Thank you very much for the trust which you have shown by purchasing this autoclave.
For 50 years now, MELAG — a medium-sized family-owned and -operated business —
has specialised in the production of sterilization equipment for medical practice. During
this period, MELAG has succeeded in becoming a leading manufacturer of sterilization
equipment. More than 335,000 MELAG units sold throughout the world testify to the exceptional quality of our products — which are manufactured exclusively in Germany.
As all other MELAG products, this autoclave was manufactured and tested according to
strict quality criteria. Before placing this unit into operation, please carefully read this Operating Manual. The long-term functional effectiveness and the preservation of the value
of your autoclave will primarily depend on careful preparation of instruments before sterilization, and on proper care of the unit.
The staff and management of MELAG
The functional effectiveness and the preservation of
value of this unit depend on:
1. Proper preparation of the instruments to be sterilized
2. Proper care of the autoclave
3. The use of sufficiently pure demineralized / distilled water
DESCRIPTION OF THE UNIT ...........................................................................................................................6
1.1 Views of the unit...................................................................................................................................6
1.4 Control panel........................................................................................................................................8
1.5 Performance features of this autoclave ...............................................................................................8
1.5.1 Preliminary air removal by sub-atmospheric pulsing ....................................................................8
1.5.2 Sterilization and disinfection functions..........................................................................................8
1.5.3 Separate steam generation...........................................................................................................8
1.5.4 One-way system / conductivity measurement / automatic re-supply of water..............................8
1.5.5 Electronic parameter control .........................................................................................................8
2.1 Requirements for installation................................................................................................................9
2.2 Transport straps ...................................................................................................................................9
2.4 Mains power supply .............................................................................................................................9
2.5 Outlet connection for one-way water .................................................................................................10
2.6 Internal water supply with demineralized / distilled water ..................................................................10
2.7 Connection of a water treatment system ...........................................................................................10
2.8 Emptying the water storage tank........................................................................................................10
3 PLACING THE AUTOCLAVE INTO OPERATION ..................................................................................10
3.1 Printer connection and initialisation (optional) ...................................................................................10
3.1.1 Connection of the MELAprint
3.1.2 Initialisation of the printer / setting for immediate printout ..........................................................10
3.2 Vacuum test .......................................................................................................................................11
3.3 Trial run ..............................................................................................................................................12
3.4 Installation log – sent to MELAG........................................................................................................12
4 FOR EACH STERILIZATION CYCLE ......................................................................................................14
4.1 Water and power................................................................................................................................14
4.1.1 Demineralized / distilled water ....................................................................................................14
4.1.2 Power supply ..............................................................................................................................14
4.2 Preparation of instruments .................................................................................................................14
4.3 Loading the autoclave ........................................................................................................................15
4.4 Closing the door .................................................................................................................................16
4.5 Selection of the program ....................................................................................................................16
4.6 Start of program .................................................................................................................................17
4.7 Program sequence.............................................................................................................................17
Printout of cycle log.......................................................................................................................................19
4.8 Removing the sterilized items ............................................................................................................19
1.2.3 Operating media __________________________________________________________________
Power ratings:
Electricity ratings 230 V AC / 10.4 A / 50 ... 60 Hz
Power ratings 2400 W; fuse = 16 A; RCCB (residual current circuit breaker); 30 mA
aqua dem / aqua dest
Demineralized / steam-distilled water in accordance with VDE 0510
Seitenansicht
Fig. 1 Views of the Vacuklav
31-B
1 Water storage tank 11 Power cable
2 Emergency overflow (OUT) 12 Front adjustable feet of the unit
3 One-way drain (OUT) 13 Serial data and printer interface (RS 232)
4 Spring-operated safety valve 14
5 Connection for demineralized / distilled
Fuses for the unit: 2 × 16 A / FF
15 Power switch
water (OUT)
6 Sterile filter 16 Sliding-lock handle
7 Connection for pressure release for
17 Hinged door (opens to the left)
water storage tank (IN)
8 Cooler 18 Operator’s control and monitoring panel
9 Connection for demineralized / distilled
water (IN) (quick-fitted threaded
connection for 6 × 1 hose)
19 Cover to water tank
20 Drain hose for water tank, drain side (OUT)
21 Drain hose for water tank, demineralized-water side (OUT)
*1 Alternative: a separate drain with stench trap, instead of the double-chamber trap.
