Dear Doctor:
Thank you very much for the trust which you have shown by purchasing this autoclave.
For nearly 50 years now, MELAG — a medium-sized family-owned and -operated
company — has specialized in the production of sterilization equipment for medical
practice. During this period, MELAG has succeeded in becoming a leading manufacturer
of sterilization equipment. More than 300 000 MELAG units sold throughout the world
testify to the exceptional quality of our sterilizers. — which are manufactured exclusively
in Germany.
As all other MELAG products, this autoclave was manufactured and tested according to
strict quality criteria. Before placing this unit into operation, please read this Operating
Manual carefully. The long-term functional effectiveness and the preservation of the
value of your autoclave will depend on careful preparation of instruments before
sterilization, and on proper care of the unit.
The staff and management of MELAG
To ensure the functional effectiveness of this
unit and to preserve its value:
1. Prepare the instruments to be sterilized carefully
2. Take proper care of the autoclave
3. Use only pure distilled or demineralized water
CONTENTS
1 DESCRIPTION OF THE UNIT........................................................................................................................3
1.1 Views of the unit.......................................................................................................................................3
1.2 Control panel............................................................................................................................................5
2.1 Setting up the autoclave..........................................................................................................................6
2.2 Transport ribbons.....................................................................................................................................6
2.4 Mains power supply.................................................................................................................................8
2.5 Outlet connection for one-way water.......................................................................................................8
2.6 Internal water supply with demineralized/distilled water..........................................................................8
2.7 Connection of a water purification system...............................................................................................8
3.3 Test run..................................................................................................................................................10
4 INSTRUCTIONS FOR ALL STERILIZATION PROCEDURES....................................................................11
4.1 Electricity and water supplies ................................................................................................................11
4.1.1 Distilled or demineralized feed water ..........................................................................................11
4.1.2 Power supply...............................................................................................................................11
4.2 Preparing instruments for sterilization...................................................................................................11
4.3 Loading the autoclave............................................................................................................................12
4.4 Closing the door.....................................................................................................................................13
4.5 Program selection..................................................................................................................................13
4.6 Program start.........................................................................................................................................14
4.7 Program progress..................................................................................................................................14
5 CLOSING DOWN / TRANSPORT / REINSTALLATION .............................................................................20
Page 2
6 SPECIAL FUNCTIONS.................................................................................................................................21
6.1 Water quality (conductivity)/ Chamber preheating temperature............................................................21
6.2 Selecting extra drying............................................................................................................................21
6.3 Records / Load documentation..............................................................................................................21
6.3.1 Record print-out...........................................................................................................................22
6.6 Total load count .....................................................................................................................................31
6.7 Distilled / demineralized water supply....................................................................................................32
6.8 Water system.........................................................................................................................................33
6.9 Program modifications...........................................................................................................................33
9.3 Instructions on drying.............................................................................................................................46
9.3.1 Drying in sterilization containers.................................................................................................. 46
4 One-way water outlet (3/4") 13 Door (left side hinge)
distilled water supply 15 Tank lid
6 Mains power cable 16 Outlet connection for wastewater
7 Adjustable front feet 17 Outlet for distilled / demineralized water
8 Serial data and printer port ( RS 232 )
Page 4
12 Sliding door lock
14 Control panel 5 Pipe connection for internal demineralized /
Operating Manual for Euroklav
1.2 Control panel _______________________________________________________
2 x 20- alphanumeric LED-display, s
program status, and parameters for basic
and special functions
Fig. 2 Control panel Euroklav
Buttons for selecting, adjusting and
displaying special functions:
Printing, date/time, pre-heating,
total loads, conductivity, re-setting
29V-S
for sterilization programs/ test programs
and for selecting options (sub-menus) of the
1.3 Technical data ______________________________________________________
Sterilization space (diameter x depth ) : 25cm x 36cm
Electric power supply : 2400W / 230V AC / 10.4 A / 50....60Hz
Sterilization pressure / temperature : 2bar/134°C; 1bar/121°C
Maximum load: : 3kg instruments or 0.9kg textiles
Further technical details are included in the Annex
howing
Program
Program selection
special functions
Start - Stop
Start - stop buttons to start
and interrupt programs,
drying, and to control special
functions
®
29V-S
1.4 Performance features of this autoclave _________________________________
1.4.1 Prevacuum procedure combined with pulsed flow method_______________________________
With the pre-vacuum procedure, most of the air is evacuated from the sterilization chamber and the load in a
first step.
