Figure 11 Material Permeability Comparison ......................................................................... 9
Figure 12 Transmitter Mounting - Pipe or Duct ................................................................... 12
Figure 13 Transmitter Mounting with Adapter ..................................................................... 14
97576Issue 1, June 2019 ii
SF82 Transmitter User’s Manual
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures
detailed in this manual. The user must not use this equipment for any other purpose than that
stated. Do not apply values greater than the maximum value stated.
This manual contains operating and safety instructions, which must be followed to ensure the
safe operation and to maintain the equipment in a safe condition. The safety instructions are
either warnings or cautions issued to protect the user and the equipment from injury or damage.
Use competent personnel using good engineering practice for all procedures in this manual.
Electrical Safety
This instrument is designed to be electrically safe when used with the options and accessories
supplied by Michell Instruments for use with it. This instrument has been independently verified
as complying with the IEC/EN 61010 Standard for Electrical Safety for Europe and for the
equivalent 61010 standards in use in N. America. The instrument is approved for use within the
operating temperature range of -40°C to +60°C, and dependant on version, as being IP66/65.
See Specification section for full details.
Pressure Safety
DO NOT permit pressures greater than the safe working pressure to be applied to the instrument.
The specified safe working pressure is 45 MPag (450 barg / 6500 psig). Refer to the Technical
Specifications in Appendix A.
Toxic Materials
The use of hazardous materials in the construction of this instrument has been minimized. During
normal operation it is not possible for the user to come into contact with any hazardous substance
which might be employed in the construction of the instrument. Care should, however, be
exercised during maintenance and the disposal of certain parts.
Repair and Maintenance
The instrument must be maintained either by the manufacturer or an accredited service agent.
For Michell Instruments’ contact information please go to www.michell.com.
Calibration
The recommended calibration interval for this instrument is 12 months unless it is to be used in
a mission-critical application or in a dirty or contaminated environment in which case the
calibration interval should be reduced accordingly. The instrument should be returned to the
manufacturer, Michell Instruments Ltd., or one of their accredited service agents for recalibration.
iii Michell Instruments
SF82 Transmitter User’s Manual
Where this hazard symbol appears in the following
Safety Conformity
This product meets the essential protection requirements of the relevant EU and US standards
and directives.
Abbreviations
The following abbreviations are used in this manual:
barg pressure unit (=100 kP or 0.987 atm) (bar gauge)
ºC degrees Celsius
ºF degrees Fahrenheit
DC direct current
g grams
in inch(es)
µm micrometer
m/sec meters per second
mA milliampere
mm millimetres
MPa megapascal
Nl/min normal liters per minute
Nm Newton meter
oz ounces
psig pounds per square inch
RH relative humidity
scfh standard cubic feet per hour
fps feet per second
T temperature
V Volts
ΩOhms
ø diameter
Warnings
The following general warning listed below is applicable to this instrument. It is repeated in the
text in the appropriate locations.
sections it is used to indicate areas where potentially
hazardous operations need to be carried out.
97576 Issue 1, June 2019 iv
SF82 Transmitter User’s Manual
1. INTRODUCTION
The Michell Instruments SF82 is a loop-powered dew-point transmitter, designed to make dew
point measurements in a flowing sample. The SF82 transmitter is available with 3 different
process connections:
• 5/8” - 18 UNF
• 3/4” – 16 UNF
• G1/2” - BSPP
The SF82 2-wire is available with a choice of electrical connections:
• DIN 43650 Form C
• M12 5-pin
1 Michell Instruments
SF82 Transmitter User’s Manual
2. INSTALLATION
2.1. Unpacking the Instrument
On delivery, please check that all the following standard components are in the packing box:
• SF82 Transmitter
• Certificate of Calibration
• Connector (for sensor/cable) for MiniDIN 43650 C version only
It is recommended that all packaging is retained, in case products are returned for service or
calibration. Alternatively, if you choose to dispose of the packaging materials, ensure they are
recycled in accordance with local legislation.
The transmitter will also be supplied with a process seal, which will be fitted to the unit.
Depending on the version, this will either be a bonded seal (5/8” or G1/2” thread versions) or an
o-ring seal (3/4” thread versions).
