Milwaukee Tool 2217-20 User Manual [en, es, fr]

Cat. No.
No de cat.
2217-20
OPERATOR'S MANUAL MANUEL de L'UTILISATEUR MANUAL del OPERADOR
TRUE RMS MULTIMETER (DMM) MULTIMÈTRE D'EFFICACITÉ VÉRITABLE (DMM) MULTÍMETRO CON RMS VERDADERO (DMM)
TO REDUCE THE RISK OF INJURY, USER MUST READ AND UNDERSTAND OPERATOR'S MANUAL.
AFIN DE RÉDUIRE LE RISQUE DE BLESSURES, L'UTILISATEUR DOIT LIRE ET BIEN COMPRENDRE LE MANUEL DE L'UTILISATEUR.
PARA REDUCIR EL RIESGO DE LESIONES, EL USUARIO DEBE LEER Y ENTENDER EL MANUAL DEL OPERADOR.
IMPORTANT SAFETY INSTRUCTIONS
WARNING READ ALL SAFETY WARNINGS AND INSTRUCTIONS. Failure to
follow the warnings and instructions may result in electric shock, fi re and/or serious injury, as well as instrument damage and/or damage to the equipment being tested. Save these instructions - This operator’s manual contains important safety and operating instructions for the MILWAUKEE True RMS Multimeters (DMM). Before using, read this operator’s manual and all labels on the True RMS Multimeters.
DANGER
Never make measurement on a circuit in which voltage over 600V rms exists. Use only leads rated 600V CATIII or better. Do not apply more than the rated voltage, as marked on the DMM, between terminals or be­tween any terminal and earth ground. Do not attempt a current measurement when the open voltage is above the fuse protection rating. Suspected open circuit voltage can be checked with voltage function. Do not use the DMM to measure voltages in circuits that could be damaged by the DMM’s low input impedance in Lo-Z Function. Impedance is approximately 4K. Do not attempt to make measurement in the presence of fl ammable gases. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion. Never attempt to use the instrument if its surface or your hand is wet. Do not exceed the maximum allowable input of any measuring range. Only test on unenergized circuits unless absolutely necessary. Check tool functionality on a known circuit fi rst. Never assume tool is working. Assume circuits are live until they can be proven de-energized. Never open the Battery cover during a measurement. Do not ground yourself while measuring. Avoid body contact with earthed or grounded sur­faces such as pipes, radiators, ranges and refrigerators. This instrument is to be used only in its intended applications or conditions. Otherwise, the instru­ment's safety functions may not work, resulting in serious personal injury and instrument damage. To reduce the risk of injury from shock and arc blasts, always wear personal protective equipment where live conductors are exposed.
WARNING
Never attempt to make measurement if any abnormal conditions, such as broken case and exposed metal parts are found on the instrument. Comply with local and national safety requirements when working in hazardous locations. Keep fi ngers behind the guards and away from test lead tips during measurements. Do not rotate the Rotary Dial while the test leads are being connected. Never attempt a voltage measurement with the test lead inserted into the A input terminal. Verify proper operation on a known source before use or taking action as a result of the in­dication of the instrument. Do not install substitute parts, replace fuse or make any modifi cation to the instrument. For repair or re-calibration, return the tool to a factory Service/Sales Support Branch or authorized service station. Do not try to replace the batteries if the surface of the instrument is wet. Disconnect all the cords and cables from the object under test and power off the instrument before opening the Battery Cover for Battery replacement. This tool is designed to be powered by 2-AA batteries properly inserted into the MILWAUKEE True RMS Multimeters. Do not attempt to use with any other voltage or power supply. Install battery according to polarity (+ and –) diagrams. Do not leave batteries within the reach of children. Do not mix new and used batteries. Do not mix brands (or types within brands) of batteries. Properly dispose of used batteries. Do not incinerate or dismantle batteries. Under abusive conditions, liquid may be ejected from the battery, avoid contact. If contact accidentally occurs, fl ush with water. If liquid contacts eyes, additionally seek medical help. Liquid ejected from the battery may cause irritation or burns.
