Michell Instruments S4000 Specifications

Chilled Mirror Instruments
www.michell.com
S4000 TRS
A precision laboratory dew-point hygrometer with the ultimate accuracy, reliability and long term performance for humidity measurement and calibration.
Highlights
0.1°Cdp accuracy (0.18°Fdp)
Measurement range: -100 to +20°Cdp
(-148 to +68°Fdp)
Precision 100 Ω 4-wire platinum resistance thermometer
Dual optics detection system
Available with VCR couplings for optimum trace
moisture sampling
Dual multi-function LED display with unit indicator
Applications
Standards Laboratory reference instrument
Research and development
Battery manufacture
Industrial gases
Chilled Mirror Instruments
S4000 TRS Precision Dew-Point Hygrometer
The Laboratory Standard
The S4000 TRS offers unmatched accuracy and reliability in dew-point measurement and calibration. The powerful three­stage Peltier thermoelectric heat pump, coupled with integrated auxiliary refrigerant cooling, gives an effective measurement range down to -100°C (-148°F) dew point.
Dual Optics for Supreme Sensitivity
At low frost points the rate of formation of frost on the mirror surface is extremely slow. As a result other chilled mirror dew­point meters may give reduced accuracy, poor control stability and extremely long response times at low moisture levels. The S4000 TRS is unique in that it utilizes a dual optics detection system. This greatly increases the sensitivity of the optical loop. Response, stability and sensitivity are improved by orders of magnitude at low frost points.
Calibration Integrity
The S4000 TRS is unique amongst dew-point hygrometers. It is the only instrument that is delivered, as standard, with a full UKAS certificate providing official traceability to the UK National Standard. UKAS is the United Kingdom member of European Co-operation on Accreditation (EA), the International Laboratory Accreditation Cooperation (ILAC) and the International Accreditation Forum (IAF). As an additional benefit, the S4000 TRS has a direct traceability path to the NIST (Washington DC, USA) National Humidity Standard.
No other hygrometer provides such comprehensive traceability to a worldwide network of standards organizations. The S4000 TRS is used by many of these organisations as part of their own humidity referencing systems.
Contamination Compensation
Any optical system carries a risk of contamination. The S4000 TRS automatically compensates for any such build­up with its ABC (Automatic Balance Compensation) System. ABC ensures continuous optimum operation of the sensor by periodically driving off condensation to allow the optical loop to be rebalanced. When the contamination level is too high a visible alarm is generated. The sensor optical system may be cleaned with distilled water or a suitable high purity solvent such as acetone. ABC cycle time, duration and recovery time can all be adjusted according to the type of application to minimize the effect of contamination risk. The S4000 TRS also features a sophisticated data hold system, which maintains the instruments signal outputs during an ABC cycle, allowing the S4000 TRS to be used for process control applications.
Unbeatable Measurement Capability
The S4000 TRS uses a highly accurate 4-wire PT100 temperature sensor and has a proven measurement capability of better than ±0.1°C (±32.18°F) dew point.
Visual Verification
Every S4000 TRS is delivered complete with an M4K Viewing Microscope. Fitting neatly into the sensor viewing port, this microscope allows the user to confirm the presence of water or ice on the mirror surface.
www.michell.com
Chilled Mirror Instruments
www.michell.com
3
Technology: Chilled Mirror
Michells chilled mirror dew-point hygrometers are precision instruments for critical measurement and control applications. Chilled mirror sensors measure a primary characteristic of moisture – the temperature at which condensation forms on a surface. This means that chilled mirror instruments:
5
have no drift: the temperature at which condensation forms is measured directly so there are no calculated variables that could shift over time
are inherently repeatable, giving reliable results every time
The chilled mirror sensor consists of a temperature controlled mirror and an advanced optical detection system.
A beam of light from an LED (1) is focused on the mirror surface (2) with a fixed intensity. As the mirror is cooled, less light is reflected due to the scattering effect of the condensate formed on the mirror surface. The levels of reflected and scattered light are measured by two photo-detectors (3 & 4) and compared against a third reference detector (5) measuring the intensity of light from the LED.
The signals from this optics system are used to precisely control the drive to a solid state thermoelectric cooler (TEC) (6), which heats or cools the mirror surface. The mirror surface is then controlled in an equilibrium state whereby evaporation and condensation are occurring at the same rate. In this condition the temperature of the mirror, measured by a platinum resistance thermometer (7), is equal to the dewpoint temperature of the gas.
An auxiliary cooling system (8) is used to remove heat from the ‘hot side of the TEC. This supplements the depression capabilities of the heat pump, and enables measurement of very low dew points.
Our chilled mirror instruments prove their reliabilty on a daily basis in our production processes and service centres, as well as in our UKAS-accredited calibration laboratory.
1
4
2
6
7
8
Chilled Mirror Instruments
Performance
Measurement technology Chilled Mirror
Measurement range TRS: -100 to +20°Cdp (-148 to +68°Fdp)
Measurement accuracy ±0.1°Cdp (±0.18°Fdp)
±0.1°C temperature (±0.18°Fdp)
Measurement units °Cdp, °Fdp; °C, °F temperature;
% RH, ppmV, ppmW, g/m3, g/kg, ppmW for SF
6
Response speed 0.5°C/sec (0.9°F/sec) + settling time
(dew point dependent)
Sensitivity 0.01°C (0.02°F)
Repeatability Better than 0.1°C (0.18°F)
Resolution 0.01 (0.1 for % RH)
Dew-Point Sensor
Mirror Gold plated copper
Temperature 4 wire Pt100, 1/10 DIN class B measurement
Sample flow rate 0.1 to 0.7 Nl/min (recommended)
(0.21 to 1.48 scfh)
Integrated flowmeter 0 to 1 Nl/min (0 to 2.12 scfh)
Sensor pressure Atmospheric
Auxiliary cooling Internal refrigeration
Remote PRT
Temperature 4 wire Pt100, 1/10 DIN class B measurement
Monitor
Resolution 0.01°C (0.02°F)
Dual optics detection Wide band red LED with dual photo sensors, all
system insulated
Outputs
Analog 2 channels
10 mV/°Cdp, 4-20 mA
Digital RS232
Logic Data hold, ABC Logic status, optics alarm
Auxiliary input 4-20 mA input for automatic pressure transducer compensation 0-0.34 MPa (0-50 psia) (optional)
Operating temperature 0 to +40°C (32 to 104°F)
Dimensions 560W x 600D x 860mm - mini rack
(22.05 x 23.62 x 33.86 inches)
Weight 85.0kg (187.39lbs)
Power supply
Monitor 90 to 265 V AC; 50-60 Hz
Sensor 100 to 115 or 220 to 240 V AC 50-60 Hz
Technical Specifications
Dimensions
© 2012 Michell Instruments
Michell Instruments Ltd, Rotronic Instruments Corp. 135 Engineers Road, Suite 150, Hauppauge NY 11788. Tel: 631 427 3898, Email: us.info@michell.com, Web:
Michell Instruments adopts a continuous development programme which sometimes necessitates specification changes without notice. Issue no: S4000 TRS_97146_V4.1_US_0718
www.michell.com/us