Michell Instruments Easidew Portable HP User Manual

Page 1
Easidew Portable
High Pressure Version
User’s Manual
97120 Issue 2.2
March 2016
Page 2
Please fi ll out the form(s) below for each instrument that has been purchased.
Use this information when contacting Michell Instruments for service purposes.
Analyzer
Serial Number
Invoice Date
Location of Instrument
Tag No
Analyzer
Serial Number
Invoice Date
Location of Instrument
Tag No
Analyzer
Serial Number
Invoice Date
Location of Instrument
Tag No
Page 3
Easidew Portable HP
For Michell Instruments' contact information please go to
www.michell.com
© 2016 Michell Instruments
This document is the property of Michell Instruments Ltd. and may not be copied or
otherwise reproduced, communicated in any way to third parties, nor stored in any Data
Processing System without the express written authorization of Michell Instruments Ltd.
Page 4
Easidew Portable HP User’s Manual
Contents
Safety ................................................................................................................................ v
Electrical Safety ........................................................................................................... v
Pressure Safety ............................................................................................................ v
Toxic Materials ............................................................................................................. v
Repair and Maintenance ............................................................................................... v
Calibration ................................................................................................................... v
Safety Conformity ........................................................................................................ v
Abbreviations ......................................................................................................................vi
Warnings ............................................................................................................................vi
1 INTRODUCTION ................................................................................................1
2 INSTALLATION ..................................................................................................2
3 OPERATION ......................................................................................................4
3.1 General Operation .............................................................................................. 4
3.2 Measuring Dew Points Below -40°C (-40°F) .......................................................... 5
3.3 User Controls ..................................................................................................... 6
4 GOOD MEASUREMENT PRACTICE .......................................................................7
5 MAINTENANCE ................................................................................................11
5.1 Calibration ....................................................................................................... 11
5.2 Sensor / Battery Replacement ........................................................................... 11
5.3 Filter Cartridge Replacement ............................................................................. 13
5.4 Checking the Easidew Portable Electronics Calibration ......................................... 13
5.5 List of Spares ................................................................................................... 13
5.6 Troubleshooting ................................................................................................ 14
Figures
Figure 1 Swagelok® Fitting Instructions .....................................................................3
Figure 2 User Controls .............................................................................................6
Figure 3 2-Wire Connection - View Showing Rear of Connector Terminal Block ...........13
Appendices
Appendix A Technical Specifi cations .............................................................................. 16
Appendix B EU Declaration of Conformity......................................................................18
Appendix C Quality, Recycling & Warranty Information ...................................................20
C.1 Recycling Policy ........................................................................... 20
C.2 WEEE And RoHS Compliance ........................................................ 20
C.3 Manufacturing Quality ..................................................................20
C.4 Calibration Facilities ..................................................................... 21
C.5 Return Policy ...............................................................................21
C.6 Warranty ..................................................................................... 21
Appendix D Decontamination Declaration ...................................................................... 23
iv 97120 Issue 2.2, March 2016
Page 5
Easidew Portable HP User’s Manual
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures detailed in this manual. The user must not use this equipment for any other purpose than that stated. Do not apply values greater than the maximum value stated.
This manual contains operating and safety instructions, which must be followed to ensure the safe operation and to maintain the equipment in a safe condition. The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage. Use qualifi ed personnel and good engineering practice for all procedures in this manual.
Electrical Safety
The instrument is designed to be completely safe when used with options and accessories supplied by the manufacturer for use with the instrument. The instrument is powered by an internally mounted rechargeable battery. The input power supply voltage limits for the battery charger supplied with the instrument are 90 to 264 V AC, 47/63 Hz.
NOTE: No other battery charger unit, other than that supplied with the instrument should be used.
NOTE: Do not allow the battery to fully discharge.
Pressure Safety
DO NOT permit pressures greater than the safe working pressure to be applied to the instrument. The specifi ed safe working pressure is 300 barg (4350 psig). Refer to the Technical Specifi cations in Appendix A.
Toxic Materials
The use of hazardous materials in the construction of this instrument has been minimized. During normal operation it is not possible for the user to come into contact with any hazardous substance which might be employed in the construction of the instrument. Care should, however, be exercised during maintenance and the disposal of certain parts.