2
3
CAUTION
Set the Rotary Dial to an appropriate position before starting measurement. Disconnect test leads from test points before changing Rotary Dial functions. Never connect to a source of voltage with the Dial function in OFF// / / / / . Firmly insert the test leads. Do not expose the instrument to the direct sun, high temperature and humidity or dew fall. Altitude 2000m or less. Appropriate operating temperature is within -10ºC and 50ºC. Keep away from excess dust and water. Be sure to power off the instrument after use. When the instrument will not be in use for a long period, place it in storage after removing the batteries. Use a damp cloth or neutral detergent for cleaning the instrument. Do not use abrasives or solvents.
General Specifi cations
Accuracy is specifi ed for 1 year after calibration, at operating temperatures of 18°C to 28°C (64°F to 82°F), with relative humidity at 45 % to 85 %.
Maximum voltage between any terminal and earth ground
Temperature Operating: -10°C to 50°C (14°F to 122°F)
Temperature Coeffi cient 0.1 x (speci ed accuracy)/°C (<18°C or >28°C) Operating Altitude 2,000 meters IP Rating (International Dust
& Water Protection) Drop Test 1 Meter Battery 2 AA, NEDA 15 A,IEC LR6 Battery Life Approx. 26 Hours all lights on Fuse 600V AC/DC, 10A Fast-acting (Milwaukee p/n 22-89-0305) Safety Compliances EN61010-1, UL 61010-1, UL 61010-031 (Probes), EN61010-031
Certifi cations cULus, CE
600 V Cat III Rated
Current measurements: -10°C to 40°C (14°F to 104°F) Storage: -40°C to 60 °C (-40°F to 140°F)
IP54
(Probes), IEC/EN 61010-1 2nd Edition for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1
FUNCTIONAL DESCRIPTION
1. Display
2. Range button
3. °F/°C button
4. Backlight button
5. V Terminal input
6. COM Terminal input
7. A mA Terminal input
8. Rotary Dial
9. Min/Max button
10. Hold button
11. Accessory bay (on back, not shown)
Read Operator’s Manual
Double insulation
Risk of electric shock
Earth
Danger, Warning, or Caution - Consult the operators manual for additional safety information.
10
9
8
7
Symbology
1
2
3
4
5
6
Battery compartment
European Conformity Mark
Underwriters Laboratories, Inc., United States and Canada
Cat III
4
Classifi cation of transient overvoltages, based on nominal line voltage to earth.
Fuse
Do not dispose of this product as unsorted municipal waste.
5
Lo-Z
Hz
Functions
Dial Position Range Resolution Accuracy
Voltage AC
Voltage DC
Low Input
Lo-Z
Impedance
Hertz
Hz
Resistance
Continuity
Capacitance
Current AC
Current DC
Temperature
† Temperature Range, Resolution and Accuracy are for the DMM. Temperature Probe may have different
specifi cations. * These instruments are True-RMS sensing. All voltage and current readings are True-RMS values. * Input impedance: Voltage DC: 10M; Voltage AC: 10M // less than 100pF; Lo-z: 4k approximately * Overload protection: Voltage DC, Voltage AC, Lo-Z Voltage and Hertz: AC/DC 720V for 10 second Current DC & Current AC: AC/DC 20A for 10 sec. Resistance, Continuity, Capacitance and Temperature: AC/DC 600V for 10 second * Maximum measurement time: 1 minute at 10A, rest time 20 minutes minimum * Minimum frequency measurement is 2Hz * Signal voltage in frequency mode: 8V-600V(RMS) * Minimum AC Current measurement is 0.40mA * Minimum AC Voltage measurement is 4.0mV * For AC Voltage, AC Current and Lo-z: Additional Accuracy by Crest Factor (C.F.): Add 6.0% for C.F. 1.0 ~ 2.0. Add 7.0% for C.F. 2.0 ~ 2.5. Add 9.0% for C.F. 2.5 ~ 3.0 Max. Crest Factor: 1.6 for 6600 ~ 5000 digits
2.0 for 5000 ~ 3000 digits
3.0 for 3000 ~ 0 digits * Measurement accuracy of square wave and truncated waveforms at 1kHz is unspecifi ed.
