Micro Motion Application Manual: 5700 Micro Motion Advanced Phase Measurement Manuals & Guides

Micro Motion™ Advanced Phase Measurement
Application Manual
Application Manual
MMI-20030076, Rev AD
September 2020
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully before proceeding to the next step.
Safety and approval information
This Micro Motion product complies with all applicable European directives when properly installed in accordance with the instructions in this manual. Refer to the EU declaration of conformity for directives that apply to this product. The following are available: the EU declaration of conformity, with all applicable European directives, and the complete ATEX Installation Drawings and Instructions. In addition the IECEx installation instructions for installations outside of the European Union and the CSA installation instructions for installations in North America are available on the internet at www.emerson.com or through your local Micro Motion support center.
Information affixed to equipment that complies with the Pressure Equipment Directive, can be found on the internet at
www.emerson.com
For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.
Return policy
Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept
your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion Customer Service department.
or through your local Micro Motion support center.
Emerson Flow customer service
Email:
Worldwide: flow.support@emerson.com
Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South America Europe and Middle East Asia Pacific
United States 800-522-6277 U.K. and Ireland 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 70 413
6666
Mexico +52 55 5809 5300 France +33 (0) 800 917
901
Argentina +54 11 4809 2700 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy +39 8008 77334 China +86 21 2892 9000
Chile +56 2 2928 4800 Central & Eastern +41 (0) 41 7686
111
Peru +51 15190130 Russia/CIS +7 495 995 9559 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
UAE 800 0444 0684
New Zealand 099 128 804
India 800 440 1468
Japan +81 3 5769 6803
2
Application Manual Contents
MMI-20030076 September 2020

Contents

Chapter 1 Before you begin........................................................................................................5
1.1 About this application manual..................................................................................................... 5
1.2 Related documentation............................................................................................................... 5
1.3 About the software...................................................................................................................... 6
1.4 APM software requirements.........................................................................................................7
1.5 Terms and definitions.................................................................................................................. 9
Chapter 2 Measurement options and configuration................................................................. 11
2.1 Liquid with gas...........................................................................................................................11
2.2 Net oil........................................................................................................................................17
2.3 Gas with liquid........................................................................................................................... 21
Chapter 3 Additional configuration.......................................................................................... 27
3.1 Configure viewing and reporting for process variables............................................................... 27
3.2 Configure APM contract totals into the totalizer history log....................................................... 30
3.3 Configure events....................................................................................................................... 31
Chapter 4 Transmitter operation..............................................................................................33
4.1 Read process variables...............................................................................................................33
4.2 Read contract totals...................................................................................................................33
Chapter 5 APM alerts............................................................................................................... 35
Appendix A Application parameters and data............................................................................. 37
A.1 Advanced Phase Measurement Modbus configuration parameters............................................37
A.2 Advanced Phase Measurement default totalizer and inventory values....................................... 38
A.3 Advanced Phase Measurement Modbus process variables......................................................... 39
A.4 Period Averaged Options current period data............................................................................40
Appendix B Typical oil and gas applications and other information............................................. 43
B.1 Advanced Phase Measurement with a two-phase separator.......................................................43
B.2 Advanced Phase Measurement with a three-phase separator.................................................... 44
B.3 Advanced Phase Measurement at the wellhead......................................................................... 45
B.4 Automatic Drive Gain Threshold determination.........................................................................46
B.5 Manual Drive Gain Threshold..................................................................................................... 47
Appendix C Best practices for two-phase measurement performance......................................... 49
C.1 Entrained gas performance........................................................................................................49
C.2 Entrained liquid (mist) performance.......................................................................................... 50
C.3 Density determination...............................................................................................................52
Application Manual 3
Contents Application Manual
September 2020 MMI-20030076
4 Micro Motion Advanced Phase Measurement
Application Manual Before you begin
MMI-20030076 September 2020

1 Before you begin

1.1 About this application manual

This application manual explains how to configure and use the Advanced Phase Measurement licensed software option on select Model 5700 transmitters. It also contains limited, supplemental installation and configuration information specifically related to the Advanced Phase Measurement software. Refer to the sensor and transmitter installation manuals and the transmitter configuration and use manual for complete information.
Important
This manual assumes that:
The transmitter has been installed correctly and completely according to the
instructions in the transmitter installation manual
Users understand basic transmitter and sensor installation, configuration, and
maintenance concepts and procedures

1.2 Related documentation

You can find all product documentation on the product documentation DVD shipped with the product or at www.emerson.com.
See any of the following documents for more information:
Model 5700 installation manualMicro Motion 5700 Transmitters with Configurable Inputs and Outputs: Installation
Manual
Micro Motion 5700 with Ethernet Transmitters: Installation Manual
Micro Motion 5700 Transmitters for FOUNDATION™ Fieldbus: Installation Manual
Micro Motion 5700 Transmitters with Intrinsically Safe Outputs: Installation Manual
Micro Motion 5700 Transmitters for PROFIBUS-PA: Installation Manual
Model 5700 configuration and use manualMicro Motion 5700 Transmitters with Configurable Inputs and Outputs: Configuration
and Use Manual
Micro Motion 5700 with Ethernet Transmitters: Configuration and Use Manual
Micro Motion 5700 Transmitters for FOUNDATION™ Fieldbus: Configuration and Use
Manual
Micro Motion 5700 Transmitters with Intrinsically Safe Outputs: Configuration and Use
Manual
Micro Motion 5700 Transmitters for PROFIBUS-PA: Configuration and Use Manual
Micro Motion Enhanced Density Application Manual
Micro Motion Modbus Map
Application Manual 5
Before you begin Application Manual
September 2020 MMI-20030076
Micro Motion ProLink III with ProcessViz Software User Manual
Micro Motion Ethernet PROFINET Siemens Integration Guide
Sensor installation manual

1.3 About the software

Micro Motion Advanced Phase Measurement software improves long-term flow reporting and measurement performance in processes with intermittent periods of two-phase flow, including liquids with entrained gas or gas with entrained liquid. And if it is combined with the Net Oil or Concentration Measurement software options, the software can also report liquid concentration, Net Oil, and/or Gas Void Fraction (GVF) during the same two-phase conditions.
There are three measurement options for the Advanced Phase Measurement software: Net Oil, Liquid with Gas, and Gas with Liquid.
Note
Each option is licensed separately in the transmitter. Field upgrades are permitted.
Table 1-1: Net oil options (choose one)
License option (ordering code)
MA - Manual Advanced Phase Measurement configuration
MW - Net Oil Computer (NOC) software - multiple wells
Description Availability
Suitable for a mixture of oil and water under predictable flow conditions. Includes manual drive gain threshold only. See Manual
Drive Gain Threshold.
This option is the Model 5700 upgrade for Production Volume Reconciliation (PVR).
Ideal for test separators shared by multiple wells.
The transmitter stores a total of three well tests in memory. The transmitter can be set up for testing up to 48 independently­configured wells; however, only the most recent three tests that have been completed remain in storage at a time.
Suitable for mixtures of oil and water, with remediation for gas.
Not available on a
transmitter with intrinsically safe or non­intrinsically safe FOUNDATION™ Fieldbus H1 outputs
Not available on a
transmitter with intrinsically safe outputs
Do not combine with APM
liquid with gas (option PL) ­basic remediation for gas is included.
Do not combine with APM
liquid with gas (option PL) ­already included.
Only available on a
transmitter with configurable outputs
6 Micro Motion Advanced Phase Measurement
Application Manual Before you begin
MMI-20030076 September 2020
Table 1-1: Net oil options (choose one) (continued)
License option (ordering code)
PO - Net Oil Suitable for mixtures of oil and
Description Availability
water. Add PL option to remediate for gas.
Table 1-2: Liquid with gas
License option (ordering code)
PL - Advanced Phase Measurement Liquid with Gas
Description Availability
Suitable for any liquid with entrained gas.
Table 1-3: Gas with liquid
License option (ordering code)
PG - Advanced Phase Measurement Gas with Liquid
Description Availability
Suitable for any gas that may contain entrained liquids (mist).
Can be combined with APM license code PL. PL is recommended since most net oil applications contain gas.
Can be combined with APM
license code PO.
Can be combined with
license code concentration measurement (CM).
Cannot be activated with any other license code.