The transmitter sensing element is protected while in transit by a cover containing a
small desiccant capsule. The connection pins are protected by a red plastic cap. None
of these plastic items are required for the operation of the transmitter. It is
recommended that the MiniDIN 43650 C connector is kept in a safe place until the
transmitter is ready for wiring.
97576 Issue 1, June 2019 2
SF82 Transmitter User’s Manual
Caution: when removing the central screw ensure that
The sensor cable is NOT supplied as standard. Cables can be obtained by contacting your local
Michell Instruments representative (see www.michell.com for details).
DIN 43650 Version
Cable connections to the SF82 transmitter are made via the removable connector. Removing the
central screw enables the connector terminal block to be removed from the outer housing by
using a small screwdriver to prise it clear.
\
the small sealing O-ring and the washer are retained on
the screw and are present during re-installation
The sensor cables are terminated as per the following diagram:
Note: The cable screen (see figure 2) should only be connected to a ground point at
either the transmitter installation side, or at the receiving equipment. Failure to
observe this precaution can result in ground loops and equipment malfunction.
3 Michell Instruments
SF82 Transmitter User’s Manual
Always connect the 4-20 m return signal to a suitable load
4-20 mA 2-wire
PIN 1
Modbus A
PIN 2
Modbus B
PIN 3
4-20 mA
PIN 4
Power Supply
PIN 5
0 V
Function
Pin
Wire Colour
Modbus A
1
Brown
Modbus B
2
White
4 -20 mA
3
Blue
Power Supply
4
Black
0 v
5
Grey
Figure 3 Sensor Connector Installation
(see figure 3) before the power is applied. Without this
connection, the transmitter may be damaged if allowed to
operate for prolonged periods.
M12 5-Pin Version
Cables with moulded M12 connectors are available from Michell Instruments in the following
lengths:
• 0.8 m
• 2 m
• 5 m
• 10 m
The other end of the sensor cable is unterminated, for straightforward connection into the
desired monitoring system.
Figure 4 Cable Connections
97576Issue 1, June 2019 4
SF82 Transmitter User’s Manual
Figure 5 Connector Installation
Figure 6 2-Wire Connection Diagram
O-ring and
washer
If longer cable runs are required, off the shelf 5-pin M12 cables can be connected between the
SF82 transmitter and the cable provided by Michell Instruments.
Note: The cable screen should only be connected to ground point at either the
transmitter installation side or at the receiving equipment. Failure to observe this
precaution can result in ground loops and equipment malfunction.
2.3. Cable Connection
DIN 43650 Version
To ensure the specified ingress protection is achieved, when installing the connector, the securing
screw (with the O-ring and washer) must be tightened to a minimum torque of 3.4 Nm (2.5 ftlbs). The sensor cable used must be a minimum diameter of 4.6 mm (0.2”).
M12 5-Pin Version
The connector should be installed by aligning the locating pin on the transmitter with the slot
on the cable. The connector can then be pushed into place and rotated until finger tight.
2.4. Electrical Schematic
Note: The cable screen should only be connected to a ground point at either the
transmitter installation side, or at the receiving equipment. Failure to observe this
precaution can result in ground loops and equipment malfunction.
5 Michell Instruments
Figure 7 Maximum Load of SF82 - Including Cable Resistance
2.4.1. Electrical Boundaries
SF82 Transmitter User’s Manual
2.4.2. Digital Communications (M12 Version Only)
Modbus RTU over RS485 communication is available on the SF82 M12, and can be used
simultaneously with the 2-wire current output. Section 2.2 describes the electrical connections to
the transmitter.
The Modbus register map can be found at the end of this manual.
2.5. Transmitter Installatiom
2.5.1. Sampling Considerations
There are two basic methods of measuring a sample with the SF82 Transmitter:
In-situ measurements are made by placing the transmitter inside the environment to be
measured.
Extractive measurements are made by installing the sensor into a block within a sample handling
system and flowing the sample outside of the environment to be measured through this system.
Extractive measurements are recommended when the conditions in the environment to be
measured are not conducive to making reliable measurements with the product.
Examples of such conditional limitations are:
• Excessive flow rate
• Presence of particulates matter
• Presence of entrained liquids
• Excessive sample temperature
97576Issue 1, June 2019 6
SF82 Transmitter User’s Manual
Figure 8 Installation Location
The basic considerations for each measurement type are as follows:
In-Situ
1. Dew-Point Sensor Position – will the sensor see an area of the environment that is
representative of what you want to measure?
