Memmert UF 55 PLUS, UF 110 PLUS, UF 260 PLUS, UF 75 PLUS Instruction Manual

PLUS
PLUS
UN
IN
SN
OPERATING INSTRUCTIONS
UF
PLUS
IF
PLUS
PLUS
PLUS
100% ATMOSAFE. MADE IN GERMANY.
www.memmert.com | www.atmosafe.net
About this manual
About this manual
/UF
/SF
and incubators IN
/IF
AtmoCONTROL, observe the respective
You will find the current version of our operating manual as pdf file if you go to
......................................................................................................................
...............................................................................................
..........................................................................................
....................................................................................................................
...........................................................................................................
..............................................................................................................
............................................................................................................................
.........................................................................................................................
..........................................................................................................................
......................................................................................................
..............................................................................................................
.................................................................................................
......................................................................................................................
....................................................................................................
............................................................................................................
..............................................................................................................
...........................................................................................................
......................................................................................................
..................................................................................................
................................................................
.............
......................................................................
.................................
.............................................................................
.................................................................................................................................
...............................................................................................................
.............................................................................................
......................................................................................................................
.......................
.............................................................................................
or
.....................................................................
........................................................................................................................
....................................................................................................................
.........................................................................................................................
........................................................................................................................
................................................................................................................................
...............................................................................................................................
............................................................................
/SN
......................................................................................................................
........................................................
.................................................................................................
............................................................................................................
..............................................................................................................................

Safety regulations

Warning signs (warning of a danger)
Prohibition signs (forbidding an action)
Regulation signs (stipulating an action)
Other icons
Safety regulations
After removing covers, live parts may be exposed. You may receive
Safety regulations
Any other use is improper, and may result in hazards and damage.
medical device
type series: The appliance serves for heating non-sterile cloths
type series: The appliance serves for heating non-sterile cloths
type series: The appliance serves for temperature control of
type series: The appliance serves for sterilising medical material
Safety regulations
You can find information on correcting malfunctions from page 35.
PLUS
ON
344.4
4444.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
N
N
Manu
44.Sept
Switch off the appliance by pressing the On/

Construction

1
Construction
keys (see page 24)
3
Steel grid
5
Nameplate (covered, see page 13)
2
1
3
9 8
7
4 5 6
Construction and description
Appliances of the
and IN
and IF
Appliance fuse: Fusible link 250 V/15 A quick-blow

Function

2
3
Air flap
5
of a maximum of 0.292 ohm at the point of transfer (service line). The operator
of a maximum of 0.292 ohm at the point of transfer (service line). The operator
Atmo-
You will find a description of how to transfer programmes via Ethernet in the en-
AtmoCONTROL manual.
USB interface
Ethernet interface
Construction and description

2.6 Designation ( nameplate)

Nameplate (example)
3
Applied standard
5
Address of manufacturer
Appliance number
The nameplate (Fig. 6) provides information about the appliance model
1
2
3
4
5
6
Typ: UN 260 plus F.-Nr.: 0109.0088 230 V
~
14.8 A 50/60 Hz 3400 W
DIN12880-Kl.3.1 Nenntemp.: 300 °C
10
9
8
7
[mm]
[mm]
[mm]
[mm]
Appliance size 30 55 75 110 160 260 450 750
Construction and description
D

Applied directives and standards

C
E
B
56
A
G
F

Ambient conditions

Ambient temperature
Altitude of installation

Scope of delivery

AtmoCONTROL manual
Converter USB to Ethernet
You may get your hands or feet squashed when transport-
30 55 75 110 160 260 450 750
≥ 5 cm ≥ 5 cm ≥ 15 cm
Minimum clearance from walls and ceiling
≥ 20 cm
Setting up Comments
justable feet
Suitable for appliance size ...
30 55 75 110 160 260 450 750

Tilt protection

Attach the appliance to a
As illustrated, fasten the

Connecting the appliance

230/115-V appliances:
230/115-V appliances:
Power connection
230/115 V
400 V CEE connection
PLUS
ON
344.4
4444.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
N
N
Manu
44.Sept
Switch on
appliance