1.4 APM software requirements

Ensure that your installation meets the requirements in this section.
Transmitter
Advanced Phase Measurement software is available only on the Model 5700 transmitter, either integrally mounted, in a 9-wire remote configuration, or in combination with any 800 Enhanced Core Processor.
Not available with:
Weights & Measures -NTEP (option NT)
Safety certification of 4-20 mA outputs per IEC 61508 (option SI)
Sensor
Advanced Phase Measurement software is compatible with any sensor that is supported by the Model 5700 transmitter except for the T-Series and R-Series sensors due to limited performance with two-phase fluids. The software is not compatible with the 700 Standard Core processor.
Installation and wiring
Follow the installation and wiring instructions in the sensor and transmitter installation
manuals.
Application Manual 7
Before you begin Application Manual
September 2020 MMI-20030076
Identify and follow any application-specific installation needs as identified in this
manual for your application type.
For the entrained gas/mist/empty-full-empty best installation practices, refer to Best
practices for two-phase measurement performance.
If you plan to use the net oil measurement option with an external water cut monitor:For the 5700 configurable transmitter, Channel D on the transmitter must be
enabled, and must be configured as a mA Input, wired to the water cut monitor, and configured appropriately. HART integration is supported on Channel A by polling.
For the 5700 Ethernet transmitter, an external host system is required to accept the
water cut monitor input and feed it to the transmitter.
For the 5700 intrinsically safe transmitter, HART integration is supported on
Channel A by polling.
Note
Water cut cannot be input into the 5700 transmitter with FOUNDATION fieldbus output if the software revision of the transmitter is 1.x.
Configuration and operation
The Advanced Phase Measurement software can be configured using the interface option already being used to configure the transmitter:
Transmitter display
ProLink III v4.2 or later
Field communicator (Model 5700 configurable and Model 5700 fieldbus transmitters
only).
A fieldbus host (Model 5700 fieldbus transmitter only)
A web browser (Model 5700 Ethernet transmitter only)
Because much of the Advanced Phase Measurement software process data is not assignable to an output (for example, contract totals), use one of the following external host systems, depending upon the transmitter output type:
For the Model 5700 configurable transmitter, a Modbus/RS-485 connection to
Channel E, that must be enabled and wired to a Modbus host to collect and process data.
For the Model 5700 Ethernet transmitter, an Ethernet/IP connection to the transmitter,
and an appropriate host program to collect the Advanced Phase Measurement software process data.
For the Model 5700 FOUNDATION fieldbus transmitter, a fieldbus H1 connection to the
transmitter and a fieldbus host that reads Advanced Phase Measurement software process data.
For the Model 5700 intrinsically safe transmitter, HART integration by polling.
8 Micro Motion Advanced Phase Measurement
Application Manual Before you begin
MMI-20030076 September 2020

1.5 Terms and definitions

The Advanced Phase Measurement software application and this manual use the terms defined here.
Contract hour
Contract period
Contract total
Corrected density
Corrected volume
Correction
Density-based water cut
Density determination
Drive gain threshold
Contract totals are recorded and then reset at the beginning of the contract hour.
The 24-hour period that monitors how much fluid the well has delivered. Contract totals are reset at the beginning of each contract period.
Up to four user-specified totals that are used to measure production.
The density of the process fluid at reference temperature.
The volume of the process fluid at reference temperature.
The process of calculating the value of a process variable at reference temperature, starting from the value of the process variable at line temperature (the measured value or observed value).
The water cut value calculated by Advanced Phase Measurement software using the measured line density referenced to a density/ water cut curve based on user-entered dry oil and water densities at reference temperature.
The procedure for obtaining the density of dry oil and water at reference temperature—both are required for use with the Net Oil software option.
Maximum drive gain value expected for single-phase fluid under typical operating conditions. Above the threshold value, the measurement will be remediated in accordance with the configured Advanced Phase Measurement software settings. See
Automatic Drive Gain Threshold determination or Manual Drive Gain Threshold for more information.
Entrained, entrainment
External water cut
Gas void fraction
Meter factor for shrinkage
Mixture
Application Manual 9
The presence of small amounts of gas in a liquid stream, or liquid in a gas stream.
A water cut value measured by an external device and supplied to the Advanced Phase Measurement software via the mA Input or any compatible digital protocol; such as HART, EtherNet/IP, etc.
The ratio of gas volume to total mixture volume at line conditions. Also called Gas Volume Fraction.
A meter factor, established by proving, that acts as a multiplier to shrinkage-factored (SF) variables. Not commonly used.
The process fluid before separation - for example, a combination of a two-phase system; two liquids (oil and water), or a three­phase system (gas, oil, and water).
Before you begin Application Manual
September 2020 MMI-20030076
Multiwell
Net
Period averaged output (PAO)
Post-mist adjustment delay
Pre-mist averaging period
Remediated
An installation where well tests can be performed on up to 48 wells. A manifold system is used to ensure that the output from a single well is routed through the test separator and the NOC system.
A measurement of a single component of the process fluid - for example, oil only, water only.
Rolling averages, with adjustable averaging period, that help identify trends in noisy data. Available PAOs are: mass flow rate, density, volume flow rate, net oil flow rate and water cut at line and reference conditions, gas void fraction, and temperature. The Average Reporting Interval is used to adjust the averaging period.
Period of stable gas flow conditions after a liquid entrainment event. After this delay, the average flow rate during Post-Mist Adjustment Delay is averaged with the Pre-Mist Averaging Period and the flow rate is adjusted a maximum of +/-10% of reading until any difference has been appropriately corrected in the totalizers. Default = 15 seconds.
Period of stable gas flow conditions prior to a liquid entrainment event. Averaging period is user configurable with units = n seconds. The average flow rate during this period will be reported until the entrained liquid event has passed. Default = 15 seconds.
An adjustment applied to a measured process variable by the Advanced Phase Measurement algorithm to correct for errors associated with two-phase fluid conditions.
Shrinkage
Shrinkage factor
Uncorrected density
Uncorrected volume
Unremediated
Water cut
The change in liquid volume between the measurement point and a stock tank due to lighter hydrocarbons.
User input multiplier used to account for shrinkage between measurement point and stock tank. Only affects shrinkage­factored (SF) variables. You can estimate your shrinkage factor by dividing the temperature-corrected metered density by the temperature-corrected oil sales density.
The density of the process fluid at line temperature and pressure.
The volume of the process fluid at line temperature and pressure.
Measured variables that are not adjusted by the Advanced Phase Measurement algorithm. In two-phase conditions, these process variables represent the whole mixture, or bulk fluid (e.g. water, oil, and gas).
The volume fraction of water in the liquid mixture, in %.
10 Micro Motion Advanced Phase Measurement

Application Manual Measurement options and configuration

MMI-20030076 September 2020
2 Measurement options and
configuration

2.1 Liquid with gas

This measurement option improves flow measurement in liquid processes with intermittent entrained gas, or with known-density liquids under continuous entrained gas conditions.
Note
The liquid with gas measurement option can also be combined with the net oil measurement option or concentration measurement. See the Micro Motion Enhanced Density Application Manual to configure concentration measurement.
Liquid with gas measurement process
The presence of entrained gas (or bubbles) can cause significant errors when measuring the volume flow of liquid through a Coriolis meter. Because bubbles displace some of the liquid in a flow stream, the measured volume of the mixture may differ from the actual amount of liquid that emerges from the pipe downstream.
So how can you tell when a liquid contains gas? When bubbles are present in a liquid stream, Coriolis meters will report an increase in drive gain coinciding with a decrease in both fluid density the liquid-gas mixture. Therefore, in order to measure only the liquid portion of the stream, the volume of the bubbles must be ignored or subtracted from the mixture total. The APM software performs exactly this function, using drive gain as the diagnostic indication that bubbles or entrained gas is present in the liquid flow stream, and then substituting a liquid-only density in place of the live measurement until the gaseous event has subsided. When the gassy portion has passed, indicated by an associated drop in drive gain, the software returns to reporting the live measured volume flow rate.
(1) High frequency sensors may erroneously report a higher fluid density when entrained gas is present, and therefore are not
recommended for use on liquids with entrained gas. High frequency sensors include the F300/H300 compact, and all T­Series sensors.
(2) The accuracy and repeatability of the mass flow and density measurements for liquids with entrained gas is dependent on
the sensor-fluid decoupling ratio, which is a complex function of fluid velocity, fluid viscosity, fluid density, the difference between the liquid and gas densities, the operating sensor frequency, and the Gas Volume Fraction (GVF) of gas. For best measurement performance, GVF should be kept below 15%.
(1)
and mass flow rate
(2)
due to the lower amount of mass contained in
Application Manual 11
Measurement options and configuration Application Manual
September 2020 MMI-20030076
Figure 2-1: How entrained gas affects drive gain and density measurement
C
B
A
ρ
D
DG%
t
DG% = drive gain percentage
t = Time
ρ = Density
A. Drive gain indication
B. Measured density C. Entrained gas ocurring during these intervals D. Drive gain threshold
Entrained gas in liquids affects drive gain and density measurement. The green line (A) shows the drive gain indication which is stable under most single-phase conditions. However, if gas is entrained in the liquid, the density reading (B) will drop and the drive gain reading will increase. When the entrained gas bubbles go away, the drive gain will return to its normal indication and the density measurement will reflect the density of the liquid.
The APM software identifies entrained gas in liquid flow by detecting the sharp increases in drive gain and corresponding decreases in density measurement. The software continuously monitors the most recent drive gain data (up to 60 minutes) to determine drive gain threshold. If the measurement exceeds the drive gain threshold, the fluid is deemed to contain entrained gas, and remediation occurs.
2.1.1
12 Micro Motion Advanced Phase Measurement

Production type options

Continuous Flow
Select this option only when flow rates are expected to be stable under normal operating conditions. The APM software assumes the liquid properties and flow velocity through the pipe is constant, and hence is able to remediate the mass flow values in addition to density and volume values.
Variable Flow (default)
Select this option when flow rates are not stable, for applications such as batching, dump valve control, beam pumps, production separators, or other variable processes.
Application Manual Measurement options and configuration
MMI-20030076 September 2020
When variable flow is selected, APM will remediate only density and volume variables.

