Marquette Hellige CardioServ User Manual

CardioServ
Operator’s Manual
Version 4.1
227 446 32 GA(e) Revision D
The product CardioServ bears the marking
-0366
indicating its conformity with the provisions of the Council Directive 93/42/EEC concerning medical devices and fulfills the essential requirements of Annex I of this directive.
The product complies with the electromagnetic immunity requirements of standard IEC 60601-1-2/EN 60601-1-2 Electromagnetic Compatibility – Medical Electrical Equip­ment”.
The device is in radio-interference protection class B in accordance with CISPR11/EN 55011.
The CE marking covers only the accessories listed in section Order Information and Accessories.
This manual describes CardioServ with all options included and reflects software version 4.1.
© Marquette Hellige GmbH 1999
Postfach 600265 D-79032 Freiburg Germany
Telephone +49 7 61 45 43-0
Revision History
This document is subject to the Marquette Hellige change order system. The revision code, a letter that follows the document part number, changes with every update of the manual. The initial version of the manual is without revision code.
Part No./Revision Code Date Comment
227 446 32 29 April 1997 Initial Release 227 446 32-A 14 May 1997 SW version 4.1 227 446 32-B 1 July 1997 ECO 202 998 227 446 32-C 9 July 1998 ECO 060 477 227 446 32-D 6 August 1999 ECO 062 982
2 CardioServ V4.1 227 446 32-D
Contents
1. Introduction to CardioServ
2. Controls and Indicators
3. Setup and Performance Test
4. Non-Synchronized Defibrillation
5 Cardioversion (Synchronized Defibrillation)
6. Displaying and Monitoring the ECG
7. The Memories of CardioServ
8. Recording
9. Oxygen Saturation SpO
2
10. Pacing
11. Configuring the Defibrillator Settings
12. Error Indications and Messages
13. Cleaning and Maintenance
14. Technical Specifications
15. Order Information and Accessories
Index
227 446 32-D CardioServ V4.1 3
General Information
General Information
* This manual is an integral part of the instrument and
describes its normal use. It should always be kept close to the equipment. Observance of the manual is a prerequi­site for proper instrument performance and correct operation and ensures patient and operator safety.
* The symbol
It serves as an indicator for important facts to be noted when operating the instrument.
* Information which refers only to certain versions of the
instrument is accompanied by the part number(s) of the instrument(s) concerned. The part number is given on the instrument nameplate.
* Patient safety, specified measuring accuracy, and interfer-
ence-free operation can be guaranteed only if original Marquette Hellige devices are interconnected (e.g. basic units and plug-in modules).
* Only use accessories which are listed in this manual or
which have been tested in combination with the device (e.g. patient cables, electrodes, transducers, sensors, consumables, etc.). If you use accessories or consumables from other manufacturers, Marquette Hellige does not guarantee safe operation or functioning of the device.
denotes: Refer to Operator’s Manual!
* Marquette Hellige considers itself responsible for the
effects on safety, reliability, and performance of the equipment, only if
assembly operations, extensions, readjustments,
modifications, or repairs are carried out by Marquette Hellige or by persons authorized by Marquette Hellige
the instrument is used in accordance with the
instructions for use.
* All publications are in conformity with the instrument
specifications and IEC publications on safety of electro­medical equipment valid at the time of printing. All rights are reserved for instruments, circuits, techniques, software programs and names appearing in the manual.
* On request Marquette Hellige will provide a service
manual.
* The Marquette Hellige quality management system
complies with the standards DIN EN ISO 9001 and EN 46001.
* The warranty does not cover damage resulting from the
use of accessories and consumables from other manufac­turers.
4 CardioServ V4.1 227 446 32-D
Introduction to CardioServ
1. Introduction to CardioServ
This section describes
The capabilities and applications of the CardioServ
defibrillator
points to note when operating CardioServ
general points to note when handling a defibrillator
227 446 32-D CardioServ V4.1 5
Introduction to CardioServ
1.1 General Information
CardioServ is a high-voltage electrotherapy unit which should be handled only by specially trained
Danger
personnel. Even though the defibrillator is equipped with various safety features, such as internal safety discharging, its operation by unqualified staff could be hazardous to the patient, the operator, and any assisting personnel.
The user instructions given in this manual refer to a
CardioServ unit equipped with pacemaker and SpO2 measuring system.
CardioServ is a light-weight, portable defibrillator with ECG monitor and built-in recorder.
The device is designed for external and internal defibrillation. It can be used both for semi-automatic and for manual defi­brillation. Furthermore CardioServ is capable of monitoring the heart rate with adjustable alarm limits.
CardioServ can be expanded with a transcutaneous pacemaker and/or SpO2 measuring system which also monitors the measured SpO2 values.
The device is easy and convenient to operate. Three operating steps are sufficient to deliver the defibrillation shock. The display can be turned 180° to allow the user to view the information when the device is standing upright.
Defibrillation can be performed with standard defibrillation paddles, defibrillation pads and internal electrodes.
In addition to line power operation, the defibrillator operates on battery power from a slot-in rechargeable battery, and on 12-Volt power supplied from an emergency vehicle.
6 CardioServ V4.1 227 446 32-D
Introduction to CardioServ
I/A
120
100
80
U
60
40
20
0
-20 0 2 4 6 8 10 12
Ra = 50 Ohm Ra = 100 Ohm Ra = 25 Ohm
c
L = 26 mH
C = 32 µF
R = 6,5
t
I
R
a
Figure 1-1. CardioServ current-discharge curve (360 J)
14
t/ms
CardioServ comes with three memories whose contents can be documented individually:
text memory (code summary documentation) event memory (16-s ECG) trend memory (trend plots of HR, SpO2, 45 min,
9 hours)
The built-in recorder is initiated automatically or manually.
Extensive safety precautions have been taken to protect the patient and the user from inadvertent delivery of defibrillation shocks.
The current-discharge curve corresponds to an approximate sinusoidal halfwave with aperiodic decay. Figure 1-1 shows the discharge curve for various external resistances.
227 446 32-D CardioServ V4.1 7
1.2 For your Safety
g
Warning
Warning
Warning
Warning
Caution
Introduction to CardioServ
The safety information given below is divided into the catego­ries Danger, Warning and Caution.
indicates an imminently hazardous situation which, if not avoided, will result in death or
Danger
serious injury.
indicates a potentially hazardous situation which,
Warnin
if not avoided, could result in death or serious injury.
indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate
Caution
injury and/or damage to the equipment.
provides application tips or other useful information
to assure that you get the most from your equipment.
CardioServ is designed to comply with IEC 60601/
EN60601 requirements. It is Class I equipment and has an internal power source.
CardioServ operates on line voltages between 95 and 240 Volts, 49 to 65 Hz. The mains plug must be connected to an appropriate power supply with a non-fused earthed wire. The use of exten­sion cords is not permitted.
Before putting the device into operation, visually check all connecting cables and electrodes for signs of damage. Damaged cables and electrodes must be replaced immediately, before use.
When disconnecting the device from the power line, first remove the plug from the wall outlet.
Disconnect the device from the power line and operate it on battery power, if the integrity of the protective earth conductor is in doubt.
Set up the device in a location which affords sufficient ventilation. The ventilation openings of the device must not be obstructed. The ambient conditions specified in the Technical Specifica­tions section must be ensured at all times.
8 CardioServ V4.1 227 446 32-D
Introduction to CardioServ
g
g
g
Warning
Warning
Warning
Warnin
Danger
Warnin
Danger
Warnin
Caution
Devices intended for emergency application must not be exposed to low temperatures during storage and transport to avoid moisture condensation at the application site. Wait until all moisture has vaporized before using the device. Avoid using the defibrillator under conditions where prolonged exposure to or excessive contact with moisture can occur.
Before putting the defibrillator into operation, make sure that the paddles and all connection cables are dry.
CardioServ is an emergency device and must be ready for operation at any time. For this reason, the defibrillator battery must always be charged. This can be achieved by leaving the defibrillator connected to the power line when it is not needed in an environment where only battery operation is possible.
Possible explosion hazard if used in the presence of concentrated oxygen.
CardioServ is suitable for application in a humid environment provided the regulations concerning drip-proof equipment of IEC 60601 are strictly observed. However, avoid defibrillation in a very moist or wet environment, unless absolutely necessary.
Use only the original Marquette Hellige batteries, as these are designed for an extended temperature range.
Magnetic and electrical fields are capable of interfering with the proper performance of the device. For this reason make sure that all external devices operated in the vicinity of the defibrillator comply with the relevant EMC requirements. X­ray equipment, MRI devices and radio systems are a possible source of interference as they may emit higher levels of electromagnetic radiation.
The defibrillator is designed for intracardiac appli-
cation.
Devices may be connected to other devices or to parts of systems only when it has been made certain that there is no danger to the patient, the operators, or the environment as a result. In those instances where there is any element of doubt concerning the safety of connected devices, the user must contact the manufacturers concerned or other informed experts as to whether there is any possible danger to the patient, the operator, or the environment as a result of the proposed combina­tion of devices. Standards IEC 60601-1-1/ EN 60601-1-1 must be complied with in all cases.
Set up the device so that the operator has a clear,
unobstructed view of the front panel.
Liquids must not be allowed to enter the device. Devices into which liquids have penetrated must be immediately cleaned and checked by a service technician.
227 446 32-D CardioServ V4.1 9
Introduction to CardioServ
Caution
Warning
g
g
Danger
Caution
Warnin
Use only the original Marquette Hellige patient cables. Do not connect other signal sources to the cables.
Patient signal inputs labelled with the symbol are protected against damage resulting from defibrillation and electrocautery voltages. Nevertheless extreme care must be taken when devices which are directly connected to the patient remain applied during electrocautery or defibril­lation. The distance between ECG electrodes should be at least 15 cm. If in doubt, disconnect the patient cable from the device while applying the defibrillation pulse or performing electrocautery.
For defibrillation of children use only the special clip-on electrodes for children listed in section 15 Order Information and Accessories.
CardioServ also operates on line power when the
battery is depleted or missing.
At the end of its service life, CardioServ and its accessories must be disposed of in compliance with the special waste control regulations for electronic parts. If you have any questions in this matter, please contact Marquette Hellige GmbH.
Dispose of the packaging material, observing the applicable waste-control regulations and keeping it out of childrens reach.
Literature
Medical Device Directive
EN 60601-1/1990 + A 1: 1993 + A2: 1995: Medical electrical equipment. General requirements for safety
EN 60601-1-1/9.1994 + A1 12/1995: General requirements for safety. Requirements for the safety of medical electrical
Warnin
The full responsibility for the use of accessories from other manufacturers lies with the user.
systems.
Check the device performance at regular intervals (once a month), strictly following the instructions
Caution
given in section 3.2. Do not select high energy
IEC Publication 513/1994: Fundamental aspects of safety standards for medical equipment.
levels for test discharges with defibrillation elec­trodes shorted together.
ROY, O.Z.: Summary of cardiac fibrillation thresholds for 60-Hz currents and voltages applied directly to the heart.
For each defibrillation, verify that the selected and the displayed/charged energy are identical.
Caution
10 CardioServ V4.1 227 446 32-D
Med. & Biol. Engn. & Computing 18: 657...659 (1980).
Introduction to CardioServ
g
g
g
g
g
Caution
Danger
Caution
Danger
Caution
General points to note when handling a defibrillator
Electromedical equipment such as the CardioServ
Warnin
defibrillator must only be handled by persons who are trained in the use of such equipment and are capable of applying it properly.
Before using the equipment, the operator must
Warnin
ascertain that it is in correct working order and operating condition.
The defibrillator paddles must be clean and dry.
Danger
The person carrying out the defibrillation should
have at least one assistant.
The operator must be trained in the use of the
Warnin
defibrillator.
All assistants must be briefed regarding the
Warnin
preparations for and execution of defibrillation.
All tasks must be assigned clearly.
Warnin
Defibrillating a patient with normal heart rhythm may induce ventricular fibrillation.
Danger
Position the patient flat on a hard, dry surface where the patient is electrically insulated. The patient must not be allowed to come into contact with metal parts, e.g., bed or litter, in order to prevent unwanted pathways for the defibrillation current which endanger the assistants. For the same reason, do not place the patient on wet ground (rain, accident in swimming pool).
Have a pacemaker at hand, if possible.
Should cardiac arrest occur or be imminent during
preparations for defibrillation, administer heart massage and artificial respiration (CPR).
Do not allow the defibrillator paddles to come into contact with other electrodes or metal parts which are in contact with the patient.
Transducers and instruments that are not defi­brillation-proof must be disconnected from the patient.
Interrupt heart massage and artificial respiration immediately before triggering the shock.
Immediately before triggering the shock discon­nect tubes and have assistants step back.
227 446 32-D CardioServ V4.1 11
Introduction to CardioServ
g
Warning
Warning
Caution
Danger
Warnin
Caution
The patients chest must be dry, as moisture causes unwanted pathways for the defibrillation current. Therefore, when using flammable skin cleansing agents, wait until they have completely dried.
Possible explosion hazard if used in the presence of concentrated oxygen, flammable substances (gasoline) or anesthetic agents. Oxygenation in the vicinity of the defibrillation paddles must be strictly avoided; if necessary, interrupt oxygen supply while defibrillating the patient.
To prevent sparking
– the electrodes should make full contact with
the body
– the electrodes should be pressed firmly onto
the thorax.
When defibrillating children it is especially important to verify that the paddles make full contact with the body surface. This is to be ob­served also when using the clip-on electrodes for children (Part No. 303 439 95).