Then the pulsed flow method is used in which steam flows into and then out of the autoclave to ensure
effective penetration of the items to be sterilized by superheated steam.
This makes it possible to carry out demanding sterilization tasks rapidly and reliably, such as for example the
sterilization of wrapped instruments or textiles.
The stream penetration can be tested by implementing a special test program for the Bowie & Dick Test,
which is the standard test for large-scale sterilization. operations..
1.4.2 Operating range for sterilization_____________________________________________________
The Euroklav
and for unwrapped loads the "Quick program", and an additional sterilization program for textiles and rubber
articles at 121°C, the "Gentle program". The test program for steam penetration ("Bowie & Dick Test“) and a
test program for leaks ("Vacuum Test“) allow the operator to check the operation of the autoclave at any
time.
The powerful steam generation in the sterilization chamber makes it possible to sterilize large loads of
instruments or textiles in a short time. This system of steam generation means that excess temperatures in
the sterilized chamber are not possible.
1.4.4 One-way/Closed-loop system/ Conductivity measurement/ Automatic water refilling _________
The Euroklav
water and all dissolved impurities are led away at the end of every sterilization cycle, and fresh
demineralized or distilled water is then used. This is particularly good for all the materials used. However, in
some instances it is also important to use less water, and the Euroklav
conserving closed-loop mode. In this case the used water flows from the left side of the double-chamber
storage container over the separating wall into the right-hand chamber. This water is then used again in the
next sterilization cycle. With the closed-loop operation it is important that the instruments to be sterilized
have been very carefully washed and swilled in purified water. The water in the autoclave must also be
exchanged once a week.
An integrated electrical conductivity meter monitors the quality of the demineralized or distilled water used to
generate the steam. Using the recommended one-way operation for the autoclave, the increased
consumption of demineralized or distilled water can be provided by a water purifier such as the
MELAdem
Provided that the instruments are prepared carefully for the sterilization, stains on the load and soiling of the
autoclave itself can be prevented.
®
29V-S has two sterilization programs at 134°C: for wrapped loads the "Universal program",
®
29V-S can be operated in the tried and tested one-way system, in which steam, condensed
®
29V-S can also be used in a water-
®
47 or MELAdem®37, which can be directly connected to the autoclave.
Page 5
1.4.5 Electronic Parameter Control EPS ___________________________________________________
The microprocessor in the Euroklav
continuously during a program by Electronic Parameter Control The overall operating time can then be
optimised according the load and the temperature of the autoclave.
The process assessment and monitoring system in the program control compare current process parameters
with standard process data and monitors the process relative to limit values for temperatures, times and
pressures. This makes it possible to identify faults as they occur, and provides quality assurance for the
sterilization process.
1.4.6 Combined pressure pulsing and vacuum drying _______________________________________
This method ensures good drying results even for wrapped instruments.
By activating the "pre-heating" function, the cold autoclave chamber can be warmed up before sterilization,
or the temperature can be maintained between sterilization runs. This reduces the duration of cycles and
considerably reduces the formation of condensation, thus improving drying results.
29V-S makes it possible to monitor pressure, temperature and time
The electronic memory stores records of the previous 40 programs.
For effective hard-copy documentation and for checking purposes a MELAprint
®
42 printer can be connected
to print out a record immediately after completion of a program or to print out records from the memory.