2.1.2 Drive gain threshold mode

There are two ways to determine when remediation occurs, either Automatic or Manual.
Automatic
The APM software will determine when flow conditions are stable or when multiphase flow is present in order to perform remediation. This mode is not available with a manual Advanced Phase Measurement configuration (option MA). See Manual Drive Gain
Threshold.
Manual
At drive gain values above the user-entered threshold, the transmitter will perform the configured remediation option. As soon as the drive gain drops below the manual drive gain threshold value, the transmitter returns to reporting unremediated values. See
Manual Drive Gain Threshold.
2.1.3

Remediation options (APM Action)

If the drive gain threshold is exceeded, you must select one of the following methods to handle the volume calculation for the period of high drive gain.
Hold Last Value (default)
APM will use a held density value from an earlier point in the process to report density, calculate volume, and calculate any other density-influenced variables during remediation. If this option is chosen, the density from the point just before the entrained gas event is held constant throughout the event.
Application Manual 13
Measurement options and configuration Application Manual
September 2020 MMI-20030076
Figure 2-2: Hold Last Value in operation
C
B
A
ρ
D
DG%
t
DG% = drive gain percentage
t = Time
ρ = Density
A. Drive gain
B. Measured density C. Held density value during entrained gas intervals D. Drive gain threshold
This figure shows how the Hold Last Value feature works in APM. The green line (A) shows the drive gain value and the blue line (B) shows the density reading. If the liquid gets entrained with gas bubbles, the drive gain increases above the drive gain threshold (D). Then the software determines a density value from recent process data that does not have a high drive gain. It then substutes that value for the measured density until the drive gain goes back below the threshold value (D). This substituted density is also referred to as the remediated density.
Density Hold Override
APM will use a user-input density value from an earlier point in the process to report density, calculate volume, and calculate any other density-influenced variables during remediation. This value should reflect the density of the liquid at line conditions.
2.1.4
14 Micro Motion Advanced Phase Measurement

Configure liquid with gas measurement

Configuring liquid with gas measurement is one part of the following required configuration for APM functionality.
Application Manual Measurement options and configuration
MMI-20030076 September 2020
Start
End
Configure units
See configuration
and use manual
Configure totalizers
and inventories
See configuration
and use manual
Configure liquid
with gas
(this procedure)
Configure channels
and outputs
See configuration
and use manual
Procedure
1. Set the mass flow cutoff and the volume flow cutoff to a non-zero value. This
ensures that totalizing stops when flow is stopped. In most installations, the default value is satisfactory.
Option Description
Display Menu Configuration Process Measurement Flow
Variables Mass Flow Settings
ProLink III Menu Configuration Process Measurement Flow
Web browser Configuration Process Measurement Flow Variables
2. Navigate to one of the following paths based on the tool you are using.
Option
Description
Display MenuConfigurationProcess Measurement
Adv Phase Measurement
ProLink III Device Tools Configuration Process
Measurement Advanced Phase Measurement
Field Communicator or enhanced FF host
Basic FF host (fieldbus
Configure Manual Setup Advanced Phase Measurement
The APM transducer block
transmitters only)
Web browser (Ethernet transmitters only)
Configuration Process Measurement Advanced Phase Measurement
3. Depending on your configuration tool, configure liquid with gas measurement:
From the display, navigate to Single Liquid and Save.
From ProLink III, set Fluid Type to Liquid with Gas.
4. Verify or make changes to the following settings:
Application Manual 15
Measurement options and configuration Application Manual
September 2020 MMI-20030076
Depending on which tool you are using, the parameter name or order may be different.
Parameter Description
Production Type Choose one of the following options:
Continuous flow if the flow in your system is generally
constant. Continuous flow assumes a constant flow rate that enables mass flow to also be remediated by subtracting GVF.
Variable flow if the flow in your system frequently starts
and stops or fluctuates between high and low flow rates.
See Production type options.
Density Hold Override Density will report this value when in remediation. This
value should reflect the density of the liquid at line conditions.
Drive Gain Threshold
From the display, enable or disable the automatic
threshold. Disabling automatic threshold defaults to manual. You must configure a manual threshold value.
From ProLink III, select automatic or manual drive gain
threshold.
See Drive gain threshold mode.
User Input DG Threshold For use with manual threshold only, this is the drive gain
value above which APM will remediate.
Gas Density @ Line The density of entrained gas. Default = 0.
Use Default (zero) if the line pressure is less than 250 psi
(17.24 bar) absolute.
If the typical line pressure is greater than 250 psi
(17.24 bar) absolute, set this parameter to the density of the entrained gas at typical line conditions. This setting will affect the GVF output (and mass flow remediation if
Process Type is set to Continuous).
Average Reporting Interval
If using the display, navigate left to select.
Transmitter Date
Not available in the display.
Transmitter Time
The time period, in minutes, over which process variables will be averaged. The averages are available for retrieval by the host system.
Displays the currently-set date.
From the display, choose Menu Configuration
Time/Date/Tag.
From ProLink III, choose Device Tools Configuration
Process Measurement Advanced Phase Measurement in the Transmitter Date field.
From the display, set the time zone.
From ProLink III, set the time zone, time zone offset,
date, and time.
16 Micro Motion Advanced Phase Measurement
Application Manual Measurement options and configuration
MMI-20030076 September 2020

2.2 Net oil

The Net Oil algorithm calculates the water fraction of the liquid stream so that net oil and net water can be determined. Net Oil is the volume of oil, corrected to a reference temperature and/or pressure, that is contained within the gross volume of produced fluid.
Note
The PO measurement option should also be combined with the standalone liquid with gas remediation function that gives it the added capability to mediate against intermittent entrained gas. The MA and MW options are not compatible with PL.
This algorithm requires the following data:
Flow rate and temperature, that are measured by the meter.
Density of both dry oil and water from this well at reference conditions. These are
determined by the operator and entered during configuration. See Density
determination for more information about density determination.
Current water cut is either:Measured by a water cut monitor and supplied to the Advanced Phase
Measurement software via the mA Input or host system
(3)
2.2.1
Calculated by the Advanced Phase Measurement software from current density
data via the density-based net oil calculation (polled via HART). If density-based water cut is chosen, the software uses the following equation to calculate the water cut.
Equation 2-1: Calculation of density-based water cut
ρ
–ρ
mix
WaterCut=
ρ
ρ
ρ
= Density of the oil and water mixture as measured by the sensor
mix
= Density of produced oil calculated from user-supplied value
oil
= Density of produced water calculated from user-supplied value
water
ρ
water
–ρ
oil
oil

