Micro Motion 9739MVD Operating Manual

Configuration and Use Manual
Micro Motion® 9739 MVD Transmitters
MMI-20016855, Rev AC
April 2013
Micro Motion customer service
Email
Asia-Pacific: APflow.support@emerson.com
North and South America Europe and Middle East Asia Pacific
United States 800-522-6277 U.K. 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 318 495 555 New Zealand 099 128 804
Mexico +41 (0) 41 7686 111 France 0800 917 901 India 800 440 1468
Argentina +54 11 4837 7000 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3238 3677 Italy 8008 77334 China +86 21 2892 9000
Venezuela +58 26 1731 3446 Central & Eastern +41 (0) 41 7686 111 Japan +81 3 5769 6803
Russia/CIS +7 495 981 9811 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudia Arabia 800 844 9564
UAE 800 0444 0684

Contents

Contents
Part I Getting Started
Chapter 1 Before you begin .............................................................................................................3
1.1 About this manual .........................................................................................................................3
1.2 Transmitter model code ................................................................................................................3
1.3 Communications tools and protocols ............................................................................................3
1.4 Additional documentation and resources ......................................................................................4
Chapter 2 Quick start .......................................................................................................................5
2.1 Power up the transmitter ...............................................................................................................5
2.2 Check flowmeter status .................................................................................................................5
2.3 Make a startup connection to the transmitter ................................................................................6
2.4 Characterize the flowmeter (if required) ........................................................................................7
2.4.1 Sample sensor tags .........................................................................................................8
2.4.2 Flow calibration parameters (FCF, FT) .............................................................................9
2.4.3 Density calibration parameters (D1, D2, K1, K2, FD, DT, TC) .........................................10
2.5 Verify mass flow measurement ....................................................................................................10
2.6 Verify the zero .............................................................................................................................11
2.6.1 Verify the zero using ProLink II ......................................................................................11
2.6.2 Terminology used with zero verification and zero calibration ........................................12
Part II Configuration and commissioning
Chapter 3 Introduction to configuration and commissioning .........................................................17
3.1 Configuration flowchart ..............................................................................................................17
3.2 Default values and ranges ............................................................................................................18
3.3 Enable access to the off-line menu of the display .........................................................................19
3.4 Disable write-protection on the transmitter configuration ..........................................................19
3.5 HART security ..............................................................................................................................19
3.6 Restore the factory configuration ................................................................................................20
Chapter 4 Configure process measurement ...................................................................................21
4.1 Configure mass flow measurement .............................................................................................21
4.1.1 Configure Mass Flow Measurement Unit .......................................................................21
4.1.2 Configure Flow Damping ..............................................................................................24
4.1.3 Configure Mass Flow Cutoff ..........................................................................................25
4.2 Configure volume flow measurement for liquid applications .......................................................26
4.2.1 Configure Volume Flow Type for liquid applications ......................................................27
4.2.2 Configure Volume Flow Measurement Unit for liquid applications ................................27
4.2.3 Configure Volume Flow Cutoff ......................................................................................30
4.3 Configure gas standard volume (GSV) flow measurement ...........................................................31
4.3.1 Configure Volume Flow Type for gas applications .........................................................32
4.3.2 Configure Standard Gas Density ....................................................................................32
4.3.3 Configure Gas Standard Volume Flow Measurement Unit .............................................33
4.3.4 Configure Gas Standard Volume Flow Cutoff ................................................................36
Configuration and Use Manual i
Contents
4.4 Configure Flow Direction .............................................................................................................37
4.4.1 Options for Flow Direction ............................................................................................38
4.5 Configure density measurement ................................................................................................. 42
4.5.1 Configure Density Measurement Unit ...........................................................................43
4.5.2 Configure slug flow parameters ....................................................................................44
4.5.3 Configure Density Damping ..........................................................................................45
4.5.4 Configure Density Cutoff .............................................................................................. 46
4.6 Configure temperature measurement .........................................................................................47
4.6.1 Configure Temperature Measurement Unit .................................................................. 47
4.6.2 Configure Temperature Damping .................................................................................48
4.7 Configure the petroleum measurement application ....................................................................49
4.7.1 Configure petroleum measurement using ProLink II ..................................................... 49
4.7.2 Configure petroleum measurement using the Field Communicator ..............................50
4.7.3 API reference tables ......................................................................................................51
4.8 Configure the concentration measurement application .............................................................. 53
4.8.1 Configure concentration measurement using ProLink II ................................................53
4.8.2 Configure concentration measurement using the Field Communicator ........................55
4.8.3 Standard matrices for the concentration measurement application ..............................57
4.8.4 Derived variables and calculated process variables ........................................................58
4.9 Configure pressure compensation ...............................................................................................60
4.9.1 Configure pressure compensation using ProLink II ........................................................60
4.9.2 Configure pressure compensation using the Field Communicator ................................ 62
4.9.3 Options for Pressure Measurement Unit ....................................................................... 63
Chapter 5 Configure device options and preferences .....................................................................65
5.1 Configure the transmitter display ................................................................................................65
5.1.1 Configure the language used for the display ................................................................. 65
5.1.2 Configure the process variables shown on the display ...................................................66
5.1.3 Configure the precision of variables shown on the display .............................................67
5.1.4 Configure the refresh rate of data shown on the display ................................................68
5.1.5 Enable or disable automatic scrolling through the display variables .............................. 68
5.1.6 Enable or disable the display backlight ..........................................................................69