Defibrillating a patient who has an implanted pacemaker is likely to impair the pacemaker function or cause damage to the pacemaker. Therefore, the following should be observed:
– Select the smallest energy level possible for the
application.
– Do not apply the defibrillation paddles in the
vicinity of the pacemaker electrodes.
– The availability of an external pacemaker is
of utmost importance in this case.
– After the defibrillation the working order of
the implanted pacemaker should be checked immediately.
Also be aware that children require less energy for a successful defibrillation than adults. For the first defibrillation pulse delivered to babies and toddlers, select an energy level of 2 joules/kg body weight. For subsequent shocks, increase the energy up to 4 joules/kg.
12 CardioServ V4.1 227 446 32-D
Controls and Indicators
2. Controls and Indicators
This section describes the CardioServ operating controls and indicators and explains their function.
When operating elements in this manual are identified with a reference number in parentheses, this number refers to Figure 2-1 in this section.
You will also find an explanation of all signs and symbols used on the CardioServ defibrillator.
227 446 32-D CardioServ V4.1 13
Controls and Indicators
3
2
3
Charge
-
+
2
2
5
7
102030
50
100
150
300
J (50 )
200
360
Charge
Shock
Shock
Sync
APEX
Apply electrode cream to both paddles
Press charge button on apex paddle (on the device, when using adhesive or internal electrodes)
Press buttons on both paddles simul­taneously (on the device, when using adhesive or internal electrodes)
Charge defibrillator
Deliver shock3
2
1 Select energy
P
a
c
e
m
a
k
e
r
S
t
i
m
u
l
a
t
e
u
r
Dem
Fix
+
P/min
P/min
+
mA
mA
Shock
Charge Shock
Autosequence
STERNUM
APEX
2
1418 17 16 15
13
11
10
12
1
13 8 975
4
6
19
20
21
22
23
Option
SpO
2
24
EKG
Figure 2-1. Controls and indicators of CardioServ VF
14 CardioServ V4.1 227 446 32-D
Controls and Indicators
1 Defibrillator paddles
Shock
2
button to trigger the defibrillation shock – together
3
with button (8)
3 keys to adjust the LCD contrast. Press both
keys simultaneously to obtain a screen copy.
4 Energy selector, on/off switch
Charge
5
key to charge the unit (manual mode) and to
Shock
trigger the defibrillation shock – together with key (6). This key assumes the function of button (8) when internal paddles or adhesive electrodes are used.
Shock
6
key to trigger the defibrillation shock – together
with key (5). This key assumes the function of button (2) when internal paddles or adhesive electrodes are used.
7 Five selection keys F1 to F5 whose functions change
with the menu displayed. The respective key functions are indicated by symbols or labels in the bottom line on the LCD. From the main menu, that appears on power up, you can access submenus which, in turn, allow the selection of further options. The back function returns you to the next higher menu level. The main menu reappears automatically if you do not depress any of the keys for about 30 seconds.
8
button to charge the unit (manual mode) and
3
2
Charge
Shock
trigger the defibrillation shock together with button (2)
11 key to start and stop the recorder. This key assumes
the function of button (9) when internal paddles or adhesive electrodes are used.
12 Green indicator is lit when the defibrillator operates
on line power.
+
13 Yellow indicator
-
is lit when the defibrillator battery
is being charged.
14 Connection for defibrillator paddles
(controls 15 to 18 only on models with pacemaker)
+
mA
15 Keys
16 Keys
17 Pacing mode selection key
mA
P/min
to adjust the pacer output
-
+
P/min
to adjust pacer rate
-
Dem
(demand/fixed rate)
Fix
18 Key to enable and disable the pacemaker
19 Aperture to open the paper compartment
20 Connector for power cord
21 Unassigned
22 1-Volt ECG output
23 Connector for SpO2 sensor
9 button to start and stop the recorder.
Sync
10
key to switch to the synchronized operating mode
24 Connector for patient cable (ECG signal input)
(section 5 “Cardioversion”)
227 446 32-D CardioServ V4.1 15
Controls and Indicators
Explanation of the signs and symbols used on the defibrillator
Type CF equipment with highly insulated patient connections, suitable for intracardiac application, connections defibrillation-proof.
Type CF equipment with highly insulated patient connections, suitable for intracardiac application, connections not defibrillation-proof.
Standby mode (for line-power operation)
Power off
+
-
Battery charging
Recorder start
ECG signal
Contrast
Standby or preparatory state only for a part of the
equipment
On, only for a part of the equipment
Hardcopy of screen image
Refer to Operators Manual
Direction indicator
Signal output
Audible alarm on/off
High voltage
16 CardioServ V4.1 227 446 32-D
Setting Up CardioServ and Testing Its Performance
3. Setting Up CardioServ and Testing Its Performance
In this section you will find information about
putting CardioServ into operation
connecting CardioServ to the 12-Volt power supply of
an ambulance vehicle
customizing the CardioServ settings to suit your per-
sonal requirements
testing the CardioServ performance before using it on a
patient
227 446 32-D CardioServ V4.1 17
Setting Up CardioServ and Testing Its Performance
3.1 Setting Up CardioServ
The defibrillator is a high-voltage electrotherapy device and must be handled by qualified person-
Danger
nel only. Improper use of this device can endanger life. Do not fail to observe the information given in this manual and only entrust the device to the hands of trained persons.
Check the electrodes and their leads for signs of damage every time before you use the defibrilla­tor. In particular, make a close visual inspection of the insulation. Replace internal electrodes or the contact inserts when you detect signs of me­chanical damage.
CardioServ operates on:
line power (95 to 240 V, 49 to 65 Hz)
battery power (rechargeable batteries), i.e., independent
of the power line
12-Volt power supplied from the emergency vehicle
(with optional defibrillator mounting system)
* Use the power cord to connect the defibrillator to the
power line (Figure 3-1).
Figure 3-1. Power input
18 CardioServ V4.1 227 446 32-D
Setting Up CardioServ and Testing Its Performance
360
x
Shock
Charge
m
a
k
e
r
S
t
i
m
u
l
a
t
e
u
mA
mA
r
+
+
P/min
m
P/min
Shock
Sync
+
-
Figure 3-2. Green indicator (indicating that defibrillator is
supplied from the power line)
Figure 3-3. Inserting the battery
* Check that the green indicator is lit (Figure 3-2).
* Check that the battery is in place (Figure 3-3).
Pull back the catch on the underside of the device to remove the battery. When inserting it, make sure that it clicks prop­erly into place.
+
When a battery is inserted, the yellow indicator
-
starts flashing as soon as the defibrillator is connected to the power line (to indicate that the battery is charging). The battery is fully charged after 16 hours and the indicator is continuously lit.
* Check that the supply of chart paper is sufficient (Figure
3-4). A stripe marks the last 3 meters of the roll.
Figure 3-4. Recording strip
227 446 32-D CardioServ V4.1 19
Setting Up CardioServ and Testing Its Performance
g
Power Supply From Emergency Vehicles
A qualified technician can be called in to connect the CardioServ to the 12-Volt supply of an emergency vehicle. The following points must be noted:
The negative terminal of the ambulance power supply
system must be connected to the ambulance chassis for grounding.
The positive lead of the ambulance power supply system
intended to supply the current must be protected with a
-
+
-
+
-
+
T
e
>
s
9
t
5
>
%
8
0
%
<
8
0
%
T
e
S
s
ta
t
rt
10-A fuse.
Use only the defibrillator mounting system listed in
section 15 Order Information and Accessories or the external charging unit, if your CardioServ has been modified accordingly.
Check that the contacts for power supply from the
Figure 3-5. Accu Service Unit
defibrillator mounting system on the underside of CardioServ (next to battery) are clean. Do not damage
We recommend our Accu Service Unit for optimal
them in any way.
care of the batteries. It prolongs the batteries service life and guarantees their operational readiness at all
Important Information on Battery-Power Operation
times.
Rechargeable batteries require special maintenance and con­tinued checks to assure they function in emergency situations. It is normal for batteries of this type to selfdischarge when not in use.
Warnin
A NiCd battery should not be charged while located in direct sunlight, over a radiator, in cold storage, or in other temperature extremes (not below 5 °C). When the instrument is charging, ambient temperatures exceeding 40 °C may adversely affect battery capacity and life.
20 CardioServ V4.1 227 446 32-D
The battery charges automatically when CardioServ is con­nected to the power line (yellow indicator (13) flashes).
Setting Up CardioServ and Testing Its Performance
In order to ensure its functioning as an emergency device, the defibrillator should not be disconnected from the power line for more than 48 hours.
A fully charged battery supplies power for 35 defibrillation shocks of 360 joules or 2 hours of monitoring (or 1.2 hours of monitoring if the CardioServ unit has pacemaker and SpO option). It takes 16 hours to charge a depleted battery.
Proceed as follows to test the battery charge level:
Disconnect CardioServ from the power line. Trigger a test discharge of 360 joules (see
section 3.2 Testing the Defibrillator Perform­ance).
If you are not prompted to charge the battery,
the charge level should be sufficient for at least 5 more 360-joule shocks.
Monthly battery maintenance and checks:
1. Disconnect CardioServ from the power line and discharge fully charged battery in the monitoring mode. To do so, set energy selector switch to , (SpO2 sensor not connected) and wait until device switches off.
2
Inserting CardioServ In Its Softcase
* Open both zips on the CardioServ softcase.
* Undo the two Velcro strips on the front of the softcase.
* Open the Velcro flap located in front of the CardioServ
printer.
* Place the softcase on the small base, so that the softcase is
positioned as though you were carrying it.
* Disconnect the paddles on the CardioServ and remove
them.
* Hold all the opened up parts of the softcase to the side
and put the CardioServ into the softcase.
* Pull the softcase by the two Velcro strips on the black
base into position, ensuring that the feet of the CardioServ are in the cut-out openings provided.
* Close the two front Velcro strips as well as the flap in
front of the printer at the side.
2. Check how long it takes before battery is depleted. If the time is less than 1.8 hours, the battery is too old or improperly maintained and must be replaced.
3. Recharge the battery. This will take 16 hours.
For easy, convenient care and maintenance of the batteries, use our Accu Service Unit”.
227 446 32-D CardioServ V4.1 21
* Close the zips.
* If the individual cut-out openings for the external con-
nections, paddles or feet are not correctly positioned, pull the material at this point into the right shape.
* Reconnect the paddles and put them back into the
CardioServ recesses.
Customizing the defibrillator settings
Setting Up CardioServ and Testing Its Performance
Further steps that can be taken while setting up the defibrilla­tor include customizing the device functions. This allows you to select defibrillator default settings which suit your personal preferences. CardioServ saves these configured settings and reactivates them automatically every time you switch the
Refer to section 11 The Defaults Menu for a detailed explanation of how to customize the CardioServ, including the language selection (available languages are English, French, German, Italian, Portuguese, Russian, Spanish and
Swedish). defibrillator on. The following chart shows the factory settings of all parameters and the optional adjustments.
retemaraP noitpircseD gnitteSyrotcaF egnartnemtsujdA sgnittesdezimotsuC
stimiLRH stimilmralaRH 061/04 ffo003ot51ffo
ytivitisneS ezislangisGCE 1 Vm/mc2,1,5.
daeL 1 selddap+sdaeldradnatslla
.qesotuA slevelygrenegnicneuqesotua J063,J002,J002 ,J063,J003,J002,J051
OpS
stimiL stimilmrala ffo/09 ffo001ot51ffo
2
kcoL-C OpSdezinorhcnys-GCE
OpS
2
resU )sretcarahc04(emanrotxeteerf
emiT.tnI OpS
mralAliaFdaeL ffo s03/ffo
peeBSRQ ffo ffo/no
enoTmralA ffo ffo/no
tuotnirPmralA tratsredrocer.motua ffo ffo/no tuotnirPkcohS tratsredrocer.motua no ffo/no
tuotnirP.tnoC tratslaunamnopupotsredrocer ffo ffo/no
rekamecaP etareslup MPB06 nim/P051ot51 yalpsiD lamron desrever/lamron yalpsiD seerged0 seerged081/seerged0
emuloV slangisoiduallafoemulov duol wol/duol
retliFCA zH05 ffo/zH06/zH05
retliFelcsuM tcafitranoitomfonoitanimile no ffo/no
emiT/etaD
tamroFetaD YYYY:MM:DD YYYY:DD:MM/YYYY:MM:DD
egaugnaL namreG ,namreG,hcnerF,hsilgnE
tluafedyrotcaF sgnittesyrotcafhsilbatseer ffo ffo/no
2
2
emitnoitargetni s8 )dednemmocerton(s21,s8,s4
tnemerusaem ffo ffo/no
,naissuR,eseugutroP,nailatI
hsidewS,hsinapS
Table 1
22 CardioServ V4.1 227 446 32-D
Setting Up CardioServ and Testing Its Performance
m
J (50 )
20
10
7
5
2
Autosequence
P
a
c
e
m
a
k
e
r
S
t
i
Figure 3-6. Switching on CardioServ
a b
Paddle
c
0
160 BPM
140 ECG
HR Alarm
d
3.2 Testing the Defibrillator
30
50
100
150
200
300
360
Charge
Shock
Shock
Sync
e
f
Performance
On power up CardioServ runs an automatic selftest. Any malfunctions identified during this test result in an error message displayed on the LCD (refer to section 12 “Error Indications and Messages). As a further performance test a trial defibrillation can be triggered.