2 Installation
When setting-up and installing the autoclave, please consult the separate instruction leaflet "Installing the
Euroklav
2.1 Setting up the autoclave ______________________________________________
The autoclave should be set up in a dry place which is protected against dust. The base should be stable,
and able to support the weight of the appliance (unloaded weight 43 kg).The space required by the autoclave
can be seen from the external dimensions (as in Section 1.1). A minimum additional space of 10 cm should
be allowed on either side and above the autoclave in order to ensure that heat can escape.
The electrical power supply should be a separate 230V AC circuit with a 16 A fuse.
Should it be decided at some stage to install an automatic one-way water system, it is necessary to have a
connection to the wastewater plumbing near the autoclave, preferably a wall outlet (NW 40) or a sink-trap
(standard length of outflow pipe 2m, 16 mm width). The work surface on which the autoclave stands must be
higher than the outlet, and the outlet pipe must be without bends and twists which could prevent water
flowing out freely. At the same time, the work surface must provide convenient access to the autoclave, and
the display must be clearly visible.
The autoclave can be supplied with demineralized/distilled water from the integrated two chamber storage
tank, with freshwater and wastewater chambers. Alternatively, the Euroklav
an external water purifier - MELAdem
However, please note that this will require additional space.
®
29V-S".
®
®
47 or MELAdem®37 (or an equivalent water purification system).
29V-S can also be connected to
2.2 Transport ribbons ___________________________________________________
Take the autoclave out of the packaging by means of the transport ribbons. The ribbons themselves are
each removed by undoing two retaining screws, which must then be screwed firmly back in place without
washers.
In order to ensure that condensate can drain out of the autoclave (which is important if it is to operate
properly) the appliance must have be higher at the front than at the back. The autoclave should first be
installed in a horizontal position (this should be checked with a spirit level at the chamber flange) and then
the front feet should be extended by giving them three (3) turns.
1
®
29V-S
Rear view
9
8
2
3
4
5
6
7
Fig. 3 Installation of the Euroklav
29V-S with wall-mounted trap
1 Tank lid
2 One-way outlet
3 Connection for internal supply of demineralized/distilled water
4 Pipe link for internal supply of demineralized/distilled water
5 Y-connection with non-return valve (included in Item 6 )
6 Wall-mounted trap (MELAG- Art.- No.: 37410)
7 Wall outlet (NW 40)
8 Supply line for external supply of demineralized/distilled water
9 Mains power supply
Page 7
2.4 Mains power supply _________________________________________________
The electric cable of the appliance is plugged into a mains socket rated at 230 V, 50 Hz. The power rating of
the autoclave is 2400 W. In order to avoid overloading the electricity supply, we recommend using a
separate electrical circuit fitted with a 16 A fuse and optionally protected with a 30mA circuit breaker.
2.5 Outlet connection for one-way water____________________________________
The connection for the one-way water outlet at the back of the appliance is connected to the drainage
system of the building by means of the outlet pipe supplied (textile-reinforced transparent pipe, DN16). It is
important that the pipe should have a steady downward gradient, without twists and kinks.
When connecting to a separate outflow pipe NW40, a wall-mounted trap should be used (MELAG Art.-No.
37410, see Fig. 3).
2.6 Internal water supply with demineralized/distilled water____________________
The internal supply of demineralized/distilled water is used for the autoclave in closed-loop operation (see
page 44, Fig. 4), the water is extracted from the right chamber of internal water storage tank. The autoclave
is supplied with a pipe link with two swivel connections installed to connect the storage tank outlet with the
inlet for demineralized/distilled water.
To fill the tank the lid must be removed and the demineralized/distilled water of suitable purity filled into the
right-hand chamber until the Max mark.
2.7 Connection of a water purification system_______________________________
The autoclave can be connected directly to a water purification unit which provides the demineralized or
distilled water required (see page 44, Fig. 4). Instead of connecting the feed water inlet to a storage
container, it is simply connected directly to the water purification unit.
The reverse-osmosis systems MELAdem
provide the quantity and quality of water needed for the Euroklav
Detailed instructions on the installation of these water purification units are provided in their operating
manuals.