5.1.7 Enable or disable Status LED Blinking ............................................................................ 69
5.2 Enable or disable operator actions from the display .....................................................................69
5.2.1 Enable or disable Totalizer Start/Stop from the display ..................................................70
5.2.2 Enable or disable Totalizer Reset from the display .........................................................70
5.2.3 Enable or disable the Acknowledge All Alarms display command ..................................71
5.3 Configure security for the display menus ....................................................................................72
5.4 Configure response time parameters .......................................................................................... 73
5.4.1 Configure Update Rate ................................................................................................. 73
5.5 Configure alarm handling ............................................................................................................75
5.5.1 Configure Fault Timeout ...............................................................................................75
5.5.2 Configure Status Alarm Severity ....................................................................................76
5.6 Configure informational parameters ........................................................................................... 79
5.6.1 Configure Descriptor .................................................................................................... 79
5.6.2 Configure Message .......................................................................................................79
5.6.3 Configure Date ............................................................................................................. 80
5.6.4 Configure Sensor Serial Number ................................................................................... 80
5.6.5 Configure Sensor Material .............................................................................................81
5.6.6 Configure Sensor Liner Material .................................................................................... 81
5.6.7 Configure Sensor Flange Type .......................................................................................81
Chapter 6 Integrate the meter with the control system ..................................................................83
ii Micro Motion® 9739 MVD Transmitters
Contents
6.1 Configure the mA output ............................................................................................................83
6.1.1 Configure mA Output Process Variable .........................................................................84
6.1.2 Configure Lower Range Value (LRV) and Upper Range Value (URV) ...............................86
6.1.3 Configure AO Cutoff .....................................................................................................87
6.1.4 Configure Added Damping ...........................................................................................89
6.1.5 Configure mA Output Fault Action and mA Output Fault Level ......................................90
6.2 Configure the frequency output ..................................................................................................91
6.2.1 Configure Frequency Output Power Source ..................................................................92
6.2.2 Configure Frequency Output Process Variable ..............................................................92
6.2.3 Configure Frequency Output Polarity ............................................................................93
6.2.4 Configure Frequency Output Scaling Method ...............................................................94
6.2.5 Configure Frequency Output Maximum Pulse Width ....................................................95
6.2.6 Configure Frequency Output Fault Action and Frequency Output Fault Level ................96
6.3 Configure the discrete output .....................................................................................................97
6.3.1 Configure Discrete Output Power Source ......................................................................98
6.3.2 Configure Discrete Output Source ................................................................................98
6.3.3 Configure Discrete Output Polarity .............................................................................100
6.3.4 Configure Discrete Output Fault Action ......................................................................102
6.4 Configure the discrete input ......................................................................................................103
6.4.1 Configure Discrete Input Action ..................................................................................104
6.4.2 Configure Discrete Input Polarity ................................................................................105
6.5 Configure the mA input .............................................................................................................106
6.5.1 Configure mA Input Process Variable ..........................................................................107
6.5.2 Configure Lower Range Value (LRV) and Upper Range Value (URV) .............................107
6.6 Configure events .......................................................................................................................108
6.6.1 Configure a basic event ...............................................................................................108
6.6.2 Configure an enhanced event .....................................................................................109
6.7 Configure digital communications ............................................................................................111
6.7.1 Configure HART/Bell 202 communications .................................................................112
6.7.2 Configure HART/RS-485 communications ..................................................................116
6.7.3 Configure Modbus/RS-485 communications ..............................................................117
6.7.4 Configure Digital Communications Fault Action .........................................................119
6.8 Set up polling for temperature ..................................................................................................120
6.9 Set up polling for pressure .........................................................................................................122
Chapter 7 Completing the configuration ......................................................................................125
7.1 Back up transmitter configuration .............................................................................................125
7.2 Enable/disable HART security ....................................................................................................125
7.3 Enable write-protection on the transmitter configuration .........................................................126
Part III Operations, maintenance, and troubleshooting
Chapter 8 Transmitter operation .................................................................................................131
8.1 Record the process variables .....................................................................................................131
8.2 View transmitter status using the status LED .............................................................................131
8.3 View and acknowledge status alarms ........................................................................................132
8.3.1 View and acknowledge alarms using the display .........................................................132
8.3.2 View and acknowledge alarms using ProLink II ............................................................135
8.3.3 View alarms using the Field Communicator ................................................................135
8.3.4 Alarm data in transmitter memory ..............................................................................136
8.4 Read totalizer and inventory values ...........................................................................................137
Configuration and Use Manual iii
Contents
8.5 Start and stop totalizers and inventories ....................................................................................137
8.5.1 Start and stop totalizers and inventories using the display ..........................................138
8.6 Reset totalizers ..........................................................................................................................139
8.6.1 Reset totalizers using the display ................................................................................139
8.7 Reset inventories .......................................................................................................................140
Chapter 9 Measurement support .................................................................................................143
9.1 Options for measurement support ............................................................................................143
9.2 Zero the flowmeter ...................................................................................................................143
