Marquette Hellige CardioServ User Manual

CardioServ
Operator’s Manual
Version 4.1
227 446 32 GA(e) Revision D
The product CardioServ bears the marking
-0366
indicating its conformity with the provisions of the Council Directive 93/42/EEC concerning medical devices and fulfills the essential requirements of Annex I of this directive.
The product complies with the electromagnetic immunity requirements of standard IEC 60601-1-2/EN 60601-1-2 Electromagnetic Compatibility – Medical Electrical Equip­ment”.
The device is in radio-interference protection class B in accordance with CISPR11/EN 55011.
The CE marking covers only the accessories listed in section Order Information and Accessories.
This manual describes CardioServ with all options included and reflects software version 4.1.
© Marquette Hellige GmbH 1999
Postfach 600265 D-79032 Freiburg Germany
Telephone +49 7 61 45 43-0
Revision History
This document is subject to the Marquette Hellige change order system. The revision code, a letter that follows the document part number, changes with every update of the manual. The initial version of the manual is without revision code.
Part No./Revision Code Date Comment
227 446 32 29 April 1997 Initial Release 227 446 32-A 14 May 1997 SW version 4.1 227 446 32-B 1 July 1997 ECO 202 998 227 446 32-C 9 July 1998 ECO 060 477 227 446 32-D 6 August 1999 ECO 062 982
2 CardioServ V4.1 227 446 32-D
Contents
1. Introduction to CardioServ
2. Controls and Indicators
3. Setup and Performance Test
4. Non-Synchronized Defibrillation
5 Cardioversion (Synchronized Defibrillation)
6. Displaying and Monitoring the ECG
7. The Memories of CardioServ
8. Recording
9. Oxygen Saturation SpO
2
10. Pacing
11. Configuring the Defibrillator Settings
12. Error Indications and Messages
13. Cleaning and Maintenance
14. Technical Specifications
15. Order Information and Accessories
Index
227 446 32-D CardioServ V4.1 3
General Information
General Information
* This manual is an integral part of the instrument and
describes its normal use. It should always be kept close to the equipment. Observance of the manual is a prerequi­site for proper instrument performance and correct operation and ensures patient and operator safety.
* The symbol
It serves as an indicator for important facts to be noted when operating the instrument.
* Information which refers only to certain versions of the
instrument is accompanied by the part number(s) of the instrument(s) concerned. The part number is given on the instrument nameplate.
* Patient safety, specified measuring accuracy, and interfer-
ence-free operation can be guaranteed only if original Marquette Hellige devices are interconnected (e.g. basic units and plug-in modules).
* Only use accessories which are listed in this manual or
which have been tested in combination with the device (e.g. patient cables, electrodes, transducers, sensors, consumables, etc.). If you use accessories or consumables from other manufacturers, Marquette Hellige does not guarantee safe operation or functioning of the device.
denotes: Refer to Operator’s Manual!
* Marquette Hellige considers itself responsible for the
effects on safety, reliability, and performance of the equipment, only if
assembly operations, extensions, readjustments,
modifications, or repairs are carried out by Marquette Hellige or by persons authorized by Marquette Hellige
the instrument is used in accordance with the
instructions for use.
* All publications are in conformity with the instrument
specifications and IEC publications on safety of electro­medical equipment valid at the time of printing. All rights are reserved for instruments, circuits, techniques, software programs and names appearing in the manual.
* On request Marquette Hellige will provide a service
manual.
* The Marquette Hellige quality management system
complies with the standards DIN EN ISO 9001 and EN 46001.
* The warranty does not cover damage resulting from the
use of accessories and consumables from other manufac­turers.
4 CardioServ V4.1 227 446 32-D
Introduction to CardioServ
1. Introduction to CardioServ
This section describes
The capabilities and applications of the CardioServ
defibrillator
points to note when operating CardioServ
general points to note when handling a defibrillator
227 446 32-D CardioServ V4.1 5
Introduction to CardioServ
1.1 General Information
CardioServ is a high-voltage electrotherapy unit which should be handled only by specially trained
Danger
personnel. Even though the defibrillator is equipped with various safety features, such as internal safety discharging, its operation by unqualified staff could be hazardous to the patient, the operator, and any assisting personnel.
The user instructions given in this manual refer to a
CardioServ unit equipped with pacemaker and SpO2 measuring system.
CardioServ is a light-weight, portable defibrillator with ECG monitor and built-in recorder.
The device is designed for external and internal defibrillation. It can be used both for semi-automatic and for manual defi­brillation. Furthermore CardioServ is capable of monitoring the heart rate with adjustable alarm limits.
