Appendix D Decontamination Declaration ...................................................................... 23
iv97120 Issue 2.2, March 2016
Easidew Portable HP User’s Manual
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures
detailed in this manual. The user must not use this equipment for any other purpose than that
stated. Do not apply values greater than the maximum value stated.
This manual contains operating and safety instructions, which must be followed to ensure the safe
operation and to maintain the equipment in a safe condition. The safety instructions are either
warnings or cautions issued to protect the user and the equipment from injury or damage. Use
qualifi ed personnel and good engineering practice for all procedures in this manual.
Electrical Safety
The instrument is designed to be completely safe when used with options and accessories supplied by
the manufacturer for use with the instrument. The instrument is powered by an internally mounted
rechargeable battery. The input power supply voltage limits for the battery charger supplied with the
instrument are 90 to 264 V AC, 47/63 Hz.
NOTE: No other battery charger unit, other than that supplied with the instrument
should be used.
NOTE: Do not allow the battery to fully discharge.
Pressure Safety
DO NOT permit pressures greater than the safe working pressure to be applied to the instrument.
The specifi ed safe working pressure is 300 barg (4350 psig). Refer to the Technical Specifi cations in
Appendix A.
Toxic Materials
The use of hazardous materials in the construction of this instrument has been minimized. During
normal operation it is not possible for the user to come into contact with any hazardous substance
which might be employed in the construction of the instrument. Care should, however, be exercised
during maintenance and the disposal of certain parts.
Repair and Maintenance
The instrument must be maintained either by the manufacturer or an accredited service agent. Refer
to www.michell.com for details of Michell Instruments’ worldwide offi ces contact information.
Calibration
The recommended calibration interval for the Easidew Portable is one year, unless otherwise
specifi ed by Michell Instruments Ltd. The instrument should be returned to the manufacturer, Michell
Instruments, or one of their accredited service agents for re-calibration (go to www.michell.com for
contact information).
Safety Conformity
This product meets the essential protection requirements of the relevant EU directives. Further
details of applied standards may be found in the product specifi cation.
Michell Instrumentsv
Abbreviations
!
The following abbreviations are used in this manual:
AC alternating current
atm pressure unit (atmosphere)
barg pressure unit (=100 kP or 0.987 atm) gauge
°C degrees Celsius
°F degrees Fahrenheit
dp dew point
ft feet
kg kilogram
lb pound
Easidew Portable HP User’s Manual
mA milliampere
m meter
mm millimeter
MPa megapascal
Nl/min normal liters per minute
psig pound(s) per square inch (gauge)
scfh standard cubic feet per hour
μm micrometer
Ω ohm
V Volts
Warnings
The following general warnings listed below are applicable to this instrument. They are
repeated in the text in the appropriate locations.
Where this hazard warning symbol appears in the following
sections, it is used to indicate areas where potentially hazardous
operations need to be carried out.
vi97120 Issue 2.2, March 2016
Easidew Portable HP User’s Manual
1 INTRODUCTION
The Easidew Portable comprises an Easidew sensor fi tted into a sample block with an
integrated fi lter cartridge, and Swagelok® tube couplings on the gas inlet and outlet.
The measured dew point from the sensor is shown on the clear red LED display on the
front panel of the instrument.
The Easidew Portable can be supplied with either °C or °F dew-point measurement
units.
A 4-20 mA analog output is provided for connection to a chart recorder, data-logger or
computer system, so dew-point trends can be analyzed over time.
The electronics are housed in a rugged Peli case, providing NEMA 6 protection when the
lid is closed. The case is supplied with a lifetime guarantee.
INTRODUCTION
The Easidew Portable is powered by a rechargeable NiMH battery pack, giving between
12 and 16 hours of measurement time from a full charge. The instrument is delivered
complete with a universal battery charger stored in the lid. It takes 16 hours to fully
charge the battery pack, during which time the instrument can be switched on or
off. The charger is suitable to connect power to the instrument indefi nitely, but it is
recommended to allow the battery pack to go through a full charge-discharge cycle at
least once per month. A battery charge indicator on the instrument front panel warns
when the battery is low.