The energy selector is used to switch CardioServ on and off. Once you have become familiar with CardioServ you can thus switch on the defibrillator and select the required energy in one single operation.
* Set the energy selector to the position. No energy
will be stored in this position of the switch.
The defibrillator beeps and displays a checkered pattern (LCD performance test). Next the main display appears (Figure 3-7).
g
The Paddle message (b, Figure 3-7) indicates that no pa­tient cable is connected and that the ECG signal is acquired via the defib paddles. Upon connection of the patient cable
FreezeSpO2ECG
Next
Menu
the selected ECG lead is displayed here. CardioServ is now ready for operation.
Figure 3-7. CardioServ main display
a heart/pulse-rate reading with alarm limits b ECG signal source c 1-mV calibration pulse d alarm message, QRS blip e signal trace, channel 1 f signal trace, channel 2 g menu
227 446 32-D CardioServ V4.1 23
Setting Up CardioServ and Testing Its Performance
Do not trigger more than 5 consecutive test discharges (or internal safety discharges) within
Caution
15 minutes at max. energy setting.
Paddle
0
160 BPM
140 ECG
HR Alarm
Figure 3-8. Display of selected energy
360 J
Next
MenuFreezeSpO2ECG
Test Discharge
A test discharge can be triggered to check the defibrillator discharge circuit. For this test the stored energy is discharged into the device via two contacts in the paddle compartments.
* Set the energy selector to 360 joules (50 joules, if inter-
nal electrodes are connected). The display first shows the selected energy (Figure 3-8).
2
Charge
* Press the
(Press the
Shock
button on the paddle to charge the unit.
3
Charge
key on the defibrillator, when using
Shock
internal electrodes.)
You may now watch the defibrillator charging.
* When the selected energy level has been reached,
CardioServ beeps and the stored energy is displayed (Figure 3-9).
* Trigger the defibrillation pulse within the next 30 sec-
onds. To do so, simultaneously press the buttons on
Charge
both paddles (or press the two
Shock
Shock
keys on the
defibrillator, when using internal electrodes).
If more than 5% of the available energy are lost before the defibrillation pulse is triggered, CardioServ recharges until the required energy level is reached.
24 CardioServ V4.1 227 446 32-D
Setting Up CardioServ and Testing Its Performance
Caution
Paddle
0
160 BPM
140 ECG
HR Alarm
Energy available
Figure 3-9. Display of available energy
If you do not trigger the defibrillation pulse within
30 seconds, an internal safety discharge is initiated automatically.
360 J
Next
MenuFreezeSpO2ECG
After defibrillation, the beeping sound stops, and the energy actually delivered into a 50-ohm resistance is displayed for 10 seconds in place of the stored energy (Figure 3-9). The delivered energy must not deviate more than ±15% or ±4 joules (whichever is greater) from the selected value. A recording is initiated at the same time (16-second strip).
Should the discharge circuit be interrupted (paddles not properly placed on contacts in the compartments, defective lead), an internal safety discharge is initiated 200 ms after the defibrillation shock has been triggered. In this case the deliv­ered energy is “0”.
If the defibrillator cannot store the selected energy so that selected and stored energy values differ, the LCD shows the message Energy high or Energy low. The defibrillation pulse can be triggeed all the same.
* Switch off CardioServ (set energy selector to ).
The message Energy high or Energy low indicates that CardioServ needs to be repaired.
Caution
If, in spite of this energy storage problem, the device has to be employed, it will display the message Self-test failed. Charge Energy Error
Testing the Pacemaker Performance
The performance of the pacemaker can be tested with a commercially available pacemaker tester (e.g. CS300 Simula­tor from Marquette Hellige, part no. 417 983-001).
upon power up. In this situation press one of the function keys and proceed as usual.
Performance Test
Test the defibrillator performance once a week: test 1 – defibrillator connected to mains, battery
removed,
test 2 – defibrillator disconnected from mains,
battery inserted
227 446 32-D CardioServ V4.1 25
For your notes
Setting Up CardioServ and Testing Its Performance
26 CardioServ V4.1 227 446 32-D
Non-Synchronized Defibrillation
4. Non-Synchronized Defibrillation
This section describes first how to perform a non-synchro­nized defibrillation, using the standard defibrillation paddles. The subsequent explanations refer to the use of internal and disposable adhesive electrodes.
At the end of this section you will find a summary of all necessary operating steps (brief operating instructions).
227 446 32-D CardioServ V4.1 27
Non-Synchronized Defibrillation
g
4.1 Defibrillation with Standard Electrodes
The information given in section 1.2 must be
Warnin
observed without fail to ensure safe and reliable application of the device.
Always switch off CardioServ before exchanging the defibrillation paddles.
Danger
Paddle
0
160 BPM
140 ECG
HR Alarm
200 J
SpO2ECG
Figure 4-1. Main display, indication of ECG lead, manual
operation and selected energy
Next
MenuFreeze
The energy selector can be set to the autosequence position, where the defibrillator automatically sequences the preset energy levels. The preset factory default setting is the sequence recommended by AHA/ERC (200 J, 200 J, 360 J). The factory default values can be changed from the defaults menu. You can choose among 150 J, 200 J, 300 J and 360 J.
* Set the energy selector (4) to “Autosequence or to the
required energy value (this turns on CardioServ).
The defibrillator beeps and displays a checkered pattern (LCD performance test). Next the main display appears (Figure 4-1). When a patient cable is connected the selected ECG lead is displayed (selectable).
* Check that the energy selector locks in on the correct
position and that the display shows the selected energy (Figure 4-1).
The energy depends on the defibrillation mode, on the patients age and constitution. In external application the thickness of the tissue is also a factor which influences the amount of energy required.
The energy necessary for successful ventricular defibrillation without damaging the myocardium has for many years been a matter of scientific controversy. The manufacturer is therefore not able to give any recommendations.
28 CardioServ V4.1 227 446 32-D
Figure 4-2. Removing the paddles
g
Do not apply the paddles over
Warnin
sternum or claviclenipplesimplanted pacemaker or defibrillator.
Non-Synchronized Defibrillation
In emergency situations AHA recommends for adult patients
1. defibrillation with 200 joules; if unsuccessful, repeat
2. defibrillation with 200 joules; if unsuccessful, repeat
3. defibrillation with max. energy setting (360 joules).
Please note that children require less energy for successful ventricular defibrillation than adults. For the first defibrilla­tion pulse delivered to babies and small children, select an energy level of 2 joules/kg body weight. For subsequent shocks, the energy may be increased to 4 joules/kg.
In compliance with IEC requirements the energy adjusted on this defibrillator is not the stored energy, but the energy released into an external resistance of 50 ohms (patient resist­ance + electrode-to-skin contact resistance). The energy selector is labelled accordingly.
* Remove the paddles from their compartments (as shown
in Figure 4-2). Carefully dry the electrodes, if they are wet. The handles, in particular, must be completely dry. Apply an ample amount of electrode cream to each paddle.
STERNUM
3
2
C h
a r
g e
3
APEX
* Apply the electrodes to the patients thorax so that the
greatest possible amount of energy flows through the myocardium (the imaginary connecting line between the two electrode centers should be identical with the cardiac median line; Figure 4-3).
* Press paddles firmly down onto the patients thorax..
Figure 4-3. Paddle application points
227 446 32-D CardioServ V4.1 29
Non-Synchronized Defibrillation
2
Charge
Shock
3
STERNUM
2
3
Charge
Shock
8
APEX
Figure 4-4. Buttons to initiate energy storage and to trigger the
defibrillation shock
Paddle
0
160 BPM
140 ECG
HR Alarm
The ECG signal trace appears on the monitor.
* Do not touch the patient any more and warn all those
present.
2
Charge
* Press the
Shock
button (8) on the apex paddle to initiate
3
charging (Figure 4-4). When using internal electrodes or defibrillation pads, press
Charge
on the device.
Shock
When the selected energy level has been reached (message Energy available), CardioServ beeps and the stored energy is displayed (Figure 4-5).
* Trigger the defibrillation shock within the next 30 sec-
onds by simultaneously pressing the buttons on both paddles (Figure 4-4). When using internal electrodes or defibrillation pads, press
Charge
Shock
and
Shock
on the device.
If more than 5% of the available energy are lost before the defibrillation shock is triggered, CardioServ recharges until the required energy level is reached.
Energy available
FreezeSpO2ECG
200 J
Next
Menu
Figure 4-5. Display of available energy
When the defibrillator is already charged you can increase the energy level simply by turning the energy selector to the new setting. To decrease the energy level, set the selector to the lower value and initiate charging again.
30 CardioServ V4.1 227 446 32-D
Non-Synchronized Defibrillation
g
Paddle
0
160 BPM
140 ECG
HR Alarm
Energy delivered
Figure 4-6. Display of delivered energy
The Check Electrode message refers to the defibril-
lation paddles only when no patient cable is con­nected. It refers to the ECG electrodes when the patient cable is plugged in.
If you do not trigger the defibrillation pulse within
30 seconds, an internal safety discharge is initiated automatically. You will then have to recharge the defibrillator.
The defibrillation energy (high voltage!) remains
Warnin
applied to the defibrillation paddles until fully discharged. Do not touch the paddle surface! The internal safety discharge is completed when the selected energy is displayed again.
198 J
Next
MenuFreezeSpO2ECG
After defibrillation, the beeping sound stops, and the
energy actually delivered is displayed for 10 seconds in place of the stored energy (Figure 4-6). At the same time the recorder writes a 16-second ECG (including a his­tory of 4 seconds) (adjustable, Figure 4-7). CardioServ saves this recording (4 s history, 5 s blanked, 10 s after release of shock), and it can be printed off again at any time. Also refer to section 7 The Memories of CardioServ.
When the electrodes are not applied at all or they are not
properly applied to the skin, the message Check Elec­trode is displayed. The defibrillation shock can be triggered all the same. Nevertheless it is recommended to reduce skin impedance (risk of skin burns!), for instance, by applying more electrode cream to the paddles or by pressing them down firmly.
When there is a break in the discharge circuit (paddles
not properly applied, leads or paddles defective), an internal safety discharge is initiated 200 ms after the defibrillation shock has been triggered. In this case the delivered energy is 0.
_ If the defibrillator cannot store the selected energy so
that selected and stored energy values differ, the LCD displays a warning. The defibrillation pulse can be triggered anyway (notify service).
* Once therapy has ended, set the selector switch to
for monitoring of the patients ECG.
227 446 32-D CardioServ V4.1 31
Non-Synchronized Defibrillation
a
DEFIB 09. 07 .1997 09 : 16 : 12 50 Hz 35 Hz 360 JOULES HR : O BPM SHOCK
PADDLE 1 CM/MV 25 MM/S ST. JOHN’S 97
g
h
i
b
k
c
<5s>
ed
Figure 4-7. Example of a recording initiated by a defibrillation pulse
a initiation e delivered energy i paper speed b date f heart rate k name of hospital/practise c time g ECG lead d active filters h sensitivity
f
PATIENT
DATE OF BIRTH.............
USER........................
COMMENTS.................
SELECTED ENERGY.........
DELIVERED ENERGY........
ALARM LIMIT HR...........
ALARM LIMIT SPO2 .......
PACEMAKER................
DATE/TIME
* After use, switch off CardioServ (set selector switch to
).
* Clean the paddles and the device as described in section
13.
* After cleaning, return the paddles to their compartments
as shown in Figure 4-8.
Figure 4-8. Returning the paddles to their compartments
32 CardioServ V4.1 227 446 32-D
Non-Synchronized Defibrillation
i
i
4.2 Defibrillation with Internal Electrodes or Single-Use Pads
Internal Electrodes
When using CardioServ with internal or adhesive
electrodes, charging is initiated with
Charge
Shock
and defi-
brillation is triggered by simultaneously pressing the
Charge
Shock
and
Shock
(Figure 4-9). Only sterile electrodes
may be used for internal defibrillation!
J (50 )
20
utosequence
S
t
i
m
u
n
n
mA
10
7
5
2
l
a
t
e
u
r
+
30
360
50
300
100
150
200
Charge
Shock
Shock
Sync
Figure 4-9. Panel keys to initiate charging and to trigger the
defibrillation shock
When electrodes for internal defibrillation are connected to CardioServ, it is not posssible to store energy above 50 joules. If you set the selector to values higher than 50 joules, release of the defibrillation pulse is blocked for reasons of patient safety (Energy high will be displayed!). Simply turn the dial back and initiate charging again, this time selecting not more than 50 joules.
Spoon-shaped electrodes are used for internal defibrillation. Their contact surface must match the dimensions of the heart, as the spoons should make full contact with the tissue.
You can choose from 3 different spoon sizes (section 15 “Order Information, Accessories”). As the spoons are in direct contact with the heart – hence the term “direct defibrillation” – energy levels considerably lower than those for transthoracic (external) defibrillation are sufficient.
Please note that internal electrodes must be sterilized before use (section 13 Cleaning and Disinfection”).
227 446 32-D CardioServ V4.1 33
Non-Synchronized Defibrillation
STERNUM
electrode + connector
electrode + connector
Figure 4-10. Anterior – anterior placement
front back
APEX
Single-Use Defibrillation Pads
* Use pads before their expiration date.
* A pair of defibrillation pads may remain attached to
the patient for up to 24 hours. They withstand up to 50 shocks of 360 joules each.
* Apply the pads (part no. 919 202 94 adult pads, part no.