CardioServ can be expanded with a transcutaneous pacemaker and/or SpO2 measuring system which also monitors the measured SpO2 values.
The device is easy and convenient to operate. Three operating steps are sufficient to deliver the defibrillation shock. The display can be turned 180° to allow the user to view the information when the device is standing upright.
Defibrillation can be performed with standard defibrillation paddles, defibrillation pads and internal electrodes.
In addition to line power operation, the defibrillator operates on battery power from a slot-in rechargeable battery, and on 12-Volt power supplied from an emergency vehicle.
6 CardioServ V4.1 227 446 32-D
Introduction to CardioServ
I/A
120
100
80
U
60
40
20
0
-20 0 2 4 6 8 10 12
Ra = 50 Ohm Ra = 100 Ohm Ra = 25 Ohm
c
L = 26 mH
C = 32 µF
R = 6,5
t
I
R
a
Figure 1-1. CardioServ current-discharge curve (360 J)
14
t/ms
CardioServ comes with three memories whose contents can be documented individually:
text memory (code summary documentation) event memory (16-s ECG) trend memory (trend plots of HR, SpO2, 45 min,
9 hours)
The built-in recorder is initiated automatically or manually.
Extensive safety precautions have been taken to protect the patient and the user from inadvertent delivery of defibrillation shocks.
The current-discharge curve corresponds to an approximate sinusoidal halfwave with aperiodic decay. Figure 1-1 shows the discharge curve for various external resistances.
227 446 32-D CardioServ V4.1 7
1.2 For your Safety
g
Warning
Warning
Warning
Warning
Caution
Introduction to CardioServ
The safety information given below is divided into the catego­ries Danger, Warning and Caution.
indicates an imminently hazardous situation which, if not avoided, will result in death or
Danger
serious injury.
indicates a potentially hazardous situation which,
Warnin
if not avoided, could result in death or serious injury.
indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate
Caution
injury and/or damage to the equipment.
provides application tips or other useful information
to assure that you get the most from your equipment.
CardioServ is designed to comply with IEC 60601/
EN60601 requirements. It is Class I equipment and has an internal power source.
CardioServ operates on line voltages between 95 and 240 Volts, 49 to 65 Hz. The mains plug must be connected to an appropriate power supply with a non-fused earthed wire. The use of exten­sion cords is not permitted.
Before putting the device into operation, visually check all connecting cables and electrodes for signs of damage. Damaged cables and electrodes must be replaced immediately, before use.
When disconnecting the device from the power line, first remove the plug from the wall outlet.
Disconnect the device from the power line and operate it on battery power, if the integrity of the protective earth conductor is in doubt.
Set up the device in a location which affords sufficient ventilation. The ventilation openings of the device must not be obstructed. The ambient conditions specified in the Technical Specifica­tions section must be ensured at all times.
8 CardioServ V4.1 227 446 32-D
Introduction to CardioServ
g
g
g
Warning
Warning
Warning
Warnin
Danger
Warnin
Danger
Warnin
Caution
Devices intended for emergency application must not be exposed to low temperatures during storage and transport to avoid moisture condensation at the application site. Wait until all moisture has vaporized before using the device. Avoid using the defibrillator under conditions where prolonged exposure to or excessive contact with moisture can occur.
Before putting the defibrillator into operation, make sure that the paddles and all connection cables are dry.
CardioServ is an emergency device and must be ready for operation at any time. For this reason, the defibrillator battery must always be charged. This can be achieved by leaving the defibrillator connected to the power line when it is not needed in an environment where only battery operation is possible.
Possible explosion hazard if used in the presence of concentrated oxygen.
CardioServ is suitable for application in a humid environment provided the regulations concerning drip-proof equipment of IEC 60601 are strictly observed. However, avoid defibrillation in a very moist or wet environment, unless absolutely necessary.
Use only the original Marquette Hellige batteries, as these are designed for an extended temperature range.
Magnetic and electrical fields are capable of interfering with the proper performance of the device. For this reason make sure that all external devices operated in the vicinity of the defibrillator comply with the relevant EMC requirements. X­ray equipment, MRI devices and radio systems are a possible source of interference as they may emit higher levels of electromagnetic radiation.
The defibrillator is designed for intracardiac appli-
cation.
Devices may be connected to other devices or to parts of systems only when it has been made certain that there is no danger to the patient, the operators, or the environment as a result. In those instances where there is any element of doubt concerning the safety of connected devices, the user must contact the manufacturers concerned or other informed experts as to whether there is any possible danger to the patient, the operator, or the environment as a result of the proposed combina­tion of devices. Standards IEC 60601-1-1/ EN 60601-1-1 must be complied with in all cases.