For example, if the sensor is to be mounted into a glove box, there are three different positions
in which it could be installed – each giving a different measurement:
• Position A is on the purge inlet. In this position the sensor will confirm the dew point of the
gas entering the glove box but will not detect any leaks in the glove box itself, or any moisture
released from the work piece.
• Position B is on the gas outlet. In this position the sensor will be exposed to the gas leaving
the glove box and will therefore be detecting any moisture which has entered into the system
(e.g. ingress/leaks) or has been released by the work piece.
• Position C is in the glovebox itself, in this position the sensor will be only detecting any moisture
in its immediate vicinity. Leaks not in close proximity to the measurement point may not be
detected as this moisture could be drawn directly to the outlet.
If the transmitter is to be mounted directly into a pipe or duct, then consider that the installation
point should not be too close to the bottom of a bend where oil or other condensate may collect.
7 Michell Instruments
SF82 Transmitter User’s Manual
Figure 9 Installation Location
2. Gas speed – if you are planning on installing the sensor in a duct, consider how fast the
sample gas is moving through it.
• If the gas speed is very low, or occasionally static, then the moisture content through the
length (and width, if it is more than a few cm across) of the duct is unlikely to be uniform.
• Extremely high gas speeds can cause damage to the sensor. Direct insertion is not
recommended in gas speeds in excess of 10m/s (32.8ft/s).
3. Particulates – Particulates travelling at speed can cause severe and irreversible damage to
the sensor. At low velocity they can cling to the sensor, reducing its’ surface area, and
therefore response speed.
The sensor is provided with a basic level of particulate protection in the form of a sintered guard;
either HMWPE (10μm pore size) or Stainless Steel (80μm pore size). If the sample stream
contains smaller particulates than this, or generally large amounts of dust; extractive
measurement is recommended to accommodate proper in-line filtration.
4. Sample Temperature – Although the sensor can be operated at sample temperatures up
to 60°C, it is advisable to keep the sample temperature as close to ambient, and as stable as
possible to keep adsorption & desorption characteristics as consistent as possible (see section
2.5.2 Sampling Hints for more information).
Extractive
If the sensor is to be mounted into a sample conditioning system, then the above points are still
of relevance, but it is important to consider the extraction point itself – make sure that the chosen
extraction point is representative of the process, i.e. that the sample of interest is flowing past
the extraction point, and it is not being pulled from a dead volume.
97576Issue 1, June 2019 8
SF82 Transmitter User’s Manual
Figure 10 Transmitter Mounting - Sensor Block
Figure 11 Material Permeability Comparison
2.5.2. Good Measuring Practice
Ensuring reliable and accurate moisture measurements requires the correct sampling techniques,
and a basic understanding of how water vapour behaves. This section aims to explain the
common mistakes and how to avoid them.
Sampling Materials – Permeation and Diffusion
All materials are permeable to water vapour since water molecules are extremely small compared
to the structure of solids, even including the crystalline structure of metals. The graph above
demonstrates this effect by showing the increase in dew point temperature seen when passing
very dry gas through tubing of different materials, where the exterior of the tubing is in the
ambient environment. If the partial water vapour pressure exerted on the outside of a
compressed air line is higher than on the inside, the atmospheric water vapour will naturally push
through the porous medium causing water to migrate into the pressurised air line. This effect
is called transpiration.
9 Michell Instruments
SF82 Transmitter User’s Manual
What this demonstrates is the dramatic effect that different tubing materials have on the humidity
levels of a gas passed through them. Many materials contain moisture as part of their structure
and when these are used as tubing for a dry gas the gas will absorb some of the moisture. Always
avoid using organic materials (e.g. rubber), materials containing salts and anything which has
small pores which can easily trap moisture (e.g. nylon).
As well as trapping moisture, porous sampling materials will also allow moisture vapour to ingress
into the sample line from outside. This effect is called diffusion and occurs when the partial water
vapour pressure exerted on the outside of a sample tube is higher than on the inside. Remember
that water molecules are very small so in this case the term ‘porous’ applies to materials that
would be considered impermeable in an everyday sense – such as polyethylene or PTFE. Stainless
steel and other metals can be considered as practically impermeable and it is surface finish of
pipework that becomes the dominant factor. Electropolished stainless steel gives the best results
over the shortest time period.