919 202 95 pediatric pads) as shown in Figure 4-10 for an anterior-anterior placement, and as shown in Figure 4-11 for an anterior-posterior placement:
– Shave any hair from each site. This improves conduc-
tivity and makes removal of the pad easier.
– Place the pads on the patient so that the connectors
point to either side of the patient. In this position the connecting cables will not hinder CPR measures.
– The electrodes are pregelled; therefore do not use
APEX STERNUM
additional contact cream or paste.
Figure 4-11. Anterior – posterior placement
Do not use pads, if the gel has dried out.
Peel off the backing from each pad and place the pad
carefully on the appropriate site.
34 CardioServ V4.1 227 446 32-D
223 383 01
Warning
Warning
Non-Synchronized Defibrillation
* Then press the connector of cable 223 383 01 on to the
electrode contact pin until you hear it click into place. Observe the connector lables: A = apex, S = sternum.
A
Figure 4-12. Connecting the cables
223 346 01
919 202 94 919 202 95 (K)
919 201 89 919 202 75 (K)
The round adhesive electrodes (part no. 919 201 89) can be used with adapter lead 223 346 01. When connecting the electrode to the lead, take care that they engage properly. To disconnect them, simply press on the rear part of the catch (Figure 4-13).
* Before defibrillating the patient, verify the position and
adhesion of the pads.
* Defibrillate the patient as described in section “Defibril-
lation with Standard Electrodes. Note, however, that energy storage and defibrillation pulse will have to be triggered with the panel keys.
* After use, carefully peel off the electrodes from the
patients skin (Figure 4-14) and discard them immedi­ately.
Figure 4-13. Connecting the defib pads to the adapter cable
When monitoring the patient with the adhesive defib pads, make sure that the energy selector is in
incorrect
correct
the monitoring position .
Discard disposable defibrillation pads immedi­ately after use. Do no reuse them!
Figure 4-14. Removing defibrillation pads
227 446 32-D CardioServ V4.1 35
Non-Synchronized Defibrillation
4.3 Brief Operating Instructions (non-synchronized)
External Electrodes/Paddles
* Set energy selector to “Autosequence” or select
the required energy
* Remove paddles from compartments and apply
electrode cream
* Apply electrodes to thorax
2
Charge
* Initiate defibrillator charging
Shock
3
* Wait for beep to sound and for stored energy to
be displayed
* Warn bystanders, do not touch the patient any
more and trigger shock; to do this, simultane-
2
Charge
Shock
ously press
Shock
on both paddles
3
3
* Watch ECG, repeat defibrillation if necessary or
set energy selector to , if defibrillation was successful
* Switch off CardioServ after use (set energy
selector to )
* Clean paddles and defibrillator
Internal Electrodes or Pads
* Connect internal electrodes or pads to
1
CardioServ
* Apply pads
* Set energy selector to “Autosequence” or select
the required energy (50 joules max. for internal
1
defibrillation)
2
* Initiate defibrillator charging
* Wait for beep to sound and for stored energy to
be displayed
3
* Warn bystanders, do not touch the patient any
more and trigger shock; to do this, simultane-
Charge
ously press the
Shock
Shock
* Watch ECG, repeat defibrillation if necessary or
set energy selector to , if defibrillation was successful
* Switch off CardioServ after use (set energy
selector to )
* Dispose of single-use electrodes, clean defibrilla-
tor
Charge
Shock
keys on the panel
2
3
* Clean/sterilize internal electrodes
36 CardioServ V4.1 227 446 32-D
Cardioversion (Synchronized Defibrillation)
5. Cardioversion
(Synchronized Defibrillation)
This section describes first how to perform a synchronized defibrillation (cardioversion). For this purpose, the ECG is picked up either via the paddles or via separate ECG elec­trodes.
At the end of this section you will find a summary of all necessary operating steps (brief operating instructions).
227 446 32-D CardioServ V4.1 37
Cardioversion (Synchronized Defibrillation)
g
5.1 General Information
Warnin
Caution
The information given in section 1.2 must be observed without fail to ensure safe and reliable application of the device. Also read section 6.1 Displaying the ECG.
The American Heart Asociation AHA recom­mends the following energy levels for cardio­version: 50 J, 100 J, 200 J, 300 j, 360 J.
For cardioversion, the defibrillation shock is delivered in synchronization with the heart action (on the R-wave), as the heart is still working. As a prerequisite the patients ECG signal must be supplied to the defibrillator. After the defibril­lator has received the “defibrillation command” from the operator who pressed the appropriate keys, the defibrillator will wait for the next R-wave to derive the trigger signal.
The ECG signal can be sensed either via a special patient cable or via defibrillation pads. Use the adhesive pads, since the standard paddles may cause major motion artifact during ECG acquisition.
We recommend ECG acquisition by means of the patient cable whenever possible.
The energy selector can be set to the autosequence position, where the defibrillator automatically sequences the preset energy levels. The preset factory default setting is the sequence recommended by AHA/ERC (200 J, 200 J, 360 J). The factory default values can be changed from the defaults menu. You can choose among 150 J, 200 J, 300 J and 360 J.
38 CardioServ V4.1 227 446 32-D
Cardioversion (Synchronized Defibrillation)
STERNUM
electrode + connector
electrode + connector
Figure 5-1. Anterior – anterior placement
front back
APEX STERNUM
APEX
5.2 Performing Cardioversion
ECG Acquisition via Defibrillation Pads
* Apply the pads as shown in Figure 5-1or 5-2 (also refer
to section 4.2 Defibrillation with Internal Electrodes or Single-Use Pads”).
* Check that no patient cable is connected to ECG signal
input (Figure 5-3).
The ECG will now be acquired via the defibrillation pads. Perform cardioversion as described below.
ECG Acquisition via ECG Electrodes and Patient Cable
Use only siver/silver chloride electrodes if you intend to acquire the ECG signal via the patient cable. This type of electrodes prevents polarization voltages which may be caused by the defibrillation shock, resulting in an ECG trace simulat­ing cardiac arrest.
Figure 5-2. Anterior – posterior placement
* Apply the electrodes (choosing the EINTHOVEN limb lead
system, for instance) and connect them to CardioServ via the patient cable (Figure 5-3).
For further information on ECG signal acquisition, please refer to our application note on electrocardiography and to
Option
SpO
EKG
2
the relevant literature.
Figure 5-3. ECG signal input
227 446 32-D CardioServ V4.1 39
Cardioversion (Synchronized Defibrillation)
I...III
aV...V
Paddle
1
cm/mV
2
100
150
200
300
360
u
r
-
3
Shock
Charge
Shock
Sync
+
Figure 5-4. Key to select synchronized defibrillation
3
APEX
Charge
Shock
Cardioversion
* Set the energy selector to (this turns on
CardioServ) and check the ECG trace.
The defibrillator beeps and displays a checkered pattern (LCD performance test). Next the main display and the label of the selected lead are shown (when patient cable is connected).
CardioServ is now ready for operation.
The defibrillator selects the following settings:
APEX
EINTHOVEN lead I (selectable) AC line filter on (selectable) sensitivity of 1 mV/cm (selectable)
If you wish to select another lead or sensitivity, proceed as follows (these are only temporary changes which will not be saved):
* Press the
* Use the
ECG
softkey to call up the ECG submenu.
softkey to select a suitable ECG lead
(shown at a, Figure 5-5).
* Use the
* Press
Sync
softkey to select the sensitivity.
(Figure 5-4).
The Sync mode is indicated on the LCD (c, Figure 5-5).
40 CardioServ V4.1 227 446 32-D
Cardioversion (Synchronized Defibrillation)
g
II
62
160 BPM
140 ECG
d
Figure 5-5. Screen display
a selected lead b Sync mark c Sync mode on d heart symbol
On the recording strip each sync pulse is identified
with a dash above and below the ECG trace.
a
Verify that the heart symbol flashes regularly on the LCD and that the sync mark appears at regular intervals along the upper
b
c
LCD margin (b, Figure 5-5). Otherwise select another lead
I...III
using softkey
aV...V
of the ECG submenu.
Paddle
When the synchronized defibrillation mode is selected, each QRS complex is identified with a SYNC mark (b, Figure
Sync
5-5). If these SYNC marks are missing, synchronized defibril­lation will not be possible, i.e., the defibrillator will not deliver a shock. Reasons for missing SYNC marks include
Next
MenuFreezeSpO2ECG
poor ECG signal quality. Select another ECG lead and/or check electrode contact.
Selecting the non-synchronized mode to defibrillate the patient is only necessary in acute emergencies, such as ven­tricular fibrillation, because this condition almost certainly precludes detection of a QRS signal to trigger the shock.
* Remove the paddles from their compartments. Carefully
dry the electrodes, if they are wet. The handles, in par­ticular, must be completely dry. Apply an ample amount of electrode cream to each paddle.
Warnin
The defibrillation shock cannot be triggered, when the SYNC marks are missing.
* Set the energy selector to “Autosequence” or to the
required energy level.
The energy depends on the defibrillation mode, on the pa­tients age and constitution. In external application the thick­ness of the tissue is also a factor which influences the amount of energy required.
The energy necessary for successful ventricular defibrillation without damaging the myocardium has for many years been a matter of scientific controversy. The manufacturer is therefore not able to give any recommendations.
227 446 32-D CardioServ V4.1 41
STERNUM
g
Figure 5-6. Paddle placement
Cardioversion (Synchronized Defibrillation)
When electrodes for internal defibrillation are connected to CardioServ, it is not possible to store energy above 50 joules. If you set the selector to values higher than 50 joules, release of the defibrillation pulse is blocked for reasons of patient safety (Energy high will be displayed!). Simply turn the dial back and initiate charging again, this time selecting not more
3
than 50 joules.
In compliance with IEC requirements the energy adjusted on
2
Charge
3
APEX
this defibrillator is not the stored energy, but the energy released into an external resistance of 50 ohms (patient resist­ance + electrode-to-skin contact resistance). The energy selector is labelled accordingly.
Do not apply the paddles over
Warnin
sternum or claviclenipplesimplanted pacemaker or defibrillator.
2
Charge
Shock
3
STERNUM
2
3
Charge
Shock
8
APEX
Figure 5-7. Buttons to inititate energy storage and to trigger the
defibrillation shock
* Apply the electrodes to the patients thorax so that the
greatest possible amount of energy flows through the myocardium (the imaginary connecting line between the two electrode centers should be identical with the cardiac median line; Figure 5-6).
* Press paddles firmly down onto the patients thorax.
* Do not touch the patient any more and warn all those
present.
2
Charge
* Press the
Shock
button (8) on the apex paddle to initiate
3
charging (Figure 5-7). When using internal electrodes or defibrillation pads, press the
Charge
key on the CardioServ
Shock
control panel.
When the selected energy level has been reached, CardioServ beeps and the stored energy is displayed (Figure 5-8). It should be identical with the selected value (±15% or ± 4 joules, whichever is greater)
42 CardioServ V4.1 227 446 32-D
Cardioversion (Synchronized Defibrillation)
g
II
62
160 BPM
140 ECG
Energy available
Figure 5-8. Display of the stored energy
The Check Electrode message refers to the defibril-
lation paddles only when no patient cable is con­nected. It refers to ECG electrodes when the patient cable is plugged in.
The defibrillation energy (high voltage!) remains
Warnin
applied to the defibrillation paddles until fully discharged. Do not touch the paddle surface! The internal safety discharge is completed when the selected energy is displayed again.
100 J
Next
MenuFreezeSpO2ECG
Sync
* Simultaneously press the butttons on both paddles
(Figure 5-7) within 30 seconds. The next synchroniza­tion pulse will release the defibrillation shock. When using internal electrodes or pads, press the panel keys
Charge
Shock
and
Shock
to deliver the shock.
After defibrillation, the defibrillator stops beeping, and
the energy actually delivered to the patient is displayed for 10 seconds in place of the stored energy. At the same time the recorder writes a 16-second ECG, including a history of 4 seconds (adjustable, Figure 5-9). CardioServ saves the ECG (4-s history, 5 s blanked, 10 s after release of shock). On the recording the blanked period of time is marked with a spike (Figure 5-9). Refer to section 7
The Memories of CardioServ”.
When the electrodes are not applied at all or they are not
properly applied to the skin, the message Check Elec­trode is displayed. The defibrillation pulse can be triggered all the same. Nevertheless it is recommended to reduce skin impedance, for instance, by applying more electrode cream to the paddles or by pressing them down firmly.
When there is a break in the discharge circuit (paddles
not properly applied, leads or paddles defective), an internal safety discharge is initiated 200 ms after the defibrillation shock has been triggered. In this case the delivered energy is not indicated.
If the defibrillator cannot store the selected energy so
that selected and stored energy values differ, the LCD shows he message Energy high or Energy low. The defibrillation pulse can be triggered all the same.
227 446 32-D CardioServ V4.1 43
Cardioversion (Synchronized Defibrillation)
a
DEFI 09. 04 .1997 09 : 16 : 12 SYNC 100 JOULES HR : 59 BPM SHOCK
I 1 CM/MV 25 MM/S ST. JOHNS DGH 97
g
Figure 5-9. Example of a recording initiated by a defibrillation pulse
II
h
i
a initiation b date c time d SYNC mode e delivered energy f heart rate g ECG lead (here: EINTHOVEN I) h sensitivity i paper speed k name of hospital/practice
b d
k
c
I
<5s>
e
f
44 CardioServ V4.1 227 446 32-D
Cardioversion (Synchronized Defibrillation)
g
Always switch off Cardio Serv before exchanging
Warnin
the defibrillation paddles.