When connecting water purification systems from other manufacturers it is very important to ensure that they
are able to provide sufficient amounts of water at the required purity. You are generally advised to consult
MELAG first.
®
47 and the ion-exchanger MELAdem®37 are ideally suited to
3.1.1 Connecting the MELAprint®42_______________________________________________________
The autoclave can be connected to an external printer, the MELAprint
with the autoclave.
In order to connect the printer to the autoclave follow the description in Section 6.3.1.1.1.
3.1.2 Initialization of the printer / Setting up immediate print-out ______________________________
Initialize the external printer (registering with the processing unit of the autoclave) as described in Section
6.3.1.1.2 .In order to select the immediate print-out option, which means that a record of each sterilization is
printed out automatically as soon as it has finished, proceed as described in Section 6.3.1.3.
®
42. This is not supplied as standard
Page 8
Operating Manual for Euroklav
3.2 Vacuum test ________________________________________________________
In order to check the operation of the autoclave in the course of the initial start-up, after long periods without
being used, or after moves, as well as periodically during routine use, a vacuum test should be carried out to
check for leaks.
The vacuum test should be carried out as follows, preferably on the cold autoclave:
Operation
1. Switch on power, then
after the message
the display will show
the basic parameters
Display message
Please wait
Door unlocking
14:27:12
0.02bar 25°C
Program
Program
Start-Stop
Start-Stop
®
29V-S
2. Close the door
Press the "Program"
3.
button several times
until the display shows
"Vacuum test"
4. Press "Start - Stop"
button
5. The evacuation
pressure has been
reached.
The equalisation period
begins
6. After waiting for the
equalisation (5 min) the
measuring period starts
(running here e.g. for 7
min 52s)
Program
Start-Stop
Door
closed
Program
Start-Stop
Vacuum test
Program
Start-Stop
Vacu test: Start
Pressure 980mbar
Program
Start-Stop
Vacu test 00:00
Press.: 240mbar
Program
Start-Stop
Vacu test 07:52
Press.: 243 mbar
After the measuring
7.
period (10 min) the
chamber is ventilated
and
then the leakage rate is
displayed
(if the immediate
printout option is
selected for an
attached printer then a
record will also be
printed)
After the message
"Please wait Door
unlocking" this
alternating message
will appear:
Program
Start-Stop
Ventilate
-0.56 bar 25 °C
Program
Start-Stop
Leak. rate 0.3 mbar
Program
Start-Stop
Open door
please
Page 9
Operation
8. Open door
Display message
Program
Start-Stop
Vacuum test
If the leakage rate is above the limit value, then the display (and the print-out) will also show "Test
unsuccessful". In this case, follow the instruction in 7.3.
3.3 Test run____________________________________________________________
In order to check the operation of the autoclave under realistic conditions, a test run should be carried out
with the "Universal Program, 134°C wrapped" and a relevant load. After loading the autoclave and selecting
the program with the "Program" button, sterilization is started by pushing the "Start/Stop“ button. If the
program runs correctly, the following message will appear on the display (see Section 4.7):
with the values for the maximum values for pressure and temperature. If the immediate printout option has
been selected for an external printer a record of the program run will be printed.
Universal-Program
run successfully
Program
Start-Stop
3.4 Installation record ___________________________________________________
As documentation that the autoclave has been set-up properly, an installation record should be produced by
an authorised person and a copy sent to MELAG. This is important in the event that you wish to make claims
under warranty provisions.
When opening the door, particularly after interrupting the drying process, residual steam can escape from
the autoclave chamber.
After opening the door, do not touch any metal surfaces - these will be hot! Danger of burns. Always use
the tray lifter to remove trays, or wear suitable hand protectors when taking out other items
If you install the optional water purifier MELAdem
water leak detector (see installation details Section 2).
If you intend to install a water purification unit from another manufacturer, then consult MELAG before
you do so.
The appliance is not suitable for sterilizing liquids.