Configure net oil measurement

Configuring net oil measurement is one part of the following required configuration for APM functionality.
(3) Not available for the Model 5700 fieldbus version 1.x.
Application Manual 17
Measurement options and configuration Application Manual
September 2020 MMI-20030076
Start
End
Configure units
See configuration
and use manual
Configure totalizers
and inventories
See configuration
and use manual
Configure net oil
measurement
(this procedure)
Configure channels
and outputs
See configuration
and use manual
Prerequisites
If you plan to use a water cut monitor:
For the 5700 configurable transmitter, Channel D on the transmitter must be enabled,
and must be configured as a mA Input, wired to the water cut monitor, and configured appropriately. HART integration is supported on Channel A by polling.
For the 5700 Ethernet transmitter, an external host system is required to accept the
water cut monitor input and feed it to the transmitter.
For the 5700 intrinsically safe transmitter, HART integration is supported on Channel A
by polling.
Note
Water cut cannot be input into the 5700 transmitter with FOUNDATION fieldbus output if the software revision of the transmitter is 1.x.
Procedure
1. Set the mass flow cutoff and the volume flow cutoff to a non-zero value. This
ensures that totalizing stops when flow is stopped. In most installations, the default value is satisfactory.
Option
Description
Display Menu Configuration Process Measurement Flow
Variables Mass Flow Settings
ProLink III Menu Configuration Process Measurement Flow
Web browser Configuration Process Measurement Flow Variables
2. Navigate to one of the following paths based on the tool you are using.
Option
Description
Display Menu Configuration Process Measurement
Adv Phase Measurement
ProLink III Device Tools Configuration Process
Measurement Advanced Phase Measurement
18 Micro Motion Advanced Phase Measurement
Application Manual Measurement options and configuration
MMI-20030076 September 2020
Option Description
Field Communicator Configure Manual Setup Advanced Phase
Measurement
Web browser (Ethernet transmitters only)
Configuration Process Measurement Advanced Phase Measurement
3. Depending on your configuration tool, configure net oil measurement:
From the display, navigate to Net Oil (NOC) and Save.
From ProLink III, set Fluid Type to Net Oil (NOC) or Net Oil (NOC) and Liquid with
Gas if both are licensed.
4. Verify or make changes to the following settings:
Depending on which tool you are using, the parameter name or order may be different.
Parameter Description
Production Type
APM Action If you are using variable flow, when the drive gain threshold
Continuous flow if the flow in your system is generally
constant. Continuous flow assumes a constant flow rate that enables mass flow to also be remediated by subtracting GVF.
Variable flow if the flow in your system frequently starts
and stops or fluctuates between high and low flow rates.
See Production type options.
is exceeded, you can configure your transmitter to hold the last value, use density oil @ line, or a density hold override, to handle the volume calculation for the period of high drive gain.
See Remediation options (APM Action).
Density Hold Override If the Density Hold Override is used, then density will report
this value when in remediation. This value should represent the liquid mixture density at line conditions.
Drive Gain Threshold
User Input DG Threshold For use with manual threshold only, this is the drive gain
Density Oil @ Line The NOC algorithm converts the density of dry oil at
Application Manual 19
From the display, enable or disable the automatic
threshold. Disabling automatic threshold defaults to manual. You must configure a manual threshold value.
From ProLink III, select automatic or manual drive gain
threshold.
See Drive gain threshold mode.
value above which APM will remediate.
reference conditions (a user-configured value) to density at line conditions, and calculates volume. This option assumes that all volume during the entrained gas event is dry oil.
Measurement options and configuration Application Manual
September 2020 MMI-20030076
Parameter Description
Gas Density @ Line The density of entrained gas. Default = 0.
Use Default (zero) if the line pressure is less than 250 psi
(17.24 bar) absolute.
If the typical line pressure is greater than 250 psi
(17.24 bar) absolute, set this parameter to the density of the entrained gas at typical line conditions. This setting will affect the GVF output (and mass flow remediation if Process Type is set to Continuous).
This setting is needed only when Net Oil is combined with Liquid with Gas to more accurately determine GVF at conditions where gases have more mass - such as at high pressure.
Water Density @ Ref The density of water, corrected to the reference
temperature obtained from density determination.
Dry Oil Density @ Ref The density of dry oil at reference conditions obtained from
a density determination.
Reference Temp The temperature to which net oil and net water
measurements will be corrected. The default is 60 °F.
View Production Meas
(display only)
The type of net oil data that will be shown on the display.
If you select Corrected to Standard, the display shows
Watercut @ Ref, Net Oil Flow @ Ref, etc.
If you select Uncorrected, the display shows Watercut
@ Line, Net Oil Flow @ Line, etc.
This parameter is applicable only if a net oil process variable is configured as a display variable.
Average Reporting Interval
If using the display, navigate left to select.
Transmitter Date and Time
Not available in the display.
The time period, in minutes, over which process variables will be averaged. The averages are available for retrieval by the host system.
Displays the currently-set date.
From the display, choose Menu Configuration
Time/Date/Tag.
From ProLink III, choose Device Tools Configuration
Process Measurement Advanced Phase Measurement in the Transmitter Date field.
From the display, set the time zone.
From ProLink III, set the time zone, time zone offset,
date, and time.
Contract Start Time
If using the display, navigate left until you can select
Contract Period to see Contract Start Hour.
20 Micro Motion Advanced Phase Measurement
The time of day at which the contract starts. Enter the time in a 24-hour HH:MM format, where 00:00 = midnight and 18:30 = 6:30 pm.
Application Manual Measurement options and configuration
MMI-20030076 September 2020
Parameter Description
Contract Total [1 - 4]
Not available from the display.
Shrinkage Factor-Adjusted Volume Flow Outputs
Shrinkage Factor A user-input value, multiplied by the measured volume to
Meter Factor for Shrinkage A meter factor, established by proving, that acts as a

2.3 Gas with liquid

This measurement option improves mass flow measurement in gaseous processes with intermittent entrained liquids (mist).
Gas with liquid measurement process
The presence of entrained liquids (or mist) can cause significant errors when measuring the mass flow of gas through a Coriolis meter. Because droplets contribute a relatively large amount of mass to a gas mixture, but do not always move uniformly through the pipe with the gas, even a small amount of condensate can cause measurement discrepancies between what is measured by the meter and what emerges from the pipe downstream.
Up to four user-specified totals that are used to measure production.
See Read contract totals.
Enable or disable shrinkage-factored variable calculations.
match the expected volumes after shrinkage.
multiplier to shrinkage-factored (SF) variables. Not commonly used.
So how can you tell when a gas contains mist or condensate? When liquids are present in a gas stream, Coriolis meters will report an increase in both drive gain and fluid density
(4)
in combination with an increase in mass flow rate – due to the higher mass of the combined gas-liquid mixture. Therefore, in order to measure only the gaseous portion of the stream, the mass of the liquids must be ignored or subtracted from the mixture total.
(5)
The Advanced Phase Measurement software performs exactly this function, using drive gain as the diagnostic indication that mist or entrained liquids are present in the gas flow stream, and then substituting a gas-only flow rate in place of the live measurement until the liquid event has subsided. When the mist event is over, indicated by an associated drop in drive gain, the software returns to reporting the live measured mass flow rate. If the flow rate after the event differs significantly from the rate prior to the event, the software will apply an adjustment to the mass flow output until the totalizers accurately represent the flow rate change that occurred during the mist event.
(6)
The following figure shows how the change in density and drive gain are affected by mist.
(4) Coriolis meters do not measure the density of gases accurately, but the density reading can be combined with drive gain as
a useful diagnostic to detect changes in fluid properties.
(5) The unmeasured liquids can be (and are often) collected and processed separately downstream if desired. (6) The outputs are adjusted by a maximum of ±10% of live reading.
Application Manual 21
Measurement options and configuration Application Manual
September 2020 MMI-20030076
Figure 2-3: Effect of transient mist on drive gain and flow measurement
B
D
A
C C
DG%
FLOW
t
FLOW = mass flow rate
DG% = drive gain percentage
t = Time
A. Drive gain indication
B. Measured mass flow rate C. Entrained liquid (mist) occurring during these intervals D. Drive gain threshold
The green line (A) shows the drive gain indication during flow. If mist gets entrained into the gas, the drive gain and the mass flow rate (B) will both increase. The red line (D) shows a drive gain threshold above which mist is entrained in the gas. APM will automatically choose an appropriate drive gain threshold (D) or you can opt to set the threshold manually.
The Advanced Phase Measurement software identifies entrained mist in gas flow by detecting the sharp increases in drive gain and corresponding increases in mass flow measurement. When automatic drive gain threshold is enabled, the software continuously monitors the most recent drive gain data (up to 60 minutes) to determine the drive gain threshold. If the measurement exceeds the drive gain threshold, the fluid is deemed to contain entrained liquid, and remediation occurs.
Gas with liquid remediation
The following figure illustrates Advanced Phase Measurement software processing when mist is detected in the gas stream.
22 Micro Motion Advanced Phase Measurement
Application Manual Measurement options and configuration
MMI-20030076 September 2020
Figure 2-4: Advanced Phase Measurement software: Gas with liquid remediation
E
C
B
F
E
G
C
F
G
D
A
DG%
FLOW
t
FLOW = mass flow rate
DG% = drive gain percentage
t = Time
A. Drive gain indication
B. Measured mass flow rate C. Pre-mist averaging period and source of average flow rate during this period D. Drive gain threshold
E. Post-mist adjustment delay and source of average flow rate during this period
F. Held flow rate during entrained mist intervals
G. Post-mist adjustment applied to measured flow rate (if applicable)
The Advanced Phase Measurement software uses the drive gain and mass flow readings to correct for the presence of entrained mist in gas. The green line (A) shows the drive gain during flow. The blue line (B) shows the measured mass flow reading both with, and without entrained gas. When mist is present, the drive gain will go above the drive gain threshold (D). This threshold value is automatically determined by APM or manually adjustable. When the drive gain exceeds this threshold, APM will look back in time (C) seconds, and determine an average, pre-mist, mass flow rate. It will then substitute mass flow rate (F) until the drive gain goes back below the threshold. Then during time period (E), a post-mist, mass flow rate average is determined. If the post- and pre-mist averages are different, a mass flow adjustment (G) is made to the mass rate after time period (E) until the total mass difference is reconciled.
The first mist event had equal mass flow rates before and after the mist event, so (G) is equal to (F). In the second mist event, the post-mist flow rate was greater than the pre-
Application Manual 23
Measurement options and configuration Application Manual
September 2020 MMI-20030076
mist rate, so a (G) was applied that was greater than (F). This held flow rate (F) is also known as the remediated mass flow rate.

