A complete, compact, portable system for automated validation of relative humidity probes.
Highlights
• Lightweight and completely self-contained
• Intuitive UI makes automating probe verifi cation
simple
• Integral battery pack means validation can be done
without access to services
• Validate 7 probes simultaneously
• Automated validation procedures for complete
hands-off probe verifi cation
• Internal calibration correction cycle ensures
continued confi dence
www.michell.com
Applications
• Validation of RH probes at pharmaceutical
manufacturing facilities
• Validation of RH probes at meteorological offi ces
• Validation of RH probes in food manufacturing
Calibration Instruments
Introducing the HygroCal100 Humidity Validator
The all-in-one validation package for
your humidity sensors
The HygroCal100 provides a stable test chamber to evaluate
the performance of relative humidity sensors across a wide
range of 5 to 95% relative humidity. The intuitive design
allows the probes under test to be fully integrated with the
chamber and user interface (UI), so up to 7 probes with
a variety of diameters and output signals can be powered,
monitored and logged simultaneously by one self-contained
unit.
A 4.3” LCD touch-screen runs a powerful UI, which displays
all measured values from the reference and probes under
test, along with a graphical trend indication of chamber
stability. It is also capable of automating complete validation
procedures with ease, and providing a complete logged
output in csv format straight to your USB memory device, to
minimize the time you spend taking readings.
Portability
The HygroCal100 is fi tted with a high capacity battery pack,
which can power the generator and 7 sensors under test for
up to 8 hours. The unit can also run from mains power while
charging the battery.
A hard carry case is supplied with the unit, which has space
for the HygroCal100 itself , in addition to the battery charger,
spare water and desiccant.
Chamber Integrity
The HygroCal100 has a test chamber milled from a solid
piece of Acetal, with minimal sealing points, ensuring the
integrity required to maintain <5% RH from laboratory
ambient temperatures, and ±0.5% RH uniformity across the
chamber.
Integrate Your Own Reference
The HygroCal interface allows you to assign any hygrometer
with an analog output as your reference device, giving you
the fl exibility to incorporate your traceable ref erence in your
validation routine.
Automated Validation
The HygroCal100’s advancedUI allows you to defi ne your
own calibration procedure, point-by-point, assigning times
to each condition to allow your probes under test to stabilize.
The system always waits until the conditions in the chamber
are completely stable before beginning the check.
Correct Chamber Control To Your Own
Reference
To ensure continual long term stability of the chamber,
the in-built calibration correction system can compare the
readings of your traceable reference to a range of pre-set
generated conditions – making adjustments to the control
sensor to ensure that your set point always matches your
own reference.
www.michell.com
Technology:
Divided Flow Mixing with
HS3 Control
The simple, low maintenance system can transition
between and stabilise on constant humidity conditions
very quickly. It features a reservoir for saturation and a
reservoir for desiccation. By driving ambient air through
either one of these reservoirs and into the chamber, the
conditions inside can be quickly altered.
Reliable, Stable Control
The HygroSmart 3 sensor utilizes the new H8000
capacitive polymer element and high resolution electronic
hybrid technology, to give outstanding accuracy across
the complete RH range, and stability throughout a wide
temperature spectrum.
• Interchangeable sensor accuracy ±0.8% RH
• ±1% RH long-term stability per year
2
C digital interface
• I
The sensor stores its own unique calibration data
within its integral electronics, ensuring 100% fi eld
interchangeability.
HS3
Set Point
Internal Reference Humidity
Humidity
56.9
%RH
Probe 1
56.9
%RH
CHAMBER
55.0%
Temperature
21.0
Temperature
21.0
Calibration Instruments
MINUTES
Responding
00 : 15 : 37
PUMP
13:04
1
Auto
TABLE VIEW
SOLENOID
100
%RH
0
0
ºC
Chamber:
Mode:
Next Point:
CALIBRATION
SETTINGS
Set Point
SYSTEM
SETTINGS
ºC
DESICCANTSATURATOR
Calibration Kits
The HygroCal100 can be supplied with several different combination packages, allowing you to begin making traceable calibr ation
checks immediately.
MDM25
The MDM25 is a portable, battery-operated hand-held hygrometer, which makes an ideal portable reference for use with the
HygroCal100. The MDM25 is supplied with a standard probe, which is calibrated traceable to national standards. Also included
is a port adaptor, to fi t this probe, in the HygroCal100 chamber.
Optidew Vision
The Optidew Vision is a fundamental chilled mirror hygrometer, its high accuracy and repeatability make it the choice for
applications where traceability and precision measurements are key. This kit includes port adaptors to fi t both the dew-point
and temperature sensors into the chamber.
www.michell.com
Calibration Instruments
Technical Specifi cations
Chamber
Generation range5 to 95% RH
RH stability±0.5%
RH uniformity±0.5%
Stabilization timeTypically <5 min for full stability from step
changes of 10% RH
Control Probe
Accuracy±0.8%
Long term stability±1% per year
Electrical Specifi cations
User interface4.3” color LCD with touchscreen
Measurement units %RH, temperature in °C, °F
Displayed resolution0.1
Data logging2Gb internal memory available for log fi les;
Battery 1500 mAh
Power supply24 V DC (100 to 240 V AC, 50/60 Hz adaptor
or 10.6yrs storage at 5s intervals
included)
Mechanical Specifi cations
Probe ports8 – sensor body diameters 5 to 25mm,
Chamber volumeApprox 1050cm
Maximum probe insertion
depth
Desiccant reservoir
capacity
Saturator reservoir
capacity
Environmental conditions+5 to +40°C
Dimensions100 x 250 x 300mm (h x w x d)
Weight3.2kg
accommodated by port adaptors
3
60mm
3
25cm
25ml
Dimensions
300mm
100mm
250mm
Related Products
OptiCal
Humidity Calibrator
Michell Instruments Ltd 48 Lancaster Way Business Park, Ely, Cambridgeshire, CB6 3NW
Tel: +44 (0) 1353 658000, Fax: +44 (0) 1353 658199, Email: uk.info@michell.com, Web: www.michell.com/uk
Michell Instruments adopts a continuous development programme which sometimes necessitates specifi cation changes without notice.
Issue no: HygroCal100_97502_V1_UK_Datasheet_0915