Operation and control

/SF
type, make sure to observe the
A
Opening and closing the door
CB
Operation and control
(see page 44).
Correct placement of the
Operation and control
PLUS
344.4
444 4.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
N
N
Manu
44.Sept
PLUS
344.4
444 4.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
N
N
Manu
44.Sept
53 4 13 14 15 161 2
/IF
/SF
appliances in operating mode (width may
PLUS
344.4
444 4.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
N
N
Manu
44.Sept
ControlCOCKPIT for UN
/IN
/SN
appliances in operating mode (width may
Activation key for temperature setpoint
adjustment
Activation key for temperature setpoint
Activation key for temperature setpoint
2
Setpoint and actual temperature display
adjustment
adjustment
3
Activation key for fan speed setting
5
Switch to menu mode (see page 38)
Activation key for fan speed setting
Activation key for fan speed setting
Activation key for timer setting
Switch to menu mode (see page 38)
Switch to menu mode (see page 38)
Air flap position display
Activation key for air flap position adjust-
Air flap position display
Air flap position display
ment
Activation key for interior lighting (ad-
Appliance state and programme display
Activation key for the appliance state
Appliance state and programme display
Appliance state and programme display
Activation key for temperature monitoring
Activation key for the appliance state
Activation key for the appliance state
20
Activation key for graphic representation
TEMP
180.4°C
Set 180
TIMER
TIMER
30m04h
44h:44m
Ende
13:30 23.11.
End 14:45
ON
ON
FAN
0
FLAP
FLAP
40%
%
20%
.4°C
1 2 5 13
TEMP
180.0 °C
Set 180.9°C
FLAP
TIMER
TIMER
30m04h
44h:44m
Ende
13:30 23.11.
End 14:45
ON
FLAP
40%
ON
ON
min
160.0°C
min
auto
000°C
ON
ON
min
160.0°C
min
auto
000°C
LIGHT
100
ALARM
ALARM
LIGHT
100
ALARM
ALARM
of °C
max
max
190.0°C
000°C
auto off +
max
max
190.0°C
000°C
auto off
+
+
-
+
5.0K
-
99K
-
%
5.0K
99K
-
%
Fr 20.10.2010 20:31
12.Sept.2012
Manual Mode
Holz trocknen aufheizen 09:12h
GRAPH
%rh°C
0 12
19 2011 12 17 186 7 8 9 10
15 1614
Fr 20.10.2010 20:31
13:44
12.Sept.2012
Manual Mode
Holz trocknen aufheizen 09:12h
GRAPH
%rh°C
0 12
13:44
off
off
6 7 8 9
10 11 12 19 2017 18
Operation and control
Activate the desired parameter (e.g. tem-
30.0
%rh
50.0
%CO2
MENU
30.0
%rh
50.0
%CO2
MENU
TEMP
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
Additional parameters (air flap position etc.) can be set accordingly.
AtmoCONTROL software at a computer / laptop and then transferred
30.0
%rh
50.0
%CO2
Programm 12
Fr 20.10.2010 20:31
min
000°C
max
000°C
.5°C100
MENU
30.0
%rh
50.0
%CO2
Programm 12
Fr 20.10.2010 20:31
min
000°C
max
000°C
MENU
22.4°C
TEMP
37.0°CSet
TEMP
22.4°C
180.0°CSet
TEMP
23.2°C
Set 180
22.4°C
37.0°CSet
.5°C100
.0°C
Operation and control
Appliance
Appliance is in manual operating mode
Adjustment options
Adjustment options
As described in chapter 5.4.2 , you can set the following parameters after pressing the cor-
Adjustment range: model dependent (see nameplate and technical
symbol.
You can select °C or °F as the temperature unit displayed (see
Air flap position
Air flap position
Adjustment range: 0 % (closed, recirculating operation) to 100 %
appliances)
Adjustment range: 0 to 100 % in steps of 10%
Adjustment range: 0 %, 100 %
12.Sept.2012