9.2.1 Zero the flowmeter using the display ..........................................................................144
9.2.2 Zero the flowmeter using ProLink II .............................................................................145
9.2.3 Zero the flowmeter using the Field Communicator .....................................................146
9.3 Validate the meter .....................................................................................................................147
9.3.1 Alternate method for calculating the meter factor for volume flow .............................149
9.4 Perform a (standard) D1 and D2 density calibration ...................................................................150
9.4.1 Perform a D1 and D2 density calibration using ProLink II .............................................150
9.4.2 Perform a D1 and D2 density calibration using the Field Communicator .....................151
9.5 Perform a D3 and D4 density calibration (T-Series sensors only) ................................................153
9.5.1 Perform a D3 or D3 and D4 density calibration using ProLink II ....................................154
9.5.2 Perform a D3 or D3 and D4 density calibration using the Field Communicator ............155
9.6 Perform temperature calibration ...............................................................................................157
9.6.1 Perform temperature calibration using ProLink II ........................................................157
Chapter 10 Troubleshooting ..........................................................................................................159
10.1 Status LED states .......................................................................................................................159
10.2 Status alarms .............................................................................................................................160
10.3 Flow measurement problems ....................................................................................................166
10.4 Density measurement problems ...............................................................................................168
10.5 Temperature measurement problems .......................................................................................169
10.6 Milliamp output problems .........................................................................................................170
10.7 Frequency output problems ......................................................................................................171
10.8 Use sensor simulation for troubleshooting ................................................................................172
10.9 Check power supply wiring ........................................................................................................173
10.10 Check sensor-to-transmitter wiring ...........................................................................................174
10.11 Check grounding .......................................................................................................................174
10.12 Perform loop tests .....................................................................................................................174
10.12.1 Perform loop tests using the display ...........................................................................175
10.12.2 Perform loop tests using ProLink II ..............................................................................176
10.12.3 Perform loop tests using the Field Communicator ......................................................178
10.13 Trim mA outputs .......................................................................................................................179
10.13.1 Trim mA outputs using ProLink II ................................................................................180
10.13.2 Trim mA outputs using the Field Communicator .........................................................180
10.14 Check the HART communication loop .......................................................................................181
10.15 Check HART Address and Loop Current Mode ............................................................................182
10.16 Check HART burst mode ............................................................................................................182
10.17 Check Lower Range Value and Upper Range Value ....................................................................182
10.18 Check mA Output Fault Action ..................................................................................................182
10.19 Check for radio frequency interference (RFI) ..............................................................................183
10.20 Check Frequency Output Maximum Pulse Width .......................................................................183
10.21 Check Frequency Output Scaling Method ..................................................................................183
10.22 Check Frequency Output Fault Action .......................................................................................184
10.23 Check Flow Direction .................................................................................................................184
10.24 Check the cutoffs ......................................................................................................................184
iv Micro Motion® 9739 MVD Transmitters
Contents
10.25 Check for slug flow (two-phase flow) .........................................................................................185
10.26 Check the drive gain ..................................................................................................................185
10.26.1 Collect drive gain data ................................................................................................186
10.27 Check the pickoff voltage ..........................................................................................................187
10.27.1 Collect pickoff voltage data ........................................................................................188
10.28 Check for electrical shorts ..........................................................................................................188
10.28.1 Check the sensor coils .................................................................................................188
Appendices and reference
Appendix A Using the transmitter display .......................................................................................191
A.1 Components of the transmitter interface ..................................................................................191
A.2 Use the optical switches ............................................................................................................193
A.3 Access and use the display menu system ...................................................................................194
A.3.1 Enter a floating-point value using the display ..............................................................195
A.4 Display codes for process variables ............................................................................................198
A.5 Codes and abbreviations used in display menus ........................................................................199
A.6 Menu maps for the transmitter display ......................................................................................202
Appendix B Using ProLink II with the transmitter ...........................................................................211
B.1 Basic information about ProLink II ..............................................................................................211
B.2 Menu maps for ProLink II ...........................................................................................................212
Appendix C Using the Field Communicator with the transmitter .....................................................217
C.1 Basic information about the Field Communicator ......................................................................217
C.2 Menu maps for the Field Communicator ....................................................................................218
Appendix D Default values and ranges ............................................................................................229
D.1 Default values and ranges ..........................................................................................................229
Appendix E Transmitter components and installation wiring .........................................................235
E.1 Transmitter components ...........................................................................................................235
E.2 Transmitter-to-sensor wiring .....................................................................................................236
E.3 Power supply terminals .............................................................................................................237
E.4 Input/output (I/O) terminals ......................................................................................................238
Index ................................................................................................................................................241
Configuration and Use Manual v
Contents
vi Micro Motion® 9739 MVD Transmitters
Part I