Under current VDE-regulations, this appliances is not suited for use in areas where there are risks of
explosion.
The appliance must only be serviced and repaired by MELAG or by its authorised representatives
(specialist dealers or customer services) using only original parts and following service in structions.
Before opening the housing always disconnect from the mains power supply!
In order to ensure effective sterilization with the autoclave observe the instructions in this operating
manual, and in particular ensure that the loading of the autoclave is appropriate for the program selected.
®
47 or MELAdem®37 we recommend the installation of a
Page 10
Operating Manual for Euroklav
®
29V-S
4 Instructions for all sterilization procedures
4.1 Electricity and water supplies ________________________________________
4.1.1 Distilled or demineralized feed water________________________________________________
The autoclave automatically monitors the availability of cooling water and purified water, as well as the
quality of the distilled / demineralized water before starting a program.
In order to allow an immediate program start and to avoid error reports or interruptions of programs (see
Sections: 7.3 and 7.4) :
Before the first sterilization at the start of the working day, check that the internal water supply has
sufficient water in the right chamber of the double internal tank. If necessary, refill with water of
appropriate quality (see Section 8.3.2).
If the feed water is drawn directly from a MELAdem
on in good time (this may be up to an hour before starting a sterilization program), if the water supply has
been turned off over night, for example.
4.1.2 Power supply ___________________________________________________________________
Switch on the power using the switch on the front of the autoclave (bottom right). About 15 seconds after the
message "Please wait Door release“ the appliance is in the start status:
(Steam)-Temperature (°C)
Time (h:min:s)
Pressure (bar)
®
47 water purifier, check that its water supply is turned
Program
Start-Stop
14:27:12
0.02bar 25°C
4.2 Preparing instruments for sterilization __________________________________
MELAG - rust-free materials
All parts of the Euroklav
pressure chamber and the door of stainless steel, steam pipes of Teflon, and screws and magnet-valves of
bronze.
Film rust
The use of these materials means that no parts of the autoclave can initiate rust formation. Where rust does
attack the autoclave or instruments sterilized in it, tests repeatedly show that this has been brought into the
autoclave on instruments (film rust).
Even top-quality stainless steel instruments can form rust if they are not handled properly, e.g. if they are
treated with the wrong chemical cleaning or disinfecting agents.
Preparing items for sterilization
The example of the formation of film rust shows how important it is to prepare items properly before
sterilization.
Handpieces and contra-angles must be cleaned before sterilization and maintained (e.g. by oiling).
Other instruments must be disinfected and cleaned immediately after use in accordance with UVV/VBG 103,
or similarly strict national codes of practice in a disinfectant and/or cleaning solution at the correct
concentration for the correct length of time
MELAG recommends the use of cleaning aids such as ultrasonic baths, cleaning and maintenance
equipment for handpieces for contra-angles, as well as thermo-disinfecting devices.
It is essential that the instruments are well cleaned in order to avoid dirt and contamination being separated
from the load in the autoclave and clogging filters, valves, and nozzles. In particular locks, joints, and hinges
must be cleaned thoroughly with a brush before sterilization. No traces of cleaning and disinfecting agents
should be allowed to enter into the sterilization chamber of the autoclave, since this can give rise to
corrosion! The instruments should be swilled off with demineralized water and then dried off before being
loaded in the autoclave. Turbines and handpieces must be oiled in accordance with the manufacturer's
instructions in order to ensure their long working life.
Brand-new instruments
The cleaning procedures described above must also be followed before sterilizing brand-new instruments.
These often carry small amounts of grease, oil and soiling from the manufacturing process.
Important: Carefully follow all instructions provided by manufacturers of instruments for the preparation of
their products for first-time sterilization and for subsequent sterilizations.