Manual Mode

13:44
TEMP
24.4°C
180.4 °CSet
FLAP
%40
FAN
LIGHT
100
50
%
%
Operation and control
30.0
%rh
50.0
%CO2
.5°C100
MENU
30.0
%rh
50.0
%CO2
MENU
TIMER
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
The status display
5.4.2
/UF
and
/IF
/SF
/UF
and incubators IN
/IF
– this determines if the timer should
-
Ende
TIMER
--h- m
9:00 23.11.
TIMER
Ende
30m04h
13:30 23.11.
04 mh 3
End
12.Sept.2012
Timer active
0
13:30 23.11.
13:44
TIMER
End
00m00h
13:30 23.11.
Operation and control
AtmoCONTROL software. Transfer to the appliance is possible using
A description of how to create and save programmes can be found in the separate
AtmoCONTROL software manual.
Starting a programme
Starting a programme
(
0
0
start
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
Test 012
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
You cannot change any parameters (e.g. the temperature) at the appliance while a
and
can still be used.
TIMER
End
--m--h
9:00 23.11.
Fr 20.10.2010 20:31
Manual mode
manueller Betrieb
Aktivate
GRAPH
8
80 40
4
17:4413.Sept.2012
10:4412.Sept.2012
Test 012 ready
10:4412.Sept.2012
Test 012 Ramp 1
Operation and control
You can cancel an active programme at any
You can cancel an active programme at any
Fr 20.10.2010 20:31
0
is highlighted.
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
A cancelled programme cannot be resumed at the point it was cancelled. It must be
when the programme
You can now
electronic temperature monitoring (TWW)
ASF)
%
%
End
Test 012
GRAPH
8
80
10:4412.Sept.2012
Test 012
manueller Betrieb
Ramp 3
10:4912.Sept.2012
10:4812.Sept.2012
Cancel program
Test 012
10:4912.Sept.2012
End
Test 012
12.Sept.2012
Manual Mode
13:44
Operation and control
display. The settings made
is shown ( Fig. 17). The type
is shown ( Fig. 17). The type
of the electronic overtemperature
is exceeded, the TWW
t
Schematic diagram of how the TWW
min
Holz trocknen aufheizen 09:12h
Fr 20.10.2010 20:31
min
000°C
ALARM
of °C
max
000°C
auto off
99K
-
+
0 12
%rh°C
GRAPH
off
ONON
0 8 16
40
80
GRAPH
344.4
4444.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
LICHT
%
100
ALARM
max
190.0°C
+
/
-
3.0K
min
160.0°C
auto
Manueller Betrieb
12.Sept.2012
13:44
180.4°C
TEMP
Set 185
.0°C
344.4
444 4.
TEMP
Set
°C
°C
FLAP
40%
LÜFTER
%
0
23.2°C
TEMP
Set 180
.0°C
TEMP
TWW Set 190.0 °C
195.4°C
160.0°C
auto
ALARM
max
190.0°C
+
5.0K
-
°C
TEMP
TEMP
TEMP
23 °C
195.4°C
180
TWW Set 190.0 °C
TIMER
Ende
23.11. 13:30
LÜFTER
0
%
FLAP
LUFTKLAPPE
30m04h
40%
20%
Emergency operation
Setting MAX
Set temperature
Controller error
Operation and control
t
°C
Schematic diagram of how the TWB
heating switched off by TWB
Setting Max Set
temperature
Controller error
Operation and control
Automatic
ASF)
ASF is a monitoring device that automatically follows the set temperature setpoint within an
180 °C
Schematic diagram of how the ASF temperature monitoring works