Repair and Maintenance
The instrument must be maintained either by the manufacturer or an accredited service agent. Refer to www.michell.com for details of Michell Instruments’ worldwide offi ces contact information.
Calibration
The recommended calibration interval for the Easidew Portable is one year, unless otherwise specifi ed by Michell Instruments Ltd. The instrument should be returned to the manufacturer, Michell Instruments, or one of their accredited service agents for re-calibration (go to www.michell.com for contact information).
Safety Conformity
This product meets the essential protection requirements of the relevant EU directives. Further details of applied standards may be found in the product specifi cation.
Michell Instruments v
Page 6
Abbreviations
!
The following abbreviations are used in this manual:
AC alternating current
atm pressure unit (atmosphere)
barg pressure unit (=100 kP or 0.987 atm) gauge
°C degrees Celsius
°F degrees Fahrenheit
dp dew point
ft feet
kg kilogram
lb pound
Easidew Portable HP User’s Manual
mA milliampere
m meter
mm millimeter
MPa megapascal
Nl/min normal liters per minute
psig pound(s) per square inch (gauge)
scfh standard cubic feet per hour
μm micrometer
Ω ohm
V Volts
Warnings
The following general warnings listed below are applicable to this instrument. They are repeated in the text in the appropriate locations.
Where this hazard warning symbol appears in the following
sections, it is used to indicate areas where potentially hazardous
operations need to be carried out.
vi 97120 Issue 2.2, March 2016
Page 7
Easidew Portable HP User’s Manual
1 INTRODUCTION
The Easidew Portable comprises an Easidew sensor fi tted into a sample block with an integrated fi lter cartridge, and Swagelok® tube couplings on the gas inlet and outlet. The measured dew point from the sensor is shown on the clear red LED display on the front panel of the instrument.
The Easidew Portable can be supplied with either °C or °F dew-point measurement units.
A 4-20 mA analog output is provided for connection to a chart recorder, data-logger or computer system, so dew-point trends can be analyzed over time.
The electronics are housed in a rugged Peli case, providing NEMA 6 protection when the lid is closed. The case is supplied with a lifetime guarantee.
INTRODUCTION
The Easidew Portable is powered by a rechargeable NiMH battery pack, giving between 12 and 16 hours of measurement time from a full charge. The instrument is delivered complete with a universal battery charger stored in the lid. It takes 16 hours to fully charge the battery pack, during which time the instrument can be switched on or off. The charger is suitable to connect power to the instrument indefi nitely, but it is recommended to allow the battery pack to go through a full charge-discharge cycle at least once per month. A battery charge indicator on the instrument front panel warns when the battery is low.
All Michell Ceramic Moisture Sensors are subject to a 13-point calibration, where their performance is characterized against a fundamental reference hygrometer. This process, and subsequent quality testing, ensures that all sensors behave optimally before they are used in the fi eld. A calibration certifi cate traceable to UK National Physical Laboratory (NPL) and the US National Institute of Standards and Technology (NIST) is provided from Michell Instruments' accredited laboratory. The certifi cate shows 7 points of the 13-point calibration.
Michell Instruments 1
Page 8
INSTALLATION
2 INSTALLATION
On delivery, please check that all the following standard components are present in the packing box:
• Easidew Portable
®
• Swagelok
• Swagelok
• Battery charger
• Mains lead
3 pole ¼” jack plug (inside lid)
• User manual
• Certifi cate of calibration
6mm metering valve
®
6mm tube port adaptor
Easidew Portable HP User’s Manual
Before using the Easidew for the fi rst time it is recommended that the battery pack is charged for a minimum of 12 hours.
The case has a pressure relief valve on the exterior below the handle. When closed, the case is completely sealed. If the instrument is taken through altitude changes, a
pressure differential will occur within the case. If this occurs: Before opening the lid,
open the black pressure relief valve by one turn (anti-clockwise) for a few seconds. This will equalize the pressures. When the case has been opened, ensure that the pressure relief valve has been fully closed.