®
29V-S which come into contact with steam are made on non-rusting materials: the
Page 11
4.3 Loading the autoclave_______________________________________________
It is of crucial importance for effective sterilization and good drying that the autoclave is loaded properly:
When loading the autoclave, take account of the following points:
Tray rack
For the Euroklav
Type "A" (MELAG-Art.-No.: 40231) for up to 6 trays and
Type "B" (MELAG-Art.-No.: 40234) for up to 4 trays or 4 standard tray-cassettes
Both types of tray rack are also suitable for the MELAG-sterilization containers Type 15K,M,G; Type 17 K,
Type 17R; Type 23R, Type28 M, Type 28G).
Normally, the autoclave should be used in conjunction with a tray rack, since this ensures that steam
penetration and drying are as good as possible. In exceptional situations (e.g. when using sterilization
containers from other manufacturers), and after consultation with your specialist dealer or with MELAG, the
tray rack can be removed and the container can be placed directly in the autoclave chamber.
For the sterilization of instruments sealed in transparent sterilization wrapping, it is recommended that you
use the foil stand MELAG-Art.-No.: 22410. This contributes considerably to the drying process for such
wrapped instruments.
Trays
Trays for objects which are to be sterilized must be perforated, in order to allow condensation to run away.
MELAG-trays are recommended. If you use dishes or trays without perforations, then the objects being
sterilized will not dry properly.
Enclosed sterilization containers
Enclosed sterilization containers must be perforated on at least one side (preferably underneath) or must
have valves, in order to ensure that steam can penetrate and condensate can run out. All MELAGsterilization containers meet these requirements with perforations on two sides and filter-cloth- inlays.
Sterilization containers which only have perforations on the top only allow limited drying.
If sterilization containers are stacked in the autoclave, it is important to ensure that the perforations are not
blocked.
Transparent sterilization packaging
If you use transparent sterilization packaging, such as MELAfol
vertically on the tray, or sterilized in foil holders (MELAG-Art.-No.: 22410). They should never be laid flat one
on top of the other.
If seals split open during sterilization it may be necessary to increase the length of the impulse on the sealing
device or to use a double-seal.
Standard tray- cassettes sealed in MELAfol
ensure that the side-seals do not split open.
Multiple wrapping
The pulsed vacuum method means it is possible to use multiple wrapping.
Maximum loads
Loads should not exceed 3 kg of instruments or 0.9 kg of textiles.
Mixed loads
If mixed loads of textiles and instruments are to be sterilized, then as far as possible the textiles should be
above the instruments and direct contact with the instruments should be avoided.
Inclusion of textiles and instruments in the same sterilization container is not desirable.
Textiles should never come into direct contact with the walls of the chamber.
If different types of packaging are included in a load, then:
Instruments and sterilization containers should be at the bottom
Transparent and paper sterilization packaging should be at the top (but lower than textiles)
Liquids
The appliance is not suitable for the sterilization of liquids!
Suitability for sterilization
Relevant information provided by manufacturers of instruments and textiles about sterilization should be
strictly observed.
®
29V-S there are 2 types of tray rack:
®
(250 mm wide) must be taped and clasped additionally to
®
, then the items should if possible be stood
Page 12
Program
Program
Program
Program
®
29V-S
Operating Manual for Euroklav
4.4 Closing the door ____________________________________________________
The door is closed by lightly applying pressure in the direction of the chamber flange and at the same time
pressing down the sliding door catch. The display shows the message:
Door
closed
Program
4.5 Program selection ___________________________________________________
A program should be selected which is appropriate for the physical properties of the items being sterilized
(and in particular their heat resistance) and the type of packaging (if any part of the load is wrapped, then
either the "Universal Program" or the "Gentle Program" must be used).