Take into consideration the gas you are measuring, and then choose materials appropriate to
the results you need. The effects of diffusion or moisture trapped in materials are more significant
when measuring very dry gases than when measuring a sample with a high level of humidity.
Temperature and Pressure effects
As the temperature or pressure of the environment fluctuates, water molecules are adsorbed and
desorbed from the internal surfaces of the sample tubing, causing small fluctuations in the
measured dew point.
Adsorption is the adhesion of atoms, ions, or molecules from a gas, liquid, or dissolved solid to
the surface of a material, creating a film. The rate of adsorption is increased at higher pressures
and lower temperatures.
Desorption is the release of a substance from or through the surface of a material. In constant
environmental conditions, an adsorbed substance will remain on a surface almost indefinitely.
However, as the temperature rises, so does the likelihood of desorption occurring.
Ensuring the temperature of the sampling components is kept at consistent levels is important
to prevent temperature fluctuation (i.e. through diurnal changes) continually varying the rates of
adsorption and desorption. This effect will manifest through a measured value which increases
during the day (as desorption peaks), then decreasing at night as more moisture is adsorbed into
the sampling equipment.
If temperatures drop below the sample dew point, water may condense in sample tubing and
affect the accuracy of measurements.
97576Issue 1, June 2019 10
SF82 Transmitter User’s Manual
Maintaining the temperature of the sample system tubing above the dew point of the sample is
vital to prevent condensation. Any condensation invalidates the sampling process as it reduces
the water vapour content of the gas being measured. Condensed liquid can also alter the humidity
elsewhere by dripping or running to other locations where it may re-evaporate.
Although ambient pressure does not change drastically in a single location, the gas sample
pressure does need to be kept constant to avoid inconsistencies introduced by adsorption or
desorption. The integrity of all connections is also an important consideration, especially when
sampling low dew points at an elevated pressure. If a small leak occurs in a high-pressure line,
gas will leak out, however, vortices at the leak point and a negative vapour pressure differential
will also allow water vapour to contaminate the flow.
Theoretically flow rate has no direct effect on the measured moisture content, but in practice it
can have unanticipated effects on response speed and accuracy. An inadequate flow rate may:
• Accentuate adsorption and desorption effects on the gas passing through the sampling system.
• Allow pockets of wet gas to remain undisturbed in a complex sampling system, which will then
gradually be released into the sample flow.
• Increase the chance of contamination from back diffusion. Ambient air that is wetter than the
sample can flow from the exhaust back into the system. A longer exhaust tube can help
alleviate this problem.
• Slow the response of the sensor to changes in moisture content.
An excessively high flow rate can:
• Introduce back pressure, causing slower response times and unpredictable changes in dew
point
• Result in a reduction in depression capabilities in chilled mirror instruments by having a cooling
effect on the mirror. This is most apparent with gases that have a high thermal conductivity
such as hydrogen and helium.
System design for fastest response times
The more complicated the sample system, the more areas there are for trapped moisture to hide.
The key pitfalls to look out for here are the length of the sample tubing and dead volumes.
The sample point should always be as close as possible to the critical measurement point to
obtain a truly representative measurement. The length of the sample line to the sensor or
instrument should be as short as possible. Interconnection points and valves trap moisture, so
using the simplest sampling arrangement possible will reduce the time it takes for the sample
system to dry out when purged with dry gas.
Over a long tubing run, water will inevitably migrate into any line, and the effects of adsorption
and desorption will become more apparent.
11 Michell Instruments
SF82 Transmitter User’s Manual
Figure 12 Transmitter Mounting - Pipe or Duct
Dead volumes (areas which are not in a direct flow path) in sample lines, hold onto water
molecules which are slowly released into the passing gas. This results in increased purge and
response times, and wetter than expected readings. Hygroscopic materials in filters, valves (e.g.
rubber from pressure regulators) or any other parts of the system can also trap moisture.
Plan your sampling system to ensure that the sample tap point and the measurement point are
as close as possible to avoid long runs of tubing and dead volumes.
Filtration
All trace moisture measurement instruments and sensors are by their nature sensitive devices.