Example 1
Vacuklav31-B rear view
Standard version = version as supplied
(with internal supply of water)
Leak detector with shutoffvalve and sensor
Example 3
Vacuklav31-B rear view
MELAdem
with manual draining of the waste-water tank
MELAjet
Leak detector with shutoff valve and sensor
(optional)
37 (optional), external water supply,
(optional)
(A) with double-chamber trap (B)
Example 2
Vacuklav31-B rear view
MELAdem
with manual draining of the waste-water tank
(optional)
Example 4
Vacuklav31-B rear view
MELAdem
automatically into a double-chamber trap
Leak detector with shutoff valve and sensor
(optional)
37 (optional), external water supply,
47 (optional), with drain of waste water
*1
Page 7
1.4 Control panel ___________________________________________________________
2 x 20-character alphanumeric LED
display For display of program status and
parameters, and for operator for basic and
special functions
Program
Start / Stop
Fig. 3: Vacuklav
Function buttons
For selections, settings, and display of
special functions: printing, date and time of
day, pre-heating, total sterilised loads, water
conductivity, error-report acknowledgement
31-B control panel
For selection of the sterilisation programs and
Program selection button
test programs, and for selection/settings of
options (sub-menus) for the special functions
Start / stop button
For starting and stopping of
programs, drying, and for
control fo special functions
1.5 Performance features of this autoclave______________________________________
1.5.1 Preliminary air removal by sub-atmospheric pulsing ________________________________________
Preliminary air removal by sub-atmospheric pulsing repeatedly evacuates the air from the autoclave and alternately
introduces steam between evacuations: providing the required penetration of the sterilized objects with saturated
steam. This ensures fast and effective execution of difficult sterilization tasks: e.g., instruments with narrow bores
and large amounts of textiles. This autoclave also features a special program for the Bowie & Dick Test (a
recognised procedure used in large-scale sterilization to confirm steam penetration.
31-B features three sterilization programs for temperatures at 134°C: the “Universal Program” (for
wrapped objects), the “Prion Program” (a special Universal Program), the “Fast Program” for unwrapped items, and
the “Gentle Program” (a sterilization program for textiles and rubber articles at 121°C). The user can at any time
perform additional functional checks of the autoclave by running the Bowie & Dick Test for steam penetration, and
the Vacuum Test for leak testing. The additional program "MELAsteam
®
" is available as an option, and functions at
a temperature of 136°C.
1.5.3 Separate steam generation______________________________________________________________
The powerful steam generator outside the actual sterilization chamber makes it possible to sterilize large loads of
instruments or textiles in a short time. No waiting periods are necessary between the sterilization cycles, and
excess temperatures within the sterilization chamber cannot occur.
1.5.4 One-way system / conductivity measurement / automatic re-supply of water ____________________
The Vacuklav
®
31-B operates with the tried-and-proven one-way system: this means that all evaporated water and
all impurities contained in this water are removed from the autoclave upon pressure release. For the following
program cycle, new demineralized / distilled water is therefore used. An integrated conductivity measurement
system monitors the quality of the demineralized / distilled water used in steam generation. For heavy use of the
autoclave, we recommend supply of the demineralized / distilled water by connection of a water-treatment system
such as the MELAdem
®
47 or MELAdem® 37. If the instruments are carefully prepared, this arrangement prevents
spots on the items sterilized, and avoids fouling of the autoclave.
1.5.5 Electronic parameter control ____________________________________________________________
The microprocessor installed in the Vacuklav
®
31-B implements electronic parameter control, which continuously
monitors pressure, temperature, and time in the programs. The process evaluation and monitoring system
contained in the program control system compares momentary process parameters with standard process data and
monitors the process with respect to limit temperatures, limit times, and limit pressures. This function detects
malfunctions in the program cycle and guarantees the security of the sterilization / disinfection results.
Activating the pre-heating function warms up the cold autoclave chamber before sterilization, or maintains the
temperature between sterilization cycles. This shortens the duration of cycles and considerably reduces the
formation of condensation, thus optimising drying results.
1.5.8 Documentation of results _______________________________________________________________
The memory of the electronic control system stores records of the previous 40 programs. For effective hard-copy
documentation and for checking purposes a MELAprint
immediately after completion of a program, or to print out records from the memory. By connection of a PC, the
program MELAwin
out these results as needed.