By pressing the "Program selection" button it is possible to review the display of the following programs for
selection:
Parameter/Application Program name/Display message
Universal program at 134°C, 2 bar, and a
sterilization time of 3.5 min
Sterilization of wrapped items, in particular
instruments (no Hollow A), or mixed loads
(unwrapped/wrapped)
Quick Program at 134°C, 2 bar and a sterilization
time of 3.5 min
Sterilization only of unwrapped instruments (no
Hollow A, no textiles) for rapid re-use (drying can
be interrupted manually)
Gentle Program at 121°C, 1 bar , sterilization time
15 min
Sterilization of all types of wrapped items (except
Hollow A), in particular large amounts of textiles or
thermolabile materials (plastic, rubber), or mixed
loads (wrapped/unwrapped)
Bowie & Dick Test Program at 134°C, 2 bar and
a sterilization time of 3.5 min
Used to check the operation of the autoclave
(Steam penetration of special indicators)
Universal Program
134°C wrapped
Quick Program
134°C unwrapped
Gentle Program
121°C wrapped
Bowie & Dick Test
134°C 2.2bar 3,5‘
Start-Stop
Vacuum test-program
Used to check the autoclave for leaks,
from a cold start
Basic display (no program selected)
Program
Vacuum test
Program
15:31:33
0.02bar 22°C
Page 13
4.6 Program start_______________________________________________________
Press the "Start-Stop" button once the desired program is shown on the display. The availability of cooling
water, and feed water will be checked automatically, with a conductivity measurement.
Start
Program
At the start of the quick program there will be an additional message "Warning: Only unwrapped
instruments". This message must be acknowledged by pressing "Start" again.
4.7 Program progress ___________________________________________________
After starting the program, it will then progress automatically. The display shows the current program status
as follows:
Program status
1. Pre-vacuum consisting of evacuation
(removal of air) until the appropriate
evacuation pressure is reached, followed
by steam introduction up to slightly above
atmospheric pressure.
Display show pressure in chamber and
steam temperature
2. Phase 1 and following
Depending on the program chosen and
the temperature of the chamber at the
start of the program, a number of cycles
of steam inflow and outflow follow to
ensure adequate penetration of super-
heated steam into the items being
sterilized.
3. Heating phase
A heating-up phase follows. The
continuous introduction of steam raises
pressure and temperature in chamber to
the values needed for the program
Display message
Pre-vacuum
-0.72 bar 60°C
1. Fractionation
0.69 bar 115°C
Heat up
1.80 bar 117°C
Start-Stop
Program
Program
Program
Start-Stop
Start-Stop
Start-Stop
4.Sterilization phase
When the required pressure and
temperature have been reached the
sterilization proper then begins.
The display shows alternately the
pressure and temperature and the time
remaining.
5. Pressure release
After completion of the sterilization time,
the pressure is released and the steam
generator emptied.
Pressure and temperature fall.
6.1st drying phase
(Flow drying)
After pressure release the drying phase
begins.
Throughout the drying phases of the
program can be ended without an error
report, since the sterilization itself has
been completed.
However, with the exception of the "Quick
Program" you should normally wait until
the drying is completed..
Program
Start-Stop
Sterilization
2.18 bar 135°C
Program
Start-Stop
Sterilization
still 2 min 12 sec
Program
Start-Stop
Pressure release
0.85 bar 96°C
Program
Start-Stop
Current drying
sin. 1' 0.9bar 85°C
Program
Start-Stop
Immediate removal
Press STOP
Page 14
7.2nd drying phase
(Vacuum drying)
The flow drying is then followed by a final
pulsed vacuum drying phase.
Throughout the drying phases the
program can be ended without an error
report, since the sterilization itself has
been completed.
However, with the exception of the "Quick
Program" you should normally wait until
the drying is completed..
Operating Manual for Euroklav
Vacuum-drying
sin. 2' -0.12 bar 60°C
Immediate removal
Press STOP
Program
Programm
Start-Stop
Start-Stop
®
29V-S
8.Program end
After ventilation of the chamber the
program is completed.
If a printer is attached and print-out "yes"
is selected, the record will be printed-out
immediately.
After the door has automatically unlocked
the door can be opened to remove the
sterilized objects.
Program
Start-Stop
Quick Program
run succesfully
Program
Start-Stop
Please wait
Door unlocking
Program
Start-Stop
Open door
please
Page 15
Loading...
+ 34 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.