Set up the device so that the operator has a clear,
unobstructed view of the front panel.
Liquids must not be allowed to enter the device. Devices into which liquids have penetrated must be immediately cleaned and checked by a service technician.
227 446 32-D CardioServ V4.1 9
Introduction to CardioServ
Caution
Warning
g
g
Danger
Caution
Warnin
Use only the original Marquette Hellige patient cables. Do not connect other signal sources to the cables.
Patient signal inputs labelled with the symbol are protected against damage resulting from defibrillation and electrocautery voltages. Nevertheless extreme care must be taken when devices which are directly connected to the patient remain applied during electrocautery or defibril­lation. The distance between ECG electrodes should be at least 15 cm. If in doubt, disconnect the patient cable from the device while applying the defibrillation pulse or performing electrocautery.
For defibrillation of children use only the special clip-on electrodes for children listed in section 15 Order Information and Accessories.
CardioServ also operates on line power when the
battery is depleted or missing.
At the end of its service life, CardioServ and its accessories must be disposed of in compliance with the special waste control regulations for electronic parts. If you have any questions in this matter, please contact Marquette Hellige GmbH.
Dispose of the packaging material, observing the applicable waste-control regulations and keeping it out of childrens reach.
Literature
Medical Device Directive
EN 60601-1/1990 + A 1: 1993 + A2: 1995: Medical electrical equipment. General requirements for safety
EN 60601-1-1/9.1994 + A1 12/1995: General requirements for safety. Requirements for the safety of medical electrical
Warnin
The full responsibility for the use of accessories from other manufacturers lies with the user.
systems.
Check the device performance at regular intervals (once a month), strictly following the instructions
Caution
given in section 3.2. Do not select high energy
IEC Publication 513/1994: Fundamental aspects of safety standards for medical equipment.
levels for test discharges with defibrillation elec­trodes shorted together.
ROY, O.Z.: Summary of cardiac fibrillation thresholds for 60-Hz currents and voltages applied directly to the heart.
For each defibrillation, verify that the selected and the displayed/charged energy are identical.
Caution
10 CardioServ V4.1 227 446 32-D
Med. & Biol. Engn. & Computing 18: 657...659 (1980).
Introduction to CardioServ
g
g
g
g
g
Caution
Danger
Caution
Danger
Caution
General points to note when handling a defibrillator
Electromedical equipment such as the CardioServ
Warnin
defibrillator must only be handled by persons who are trained in the use of such equipment and are capable of applying it properly.
Before using the equipment, the operator must
Warnin
ascertain that it is in correct working order and operating condition.
The defibrillator paddles must be clean and dry.
Danger
The person carrying out the defibrillation should
have at least one assistant.
The operator must be trained in the use of the
Warnin
defibrillator.
All assistants must be briefed regarding the
Warnin
preparations for and execution of defibrillation.
All tasks must be assigned clearly.
Warnin
Defibrillating a patient with normal heart rhythm may induce ventricular fibrillation.
Danger
Position the patient flat on a hard, dry surface where the patient is electrically insulated. The patient must not be allowed to come into contact with metal parts, e.g., bed or litter, in order to prevent unwanted pathways for the defibrillation current which endanger the assistants. For the same reason, do not place the patient on wet ground (rain, accident in swimming pool).
Have a pacemaker at hand, if possible.
Should cardiac arrest occur or be imminent during
preparations for defibrillation, administer heart massage and artificial respiration (CPR).
Do not allow the defibrillator paddles to come into contact with other electrodes or metal parts which are in contact with the patient.
Transducers and instruments that are not defi­brillation-proof must be disconnected from the patient.
Interrupt heart massage and artificial respiration immediately before triggering the shock.
Immediately before triggering the shock discon­nect tubes and have assistants step back.
227 446 32-D CardioServ V4.1 11
Introduction to CardioServ
g
Warning
Warning
Caution
Danger
Warnin
Caution
The patients chest must be dry, as moisture causes unwanted pathways for the defibrillation current. Therefore, when using flammable skin cleansing agents, wait until they have completely dried.
Possible explosion hazard if used in the presence of concentrated oxygen, flammable substances (gasoline) or anesthetic agents. Oxygenation in the vicinity of the defibrillation paddles must be strictly avoided; if necessary, interrupt oxygen supply while defibrillating the patient.
To prevent sparking
– the electrodes should make full contact with
the body
– the electrodes should be pressed firmly onto
the thorax.