®
can also transfer the records to the PC, for permanent archiving of all results, and for printing
®
42 printer can be connected to print out a record
®
31-B
2 Installation
For the steps required to prepare for and carry out the setting up and installation of this autoclave, see the separate
instructions contained in “Setting up the Vacuklav
2.1 Requirements for installation ______________________________________________
The autoclave should be set up in a dry place which is protected against dust. The base should be stable and
capable of supporting the weight of the appliance (unloaded weight = 43 kg). The space required by the autoclave
can be seen from the external dimensions (see Section 1.2.1 here). A minimum additional space of 10 cm should
be allowed on both sides and above the autoclave in order to ensure that heat can escape.
The electrical power supply should be a separate 230V AC circuit with a 16 A fuse.
This autoclave operates with a blower for the cooling system, on the rear side of the unit. If there is a restriction of
the escape of heat above the unit, this can impair the functioning and shorten the service life of the autoclave.
IMPORTANT: We therefore urgently advise our users not to install the autoclave in an area where there is
not sufficient circulation of air.
If the user decides to install an automatic one-way water system, it is necessary to have a connection to the
wastewater plumbing near the autoclave, preferably a wall outlet (NW 40) or a sink drain (standard length of
outflow pipe = 2 m, with 16 mm diameter). The work surface on which the autoclave stands must be higher than the
outlet, and the outlet pipe must be without bends and twists which could prevent water flowing out freely. At the
same time, the work surface must provide convenient access to the autoclave, and the display must be clearly
visible.
The autoclave can be supplied with demineralized / distilled water from the integrated dual-chamber storage tank,
with freshwater and wastewater chambers. Alternatively, the Vacuklav
water purifier – the MELAdem
note that this will require additional space.
®
47 or MELAdem®37 (or an equivalent water purification system). However, please
®
31-B”.
®
31-B can also be connected to an external
2.2 Transport ribbons _______________________________________________________
Lift the autoclave out of the packaging by using the transport straps. The straps themselves are each removed by
unscrewing the two retaining screws, which must then be screwed firmly back in place without washers.
In order to operate properly, and to allow unimpeded return flow of condensate into the steam generator, the autoclave must be installed so that it definitely slopes toward the rear. First install the autoclave in a horizontal (level)
position, and check that this position is in fact horizontal with a spirit level at the chamber flange. Then extend the
front feet to lift up the front side of the autoclave, by screwing the feet out by at least three (3) turns.
2.4 Mains power supply _____________________________________________________
The electric cable of the appliance must be plugged into a mains socket rated at 230 V / 50 Hz. The power rating of
the autoclave is 2400 W. In order to avoid overloading the building electricity supply, we recommend using a
separate electrical circuit fitted with a 16 A fuse and optionally protected with a type RCCB (residual current circuit
breaker) with a tripping current of 30 mA.
Page 9
2.5 Outlet connection for one-way water _______________________________________
The connection for the one-way water outlet at the back of the unit (3) must be hooked up to the drainage system
of the building by means of the outlet pipe supplied (transparent pipe). It is important that the pipe should have a steady downward gradient, without twists and kinks. The outlet must be connected to an existing sink drain, or
to a separate NW40 drain pipe with use of the wall-mounted drain (MELAG article no. 37410).
2.6 Internal water supply with demineralized / distilled water ______________________
If the user decides for internal supply of the autoclave with demineralized / distilled water, this water is taken from
the right chamber of the built-in water storage tank. To install the internal system for supply of demineralized /
distilled water, connect the hose at the demineralized / distilled water connection of the supply tank to the swivell
threaded connection at the supply point for demineralized / distilled water on the rear side of the autoclave. For this
variation of water supply, the autoclave is delivered ready for installation. To fill the feed water tank, remove the
cover of the tank and fill the right chamber with demineralized / distilled water up to the maximum mark.
2.7 Connection of a water treatment system ____________________________________
The autoclave can be directly connected to a water treatment system for direct supply with demineralized / distilled
water. Instead of being connected to the water storage tank (as in Section 2.6), the demineralized / distilled water
connection is hooked up directly to the water treatment system. The reverse-osmosis system MELAdem
the ion-exchanger MELAdem
consult the operating manuals of these reverse-osmosis systems for detailed instructions on installation and
placing into operation. If you decide to use water treatment systems made by other
they deliver water with sufficient quality and quantity. Please get in touch with the company MELAG before
connecting such systems made by other manufacturers. In addition, a water stop valve (MELAG article no. 01056)
should be installed upstream of such water treatment systems, in order to satisfy insurance requirements.