Adjusting temperature monitoring

display. The
setting (under-
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
Holz trocknen aufheizen 09:12h
Fr 20.10.2010 20:31
0 12
%rh°C
DRAPH
off
ON
ON
183 °C
177 °C
A B C D E
183 °C
176 °C
170 °C
ASF active
177 °C
AUTO AUTO AUTO
ASF alarm
ASF active ASF active
min
60 0.
min
auto
000°C
ALARM
ALARM
°C
t
max
max
120.0°C
000°C
auto off
+
+
0.0K
-
99K
-
Operation and control
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
min
display (overtemperature pro-
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
Accept the upper alarm limit value by
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
With the turn control, adjust the desired
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
1600.
auto
min
1600.
auto
min
1600.
auto
min
160.0°C
auto
min
160.0°C
auto
ALARM
°C
ALARM
°C
ALARM
°C
ALARM
ALARM
max
120.0°C
+
0.0K
-
max
120 0.
+
0.0K
-
max
190 0.
+
0.0K
-
max
190.0°C
+
0.0K
-
max
190.0°C
+
0.0K
-
°C
°C
min
160 0.
auto
min
160 0.
auto
min
160.0°C
auto
ALARM
°C
ALARM
°C
ALARM
max
190.0°C
+
0.3
-
max
190.0°C
+
5.0
-
max
190.0°C
+
5.0K
-
K
K
Operation and control
display provides an overview of the chronological sequence of the actual values
display.
80
Fr 20.10.2010 20:34
12.09.2012
ar-
20:34
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
PLUS
344.4
4444.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
N
N
Manu
44.Sept
Switch off
appliance
60 40 20
0 4 8 12 16 20 24
14.00 16.00 18.00
0 20:34
12.09.2012
°C 100
80 60 40 20
0 4 8 12 16 20 24
14.00 16.00 18.00
Fr 20.10.2010
ON
After removing covers, live parts may be exposed. You may receive
the acoustic
menu (see
in the
Description Cause Action See
Automatic
/ overtemperature
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
TEMP
185.4°C
ASF Set 190.0 °C
TEMP
195.4°C
TWW Set 190.0 °C
Description Cause Action See
Appliance locked by user ID
Appliance is in programme
Appliance is in "wrong"
key
Appliance error
TEMP
195.4°C
TWB Set 190.0 °C
TEMP
230.4 °C
TB
Error description Cause of error Rectifying errors See
Error 23 Pt100 Error Contact Service
Additionally, depending on the duration of the power loss, the
After power supply has been restored, operation is continued with the parameters set. The
After power supply has been restored, the timer or programme always starts again from the
enter menu mode, press the MENU key.
PROGRAM
USER ID
PROTOCOL
PLUS
344.4
4444.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
N
N
Manu
44.Sept
1 2 3 4 5 12 14 1513
2
3
5
mode
Setup activation key (basic appliance set-
Setup display (basic appliance settings)
Adjustment display
Setup display (basic appliance settings)
Setup display (basic appliance settings)
Adjustment activation key
Adjustment display
Adjustment display
Acoustic signal adjustment activation key
Acoustic signal adjustment display
Acoustic signal adjustment activation key
Acoustic signal adjustment activation key
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
TIME
LANGUAGE
SIGNALTÖNE
PROG
PROTOCOL
SETUP
ON
CALIB
6 7 18 198 9 10 11 16 17
ZEIT UND DATUM
SOUND
USER ID
Activate the desired parameter (in this
LANGUAGE
OG
D
O
OCO
PROG
ZEIT UND DATUM
USER ID
PROTOCOL
LANGUAGE
PROGRAM
USER ID
PROTOCOL
SOUND
PLUS
344.4
4444.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
N
N
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SPRACHE
SIGNALTÖNE
PROG
ZEIT UND DATUM
CALIB
USER ID
PROTOCOL
PROGRAM
TIME
USER ID
PROTOCOL
SOUND
PLUS
344.4
4444.
TEMP
Set
°C
°C
FLAP
40%
TIMER
hd
1002
End Sept.29 22 24
FAN
%0
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
You can now
SETUP
PROGRAM
TIME
USER ID
PROTOCOL
PROGRAM
ER I
D
R
OTOCO
L
+30%
192.168.100.100 °C F TWW TWB
Setup
192.168.100.100
255.255.0.0 °C F TWW TWB
Grid Shelf
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SPRACHE
ENGLISH
DEUTSCH FRANCAIS ESPANOL
SIGNALTÖNE
TIME
SETUP
ON
CALIB
ENGLISH DEUTSCH
ENGLISH
FRANCAIS
DEUTSCH
ESPANOL
FRANCAIS
ESPANOL
LANGUAGE
SETUP
IP address
Balance
Subnet mask
IP Adresse
Unit
Einheit
Alarm temp
Alarm Temp
Timer mode
Timer Mode
Slide-in unit
All other settings can be made accordingly. The settings possible are described in the
display, you can set the following parameters:
and
of the appliance's Ethernet interface (for connection to
of the temperature display (°C or °F, see page 41)
(see page 43)
Programm negnulletsniEllokotorPtäreG efliHnekcurD
network (schematic example)

