Melag Vacuklav 30 B User manual

Operating Manual
for the
Autoclave
Vacuklav30 B
Dear Doctor:
Thank you very much for the trust which you have shown by purchasing this autoclave.
As all other MELAG products, this autoclave was manufactured and tested according to strict quality criteria. Before placing this unit into operation, please read this Operating Manual carefully. The long-term functional effectiveness and the preservation of the value of your autoclave will depend on careful preparation of instruments before sterilization, and on proper care of the unit.
The staff and management of MELAG
To ensure the functional effectiveness of this unit and to preserve its value:
1. Prepare the instruments to be sterilized carefully
2. Take proper care of the autoclave
3. Use only pure distilled or demineralised water
1 Description of the unit................................................................................................................. ..................4
1.1 Views of the unit .................................................................................................................. ....................4
1.2 Control panel ............................................................................................................................................ 5
1.3 Technical data ............................................................................................................................ ..............5
1.4 Performance features of this autoclave ............................................................................................. ....5
1.4.1 Fractionating pulsed vacuum procedure ................................................................................... ........5
1.4.2 Programs for sterilization / disinfection ............................................................................ ..................5
1.4.3 Separate steam generation .................................................................................... ..........................5
1.4.4 One-way system/ Conductivity measurement/ Automatic water supply ............................................5
1.4.5 Electronic Parameter Control (EPS) ................................................................................. .................6
1.4.6 Vacuum drying ....................................................................................................................... ............6
1.4.7 Pre-heating ......................................................................................................................... ...............6
1.4.8 Documentation ...................................................................................................... ...........................6
2 Installation................................................................................................................................... ...................6
2.1 Setting up the autoclave .................................................................................................. ......................6
2.2 Transport ribbons ....................................................................................................................... .............6
2.3 Levelling the autoclave ...................................................................................................... .....................7
2.4 Power supply ................................................................................................................... .......................8
2.5 Cooling water inlet connection ............................................................................................................ ..8
2.6 Cooling water outlet connection ........................................................................................ ...................8
2.7 Connection of the feed water storage container ................................................................................ ..8
2.8 Connection of a water purification system ........................................................................................... 8
3 Initial start-up ...................................................................................................................... ..........................8
3.1 Printer connection/ Initialisation (optional) ............................................................................... ............8
3.1.1 Connecting the MELAprint®42 ...................................................................................... ....................8
3.1.2 Initialisation of the printer / Setting up immediate print-out ............................................ ...................8
3.2 Vacuum test ................................................................................................................. ............................8
3.3 Test run ........................................................................................................................................... .......10
3.4 Installation record .............................................................................................................................. ...10
3.5 Safety instructions ..................................................................................................... ..........................10
4 Instructions for all sterilization procedures .................................................................................... ..........11
4.1 Electricity and water supplies .............................................................................................................. .11
4.1.1 Cooling water and purified feed water ......................................................................................... ....11
4.1.2 Power supply ........................................................................................................... .......................11
4.2 Loading the autoclave ............................................................................................................ ...............11
4.3 Closing the door ................................................................................................................ ...................12
4.4 Program selection ....................................................................................................... .........................12
4.5 Basic display (no program selected)Program start .................................................................. .........13
4.6 Program progress .............................................................................................................................. ...13
4.7 Print-out record ................................................................................................................................... ..16
4.8 Removing the sterilized items ........................................................................................... ..................16
4.9 Sterile storage ................................................................................................................................ .......16
4.10 Sterilization frequency / pauses .......................................................................................... ..............17
4.11 Manual termination of program ....................................................................................... ...................17
4.11.1 Termination of sterilization/ disinfection .................................................................................... ......17
3.To undo the program termination, press the " - " button. ................................................................... ...17
4.11.2 Terminate drying ................................................................................................ ...........................18
4.12 4This is followed by ventilation of the autoclave chamber................................................... ............20
4.13 Disinfection ........................................................................................................................ ..................21
4.14 Reaction to warnings / error messages ................................................................................. ...........21
4.15 Operational pauses ................................................................................................... ..........................21
5 Closing down / Transport / Reinstallation ........................................................................................ .........21
6 Special functions ............................................................................................................ ............................22