®
The supplied Swagelok
port adaptor needs to be assembled before the metering valve
can be connected to the instrument. See below:
1. Slip the nut over the machined ferrule end of the port connector
2. Insert the port connector into the end connection and finger-tighten the nut
3. While holding the fitting body steady, tighten the nut one-quarter turn
Do not use the Swagelok gap inspection
!
gauge with machined ferrule ends.
2 97120 Issue 2.2, March 2016
Page 9
Easidew Portable HP User’s Manual
NOTE: Connect the machined ferrule end before connecting the tube adapter end
INSTALLATION
4. Insert the tube adapter until it rests firmly on the shoulder of the Swagelok
®
tube fitting body.
Finger-tighten the nut.
5. Mark the nut at the 6 o’clock position. While holding the fitting body steady, tighten the nut 1¼ turns to the 9 o’clock position.
Figure 1
Swagelok® Fitting Instructions
© 2001-2010 Swagelok Company September 2010, R4 MS-12-12
Michell Instruments 3
Page 10
OPERATION
!
!
3 OPERATION
Easidew Portable HP User’s Manual
Maximum operating pressure is 300 barg (4350 psig).
It is important that both the gas fi ttings and the Swagelok
Failure to do so will affect the instrument's pressure rating.
3.1 General Operation
To operate the Easidew Portable, follow these instructions:
1. Ensure that both the gas fi ttings and the Swagelok® tube fi ttings are fully tightened before use.
2. For measurement at line pressure, connect the supplied metering valve to the outlet of the instrument. For measurement at atmospheric pressure, connect the metering valve to the inlet.
3. Gas under test should be restricted to fl ow rates between 1 and 5 Nl/min (2 and 10.6 scfh) and pressures from atmospheric to 30 MPa (300 barg / 4350 psig).
4. Connect the sample gas supply line to the instrument
tube fi ttings are correctly tightened before use.
Gas In port.
®
5. Connect the sample gas vent line to the instrument
6. Switch the instrument on. The display will begin to change as the sensor responds to the applied dew point.
7. Allow the gas to fl ow until the display shows a stable reading. Typically this would be around 15 to 30 minutes for spot checks at dew points of
-40°C (-40°F) and above.
8. Switch the unit off and disconnect the sample lines.
Before disconnecting the Easidew Portable from the gas line
it is essential to vent the system to atmospheric pressure,
otherwise severe injury could result.
NOTE: Filters are essential for potentially dirty/contaminated gases – the tted fi lter should be checked before and after use and replaced regularly - as required.
NOTE: The High Pressure version features two gas fi ttings with 4mm orifi ces that give no restriction and are engraved with a number 1. Atmospheric or line pressure measurement is achieved by changing the positioning of the metering valve as per the General Operation instructions above.
Gas Out port.
4 97120 Issue 2.2, March 2016
Page 11
Easidew Portable HP User’s Manual
3.2 Measuring Dew Points Below -40°C (-40°F)
Due to the signifi cantly lower levels of moisture present at dew points of this level, the response times of the sensor will be signifi cantly increased. The table below offers an approximate guide to the times taken for the instrument to stabilize at a given dew point (from a starting point of 10°Cdp (50°Fdp) ambient):
OPERATION
Target Dew Point
°C °F
-30 -22 15
-40 -40 30
-50 -58 60
-60 -76 180
The Easidew sensor fi tted to the Easidew Portable is designed and calibrated to measure dew points as low as -100°C (-148°F). However, due to the large internal volume of the sample block, the hygroscopic nature of the built-in fi lter, and the potential of the gas fi ttings to trap moisture at very low dew points, we recommend that the Easidew Portable is not used to measure dew points below -60°C (-76°F). Suitable alternative solutions in our range would be the Easidew Online, Pura fi xed installations, MDM300 Portable, or our cooled mirror range of instruments.