Setup

Unit
IP adress 255. 145.136.225
Subnet mask 255.255. 0.0
°C F
1/2
AtmoCONTROL
INP 250 Test 01
180
.0°C
i
HPP 250 Labor
37
.0°C
i
44
.4%rh
-
+-+
Programmname
Editor
Simulation Protokoll
STAND BY
151
192.168.1.216
LAN 1: 192.168.1.233
LAN 2: 192.168.1.215
LAN 3: 192.168.1.241
Activate the
display. The entry
PROGRAM
TIME
PROGRAM
D
O
OCO
Setup
Accept the selection by pressing the
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
Accept the selection by pressing the
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
Grid Shelf
After setting the last three digits, accept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
Slide-in unit
Grid Shelf

Unit

Activate the
display and select
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
Accept the selection by pressing the
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
SETUP
IP address 192.168.100.100
Balance
Subnet mask 255. 255.0.0
IP Adresse
Unit
Einheit
Alarm temp
Alarm Temp
Timer mode
Timer Mode
Slide-in unit
+30%
192.168.100.100
°C F
°C F
TWW TWB
TWW TWB
Grid Shelf
IP address 192. 168.100.100
Subnet mask 255.255. 0.0 Unit Alarm temp
°C F
TWW TWB Timer mode Slide-in unit
Grid Shelf
IP address 255. 168.100.100
Subnet mask 255.255. 0.0 Unit Alarm temp
°C F
TWW TWB Timer mode Slide-in unit
Grid Shelf
IP address 255. 168.100.100
Subnet mask 255.255. 0.0 Unit Alarm temp
°C F
TWW TWB Timer mode Slide-in unit
IP address 255. 145.136.225
Subnet mask 255.255. 0.0 Unit Alarm temp
°C F TWW TWB
Timer mode
IP address 255. 145.136.225 Subnet mask 255.255. 0.0
Unit
Alarm temp
°C °F
TWW TWB Timer mode Slide-in unit
Grid Shelf
IP address 255. 145.136.225 Subnet mask 255.255. 0.0
Unit
Alarm temp
°C °F
TWW TWB Timer mode Slide-in unit
Grid Shelf
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
IP address 255. 145.136.225
Slide-in unit
Grid Shelf
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
Activate the
display and select
with the turn control.
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
Accept the selection by pressing the
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
Subnet mask 255.255. 0.0
Unit
Alarm temp
°C °F
TWW TWB Timer mode
IP address 255. 145.136.225 Subnet mask 255. 255.0.0
Unit
Alarm temp Timer mode Slide-in unit
°C °F
TWW TWB
Grid Shelf
IP address 255.145.136. 225 Subnet mask 255.255. 0.0
Unit
Alarm temp
°C F
TWW TWB
Timer mode Slide-in unit
Grid Shelf
IP address 255.145.136. 225 Subnet mask 255.255. 0.0 Unit
Alarm temp
°C F
TWW TWB
Timer mode Slide-in unit
Grid Shelf
IP address 255.145.136. 225 Subnet mask 255.255. 0.0 Unit
Alarm temp
°C F
TWW TWB
Timer mode Slide-in unit
Grid Shelf
IP address 255.145.136. 225 Subnet mask 255.255. 0.0
Unit
Alarm temp
°C F
TWW TWB
Timer mode Slide-in unit
Grid Shelf
/UF
and incubators IN
/IF
A
t