6.1 Water quality (conductivity)/ Chamber temperature .................................................................. ........22
Page 2
Operating manual - Vacuklav
®
30-B
6.2 Selecting extra drying ..................................................................................................... .....................22
6.3 Records / Load documentation ......................................................................................... ...................22
6.3.1 Record print-out ................................................................................................................... ...........23
6.3.2 Immed. print-out? Yes/No ........................................................................................... ....................24
6.3.3 Printing out stored records .......................................................................................................... ....25
6.3.4 Print all stored cycle records ....................................................................................................... ....26
6.3.5 Display printer memory ........................................................................................................... ........27
6.3.6 Deleting cycle records ........................................................................................... .........................28
6.3.7 Test print-out ........................................................................................................................ ............29
6.4 Resetting date and time .................................................................................................. ......................29
6.5 Automatic preheating ............................................................................................................... .............31
6.6 Total load count ............................................................................................................... .....................32
6.7 Program modifications ...................................................................................................................... ....32
7 Operational errors / Malfunctions .................................................................................................. ............32
7.1 What to do if the autoclave malfunctions ............................................................................... ............32
7.2 Malfunctions without display messages ....................................................................... ......................33
7.2.1 No display .............................................................................................................................. .........33
7.2.2 Excessive water consumption .............................................................................. ...........................33
7.2.3 Poor drying ................................................................................................................ .....................33
7.3 Warning messages ............................................................................................................. ..................33
7.4 Error reports .......................................................................................................................................... .36
8 Taking care of your autoclave......................................................................................................... ............39
8.1 Preparation of instruments ...................................................................................................... ............39
8.2 Rust formation = Drag-in rust .................................................................................................... ..........40
8.3 Taking care of the Vacuklav®30-B ..................................................................................................... ...40
8.3.1 Cleaning ............................................................................................................... ..........................40
8.3.2 Use of demineralised or distilled water ........................................................................ ...................40
8.4 Checking the operation of the autoclave ........................................................................................ .....41
8.4.1 Safety with automatic monitoring ........................................................................................ ............41
8.4.2 Periodical bacteriological testing (twice a year) ............................................................... ................41
8.4.3 Maintenance recommendations ................................................................................................. .....41
9 Annex ..................................................................................................................................... ......................42
9.1 Installation options ...................................................................................................................... .........42
9.2 Additional technical data ........................................................................................... ..........................43
9.2.1 Capacity/Weight ............................................................................................................................ ..43
9.2.2 External supplies ..................................................................................................... .......................43
9.2.3 Operational parameters ..................................................................................................... ..............43
9.3 Instructions on drying ................................................................................................................. ..........44
9.3.1 Drying in sterilization containers .................................................................................... .................44
9.3.2 Textiles ........................................................................................................................................ .....45
9.3.3 Instruments ............................................................................................................................ ..........46
9.3.4 Loading the autoclave ................................................................................................................... ..46
9.3.5 Loading containers with soft sterilization packing material .................................... .........................46
9.3.6 Stacking sterilization containers ......................................................................... ............................47
9.3.7 Removing the sterilized items ............................................................................................... ..........47
9.3.8 Improving the drying ..................................................................................................... ..................47
Page 3

1 Description of the unit

1.1 Views of the unit_______________________________________________________
1 Safety valve 9 Fuses - 2 x 16 A / FF 2 Sterile filter 10 Power switch 3 Cooling water inlet (3/4" external thread) 11 Motor circuit breaker (vacuum pump) 4 Cooling water outlet (3/4" external thread) 12 Sliding door lock 5 Purified water inlet
(Swivel connection for 6x1 pipe)
13 Door (left-side hinge) 14 Control panel
6 Mains power cable 7 Adjustable front feet 8 Serial data and printer port
( RS 232 )
Page 4
Fig. 1 Views of the Vacuklav®30-B
Operating manual - Vacuklav
®
30-B
1.2 Control panel_________________________________________________________
Fig. 2 Control panel Vacuklav30-B
1.3 Technical data________________________________________________________
Sterilization space (diameter x depth ) : 25cm x 36cm
Electric power supply : 2400W / 230V AC / 10.4A / 50....60Hz
Sterilization pressure / temperature : 2bar/134°C; 1bar/121°C Disinfection pressure/ temperature : 0.3 bar/105°C Maximum load: : 5kg instruments or 2kg textiles
Further technical details are included in the Annex
1.4 Performance features of this autoclave___________________________________
1.4.1 Fractionating pulsed vacuum procedure _______________________________________________
A fractionating pulsed vacuum method involved the repeated evacuation of the sterilization chamber, alternating with the introduction of steam in order to ensure the necessary penetration of the items undergoing sterilization with superheated steam. This means that even demanding sterilization tasks, involving for example intricate instruments or large amounts of textiles can be completed quickly and reliably. Steam penetration can be demonstrated using the established Bowie & Dick Test, for which a special test program can be implemented.