T100 Response Time
(mins)
Michell Instruments 5
Page 12
OPERATION
3.3 User Controls
Easidew Portable HP User’s Manual
2
1
EASIDEW Portable Hygrometer
h
Power Fuse
°Cdp
4-20
mA
O/P
BATTERY
Battery
Level
34
h
h
o
-
h
Battery Charger
h
h
h
5
6
Gas Out
Figure 2
§
User Controls
Filter Fitted
Indicates the measured dew point in °C from -100 to +20 (-148 to +68°F). Under certain conditions the following error messages may be
1 Digital Display
displayed:
ErrL = Sensor under range ErrH = Sensor over range Err I = Sensor fault or sensor disconnected
2 Power Switch Switches the Easidew Portable ON or OFF.
This 1A quick blow fuse provides protection for the display
3 Fuse
electronics in the event of a fault with the charger or battery pack. Another protection fuse is located on the display PCB.
Located in the top right hand corner of the instrument, it can be accessed by removing two screws (5). The battery pack can be recharged using the supplied charger, via the battery-charging
NiMH Battery
4
Pack
socket and should be recharged whenever the battery level meter is in the red region.
The battery pack will charge if the instrument is switched ON or OFF. However, the battery level meter will only indicate when the instrument is switched on.
The Easidew portable features an analog output socket that
Analog Output
6
Socket
provides a linear 4-20 mA current loop, scaled to -100 to +20°C (-148 to +68°F) dew point. This allowed the instrument to be connected to a chart recorder, data logger or PC to enable dew­point trends to be analyzed over time.
The Easidew Portable can be used for measurements at line pressure (up to 300 barg (4350 psig)) or atmospheric pressure.
®
tube fi ttings are supplied for use with 6mm OD
7 Gas Fittings
Swagelok stainless steel tube. There is a 32mm particulate fi lter (99.5% removal of 0.1 micron particles) fi tted as standard under the
Gas In port position.
§
h
Gas In
7
6 97120 Issue 2.2, March 2016
Page 13
Easidew Portable HP User’s Manual
4 GOOD MEASUREMENT PRACTICE
Measurement of moisture content is a complex subject, but does not need to be diffi cult. This section aims to explain the common mistakes made in measurement situations, the causes of the problem, and how to avoid them. Mistakes and bad practices can cause the measurement to vary from the expectation; therefore a good sampling technique is crucial for accurate and reliable results.
Transpiration and Sampling Materials
- 20
- 30
GOOD MEASUREMENT PRACTICE
- 40
- 50
Dew point (ºC)
- 60
- 70
12345
nickel
stainless steel
Time (hours)
nylon
copper
polyethylene
PTFE
All materials are permeable to water vapor, as the water molecule is extremely small compared to the structure of solids, even when compared to the crystalline structure of metals. The graph to the right shows the dew point inside tubing of different materials when purged with very dry gas, where the exterior of the tubing is in the ambient environment.
Many materials contain moisture as part of their structure, particularly organic materials (natural or synthetic), salts (or anything which contains them) and anything which has small pores. It is important to ensure that the materials used are suitable for the application.
If the partial water vapor pressure exerted on the outside of a compressed air line is higher than on the inside, the atmospheric water vapor will naturally push through the porous medium causing water to migrate into the pressurized air line. This effect is called transpiration.
Michell Instruments 7
Page 14
GOOD MEASUREMENT PRACTICE
Adsorption and Desorption
Adsorption is the adhesion of atoms, ions, or molecules from a gas, liquid, or dissolved solid to the surface of a material, creating a fi lm. The rate of adsorption is increased at higher pressures and lower temperatures.
Desorption is the release of a substance from or through the surface of a material. In constant environmental conditions, an adsorbed substance will remain on a surface almost indefi nitely. However, as the temperature rises, so does the likelihood of desorption occurring.
In practical terms, as the temperature of the environment fl uctuates, water molecules are adsorbed and desorbed from the internal surfaces of the sample tubing, causing small fl uctuations in the measured dew point.
Easidew Portable HP User’s Manual
Sample Tubing Length
The sample point should always be as close to the critical measurement point as possible, in order to obtain a truly representative measurement. The length of the sample line to the sensor or instrument should be as short as possible. Interconnection points and valves trap moisture, so using the simplest sampling arrangement possible will reduce the time it takes for the sample system to dry out when purged with dry gas.
Over a long tubing run, water will inevitably migrate into any line, and the effects of adsorption and desorption will become more apparent. It is clear from the graph shown above that the best materials to resist transpiration are stainless steel and PTFE.
