SUPPLEMENT TO THE
OPERATING INSTRUCTIONS
for climatic test chambers CTC and temperature test chambers
TTC with additional accessories
About this manual
This is a supplement to the operating instructions of climatic test chambers CTC and
temperature test chambers TTC with additional accessories. Function and application of the
following optional accessories are described:
► Temperature sensor PT100 - freely positionable in the chamber for load monitoring and
registration
► Compressed air dehumidification
► Connection for central water supply
Manufacturer and customer service
MEMMERT GmbH + Co. KG
PO Box 17 20
91107 Schwabach, Germany
Äußere Rittersbacherstr. 38
91126 Schwabach
Germany
Phone: +49 (0) 09122 / 925-0
Fax: +49 (0) 09122 / 14585
E-mail: sales@memmert.com
Online: www.memmert.com
Customer service:
Phone: +49 (0) 09122/925-128
and: +49 (0) 09122/925-126
E-Mail: service@memmert.com
For service enquiries, please always specify the appliance number on the nameplate.
© 2012 Memmert GmbH + Co. KG
Date 11/2012
We reserve the right to make changes
Supplement to Operating Instructions CTC/TTC
1. Temperature sensor PT100 - freely positionable in the
chamber for load monitoring and registration
1.1 Functions
One of the two temperature sensors of the controller is a freely positionable temperature
sensor PT100. The PT100 sensor is plugged into the 4-pin LEMO connector on the right side
of the shutter box and placed in the interior of the climate chamber through the silicone feedthrough.
Thanks to the second temperature sensor, the programme sequence can be controlled in a
way keeping the programme from switching to the next ramp until the chamber load has
reached the desired setpoint temperature.
To control the temperature inside the chamber, the temperature value of the temperature
monitoring sensor is used. The monitoring function is fully maintained.
1.2 Displaying the temperature of the additional, freely
positionable temperature sensor PT100
Turn the turn control to the left until „SENSOR 1“ appears in the display. The temperature display shows the current temperature of the additional, freely positionable temperature sensor
PT100.
1.3 Display of the sensor temperature in the internal flash log
memory
RED: Interior temperature in the test chamber
YELLOW:
Temperature of the freely positionable PT100 sensor (e. g. core temperature of
the test object)
1.4 SPWT function in combination with the freely positionable
PT100 sensor
Default sensor
configuration
Without freely
positionable PT100 sensor
Criterion for reached setpoint temperature
Temperature inside the chamber must have reached the setpoint temperature (± 2 K).
With additional, freely
positionable PT100 sensor
Temperature at the freely positionable PT100 sensor must have
reached the set ASF tolerance band; adjustment range (±0.5 to
±10 K).
3