2.3.1 Configure gas with liquid measurement

Configuring gas with liquid measurement is one part of the following required configuration for APM functionality.
Start
End
Configure units
See configuration
and use manual
Configure totalizers
and inventories
See configuration
and use manual
Configure gas with
liquid measurement
(this procedure)
Configure channels
and outputs
See configuration
and use manual
Procedure
1. Set the mass flow cutoff and the volume flow cutoff to a non-zero value. This
ensures that totalizing stops when flow is stopped. In most installations, the default value is satisfactory.
Option
Description
Display Menu Configuration Process Measurement Flow
Variables Mass Flow Settings
ProLink III Menu Configuration Process Measurement Flow
Web browser Configuration Process Measurement Flow Variables
2. Navigate to one of the following paths based on the tool you are using.
Option
Description
Display MenuConfigurationProcess Measurement
Adv Phase Measurement
ProLink III Device Tools Configuration Process
Measurement Advanced Phase Measurement
Field Communicator or enhanced FF host
Basic FF host (fieldbus
Configure Manual Setup Advanced Phase Measurement
The APM transducer block
transmitters only)
24 Micro Motion Advanced Phase Measurement
Application Manual Measurement options and configuration
MMI-20030076 September 2020
Option Description
Web browser (Ethernet transmitters only)
Configuration Process Measurement Advanced Phase Measurement
3. Depending on your configuration tool, configure net oil measurement:
From the display, navigate to Gas with Liquid and Save.
From ProLink III, set Fluid Type to Gas with Liquid.
4. Verify or make changes to the following settings:
Depending on which tool you are using, the parameter name or order may be different.
ProLink III Description
Drive Gain Threshold
User Input DG Threshold For use with manual threshold only, this is the drive gain
Pre-Mist Ave Period Period of stable gas flow conditions prior to a liquid
From the display, enable or disable the automatic
threshold. Disabling automatic threshold defaults to manual. You must configure a manual threshold value.
From ProLink III, select automatic or manual drive gain
threshold.
See Drive gain threshold mode.
value above which APM will remediate.
entrainment event. Averaging period is user configurable with units = n seconds. The average flow rate during this period will be reported until the entrained liquid event has passed. Default = 15 seconds.
Post-Mist Adj Delay Period of stable gas flow conditions after a liquid
entrainment event. After this delay, the average flow rate during Post-Mist Adjustment Delay is averaged with the Pre­Mist Averaging Period and the flow rate is adjusted a maximum of +/-10% of reading until any difference has been appropriately corrected in the totalizers. Default = 15 seconds.
Average Reporting Interval
If using the display, navigate left to select.
Transmitter Date and Time
Not available in the display.
Application Manual 25
The time period, in minutes, over which process variables will be averaged. The averages are available for retrieval by the host system.
Displays the currently-set date.
From the display, choose Menu Configuration
Time/Date/Tag.
From ProLink III, choose Device Tools Configuration
Process Measurement Advanced Phase Measurement in the Transmitter Date field.
From the display, set the time zone.
From ProLink III, set the time zone, time zone offset,
date, and time.
Measurement options and configuration Application Manual
September 2020 MMI-20030076
ProLink III Description
Contract Start Time
If using the display, navigate left until you can select
Contract Period to see Contract Start Hour.
Contract Total [1 - 4]
Not available from the display.
The time of day at which the contract starts. Enter the time in a 24-hour HH:MM format, where 00:00 = midnight and 18:30 = 6:30 pm.
Up to four user-specified totals that are used to measure production.
See Read contract totals.
26 Micro Motion Advanced Phase Measurement
Application Manual Additional configuration
MMI-20030076 September 2020

3 Additional configuration

3.1 Configure viewing and reporting for process variables

When Advanced Phase Measurement software is enabled, additional process variables are available.
Procedure
To configure a process variable as a display variable, or to report a process variable over
an output, see the transmitter configuration and use manual.
To query a process variable using Modbus, follow standard Modbus programming
techniques.
3.1.1

Advanced Phase Measurement specific process variables

The process variables listed here are available only when Advanced Phase Measurement software is enabled.
The following table lists the process variables by measurement option, and provides information on reporting.
Process variable
Gas Void Fraction
DensityOil
DensityOil
NetFlowOil
NetFlowOil
NetTotalOil
NetTotalOil
NetFlowWater
NetFlowWater
NetTotalWater
NetTotalWater
Watercut
Watercut
@Line
@Ref
@Line
@Ref
@Line
@Ref
@Line
@Ref
@Line
@Ref
@Line
@Ref
Advanced Phase Measurement
option
Liquid
with gas
Net Oil Gas with
liquid
Viewing and reporting
Display Modbus
and
Ethernet
Fieldbus
1)
(
mAO1, mAO2,
mAO3
FO1, FO2
Application Manual 27
Additional configuration Application Manual
September 2020 MMI-20030076
Process variable Advanced Phase Measurement
option
Liquid
with gas
APM contract period
Today's total
Yesterday's total
(1) For fieldbus version 1.x transmitters, you can only publish through the AI Blocks four process variables and two
totalizers/inventories.
3.1.2

Default display variables

Net Oil Gas with
liquid
Display Modbus
Viewing and reporting
Fieldbus
and
Ethernet
1)
(
mAO1, mAO2,
mAO3
The following table lists the default display variables for Advanced Phase Measurement software. In all cases, the two-line display option is enabled. All of these process variables can be configured as display variables.
Display variable Advanced Phase Measurement option
Liquid with gas NOC with gas Gas with liquid
Two-line display, Variable 1
Volume Flow Rate Volume Flow Rate Mass Flow Rate
FO1, FO2
Two-line display, Variable 2
Display Variable 1 Density Net Oil Flow Rate
Display Variable 2 Temperature Net Oil Total
Display Variable 3 GVF Net Water Total
Display Variable 4 Empty Temperature Empty
Display Variable 5 Empty GVF Empty
Display Variable 6–15 Empty Empty Empty
Volume Total Water Cut
@Line
@Line
@Line
@Line
Mass Total
Temperature
Density
Empty

3.1.3 Period averaged output (PAO) configuration

Period averaged outputs provide an easy and reliable way to prevent data collection systems from recording outliers in erratic data. This is included with Advanced Phase Measurement software since two-phase conditions cause more volatile measurement outputs. The period over which the PAOs are averaged can be adjusted using the Average Reporting Interval parameter.
28 Micro Motion Advanced Phase Measurement
Application Manual Additional configuration
MMI-20030076 September 2020
Figure 3-1: Example density measurement without averaged outputs and with average outputs
ρ
ρ
0.9
0.8
0.7
0.6
0.9
0.8
0.7
0.6
A
t
1
2
3
B
t
1
2
3
ρ = Density
t = Time
A. Density measurement without averaged outputs
B. Density measurement with averaged outputs
3.1.4
Application Manual 29

Configure PAO

Procedure
Navigate to one of the following paths to set the interval.
Option
Display Configuration Adv Phase Measurement Reporting Interval
ProLink III Device Tools Configuration Process Measurement Advanced Phase
Description
Measurement Average Reporting Interval
Additional configuration Application Manual
September 2020 MMI-20030076

3.2 Configure APM contract totals into the totalizer history log

The transmitter can be configured to store Advanced Phase Measurement contract totals to the totalizer history log. This allows you to access totals from earlier contract periods. Otherwise, the transmitter maintains data for only the current contract period (today) and the immediately preceding contract period (yesterday).
Procedure
1. Ensure that you have configured the Advanced Phase Measurement contract totals as desired.
Option Description
Display Menu Configuration Process Measurement Adv Phase
Measurement Contract Period
ProLink III Device Tools Contract Period Totals
3.2.1
2. Navigate to the Totalizer Log.
Option Description
Display Menu Configuration Totalizer Log
ProLink III Device Tools Configuration Totalizer Log
Field Communicator Not available
3. Set Log Total 1, Log Total 2, Log Total 3, and/or Log Total 4 to the desired Advanced Phase Measurement contract total.
You can configure the totalizer history log to include both Advanced Phase Measurement and standard totals.
Related information
Read contract totals
(7)

