Miele E 505 Operating instructions

0 (0)
Module E 505 for intubation equipment
Please read the Warning and Safety instructions in the Operating
,
Instructions for your washer-disinfector carefully. You should only process instruments/medical products in this module if they have been declared as suitable for machine processing by the manufacturer. Please also observe the manufacturer's care instructions. The programme selected must be suitable for the materials from which the items to be processed are made. Where possible the programme should include thermal disinfection.
Areas of application:
This module is designed for use in mobile unit E 501 (which is supplied with separate operating instructions). It is designed for cleaning, disinfecting and drying hollow anaesthetic instruments and equipment. Special holders and injector jets are used for securing breathing tubes, breathing bags, Guedel tubes and endotracheal tubes, laryngeal masks and bellows.
To ensure a suitable standardised spray pressure for all connections, all
,
screw connectors must be fitted with jets, adapters, sleeves or blanking screws. Damaged jets, adapters or sleeves etc. must not be used.
Empty jets, etc. do not have to be replaced with blanking screws.
To ensure that hollow instruments are thoroughly cleaned and disinfected
,
internally, it is essential to connect them to the correct holders/jets in the mobile injector unit. They must not detach themselves from their holders during the programme. If, however, a hollow instrument does become dislodged from its holder during a programme, it must be processed again.
To avoid damage to certain items, observe the maximum permitted drying
,
temperature. The maximum temperature permitted for the elastomers in some breathing bags and masks is below 85°C. Drying results must be inspected after the cleaning programme has finished. If necessary, adjust the drying time at the end of the cleaning programme.
The illustration above shows module E 505 in mobile unit E 501.
Fit module E 505 into mobile unit
^
E 501.
To do this:
Place the two positioning guides over
^
the middle water inlet.
The inlet pipe needs to be positioned in front of the cover in the water inlet.
^ Press the bracket on the left hand
side down until the handle engages behind the rail on the mobile unit.
^
Attach Guedel tubes to injector jets with spring clips.
^
Expel air from the balloon lead.
^
Endotracheal tubes are fitted onto the injector jets so that spring clips secure them to the jets.
^
Place the balloon lead with the valve behind the spring clip.
^
Fit the breathing bags onto the injector jets.
^
Make sure the opening of the breathing bag fits securely over the plastic cone. This will prevent the breathing bag from sticking together during the cleaning and drying process and ensure it is thoroughly cleaned internally.
The breathing bag is attached correctly when it cannot dislodge itself from the plastic cone.
^
Remove the stopper from breathing masks which have a closable air cushion. Expel as much air as possible and then replace the stopper.
^
Several breathing masks can be placed onto one breathing bag injector jet.
Miele Company Ltd., Fairacres, Marcham Road, Abingdon, Oxon, OX14 1TW, Tel 0330 160 6600, Internet: http://www.miele.co.uk / ! 0297 Alteration rights reserved / Publication date 2015-11-01
M.-Nr. 10 338 880 / 00 en
Loading...
+ 1 hidden pages