All Michell Ceramic Moisture Sensors are subject to a 13-point calibration, where their
performance is characterized against a fundamental reference hygrometer. This process,
and subsequent quality testing, ensures that all sensors behave optimally before they
are used in the fi eld. A calibration certifi cate traceable to UK National Physical Laboratory
(NPL) and the US National Institute of Standards and Technology (NIST) is provided
from Michell Instruments' accredited laboratory. The certifi cate shows 7 points of the
13-point calibration.
Michell Instruments1
INSTALLATION
2 INSTALLATION
On delivery, please check that all the following standard components are present in the
packing box:
• Easidew Portable
®
• Swagelok
• Swagelok
• Battery charger
• Mains lead
• 3 pole ¼” jack plug (inside lid)
• User manual
• Certifi cate of calibration
6mm metering valve
®
6mm tube port adaptor
Easidew Portable HP User’s Manual
Before using the Easidew for the fi rst time it is recommended that the battery pack is
charged for a minimum of 12 hours.
The case has a pressure relief valve on the exterior below the handle. When closed,
the case is completely sealed. If the instrument is taken through altitude changes, a
pressure differential will occur within the case. If this occurs: Before opening the lid,
open the black pressure relief valve by one turn (anti-clockwise) for a few seconds. This
will equalize the pressures. When the case has been opened, ensure that the pressure
relief valve has been fully closed.
®
The supplied Swagelok
port adaptor needs to be assembled before the metering valve
can be connected to the instrument. See below:
1. Slip the nut over the
machined ferrule end of
the port connector
2. Insert the port connector
into the end connection and
finger-tighten the nut
3. While holding the fitting
body steady, tighten the
nut one-quarter turn
Do not use the Swagelok gap inspection
!
gauge with machined ferrule ends.
297120 Issue 2.2, March 2016
Easidew Portable HP User’s Manual
NOTE: Connect the machined ferrule end before
connecting the tube adapter end
INSTALLATION
4. Insert the tube adapter until it rests firmly on
the shoulder of the Swagelok
®
tube fitting body.
Finger-tighten the nut.
5. Mark the nut at the 6 o’clock position. While holding
the fitting body steady, tighten the nut 1¼ turns to the
9 o’clock position.
Maximum operating pressure is 300 barg (4350 psig).
It is important that both the gas fi ttings and the Swagelok
Failure to do so will affect the instrument's pressure rating.
3.1 General Operation
To operate the Easidew Portable, follow these instructions:
1. Ensure that both the gas fi ttings and the Swagelok® tube fi ttings are fully
tightened before use.
2. For measurement at line pressure, connect the supplied metering valve to
the outlet of the instrument. For measurement at atmospheric pressure,
connect the metering valve to the inlet.
3. Gas under test should be restricted to fl ow rates between 1 and 5 Nl/min
(2 and 10.6 scfh) and pressures from atmospheric to 30 MPa (300 barg
/ 4350 psig).
4. Connect the sample gas supply line to the instrument
tube fi ttings are correctly tightened before use.
Gas In port.
®
5. Connect the sample gas vent line to the instrument
6. Switch the instrument on. The display will begin to change as the sensor
responds to the applied dew point.
7. Allow the gas to fl ow until the display shows a stable reading. Typically
this would be around 15 to 30 minutes for spot checks at dew points of
-40°C (-40°F) and above.
8. Switch the unit off and disconnect the sample lines.
Before disconnecting the Easidew Portable from the gas line
it is essential to vent the system to atmospheric pressure,
otherwise severe injury could result.
NOTE: Filters are essential for potentially dirty/contaminated gases – the
fi tted fi lter should be checked before and after use and replaced regularly - as
required.
NOTE: The High Pressure version features two gas fi ttings with 4mm orifi ces
that give no restriction and are engraved with a number 1. Atmospheric or
line pressure measurement is achieved by changing the positioning of the
metering valve as per the General Operation instructions above.
Gas Out port.
497120 Issue 2.2, March 2016
Loading...
+ 22 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.