®
37 are designed to optimally satisfy the requirements of the Vacuklav®31-B. Please
manufacturers, be sure that
®
47 and
2.8 Emptying the water storage tank __________________________________________
It is easy to completely empty the water storage tank of this autoclave. This is required, for example, to clean or
repair the autoclave, or to place it out of operation. To drain the tank, pull the drain hoses out of the front panel at
the plug (see Fig. 1: drain side [21], and tank side, with demineralized water [22]). Pull out the plugs, and allow the
water to drain into a separate container.
3 Placing the autoclave into operation
3.1 Printer connection and initialisation (optional) _______________________________
3.1.1 Connection of the MELAprint®42 _________________________________________________________
The accessory printer MELAprint
item which is not supplied with the normal scope of delivery for the autoclave. Follow the instructions in Section
6.3.1.1.1 of this manual to connect the printer to the autoclave.
3.1.2 Initialisation of the printer / setting for immediate printout____________________________________
Follow the instructions in Section 6.3.1.1.2 of this manual to initialise the external printer after it has been
connected. This initialisation will log the printer on to the computer control system in the autoclave. Follow the
instructions in Section 6.3.2 of this manual to set the immediate-printout option to "Yes". This means that the
system will automatically print out a report when the system program has come to an end.
®
42 is designed for connection to the autoclave Vacuklav®31-B. This is an optional
Page 10
Operating Manual for the Vacuklav
3.2 Vacuum test ____________________________________________________________
A vacuum test (leak test) should be conducted under the following conditions to assure proper functioning of the
autoclave:
− Once when a new autoclave is set up for the first time
− When the autoclave is being put back into service after a long period of inactivity
− When the autoclave is moved to a new location
− Periodically as part of routine operations.
The vacuum test should be performed on the cold
Operator action:
1. Switch on the autoclave
at the power switch: After
the following message ...
... the display will show the
initial state:
Shown on the display:
autoclave as follows:
Program
Please wait
Door unlocking
Program
14:27:12
®
31-B
Start-Stop
Start-Stop
0.02bar 25°C
2. Press the “Program” button
several times until the program “Vacuum test” appears.
3. Close the door
4. Press the “Start – Stop”
button
5. The vacuum pressure has
been reached. The equilibration time begins to run (waiting time until beginning of
measurement).
6. After the waiting time has
run out (5 min), the measuring time will begin (here:
7 min 52 s).
Program
Start-Stop
Vacuumtest
Program
Start-Stop
Door
closed
Program
Start-Stop
Vacuumtest: Start
Press. = 980mbar
Program
Start-Stop
Vacuumtest 00:00
Press. = 80mbar
Program
Start-Stop
Vacuumtest 07:52
Pressure = 82mbar
Page 11
Operator action:
After the measuring time has
7.
run out (10 min), the cham-
ber is vented, and ...
... then the leak rate is given
(with connected printer and
setting for immediate printout = “Yes”, a record will
now be printed out. After the
message “Please wait –
door unlocking”, the following request will appear:
Shown on the display:
Program
Start-Stop
Ventilate
–0.56 bar 25 °C
Program
Start-Stop
Leak rate = 0.2 mbar
Program
Start-Stop
Open door
please
8. Open the door.
Program
Start-Stop
Vacuum test
If the leak rate determined is greater than the maximum permissible value, the following will appear in the display
and on the printed log: “Test not successful”. In such a case, follow the instructions given in this manual under
Section 7.3.
3.3 Trial run _______________________________________________________________
To continue the functional check of the autoclave, carry out the trial run by loading the autoclave with a typical load
for your daily work, and then run the program “Universal program, 134°C, wrapped”. Load the autoclave and use
the “Program” button to select this program. Then press the button “Start – Stop”. If the program runs correctly
(also see Section 4.7 here), the following message will appear at the end of the program:
Universal program
run successfully
Program
This display will also show the maximum values achieved for pressure and temperature. If a printer is connected,
and with setting for immediate printout = “Yes”, a record will now be printed out.
3.4 Installation log – sent to MELAG ___________________________________________
As documentation that the autoclave has been set up properly, an installation record should be produced by an
authorised person and a copy sent to MELAG. This is important in the event that you wish to make claims under
warranty provisions.
*Danger: Be careful when opening the door of the autoclave, especially if you are
interrupting a drying cycle. Small amounts of residual steam can escape through the door
and can burn operating personnel.
* Warning: After opening the autoclave door, do not touch exposed hot metal parts.