Many processes contain dust, dirt or liquid droplets. Particulate filters are used for removing dirt,
rust, scale and any other solids that may be in a sample stream. For protection against liquids, a
coalescing or membrane filter should be used. The membrane provides protection from liquid
droplets and can even stop flow to the analyser completely when a large slug of liquid is
encountered, saving the sensor from potentially irreparable damage.
2.5.3. Transmitter Mounting
Once an installation location has been chosen, this point will require a thread to match the
transmitter thread. Fixing dimensions are shown in Figure 6. For circular pipework, to ensure the
integrity of a gas tight seal, a mounting flange will be required on the pipework in order to provide
a flat surface to seal against.
97576Issue 1, June 2019 12
SF82 Transmitter User’s Manual
5/8” 18 UNF Version
1. Remove the protective cover and desiccant capsule from the transmitter and retain for future
use
2. Prevent any contamination of the sensor before installation by handling the transmitter by
the main body only, avoiding contact with the sensor guard.
3. Pass the bonded seal over the 5/8”- 18 UNF mounting thread.
4. Screw the transmitter into the sampling location or sample block by hand using the wrench
flats only. DO NOT grip and twist the sensor cover when installing the sensor.
5. When installed, fully tighten using a wrench to a torque setting of 30.5 Nm (22.5 ft-lbs)
3/4” - 16 UNF Version
1. Remove the protective cover and desiccant capsule from the transmitter and retain for future
use.
2. Prevent any contamination of the sensor before installation by handling the transmitter by
the main body only, avoiding contact with the sensor guard.
3. Ensure that the O-ring is seated in the recess at the top of the transmitter body.
4. Screw the transmitter into the sampling location or sample block by hand using the wrench
flats only. DO NOT grip and twist the sensor cover when installing the sensor.
5. When installed, fully tighten using a wrench to a torque setting of 40 Nm (29.5 ft-lbs).
G1/2" BSPP Version
1. Remove the protective cover and desiccant capsule from the transmitter and retain for future
use
2. Prevent any contamination of the sensor before installation by handling the transmitter by
the main body only, avoiding contact with the sensor guard.
3. Pass the bonded seal over the G1/2" mounting thread.
4. Screw the transmitter into the sampling location or sample block by hand using the wrench
flats only. DO NOT grip and twist the sensor cover when installing the sensor.
5. When installed, fully tighten using a wrench to a torque setting of 30.5 Nm (22.5 ft-lbs)
13 Michell Instruments
SF82 Transmitter User’s Manual
Figure 13 Transmitter Mounting with Adapter
Installation using Additional Thread Adaptor
1. Remove the protective cover and desiccant capsule from the transmitter and retain for future
use
2. Prevent any contamination of the sensor before installation by handling the transmitter by
the main body only, avoiding contact with the sensor guard.
3. Pass the bonded seal over the 5/8”- 18 UNF mounting thread.
4. Screw the transmitter into the adaptor, and tighten to 30.5 Nm (22.5 ft-lbs)
5. NOTE: Use the flats of the hexagonal nut and not the sensor body.
6. Screw the transmitter (1) with its seal (3) and adapter (4) into the sampling location block
(and fully tighten using a wrench to the following torque settings:
• G 1/2” BSP 56 Nm (41.3 ft-lbs)
• 3/4” - 16 UNF 40 Nm (29.5 ft-lbs)
97576Issue 1, June 2019 14
SF82 Transmitter User’s Manual
Figure 14 Replacement HMWPE Guard
Handle, using
Calibration
Annual recalibration of the SF82 is recommended to maintain the performance. Calibration
services traceable to the UK National Physical Laboratory (NPL) and the US National Institute of
Standards and Technology (NIST) are provided by Michell Instruments.
Michell Instruments offers a variety of re-calibration and exchange sensor schemes to suit specific
needs. A Michell representative can provide detailed, custom advice (for Michell Instruments’
contact information go to www.michell.com).
Sensor Guard Replacement
The sensor is supplied with a white HMWPE guard (standard) or a stainless steel guard (if
specified at time or order).
The sensor guard should be replaced if the surface shows any damage or signs of discolouration.
When replacing a guard, make sure to wear clean disposable gloves, and handle by the threaded
base section only.