Getting Started

Chapters covered in this part:
Before you begin
Quick start
Getting Started
Configuration and Use Manual 1
Getting Started
2 Micro Motion® 9739 MVD Transmitters
1 Before you begin
Topics covered in this chapter:

About this manual

Transmitter model code

Communications tools and protocols

Additional documentation and resources
1.1 About this manual
This manual provides information to help you configure, commission, use, maintain, and troubleshoot the Micro Motion 9739 MVD transmitter.
Important
This manual assumes that the transmitter has been installed correctly and completely, according to the instructions in the transmitter installation manual, and that the installation complies with all applicable safety requirements.

Before you begin

1.2
1.3
Communications tools, protocols, and related informationTable 1-1:
Communica­tions tool Supported protocols Scope In this manual For more information
Display Not applicable Basic configuration and
ProLink II HART/RS-485
Transmitter model code
Your transmitter can be identified by the model number on the transmitter tag.
Communications tools and protocols
You may use different tools in different locations or for different tasks.
HART/Bell 202
Modbus/RS-485
Service port
commissioning
Complete configuration and commissioning
Complete user informa­tion. See Appendix A.
Basic user information. See Appendix B.
Not applicable
User manual
Installed with soft-
ware
On Micro Motion
user documentation CD
On Micro Motion
web site (www.mi-
cromotion.com
Configuration and Use Manual 3
Before you begin
Communications tools, protocols, and related information (continued)Table 1-1:
Communica­tions tool Supported protocols Scope In this manual For more information
Field Commu­nicator
HART/Bell 202 Complete configuration
and commissioning
Tip
You may be able to use other communications tools from Emerson Process Management, such as AMS Suite: Intelligent Device Manager, or the Smart Wireless THUM™ Adapter. Use of AMS or the Smart Wireless THUM Adapter is not discussed in this manual. The AMS interface is similar to the ProLink II interface. For more information on the Smart Wireless THUM Adapter, refer to the documentation available at www.micromotion.com.
Basic user information. See Appendix C.
User manual on Micro Motion web site (www.micromo-
tion.com
1.4 Additional documentation and resources
Micro Motion provides additional documentation to support the installation and operation of the transmitter.

Additional documentation and resourcesTable 1-2:

Topic Document
Sensor Sensor documentation
Transmitter installation
Hazardous area installation See the approval documentation shipped with the transmitter, or
Transmitter electronics mod­ule upgrade
Micro Motion 9739 MVD Transmitters: Installation Manual
download the appropriate documentation from the Micro Motion web site at www.micromotion.com.
Micro Motion 9739 MVD Transmitter Electronics Module Installation Guide
All documentation resources are available on the Micro Motion web site at
www.micromotion.com or on the Micro Motion user documentation CD.
4 Micro Motion® 9739 MVD Transmitters
2 Quick start
Topics covered in this chapter:

Power up the transmitter

Check flowmeter status

Make a startup connection to the transmitter
Characterize the flowmeter (if required)
Verify mass flow measurement
Verify the zero
2.1 Power up the transmitter
The transmitter must be powered up for all configuration and commissioning tasks, or for process measurement.
1. Ensure that all transmitter and sensor covers and seals are closed.