When defibrillating children it is especially important to verify that the paddles make full contact with the body surface. This is to be ob­served also when using the clip-on electrodes for children (Part No. 303 439 95).
Defibrillating a patient who has an implanted pacemaker is likely to impair the pacemaker function or cause damage to the pacemaker. Therefore, the following should be observed:
– Select the smallest energy level possible for the
application.
– Do not apply the defibrillation paddles in the
vicinity of the pacemaker electrodes.
– The availability of an external pacemaker is
of utmost importance in this case.
– After the defibrillation the working order of
the implanted pacemaker should be checked immediately.
Also be aware that children require less energy for a successful defibrillation than adults. For the first defibrillation pulse delivered to babies and toddlers, select an energy level of 2 joules/kg body weight. For subsequent shocks, increase the energy up to 4 joules/kg.
12 CardioServ V4.1 227 446 32-D
Controls and Indicators
2. Controls and Indicators
This section describes the CardioServ operating controls and indicators and explains their function.
When operating elements in this manual are identified with a reference number in parentheses, this number refers to Figure 2-1 in this section.
You will also find an explanation of all signs and symbols used on the CardioServ defibrillator.
227 446 32-D CardioServ V4.1 13
Controls and Indicators
3
2
3
Charge
-
+
2
2
5
7
102030
50
100
150
300
J (50 )
200
360
Charge
Shock
Shock
Sync
APEX
Apply electrode cream to both paddles
Press charge button on apex paddle (on the device, when using adhesive or internal electrodes)
Press buttons on both paddles simul­taneously (on the device, when using adhesive or internal electrodes)
Charge defibrillator
Deliver shock3
2
1 Select energy
P
a
c
e
m
a
k
e
r
S
t
i
m
u
l
a
t
e
u
r
Dem
Fix
+
P/min
P/min
+
mA
mA
Shock
Charge Shock
Autosequence
STERNUM
APEX
2
1418 17 16 15
13
11
10
12
1
13 8 975
4
6
19
20
21
22
23
Option
SpO
2
24
EKG
Figure 2-1. Controls and indicators of CardioServ VF
14 CardioServ V4.1 227 446 32-D
Controls and Indicators
1 Defibrillator paddles
Shock
2
button to trigger the defibrillation shock – together
3
with button (8)
3 keys to adjust the LCD contrast. Press both
keys simultaneously to obtain a screen copy.
4 Energy selector, on/off switch
Charge
5
key to charge the unit (manual mode) and to
Shock
trigger the defibrillation shock – together with key (6). This key assumes the function of button (8) when internal paddles or adhesive electrodes are used.
Shock
6
key to trigger the defibrillation shock – together
with key (5). This key assumes the function of button (2) when internal paddles or adhesive electrodes are used.
7 Five selection keys F1 to F5 whose functions change
with the menu displayed. The respective key functions are indicated by symbols or labels in the bottom line on the LCD. From the main menu, that appears on power up, you can access submenus which, in turn, allow the selection of further options. The back function returns you to the next higher menu level. The main menu reappears automatically if you do not depress any of the keys for about 30 seconds.
8
button to charge the unit (manual mode) and
3
2
Charge
Shock
trigger the defibrillation shock together with button (2)
11 key to start and stop the recorder. This key assumes
the function of button (9) when internal paddles or adhesive electrodes are used.
12 Green indicator is lit when the defibrillator operates
on line power.
+
13 Yellow indicator
-
is lit when the defibrillator battery
is being charged.
14 Connection for defibrillator paddles
(controls 15 to 18 only on models with pacemaker)
+
mA
15 Keys
16 Keys
17 Pacing mode selection key
mA
P/min
to adjust the pacer output
-
+
P/min
to adjust pacer rate
-
Dem
(demand/fixed rate)
Fix
18 Key to enable and disable the pacemaker
19 Aperture to open the paper compartment
20 Connector for power cord
21 Unassigned
22 1-Volt ECG output
23 Connector for SpO2 sensor
9 button to start and stop the recorder.
Sync
10
key to switch to the synchronized operating mode
24 Connector for patient cable (ECG signal input)
(section 5 “Cardioversion”)
227 446 32-D CardioServ V4.1 15
Controls and Indicators
Explanation of the signs and symbols used on the defibrillator
Type CF equipment with highly insulated patient connections, suitable for intracardiac application, connections defibrillation-proof.
Type CF equipment with highly insulated patient connections, suitable for intracardiac application, connections not defibrillation-proof.