History log variables

The following variables are added to the history log when APM is enabled. The output type is not equal to zero.
Liquid with gas
Gas Void Fraction
Unremediated Density
For fieldbus version 1.x transmitters, any two of the publishable totalizers and inventories can be used, but only two at a
(7)
time.
30 Micro Motion Advanced Phase Measurement
Application Manual Additional configuration
MMI-20030076 September 2020
Unremediated Mass Flow
External Temperature
Velocity
Net oil with gas and NOC multiple wells
Net Flow Oil @ Ref
Watercut @ Ref
Density of Oil @ line
Unremediated Density
Unremediated Mass Flow
Gas with liquid
Unremediated Mass Flow
GSV Volume Flow
TMR Time
Liquid mass flow estimate
Extended drive gain
Net oil without gas
Net Flow Oil @ Ref
Watercut @ Ref

3.3 Configure events

When Advanced Phase Measurement software is enabled, additional process variables are available to use in event configuration.
Procedure
See the transmitter configuration manual for instructions on configuring events.
Application Manual 31
Additional configuration Application Manual
September 2020 MMI-20030076
32 Micro Motion Advanced Phase Measurement
Application Manual Transmitter operation
MMI-20030076 September 2020

4 Transmitter operation

4.1 Read process variables

Advanced Phase Measurement process variables can be configured as display variables or assigned to outputs. See the transmitter configuration manual for information on using the display to read process variables. You can also use the host system for your Modbus, fieldbus, or Ethernet/IP network to read the variables.

4.2 Read contract totals

You can read contract totals for the current 24-hour contract period and for the previous 24-hour contract period. Depending on the configuration of the totalizer log, you may be able to read contract totals for earlier periods.
The contract totals are derived from existing inventories. However, they are reset automatically at the beginning of each contract period. Therefore, the values shown will probably not match the values shown for the inventories.
Important
You can reset inventories manually, and you can stop and start inventories manually. However, if you do this, data for the current contract period will not reflect the entire 24­hour period. Data for earlier contract periods is not affected.
The contract totals for the current contract period are stored in the Today's Total [1-4]
parameters.
The contract totals for the previous contract period are stored in the Yesterday's Total
[1-4] parameters.
The contract totals from earlier contract periods can be read in the totalizer log.
Related information
Configure APM contract totals into the totalizer history log
Application Manual 33
Transmitter operation Application Manual
September 2020 MMI-20030076
34 Micro Motion Advanced Phase Measurement
Application Manual APM alerts
MMI-20030076 September 2020

5 APM alerts

This section provides information on the status alerts associated with the APM application. For information on all other Model 5700 alerts, see the appropriate Micro Motion Model 5700 configuration and use manual.
Alert Cause
A138 APM Remediation Remediation is active.
Watercut limited to 0% Watercut has exceeded the upper limit. Check base oil density.
Watercut limited to 100% Watercut has exceeded the lower limit. Check base water
density.
Watercut Unavailable Watercut unavailable due to high gas. Consider using external
watercut meter.
All alerts are configurable
All alert severities default to Out of Specification
No alerts are affected by fault timeout
Application Manual 35
APM alerts Application Manual
September 2020 MMI-20030076
36 Micro Motion Advanced Phase Measurement
Application Manual Application parameters and data
MMI-20030076 September 2020

A Application parameters and data

This section includes only Modbus registers that are specific to or redefined for the Advanced Phase Measurement software configuration. To use Modbus to configure other parameters, see the Micro Motion Modbus Map.
Typically, Advanced Phase Measurement configuration is performed using either ProLink III or the display. This information is provided for completeness.
A.1 Advanced Phase Measurement Modbus
configuration parameters
Parameter APM option Modbus
Liquid
with gas
Average Reporting Interval
Contract Start Time 3966 U16 hours (0–23)
Contract Total [1-4] 3967-3970U16 4 = Inventory 1
Density Corrective Action
Drive Gain Threshold Override
3900 U16 minutes (1–1440)
3971 U16 0 = manual
Net Oil
with gas
Gas with
liquid
4450 U16 0 = Hold Last
Address Data
type
Integer codes / Unit (Range)
7 = Inventory 2
18 = Inventory 3
64 = Inventory 4
25 = Inventory 5
28 = Inventory 6
31 = Inventory 7
Value
1 = Density Oil @
Line
1 = auto
Drive Gain Threshold Override Value
Dry Oil Density @ Ref
Gas Density
Application Manual 37
(1)
3998-39
99
1959 Float g/cm³ (0.2–1.5)
3935 Float Configured unit
Application parameters and data Application Manual
September 2020 MMI-20030076
Parameter APM option Modbus
Liquid
with gas
Output Type 3940 U16 0 = No
Post-Mist Averaging Period
Pre-Mist Averaging Period
Net Oil
with gas
Gas with
liquid
Address Data
type
620 U16 seconds (2–128)
619 U16 seconds (2–128)
Integer codes / Unit (Range)
remediation
1 = Liquid with
gas, continuous flow
2 = NOC with
gas, continuous flow
3 = Liquid with
gas, variable flow
4 = NOC with
gas, variable flow
5 = Gas with
liquid
6 = NOC only
Reference Temperature
Water Density @ Ref 1831 Float g/cm³ (0.5–1.5)
(1) At line conditions
319 Float Configured unit
A.2 Advanced Phase Measurement default totalizer
and inventory values
Totalizer/Inventory Liquid with Gas Gas with Liquid Net Oil
1 Mass Flow
(Remediated)
2 Volume Flow
(Remediated)
3 Temperature
Corrected Volume
4 Gas Standard Volume Gas Standard Volume Net Oil @ Line
5 Standard Volume Standard Volume Net Water @ Ref
6 Net Mass Net Mass Net Water @ Line
Mass Flow (Remediated)
Volume Flow (Remediated)
Temperature Corrected Volume
Mass Flow (Remediated)
Volume Flow
Net Oil @ Ref
38 Micro Motion Advanced Phase Measurement
Application Manual Application parameters and data
MMI-20030076 September 2020
Totalizer/Inventory Liquid with Gas Gas with Liquid Net Oil
7 Net Volume Net Volume Net Volume
A.3 Advanced Phase Measurement Modbus
process variables
Process variable
Gas Void Fraction 3907 Float %
Density Oil @ Line 345 Float SGU
Density Oil @ Line 347 Float °API
Density Oil @ 60F 1665 Float °API
Net Oil Flow @ Line 1553 Float Configured unit
Net Oil Flow @ Ref 1547 Float Configured unit
Net Oil Inventory @
(2)
Line
Net Oil Inventory @
(3)
Line
Net Oil Total @
(2)
Ref
Net Oil Total @
(4)
Ref
(1)
Advanced Phase
Measurement option
Liquid
with gas
Net Oil
with gas
1665 Float Automatically
4240 Double Automatically
1661 Float Automatically
4236 Double Automatically
Gas with
liquid
Address Data
type
Modbus
Unit
derived from configured unit
derived from configured unit
derived from configured unit
derived from configured unit
Net Water Flow @ Line
Net Water Flow @ Ref
Net Water Total @
(2)
Line
Net Water Total @
(5)
Line
Net Water Total @ Ref
Application Manual 39
1561 Float Configured unit
1549 Float Configured unit
1667 Float Automatically
derived from configured unit
4248 Double Automatically
derived from configured unit
1663 Float Automatically
derived from configured unit
Application parameters and data Application Manual
September 2020 MMI-20030076
(7)
(1)
Advanced Phase
Measurement option
Liquid
with gas
Net Oil
with gas
4244 Double Automatically
Modbus
Gas with
liquid
Address Data
type
989 U32 Seconds
433, Bit
#12
U16 0 = Inactive
Unit
derived from configured unit
1 = Active
Process variable
Net Water Total @
(6)
Ref
Watercut @ Line 1555 Float %
Watercut @ Ref 1557 Float %
Total Mist Time
APM Liquid with Gas remediation status
(1) For fieldbus version 1.x transmitters, you can only publish through the AI Blocks four process
variables and two totalizers/inventories.
(2) Legacy NOC register (3) Inventory 4 (only if you are using the default configuration) (4) Inventory 3 (only if you are using the default configuration) (5) Inventory 6 (only if you are using the default configuration) (6) Inventory 5 (only if you are using the default configuration) (7) Automatically set to 0 on a power cycle

