Milwaukee 2236-20, 2237-20 Service Instructions

MILWAUKEE ELECTRIC TOOL CORPORATION – TEST & MEASUREMENT PRODUCT
REPAIR SERVICE INSTRUCTIONS — CALIBRATION 2236-20 Clamp Meter for HVAC/R
2237-20 Clamp Meter
Environmental Condition
Perform all calibration at an ambient temperature of 23°C ± 2°C / 73.4°F ± 3.6°F and relative humidity of 80% - allow the fork meter to sit at this temperature for a minimum of 30 minutes before proceeding.
REPAIR SERVICE
INSTRUCTIONS
Calibration
MILWAUKEE ELECTRIC TOOL CORPORATION
13135 WEST LISBON ROAD BROOKFIELD, WISCONSIN 53005-2550 USA
58-92-2236d2 06/2012
STNETNOC FO ELBAT Page
Introduction 1 Precautions and Safety Information 1 Symbols 1 Safety 2 Specifications 3 General Specifications 3 Voltage Specifications 4 Current Specifications 4 Resistance Specifications 4 Continuity Specifications 4 Capacitance Specifications 4 Temperature Specifications 5 Physical and Environment Characteristics 6 Certification and Compliance 6 Basic Maintenance 6 Opening the Meter Case 6 Replacing the Battery 6 Performance Tests 7 Testing the Display 7 Component Layout and Spare Parts List 8 Disassembly 9 Removal of Battery Cover 9 Removal of Panel 9 Removal of PCB 9 Removal of Jaws 9 Adjustment Procedures 9 Required Equipment and Tools 9 Current Value for Adjustment 10 Check for Low Battery warning 10 DC Voltage Adjustment 10 AC Voltage Adjustment 10 AC Current Adjustment 11 Jaw Balance Adjustment 11 Off-Set adjustment for DC Current (2237-20) 11 Check for Accuracy at 600A Function 11 DC uA Adjustment (2236-20) 11 Electric Capacity Function 11 Temperature Measurement Function (2236-20) 12 Operation Check 12 Resistance Function 12 Continuity Function 12 Frequency Function (2237-20) 12 Current Function 13 DC Current Function (2237-20) 13 DC uA Function (2236-20) 13 Trouble Shooting 13 Trouble Shooting - continued 14
i
Introduction
WARNING
To avoid shock or injury, do not perform the verification tests or calibration procedures
described in the manual unless you are qualified to do so.
The information provided in this document is for the use of qualified personnel only.
CAUTION
The 2236-20 and 2237-20 contain parts that can be damaged by static discharge.
Follow the standard practices for handling static sensitive devices.
Precautions and Safety Information
Use the Meter only as described in the Operator’s Manual. If you do not do so, the protection provided
by the Meter may be impaired. Read the “Safety Information” page before servicing this product.
In this manual, a WARNING identifies conditions and actions that pose hazard (s) to the user; a
CAUTION identifies conditions and actions that may damage the Meter or the test instru
The Symbols
The symbols used on the Meter and in this manual are explained in Table 1.
ments.
Table A. The Symbols
Risk of electric shock
See instruction card
DC measurement
Equipment protected by double or
reinforced insulation
++
Battery
Earth
AC measurement
Conforms to EU directives
Application around and removal from hazardous
live conductors is permitted.
Do not discard this product or throw away.
1
SAFETY
Review the following safety precautions to avoid injury and prevent damage to this product or
products connected to it. To avoid potential hazards, use the product only as specified.
WARNING: These statements identify conditions or practices that could result in personal
injury or loss of life.
CAUTION: These statements identify conditions or practices that could result in damage to
the equipment or other property.
Specific precautions
To reduce the risk of injury, user must read and understand operator’s manual.
Do not operate without covers. To avoid personal injury, do not apply any voltage or current to
the product without covers in place.
Electric overload. Never apply a voltage to a connector on the product that is outside the range
specified for that connector.
Avoid electric shock. To avoid injury o
leads from the meter while they are connected to a voltage source.
Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in
wet or damp conditions.
r loss of life, do not connect or disconnect probes or test
2
SPECIFICATIONS
All specifications are warranted unless noted typical and apply to the 2236-20 and 2237-20.
Stated accuracies are at 23°C±5°C at than 80% relative humidity and without the battery
indicator displayed.
General specifications
noitpircseD scitsiretcarahC
stnuoc 0006 tnuoc yalpsiD
ces / semit 3 etar etadpu ciremuN
citamotuA yalpsid ytiraloP
Overrange display
Low voltage indicator is indicated
Display "OL" when the reading exceeds
range by 10%
++
setunim 02 emit ffo-rewop citamotuA
ezis AA ro 6RL CEI 2 * V5.1 ecruos rewoP
Maximum input voltage
600V CAT lll between V and COM
600V CAT lll between any terminal and earth
Maximum floating voltage
ground
V connector V , V , , ,
C°82> ro C°81< ,C° / )ycaruccA .cepS1.0 tneiciffeoC erutarepmeT
)enilaklA( lacipyt sruoh 30 efiL yrettaB
3