Trapped Moisture
Dead volumes (areas which are not in a direct fl ow path) in sample lines, hold onto water molecules which are slowly released into the passing gas; this results in increased purge and response times, and wetter than expected readings. Hygroscopic materials in fi lters, valves (e.g. rubber from pressure regulators) or any other parts of the system can also trap moisture.
8 97120 Issue 2.2, March 2016
Page 15
Easidew Portable HP User’s Manual
Sample Conditioning
Sample conditioning is often necessary to avoid exposure of sensitive measuring components to liquids and other contaminants which may cause damage or affect the accuracy over time, depending on the measurement technology.
Particulate fi lters are used for removing dirt, rust, scale and any other solids that may be in a sample stream. For protection against liquids, a coalescing fi lter should be used. The membrane fi lter is a more expensive but highly effective alternative to a coalescing lter. It provides protection from liquid droplets, and can even stop fl ow to the analyzer completely when a large slug of liquid is encountered.
Condensation and Leaks
GOOD MEASUREMENT PRACTICE
Dewpoint > T Dewpoint < T
Maintaining the temperature of the sample system tubing above the dew point of the sample is vital to prevent condensation. Any condensation invalidates the sampling process as it changes the water vapor content of the gas being measured. Condensed liquid can alter the humidity elsewhere by dripping or running to other locations where it may re-evaporate.
The integrity of all connections is also an important consideration, especially when sampling low dew points at an elevated pressure. If a small leak occurs in a high pressure line, gas will leak out but vortices at the leak point and a negative vapor pressure differential will also allow water vapor to contaminate the fl ow.
Michell Instruments 9
Page 16
GOOD MEASUREMENT PRACTICE
!
!
Flow Rates
Theoretically fl ow rate has no direct effect on the measured moisture content, but in practice it can have unanticipated effects on response speed and accuracy. The optimal ow rate varies depending on the measurement technology, and can always be found in the instrument or sensor manual.
An inadequate fl ow rate can:
Accentuate adsorption and desorption effects on the gas passing through the sampling system.
Allow pockets of wet gas to remain undisturbed in a complex sampling system, which will then gradually be released into the sample fl ow.
Increase the chance of contamination from back diffusion: ambient air that is wetter than the sample can fl ow from the exhaust back into the system. A longer exhaust (sometimes called a pigtail) can also help alleviate this problem.
Easidew Portable HP User’s Manual
Slow the response of the sensor to changes in moisture content.
An excessively high fl ow rate can:
Introduce back pressure, causing slower response times and unpredictable effects on equipment such as humidity generators.
Result in a reduction in heating capabilities of the sensor tile during the initialization period. This is most apparent with gases that have a high thermal conductivity such as hydrogen and helium.
Which Gases to Measure?
The Easidew Portable is suitable for measurement of the moisture content of a wide variety of gases. In general, if the gas (in conjunction with water vapor) is not corrosive to ceramics or base metals then it will be suitable for measurement.
POSSIBLE INJURY! The tubing, valves and other apparatus
attached to this instrument must be adequate for the
maximum pressure which will be applied, otherwise physical
injury to the operator or bystander is possible.
Before disconnecting the Easidew Portable from the gas line
it is essential to vent the system to atmospheric pressure,
otherwise severe injury could result.
10 97120 Issue 2.2, March 2016
Page 17
Easidew Portable HP User’s Manual
5 MAINTENANCE
Routine maintenance of the Easidew Portable is confi ned to regular re-calibration of the internal, removable Easidew sensor and replacement of the fi lter cartridge.
5.1 Calibration
The calibration of the internal sensor is traceable to national standards. For this reason it should only be calibrated in an accredited, e.g. UK
Service (UKAS)
laboratory. If these facilities do not exist it is recommended that the sensor is returned to the manufacturer, Michell Instruments, or an approved agent. A calibration certifi cate bearing a seven-point calibration is issued with each sensor.
In most applications, annual re-calibration ensures that the stated accuracy of the Easidew sensor is maintained.
or US
National Institute of Standards and Technology
MAINTENANCE
United Kingdom Accreditation
(NIST) standards
5.2 Sensor / Battery Replacement
To remove the sensor, or battery pack:
1. Ensure instrument is switched off and that the battery charger, current output and any sampling components are also disconnected.