Quick start

2.2
CAUTION!
To prevent ignition of flammable or combustible atmospheres, ensure that all covers and seals are tightly closed. For hazardous area installations, applying power while housing covers are removed or loose can cause an explosion.
2. Turn on the electrical power at the power supply.
The transmitter will automatically perform diagnostic routines. During this period, Alarm 009 is active. The diagnostic routines should complete in approximately 30 seconds. For transmitters with a display, the status LED will turn green and begin to flash when the startup diagnostics are complete. If the status LED exhibits different behavior, an alarm condition is present.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if power is been off long enough to allow components to reach ambient temperature, allow the electronics to warm up for approximately 10 minutes before relying on process measurements. During this warm-up period, you may observe minor measurement instability or inaccuracy.
Check flowmeter status
Check the flowmeter for any error conditions that require user action or that affect measurement accuracy.
Configuration and Use Manual 5
Quick start
1. Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and checks for error conditions. During the power-up sequence, Alarm A009 is active. This alarm should clear automatically when the power-up sequence is complete.
2. Check the status LED on the transmitter.
Transmitter status reported by status LEDTable 2-1:
LED state Description Recommendation
Green No alarms are active. Continue with configuration or process meas-
urement.
Yellow One or more low-severity alarms are active,
and have been acknowledged.
Flashing yellow
Red One or more high-severity alarms are active,
(1)
One or more low-severity alarms are active and have not been acknowledged.
and have been acknowledged.
A low-severity alarm condition does not affect measurement accuracy or output behavior. You can continue with configuration or proc­ess measurement. If you choose, you can iden­tify and resolve the alarm condition.
A low-severity alarm condition does not affect measurement accuracy or output behavior. You can continue with configuration or proc­ess measurement. If you choose, you can iden­tify and resolve the alarm condition. You may also acknowledge the alarm.
A high-severity alarm condition affects meas­urement accuracy and output behavior. Re­solve the alarm condition before continuing.
Postrequisites
For information on viewing the list of active alarms, see Section 8.3.
For information on individual alarms and suggested resolutions, see Section 10.2.
2.3

Make a startup connection to the transmitter

For all configuration tools except the display, you must have an active connection to the transmitter to configure the transmitter. Follow this procedure to make your first connection to the transmitter.
Identify the connection type to use, and follow the instructions for that connection type in the appropriate appendix. Use the default communications parameters shown in the appendix.
Communications tool
ProLink II Modbus/RS-485 Appendix B
(1) If Status LED Blinking is disabled, the LED will show solid yellow rather than flashing.
Connection type to use Instructions
6 Micro Motion® 9739 MVD Transmitters
Quick start
Communications tool Connection type to use Instructions
Field Communicator HART Appendix C
Postrequisites
(Optional) Change the communications parameters to site-specific values.
To change the communications parameters using ProLink II:
To change the protocol, baud rate, parity, or stop bits, choose ProLink > Configuration >
RS-485.
To change the address, choose ProLink > Configuration > Device.
To change the communications parameters using the Field Communicator, choose On-Line Menu > Configure > Manual Setup > Inputs/Outputs > Communications.
Important
If you are changing communications parameters for the connection type that you are using, you will lose the connection when you write the parameters to the transmitter. Reconnect using the new parameters.