Timer mode

A Timer independent of setpoint: Timer starts right after activation
/SF
/UF
and incubators IN
/IF
Setting
Setting
Activate the
display and select
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
Accept the selection by pressing the
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
).
).
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
°C/°F
6 K
B
t
Timer
t
Timer
IP address 255. 145.136.225 Subnet mask 255.255. 0.0 Unit Alarm temp
°C F
TWW TWB
Timer mode
Slide-in unit
Grid Shelf
IP address 255. 145.136.225 Subnet mask 255.255. 0.0
Unit Alarm temp
°C F
TWW TWB
Timer mode
Slide-in unit
Grid Shelf
IP address 255. 145.136.225 Subnet mask 255.255. 0.0
Unit Alarm temp
°C F
TWW TWB
Timer mode
Slide-in unit
Grid Shelf
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
IP address 255. 145.136.225
or
)
Activate the
display and select
with the turn control.
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
Accept the selection by pressing the con-
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
Grid Shelf
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
SETUP
Subnet mask 255.255. 0.0
Unit Alarm temp
°C F
TWW TWB
Timer mode
Slide-in unit
Grid Shelf
IP adress 255. 145.136.225 Subnet mask 255.255. 0.0
Unit Alarm temp
°C °F
TWW TWB Timer mode
Slide-in unit
Grid Shelf
IP adress 255. 145.136.225 Subnet mask 255.255. 0.0
Unit
Alarm temp
°C °F TWW TWB
Timer mode
Slide-in unit
IP adress 255. 145.136.225 Subnet mask 255.255. 0.0
Unit Alarm temp
°C °F
TWW TWB Timer mode
Slide-in unit
Grid Shelf
IP adress 255. 145.136.225 Subnet mask 255.255. 0.0 Unit
Alarm temp
°C °F
TWW TWB Timer mode
Slide-in unit
Grid Shelf
upper heat output
upper heat output
groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower
heating groups to emit 30 % more heat than the upper ones. The 0 % setting restores the
groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower
groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower
Setting
Setting
Activate the
display and select
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
lower heat output
-20%
2/2

Balance

-20 %
SETUP
lower heat output
+30%
2/2
Balance
+30 %
2/2
2/2
Balance
Balance
SETUP
SETUP
+30 %
SETUP
+30 %
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
2/2
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
display, you can set the date and time, time zone and daylight savings.
Activate the time setting To do so, press
display. The display is enlarged
– e.g. the time zone: Turn the turn
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
TIME
Accept the selection by pressing the
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
TIME
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
TIME
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
TIME
Balance
SETUP
+50 %
Date
Time Time zone Daylight savings

Date and time

12. 05. 2012
12:00 GMT +1
2/2
Balance
Date Time
Time zone
Daylight savings
Date 12
Time Time zone Daylight savings
SETUP
+50 %
12.05 .2012 12:00 GMT +1
.05. 2012 12:00 GMT +1
Date 27
Time Time zone
.05. 2012 12:00 GMT +1
Daylight savings
Date 27 05
Time Time zone
. . 2012 12:00 GMT +1
Daylight savings
Accordingly, you can also set:
Year
-11 -10 -9 -8 -7 -6 -5 -4 -3 -2 -1 0 +1 +2 +3 +4 +5 +6 +7 +8 +9 +10 +11
+12 -12-11
-11 -10 -9 -8 -7 -6 -5 -4 -3 -2 -1 0 +1 +2 +3 +4 +5 +6 +7 +8 +9 +10 +11
+12 -12-11
Time zones
CAL 1
Schematic example of temperature adjustment
Activate the adjustment setting. To do
display. The display
PROGRAM
TIME
USER ID
PROTOCOL
PROGRAM
ER I
D
R
O
T
OCO
L
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
JUSTIEREN
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
JUSTIEREN
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
JUSTIEREN
+0,5 K
CAL 3
-0,8 K
CAL 2
+2,6 K
factory calibration
0°C
40°C
120°C
Temperatur
Temperature
Last updated 12.10.2012 12:00:00
letzte Justierung 12.10.2012 12:00
180°C
JUSTIEREN
Calibration
Cal1
°C -0,2
40.0
C K0,2-
Cal1
40.0
Cal2 100.0 °C +0,1K
C K0,1+
Cal2
100.0
Cal3 180.0 °C -0,2 K
C K0,2-
180.0
Cal3
K
Temperature
Temperature
Temperature
Cal1
Cal2
100.0
180.0
Cal3
Cal1
Cal2
120.0
Cal3
180.0
Cal1
Cal2
120.0 0,1+
Cal3
180.0
40.0
40.0
40.0
C K0,2- C K0,1+ C K0,2-
C K0,2- C K0,1+ C K0,2-
C K0,2-
C K
C K
0,2-
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
Calibration
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
JUSTIEREN
After the calibration procedure, the
°C
/UF
as well as sterilisers SN/SF and 10 K for incubators IN/IF.
TEMP
120.0°C
Set 120
.0°C
TEMP
Temperature
Temperature
C K0,2-
Cal1
40.0
C K
Cal2
120.0 0,0
C K
Cal3
180.0
TEMP
21.4°C
120.0 °CSet
122.6 °C
C K0,2-
Cal1
40.0
C K
Cal2
120.0 2,6+
C K
Cal3
180.0
0,2-
0,2-
120.0°C
Set 120
.0°C
120.0 °C
display, you can transfer programmes created using
AtmoCONTROL software
Activate the programme display. To do so,
display. The display is enlarged
Select
automatically
– is highlighted
Select
function by pressing
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
PROGRAMM
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
PROGRAMM
Accept the selection by pressing the
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
PROGRAMM
As soon as the programme is ready, the
To start the
PROGRAMM
Test 015
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
You can now remove the USB storage medium.
Delete
with the turn control and select the programme to be
Delete
Test 012
Test 013 Test 014 Test 015 Test 016 Test 017
Program
Select
Delete
Test 012 Test 022 Test 013 Test 014
Test 023
Select
Delete
Select
Delete
Select
Delete
Test 012
Test 022 Test 013 Test 014 Test 022
Test 012
Test 022 Test 013 Test 014
Test 023
Test 012
Test 022 Test 013 Test 014
Test 023