Following each synchronized defibrillation, CardioServ reverts to the non-synchronized mode. Please note that the synchronized mode must be deliberately activated with
Sync
every time you want to perform cardioversion. This measure is to ensure that in emergencies the defibrillator is always ready to deliver nonsynchronized shocks.
* Once therapy has ended, set the selector switch to
for monitoring of the patients ECG.
* After use, switch off CardioServ (Set selector switch to
).
* Clean electrodes and the defibrillator as described in
section 13.
* After cleaning, return the paddles to their compartments.
* When using CardioServ with internal electrodes or pads,
please refer to section 4.2 Defibrillation with Internal Electrodes or Single-Use Pads”.
227 446 32-D CardioServ V4.1 45
Cardioversion (Synchronized Defibrillation)
5.3 Brief Operating Instructions (cardioversion)
ECG Acquisition with ECG Electrodes (this is the preferred method)
* Apply ECG electrodes and connect them to
CardioServ, using the patient cable
* Set energy selector to required energy
* Select synchronized operating mode (
Sync
) and check LCD for regular trigger pulses; if they do not appear, select another lead (F1
ECG
)
* Remove paddles from their compartments and
apply electrode cream
* Apply paddles to patient and initiate defibrillator
2
Charge
charging
Shock
3
* Wait for beep to sound and for available energy
to be displayed
* Warn bystanders, do not touch the patient any
more and trigger pulse; to do this, simultaneously
2
Charge
Shock
press the buttons
Shock
3
on both paddles; wait
3
for shock to be delivered
* Watch ECG, repeat defibrillation if necessary or
set energy selector to , if defibrillation was successful
ECG Acquisition with Defib Pads
* Apply pads to patient and connect them to
CardioServ; a patient cable must not be con- nected to the unit (24)!
1
* Set energy selector to required energy
* Select synchronized operating mode (
Sync
)
1
* Check LCD for regular trigger pulses; if they do
not appear, apply ECG electrodes for signal acquisition
2
* Initiate defibrillator charging
Charge
Shock
2
* Wait for beep to sound and for available energy
to be displayed
* Warn bystanders, do not touch the patient any
3
press
Charge
Shock
Shock
; wait for shock to be delivered
more and trigger pulse; to do this, simultaneously
3
* Switch off CardioServ after use
* Switch off CardioServ after use (energy selector
at )
* Discard disposable electrodes, clean paddles and
defibrillator
46 CardioServ V4.1 227 446 32-D
Displaying and Monitoring the ECG
6. Displaying and Monitoring the ECG
This section describes how to apply the ECG electrodes in order to display all 12 standard leads, and how to monitor the heart rate (adjusting alarm tone, QRS beep and alarm limits).
At the end of this section you will find a summary of all nesessary operating steps (brief operating instructions).
227 446 32-D CardioServ V4.1 47
Displaying and Monitoring the ECG
red
R
yellow
L
white
C
IEC
red
configuration
NF
black green
white
RA
black
LA
brown
V
AHA
white
configuration
LL
RL
green red
Figure 6-1. Electrode application points
5-lead cable (left) 3-lead cable (right)
With a 3-lead cable only ECG leads I,II and III can
be displayed.
When monitoring the patient via the adhesive defib pads, make sure that the energy selector is in
Caution
the monitoring position .
R
yellow
L
6.1 Displaying the ECG
For a quick diagnosis the ECG signal can be sensed via the defibrillation paddles (see below). For more accurate examina­tions and heart-rate monitoring, however, ECG electrodes
N
black
RA
black
LA
must be applied.
Either 3 or 5 electrodes can be used for ECG acquisition. In emergencies, 3 electrodes are sufficient. Use only silver/silver chloride electrodes. This type of electrodes prevents polariza­tion voltages which may be caused by the defibrillation pulse, resulting in an ECG trace simulating cardiac arrest.
LL
red
For detailed information on ECG signal acquisition, please refer to our application note on electrocardiography and to the relevant literature.
* Apply the electrodes as shown in Figure 6-1.
* Connect the patient cable to the electrodes and to the
ECG signal input (Figure 6-2).
If you prefer to use 5 ECG electrodes, apply 4 limb-lead
electrodes, for instance, and one precordial electrode.
Option
SpO
EKG
2
Figure 6-2. ECG signal input
48 CardioServ V4.1 227 446 32-D
C1
C2
C3
Displaying and Monitoring the ECG
To obtain a recording of all 12 standard leads, attach the
C5
C4
C8
C6
C7
The C-electrodes of the IEC system shown here are
the V-electrodes of the AHA system.
4 limb-lead electrodes and apply suction electrode 217 144 01 to the thorax. This electrode is easy to move from one pick-up point to the next (C1 through C6, Figure 6-3). Use electrode lead 223 404 10 to connect the suction electrode to the patient cable. While record-
I...III
ing the chest leads, leave softkey
aV...V
Paddle
set to V and simply move the chest electrode to application points C1 through C6 (C8).
C1
C2C3
C4
Figure 6-3. Chest electrode application points
C1 in the 4th intercostal space at the right sternal
edge C2 in the 4th intercostal space at the left sternal edge C3 at the level of the 5th rib midway between C2
and C4 C4 in the 5th intercostal space on the left midclavi-
cular line C5 between C4 and C6 on the left anterior axillary
line C6 on the mid-axillary line at the level of C4 C7 in the 5th intercostal space on the left posterior
axillary line C8 in the 5th intercostal space on the left scapular
line
C5
C6
As an alternative, you can connect the 3-lead monitoring
cable, Part No. 223 287 01 or Part No. 223 288 01 (with HF protection).
In this case only leads I,II and III can be displayed.
* Set the energy selector (4) to : This turns on
CardioServ.
227 446 32-D CardioServ V4.1 49
Displaying and Monitoring the ECG
I...III
aV...V
Paddle
1
cm/mV
g
a
b
I
36
160 BPM
140 ECG
HR Alarm
97
- % 90 SpO2
Figure 6-4. Main display
a heart/pulse-rate reading with alarm limits b lead c 1-mV calibration pulse d alarm message, QRS blip e channel 1: ECG f channel 2: plethysmogram g menu
c
The defibrillator beeps and displays a checkered pattern (LCD performance test). Next the software version is displayed, followed by the main display (Figure 6-4).
A field is reserved for technical alarms and messages above the heart and pulse rate reading.
d
f
g
FreezeSpO2ECG
e
The defibrillator selects the following default settings:
EINTOVEN lead I (selectable) AC filter on/ muscle filter on (selectable) sensitivity of 1 cm/mV (selectable)
Next
Menu
If you wish to select another lead or sensitivity, proceed as follows (these are only temporary changes which will not be saved):
* Press the
* Use the
ECG
softkey to call up the ECG submenu.
softkey to select a suitable ECG lead
(shown at b, Figure 6-4).
* Use the
softkey to select the sensitivity.
Filters (AC line filter / muscle filter) enabled during signal acquisition render the ECG display insensitive to signal noise arising from the mains or from muscle tremor or motion. The
Warnin
Filters will falsify the ECG signal. For diagnostic purposes, all filters should be disabled.
filters, however, falsify the ECG signal, making it unsuitable for diagnostic purposes. The heart rate is always calculated from the unfiltered ECG. An artifactual ECG signal may thus lead to a wrong heart rate reading, even though the displayed, filtered ECG is “clean”.
50 CardioServ V4.1 227 446 32-D
Displaying and Monitoring the ECG
Freeze
The Freeze function may mask messages on the
display.
* For a more detailed assessment of the ECG, you can
freeze the trace with the
Freeze
softkey.
The current ECG continues to be displayed in the upper section of the LCD for monitoring, while the frozen segment appears in the bottom section. The monitoring function is still on.
When you press
Freeze
the erase bar moves to the right-hand screen edge, i.e., the entire display shows the period of time immediately prior to depression of
Freeze
.
The frozen ECG segment can be:
Freeze
printed out with the
Print
key (incl. a history of
4 seconds)
Store
stored with the
released with the
Freeze
Freeze
Un-
key
key.
Rotating the display (temporary)
The screen display can be rotated 180°.
Next
* Press
* Press
Menu
to display the submenu.
Display
Flip
to rotate the display.
Pressing the same key again restores the original orientation.
227 446 32-D CardioServ V4.1 51
Displaying and Monitoring the ECG
6.2 Monitoring the Heart Rate
CardioServ can be set up (Section 11 “Configuring
the Defibrillator Settings) to automatically enable the audible alarm on power-up. The default alarm limits can also be preset.
a
b
c
I
36
160 BPM
140 ECG
HR Alarm
d
f
97
- % 90 SpO2
Figure 6-5. Main display
a heart/pulse-rate reading with alarm limits b lead c 1-mV calibration pulse d alarm message, QRS blip e channel 1: ECG f channel 2: plethysmogram g menu
g
FreezeSpO2ECG
With the factory settings unchanged, there will be no audible alarm upon power up (F4 ). You enable the alarm tone by pressing F4 once.
If the device sounds an alarm, you can press to silence the alarm tone. Should the alarm cause persist, the alarm will recur after 120 s (the remaining time is displayed in the softkey). To permanently disable the audible alarm, press the key longer than 2 seconds. As a result the crossed-out bell symbol will appear.
The alarm limits can be modified either permanently via the defaults menu (section 11 The Defaults Menu) or tempo­rarily (see below).
e
Next
Menu
If the heart rate violates one of the set limits for more than 5 seconds CardioServ gives alarm:
the alarm tone sounds (selectable) the alarm cause is shown in the display (d, Figure 6-5) the violated limit value flashes the alarm recorder is started (selectable) the ECG is saved (4-s history, 6 seconds after alarm
release). See section 7 The Memories of CardioServ”.
* Press to silence the alarm.
If the alarm cause has been eliminated, the alarm is cleared; if it persists, the alarm recurs after 120 seconds.
52 CardioServ V4.1 227 446 32-D
Displaying and Monitoring the ECG
When an electrode required for the selected ECG lead drops off, an alarm sounds and the Check Electrode message appears on the display.
Modifying Device Settings
Using keys F1 to F5 you can change the settings for ECG monitoring. In this case, however, the settings are only tem­porary and will not be saved. Permanent adjustments are only possible via the defaults menu (section 11 Configuring the Defibrillator Settings”).
Adjusting Alarm Limits
I...III
aV...V
Paddle
1
cm/mV
Alarm
Limits
Alarm
Printout back
Figure 6-6. Next menu
High Alarm
- + ECG
Low Alarm
- + ECG
Figure 6-7. Alarm menu
You can disable the alarm limits by selecting a value
outside the adjustment range.
back
* Press the
ECG
softkey to display the ECG menu
(Figure 6-6).
Alarm
* Press
Limits
to change the alarm limits. The alarm
menu will be displayed (Figure 6-7).
* The + keys increase the values, the - keys decrease them
in steps of 5 BPM. If you hold the key depressed, the digits change at a higher rate.
227 446 32-D CardioServ V4.1 53
Displaying and Monitoring the ECG
QRSPulse
Beep
OFF
Selecting the HR Source
CardioServ automatically returns to the main menu
when no function keys are pressed for more than 30 s.
HR
Source
ECG
QRSPulse
Beep
OFF
Memory
Display
Flip
Figure 6-8. The submenu
With the C-LOCK ECG synchronization function
enabled, the QRS beep is triggered by the ECG signal even when the selected heart rate source is the pulse signal.
back
CardioServ devices with SpO2 measuring system allow the selection of the heart/pulse-rate source (heart rate from ECG or pulse rate from the SpO2 signal). The source is indicated by ECG or Pulse below the parameter reading.
Next
* Display the submenu with the
Menu
key.
Figure 6-8 will appear.
HR
* Select ECG or SpO2” with the
Source
ECG
key.
Enabling/Disabling the QRS Beep
Next
* Display the submenu (Figure 6-8) with the
Menu
key.
* Enable or disable the QRS or pulse beep with the
key, as required.
54 CardioServ V4.1 227 446 32-D
Displaying and Monitoring the ECG
g
g
6.3 Monitoring Pacemaker Patients
Warnin
Warnin
If several adverse conditions exist at once, the possibility that pacing pulses are interpreted as QRS complexes should be considered. For safety, always watch pacemaker patients closely.
The device does not recognize pacing pulses with an amplitude below 20 mV.
When monitoring pacemaker patients, it is important that the device counts only the QRS complexes and not the pacing pulses from the pacemaker. For this reason, CardioServ comes with an electronic circuit that filters out pacemaker pulses. However, in exceptional cases, depending on the pacemaker model involved and the electrode placement, the compensa­tion algorithm following every pacing pulse may simulate a QRS complex. Every pacemaker must provide an oppositely charged current (reverse current) after delivering a pacing pulse. Ineffective stimulation (absence of QRS complexes) can lead to misinterpretation, so that alarm is not released in cases of bradycardia (slow heart rate) or cardiac arrest.
For this reason we recommend the plethysmogram (SpO2) for monitoring of pacemaker patients. Also, set the HR source to Pulse.
You should disable the C-LOCK synchronization to prevent that frustraneous pacing pulses are counted as QRS com­plexes.
Whether or not the device interprets the pacemaker compen­sation algorithm as a QRS complex depends on the pace­maker pulse parameters (See section 14 Technical Specifica­tions).
For pacemaker patients, the ECG amplitude should be greater than 1 mV.
227 446 32-D CardioServ V4.1 55
Displaying and Monitoring the ECG
6.4 Brief Operating Instructions
Displaying the ECG
* Apply ECG electrodes and connect them to CardioServ.