Standby mode (for line-power operation)
Power off
+
-
Battery charging
Recorder start
ECG signal
Contrast
Standby or preparatory state only for a part of the
equipment
On, only for a part of the equipment
Hardcopy of screen image
Refer to Operators Manual
Direction indicator
Signal output
Audible alarm on/off
High voltage
16 CardioServ V4.1 227 446 32-D
Setting Up CardioServ and Testing Its Performance
3. Setting Up CardioServ and Testing Its Performance
In this section you will find information about
putting CardioServ into operation
connecting CardioServ to the 12-Volt power supply of
an ambulance vehicle
customizing the CardioServ settings to suit your per-
sonal requirements
testing the CardioServ performance before using it on a
patient
227 446 32-D CardioServ V4.1 17
Setting Up CardioServ and Testing Its Performance
3.1 Setting Up CardioServ
The defibrillator is a high-voltage electrotherapy device and must be handled by qualified person-
Danger
nel only. Improper use of this device can endanger life. Do not fail to observe the information given in this manual and only entrust the device to the hands of trained persons.
Check the electrodes and their leads for signs of damage every time before you use the defibrilla­tor. In particular, make a close visual inspection of the insulation. Replace internal electrodes or the contact inserts when you detect signs of me­chanical damage.
CardioServ operates on:
line power (95 to 240 V, 49 to 65 Hz)
battery power (rechargeable batteries), i.e., independent
of the power line
12-Volt power supplied from the emergency vehicle
(with optional defibrillator mounting system)
* Use the power cord to connect the defibrillator to the
power line (Figure 3-1).
Figure 3-1. Power input
18 CardioServ V4.1 227 446 32-D
Setting Up CardioServ and Testing Its Performance
360
x
Shock
Charge
m
a
k
e
r
S
t
i
m
u
l
a
t
e
u
mA
mA
r
+
+
P/min
m
P/min
Shock
Sync
+
-
Figure 3-2. Green indicator (indicating that defibrillator is
supplied from the power line)
Figure 3-3. Inserting the battery
* Check that the green indicator is lit (Figure 3-2).
* Check that the battery is in place (Figure 3-3).
Pull back the catch on the underside of the device to remove the battery. When inserting it, make sure that it clicks prop­erly into place.
+
When a battery is inserted, the yellow indicator
-
starts flashing as soon as the defibrillator is connected to the power line (to indicate that the battery is charging). The battery is fully charged after 16 hours and the indicator is continuously lit.
* Check that the supply of chart paper is sufficient (Figure
3-4). A stripe marks the last 3 meters of the roll.
Figure 3-4. Recording strip
227 446 32-D CardioServ V4.1 19
Setting Up CardioServ and Testing Its Performance
g
Power Supply From Emergency Vehicles
A qualified technician can be called in to connect the CardioServ to the 12-Volt supply of an emergency vehicle. The following points must be noted:
The negative terminal of the ambulance power supply
system must be connected to the ambulance chassis for grounding.
The positive lead of the ambulance power supply system
intended to supply the current must be protected with a
-
+
-
+
-
+
T
e
>
s
9
t
5
>
%
8
0
%
<
8
0
%
T
e
S
s
ta
t
rt
10-A fuse.
Use only the defibrillator mounting system listed in
section 15 Order Information and Accessories or the external charging unit, if your CardioServ has been modified accordingly.
Check that the contacts for power supply from the
Figure 3-5. Accu Service Unit
defibrillator mounting system on the underside of CardioServ (next to battery) are clean. Do not damage
We recommend our Accu Service Unit for optimal
them in any way.
care of the batteries. It prolongs the batteries service life and guarantees their operational readiness at all
Important Information on Battery-Power Operation
times.
Rechargeable batteries require special maintenance and con­tinued checks to assure they function in emergency situations. It is normal for batteries of this type to selfdischarge when not in use.
Warnin
A NiCd battery should not be charged while located in direct sunlight, over a radiator, in cold storage, or in other temperature extremes (not below 5 °C). When the instrument is charging, ambient temperatures exceeding 40 °C may adversely affect battery capacity and life.
20 CardioServ V4.1 227 446 32-D
The battery charges automatically when CardioServ is con­nected to the power line (yellow indicator (13) flashes).
Setting Up CardioServ and Testing Its Performance
In order to ensure its functioning as an emergency device, the defibrillator should not be disconnected from the power line for more than 48 hours.
A fully charged battery supplies power for 35 defibrillation shocks of 360 joules or 2 hours of monitoring (or 1.2 hours of monitoring if the CardioServ unit has pacemaker and SpO option). It takes 16 hours to charge a depleted battery.