2.4 Characterize the flowmeter (if required)

Display
ProLink II ProLink > Configuration > Density ProLink > Configuration > Flow
Field Communicator Configure > Manual Setup > Characterize
Overview
Characterizing the flowmeter adjusts your transmitter to match the unique traits of the sensor it is paired with. The characterization parameters (also called calibration parameters) describe the sensor’s sensitivity to flow, density, and temperature. Depending on your sensor type, different parameters are required. Values for your sensor are provided by Micro Motion on the sensor tag or the calibration certificate.
Tip
If your flowmeter was ordered as a unit, it has already been characterized at the factory. However, you should still verify the characterization parameters.
Procedure
1. Specify Sensor Type.
Not available
Straight-tube (T-Series)
Curved-tube (all sensors except T-Series)
Configuration and Use Manual 7
Quick start
2. Set the flow characterization parameters. Be sure to include all decimal points.
For straight-tube sensors, set FCF (Flow Cal or Flow Calibration Factor), FTG, and FFQ.
For curved-tube sensors, set Flow Cal (Flow Calibration Factor).
3. Set the density characterization parameters.
For straight-tube sensors, set D1, D2, DT, DTG, K1, K2, FD, DFQ1, and DFQ2.
For curved-tube sensors, set D1, D2, TC, K1, K2, and FD. (TC is sometimes shown
as DT.)
2.4.1 Sample sensor tags
Tag on older curved-tube sensors (all sensors except T-Series)Figure 2-1:
Tag on newer curved-tube sensors (all sensors except T-Series)Figure 2-2:
8 Micro Motion® 9739 MVD Transmitters
Quick start
Tag on older straight-tube sensor (T-Series)Figure 2-3:
Tag on newer straight-tube sensor (T-Series)Figure 2-4:
2.4.2 Flow calibration parameters (FCF, FT)
Two separate values are used to describe flow calibration: a 6-character FCF value and a 4­character FT value. They are provided on the sensor tag.
Both values contain decimal points. During characterization, these may be entered as two values or as a single 10-character string. The 10-character string is called either Flowcal or FCF.
If your sensor tag shows the FCF and the FT values separately and you need to enter a single value, concatenate the two values to form the single parameter value.
If your sensor tag shows a concatenated Flowcal or FCF value and you need to enter the FCF and the FT values separately, split the concatenated value:
FCF = The first 6 characters, including the decimal point
FT = The last 4 characters, including the decimal point
Configuration and Use Manual 9
Quick start
Example: Concatenating FCF and FT
FCF = x.xxxx FT = y.yy Flow calibration parameter: x.xxxxy.yy
Example: Splitting the concatenated Flowcal or FCF value
Flow calibration parameter: x.xxxxy.yy FCF = x.xxxx FT = y.yy
2.4.3 Density calibration parameters (D1, D2, K1, K2, FD, DT, TC)
Density calibration parameters are typically on the sensor tag and the calibration certificate.
If your sensor tag does not show a D1 or D2 value:
For D1, enter the Dens A or D1 value from the calibration certificate. This value is the
line-condition density of the low-density calibration fluid. Micro Motion uses air. If you cannot find a Dens A or D1 value, enter 0.001 g/cm3.
2.5
For D2, enter the Dens B or D2 value from the calibration certificate. This value is the
line-condition density of the high-density calibration fluid. Micro Motion uses water. If you cannot find a Dens B or D2 value, enter 0.998 g/cm3.
If your sensor tag does not show a K1 or K2 value:
For K1, enter the first 5 digits of the density calibration factor. In the sample tag, this
value is shown as 12500.
For K2, enter the second 5 digits of the density calibration factor. In the sample tag,
this value is shown as 14286.
If your sensor does not show an FD value, contact Micro Motion customer service.
If your sensor tag does not show a DT or TC value, enter the last 3 digits of the density calibration factor. In the sample tag, this value is shown as 4.44.

Verify mass flow measurement

Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available method.
Read the value for Mass Flow Rate on the transmitter display.
Connect to the transmitter with ProLink II and read the value for Mass Flow Rate in the
Process Variables window (ProLink > Process Variables).
Connect to the transmitter with the Field Communicator and read the value for Mass
Flow Rate in the Process Variables menu (On-Line Menu > Overview > Primary Purpose Variables).
10 Micro Motion® 9739 MVD Transmitters
Postrequisites
If the reported mass flow rate is not accurate:
Check the characterization parameters.
Review the troubleshooting suggestions for flow measurement issues. See
Section 10.3.

