Oxygen Analyzers
XZR200
Oxygen Analyzer
A cost effective zirconium-dioxide analyzer to measure percentage level oxygen in combustion 
processes, ambient air monitoring and many more applications. The unit is configurable to 
measure either 0-25% or 0-100% oxygen and offers manual or automatic calibration to suit the 
customer’s needs. Two probe lengths are available (210mm & 400mm) as well as two maximum 
sample temperatures (250°C & 400°C) for greater flexibility.
Highlights
• Configurable outputs: 4-20 mA and 0 to 10 V DC or 
RS232 comms interface
• Cycling 3.3 V DC logic output allows direct monitoring 
of the O
• Can be calibrated in normal air (20.7% O 
other known O
• Selectable output filtering allows fast and dynamic or 
slow and stable output
• Externally triggered automatic or manual calibration
• Diecast aluminium case IP65 with stainless steel 
probe
• Sample temperature up to +400°C
www.michell.com
 sensor for diagnostic purposes
2
 concentration
2
) or in any 
2
Applications
• Combustion control including oil, gas and biomass 
boiler applications
• Laboratory & building air quality monitoring including 
confined space personnel safety 
• Composting
• Scientific including respiratory studies of a 
community or an organism, plants and animals
• Food and beverage packaging

Oxygen Analyzers 
Technical Specifications
Performance
Measurement technology Zirconium Dioxide 
Gas  Oxygen 
Measurement range 0-25% or 0-100% 
Output resolution 0.01 V, 0.01 mA or 0.01% O 
Accuracy (0-25%) 1% of reading < 0.5% O
2
2
Accuracy (0-100%) < 1% of reading 
Response time (T90) < 5 seconds 
Repeatability  < 0.5%  
Sample flow rate  0 to 1.7 m/sec 
Sample flow effect ±0.5% of full scale 
Sample pressure Atmospheric 
Sample temperature +250°C or +400°C 
Sample cell temperature +700°C 
Background gas Air, N
, CO2, Ar or He
2
Warning: Probe tip gets hot, do not touch without PPE
Dimensions
Electrical Input/Output
Power supply 24 V DC, ±10% 
Power consumption 500 mA maximum @ 24 V DC 
Analog outputs 4-20 mA and 0 to 10 V DC 
Output ranges 0-25% or 0-100%
Digital communications RS232 (not available if 4-20 mA output 
selected)
Operating Conditions
Ambient temperature -10 to +85°C
Mechanical Specifications
Warm Up time Approx. 10 minutes 
Stabilization time Included in the above 
Dimensions 52 x 75 x 80mm (h x w x d) excluding probe 
Probe dimensions 210 or 400mm (length) ø12mm 
Weight < 0.5kg 
Wetted materials Stainless steel 
Process connection 12mm Swagelok 
Ingress protection IP65 
Housing material Waterproof die-cast aluminium housing
®
 connector
75mm
52mm
62.60mm
80mm
52.40mm
210mm or 400mm
Top View
Side View
Mounting Holes ø4.5mm (x2) 
Mounting holes accessible on removal of housing lid
Bottom View
ø12mm
Bar Probe
O-Ring
Weld Socket
52mm
Partition
Bar Probe
Nut
Housing
O-Ring
Partition
Nut
Weld Socket
Housin
Atmospheric Air 
+60°C Max
15°
Partition
Gas Mixture 
+250°C Max 
+400°C Max
Michell Instruments Ltd 48 Lancaster Way Business Park, Ely, Cambridgeshire, CB6 3NW 
Tel: +44 (0) 1353 658000, Fax: +44 (0) 1353 658199, Email: uk.info@michell.com, Web: 
Michell Instruments adopts a continuous development programme which sometimes necessitates specification changes without notice. 
Issue no: XZR200_97338_V1_UK_0613
www.michell.com/uk