* Set energy selector to .
* Display the ECG menu (key F1
ECG
) and select ECG
lead (F1).
* Adjust amplitude (F2).
* Freeze and release the ECG, if desired, with F3 (main
menu).
Monitoring the Heart Rate
* Display the ECG as described above.
* Enable alarm tone with F4.
Modifying the Heart-Rate Alarm Limits
* Press F1
ECG
to display the ECG menu.
* Press F3 to display the alarm limits menu.
* Decrease the values with F1 and F3, increase them with
F2 and F4.
* Silence alarm with F4, if alarm cause persists, the alarm
recurs after 120 seconds (or disable the audible alarm permanently by pressing F4 longer than 2 s).
56 CardioServ V4.1 227 446 32-D
The Memories of CardioServ
7. The Memories of CardioServ
This section of the manual informs you
about the memories of CardioServ
how to print out the stored information
how to clear the memories
227 446 32-D CardioServ V4.1 57
HR
Source
ECG
QRSPulse
Beep
OFF
Figure 7-1. Submenu
Print
Text
Print
Event
Figure 7.2. Memory menu
Memory
Print
Trend
Display
Flip
Memory
Clear
The Memories of CardioServ
CardioServ comes with 3 different memories:
a text memory an event memory a trend memory
All information concerning device operation goes into the text memory where it is saved with the time of day (e.g., device
on, device off, alarms, shocks, etc.). CardioServ can save up to 80 such events. When the memory is full, the device updates the information automatically by saving new events and deleting old ones. Moreover, all three memories can be cleared in one action (see below).
To print out the stored information, proceed as follows:
back
* Press
* Press
back
* Press
text memory (Figure 7-3).
Next
Menu
to display the submenu (Figure 7-1).
Memory
to display the memory menu (Figure 7-2).
Print
Text
to print out the information stored in the
DEVICE OFF . . . . . . : 17:15:02 28.06.1997
DEVICE ON . . . . . . : 17:15:08 28.06.1997
HR ALARM 335BPM : 17:15:09 28.06.1997 SHOCK 349JOULES: 17:15:25 28.06.1997 HR ALARM 61BPM: 17:16:20 28.06.1997 HR ALARM 60BPM: 17:17:44 28.06.1997 HR ALARM 61BPM: 17:19:31 28.06.1997 SHOCK 205JOULES: 17:20:12 28.06.1997
DEVICE OFF . . . . . . : 17:38:17 28.06.1997
DEVICE ON . . . . . . : 17:52:09 28.06.1997
Figure 7-3. Printout from text memory
58 CardioServ V4.1 227 446 32-D
The Memories of CardioServ
I
65
160 BPM 40 ECG
Event SHOCK #6 30.03.1998 11:41
Print
all
Figure 7-4. Print Event menu
Select
-
Event
The event memory contains up to 40 ECG strips of 16 sec- onds duration, each with a history of 4 seconds. The event strips are recorded either automatically (upon violation of a heart-rate limit, delivery of the shock and modification of the pacemaker settings, provided the pacemaker is turned on) or
Store
a
device updates the information automatically by saving new
manually with the
Freeze
key. When the memory is full, the
events and deleting old ones. Moreover, all three memories can be cleared in one action (see below). Recordings initiated by a defibrillation shock are blanked for 5 seconds after the
200 J
+
Print
Event
back
4-second strip before the shock. After the blank, the recording continues for another 10 seconds.
To print out the stored events, proceed as follows:
Next
* Press
* Press
* Press
Menu
to display the submenu (Figure 7-1).
Memory
to display the memory menu (Figure 7-2).
Print
Event
to display the print-event menu (Figure
7-4).
Now you can
* select an event with the + and - keys (The event is dis-
played at a in Figure 7-4) and initiate the printout with
Print
Event
or
Print
* print all stored events with
all
(starting with the
event displayed at a, Figure 7-4) or
* press
227 446 32-D CardioServ V4.1 59
back
to return to the memory menu.
The Memories of CardioServ
HR
Source
ECG
QRSPulse
Beep
OFF
Memory
Figure 7-1. Submenu (repeated)
Print
Text
Print
Event
Print
Trend
Figure 7-2. Memory menu (repeated)
45 min
HR
Print
9 h
HR
Print
45 min
SpO2
Print
Figure 7-5. Print trend menu
Display
Flip
Memory
Clear
9 h
SpO2
Print
back
back
back
The CardioServ trend memory contains the HR and SpO readings of the past 45 minutes and 9 hours. This memory, too, is updated automatically. The stored trends can be printed out as follows:
Next
* Press
* Press
* Press
Menu
to display the submenu (Figure 7-1).
Memory
to display the memory menu (Figure 7-2).
Print
Trend
to display the print-trend menu (Figure
7-5).
Now you can
45 min
* print the 45-min HR trend with
* print the 9-hour HR trend with
* print the 45-min SpO2 trend with
Print
9 h HR
Print
HR
45 min
SpO2 Print
2
9 h
* print the 9-hour SpO2 trend with
* press
back
to return to the memory menu.
SpO2
Print
Clearing the Memories
Next
* Press
* Press
* Press
Menu
to display the submenu (Figure 7-1).
Memory
to display the memory menu (Figure 7-2).
Memory
Clear
for 2 seconds to clear all three memories in
one action.
60 CardioServ V4.1 227 446 32-D
Recording
8. Recording
This section describes
how to initiate a manual recording
in which situations the recorder operates automatically
how to load chart paper
227 446 32-D CardioServ V4.1 61
Recording
100
8.1 Manual Recordings
The keys are used to start and stop manual record­ings. When Cont. Printout in the defaults menu is off (Default), CardioServ will record the ECG shown on the display for 16 seconds. The recorder can be stopped before that with . When Cont. Printout is on, CardioServ will record the ECG until stopped with
360
300
150
200
Charge
Shock
APEX
3
Shock
Sync
3
Charge
Shock
2
recorded first (Figure 8-2). The following information is
+
-
annotated in the margin of the recording strip:
a text indicating that this
Figure 8-1. Keys to initiate a recording
b date c time d active filters e delivered energy
a
MANUAL 09. 04 .1997 09 : 16 : 12 50 Hz 35 Hz 0 JOULES HR:
II 1 CM/MV 25 MM/S ST. JOHNS DGH 97
b
c
. The paper speed is 25 mm/s. A calibration pulse is
f heart rate
is a manual recording
g ECG lead h sensitivity i paper speed k hospital/department name
ed
f
PATIENT
DATE OF BIRTH.............
USER........................
COMMENTS.................
SELECTED ENERGY.........
DELIVERED ENERGY........
ALARM LIMIT HR...........
ALARM LIMIT SPO2 .......
PACEMAKER................
DATE/TIME..................
g
h
i
k
Figure 8-2. Manual recording
62 CardioServ V4.1 227 446 32-D
Recording
g
8.2 Automatic Recordings
It is important that the chart paper exit is unob-
Warnin
structed. If this is not ensured, the paper may be pulled back into the device and wrap around the transport roller. When this happens, remove the paper jam as follows:
open the paper compartmentcarefully pull out the tangled paper and tear
it off the strip
– re-load the chart paper (section 8.3).
You can obtain a copy of the displayed image by
simultaneously pressing the two contrast adjust­ment keys.
A recording is initiated automatically with each delivered defibrillation shock. Via the defaults menu (section 11) you can select automatic recordings to be triggered by each viola­tion of an alarm limit. An automatic recording covers a period of 16 seconds, including a history of 4 seconds. The following information is annotated in the margin of the recording strip (Figure 8-2):
a Message indicating the reason for the recording b Date c Time d Active filters e Delivered energy f Heart rate g ECG lead h Sensitivity i Paper speed k Hospital/department name
227 446 32-D CardioServ V4.1 63
Figure 8-3. Opening the paper compartment
Recording
8.3 Loading Chart Paper
To prevent damage to the printhead use the original HELLIGE CONTRAST® chart paper only (Part No. 226 130 02).
* Push the cover of the paper compartment upward and
fold it out (Figure 8-3).
* Remove the empty sleeve of the old paper roll by pulling
on the white plastic tab.
* Insert the new roll as shown in Figure 8-4.
Figure 8-4. Inserting the paper roll
64 CardioServ V4.1 227 446 32-D
Figure 8-5. Inserting the leading edge
Recording
* Feed the leading edge of the paper under the paper
transport roller (Figure 8-5) and keep pushing it forward until it appears above the roller.
* Feed the leading edge of the paper through the exit in
the cover as shown in Figure 8-6.
Figure 8-6. Feeding the leading edge through the aperture
227 446 32-D CardioServ V4.1 65
Figure 8-7. Closing the cover
Thermorecordings should only be stored in transpar-
ent envelopes made of polyethylene, since PVC bleaches the text and traces (if in doubt, insert a sheet of tissue paper in between).
Recording
* Close the cover. While doing so, pull firmly on the paper
strip to ensure that it does not get caught behind the cover (Figure 8-7).
The last 3 meters of the roll are marked with a red stripe. Insert a new roll in time to ensure that all alarm recordings are documented.
66 CardioServ V4.1 227 446 32-D
Oxygen Saturation SpO
2
9. Oxygen Saturation SpO
2
This section of the manual explains
the oxygen saturation measuring method
how to apply the sensors
points to note during SpO2 measurement
227 446 32-D CardioServ V4.1 67
Oxygen Saturation SpO
9.1 General Information
2
II
60
160 BPM
140 ECG
SpO2 Alarm
89
100 % 90 SpO2
Next
MenuFreezeSpO2ECG
Figure 9-1. Display of the SpO2 reading and the plethysmogram
Pulse oximetry is not suitable for oxygen monitor­ing in fetuses before or during birth. Moreover, it
Caution
is not suitable for use on patients with carbon­monoxide poisoning.
SpO2 measurement is employed to determine the oxygen saturation of hemoglobin in the arterial blood. Alarm limits can be adjusted to monitor levels of oxygen saturation. A plethysmographic waveform is displayed in channel 2 (Figure 9-1).
Arterial oxygen saturation is measured by a method called pulse oximetry. It is a continuous, non-invasive method based on the different absorption spectra of reduced hemoglobin and oxyhemoglobin.
Therefore the sensors consist of a light source (two LEDs in most cases) and a photodetector on the opposite side which collects the incident light. The light (red and infrared range between 660 nm and 940 nm) from the LEDs is transmitted through the blood and tissue components of the finger, ear lobe or foot. The photodiode in the sensor measures the light that passes through and converts it into an electrical signal. The pulsatile component of the signal is used to build the plethysmogram.
Elevated levels of CO-Hb and Met-Hb can influence the SpO2 readings. Also dyes in the
Caution
blood (e.g. Cardiogreen) can impair the measur­ing accuracy of the system.
Literature
WUKITSCH, M.W.; PETTERSON, M.T.; TOBLER, D.R.;POLOGE, J.A.: Pulse Oximetry: Analysis of Theory, Technology, and Practice, J. Clin. Monit. 4: 290– 301 (1988)
CECIL, W.T.; THORPE, K.J.; FIBUCH E.E.; TUOHY, G.F.: A Clinical Evaluation of the Accuracy of the Nellcor N-100 and Ohmeda 3700 Pulse Oximeters, J. Clin. Monit. 4: 31–36 (1988)
68 CardioServ V4.1 227 446 32-D
Oxygen Saturation SpO
2
C-LOCK ECG Synchronization
The C-LOCK ECG synchronization feature enables the monitor to use an ECG signal as a reference point for identi­fying the pulse and synchronizing saturation measurements. This enhances the performance of the monitor in the presence of patient movement and when the patients perfusion is poor.
When an ECG signal is present during SpO2 measurement, the monitor is receiving two separate signals that reflect cardiac activity: an optical signal from the sensor and an electrical signal from the ECG. The time that elapses between the ECG R-wave and the optical pulse detected at the sensor site depends on the patients physiology, the heart rate and the location of the sensor. However, for a given patient the length of the delay is relatively stable. Through C-LOCK ECG synchronization, the monitor uses that time relationship to identify good pulses and reject nonsynchronized artifacts.
Application Hints
Use only the sensors listed in section 15 “Order Informa-
tion and Accessories. Apply the sensors as described in their instructions for use. Carefully observe all informa­tion and cautions given in these instructions.
Take care that the sensor does not exert too much pres-
sure on the tissue, as this would result in wrong readings and blistering. The blisters are not caused by overheating but by lack of ventilation.
Exercise extreme care to assure continued circulation
distal to the sensor site after application.
Change the application site at least every 24 hours to
allow the skin to breathe.
Excessive ambient light impairs signal quality. This can
be prevented by covering the sensor site with a cloth.
Simultaneously determining the cardiac output by means
of dye dilution may prevent SpO2 measurement.
When circulation of blood is impaired (blood-pressure
cuff or extreme vascular resistance), it may not be possi­ble to determine SpO2 values or the pulse rate.
Remove nail polish and artificial fingernails before
applying the sensor, as these may affect the reading.
Do not apply the finger sensor to the same arm as a
blood-pressure cuff.
227 446 32-D CardioServ V4.1 69
Oxygen Saturation SpO
2
In order to minimize motion artifact:
take care to provide an ECG of good quality (C-LOCK
ECG synchronization)
use a new sensor with fresh adhesive backing
move the sensor to a less active site
select a slow integration time.
When monitoring SpO2 during electrosurgical intervention, take care that:
CardioServ is powered from the built-in battery or from
a different power circuit than the electrosurgical unit
the ground pad is close to the surgical site
the sensor is applied as far from the surgical site, the
ground pad and the electrosurgical unit as possible.