2.6 Verify the zero

Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field zero can improve measurement accuracy.
The zero verification procedure analyzes the Live Zero value under conditions of zero flow, and compares it to the Zero Stability range for the sensor. If the average Live Zero value is within a reasonable range, the zero value stored in the transmitter is valid. Performing a field calibration will not improve measurement accuracy.
2.6.1 Verify the zero using ProLink II
Quick start
Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field zero can improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the flowmeter unless one of the following is true:
The zero is required by site procedures.
The stored zero value fails the zero verification procedure.
Prerequisites
ProLink II v2.94 or later
Important
Do not verify the zero or zero the flowmeter if a high-severity alarm is active. Correct the problem, then verify the zero or zero the flowmeter. You may verify the zero or zero the flowmeter if a low­severity alarm is active.
Procedure
1. Prepare the flowmeter:
a. Allow the flowmeter to warm up for at least 20 minutes after applying power.
b. Run the process fluid through the sensor until the sensor temperature reaches
the normal process operating temperature.
c. Stop flow through the sensor by shutting the downstream valve, and then the
upstream valve if available.
Configuration and Use Manual 11
Quick start
d. Verify that the sensor is blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
2. Choose ProLink > Calibration > Zero Verification and Calibration > Verify Zero and wait until
the procedure completes.
3. If the zero verification procedure fails:
a. Confirm that the sensor is completely blocked in, that flow has stopped, and that
the sensor is completely full of process fluid.
b. Verify that the process fluid is not flashing or condensing, and that it does not
contain particles that can settle out.
c. Repeat the zero verification procedure.
d. If it fails again, zero the flowmeter.
For instructions on zeroing the flowmeter, see Zero the flowmeter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
2.6.2 Terminology used with zero verification and zero calibration
Terminology used with zero verification and zero calibrationTable 2-2:
Term Definition
Zero In general, the offset required to synchronize the left pickoff and the right pickoff under
conditions of zero flow. Unit = microseconds.
Factory Zero The zero value obtained at the factory, under laboratory conditions.
Field Zero The zero value obtained by performing a zero calibration outside the factory.
Prior Zero The zero value stored in the transmitter at the time a field zero calibration is begun. May
be the factory zero or a previous field zero.
Manual Zero The zero value stored in the transmitter, typically obtained from a zero calibration proce-
dure. It may also be configured manually. Also called “mechanical zero” or “stored zero.”
Live Zero The real-time bidirectional mass flow rate with no flow damping or mass flow cutoff ap-
plied. An adaptive damping value is applied only when the mass flow rate changes dra­matically over a very short interval. Unit = configured mass flow measurement unit.
Zero Stability A laboratory-derived value used to calculate the expected accuracy for a sensor. Under
laboratory conditions at zero flow, the average flow rate is expected to fall within the range defined by the Zero Stability value (0 ± Zero Stability). Each sensor size and model has a unique Zero Stability value. Statistically, 95% of all data points should fall within the range defined by the Zero Stability value.
Zero Calibration The procedure used to determine the zero value.
Zero Time The time period over which the Zero Calibration procedure is performed. Unit = seconds.
Field Verification Zero A 3-minute running average of the Live Zero value, calculated by the transmitter. Unit =
configured mass flow measurement unit.
12 Micro Motion® 9739 MVD Transmitters
Quick start
Terminology used with zero verification and zero calibration (continued)Table 2-2:
Term Definition
Zero Verification A procedure used to evaluate the stored zero and determine whether or not a field zero
can improve measurement accuracy.
Configuration and Use Manual 13
Quick start
14 Micro Motion® 9739 MVD Transmitters

Configuration and commissioning

Part II
Configuration and commissioning
Chapters covered in this part:
Introduction to configuration and commissioning
Configure process measurement
Configure device options and preferences
Integrate the meter with the control system
Completing the configuration
Configuration and Use Manual 15
Configuration and commissioning
16 Micro Motion® 9739 MVD Transmitters

Introduction to configuration and commissioning

3 Introduction to configuration and
commissioning
Topics covered in this chapter:

Configuration flowchart

Default values and ranges
Enable access to the off-line menu of the display
Disable write-protection on the transmitter configuration
HART security
Restore the factory configuration
3.1 Configuration flowchart
Use the following flowchart as a general guide to the configuration and commissioning process.
Some options may not apply to your installation. Detailed information is provided in the remainder of this manual. If you are using the Weights & Measures application, additional configuration and setup are required.
Configuration and Use Manual 17
Introduction to configuration and commissioning
Configuration flowchartFigure 3-1:
Configure process measurement
Configure mass flow
measurement
Configure volume flow
meaurement
Configure device options and preferences
Configure display
parameters
Configure fault handling
parameters
Test and move to production
Test or tune transmitter using sensor simulation
Back up transmitter
configuration
Volume flow type
Liquid
Configure flow direction
Configure density
measurement
Configure temperature
measurement
Configure petroleum
measurement (API)
application (if available)
Configure concentration
measurement application
(if available)
Configure pressure
compensation (optional)
Gas
Define gas properties
Configure sensor
parameters
Configure device
parameters
Integrate device with control system
Configure the mA
output(s)
Configure the frequency
output(s)
Configure the discrete
output(s)
Configure the discrete
input
Configure the mA input
Enable write-protection on
transmitter configuration
Done
Configure events
Configure digital communications