Replacement HMWPE or stainless steel guards can be ordered from your Michell Instruments
representative.
gloves, by white
knurled part only
Bonded Seal
If the supplied bonded seal is damaged or lost, a pack of 5 replacement bonded seals can be
obtained by contacting your Michell Instruments representative.
O-ring Seal
If the supplied O-ring seal is damaged or lost a pack of 5 replacement O-ring seals
can be obtained by contacting your Michell Instruments representative.
15 Michell Instruments
SF82 Transmitter User’s Manual
APPENDIX A
Technical Specifications
97576 Issue 1, June 2019 16
SF82 Transmitter User’s Manual
Measurement Range (Dew
Point)
63% at room temperature at 1 bara
-20 ºC to -60 ºC dew point: 40 s
User configurable over range;
communications
-60 °C to +60 °C dew point
Analog only 23 mA max,
digital only 6 mA max
IEC61010-1, UL61010-1 &
EN50121-3-2 Rail EMC/RFI
Technical Specifications
Performance
Product
SF82 MiniDIN 43650 SF82 M12
-60 ºC to +60 ºC dew point
Accuracy (Dew Point)
Response Time
-60 ºC to -20 ºC dew point: 6 s
±2 °C dew point*
Repeatability 0.5 ºC dew point
Calibration 9-point calibration certificate traceable to national standards
Electrical Specifications
Output Signal
User configurable over range;
4-20 mA (2 wire connection,
current source)
4-20 mA (2 wire connection,
current source) Modbus RTU
over RS485 digital
Moisture Output Dew point or moisture content
Temperature Output Not available Data via Modbus RTU
Analog output scaled range
4-20 mA (Dew point)
Analog output scaled range
4-20 mA (Moisture content
in gas)
1 to 5 Nl/min mounted in standard sampling block; 0 to 10
-20 °C to +60 °C
-20 °C to +50 °C
-40 °C to +60 °C
m/sec direct insertion
Mechanical Specifications
60529 (current version):
Ingress protection
NEMA 4 ingress protection in
accordance with NEMA 250
Housing material 316 stainless steel
IP65
Dimensions
Filter (sensor protection)
Process connection
3/4” - 16 UNF
G1/2” - BSP
Weight 150 g (excluding connector cable)
Electrical connections MiniDIN 43650 form CM12 5 pin (A coded)
Optional 0.8 , 2, 5, 10 meter
M12 A coded
connector/cable available
Mating Electrical
Connectors
Diagnostic conditions
standard
Optional 0.8, 2, 5, 10 metre
MiniDIN connector/cable
Under-range dew point: mA
(factory programmed)
NOTES: * Over Compensated Temperature Range
97576Issue 1, June 2019 18
SF82 Transmitter User’s Manual
M12, 5/8” UNF
M12, G1/2
M12, 3/4" UNF
Product Dimensions
19 Michell Instruments
SF82 Transmitter User’s Manual
MiniDIN, 5/8” UNF
MiniDIN, G1/2
MiniDIN, 3/4” UNF
97576 Issue 1, June 2019 20
SF82 Transmitter User’s Manual
5/8” UNF
G1/2
Quick Connect
3/4” UNF
21 Michell Instruments
SF82 Transmitter User’s Manual
APPENDIX B
Quality, Recycling &
Warranty Information
97576 Issue 1, June 2019 22
SF82 Transmitter User’s Manual
Quality, Recycling & Warranty Information
Michell Instruments is dedicated to complying to all relevant legislation and directives. Full
information can be found on our website at:
www.michell.com/compliance
This page contains information on the following directives:
• ATEX Directive
• Calibration Facilities
• Conflict Minerals
• FCC Statement
• Manufacturing Quality
• Modern Slavery Statement
• Pressure Equipment Directive
• REACH
• RoHS2
• WEEE2
• Recycling Policy
• Warranty and Returns
23 Michell Instruments
SF82 Transmitter User’s Manual
APPENDIX C
Return Document &
Contamination Declaration
97576 Issue 1, June 2019 24
SF82 Transmitter User’s Manual
Return Document & Decontamination Declaration
25 Michell Instruments
SF82 Transmitter User’s Manual
APPENDIX D
Modbus Register Map
97576 Issue 1, June 2019 26
SF82 Transmitter User’s Manual
16-bit unsigned integer, can contain options list e.g. 0 = Dew Point, 1 =
Temperature.
http://www.simplymodbus.ca/FAQ.htm is an excellent
be found in the sidebar.
https://www.scadacore.com/tools/programming-
Modbus Register Map
All the data values relating to the SF82 are stored in 16-bit wide holding registers. Registers can
contain either measured or calculated values (dew-point, temperature, etc.), or configuration
data (output settings).
