Appendix D Return Document & Decontamination Declaration ........................................24
Michell Instrumentsv
VDS3 User’s Manual
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures
detailed in this manual. The user must not use this equipment for any other purpose than that
stated. Do not apply values greater than the maximum value stated.
This manual contains operating and safety instructions, which must be followed to ensure the safe
operation and to maintain the equipment in a safe condition. The safety instructions are either
warnings or cautions issued to protect the user and the equipment from injury or damage. Use
qualifi ed personnel and good engineering practice for all procedures in this manual.
Electrical Safety
The instrument is designed to be completely safe when used with options and accessories supplied
by the manufacturer for use with the instrument. The input power supply voltage limits are 220 to
240 V AC OR 100 to 120 V AC 50/60Hz.
Pressure Safety
DO NOT permit pressures greater than the safe working pressure to be applied to the instrument.
The maximum safe inlet pressure is 4.8 barg (70 psig).
Toxic Materials
The use of hazardous materials in the construction of this instrument has been minimized. During
normal operation it is not possible for the user to come into contact with any hazardous substance
which might be employed in the construction of the instrument. Care should, however, be exercised
during maintenance and the disposal of certain parts.
Repair and Maintenance
The instrument must be maintained either by the manufacturer or an accredited service agent. Refer
to www.michell.com for details of Michell Instruments’ worldwide offi ces contact information.
Safety Conformity
This product meets the essential protection requirements of the relevant EU directives.
vi97433 Issue 1.2, July 2016
VDS3 User’s Manual
!
Abbreviations
The following abbreviations are used in this manual:
AC alternating current
atm pressure unit (atmosphere)
barg pressure unit (=100 kP or 0.987 atm) (gauge)
°C degrees Celsius
°F degrees Fahrenheit
dp dew point
EU European Union
Hz Hertz
IEC International Electrotechnical Commission
Nl/min normal liters per minute
mm millimeters
lb pound
psig pound(s) per square inch (gauge)
ppb
scfh standard cubic feet per hour
USB Universal Serial Bus
V Volts
W Watts
" inches
Warnings
The following general warnings listed below are applicable to this instrument. They are
repeated in the text in the appropriate locations.
parts per billion (volume)
V
Where this hazard warning symbol appears in the following
sections, it is used to indicate areas where potentially hazardous
operations need to be carried out.
Michell Instrumentsvii
VDS3 User’s Manual
This page intentionally left blank
viii97433 Issue 1.2, July 2016
VDS3 User’s Manual
1 INTRODUCTION
1.1 Overview
The Michell VDS3 Vapor Delivery System is designed for use as part of a hygrometry
calibration system. It is capable of repeatable generation of set dew-point levels over
the range -100 to +20°C (-148 to +68°F) dew point.
The system is assembled within a small fl oor-mounted 19” equipment rack enclosure fi tted with castors for ease of movement. The system requires a supply of high quality
dry gas with a moisture content of better than 1 ppbV (-100°C (-148°F) atmospheric
dew point) to enable the maximum range of dew points to be generated. The supply of
high quality dry gas can be provided by a Michell PSD4 Pressure Swing Dryer.
The vapor delivery system divides the dry gas into three controllable gas streams. One
of these streams passes through a controlled evaporator mixer and is then mixed by
volumetric ratios with the other streams to produce different factory pre-set dew point
values over the range -100 to +20°C (-148 to +68°F) (user confi gurable). These dew
points are selected remotely via a PC.
INTRODUCTION
Michell Instruments1
INSTALLATION
2 INSTALLATION
2.1 Enclosure
The instrument enclosure is designed for fl oor mounting (19” equipment mini-rack).
Castors are fi tted for ease of movement.
Access at the front, left hand side and rear is required for operation. Front and side
panels are removable for maintenance & servicing.
2.2 Gas Connections
2.2.1 Dry Air Inlet
The system must be supplied with high quality (oil free) dry compressed air to the same
specifi cation as used during its calibration/setup (< 1ppbV (-100°C (148°F) moisture
content).
The generator was calibrated/set -up using compressed air supplied from a Michell PSD4
Pressure Swing Dryer.
VDS3 User’s Manual
The input gas pressure is further regulated and controlled by an factory set pressure
regulator, located internally. NOTE: This regulator MUST NOT be adjusted!
The gas inlet connection is a stainless steel ¼” Swagelok VCR face seal bulkhead union
fi tting located on the rear panel of the unit marked
2.2.2 Dew-Point Outlet
The gas outlet connection is a stainless steel ¼” Swagelok VCR face seal bulkhead
union fi tting located on the rear panel of the unit marked
2.2.3 Compressed Air
The system requires a separate supply of oil-free compressed air for operating pneumatic
valves and pressurizing the water supply. The compressed air supply can be a normal
industrial compressed air quality (i.e. 5 to 7 barg (72 to 101 psig) @ -40°Cdp (-40°Fdp).
The compressed air inlet connection is a stainless steel 6mm Swagelok
tube fi tting located on the rear panel of the unit marked COMPRESSED AIR INLET.
GAS INLET.
GAS OUTLET.
®
bulkhead union
GAS OUTLET
(1/4” VCR FACE
COMPRESSED AIR
INLET
(6mm TUBE FITTING)
Figure 1
g
g
g
REAR VIEW
Gas Connections
SEAL FITTING)
GAS INLET
(1/4” VCR FACE
SEAL FITTING)
297433 Issue 1.2, July 2016
VDS3 User’s Manual
2.3 Electrical Connections
2.3.1 Power Supply
100 to 120 V AC OR 220 to 240 V AC 50/60Hz power supply is required to operate this
instrument. Refer to the yellow label located on the rear panel of the unit for the correct
supply voltage required.
The power supply connection is via the 3-pin IEC fused plug located on the rear panel
of the unit. A 3 core power cable is provided, the free end of which should be wired to
a suitable earthed plug or directly via a fused power spur.
Power cable conductors are colored according to the convention:
Brown L (Live)
Blue N (Neutral)
Green/Yellow E (Earth/Ground)
INSTALLATION
2.3.2 PC Control Connection
Connect the host PC to the VDS via a USB connector located on the rear panel of the
unit marked
PC (CONTROL). A 3 meter USB cable assembly is provided and should be
connected as shown below:
VDS3
RS485 (SET UP)
PC (CONTROL)
S4000 Dewpointmeter
RS232
USB Cable
Serial Cable
Connect to spare
USB port
Host PC
g
g
Connect to spare
COMM Serial port
Figure 2
PC Control Connection
NOTE: Ensure that the S4000 Dewpointmeter is connected to the host PC to
enable the software to read and record dew points. If it is not connected the
software will not operate correctly.
Michell Instruments3
Loading...
+ 23 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.