In the presence of AC line interference
When interference signals from the power line are
present, square waves may be displayed instead of the plethysmogram. In this situation we recommend to disconnect the device from the power line and operate it on battery power.
70 CardioServ V4.1 227 446 32-D
Oxygen Saturation SpO
2
The ECG signal (section 6.1 Displaying the ECG”)
is a prerequisite for the proper functioning of the C-LOCK ECG synchronization feature.
Option
SpO
EKG
2
Figure 9-2. SpO2 sensor connection
II
60
160 BPM
140 ECG
9.2 Measuring and Monitoring Oxygen Saturation
* Turn on CardioServ (switch position ).
* Apply the sensor as described in the instructions for use
enclosed with the sensor.
* Connect the sensor to CardioServ (SpO2 connection,
Figure 9-2).
Within a few seconds the SpO2 reading and the plethysmo­gram will be displayed (Figure 9-3).
Please note: The following adjustments are temporary and, contrary to the settings of the default menu (Section 11 Configuring the Defibrillator Settings), will not be saved.
a
b
97
--- % 90 SpO2
c
I...III
aV...V
Paddle
1
cm/mV
Figure 9-3. SpO2 value and plethysmogram
a SpO2 value b plethysmogram c SpO2 alarm limits (upper limit disabled)
227 446 32-D CardioServ V4.1 71
Alarm
Limits
back
Oxygen Saturation SpO
Enabling/Disabling the C-LOCK ECG Synchronization
2
Disable the C-LOCK ECG synchronization func-
tion for monitoring of pacemaker patients (section
6.3)
C-Lock
ON
Integ.
Time
12 s
Alarm
Limits
Figure 9-4. SpO2 menu
If, in monitoring the patient, several adverse conditions exist at once, a disturbed signal may go
Caution
unnoticed. Artifacts could then be capable of simulating a plausible reading, and no alarm would be released. To assure reliable monitoring, the sensor application and signal quality should be verified from time to time.
When monitoring the pulse rate derived from the SpO2 signal instead of the heart rate, select an
Caution
integration time of 4 s or 8 s. Do not select 12 s.
back
With the factory settings unchanged, the C-LOCK ECG synchronization feature is inactive when the CardioServ is switched on. If an ECG signal is available, enable the synchro­nization feature as follows:
* Press the
SpO2
softkey to display the SpO2 menu (Figure 9-4) (only available when SpO2 sensor is con­nected to the unit).
C-Lock
* Press the
OFF
softkey to enable (ON) the C-LOCK feature (or disable the feature by pressing the key again).
Selecting the Integration Time
The integration time is the time over which the SpO2 readings are averaged. Selectable times are 4, 8 (default) and 12 sec­onds. The integration time of 12 seconds should only be selected in exceptional cases.
* Press the
SpO2
softkey to display the SpO2 menu (Figure 9-4).
Integ.
* Press
Time
to select an integration time.
12 s
72 CardioServ V4.1 227 446 32-D
Oxygen Saturation SpO
Adjusting SpO2 Limits
2
C-Lock
ON
Integ.
Time
12 s
Alarm
Limits
Figure 9-4. SpO2 menu (repeated)
High
-
Alarm
SpO2 SpO2
+
Low
-
Alarm
+
Figure 9-5. Alarm limits menu
SpO2 alarms are similar to HR alarms. Please refer
to section 6.2 Monitoring the ECG”.
MANUAL 29.06.1997 09:16:12 50 Hz 35 H
back
back
* Press
* Press
SpO2
to display the SpO2 menu (Figure 9-4).
Alarm
Limits
to display the alarm limits menu (Figure
9-5).
* The + key increases the limit value, the - key decreases
it.
SpO2 alarms neither initiate an alarm recording nor are they saved to the event memory.
Printing the Plethysmogram
The recordings, no matter whether they are initiated manually or automatically, always display the plethysmorgram in chan­nel 2.
II 1 CM/MV 25 MM/S CARDIOSERV PACEMAKER SPO2
Figure 9-6. ECG and plethysmogram
227 446 32-D CardioServ V4.1 73
For your notes
Oxygen Saturation SpO
2
74 CardioServ V4.1 227 446 32-D
Pacing
10. Pacing
This section of the manual explains
the pacemaker uses
the points to note during pacing
pacemaker operation
227 446 32-D CardioServ V4.1 75
10.1 General Information
Pacing
Application and Functional Description
The transcutaneous pacemaker of CardioServ is applied in emergencies for external (transcutaneous) cardiac stimulation. It is also used as a temporary aid in cases of acute arrhythmias or Stokes-Adams attacks. Some forms of bradycardia and tachycardia can be treated as well.
The pacemaker offers two modes of operation: demand and fixed-rate pacing.
The pacing pulses are delivered through the defibrillation pads.
Rules for Application of External Pacemakers
These rules are valid for all pacemakers, regardless of model and manufacturer.
All electrical devices that deliver energy to patients in any form or that have an electrically conductive connection to the patient present a possible hazard.
The safe application of the device lies in the hands of the user; thus, it is very important to observe the following rules:
* Pacemakers may only be used under the supervision of
qualified medical staff.
* The safe application necessitates expert knowledge, good
organization, special care in selecting the technical equipment and regular maintenance.
* Medical-technical devices such as the CardioServ must
only be applied by persons who are adequately trained in the use of such equipment.
* Before application, the user must check the unit for
functional safety.
* The patients ECG must be monitored to allow the user
to determine capture. Furthermore, at least one of the persons present must be familiar with the application of the defibrillator.
76 CardioServ V4.1 227 446 32-D
Pacing
Due to their functional requirements pacemakers operate with high voltages and are thus equipped
Caution
with special non-accessible outputs. Nevertheless, it is important not to come into contact with the voltage-conducting contacts via conductive metal objects, such as tweezers, as long as the pacemaker is operating. Currents passing through the heart which exceed 10 µA may induce ventricular fibrillation.
Switch the pacemaker on and off as follows:
Caution
Switching the pacemaker on:
1. Apply the pace pads
2. Connect pads to CardioServ via the adapter lead
3. Switch on CardioServ
4. Switch on the pacemaker
Switching the pacemaker off:
1. Switch off the pacemaker
2. Switch off CardioServ
3. Disconnect adapter lead from CardioServ
4. Remove pace pads
Each of the following rules must be observed:
* Check the performance of the defibrillator at hand.
* When positioning the patient ensure that no electrically
conductive connections are created between the patient and earthed metal parts (also beware of puddles, etc.). Although the pacemaker pulse current output is re­quired to be electrically isolated (floating), this is an additional safety precaution to ensure that the pacemaker current pulse only flows between the pacemaker elec­trodes.
* Apply electrodes, arrange electrode leads and connect
them to CardioServ. Use only the electrodes and leads listed in section 15 Order Information and Accessories.
* Operate the pacemaker as described in this manual.
* The pacemaker is switched off after defibrillation (pacer
rate setting is unchanged, current is reset to 0 mA).
Determine capture by measuring the pulse rate, not the heart rate.
Caution
227 446 32-D CardioServ V4.1 77
Pacing
10.2 Application of Single-Use Defibrillation Pads
STERNUM
electrode + connector
electrode + connector
Figure 10-1. Defibrillation pad placement
223 383 01
The single-use defibrillation pads (part no. 919 202 94 adult pads, part no. 919 202 95 pediatric pads) can be used for
defibrillation/cardioversion (refer to section 4.2 Defi-
brillation with Internal Electrodes or Defibrillation Pads”)
ECG monitoring
transcutaneous pacing.
APEX
Below we will explain the correct pad application for transcutaneous pacing
* Use pads before their expiration date.
* Apply the pads as follows:
Shave each site. This improves conductivity and makes
removal of the pad easier.
Place the pads on the patient so that the connectors
A
point to either side of the patient and that the cables are not hindering patient treatment.
919 202 94 919 202 95 (K)
Figure 10-2. Connecting the pad to the cable
The electrodes are pregelled; therefore do not use addi-
tional contact cream or paste.
Do not use pads, if the gel has dried out.
Peel off the backing from each pad and place the pad
carefully on the appropriate site.
78 CardioServ V4.1 227 446 32-D
g
223 346 01
919 201 89 919 202 75 (K)
Figure 10-3. Connecting the defib pads to the adapter cable
Pacing
* Then press the connector of cable 223 383 01 on to the
electrode contact pin until you hear it click into place. Observe the connector labels: “A” =apex, “S” = sternum.
The round adhesive electrodes (part. no. 919 201 89) can be used with adapter lead 223 346 01. When connecting the electrode to the lead, take care that they engage properly. To disconnect them, simply press on the rear part of the catch (Figure 10-3).
* After use, carefully peel off the pads from the patient’s
skin (Figure 10-4) and discard them immediately.
Should a patient require defibrillation during transcutaneous pacing, immediately select the
Caution
required energy with the energy selector, and push
Charge
the
button. The pacemaker automatically
Shock
selects a current of 0 mA in this situation. The shock is delivered by pressing the described in section 4.2). Should the patient require further pacing, resume pacing by following the instructions in sections
10.3 (Demand Mode) or 10.4 (Fixed-Rate Mode).
incorrect
Figure 10.4. Removing defibrillation pads
Shock
buttons (as
correct
Discard disposable defibrillation pads immedi-
Warnin
ately after use. Do not reuse them!
If the ECG signal is extremely noisy and the AC line filter is not sufficient to provide a clean signal, the demand mode operation of the CardioServ may be disturbed. In this situation the message Check pace pads will be displayed. To reduce the interference signal, reapply the pads, carefully observing the application instructions.
227 446 32-D CardioServ V4.1 79
Pacing
g
g
10.3 Demand Mode
Warnin
Warnin
Observance of the pacemaker application rules stated in section 10.1 is an absolute must to ensure the safe and successful use of the pace­maker.
During pacing, always set the energy selector to the position. This prevents that a defibrilla­tion pulse is triggered inadvertently.
Caution: The pacing pulses are delivered via the defibrillation pads. These must be applied to the patient as explained in section 10.2.
The pacemaker can be switched on only when the pace pads are connected (otherwise a message will be displayed to in­form the user of missing electrodes).
In the demand mode the pacemaker does not deliver a pacing pulse as long as the patients intrinsic heart rate exceeds the set pacer rate. When the heart rate drops below the pacer rate, the pacemaker starts delivering pacing pulses. For this reasons the ECG must be continuously monitored. The necessary syn­chronization pulses are automatically transmitted to the pacemaker.
Demand mode is the method of choice when bradycardia or asystole is expected to develop after critical events. Control­ling the pacemaker in this manner precludes the possible competition between intrinsic excitation and external pacing pulses which could result in ventricular fibrillation.
* Acquire the ECG signal as described in section 6.1
Displaying the ECG.
* Verify that the pads are properly applied and connected
to CardioServ.
80 CardioServ V4.1 227 446 32-D
Defib
n
P
a
c
e
m
a
k
e
r
S
t
i
m
u
+
P/min
P/min
mA
mA
Dem
Fix
STERNUM
18 17 16 15
Figure 10-5. Pacemaker operating controls
15 Buttons for selection of the pacer output 16 Buttons for selection of the pacer rate 17 Pacing mode selection button (Fix/Demand) 18 toggles the pacemaker on and off (press button
longer than 2 seconds to turn the pacemaker off)
The default pacer rate can be preset (Defaults
menu).
Pacing
* Press the button to switch on the pacemaker (Fig-
l
a
t
e
u
r
+
+
-
Sy
* Using the
ure 10-5). The pacemaker defaults to the demand mode and to a pacer rate of 60 BPM (selectable).
+
mA
button, select a low pacer output (e.g.
20 mA)
+
* Increase the pacer rate with the
P/min
button until the
asterisk (a, Figure 10-6) just begins to flash: The pacer rate is now identical with or just above the intrinsic heart rate.
+
* Press the
mA
button to slowly increase the pacer output to a level which ensures consistent responses from the heart.
+
* Now select the required pacer rate with buttons
P/min
and
.
-
P/min
* Increase the pacer output by another 5 mA to achieve
reliable stimulation.
* To verify the success of the treatment, watch the ECG
on the screen. You can press the
Sync
button to view the
trigger marks on the display.
227 446 32-D CardioServ V4.1 81
160 BPM
140 ECG
Pacing
* At the end of the intervention, first turn off the pace-
0
maker, then remove the pads carefully.
In order to turn off the pacemaker, button must be held down for at least 2 seconds. This safety precaution has been taken to prevent the pacemaker being turned off inadvertently. For this reason it is not possible to switch off CardioServ while the pacemaker is still on.
a
Pacemaker Demand 60 BPM 65 mA
*
b
c
d
Figure 10-6. Screen display with pacemaker switched on
a asterisk, flashing with every delivered pacing
pulse b pacing rate c pacing current d pacing mode
If the defibrillator is put into operation during
pacing, the pacemaker automatically selects an output of 0 mA
Next
MenuFreezeSpO2ECG
82 CardioServ V4.1 227 446 32-D
Pacing
g
Defib
n
g
10.4 Fixed-Rate Mode
Observance of the pacemaker application rules
Warnin
stated in section 10.1 is an absolute must to ensure the safe and successful use of the pace­maker.
