Memmert TTC User Manual

SUPPLEMENT TO THE OPERATING INSTRUCTIONS
for climatic test chambers CTC and temperature test chambers TTC with additional accessories
About this manual
This is a supplement to the operating instructions of climatic test chambers CTC and temperature test chambers TTC with additional accessories. Function and application of the following optional accessories are described:
registration
► Compressed air dehumidification ► Connection for central water supply
Manufacturer and customer service
MEMMERT GmbH + Co. KG PO Box 17 20 91107 Schwabach, Germany
Äußere Rittersbacherstr. 38 91126 Schwabach Germany
Phone: +49 (0) 09122 / 925-0 Fax: +49 (0) 09122 / 14585 E-mail: sales@memmert.com Online: www.memmert.com
Customer service: Phone: +49 (0) 09122/925-128 and: +49 (0) 09122/925-126 E-Mail: service@memmert.com
For service enquiries, please always specify the appliance number on the nameplate. © 2012 Memmert GmbH + Co. KG Date 11/2012 We reserve the right to make changes
Supplement to Operating Instructions CTC/TTC
1. Temperature sensor PT100 - freely positionable in the chamber for load monitoring and registration
1.1 Functions
One of the two temperature sensors of the controller is a freely positionable temperature sensor PT100. The PT100 sensor is plugged into the 4-pin LEMO connector on the right side of the shutter box and placed in the interior of the climate chamber through the silicone feed­through.
Thanks to the second temperature sensor, the programme sequence can be controlled in a way keeping the programme from switching to the next ramp until the chamber load has reached the desired setpoint temperature.
To control the temperature inside the chamber, the temperature value of the temperature monitoring sensor is used. The monitoring function is fully maintained.
1.2 Displaying the temperature of the additional, freely positionable temperature sensor PT100
Turn the turn control to the left until „SENSOR 1“ appears in the display. The temperature dis­play shows the current temperature of the additional, freely positionable temperature sensor PT100.
1.3 Display of the sensor temperature in the internal flash log memory
RED: Interior temperature in the test chamber
YELLOW:
Temperature of the freely positionable PT100 sensor (e. g. core temperature of the test object)
1.4 SPWT function in combination with the freely positionable PT100 sensor
Default sensor configuration
Without freely positionable PT100 sensor
Criterion for reached setpoint temperature
Temperature inside the chamber must have reached the set­point temperature (± 2 K).
With additional, freely positionable PT100 sensor
Temperature at the freely positionable PT100 sensor must have reached the set ASF tolerance band; adjustment range (±0.5 to ±10 K).