Sound

display, you can define whether or not the appliance should emit acoustic
Activate the acoustic signal adjustment.
display. The
Keysound
OG
D
O
OCO
– e.g.
(Special accessories) –
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
Sound
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
At the end On alarm If door open
Keysound At the end On alarm
If door open
Keysound At the end On alarm
If door open
Keysound At the end On alarm
If door open
Keysound At the end On alarm
If door open
Sound
Sound
Sound
Sound

Protocol

You can export the protocol data for different periods to a USB storage medium
AtmoCONTROL software for graphical
Activate the protocol. To do so, press
display. The display
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
As soon as the transfer is complete, a
AtmoCONTROL or
AtmoCONTROL manual.
Protocol
Protocol
1 Week
1 Month Complete control range
Protocol
1 Week
1 Month Complete control range
lock symbol in the respective display (Fig. 28 ).
AtmoCONTROL
A description of how to create a user ID in
AtmoCONTROL is provided in the separate
AtmoCONTROL manual.
Activate the user ID. To do so, press the
display. The display is enlarged
automatically
%
100
ALARM
max
444.4°C
auto
+
/
-
0.0K
min
444.4°C
Manu
44.Sept
entry and
TEMP
Temperature adjustment
at appliance locked (example)
Temperature adjustment
Temperature adjustment
Activate
Deactivate
22.4
USER-ID
°C
USER-ID
Activate
Deactivate

Sterilisers SFplus/SNplus

/SN
/SN
appliance serves for sterilising medical material through dry heated air at
According to WHO: 180 °C with a minimum hold time of 30 min
According to the European Pharmacopoeia: 160 °C with a minimum hold time of 120 min
destroy
/SN
feature
PROGRAM
Activate
Delete
Test 012
Test 022 Test 013 Test 014
Steri 180 °C
Test 015

Maintenance and service

Maintenance and service
After removing covers, live parts may be exposed. You may

Storage and disposal

Storage

August 13

Index

A
Accessories 16
Acoustic signals 47
Activation key 25
Adjustment 47
Air flap position 26
Air supply 11
Alarm 32
Alarm temperature 42
Ambient conditions 15
Ambient temperature 16
Appliance error 36
ASF 29
AtmoCONTROL 3
Automatic temperature
Universal ovens
Incubators
Sterilisers
PLUS
PLUS
PLUS
24.06.2013
D24026 // englisch
Memmert GmbH + Co. KG Willi-Memmert-Straße 90-96 | D-91186 Büchenbach Tel. +49 9122 925-0 | Fax +49 9122 14585 E-Mail: sales@memmert.com facebook.com/memmert.family Die Experten-Plattform: www.atmosafe.net