800 832 1949
IMPORTED AND COMMERCIALIZED BY:
TECHTRONIC INDUSTRIES MEXICO, .S.A. DE C.V.
Av. Santa Fe 481 piso 6, Col. Curz Manca.
CP 05349, Cuajimalpa, D.F.
8
9
Loading...
+ 10 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.