This document is the property of Michell Instruments Ltd. and may not be copied or otherwise reproduced,
communicated in any way to third parties, nor stored in any Data Processing System without the express
written authorization of Michell Instruments Ltd.
97300 Issue 1, March 2012 iii Michell Instruments
PCMini52 User’s Manual
Contents
Safety .............................................................................................................................................. v
Electrical Safety ......................................................................................................................... v
Toxic Materials .......................................................................................................................... v
Repair and Maintenance............................................................................................................. v
Calibration ................................................................................................................................ v
Safety Conformity ...................................................................................................................... v
Appendix A Technical Specications ...........................................................................................5
Appendix B List of Worldwide Michell Instruments’ Ofces ........................................................... 7
97300 Issue 1, March 2012 iv Michell Instruments
PCMini52 User’s Manual
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures detailed in this
manual. The user must not use this equipment for any other purpose than that stated. Do not apply values
greater than the maximum value stated.
This manual contains operating and safety instructions, which must be followed to ensure the safe operation
and to maintain the equipment in a safe condition. The safety instructions are either warnings or cautions
issued to protect the user and the equipment from injury or damage. Use competent personnel using good
engineering practice for all procedures in this Manual.
Electrical Safety
The instrument is designed to be completely safe when used with options and accessories supplied by the
manufacturer for use with the instrument.
Toxic Materials
The use of hazardous materials in the construction of this instrument has been minimized. During normal
operation it is not possible for the user to come into contact with any hazardous substance which might be
employed in the construction of the instrument. Care should, however, be exercised during maintenance and
the disposal of certain parts.
Repair and Maintenance
The instrument must be maintained either by the manufacturer or an accredited service agent. Refer to
Appendix B for details of Michell Instruments’ worldwide ofces contact information.
Calibration
Michell Instruments recommends an annual calibration for an accuracy requirement of ±2% RH under
ambient conditions where temperature is 0 to +50°C (+32 to +122°F) and relative humidity is 0 to 70%
RH. For environments with airborne chemicals or for high humidity and high temperature conditions, Michell
recommends more frequent calibration.
Safety Conformity
This product meets the essential protection requirements of the relevant EU directives. Further details of
applied standards may be found in the product specication.
Michell Instruments v 97300 Issue 1, March 2012
PCMini52 User’s Manual
Abbreviations
The following abbreviations are used in this Manual:
AC alternating current
°C degrees Celsius
°F degrees Fahrenheit
DC direct current
g Gram(s0
oz ounce(s)
sec second(s)
temp temperature
V Volts
" inch(es)
ø diameter
μ micro
Recycling Policy
Michell Instruments is concerned with the protection of the environment. It is our commitment to reduce
and eliminate from our operations, wherever possible, the use of substances which may be harmful to the
environment. Similarly, we are increasingly using recyclable and/or recycled material in our business and
products wherever it is practical to do so.
The product that you have purchased may contain recyclable and/or recycled parts and we will be happy to
provide you with information on these components if required.
WEEE And RoHS Compliance
The Waste Electronic and Electrical Equipment (WEEE) Directive, and the Restriction of Hazardous Substances
(RoHS) Directive place rules upon European manufacturers of electrical and electronic equipment. The
directive’s aim is to reduce the impact that electronic devices have on the environment.
Michell products are currently exempt from the RoHS directive, however all future products will be developed
entirely using compliant materials. Furthermore, Michell is taking active steps to remove non-compliant
materials and components from existing products wherever possible.
Michell is in full compliance with the WEEE Directive (Registration No. WEE/JB0235YW). Customers may be
required to return certain instruments for treatment at the end of their working life.
June 2010
97300 Issue 1, March 2012 vi Michell Instruments
PCMini52 User’s Manual
Warranty
Unless otherwise agreed, the Supplier warrants that, as from the date of delivery for a period of 12 months,
the goods and all their component parts, where applicable, are free from any defects in design, workmanship,
construction or materials.
The Supplier warrants that the services undertaken shall be performed using reasonable skill and care, and of
a quality conforming to generally accepted industry standards and practices.
Except as expressly stated all warranties, whether express or implied, by operation of law or otherwise, are
hereby excluded in relation to the goods and services to be provided by the Supplier.
All warranty services are provided on a return to base basis. Any transportation costs for the return of a
warranty claim shall reside with the Customer.
Return Policy
If a Michell Instruments’ product malfunctions within the warranty period, the following procedure must be
completed:
1. Notify a Michell Instruments’ representative, giving full details of the problem,
the model variant and the serial number of the product.
2. If the nature of the problem indicates the need for factory service then the
instrument should be returned to Michell Instruments, carriage prepaid,
preferably in the original packaging, with a full description of the fault and the customer contact information.
3. Upon receipt, Michell Instruments will evaluate the product to determine the
cause of the malfunction. Then, one of the following courses of action will
be taken:
• If the fault is covered under the terms of the warranty, the
instrument will be repaired at no cost to the owner and returned.
• If Michell Instruments determines that the fault is not covered
under the terms of the warranty, or if the warranty has expired,
an estimate for the cost of the repairs, at standard rates, will be
provided. Upon receipt of the owner’s approval to proceed, the
product will be repaired and returned.
Michell Instruments vii 97300 Issue 1, March 2012
PCMini52 User’s Manual
1 INTRODUCTION
The micro-processor control and multi-point calibration give the PCMini52 RH probe excellent performance in
terms of accuracy and linearity. The mini probe can provide two linear analog outputs for temperature and
relative humidity, dew point, absolute humidity or wet bulb temperature.