1.4.2 Programs for sterilization / disinfection________________________________________________
The Vacuklav® 30-B has two sterilization programs, the "universal program" at 134°C: for packed utensils and the "fast program" for unpacked utensils, as well as a sterilization program for textiles and rubber articles at 121°C, and a disinfection program at 105°C. In addition, it is possible to check at any time that the autoclave is functioning properly by using the test program for steam penetration (the Bowie & Dick Test Program), as well as to test for leaks (the Vacuum Test Program).
1.4.3 Separate steam generation __________________________________________________________
The generation of steam by a high-performance unit outside the sterilization chamber makes it is possible to sterilise large loads of instruments or textiles quickly. It is not necessary to have any pauses between the sterilization runs, and overheating in the sterilization chamber is impossible.
1.4.4 One-way system/ Conductivity measurement/ Automatic water supply ______________________
The Vacuklav® 30-B operates with a reliable one-way system, in which all the condensate from a sterilization, including any impurities which this may contain is completely eliminated and purified water used for the next program run. An integrated conductivity measurement system monitors the quality of the distilled or demineralised water used for the steam generation. For frequent operations, the larger demand for purified water can be met by a water purifier, e.g. MELAdem®47 or MELAdem® 37, which can be directly connected to the autoclave. Provided the instruments to be sterilized are then also prepared carefully, stains on the load and soiling of the autoclave itself can be prevented.
Page 5
Program
Start - Stop
Buttons for selecting, adjusting and displaying special functions: Printing, date/time, pre-heating, total loads, conductivity, re-setting
Program selection
for sterilization programs/ test programs
and for selecting options (sub-menus) of the
special functions
Start - stop buttons to start and interrupt programs, drying, and to control special functions
2 x 20- alphanumeric LED-display, showing program status, and parameters for basic and special functions
1.4.5 Electronic Parameter Control (EPS)____________________________________________________
The microprocessor in the Vacuklav® 30-B makes it possible to monitor pressure, temperature and time continuously during a program by Electronic Parameter Control The overall operating time can then be optimised according the load and the temperature of the autoclave. The process assessment and monitoring system in the program control compare current process parameters with standard process data and monitors the process relative to limit values for temperatures, times and pressures. This makes it possible to identify faults as they occur, and provides quality assurance for the sterilization and disinfection processes.
1.4.6 Vacuum drying_____________________________________________________________________
Vacuum drying ensures the best possible drying results for packed utensils.
1.4.7 Pre-heating________________________________________________________________________
By activating the "pre-heating" function, the cold autoclave chamber can be warmed up before sterilization, or the temperature can be maintained between sterilization runs. This reduces the duration of cycles and considerably reduces the formation of condensation, thus improving drying results.
1.4.8 Documentation____________________________________________________________________
The electronic memory stores records of the previous 40 programs. For effective hard-copy documentation and for checking purposes a MELAprint®42 printer can be connected to print out a record immediately after completion of a program or to print out records from the memory.

2 Installation

When setting-up and installing the autoclave, please consult the separate instructions leaflet "Installing the Vacuklav® 30-B".
2.1 Setting up the autoclave _______________________________________________
The autoclave should be set up in a dry place which is protected against dust. The base should be stable, and able to support the weight of the appliance (unloaded weight 43 kg). The space required by the autoclave can be seen by the external dimensions (as in Section 1.1). A minimum additional space of 10 cm should be allowed on either side and above the autoclave in order to ensure that heat can escape. The electrical power supply should be a separate 230V AC circuit with a 16 A fuse. The cooling water should be obtained from a mains water supply, to which the autoclave is connected by means of a water tap or stop cock with a 3/4" external thread. This water supply should be accessible close to point where the autoclave is set up. (The standard connection pipe supplied is 2 m long). The disposal of the cooling water outflow requires either a wall-outlet (NW 40) or a sink outflow, and this should also be situated close to the autoclave (the standard 16 mm connection pipe provided is 2m long). The outflow must be lower than the working surface on which the autoclave stands, so that the outflow pipe can be installed without bends and kinks, ensuring a constant downward flow. The autoclave should set-up at a level which allows easy loading and at which the display can be read clearly without bending. The autoclave must also be supplied with purified water for steam generation. The distilled or demineralised water can come either from a storage container (MELAG Art.-No. 244) or directly from the container in which purchased water is supplied. This should also be situated close to the autoclave (length of feed pipe supplied
1.5m, maximum height difference 1.5 m). If the purified water comes directly from a water purification system MELAdem®47 or MELAdem®37 (or another purification system) then a storage container is not necessary, but space must be allowed to accommodate the water purification system.