2. Remove and retain the 10 cross-head screws from the top plate.
3. Lift the small battery pack cover in the top right corner of the instrument. A small fl at bladed screwdriver may be required to gently pry the edges of the cover if it has become stuck to the waterproof seal underneath. Removing this cover fi rst will make it easier to lift the entire top panel out of the instrument.
4. Lift the top plate of the instrument out of the case. Disconnect the battery pack before removing the top plate completely to prevent it from straining on the attached cable. If only replacing the battery pack, do not follow the next two steps.
5. Undo the screw from the centre of the sensor connecting plug and pull off the connector.
6. Unscrew the sensor from the block.
Fitting the replacement is simply a reversal of the above procedure. When fi tting a new sensor it may be necessary to change the alignment of the GDSN connector (see next page).
Michell Instruments 11
Page 18
MAINTENANCE
If the front panel prevents the connector from fi tting onto the sensor, follow the instructions below:
Completely unscrew the cable gland on the GDSN connector to release the cable tension.
Remove the locking screw from the back of the connector (retaining the metal O-ring).
Easidew Portable HP User’s Manual
Remove the connector block using a small screwdriver.
Rotate the connector block and push it back into the connector housing. Take care not to trap any wires against the hole for the locking screw.
X
12 97120 Issue 2.2, March 2016
Page 19
Easidew Portable HP User’s Manual
Replace the locking screw and cable gland.
5.3 Filter Cartridge Replacement
Unscrew the gas inlet fi tting to reveal the fi lter, which can then simply be removed for checking or replacement.
MAINTENANCE
5.4 Checking the Easidew Portable Electronics Calibration
Supply 4-20 mA
h
PIN 1
PIN 3
Figure 3
To verify the display electronics are within calibration, a 4-20mA current source can be connected in place of the sensor.
At 4 mA the display should show -100.0°Cdp ±0.5°C (-148°Fdp ±0.9°F)
At 20 mA the display should show +20.0°Cdp ±0.5°C ( +68°Fdp ±0.9°F)
If the displayed values are outside of this range, please contact Michell Instruments' Service Department.
2-Wire Connection - View Showing Rear of Connector Terminal Block
h
(GND)
Screen/Shield
Optional
5.5 List of Spares
P/N Description
SSF-PF-10PK Pack of 10 particulate fi lter cartridges EAP-BAT Replacement battery pack EAP-CHR Replacement battery charger
Michell Instruments 13
Page 20
MAINTENANCE
5.6 Troubleshooting
Symptom Cause Actions
Display shows “
Display shows “
ErrI
ErrL
Easidew Portable HP User’s Manual
Sensor disconnected or sensor tile / sensor thermistor fault.
NOTE: This error is normally displayed during the fi rst few seconds while the instrument is starting up.
Sensor reading under­range.
Check sensor cable is securely connected inside instrument.
Check instrument electronics with 4-20mA source.
Exchange sensor.
Check instrument electronics with 4-20mA source.
Display shows “
Display fl ickers on and off.
Display off but battery meter full.
ErrH
Sensor current signal < 4mA
Sensor tile open circuit. Sensor reading over-
range.
Sensor current signal > 20mA
Sensor tile short circuit.
Battery voltage low. Connect charger.
Battery voltage low. Connect charger.
Exchange sensor.
Sensor may have been exposed to saturation conditions or liquid contamination. Check that fi lter and sensor guard are clean and dry.
If liquid water has contacted sensor, or sensor block, then disassemble and dry thoroughly. After drying, it is recommend to purge assembled instrument with very dry (-75°Cdp (-103°Fdp) / 1ppm 12 to 24 hours.
Exposure to other contaminants can cause lasting damage and may require sensor to be exchanged.