Proceed as follows to test the battery charge level:
Disconnect CardioServ from the power line. Trigger a test discharge of 360 joules (see
section 3.2 Testing the Defibrillator Perform­ance).
If you are not prompted to charge the battery,
the charge level should be sufficient for at least 5 more 360-joule shocks.
Monthly battery maintenance and checks:
1. Disconnect CardioServ from the power line and discharge fully charged battery in the monitoring mode. To do so, set energy selector switch to , (SpO2 sensor not connected) and wait until device switches off.
2
Inserting CardioServ In Its Softcase
* Open both zips on the CardioServ softcase.
* Undo the two Velcro strips on the front of the softcase.
* Open the Velcro flap located in front of the CardioServ
printer.
* Place the softcase on the small base, so that the softcase is
positioned as though you were carrying it.
* Disconnect the paddles on the CardioServ and remove
them.
* Hold all the opened up parts of the softcase to the side
and put the CardioServ into the softcase.
* Pull the softcase by the two Velcro strips on the black
base into position, ensuring that the feet of the CardioServ are in the cut-out openings provided.
* Close the two front Velcro strips as well as the flap in
front of the printer at the side.
2. Check how long it takes before battery is depleted. If the time is less than 1.8 hours, the battery is too old or improperly maintained and must be replaced.
3. Recharge the battery. This will take 16 hours.
For easy, convenient care and maintenance of the batteries, use our Accu Service Unit”.
227 446 32-D CardioServ V4.1 21
* Close the zips.
* If the individual cut-out openings for the external con-
nections, paddles or feet are not correctly positioned, pull the material at this point into the right shape.
* Reconnect the paddles and put them back into the
CardioServ recesses.
Customizing the defibrillator settings
Setting Up CardioServ and Testing Its Performance
Further steps that can be taken while setting up the defibrilla­tor include customizing the device functions. This allows you to select defibrillator default settings which suit your personal preferences. CardioServ saves these configured settings and reactivates them automatically every time you switch the
Refer to section 11 The Defaults Menu for a detailed explanation of how to customize the CardioServ, including the language selection (available languages are English, French, German, Italian, Portuguese, Russian, Spanish and
Swedish). defibrillator on. The following chart shows the factory settings of all parameters and the optional adjustments.
retemaraP noitpircseD gnitteSyrotcaF egnartnemtsujdA sgnittesdezimotsuC
stimiLRH stimilmralaRH 061/04 ffo003ot51ffo
ytivitisneS ezislangisGCE 1 Vm/mc2,1,5.
daeL 1 selddap+sdaeldradnatslla
.qesotuA slevelygrenegnicneuqesotua J063,J002,J002 ,J063,J003,J002,J051
OpS
stimiL stimilmrala ffo/09 ffo001ot51ffo
2
kcoL-C OpSdezinorhcnys-GCE
OpS
2
resU )sretcarahc04(emanrotxeteerf
emiT.tnI OpS
mralAliaFdaeL ffo s03/ffo
peeBSRQ ffo ffo/no
enoTmralA ffo ffo/no
tuotnirPmralA tratsredrocer.motua ffo ffo/no tuotnirPkcohS tratsredrocer.motua no ffo/no
tuotnirP.tnoC tratslaunamnopupotsredrocer ffo ffo/no
rekamecaP etareslup MPB06 nim/P051ot51 yalpsiD lamron desrever/lamron yalpsiD seerged0 seerged081/seerged0
emuloV slangisoiduallafoemulov duol wol/duol
retliFCA zH05 ffo/zH06/zH05
retliFelcsuM tcafitranoitomfonoitanimile no ffo/no
emiT/etaD
tamroFetaD YYYY:MM:DD YYYY:DD:MM/YYYY:MM:DD
egaugnaL namreG ,namreG,hcnerF,hsilgnE
tluafedyrotcaF sgnittesyrotcafhsilbatseer ffo ffo/no
2
2
emitnoitargetni s8 )dednemmocerton(s21,s8,s4
tnemerusaem ffo ffo/no
,naissuR,eseugutroP,nailatI
hsidewS,hsinapS
Table 1
22 CardioServ V4.1 227 446 32-D
Setting Up CardioServ and Testing Its Performance
m
J (50 )
20
10
7
5
2
Autosequence
P
a
c
e
m
a
k
e
r
S
t
i
Figure 3-6. Switching on CardioServ
a b
Paddle
c
0
160 BPM
140 ECG
HR Alarm
d
3.2 Testing the Defibrillator
30
50
100
150
200
300
360
Charge
Shock
Shock
Sync
e
f
Performance
On power up CardioServ runs an automatic selftest. Any malfunctions identified during this test result in an error message displayed on the LCD (refer to section 12 “Error Indications and Messages). As a further performance test a trial defibrillation can be triggered.