1.1 Features
• Output can be % RH, calculated absolute humidity, dew point or wet bulb
temperature + temperature
• Excellent linearity, (μP) micro-processor temperature compensated
• Low power consumption, fast settling time
• Small size: L = 80mm, ø12mm (3.15”, ø0.47”)
1.2 Dimensions
80mm
(3.15”)
11mm
(0.43”)
Ø12mm
(0.47”)
12.5mm
(0.49”)
M8 x 1
M12 x 1
Figure 1.1 Dimensions
Michell Instruments 1 97300 Issue 1, March 2012
PCMini52 User’s Manual
2 INSTALLATION
2.1 Wall Mounting
The PCMini52 can be used for wall mounting and must always be installed with the sensing element facing
downwards so as to minimize the propagation of heat between the transmitter and the sensitive element. It
should be attached to the wall with a mounting clip.
Figure 2.1 Wall Mount Position
2.2 Duct Mounting
The PCMini52 can be used for duct mounting and must be installed so that the end of the sensor body (the
lter) comes into contact, perpendicularly, with the ow of air from which the relative humidity is to be
measured.
An extra opening should be made in the air duct close to the opening for the installation of the sensor in order
that checks and further calibration may be performed, when required.
If necessary, an appropriate form of thermal insulation should be placed between the walls of the duct and
the transmitter housing.
DUCT
Figure 2.2 Duct Mount Position
97300 Issue 1, March 2012 2 Michell Instruments
3 CALIBRATION
The transmitter is calibrated at two points.
Cable color
BrownSupply voltage V +
WhiteOutput RH, Dew Point or Absolute Humidity
BlackOutput temperature
BlueCommon ground
PCMini52 User’s Manual
GND
OUTPUT TEMPOUTPUT RH
+ SUPPLY (8 - 35 V DC)
Figure 3.1 Cable Connections
Michell Instruments 3 97300 Issue 1, March 2012
PCMini52 User’s Manual
Appendix A
Technical Specications
97300 Issue 1, March 2012 4 Michell Instruments
PCMini52 User’s Manual
Appendix A Technical Specications
Performance
Measurement range (RH)0–100% RH
Measurement range (T)-20 to +80°C (-4 to +176°F)
Accuracy at 23°C (73°F)
Humidity
Accuracy at 23°C (73°F)
Temperature
Stability – RH Sensor±1% RH/year
Response time – RH Sensor<10 sec typical (for 90% of the step change)
<±2% RH (10–90% RH)
±0.2°C (±0.36°F)
Electrical output/input
Output signal options0–1, 0–5, 0–10 V
Supply voltage14–35 V DC (for 0–5 / 10 V output)
4.5–35 V DC (for 0–1 V output)
or 14–26 V AC (for all output ranges)
Operating conditions
Operating temperature
Sensing element
Housing
Storage
-30 to +85°C (-22 to +185°F)
-30 to +85°C (-22 to +185°F)
-40 to +85°C (-40 to +185°F)
Mechanical specication
Ingress protectionIP65 (NEMA 4 level)
Housing materialMolded polymer or stainless steel (ordering option)
DimensionsL=80mm, ø12mm(L=3.15”, ø0.47”)
Weight10g (0.35oz)
Electrical connections4 pin, M8 connector
Michell Instruments 5 97300 Issue 1, March 2012
PCMini52 User’s Manual
Appendix B
List of Worldwide Michell
Instruments’ Ofces
97300 Issue 1, March 2012 6 Michell Instruments
PCMini52 User’s Manual
Appendix B List of Worldwide Michell Instruments’ Ofces
Asia
Michell Asia
PO Box 3149
Joondalup
WA 6027
Australia
Tel: +61 893 046587
E-mail: au.info@michell.com
Web: www.michell.com/au
China
Michell Instruments (Shanghai) Ltd
Room 1007, Qilai Building
889 Yishan Road
Shanghai, 200233
P R China
Tel: +86 21 5401 2255
Fax: +86 21 5401 2085
E-mail:
Michell Instruments SAS
2-4, rue Jean Desparmet
69008 Lyon
France
Tel: +33 437 53 88 20
Fax: +33 437 53 88 21
E-mail: fr.info@michell.com
Web: www.michell.com/fr
Italy
Michell Italia Srl
Via Capecelatro, 10
20148 Milano
Italy
Tel: +39 02 4047194
Fax: + 39 02 40010565
E-mail: it.info@michell.com
Web: www.michell.com/it
Japan
Michell Japan KK
Musashino Center Building
1-19-18 Nakacho, Musashino
Tokyo 180-0006
Japan
Tel: +81 422 502600
Fax: +81 422 521700
E-mail: info@michell-japan.co.jp
Web: www.michell-japan.co.jp
North America
Michell Instruments Inc
319 Newburyport Turnpike, Suite 207
Rowley, MA 01969
USA
Tel: +01 978 484 0005
Fax: +01 978 843 7669
E-mail:
us.info@michell.com
Web: www.michell.com/us
Middle East
Michell Instruments Middle East
P-06, #097
Sharjah Airport Int’l free zone
Sharjah,
United Arab Emirates
Tel: +971 6 5575028
+971 6 5575029
Fax:
E-mail: me.info@michell.com
United Kingdom
Michell Instruments Ltd
48 Lancaster Way Business Park
Ely, CB6 3NW
Cambridgeshire
England
Tel: +44 1353 658000
Fax: +44 1353 658199
E-mail: info@michell.com
Web: www.michell.com/uk
Michell Instruments 7 97300 Issue 1, March 2012
PCMini52 User’s Manual
NOTES:
97300 Issue 1, March 2012 8 Michell Instruments
http://www.michell.com