A.4 Period Averaged Options current period data

Process variable
PAO Mass Flow 3949 Float Configured unit
PAO Density 3951 Float Configured unit
PAO Volume Flow 3953 Float Configured unit
PAO Net Oil Flow @ Line
PAO Net Oil Flow @ Ref
PAO Watercut @ Line
PAO Gas Void Fraction
PAO Temperature 3963 Float Configured unit
Unremediated Mass Flow
(1)
Advanced Phase
Measurement option
Liquid
with gas
3961 Float Configured unit
3943 Float Automatically
Net Oil
with gas
Gas with
liquid
3955 Float Configured unit
3957 Float Configured unit
3959 Float Configured unit
Address Data
Modbus
Unit
type
derived from configured uni
40 Micro Motion Advanced Phase Measurement
Application Manual Application parameters and data
MMI-20030076 September 2020
Process variable
(1)
Advanced Phase
Modbus
Measurement option
Unremediated Density
Liquid
with gas
3945 Float Automatically
Net Oil
with gas
Gas with
liquid
Address Data
type
Unit
derived from configured uni
Unremediated Volume Flow
3947 Float Automatically
derived from configured uni
(1) For fieldbus version 1.x transmitters, you can only publish through the AI Blocks four process
variables and two totalizers/inventories.
Application Manual 41
Application parameters and data Application Manual
September 2020 MMI-20030076
42 Micro Motion Advanced Phase Measurement
Application Manual Typical oil and gas applications and other information
MMI-20030076 September 2020
B Typical oil and gas applications and
other information
B.1 Advanced Phase Measurement with a two-
phase separator
A. From wellhead
B. Separator C. Gas leg D. Oil/water leg
E. Coriolis sensor and transmitter with Advanced Phase Measurement (NOC with gas)
F. Coriolis sensor and transmitter with Advanced Phase Measurement (Gas with liquid)
G. Modbus host (flow computer)
Related information
Best practices for two-phase measurement performance
Application Manual 43
Typical oil and gas applications and other information Application Manual
September 2020 MMI-20030076
B.2 Advanced Phase Measurement with a three-
phase separator
A. From wellhead
B. Separator C. Water leg D. Oil leg
E. Gas leg
F. Coriolis sensor and transmitter with Advanced Phase Measurement (NOC with gas,
variable flow)
G. Coriolis sensor and transmitter with Advanced Phase Measurement (Liquid with gas,
variable flow)
H. Coriolis sensor and transmitter with Advanced Phase Measurement (Gas with liquid)
I. Modbus host (flow computer)
44 Micro Motion Advanced Phase Measurement
Application Manual Typical oil and gas applications and other information
MMI-20030076 September 2020

B.3 Advanced Phase Measurement at the wellhead

A. Wellhead
B. Coriolis sensor C. Transmitter with Advanced Phase Measurement (NOC with gas) D. Modbus host (flow computer)
Application Manual 45
Typical oil and gas applications and other information Application Manual
September 2020 MMI-20030076

B.4 Automatic Drive Gain Threshold determination

An essential function of the Advanced Phase Measurement software is to monitor drive gain, and then use drive gain data to define and adjust the Drive Gain Threshold, which ultimately determines when and how measurement remediation is needed.
E
A
B
D
C
A
B
C
D
DG%
t
1
t
2
DG% = Drive Gain Percentage
t1 = Drive Gain Threshold interval 1
t2 = Drive Gain Threshold interval 2
A. Drive Gain Threshold (measurement will be remediated if drive gain exceeds this
threshold)
B. Drive gain indication C. Minimum drive gain detected at the beginning of each drive gain threshold interval. D. Drive Gain Threshold Addition. This value is added to the minimum drive gain (C) to
establish a small buffer so the software only remediates values that exceed the typical drive gain fluctuations for each unique process.
E. Advanced Phase Measurement software remediates density during these intervals.
This figure illustrates how the software determines the Drive Gain Threshold (A) under live conditions if automatic drive gain threshold is configured. The green dotted line shows the Coriolis sensor’s live drive gain indication (B) over time (t). The software continuously analyzes the live drive gain using the most recent data (duration set by the Drive Gain Threshold Interval) to determine the lowest drive gain that is typical for the process.
(8)
The live drive gain indication for most applications fluctuates a bit under normal operating conditions, typically within a small range that is not attributable to two-phase flow or other process upsets. To avoid remediating during this typical process noise, a small Drive Gain Threshold Addition (D) is added to the lowest sample point (C). The newly established Drive Gain Threshold (A) represents the sampled minimum drive gain plus the drive gain threshold addition. During the remainder of the threshold Interval, if the indicated drive gain (B) exceeds the established Drive Gain Threshold (A), the transmitter will appropriately remediate the measured flow rate and/or the fluid density.
(9)
After the
threshold interval is over, the process starts again.
The factory default for Drive Gain Threshold Interval is 60 minutes, which is suitable for most continuous processes, but it
(8)
may be shortened or lengthened as required for each application. Contact customer support for assistance.
46 Micro Motion Advanced Phase Measurement
Application Manual Typical oil and gas applications and other information
MMI-20030076 September 2020

B.5 Manual Drive Gain Threshold

When using the manual drive gain threshold option, select a value above the typical variation in the drive gain under normal flowing conditions. Remediation will occur when the drive gain exceeds this threshold limit. If the threshold is set too high, it is possible that some gas entrainment may occur without exceeding the established threshold, and hence no remediation will occur. Alternatively, if the threshold is too low, the process may be remediated even under normal flowing conditions.
If you are using ProLink III, view variations and maximum values during normal operation by using trends or data logs on the drive gain data. If your process is batched or cyclical, you should observe and/or record several cycles to establish normal drive gain levels.
If data collection is not available, view the drive gain on the 5700 display by configuring the display variables. For more information, see the Model 5700 configuration and use manual.
DG% = Drive Gain Percentage
t = Time
A. Drive gain indication
B. Drive gain threshold
C. Remediation occurring during these intervals
(9) The measurement variables remediated by APM are configurable per license type and software and output configuration.
Application Manual 47
Typical oil and gas applications and other information Application Manual
September 2020 MMI-20030076
48 Micro Motion Advanced Phase Measurement
Application Manual Best practices for two-phase measurement performance
MMI-20030076 September 2020
C Best practices for two-phase
measurement performance
Related information
Advanced Phase Measurement with a two-phase separator

C.1 Entrained gas performance

Measurement accuracy for liquids with entrained gas is a complex function of GVF, viscosity, velocity, sensor geometry, drive frequency, and orientation. The best measurement performance will always be achieved if fluid can be measured in single­phase. Add a free-gas knockout upstream if possible. The following guidelines apply regardless if APM options are licensed or not. When gas entrainment is inevitable, APM will improve the measurement performance.
Common sources for unintentional gas entrainment
Long drops from fill point to liquid level in tanks
Agitators and mixers
Leaks in seals or pumps
Pumping out of nearly empty tanks
Pressure loss (flashing) for volatile liquids
Pumping through nearly empty piping
Ways to minimize flow errors
Use ELITE® (low frequency) sensors whenever possible. F-Series and H-Series sensors
are also acceptable, but less accurate.
Do not use T-Series sensors or Models F300/H300 compact because they have a high
operating frequency.
Orient the meter properly:
Table C-1: Preferred sensor orientation for liquids with entrained gas
Process Preferred orientation
Delta-shaped sensors (CMF010, CMF025, CMF050, CMF100)
Application Manual 49
Best practices for two-phase measurement performance Application Manual
September 2020 MMI-20030076
Table C-1: Preferred sensor orientation for liquids with entrained gas (continued)
Process Preferred orientation
Any F-Series or CMFS sensor, and CMF200 or larger (flow should go up)
Ensure sensor is filled as quickly as possible, and stays full during measurement:
For horizontal pipes, maintain a minimum flow velocity of 1 m/s to purge air from
an empty pipe and keep it full.
For vertical pipes, flow upward and maintain minimum velocity of 1 m/s to prevent
solids from settling out of the fluid.
Add back pressure, or increase line pressure, to minimize size of bubbles in flow
stream.
Size the meter appropriately to operate normally as close to the sensor nominal flow
rate as is practical. Higher velocity leads to better performance, as long as pressure drop does not cause liquids flash.
Ensure fluid is well mixed. If needed, you can install a blind “T” and/or static mixer just
upstream of the sensor to evenly distribute bubbles through both sensor tubes. If using a blind “T”, install it in the same plane as the sensor tubes.
If re-zeroing in the field is necessary, zeroing must be done on a pure liquid without
bubbles in order to avoid error. If this cannot be done, use the factory zero.
Minimize damping on outputs to minimize processing delay from electronics.
Do not stop the totalizer immediately after batch; allow the totalizer to stabilize for
approximately 1 second.
Set Flow Cutoff as high as is practical to avoid totalizing at no flow condition if bubbles
remain in the sensor.