2.2 Transport ribbons_____________________________________________________
Take the autoclave out of the packaging by means of the transport ribbons. The ribbons themselves are each removed by undoing the two retaining screws, which must then be screwed firmly back in place without washers.
Page 6
Operating manual - Vacuklav
®
30-B
2.3 Levelling the autoclave________________________________________________
In order to ensure that condensate can drain out of the autoclave (which is important if it is to operate properly) the appliance must have be clearly higher at the front than at the back. The autoclave should first be installed in a horizontal position (this should be checked with a spirit level at the chamber flange) and then the front feet should be extended by giving them at least five (5) turns.
5
3211
78916410
210x230
340
Fig. 3 Typical installation of the Vacuklav30-B with storage container for distilled/demineralised water, with a wall-mounted trap
1
Vacuklav30-B rear view 2 Cooling water inlet pipe (pressure pipe) 3 Cooling water outlet pipe DN16 4 Water stop cock with safety combination in accordance with DIN 1988 (MELAG Art.-No.: 37310) 5 Leak detector with stop valve and sensor (MELAG Art.-No.: 1056) 6 Wall- mounted trap (MELAG- Art.- No.: 37410) 7 Wall outlet (NW 40) 8 Storage container for distilled or demineralised water (MELAG Art.-No. : 244) 9 Feed water intake with filter 10 Y-junction with non-return valve (included in (6) ) 11 Power connection
Page 7
2.4 Power supply ________________________________________________________
The electric cable of the appliance is plugged into a mains socket rated at 230 V, 50 Hz. The power rating of the autoclave is 2400 W. In order to avoid overloading the electricity supply, we recommend using a separate electrical circuit fitted with a 16 A fuse and optionally protected with a 30mA circuit breaker.
2.5 Cooling water inlet connection __________________________________________
The connection for the cooling water inlet at the back of the appliance is connected to the mains water supply using the pressure pipe supplied (see page 7, Fig. 3.). Make sure that the washers are fitted correctly. In order to avoid water damage in the event of leaks, we also recommend the installation of a leak detector ("Water-stop", MELAG-Art.-No.:1056).
2.6 Cooling water outlet connection __________________________________________
The connection for the cooling water outlet at the back of the appliance is connected to the drainage system of the building by means of the outlet pipe supplied (textile-reinforced transparent pipe, DN16). It is important that the pipe should have a steady downward gradient, without twists and kinks. When connecting to an existing sink outflow, the use of the double-chamber trap supplied helps to reduce noise due to water discharging. When connecting to a separate outflow pipe NW40, a wall-mounted trap should be used (MELAG Art.-No. 37410 ,see Fig. 3).
2.7 Connection of the feed water storage container ____________________________
If the autoclave is to be supplied with distilled or demineralised water from a storage container (see Fig. 3) then this will be drawn up through the suction pipe. The one end of the pipe is attached to the inlet at the rear of the autoclave, either directly or by means of the swivel connection. The pipe is then fed into the storage container until its other end with the inlet filter reaches the bottom. If you use the MELAG storage container (MELAG Art.-No.: 244) you will find the necessary instructions on the label. This container has a volume of 11.5 litres, enough for an average of 25 sterilization cycles.
2.8 Connection of a water purification system ________________________________
The autoclave can be connected directly to a water purification unit which provides the demineralised or distilled water required (see page 39, Fig. 4). Instead of connecting the feed water inlet to a storage container, it is simply connected directly to the water purification unit. The reverse-osmosis systems MELAdem®47 and MELAdem®37 are ideally suited to provide the quantity and quality of water needed for the Vacuklav®30-B. Detailed instructions on the installation of these water purification units are provided in their operating manuals. When connecting water purification systems from other manufacturers it is very important to ensure that they are able to provide sufficient amounts of water at the required purity. You are generally advised to consult MELAG first.