Warning: These parts can burn operating personnel. Be sure to use a MELAG or standard
tray handle to remove still-hot trays from the autoclave. Be sure to use suitable hand
protection (for example, potholders or gloves) when removing other sterilizing containers.
*We recommend the installation of a leak monitor in the water supply line (see Section 2,
Installation, for instructions).
*If you wish to connect to this autoclave water-treatment systems which are made by other
equipment manufacturers, get in touch with the MELAG company before you install this
equipment.
* This device is not suitable for the sterilization of liquids.
* According to presently valid regulations published by the Society of German Electrical
Engineers (VDE), this device is not suitable for operation in hazardous areas which are
endangered by the possibility of explosion (for example, in hazardous atmospheres).
*Important: Empty the water supply tank before moving the autoclave. If the autoclave has
water inside, do not tip it.
*This device may be repaired only by the company MELAG or by persons authorised by
MELAG (specialist personnel or customer-care staff). These persons must use original
spare parts and must observe the instructions for service published by MELAG.
*Danger: Do not open the enclosure to this device before disconnecting the electrical power
plug from its outlet.
*This autoclave will properly sterilize only if the operating personnel observe the instructions
in this Operating Manual – especially the instructions for loading the autoclave in
accordance with the program required.
Page 13
4 For each sterilization cycle
4.1 Water and power _______________________________________________________
This autoclave automatically checks to make sure that there is enough demineralized / distilled water, and it checks
that the quality of the demineralized / distilled water is sufficient for starting a program. In order to start a program
without delay, to avoid error messages, and to prevent a program from stopping before it is finished (see Sections
7.3 and 7.4), please make sure to carry out the following:
1. Before starting the first sterilization cycle of a day, check to make sure that there is enough water in the right
side of the dual-chamber water supply tank (this applies if you are using the system’s own internal supply: i.e., if
you are not using a connected water-treatment system). If there is not enough water in the tank, fill with water of
sufficient quality (see Section 8.3.2 of this manual).
2. If you use a direct water supply from the MELAdem
to the autoclave is open in time. If the pressure tank is empty, this feed must be opened at least one
starting a program. This step is not necessary if the water supply remains open continuously .
4.1.2 Power supply________________________________________________________________________
Switch on the power switch at the bottom right of the front of the autoclave. The message “Please wait – Door
unlocking” will appear in the display. After approx. 15 s, the device will then be in its initial state.
Time of day (h:min:s)
Chamber pressure (bar)
Temperature (steam) (°C)
®
47 water-treatment system, make sure that the water feed
hour before
Program
Start-Stop
14:27:12
0.02bar 25°C
4.2 Preparation of instruments ________________________________________________
MELAG — use of non-rusting materials for parts which come into contact with steam
All parts of the Vacuklav
sterilization chamber and the chamber door are of stainless steel, steam lines are Teflon
connections and solenoid valves are brass.
Drag-in rust
The non-rusting parts used in the Vacuklav
cases in which rust has developed on the autoclave or on the sterilized items inside, investigations have
consistently shown that this rust has been dragged into the autoclave from the outside by rusty instruments. We must point out that rust can form on stainless-steel instruments made by even the best manufacturers.
One main cause of instrument rusting is improper treatment with chemical cleaning and disinfecting agents during
preparation for sterilization.
Preparation of instruments for sterilization
The problem of drag-in rust shows how important it is to properly prepare items for sterilization. It is urgently
necessary to observe the following procedures:
1. Clean and prepare (for example, by oiling) handpieces and contra angles as described in the instructions
provided by the manufacturers of these items.
2. Immediately after their medical use, all other instruments must be placed in a disinfection and/or cleaning
solution in accordance with UVV/VBG 103 and properly cleaned and disinfected there.
3. Always be sure to use the correct dosage of disinfection and/or cleaning agents in these solutions.
4. Important: Be sure to carefully follow the instructions for the length of time that the instruments must soak in
the solutions.
5. We also recommend the use of ultrasonic equipment, recommended cleaning and care equipment for
handpieces and contra angles, as well as thermal disinfecting techniques.
6. The proper cleaning of instruments is also essentially important in preventing particles of soiled matter from
separating from the instruments during sterilization and from clogging the filters, nozzles, and valves of the
autoclave. It is extremely important to use a brush to very carefully clean the locks, joints, and hinges of the
instruments before sterilization.
