Appendix D Return Document & Decontamination Declaration ........................................ 43
Michell Instrumentsv
OptiCal User’s Manual
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures
detailed in this manual. The user must not use this equipment for any other purpose than that
stated. Do not apply values greater than the maximum value stated.
This manual contains operating and safety instructions, which must be followed to ensure the safe
operation and to maintain the equipment in a safe condition. The safety instructions are either
warnings or cautions issued to protect the user and the equipment from injury or damage. Use
qualifi ed personnel and good engineering practice for all procedures in this manual.
Electrical Safety
The instrument is designed to be completely safe when used with options and accessories supplied
by the manufacturer for use with the instrument. The input power supply voltage is 85 to 264 V AC,
47/63 Hz.
Toxic Materials
The use of hazardous materials in the construction of this instrument has been minimized. During
normal operation it is not possible for the user to come into contact with any hazardous substance
which might be employed in the construction of the instrument. Care should, however, be exercised
during maintenance and the disposal of certain parts.
Repair and Maintenance
The instrument must be maintained either by the manufacturer or an accredited service agent. Refer
to www.michell.com for details of Michell Instruments’ worldwide offi ces contact information.
Calibration
The OptiCal contains a high precision Optidew chilled mirror hygrometer. In order to maintain full
traceability the complete OptiCal instrument should be returned annually for a service and calibration
to the manufacturer, Michell Instruments, or one of their accredited service agents.
Safety Conformity
This product meets the essential protection requirements of the relevant EU directives. Further
details of applied standards may be found in Appendix A, Technical Specifi cations.
vi97079 Issue 9.2, March 2016
OptiCal User’s Manual
Abbreviations
The following abbreviations are used in this manual:
°C degrees Celsius
°F degrees Fahrenheit
AC alternating current
DC direct current
3
cubic centimeters
cm
-1
grams per kilogram
gkg
-3
grams per cubic meter
gm
Hz Hertz
in inch(es)
3
cubic inches
in
kg kilogram
lbs pounds
mA milli Ampere
max maximum
min minimum
mm millimeter
N/C normally closed
N/O normally open
% percentage
RH relative humidity
RS232 serial data transmission standard
T temperature
USB Universal Serial Bus
V Volts
Warnings
The following general warning listed below is applicable to this instrument. It is repeated
in the text in the appropriate locations.
Where this hazard warning symbol appears in the following
sections it is used to indicate areas where potentially hazardous
operations need to be carried out.
Michell Instrumentsvii
OptiCal User’s Manual
1 INTRODUCTION
1.1 Description
The Michell OptiCal is a stable and accurate calibration solution for humidity sensors
over the 10 to 90%RH and +10 to +50°C (+50 to +122°F) range. The stand-alone,
transportable calibrator requires no external services other than mains power, and
features an integrated chilled mirror reference instrument to enable the operator to
perform calibrations that are traceable to national standards.
The calibration chamber features fi ve interchangeable ports to accommodate virtually
any brand, type or model of sensor. The environment within the insulated calibration
chamber is temperature controlled using a four-zone fan-assisted peltier arrangement
for maximum stability and minimum temperature gradient. The humidity of the
circulating air is precisely regulated using a closed-loop control system that functions
by proportionally mixing fl ows of dry and saturated air.
A bright and clear VFD (vacuum fl uorescent display) displays the parameters measured
by the reference instrument in various relative and absolute humidity units, alongside
the temperature within the chamber.
INTRODUCTION
The humidity and temperature set-points can be controlled either manually or
automatically as part of a calibration program. Manual control is achieved by the switches
on the front panel. Response time to a humidity or temperature step change is typically
quicker than ten minutes. The supplied application software allows calibration programs
to be created, enabling automatic time-based control of temperature and humidity set
points. The software also allows the user to monitor, chart and log calibration reference
data on a PC for later analysis.
The desiccant changes color to indicate when it needs to be recharged, and is visible
through a clear window on the front of the unit. Recharge the desiccant following the
instructions in Section 2.4. The water reservoir at the front of the unit shows the current
saturator fi ll level, and makes it easy to top up with distilled water when required (refer
to Section 2.2). Apart from periodic calibration of the chilled mirror reference, no other
maintenance is necessary. The only external service required is a single phase power
supply.
Michell Instruments1
INTRODUCTION
1.2 System Components
1.2.1 Front Panel
OPTI-CAL
integrated Humidity Calibrator
OptiCal User’s Manual
h
AA
Auto OnOn Auto
Man. OffOff Man.
h
BB
7
h
Chamber Set
Temperature
126438
h
h
5
h
Display Control
495206
Chamber Set
Humidity
%rh
h
h
°C
Max.
h
Min.
Data AcquisitionControl Outputs
h
h
Desiccant Charge
Humidity = Yellow
De-Humidify - Green
h
Heating = Yellow
Cooling = Green
h
12
11
Figure 1
OptiCal - Front View
1. Chamber door
2. Water reservoir
3. Optidew chilled mirror reference VFD display
4. Desiccant cell and indicator window
5. Display control
6. Relative humidity set point (%RH)
A Manual/auto switches for relative humidity / temperature control
7.
