Michell Instruments XZR200 Specifications

Oxygen Analyzers
XZR200
Oxygen Analyzer
A cost effective zirconium-dioxide analyzer to measure percentage level oxygen in combustion processes, ambient air monitoring and many more applications. The unit is configurable to measure either 0-25% or 0-100% oxygen and offers manual or automatic calibration to suit the customer’s needs. Two probe lengths are available (210mm & 400mm) as well as two maximum sample temperatures (250°C & 400°C) for greater flexibility.
Highlights
• Configurable outputs: 4-20 mA and 0 to 10 V DC or RS232 comms interface
• Cycling 3.3 V DC logic output allows direct monitoring of the O
• Can be calibrated in normal air (20.7% O other known O
• Selectable output filtering allows fast and dynamic or slow and stable output
• Externally triggered automatic or manual calibration
• Diecast aluminium case IP65 with stainless steel probe
• Sample temperature up to +400°C
www.michell.com
sensor for diagnostic purposes
2
concentration
2
) or in any
2
Applications
• Combustion control including oil, gas and biomass boiler applications
• Laboratory & building air quality monitoring including confined space personnel safety
• Composting
• Scientific including respiratory studies of a community or an organism, plants and animals
• Food and beverage packaging
Oxygen Analyzers
g
Technical Specifications
Performance
Measurement technology Zirconium Dioxide Gas Oxygen Measurement range 0-25% or 0-100% Output resolution 0.01 V, 0.01 mA or 0.01% O Accuracy (0-25%) 1% of reading < 0.5% O
2
2
Accuracy (0-100%) < 1% of reading Response time (T90) < 5 seconds Repeatability < 0.5% Sample flow rate 0 to 1.7 m/sec Sample flow effect ±0.5% of full scale Sample pressure Atmospheric Sample temperature +250°C or +400°C Sample cell temperature +700°C Background gas Air, N
, CO2, Ar or He
2
Warning: Probe tip gets hot, do not touch without PPE
Dimensions
Electrical Input/Output
Power supply 24 V DC, ±10% Power consumption 500 mA maximum @ 24 V DC Analog outputs 4-20 mA and 0 to 10 V DC Output ranges 0-25% or 0-100%
Digital communications RS232 (not available if 4-20 mA output
selected)
Operating Conditions
Ambient temperature -10 to +85°C
Mechanical Specifications
Warm Up time Approx. 10 minutes Stabilization time Included in the above Dimensions 52 x 75 x 80mm (h x w x d) excluding probe Probe dimensions 210 or 400mm (length) ø12mm Weight < 0.5kg Wetted materials Stainless steel Process connection 12mm Swagelok Ingress protection IP65 Housing material Waterproof die-cast aluminium housing
®
connector
75mm
52mm
62.60mm
80mm
52.40mm
210mm or 400mm
Top View
Side View
Mounting Holes ø4.5mm (x2) Mounting holes accessible on removal of housing lid
Bottom View
ø12mm
Bar Probe
O-Ring
Weld Socket
52mm
Partition
Bar Probe
Nut
Housing
O-Ring
Partition
Nut
Weld Socket
Housin
Atmospheric Air +60°C Max
15°
Partition
Gas Mixture +250°C Max +400°C Max
Michell Instruments Ltd 48 Lancaster Way Business Park, Ely, Cambridgeshire, CB6 3NW Tel: +44 (0) 1353 658000, Fax: +44 (0) 1353 658199, Email: uk.info@michell.com, Web:
Michell Instruments adopts a continuous development programme which sometimes necessitates specification changes without notice. Issue no: XZR200_97338_V1_UK_0613
www.michell.com/uk
Recommended Installation
© 2013 Michell Instruments