The energy selector is used to switch CardioServ on and off. Once you have become familiar with CardioServ you can thus switch on the defibrillator and select the required energy in one single operation.
* Set the energy selector to the position. No energy
will be stored in this position of the switch.
The defibrillator beeps and displays a checkered pattern (LCD performance test). Next the main display appears (Figure 3-7).
g
The Paddle message (b, Figure 3-7) indicates that no pa­tient cable is connected and that the ECG signal is acquired via the defib paddles. Upon connection of the patient cable
FreezeSpO2ECG
Next
Menu
the selected ECG lead is displayed here. CardioServ is now ready for operation.
Figure 3-7. CardioServ main display
a heart/pulse-rate reading with alarm limits b ECG signal source c 1-mV calibration pulse d alarm message, QRS blip e signal trace, channel 1 f signal trace, channel 2 g menu
227 446 32-D CardioServ V4.1 23
Setting Up CardioServ and Testing Its Performance
Do not trigger more than 5 consecutive test discharges (or internal safety discharges) within
Caution
15 minutes at max. energy setting.
Paddle
0
160 BPM
140 ECG
HR Alarm
Figure 3-8. Display of selected energy
360 J
Next
MenuFreezeSpO2ECG
Test Discharge
A test discharge can be triggered to check the defibrillator discharge circuit. For this test the stored energy is discharged into the device via two contacts in the paddle compartments.
* Set the energy selector to 360 joules (50 joules, if inter-
nal electrodes are connected). The display first shows the selected energy (Figure 3-8).
2
Charge
* Press the
(Press the
Shock
button on the paddle to charge the unit.
3
Charge
key on the defibrillator, when using
Shock
internal electrodes.)
You may now watch the defibrillator charging.
* When the selected energy level has been reached,
CardioServ beeps and the stored energy is displayed (Figure 3-9).
* Trigger the defibrillation pulse within the next 30 sec-
onds. To do so, simultaneously press the buttons on
Charge
both paddles (or press the two
Shock
Shock
keys on the
defibrillator, when using internal electrodes).
If more than 5% of the available energy are lost before the defibrillation pulse is triggered, CardioServ recharges until the required energy level is reached.
24 CardioServ V4.1 227 446 32-D
Setting Up CardioServ and Testing Its Performance
Caution
Paddle
0
160 BPM
140 ECG
HR Alarm
Energy available
Figure 3-9. Display of available energy
If you do not trigger the defibrillation pulse within
30 seconds, an internal safety discharge is initiated automatically.
360 J
Next
MenuFreezeSpO2ECG
After defibrillation, the beeping sound stops, and the energy actually delivered into a 50-ohm resistance is displayed for 10 seconds in place of the stored energy (Figure 3-9). The delivered energy must not deviate more than ±15% or ±4 joules (whichever is greater) from the selected value. A recording is initiated at the same time (16-second strip).
Should the discharge circuit be interrupted (paddles not properly placed on contacts in the compartments, defective lead), an internal safety discharge is initiated 200 ms after the defibrillation shock has been triggered. In this case the deliv­ered energy is “0”.
If the defibrillator cannot store the selected energy so that selected and stored energy values differ, the LCD shows the message Energy high or Energy low. The defibrillation pulse can be triggeed all the same.
* Switch off CardioServ (set energy selector to ).
The message Energy high or Energy low indicates that CardioServ needs to be repaired.
Caution
If, in spite of this energy storage problem, the device has to be employed, it will display the message Self-test failed. Charge Energy Error
Testing the Pacemaker Performance
The performance of the pacemaker can be tested with a commercially available pacemaker tester (e.g. CS300 Simula­tor from Marquette Hellige, part no. 417 983-001).
upon power up. In this situation press one of the function keys and proceed as usual.
Performance Test
Test the defibrillator performance once a week: test 1 – defibrillator connected to mains, battery
removed,
test 2 – defibrillator disconnected from mains,
battery inserted
227 446 32-D CardioServ V4.1 25
For your notes
Setting Up CardioServ and Testing Its Performance
26 CardioServ V4.1 227 446 32-D
Non-Synchronized Defibrillation
4. Non-Synchronized Defibrillation
This section describes first how to perform a non-synchro­nized defibrillation, using the standard defibrillation paddles. The subsequent explanations refer to the use of internal and disposable adhesive electrodes.