Check instrument electronics with 4-20mA source.
moisture) air for
V
14 97120 Issue 2.2, March 2016
Page 21
Easidew Portable HP User’s Manual
APPENDIX A
Appendix A
Technical Speci cations
Michell Instruments 15
Page 22
APPENDIX A
Easidew Portable HP User’s Manual
Appendix A Technical Specifi cations
Performance
Measurement Technology Impedance Measurement Ranges
Spot Checks Online
Accuracy ±2°Cdp (±3.6°Fdp) Run time 12 - 16 hours Charge time 16 hours for maximum charge Flow Rate 1 to 5 Nl/min (2.1 to 10.5 scfh)
-40 to +20°Cdp (-40 to +68°Fdp)
-100 to -40°Cdp (-148 to -40°Fdp)
Electrical Input/Output
Output
Power Supply Rechargeable NiMH battery pack, charger included
4-20 mA current maximum load resistance 400 Ω
Operating Conditions
Operating Temperature -20 to +50°C (-4 to +122°F) Storage Temperature -40 to +75°C (-40 to +167°F) Operating Pressure Up to 30 MPa (300 barg / 4350 psig)
Mechanical Specifi cations
Display Flush mounted 3.5 digit red LED
Case
Weight 4kg (8.8lbs) total weight Enclosure Rating
Case Closed Sample Connections 6mm Swagelok
Sample Block
Filter Cartridge
Interchangeability Sensor is fully interchangeable with other Easidew products
Yellow propylene with charger and output connector stored in the lid
NEMA Type 6
®
stainless steel tube fi ttings
Stainless steel, fully self-contained sample system with fi xed orifi ce ports for fl ow control/pressure or atmospheric measurement and built-in fi ltration using a standard drop-in cartridge
Removes 99.5% of particles 0.3 μm supplied with cartridge installed. Spare cartridges are available (part no: SSF-PF-10PK)
16 97120 Issue 2.2, March 2016
Page 23
Easidew Portable HP User’s Manual
APPENDIX B
Appendix B
EC Declaration of Conformity
Michell Instruments 17
Page 24
APPENDIX B
Appendix B EU Declaration of Conformity
Easidew Portable HP User’s Manual
18 97120 Issue 2.2, March 2016
Page 25
Easidew Portable HP User’s Manual
APPENDIX C
Appendix C
Quality, Recycling
& Warranty
Information
Michell Instruments 19
Page 26
APPENDIX C
Easidew Portable HP User’s Manual
Appendix C Quality, Recycling & Warranty Information
C.1 Recycling Policy
Michell Instruments is concerned with the protection of the environment. It is our commitment to reduce and eliminate from our operations, wherever possible, the use of substances which may be harmful to the environment. Similarly, we are increasingly using recyclable and/or recycled material in our business and products wherever it is practical to do so.
To protect natural resources and to promote material reuse, please separate batteries from other types of waste and recycle responsibly. If batteries are not properly disposed of, these substances can cause harm to human health and the environment
The product that you have purchased may contain recyclable and/or recycled parts and we will be happy to provide you with information on these components if required.
C.2 WEEE And RoHS Compliance
The Waste Electronic and Electrical Equipment (WEEE) Directive, and the Restriction of Hazardous Substances (RoHS) Directive place rules upon European manufacturers of electrical and electronic equipment. The directives’ aim is to reduce the impact that electronic devices have on the environment.
Michell products are currently exempt from the RoHS directive, however all future products will be developed entirely using compliant materials. Furthermore, Michell is taking active steps to remove non-compliant materials and components from existing products wherever possible.
Michell is in full compliance with the WEEE Directive (Registration No. WEE/JB0235YW). Customers may be required to return certain instruments for treatment at the end of their working life.
June 2010
C.3 Manufacturing Quality
Michell Instruments is registered with the British Standards Institute for Quality Assurance to:
BS EN ISO 9001: 2008
Rigorous procedures are performed at every stage of production to ensure that the materials of construction, manufacturing, calibration and fi nal test procedures meet the requirements laid down by our BSI approved Quality System.
Please contact Michell Instruments (www.michell.com) if the product does not arrive in perfect working order.
20 97120 Issue 2.2, March 2016
Page 27
Easidew Portable HP User’s Manual
C.4 Calibration Facilities
Michell Instruments’ calibration facilities are among the most sophisticated in the world and have been recognized for their excellence.
Traceability to the National Physical Laboratory (NPL) UK is achieved through our UKAS Accreditation (Number 0179). This covers dew point over the range -90 to +90°C (-130 to +194°F) and also Relative Humidity.