3 Initial start-up

3.1 Printer connection/ Initialisation (optional)________________________________
3.1.1 Connecting the MELAprint®42________________________________________________________
The autoclave can be connected to an external printer, the MELAprint®42. This is not supplied as standard with the autoclave. In order to connect the printer to the autoclave follow the description in Section 6.3.1.1.1.
3.1.2 Initialisation of the printer / Setting up immediate print-out ________________________________
Initialise the external printer (registering with the processing unit of the autoclave) as described in Section
6.3.1.1.2. In order to select the immediate print-out option, which means that a record of each sterilization is printed out automatically as soon as it has finished, proceed as described in Section 6.3.1.3.
3.2 Vacuum test__________________________________________________________
In order to check the operation of the autoclave in the course of the initial start-up, after long periods without being used, or after moves, as well as periodically during routine use, a vacuum test should be carried out to check for leaks. The vacuum test should be carried out as follows, preferably on the cold autoclave:
Page 8
Operating manual - Vacuklav
®
30-B
Operation Display message
1.
S
wit ch on po we r,
2.
Pr es s th e "P ro
3.
Cl os e th e do or
4. Pr es s "St art
­St op " bu tto n
5.The evacuation pressure has been reached.6.After waiting for the equalisation (5 min) the measuring period starts (running
here e.g. for 7min 52s)7.After the measuring period (10 min) the chamber is ventilated and
Page 9
Program
Start-Stop
Leak rate 0. 2 m b a r
Program
Start-Stop
V e n t i l a t i o n - 0 . 5 6 b a r 2 5 ° C
Program
Start-Stop
V a c u u m t e s t
Program
Start-Stop
D o o r c l o s e d
Program
Start-Stop
V a c u u m t e s t : S t a r t P r e s s u r e
Program
Start-Stop
V a c u u m t e s t 0 0 : 0 0 P r e s s u r e 8 0 m b a r
Program
Start-Stop
V a c u u m t e s t 0 7 : 5 2 P r e s s u r e 8 2 m b a r
O p e n d o o r p l e a s e
Program
Start-Stop
Program
Start-Stop
P l e a s e w a i t D o o r u n l o c k i n g
Program
Start-Stop
1 4 : 2 7 : 1 2 0 . 0 2 b a r 2 5 ° C
Operation Display message
8. Open door
If the leakage rate is above the limit value, then the display (and the print-out) will also show "Test unsuccessful". In this case, follow the instruction in Section 7.3.
3.3 Test run _____________________________________________________________
In order to check the operation of the autoclave under realistic conditions, a test run should be carried out with the "Universal Program, 134°C wrapped" and a relevant load. After loading the autoclave and selecting the program with the "Program" button, sterilization is started by pushing the "Start/Stop“ button. If the program runs correctly, the following message will appear on the display (see Section 4.6 ):
with the values for the maximum values for pressure and temperature. If the immediate printout option has been selected for an external printer a record of the program run will be printed.
3.4 Installation record ____________________________________________________
As documentation that the autoclave has been set-up properly, an installation record should be produced by an authorised person and a copy sent to MELAG. This is important in the event that you wish to make claims under warranty provisions.
3.5 Safety instructions ____________________________________________________
When opening the door, particularly after interrupting the drying process, residual steam can escape from
the autoclave chamber.
After opening the door, do not touch any metal surfaces - these will be hot! Danger of burns. Always use
the tray lifter to remove trays, or wear suitable hand protectors when taking out other items
We recommend the installation of a water leak detector (see installation details Section 2).
If you intend to install a water purification unit from another manufacturer, then consult MELAG before you
do so.
The appliance is not suitable for sterilising liquids.
Under current VDE-regulations, this appliances is not suited for use in areas where there are risks of
explosion.
The appliance must only be serviced and repaired by MELAG or by its authorised representatives
(specialist dealers or customer services) using only original parts and following service instructions.
Before opening the housing always disconnect from the mains power supply!
In order to ensure effective sterilization with the autoclave observe the instructions in this operating manual,
and in particular ensure that the loading of the autoclave is appropriate for the program selected.