7. Important: Use a brush under running water to completely rinse all cleaning and disinfecting agents from
instruments before placing them in the autoclave. Residue of chemical cleaning and disinfecting agents must by no means be allowed to enter the autoclave, since they will cause corrosion there.
8. As a final step before placing the instruments into the autoclave, rinse them with demineralized water, and dry
them well.
9. Be sure to use oil (as indicated in the manufacturers’ instructions) to prepare turbines and handpieces for
sterilization, in order to ensure their long service life.
®
31-B which come into contact with steam are made of non-rusting materials. The
®
31-B ensure that the autoclave itself cannot cause rust to form. In any
®
, and threaded
Page 14
Operating Manual for the Vacuklav
Instruments new from the factory:
Be sure also to follow the above-described pre-cleaning procedure even for instruments new from the factory.
Often there are small amounts of oil, grease, and impurities still on new instruments, from their production.
Important
procedures which are provided by manufacturers of instruments.
4.3 Loading the autoclave ____________________________________________________
The correct loading of the autoclave is essentially important in guaranteeing the sterilization effectiveness and the
proper drying of the items being sterilized. Be sure to observe the following basic instructions in loading the
autoclave:
Tray mounts:
There are 2 different tray mounts for the Vacuklav
1. Tray mount “B” (MELAG art.-no. 40234) for loading of either a maximum of 4 trays, or 4 standard tray cassettes
2. Tray mount “C” (MELAG art.-no. 40232) for loading of either a maximum of 6 trays, or 3 standard tray
In normal cases the autoclave should always be operated with a tray mount, since this ensures optimal steam
penetration and drying. In exceptional cases, it may be possible to remove the tray mount and place the suitable
containers directly into the sterilization chamber (for example, with use of sterilization containers of other
manufacturers). But please check with your specialist dealer or with MELAG before attempting this solution.
For sterilization of instruments sealed in see-through sterilization packaging, we recommend the use of the MELAG
Package Stand (MELAG article no. 22410). The Package Support considerably improves the results of the drying
process for instruments wrapped in this way.
Trays:
Trays used for the loading of sterilized items must be perforated, in order to allow the condensate to run out.
MELAG trays fulfil this requirement. We cannot advise the use of non-perforated trays or bowls (e.g., unperforated
standard-tray supports), since they do not allow proper trying.
Closed sterilization containers:
Closed sterilization containers must be perforated on at least one side (preferably the bottom), or they must have
valves, in order to allow the steam to properly penetrate and the condensate to flow out. All MELAG sterilization
containers satisfy this requirement, with perforations on two sides and with a sterile filter cloth insert. We cannot
recommend sterilization with sterilization containers with holes only on the top, since they do not allow proper
drying.
See-through sterilization packaging:
If you use see-through sterilization packaging (for example, MELAfol
tray for sterilization. We recommend the use of MELAG Foil Holders (MELAG article no. 22410).
Important: Do not stack the see-through sterilization packages horizontally on top of each other. If the sealing
seam of the package breaks during sterilization, please use a longer sealing pulse on the MELAG MELASEAL
Package Sealing Device, or seal with a double seam.
If standard tray cassettes are sealed in MELAfol
side seam and to prevent it from breaking open. In addition, be sure to press excess air out of the package before
sealing it.
Multiple packaging:
Air removal by sub-atmospheric pulsing allows multiple-wrapping.
Loading limits:
Be careful not to exceed the prescribed maximum loading amounts: 5kg of instruments or 1.8kg of textiles.
Mixed loads of textiles and instruments:
If you are sterilizing textiles and instruments at the same time, place the textiles above the instruments and do not
allow the textiles to come into direct contact with the instruments. Do not sterilize textiles and instruments in the
same sterilization container. Important: Do not by any means allow the textiles to come into direct contact with the
wall of the sterilization chamber. Please observe the following instructions for different types of packaging in one
sterilization cycle:
Place the see-through sterilization packaging and paper packing on the top (but if you are using a combination with
textiles: then arrange these packages on the bottom)
Sterilization of liquids:
Danger: The Vacuklav
: It is absolutely necessary to carefully follow the instructions for preparation and re-sterilization
®
31-B:
cassettes.
Both of these tray mounts are also designed to support the following MELAG sterilization containers:
Types 15K,M,G; Type 17K,R; Type 23R, Type 28M,G.
®
), these should be placed vertically on the
®
(250 mm wide), clamps or tape must be used to reinforce the
®
31-B is not suitable for the sterilization of liquids.
®
31-B
Page 15
Loading...
+ 33 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.