At the end of this section you will find a summary of all necessary operating steps (brief operating instructions).
227 446 32-D CardioServ V4.1 27
Non-Synchronized Defibrillation
g
4.1 Defibrillation with Standard Electrodes
The information given in section 1.2 must be
Warnin
observed without fail to ensure safe and reliable application of the device.
Always switch off CardioServ before exchanging the defibrillation paddles.
Danger
Paddle
0
160 BPM
140 ECG
HR Alarm
200 J
SpO2ECG
Figure 4-1. Main display, indication of ECG lead, manual
operation and selected energy
Next
MenuFreeze
The energy selector can be set to the autosequence position, where the defibrillator automatically sequences the preset energy levels. The preset factory default setting is the sequence recommended by AHA/ERC (200 J, 200 J, 360 J). The factory default values can be changed from the defaults menu. You can choose among 150 J, 200 J, 300 J and 360 J.
* Set the energy selector (4) to “Autosequence or to the
required energy value (this turns on CardioServ).
The defibrillator beeps and displays a checkered pattern (LCD performance test). Next the main display appears (Figure 4-1). When a patient cable is connected the selected ECG lead is displayed (selectable).
* Check that the energy selector locks in on the correct
position and that the display shows the selected energy (Figure 4-1).
The energy depends on the defibrillation mode, on the patients age and constitution. In external application the thickness of the tissue is also a factor which influences the amount of energy required.
The energy necessary for successful ventricular defibrillation without damaging the myocardium has for many years been a matter of scientific controversy. The manufacturer is therefore not able to give any recommendations.
28 CardioServ V4.1 227 446 32-D
Figure 4-2. Removing the paddles
g
Do not apply the paddles over
Warnin
sternum or claviclenipplesimplanted pacemaker or defibrillator.
Non-Synchronized Defibrillation
In emergency situations AHA recommends for adult patients
1. defibrillation with 200 joules; if unsuccessful, repeat
2. defibrillation with 200 joules; if unsuccessful, repeat
3. defibrillation with max. energy setting (360 joules).
Please note that children require less energy for successful ventricular defibrillation than adults. For the first defibrilla­tion pulse delivered to babies and small children, select an energy level of 2 joules/kg body weight. For subsequent shocks, the energy may be increased to 4 joules/kg.
In compliance with IEC requirements the energy adjusted on this defibrillator is not the stored energy, but the energy released into an external resistance of 50 ohms (patient resist­ance + electrode-to-skin contact resistance). The energy selector is labelled accordingly.
* Remove the paddles from their compartments (as shown
in Figure 4-2). Carefully dry the electrodes, if they are wet. The handles, in particular, must be completely dry. Apply an ample amount of electrode cream to each paddle.
STERNUM
3
2
C h
a r
g e
3
APEX
* Apply the electrodes to the patients thorax so that the
greatest possible amount of energy flows through the myocardium (the imaginary connecting line between the two electrode centers should be identical with the cardiac median line; Figure 4-3).
* Press paddles firmly down onto the patients thorax..
Figure 4-3. Paddle application points
227 446 32-D CardioServ V4.1 29
Non-Synchronized Defibrillation
2
Charge
Shock
3
STERNUM
2
3
Charge
Shock
8
APEX
Figure 4-4. Buttons to initiate energy storage and to trigger the
defibrillation shock
Paddle
0
160 BPM
140 ECG
HR Alarm
The ECG signal trace appears on the monitor.
* Do not touch the patient any more and warn all those
present.
2
Charge
* Press the
Shock
button (8) on the apex paddle to initiate
3
charging (Figure 4-4). When using internal electrodes or defibrillation pads, press
Charge
on the device.
Shock
When the selected energy level has been reached (message Energy available), CardioServ beeps and the stored energy is displayed (Figure 4-5).
* Trigger the defibrillation shock within the next 30 sec-
onds by simultaneously pressing the buttons on both paddles (Figure 4-4). When using internal electrodes or defibrillation pads, press
Charge
Shock
and
Shock
on the device.
If more than 5% of the available energy are lost before the defibrillation shock is triggered, CardioServ recharges until the required energy level is reached.
Energy available
FreezeSpO2ECG
200 J
Next
Menu
Figure 4-5. Display of available energy
When the defibrillator is already charged you can increase the energy level simply by turning the energy selector to the new setting. To decrease the energy level, set the selector to the lower value and initiate charging again.
30 CardioServ V4.1 227 446 32-D
Loading...
+ 86 hidden pages