Page 10
Program
Start-Stop
Vacuumtest
Universal-progr.run successfully
Program
Start-Stop
Operating manual - Vacuklav
®
30-B

4 Instructions for all sterilization procedures

4.1 Electricity and water supplies___________________________________________
4.1.1 Cooling water and purified feed water _________________________________________________
The autoclave automatically monitors the availability of cooling water and purified water, as well as the quality of the distilled / demineralised water before starting a program. In order to allow an immediate program start and to avoid error reports or interruptions of programs (see Sections: 7.3 and 7.4) :
Before the first sterilization at the start of the working day, check that the water supply is turned on (stop
cock)
If purified feed water is drawn from a storage container, check the water level, and if necessary fill up with
purified water of appropriate quality (see Section 8.3.2)
If the feed water is drawn directly from a MELAdem
®
47 water purifier, check that its water supply is turned on in good time (this may be up to an hour before starting a sterilization program), if the water supply has been turned off over night, for example.
4.1.2 Power supply _____________________________________________________________________
Switch on the power using the switch on the front of the autoclave (bottom right). About 15 seconds after the message "Please wait Door release“ the appliance is in the start status :
4.2 Loading the autoclave_________________________________________________
It is of crucial importance for effective sterilization and good drying that the autoclave is loaded properly:
Tray rack
For the Vacuklav® 30-B there are 2 types of tray rack: Type "A" (MELAG-Art.-No.: 40231) for up to 6 trays and Type "B" (MELAG-Art.-No.: 40234) for up to 4 trays or 4 standard tray-cassettes. Both types of tray rack are also suitable for the MELAG-sterilization containers Type 15K,M,G; Type 17 K,M,G; Type 17R; Type 23R, Type 28 M, Type 28G). Normally, the autoclave should be used in conjunction with a tray rack, since this ensures that steam penetration and drying are as good as. In exceptional situations (e.g. when using sterilization containers from other manufacturers), and after consultation with your specialist dealer or with MELAG, the tray rack can be removed and the container can be placed directly in the autoclave chamber.
Trays
Trays for objects which are to be sterilized must be perforated, in order to allow condensate to run away. MELAG-trays are recommended. If you use dishes or trays without perforations, then the objects being sterilized will not dry properly.
Enclosed sterilization containers
Enclosed sterilization containers must be perforated on at least one side (preferably underneath) or must have valves, in order to ensure that steam can penetrate and condensate can run out. All MELAG-sterilization containers meet these requirements with perforations on two sides and filter-cloth- inlays. Sterilization containers which only have perforations on the top only allow limited drying. If sterilization containers are stacked in the autoclave, it is important to ensure that the perforations are not blocked.
Transparent sterilization packaging
If you use transparent sterilization packaging, such as MELAfol® , then the items should if possible be stood vertically on the tray, or sterilized in foil holders (MELAG-Art.-No.: 238). They should never be laid flat one on top of the other. If seals split open during sterilization it may be necessary to increase the length of the sealing impulse or to use a double-seal.
Page 11
Program
Start-Stop
1 4 : 2 7 : 1 2 0 . 0 2 b a r 2 5 ° C
Time (24-hour)
Chamber pressure (bar)
(steam)-temperature (°C)
Standard tray- cassettes sealed in MELAfol® (250 mm wide) must be taped and clasped additionally to ensure that the side-seals do not split open.
Multiple wrapping
The pulsed vacuum method means it is possible to use multiple wrapping.
Maximum loads
Loads should not exceed 5kg of instruments or 2kg of textiles.
Mixed loads
If mixed loads of textiles and instruments are to be sterilized, then as far as possible the textiles should be above the instruments and direct contact with the instruments should be avoided. Inclusion of textiles and instruments in the same sterilization container is not desirable. Textiles should never come into direct contact with the walls of the chamber. If different types of packaging are included in a load, then:
Instruments and sterilization containers should be at the bottom
Transparent and paper sterilization packaging should be at the top (but lower than textiles)
Liquids
The appliance is not suitable for the sterilization of liquids!
Suitability for sterilization
Relevant information provided by manufacturers of instruments and textiles about sterilization should be strictly observed.
4.3 Closing the door _____________________________________________________
The door is closed by lightly applying pressure in the direction of the chamber flange and at the same time pressing down the sliding door catch. The display shows the message:
4.4 Program selection ____________________________________________________
A program should be selected which is appropriate for the physical properties of the items being sterilized (and in particular their heat resistance) and the type of packaging (if any part of the load is wrapped, then either the "Universal Program" or the "Gentle Program" must be used). By pressing the "Program selection" button it is possible to review the display of the following programs for selection:
Program name/Display message Parameter/Application
Universal-progr. at 134°C, 2 bar, and a sterilization time of 3.5 min Sterilization of all types of wrapped items, in particular instruments, or mixed loads (unwrapped/wrapped)
Quick-program at 134°C, 2bar and a sterilization time of 3.5 min Sterilization only of unwrapped instruments (no textiles) for rapid re-use (drying can be interrupted manually)
Page 12
Program
U n i v e r s a l - p r o g r .