600.0mV/6.000V
60.00V/600.0V
600.0mV/6.000V
60.00V/600.0V
600.0V 0.1V ±(2.0% + 3 dgt) DC,AC : 45~500Hz
99.99/999.9Hz
9.999/50.00kHz
600.0
6.000/60.00/600.0k
6.000/40.00M
Cont Buzzer
0-600.0
100.0F
0.40mA-60.00mA/600.0mA
0.40mA-60.00mA/600.0mA
1000F
6.000A/10.00A
6.000A/10.00A
- 40.0ºC ~ 400.0ºC
-40.0ºF ~ 752.0ºF
0.1mV/0.001V/
0.01V/0.1V
0.1mV/0.001V/
0.01V/0.1V
0.01Hz/0.1Hz/
0.001kHz/0.01kHz
0.1
0.001/0.01/0.1k .001/.01M ±(1.9% + 5 dgt)
0.1F 1F
0.01mA/0.1mA
0.001A/0.01A
0.01mA/0.1mA
0.001A/0.01A
0.1°C
0.2°F
±(1.0% + 3 dgt) (45~500Hz)
±(2.0% + 3 dgt) (500~1000Hz)
±(0.5% + 2 dgt)
±(0.1% + 2 dgt)
±(1.0% + 5 dgt)
Buzzer sounds at 30 or less
±(1.9% + 2 dgt)
±(1.5% + 3 dgt) (45~500Hz)
±(1.0% + 3 dgt)
±(1.0% + 10 dgt) ±(1.0% + 18 dgt)
ASSEMBLY
WARNING To avoid an electrical hazard,
turn the Rotary Dial to OFF and disconnect the test leads before replacing batteries.
Loading/Changing the Batteries
Replace batteries when the Low Battery indicator
is displayed.
1. Turn Rotary Dial to OFF and disconnect the test leads.
2. Unscrew and remove battery door.
3. Insert two (2) AA batteries, according to the polarity marked in the battery compartment
4. Close the battery door and tighten four (4) screws securely.
OPERATION
WARNING Only use Milwaukee test
leads with the MILWAUKEE DMM. Inspect test leads for continuity before each use. Do not use if the readings are high or noisy.
Before Use
Confi rm the Rotary Dial is set to the correct position, the instrument is set to the correct measurement mode, and the Data hold function is disabled. Oth­erwise, desired measurement cannot be made.
LCD Backlight
The LCD backlight will turn off after about 10 min­utes of inactivity. Press the backlight button to turn the backlight on again.
Making a Measurement
AC Voltage
DANGER To avoid electrical shock:
Never make measurement on a circuit in which voltage over AC600V exists. Do not use with the Battery Cover removed. Keep fi ngers behind the guards and away from test lead tips during measurements.
1. Set the Dial to position.
2. Connect the red test lead to the V terminal and the black test lead to the COM terminal.
3. Connect the test leads to the circuit under test. The read­ing is displayed.
CAUTION Readings may fl uctuate or
be infl uenced in noisy environment.
6
7
DC Voltage
DANGER To avoid electrical shock:
Never make measurement on a circuit in which voltage over DC600V exists. Do not use with the Battery Cover removed. Keep fi ngers behind the guards and away from test lead tips during measurements.
1. Set the Dial to position.
2. Connect the red test lead to the V terminal and the black test lead to the COM terminal.
3. Connect the red test lead to the positive (+) side and black test leads to the negative (-) side of the cir­cuit under test. The reading is displayed. A reversed connection is indicated as a negative value.
Lo-Z Low Input Impedance
Automatic voltage detection (AC or DC).
1. Set the Dial to Lo-Z position.
2. Connect the red test lead to the V terminal and the black test lead to the COM terminal.
3. AC: Connect the test leads to the circuit under test. The reading is displayed.
DC: Connect the red test lead
to the positive (+) side and black test leads to the negative (-) side of the circuit under test. The reading is displayed. A reversed connection is indicated as a negative value.