Dew-point calibrations are also traceable to the National Institute for Standards & Technology (NIST) USA over the range -75 to +20°C (-103 to +68°F).
NOTE: Standard traceable calibration certifi cates for instruments and sensors are not
issued under our UKAS accreditation. UKAS certifi cates are usually to special order and are clearly identifi ed.
C.5 Return Policy
If a Michell Instruments’ product malfunctions within the warranty period, the following procedure must be completed:
APPENDIX C
1. Notify a Michell Instruments’ distributor, giving full details of the problem, the model variant and the serial number of the product.
2. If the nature of the problem indicates the need for factory service then the instrument should be returned to Michell Instruments, carriage prepaid, preferably in the original packaging, with a full description of the fault and the customer contact information.
3. Upon receipt, Michell Instruments will evaluate the product to determine the cause of the malfunction. Then, one of the following courses of action will be taken:
If the fault is covered under the terms of the warranty, the
If Michell Instruments determines that the fault is not covered
C.6 Warranty
Unless otherwise agreed, the Supplier warrants that as from the date of delivery for a period of 12 months the goods and all their component parts, where applicable, are free from any defects in design, workmanship, construction or materials.
instrument will be repaired at no cost to the owner and returned.
under the terms of the warranty, or if the warranty has expired, an estimate for the cost of the repairs, at standard rates, will be provided. Upon receipt of the owner’s approval to proceed, the product will be repaired and returned.
The Supplier warrants that the services undertaken shall be performed using reasonable skill and care, and of a quality conforming to generally accepted industry standards and practices.
Except as expressly stated all warranties whether express or implied, by operation of law or otherwise, are hereby excluded in relation to the goods and services to be provided by the Supplier.
All warranty services are provided on a return to base basis. Any transportation costs for the return of a warranty claim shall reside with the Customer.
Michell Instruments 21
Page 28
APPENDIX D
Easidew Portable HP User’s Manual
Appendix D
Decontamination Declaration
22 97120 Issue 2.2, March 2016
Page 29
Easidew Portable HP User’s Manual
Appendix D Decontamination Declaration
'HFRQWDPLQDWLRQ&HUWL¿FDWH
IMPORTANT NOTE: Please complete this form prior to this instrument, or any components, leaving your site and being returned to us, or, where applicable, prior to any work being carried out by a Michell engineer at your site.
Instrument Serial Number Warranty Repair? YES NO Original PO # Company Name Contact Name
Address
Telephone # E-mail address Reason for Return /Description of Fault:
APPENDIX D
Has this equipment been exposed (internally or externally) to any of the following? Please circle (YES/NO) as applicable and provide details below
Biohazards YES NO Biological agents YES NO Hazardous chemicals YES NO Radioactive substances YES NO Other hazards YES NO Please provide details of any hazardous materials used with this equipment as indicated above (use continuation sheet
if necessary)
Your method of cleaning/decontamination
Has the equipment been cleaned and decontaminated? YES NOT NECESSARY Michell Instruments will not accept instruments that have been exposed to toxins, radio-activity or bio-hazardous
PDWHULDOV)RUPRVWDSSOLFDWLRQVLQYROYLQJVROYHQWVDFLGLFEDVLF ÀDPPDEOHRUWR[LFJDVHVDVLPSOHSXUJHZLWKGU\ JDVGHZSRLQW&RYHUKRXUVVKRXOGEHVXI¿FLHQWWRGHFRQWDPLQDWHWKHXQLWSULRUWRUHWXUQ
Work will not be carried out on any unit that does not have a completed decontamination declaration.
Decontamination Declaration
I declare that the information above is true and complete to the best of my knowledge, and it is safe for Michell personnel to service or repair the returned instrument.
Name (Print) Position
Signature Date
F0121, Issue 2, December 2011
Michell Instruments 23
Page 30
NOTES:
Easidew Portable HP User’s Manual
24 97120 Issue 2.2, March 2016
Page 31
Easidew Portable HP User’s Manual
NOTES:
Michell Instruments 25
Page 32
http://www.michell.com
Loading...