1 3 4 ° C
Program
Q u i c k - p r o g r a m
1 3 4 ° C u n w r a p p e d
Program
Start-Stop
D o o r c l o s e d
Operating manual - Vacuklav
®
30-B
Gentle-program at 121°C, 1bar and a sterilization time of 15min
Program name/Display message Parameter/Application
Disinfection program at 105°C, 0.3 bar and a disinfection time of 8 min for steam-pressure disinfection of materials with physical properties which mean that they cannot be sterilized.
Warning this is not sterilization!
Bowie&Dick Test Program at 134°C, 2 bar and a
sterilization time of 3.5 min Used to check the operation of the autoclave (Steam penetration of special indicators)
Vacuumtest-program Used to check the autoclave for leaks, from a cold start
4.5 Basic display (no program selected)Program start_________________________
Press the "Start-Stop" button once the desired program is shown on the display. The availability of cooling water, and feed water will be checked automatically, with a conductivity measurement.
At the start of the quick program there will be an additional message "Warning: Only unwrapped instruments“, at the start of the disinfection program the warning "Warning - Not sterilization“. These messages must be acknowledged by pressing "Start“ again.
4.6 Program progress ____________________________________________________
After starting the program, it will then progress automatically. The display shows the current program status as follows:
Page 13
G e n t l e - p r o g r a m
1 2 1 ° C w r a p p e d
Program
D i s i n f e c t i o n
1 0 5 ° C 0 . 3 b a r 8 ‘
Program
B o w i e & D i c k T e s t
1 3 4 ° C 2 . 2 b a r 3 . 5 m i n ‘
Program
V a c u u m t e s t
Program
Start-Stop
S t a r t
Program
1 5 : 3 1 : 3 3
0 . 0 2 b a r 2 2 ° C
Program
Program status Display message
1. Phase 1 consisting of evacuation
(removal of air) until the appropriate evacuation pressure is reached, followed by steam introduction up to slightly above atmospheric pressure. Display show pressure in chamber and steam temperature
2.
Ph as e 2 an d fol lo wi ng
De pe ndi ng on th e pr og ra m ch os en an d th e te m pe rat ur e of th e ch a m be r at th e sta rt of th e te m pe rat ur e, ad diti on al cy cle s m ay be ne ce ss ar y
Page 14
Program Start-Stop
1 . s u b - a t m . p u l s i n g
- 0 . 0 8 5 b a r 2 2 ° C
Program Start-Stop
2 . s u b - a t m . p u l s i n g
- 0 . 8 5 b a r 7 0 ° C
Operating manual - Vacuklav
®
30-B
Program status Display message
3.
He ati ng ph as e
A
he ati ng
­up ph as
4.
St eri liz ati on ph as e
W he n th e re
5.
Pr es su re rel ea se
A
fte r co m ple tio n of th e ste rili zat
6. Drying phase After pressure release the drying phase begins. At this point it is possible to stop the program without this leading to a fault being reported, since the sterilization itself is now completed. However, with the exception of the "Quick Program", the drying phase should be allowed to run to completion.
7. Ventilation 8. Program end
Page 15
Program Start-Stop
V e n t i l a t i o n
- 0 . 1 2 b a r 6 0 ° C
Program Start-Stop
O p e n d o o r
p l e a s e B i t t e T ü r
Program Start-Stop
Q u i c k - p r o g r a m
R u n s u c c e s s f u l l y
Program Start-Stop
H e a t u p
0 . 7 6 b a r 1 1 7 ° C
Program Start-Stop
S t e r i l i z a t i o n
S t i l l 2 m i n 1 2 s e c
Program Start-Stop
S t e r i l i z a t i o n
2 . 1 8 b a r 1 3 5 ° C
Program Start-Stop
P r e s s u r e r e l e a s e
0 . 8 5 b a r 9 6 ° C
Program
Start-Stop
I m m e d i a t e r e m o v a l
P r e s s S T O P
Program Start-Stop
V a c u u m - d r y i n g
1 ' - 0 . 9 b a r 9 9 ° C
Program Start-Stop
P l e a s e w a i t
D o o r u n l o c k i n g
Loading...
+ 33 hidden pages