Modbus RTU Implementation
This is a partial implementation of the Modbus RTU Standard with the following codes
implemented:
Function Code Description
3 Read Holding Register
6 Write Holding Register
16 Write Multiple Holding Registers
Register Types
Data Type Description
uint16
int16 16-bit signed integer.
int32 32-bit signed integer, stored across 2 16-bit registers.
float IEEE754 single precision floating pint, stored across 2 16-bit registers
Serial Port Settings (RS485)
9600 Baud Rate, 8 Data Bits, No Parity, 1 Stop Bit, No Flow Control
resource covering the basics of the Modbus protocol. Full
descriptions of the function codes (FC03/FC06/FC16) can
calculators/online-hex-converter/ is an excellent resource
for determining register types/byte order issues in raw
received Modbus data.
27 Michell Instruments
Dec
Hex
Access
Data
Type
Description
Comment
Instrument Modbus
Address
1
01 R uint16
Instrument ID
Batch 0xA123
A123-001
3
03 R uint16
Sensor Serial Number
4
04 R uint16
Firmware Version
Divide by 1000, ie 12003 = V12.003
5
05 R uint16
Register Map Version
Divide by 1000, ie 12003 = V12.003
6
06 R uint16
Year of Calibration
7 07 R uint16
Month of Calibration
8 08 R uint16
Day of Calibration
… …
bit0 = Dew-point Sensor Short
bit15 = Hardware Fault
… …
17
11 R float
Dew Point (High Word)
18
12
Dew Point (Low Word)
19
13 R float
Temperature (High
Word)
20
14
Temperature (Low
Word)
Register Address
SF82 Transmitter User’s Manual
0 00 R uint16
2 02 R uint16 Sensor Batch Number
Serial 0x0001
Complete sensor serial would be
14 0E R special Status
bit1 = Dew-point Sensor Open
bit2 = Temperature Sensor Short
bit3 = Temperature Sensor Open
bit4 = Analogue Output UnderRange
bit5 = Analogue Output Over-Range
bit6 = Analogue Output Out-OfRange
...
bit14 = Memory Fault
97576Issue 1, June 2019 28
SF82 Transmitter User’s Manual
21
15 R float
ppmV Ideal Gas (High
Word)
22
16
ppmV Ideal Gas (Low
Word)
… …
101
65
R/W
float
Pressure Value (High
Word)
Used for ppmV Ideal Gas calculation
102
66
Pressure Value (Low
Word)
… …
110
6E
R/W
uint16
Analogue Output
0 = Off
3 = ppmV Ideal Gas
111
6F
R/W
float
Analogue Output Range
This value is clipped when
parameter ranges below
112
70
Analogue Output Range
Low (Low Word)
113
71
R/W
float
Analogue Output Range
This value is clipped when
parameter ranges below
114
72
Analogue Output Range
High (Low Word)
… …
R/W
uint16
Analogue Output,
R/W
uint16
Analogue Output, OverRange Output
R/W
uint16
Analogue Output, DewPoint Sensor Fault
R/W
uint16
Analogue Output,
Fault
120 78
Parameter
Low (High Word)
High (High Word)
Under-Range Output
1 = Dew Point
2 = Temperature
parameter is changed. See
parameter is changed. See
0 = None
1 = Low Alarm (3.5ma)
121 79
122 7A
123 7B
29 Michell Instruments
Temperature Sensor
2 = High Alarm (23ma)
3 = Minimum Scale (4ma)
4 = Maximum Scale (20ma)
5 = Namur Low Alarm (3.7ma)
6 = Namur High Alarm (20.5ma)
Parameter Ranges
Min
Max
Dew Point
-150
250
Temperature
-150
250
ppmV
0
30000
SF82 Transmitter User’s Manual
97576Issue 1, June 2019 30
http://www.michell.com
Aufgrund laufender Weiterentwicklungen sind Änderungen der Spezikationen vorbehalten. Alle Angaben vorbehaltlich Satz- und Druckfehler.