3.2 Default values and ranges

See Section D.1 to view the default values and ranges for the most commonly used parameters.
18 Micro Motion® 9739 MVD Transmitters
Introduction to configuration and commissioning

3.3 Enable access to the off-line menu of the display

Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > OFFLN
ProLink II ProLink > Configuration > Display > Display Options > Display Offline Menu
Field Communicator
Overview
By default, access to the off-line menu of the display is enabled. If it is disabled, you must enable it if you want to use the display to configure the transmitter.
Restriction
You cannot use the display to enable access to the off-line menu. You must make a connection from another tool.
Not available

3.4 Disable write-protection on the transmitter configuration

Display OFF-LINE MAINT > CONFG > LOCK
ProLink II ProLink > Configuration > Device > Enable Write Protection
Not available
3.5
Field Communicator
Overview
If the transmitter is write-protected, the configuration is locked and you must unlock it before you can change any configuration parameters. By default, the transmitter is not write-protected.
Tip
Write-protecting the transmitter prevents accidental changes to configuration. It does not prevent normal operational use. You can always disable write-protection, perform any required configuration changes, then re-enable write-protection.

HART security

HART security may be enabled on your transmitter. To configure the transmitter using HART protocol, you must disable HART security.
Configuration and Use Manual 19
Introduction to configuration and commissioning

3.6 Restore the factory configuration

Display
ProLink II ProLink > Configuration > Device > Restore Factory Configuration
Field Communicator
Not available
Not available
Overview
Restoring the factory configuration returns the transmitter to a known operational configuration. This may be useful if you experience problems during configuration.
Tip
Restoring the factory configuration is not a common action. You may want to contact Micro Motion to see if there is a preferred method to resolve any issues.
20 Micro Motion® 9739 MVD Transmitters

Configure process measurement

4 Configure process measurement
Topics covered in this chapter:

Configure mass flow measurement

Configure volume flow measurement for liquid applications
Configure gas standard volume (GSV) flow measurement
Configure Flow Direction
Configure density measurement
Configure temperature measurement
Configure the petroleum measurement application
Configure the concentration measurement application
Configure pressure compensation
4.1 Configure mass flow measurement
The mass flow measurement parameters control how mass flow is measured and reported.
The mass flow measurement parameters include:
Mass Flow Measurement Unit
Flow Damping
Mass Flow Cutoff
4.1.1
Configure Mass Flow Measurement Unit
Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > MASS
ProLink II ProLink > Configuration > Flow > Mass Flow Units
Field Communicator Configure > Manual Setup > Measurements > Flow > Mass Flow Unit
Overview
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass flow rate. The unit used for mass total and mass inventory is derived from this unit.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
The default setting for Mass Flow Measurement Unit is g/sec (grams per second).
Configuration and Use Manual 21
Configure process measurement
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit, plus one user-defined special measurement unit. Different communications tools may
use different labels for the units.
Options for Mass Flow Measurement UnitTable 4-1:
Unit description
Grams per second
Grams per minute
Grams per hour
Kilograms per second
Kilograms per minute
Kilograms per hour
Kilograms per day
Metric tons per minute
Metric tons per hour
Metric tons per day
Pounds per second
Pounds per minute
Pounds per hour
Pounds per day
Short tons (2000 pounds) per minute
Short tons (2000 pounds) per hour
Short tons (2000 pounds) per day
Long tons (2240 pounds) per hour
Long tons (2240 pounds) per day
Special unit
Label
Display ProLink II ProLink III Field Communi-
cator
G/S g/sec g/sec g/s
G/MIN g/min g/min g/min
G/H g/hr g/hr g/h
KG/S kg/sec kg/sec kg/s
KG/MIN kg/min kg/min kg/min
KG/H kg/hr kg/hr kg/h
KG/D kg/day kg/day kg/d
T/MIN mTon/min mTon/min MetTon/min
T/H mTon/hr mTon/hr MetTon/h
T/D mTon/day mTon/day MetTon/d
LB/S lbs/sec lbs/sec lb/s
LB/MIN lbs/min lbs/min lb/min
LB/H lbs/hr lbs/hr lb/h
LB/D lbs/day lbs/day lb/d
ST/MIN sTon/min sTon/min STon/min
ST/H sTon/hr sTon/hr STon/h
ST/D sTon/day sTon/day STon/d
LT/H lTon/hr lTon/hr LTon/h
LT/D lTon/day lTon/day LTon/d
SPECL special special Spcl
22 Micro Motion® 9739 MVD Transmitters
Loading...
+ 226 hidden pages