C.2 Entrained liquid (mist) performance

Measurement accuracy for gases with entrained liquids (mist) is mostly related to the amount of mass contained in liquid droplets compared to an equivalent volume of gas containing the same mass. It is important to choose the correct sensor. Otherwise, sensor geometry, drive frequency, and orientation can cause errors that reduce performance. The best measurement performance will always be achieved if fluid can be measured in single­phase. Add a liquid trap upstream if possible. The following guidelines apply regardless if APM options are licensed or not. When liquid entrainment is inevitable, APM will improve the measurement performance.
50 Micro Motion Advanced Phase Measurement
Application Manual Best practices for two-phase measurement performance
MMI-20030076 September 2020
Common sources for unintentional liquid entrainment
Temperature loss (condensation)
Pressure increase
Poorly managed level control in separators or GLCCs
Malfunctioning or over-filled liquid traps
Ways to minimize measurement errors
Use ELITE® (low frequency) sensors whenever possible. F-Series and H-Series sensors
are also acceptable, but less accurate.
Size the meter appropriately for gas flow. Avoid high turndowns where sensor
sensitivity may be reduced.
Do not use T-Series sensors or compact Models F300/H300 because they have a high
operating frequency.
Use the enhanced core processor (Model 800) or : they perform best in applications
with entrained liquid.
Orient the meter properly:
Table C-2: Preferred sensor orientation when there could be entrained liquid
Process Preferred orientation
Delta-shaped sensors (CMF010, CMF025, CMF050, CMF100)
Any F-Series or CMFS sensor, and CMF200 or larger (flow should go down)
Ensure sensor is dried (blown-out) as quickly as possible, and stays dry during
measurement.
Avoid temperature losses; insulation is highly recommended if condensate is caused by
cooling temperatures.
Avoid pressure increases in the system; Ensure that pressure regulators are functioning
properly.
If entrained liquid is unavoidable, try to ensure that the process is well mixed.
Application Manual 51
Best practices for two-phase measurement performance Application Manual
September 2020 MMI-20030076
Avoid elbows, valves, or other components that may introduce a flow profile affecting
one tube (for example, a swirling motion entering the flow tubes)
If re-zeroing in the field is necessary, zeroing must be done on a pure gas without liquid
in order to avoid error. If this cannot be done, use the factory zero.
Minimize damping on outputs to minimize processing delay from electronics.
Do not stop the totalizer immediately after batch; allow the totalizer to stabilize for
approximately 1 second.
Set Flow Cutoff as high as is practical to avoid totalizing at no flow condition if droplets
remain in the sensor.

C.3 Density determination

If you are using either the density of water from the well, corrected to reference temperature, and the density of dry oil from the well, corrected to reference temperature.
Important
Micro Motion recommends working with a laboratory to obtain the most accurate values. The accuracy of the data depends upon the accuracy of these two density values.
the PVR application or the Net Oil application, you must know

C.3.1 Density determination using a three-phase separator

To configure net oil measurement, you must know the density of dry oil at reference temperature, and the density of produced water at reference temperature. If you have a three-phase separator, you can use density data and the Oil & Water Density Calculator to obtain these values.
Note
Even after separation, oil typically contains some amount of interstitial water. The water cut may be as high as 1% to 3%. For purposes of this application, this is considered dry oil.
Prerequisites
You must have a three-phase separator in the process. You can use a mobile three-phase test separator.
You must have a sensor and transmitter installed on the oil leg, and a sensor and transmitter installed on the water leg or determine the water density separately by manual sampling.
You must have the Oil & Water Density Calculator. This is a spreadsheet tool developed by Micro Motion. You can obtain a copy from your Micro Motion representative or by visiting
https://www.emersonflowsolutions.com/oildensityref.
Important
The accuracy of net oil data depends on the accuracy of the density data. Never use an unstable density value, or any density value that has an elevated drive gain.
Procedure
1.
Wait until separation has occurred.
52 Micro Motion Advanced Phase Measurement
Application Manual Best practices for two-phase measurement performance
MMI-20030076 September 2020
2. At the transmitter on the oil leg, do one of the following options:
Read and record the density value and the temperature value
If logging the live variable data, monitor the live density at line conditions, or the
corrected density at 60 °F (15.6 °C) (modbus register 1655)
3. At the transmitter on the water leg, read and record the density value and the temperature value. Alternatively, enter the density of the water obtained by another method, such as sampling.
4. Use the Oil & Water Density Calculator to calculate the density of dry oil at reference temperature and the density of produced water at reference temperature. You can obtain a copy from your Micro Motion representative or by visiting https://
www.emersonflowsolutions.com/oildensityref.
Tip
Unless the oil is light hot condensate, the oil will almost always contain some interstitial water. This is generally acceptable for allocation measurements. However, if further accuracy is desired, you can determine the water cut and use it in the calculation. To determine or estimate the water cut, take a shakeout sample from one of the following:
C.3.2
The current flow/dump cycle, at the time of minimum density
Similar oils produced from the same reservoir
The tank or tanks that the separator flows into
Enter this water cut into the Oil & Water Density Calculator to calculate the density of dry oil at reference temperature.
Density determination using a petroleum laboratory
To configure APM for net oil measurement, you must know the density of oil at reference temperature, and the density of produced water at reference temperature. You can obtain these values from a petroleum laboratory.
Note
Even after separation, oil typically contains some amount of interstitial water. The water cut may be as high as 1% to 3%.
Important
If you are using a three-phase separator, you can collect the oil sample and the water sample separately, after separation, or you can collect one sample before separation and have the laboratory perform the separation.
If you are using a two-phase separator, you should collect one sample before separation and have the laboratory perform the separation.
Prerequisites
Sample collection must meet these requirements:
You must be able to collect a sample that is representative of your process.
Application Manual 53
Best practices for two-phase measurement performance Application Manual
September 2020 MMI-20030076
The sample must be collected by a qualified person, using industry-accepted safety
standards.
You must know the minimum required sample size. This varies depending on the water
cut and the volume of the sample cylinder. Consult the petroleum laboratory for specific values.
If the sample contains oil, you must be able to collect and maintain the sample at line
pressure, so that the oil will not lose pressure and outgas. This will change the laboratory-measured density.
If you collect the water sample separately, you must be able to protect it from
contamination and evaporation.
You must know the reference temperature that you plan to use.
The petroleum laboratory must be able to meet these requirements:
The laboratory density meter must be able to keep the oil sample pressurized at line
pressure during the density measurement.
The sample cylinder must be a constant-pressure type, and must be properly rated for
the oil–water composition and for sample pressure.
The laboratory report must include the oil density, water density, and the reference
temperature.
Procedure
1. Communicate the handling and measurement requirements and the reference
temperature to the petroleum laboratory.
2. If you are collecting one sample that contains both oil and water, identify the point
in the line where the sample will be taken.
Recommendations:
Collect the sample at a point where the fluid is well mixed.
The line pressure at the sample point should be close to the line pressure at the
sensor.
The line temperature at the sample point should be close to the line
temperature at the sensor.
3. If you are using a three-phase separator and collecting the oil and water samples separately:
a) Identify the points where the samples will be taken.
Recommendations:
The sample point for oil must be on the oil leg, as close to the sensor as
possible.
The line pressure at the oil sample point should be similar to the line
pressure at the sensor.
The sample point for water must be on the water leg, as close to the
sensor as possible.
54 Micro Motion Advanced Phase Measurement
Application Manual Best practices for two-phase measurement performance
MMI-20030076 September 2020
The line temperature at the water sample point should be similar to the
line temperature at the sensor.
b) Wait until separation has occurred.
4. Collect the sample or samples, meeting all requirements for pressure and protection from contamination or evaporation.
5. Mark and tag the sample or samples with the well name or number, time and date, sample type, line pressure, and line temperature.
6. Transport the samples to the laboratory safely, as soon as is practical.
Postrequisites
If the laboratory measurements were not corrected to your reference temperature, use the Oil & Water Density Calculator to calculate density at reference temperature. This is a spreadsheet tool developed by Micro Motion. You can obtain a copy by visiting https://
www.emersonflowsolutions.com/oildensityref or from your Micro Motion representative.
Application Manual 55
*MMI-20030076*
MMI-20030076
Rev. AD
2020
Micro Motion Inc. USA
7070 Winchester Circle Boulder, Colorado USA 80301 T +1 303-527-5200 T +1 800-522-6277 F +1 303-530-8459
www.emerson.com
Micro Motion Asia
Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Republic of Singapore T +65 6363-7766 F +65 6770-8003
©
2020 Micro Motion, Inc. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Micro Motion, ELITE, ProLink, MVD and MVD Direct Connect marks are marks of one of the Emerson Automation Solutions family of companies. All other marks are property of their respective owners.
Micro Motion Europe
Emerson Automation Solutions Neonstraat 1 6718 WX Ede The Netherlands T +31 (0) 318 495 555 T +31 (0) 70 413 6666 F +31 (0) 318 495 556
www.emerson.com/nl-nl
Micro Motion United Kingdom
Emerson Automation Solutions Emerson Process Management Limited Horsfield Way Bredbury Industrial Estate Stockport SK6 2SU U.K. T +44 0870 240 1978 F +44 0800 966 181
Loading...