9
10
MAN - Front panel switches control relative humidity / temperature set
points
AUTO - Remote RS232 / OptiSoft controls relative humidity / temperature
set points (see Section 3.4)
B ON/OFF switches for relative humidity / temperature control
8. Temperature set-point (°C)
9. Humidity control indication LED
Humidify (yellow)
De-humidify (green)
10. 4-Zone chamber temperature control indication LEDs
Heating (yellow)
Cooling (green)
11. Control outputs
12. RS232 digital communications connector
297079 Issue 9.2, March 2016
OptiCal User’s Manual
Back Panel
INTRODUCTION
1
h
On
Off
h
h
4
Remote Communications
Optics Adjustment
Figure 2
2
3
h
h
OptiCal - Back Panel
1. Remote Communications Switch to enable remote chamber control
4. Electrical mains connector, on/off switch and power input fuse
Accessories
The OptiCal humidity and temperature generator is delivered with the following standard
accessories:
• Bottle of distilled water
• Desiccant cell fi lled with indicating silica gel
• HT961T00 control probe + temperature probe
• Chamber door (according to customer specifi cation)
• IEC mains cord set
• Calibration certifi cate for internal reference probe
• Certifi cate of conformity (graph)
• User’s manual
• RS232 cable
• Application software
• Calibration certifi cate for Optidew chilled mirror reference instrument
Michell Instruments3
INSTALLATION
2 INSTALLATION
Before using the OptiCal make sure that Sections 2.1, 2.2 and
The OptiCal enclosure is designed for bench top mounting in a laboratory type
environment. It must be positioned in a clean and level location with suffi cient clearance
at the rear of the enclosure for adequate ventilation.
The OptiCal is not designed to be fully portable. However, it can
easily be moved to any suitable location for use. Before moving,
ensure that any water in the reservoir is drained and that the
RH control probe and the temperature probe in the chamber are
OptiCal User’s Manual
2.4 are read thoroughly.
removed.
The OptiCal should NOT be moved while in operation.
2.1 Installing the Relative Humidity and Temperature Control Probes
The relative humidity control probe and the Pt100 temperature probe are both supplied
as accessories with the OptiCal, and are removed during transportation.
To install the probes, remove the chamber door and plug in the probes as shown in
Figure 3.
This internal RH control probe is delivered with its own calibration certifi cate.
RH Control
Probe
Temperature Probe
Figure 3
The RH control probe and the temperature probe must always
Probe Installation
be removed during transportation.
497079 Issue 9.2, March 2016
OptiCal User’s Manual
Control
Probe
Connector
INSTALLATION
Temperature
Probe
Connector
h
h
TOP VIEW
GND
OUTPUT TEMP
Figure 4
+ SUPPLY (8 - 35 VDC)
OUTPUT RH
BOTTOM VIEW
Location of Probes Inside the Climate Chamber
Michell Instruments5
INSTALLATION
2.2 Filling the Water Reservoir
Before operation the water reservoir located on the front panel must be fi lled with
distilled water (supplied with the instrument).
Do not use tap water or demineralized water!
OptiCal User’s Manual
Figure 5
Use the fi lter bottle supplied to fi ll the water reservoir.
1. Remove the red plastic cap from the top of the reservoir.
2. Carefully fi ll with clean distilled water to a level between the two indicator
lines.
3. Replace the red cap on the water reservoir after fi lling.
Do NOT fi ll above the MAX indicated line as this may cause liquid
to enter the humidity chamber and adversely affect the control
Filling the Water Reservoir
process.
697079 Issue 9.2, March 2016
OptiCal User’s Manual
2.3 Draining the Water Reservoir
Drain the water reservoir before transporting, or if the system is accidentally overfi lled.
To empty the water reservoir:
1. Remove the red caps from the bottom and top of the water reservoir.
2. Drain the water into a suitable container.
3. Tilt the OptiCal to empty it completely.
4. Re-fi t the red cap after emptying.
INSTALLATION
Figure 6
It is essential to drain the water reservoir before transporting or
when the system is not going to be used for a few weeks.
Do not re-use any of the drained water within the system.
Emptying the Water Reservoir
Michell Instruments7
INSTALLATION
2.4 Desiccant
The OptiCal has a container fi lled with a desiccant which is used to dry the chamber air.
The desiccant container can be accessed by following these steps:
Figure 7
OptiCal User’s Manual
Desiccant Replacement
1. Remove the clear plastic screw cap on the front panel.
2. Pull out the desiccant container using fi nger tips.
3. Fill with desiccant up to the tapered thread as shown below.
Figure 8
The natural color of the desiccant (dry silica gel) should be orange. As moisture is
absorbed the color will gradually change to transparent.
When the desiccant has become transparent, this is an indication that the instrument’s
ability to generate low humidity is reduced. It is recommended that the silica gel is
replaced or regenerated at this time.
The silica gel can be regenerated by emptying it completely from the desiccant chamber
and drying it in an oven for approximately 3 hours at a temperature between 130°C
(266°F) and 160°C (320°F). After drying it in an oven, allow the silica gel to cool before
refi lling the desiccant chamber and refi tting the chamber into the generator. Ensure that
the screw lid is correctly fi tted.
Level of Silica Gel Required
Frequency of desiccant regeneration or replacement is dependant entirely upon the
length of time in operation. Typically, given a normal calibration profi le and operation
cycle, the OptiCal can continually generate accurate and stable chamber humidity for a
period of several weeks before regeneration or replacement of the silica gel is necessary.
897079 Issue 9.2, March 2016
OptiCal User’s Manual
2.5 Power Supply
A single mains power supply between 85 to 264 V AC is required to operate the unit.
The power supply connection is a 3 pin IEC plug located on the rear panel of the
instrument. The
adjacent to the power socket.
A 3-core power cable is provided, the free end of which should be wired to a suitable
earthed plug or directly connected via a fused power spur. The power cable conductors
are color coded according to the international convention:
Brown L (Live)
Blue N (Neutral)
Green/Yellow E (Earth/GND)
ON/OFF switch and the power input fuse are in the same location,
INSTALLATION
The instrument must be connected to an electrical earth for
safety purposes.
Michell Instruments9
Loading...
+ 36 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.