CAUTION Do not use the DMM to measure
voltages in circuits that could be damaged by the DMM’s low input impedance less than approximately 4k.
Hz Frequency
1. Set the Dial to Hz position.
2. Connect the red test lead to the V terminal and the black test lead to the COM terminal.
3. Connect the test leads to the circuit under test. The reading is displayed.
Resistance/Continuity/Capacitance Measurements
DANGER To reduce the risk of
electric shock for Resistance, Continuity, and Capacitance measurements, never use the DMM on an energized circuit. Make sure a capacitor is fully discharged before touching or attempting to make a measurement.
Do not use with the Battery Cover removed.
Resistance
1. Set the Dial to position.
2. Connect the red test lead to the V terminal and the black test lead to the COM terminal.
Confi rm “OL” is indicated on
the display, and then short­circuit the tips of test leads to make the indication zero.
3. Connect the test leads to the both ends of the resistor under test.
4. The reading is displayed.
CAUTION After shorting the test leads,
the displayed value may not be zero due to the resistance of test leads themselves.
Continuity
1. Set the Dial to position.
2. Connect the red test lead to the V terminal and the black test lead to the COM terminal.
Confi rm “OL” is indicated on
the display, and then short­circuit the tips of test leads to make the indication zero. A buzzer will sound.
3. Connect the test leads to the both ends of the conductor under test. If the resistance under test is 30 or less, the buzzer will sound.
Capacitance
1. Set the Dial to position.
2. Use the Range button to se­lect either 100F or 1000F.
3. Connect the red test lead to the V terminal and the black test lead to the COM terminal.
4. Discharge capacitor.
5. Connect the test leads to the both ends of the capacitor under test.
6. The reading is displayed.
AC Current
DANGER To avoid electrical shock:
Never make measurement on a circuit in which voltage over AC600V exists.
Do not use with the Battery Cover removed. Keep fi ngers behind the guards and away
from test lead tips during measurements.
1. Set the Dial to position.
2. Connect the red test lead to the A terminal and the black test lead to the COM terminal.
3. Turn circuit power off, dis­connect the circuit, connect the test leads in series with the circuit under test, and then turn circuit power on. The reading is displayed.
DC Current
DANGER To avoid electrical shock:
Never make measurement on a circuit in which voltage over AC600V exists. Do not use with the Battery Cover removed. Keep fi ngers behind the guards and away from test lead tips during measurements.
1. Set the Dial to
2. Connect the red test lead to the A terminal and the black test lead to the COM terminal.
3. Turn circuit power off, dis­connect the circuit, connect the test leads in series with the circuit under test, and then turn circuit power on. The reading is displayed.
position.
Temperature
1. Set the Dial to position.
2. Connect the K-type Tempera­ture Probe to the input termi­nal. The positive (+) side of Probe should be connected to V.
3. Contact the probe sensor to the object under test.
4. The reading is displayed.
WARNING
Never connect the Temperature Probe to an energized circuit.
CAUTION
When the dial is set to , OL should be dis­played. If anything else is displayed, some­thing may be wrong with the DMM. Stop using the DMM immediately.
Using Bar Graph Display
The bar graph is like the needle on an analog me­ter, it updates much faster than the digital display. The number of segments indicates the measured value and is relative to the full-scale value of the selected range.
HOLD Button
Data Hold Function - Freezes the value on the dis­play. Press the HOLD button to freeze the reading. The reading will be held regardless of subsequent variation in input. HOLD is displayed with the read­ing. To exit Data Hold mode, press the HOLD button again or change the dial.
SMART HOLD: The meter will beep continuously and the display will fl ash if the measured signal is 50 counts larger than the display reading. (However, it can not detect across the AC and DC Voltage/ Current).
CAUTION
The Data Hold readings of maximum / mini­mum are released when the DMM enters Sleep Mode.
MIN/MAX Button
The MIN MAX recording mode captures the mini­mum and maximum input values. When a new high or low is detected, the DMM beeps. Put the DMM in the desired measurement function and range, then press MIN/MAX button to enter MIN MAX mode, and present readings and MAX MIN are displayed.
8
9
Loading...
+ 10 hidden pages