P
a
c
e
m
a
k
e
r
S
t
i
m
u
l
+
P/min
P/min
mA
mA
Dem
Fix
STERNUM
18 17 16 15
Figure 10-7. Pacemaker operating controls
15 Buttons for selection of the pacer output 16 Buttons for selection of the pacer rate 17 Pacing mode selection button (Fix/Demand) 18 toggles the pacemaker on and off (press button
longer than 2 seconds to turn the pacemaker off)
Caution: The pacing pulses are delivered via the defibrillation pads. These must be applied to the patient as explained in section 10.2.
The pacemaker can be turned on only when the pace pads are connected (otherwise a message will be displayed to warn the
Sy
a
t
e
u
r
+
+
-
user of missing electrodes).
In the fixed-rate mode the pacemaker delivers pacing pulses at a selectable rate and output setting. The selected rate is fixed, i.e., it does not take into account intrinsic action of the heart. This mode of operation should be selected in cases of asystole. Some forms of bradycardia and tachycardia can be corrected with fixed-rate pacing, but transcutaneous emer­gency pacing is not the therapy of choice in these instances.
* Verify that the electrodes are properly applied and
connected to CardioServ.
* Press the button, to switch on the pacemaker
(Figure 10-7). The pacemaker defaults to the demand mode and to a pacing rate of 60 BPM (selectable).
During pacing, always set the energy selector to
Warnin
the , position. This prevents that a defibril­lation pulse is triggered inadvertently.
227 446 32-D CardioServ V4.1 83
Pacing
0
160 BPM
140 ECG
a
Pacemaker Fix 60 BPM
65 mA
d
Figure 10-8. Screen display with pacemaker switched on
a asterisk, flashing with every delivered pacing
b pacing rate c pacing current d pacing mode
The default pacer rate can be preset (Defaults
menu).
pulse
*
b
c
Next
MenuFreezeSpO2ECG
* Press the
Dem
button for 3 seconds to select the fixed-
Fix
rate mode (“Demand” is replaced with “Fix” at d in Figure 10-8).
The asterisk (a; Figure 10-8) flashes each time a pacing pulse is delivered.
* Select the pacing rate with the buttons
+
* Press the
mA
button to slowly increase the pacer output
+
P/min
and
P/min
-
to a level which ensures consistent reponses from the heart.
* To verify the success of the treatment, watch the ECG
on the screen.
* Increase the current by another 5 mA to ensure reliable
pacing.
* At the end of the intervention, first turn off the pace-
maker, then remove the pads carefully.
In order to turn off the pacemaker, button must be held down for at least 2 seconds. This safety precaution has been taken to prevent the pacemaker being turned off inadvertently. For this reason it is not possible to switch off CardioServ while the pacemaker is still on.
.
84 CardioServ V4.1 227 446 32-D
Configuring the Defibrillator Settings
11. Configuring the Defibrillator Settings
This section describes the instrument settings that can be modified to suit your personal needs and preferences, and how this is done.
227 446 32-D CardioServ V4.1 85
Configuring the Defibrillator Settings
Defaults
HR Limits Sensitivity Lead
Autosequence SPO2 Limits C-LOCK SPO2-Int. Time Lead Fail Alarm QRS Beep Alarm Tone Alarm Printout Shock Printout Cont. Printout
Defaults
Cont. Printout Pacemaker Display Display Volume AC Filter Muscle Filter Date Format Date/Time Language
User Factory Setting
Figure 11-1. The menu
low : 40 high : 160
0.5 1 2 cm/mV
I II III aVR aVL aVF V
Paddle 200J 200J 360J low : 40 high : OFF
OFF ON
4s 8s 12s
OFF ON OFF ON OFF ON OFF ON
OFF ON
OFF ON
Next
EXIT
OFF ON
60 BPM
normal reverse 0 deg. 180 deg. high low 50Hz 60Hz OFF
OFF ON DD:MM:YYYY MM:DD:YYYY
06.01.1998 11:35:47
deutsch English français
усск espanol italiano Svenska Portugues
OFF ON
Menu
Next MenuEXIT
The defaults menu allows you to customize device settings, which means that they are retained in memory and are acti­vated automatically on power up.
* Simultaneously press F1 and F5 to display the defaults
menu.
Page 1 of the menu appears (Figure 11-1 top).
The cursor keys F1, F2 (up/down) and F4 (right) are used to move the cursor to the parameter whose setting is to be modified. Then you can change the setting with function keys F1 through F5. The functions and labels of these keys change with each parameter.
Next
With F5
Confirm each selection with F5
Menu
you scroll to the next page of the menu.
EXIT
. This will also take
you back to the main menu.
Defaults
HR Limits
Sensitivity Lead
Autosequence SPO2 Limits C-LOCK SPO2-Int. Time Lead Fail Alarm QRS Beep Alarm Tone Alarm Printout Shock Printout Cont. Printout
low : 40 high : 160
0.5 1 2 cm/mV
I II III aVR aVL aVF V
Paddle 200J 200J 360J low : 40 high : OFF
OFF ON
4s 8s 12s
OFF ON OFF ON OFF ON OFF ON
OFF ON
OFF ON
+
F1 F2 F3 F4 F5
Enter
Figure 11-2. The configuration menu
86 CardioServ V4.1 227 446 32-D
Configuring the Defibrillator Settings
g
With F3
EXIT
you leave the configuration menu and save
the changes.
The new settings are saved only if you quit the
defaults menu with F3
EXIT
.
ECG monitoring is suspended during configura-
Warnin
tion.
It is not possible to call up the defaults menu, while the defibrillator is charged, during ECG analysis
Analyse
(
button) or while the pacemaker operates.
When you display the defaults menu, the values preset in this menu will overwrite the temporary
Caution
device settings (alarm limits, leads, etc.).
HR Alarm Limits
The limit values for heart-rate monitoring selected here are automatically activated on power up.
* Using F4, move the cursor to the right.
* Adjust the low limit with F1 and F2 (if you keep the
keys depressed, the digits change at a higher rate) (Figure 11-2). Alarm limits can be disabled by selecting a value outside the adjustment range.
* Using F4, move the cursor to the high limit and adjust it
in the same way as the low limit.
* Confirm the new limit values with F5.
Sensitivity
The setting selected here (.5,1,2 cm/mV) is the CardioServ default sensitivity.
* Using F4, move the cursor to the right.
* Select the desired setting with F3 or F4 and confirm the
selection with F5.
227 446 32-D CardioServ V4.1 87
Configuring the Defibrillator Settings
Lead
The setting selected here is the CardioServ default ECG lead.
* Using F4, move the cursor to the right.
* Select the desired lead with F3 or F4 and confirm the
selection with F5.
Autosequence
This menu item defines the energy levels for the 1st, 2nd and 3rd defibrillation shock in autosequence mode (default: 200 J, 200 J, 360 J).
You can choose a different energy level (150 J, 200 J, 300 J, 360 J) for each shock.
When the cursor highlights a value, this value can be in­creased with F1 and decreased with F2.
SpO2 Int. Time
Here you select the default integration time for averaging of the SpO2 values. The 12 second integration time should only be selected in exceptional cases.
Lead Fail Alarm
Do you wish the alarm to sound after 30 s when an ECG electrode drops off?
QRS Beep
Do you wish to have the QRS beep enabled or disabled on power up?
Alarm Tone
Do you wish to have the audible alarm enabled or disabled on power up?
SpO2 Limits
Here you adjust the default alarm limits for monitoring of SpO2.
C-LOCK
Do you wish to have the C-LOCK ECG Synchronization feature enabled or disabled on power up?
88 CardioServ V4.1 227 446 32-D
Alarm Printout
Do you want the recorder to start automatically in an alarm situation (violation of HR limits)?
Shock Printout
Do you want the recorder to start automatically when a defibrillation shock is delivered?
Configuring the Defibrillator Settings
Continuous Printout
When the function is disabled, the recorder will stop auto­matically after 16 seconds. When it is enabled, the recorder must be stopped manually.
Pacemaker
This is to select the default pacing rate.
Display
Normal (black on white) or reverse (white on black) display on the LCD screen.
Normal or flipped display (rotating the display 180° may prove useful for CardioServ units operated in the defibrillator mounting system).
Volume
For selection of the volume of all audio signals emitted by the unit (prompts, alarms).
AC Filter
for elimination of signal noise from the power line. Default setting for 50 Hz (Europe) or 60 Hz (USA) or no filter (OFF).
Muscle Filter
For elimination of motion artifact and muscle action potentials (35 Hz).
Date Format
DD.MM.YYYY (European format) or MM.DD.YYYY (US format).
Date / Time
For adjustment of date and time.
* Using F4, move the cursor to the right (day).
* Using F1 and F2, adjust the day.
* Using F4, move the cursor to the right (month).
* Using F1 and F2, adjust the month.
* Proceed in the same manner to adjust year, hours,
minutes and seconds.
* Confirm the entries with F5.
Depending on the CardioServ model purchased, not
all menu items may be available.
227 446 32-D CardioServ V4.1 89
Configuring the Defibrillator Settings
Language
This menu item is used to select the language for the screen texts and printed documents.
* Using F4, move the cursor to the right.
* Select the desired language with F3 or F4.
* Confirm your selection with F5.
User
Via this menu item you can enter a text or a name which will be printed in the margin of the recording strip.
* Using F4, move the cursor to the right.
You can now use F1 and F2 to select the first numeral, sym­bol or letter from a character set for the name or text to be entered. With F1 you scroll forward through the set, with F2, backward. F4 moves the cursor to the right, allowing you to enter the second character, etc. The first character is a blank.
Factory Settings
You can restore the factory settings by selecting ON (language and operating mode will not be changed).
Restoring the factory settings will delete the informa-
tion entered with menu item “User”.
* Confirm entries with F5.
90 CardioServ V4.1 227 446 32-D
Error Indications and Messages
12. Error Indications and Messages
In this section of the manual you will find
the error indications and messages which CardioServ
may display after the power-on self-test
the error indications and messages which CardioServ
may display while in use
an explanation of the meaning of each message, and
troubleshooting tips
227 446 32-D CardioServ V4.1 91
Error Indications and Messages
Error Messages during power-on self-test (POST)
egasseMtceffEnoitanalpxEydemeR
rorrEMARehtesutonod,tcefedecived
rorrEMORehtesutonod,tcefedecived
rorrEMARyalpsiD
yekynasserp
yromemdexifnirorrE
yekynasserp
rorrEesaBemiT
yekynasserp
yek”kcohSkcehC
yekynasserp
eciffoecivresyfitoN
ecived
eciffoecivresyfitoN
ecived
seicnegremerofesudetcirtser
elbissopnoitallirbifed,ylno
llatadaer
seicnegremerofesudetcirtser
elbissopnoitallirbifed,ylno
)devasebtonnac
seicnegremerofesudetcirtser
elbissopnoitallirbifed,ylno
suoenorre
(snottubrehtie 2 (ro/dna) 8)
,purewopnodesserpederew
nehw;evitcefed)s(nottubro
ecived,evitcefednottub
rofdesuebtonnac
roftubsnoitallirbifed
siyalpsideht,tcefedyalpsid
ebtonnacrodaerottluciffid
(syek 7 ecivresyfitondna)
noitcnufehtfoenosserP
eciffo
tcefednoitarugifnocresu
snoitarugifnocmotsuc,.g.e(
(syek 7 ecivresyfitondna)
noitcnufehtfoenosserP
eciffo
sahcus(ataddetaler-emitlla
ebyam)etartraeheht
(syek 7 rorrefi;emittsujda,)
noitcnufehtfoenosserP
eciffoecivresyfiton,stsisrep
noitcnufehtfoenosserP
(syek 7 ,ecivedffohctiws,)
(syeksserptonod 2 (dna) 8)
.niaganoecivedhctiwsdna
,sruceregassemrorrefI
.eciffoecivresyfiton
ylnognirotinom
yekEGRAHCkcehC
)evobayek”kcohSees()evobayek”kcohSees()evobayek”kcohSees(
yekynasserp
)evobayek”kcohSees(
rorrEygrenEegrahC
yekynasserp
morfsreffidygrenederots
niecivedesu,ygrenedetceles
!ylnoseicnegreme
devas
ehtemittsalderruccororre
tierofeb(desusawecived
(syek 7 -sidtsetreviled,)
sawdna)ffodenrutsaw
noitcnufehtfoenosserP
dnaffoecivedhctiws,egrahc
egassemrorrefI.niagano
.eciffoecivresyfiton,srucer
92 CardioServ V4.1 227 446 32-D
Messages/problems during operation
egasseMnoitanalpxEydemeR
yrettabwoLdetelpedtsomlayrettabsniamotecivedtcennoc
Error Indications and Messages
mralaRHro,timilRHwolrohgihfonoitaloiv
edortcelekcehCelddaproedortceleGCEhgih
degrahcbifeDs03rofderotsygrenenoitallirbifed
rotceleskcehC;ylreporpgnixednitonsirotcelesygrene
hgihygrenEneve,seluoj05sdeecxeygrenedetceles
wolygrenE;ygrenedetceleswolebygrenederots
edortcelEgnirotinomedortceleroelddapbifed
kcehc,cilotsysatonsitneitapnehw
s4rofdeifitnedievaw-Ron
ehtnehwdaelrehtonatceles;sedortcele
llamsootsiedutilpmalangis
euqinhcetedortcelekcehc
ecnadepmi
;noitisoptcaxeehtotrotcelesygrenetes
elbissoptonnoitallirbifed
ecivresyfiton,sruceregassemnehw
eciffo
,)selddaplanretni(eulavrewolatceles
;detcennoceraselddaplanretnihguoht
detcelesnahtrehgihygrenederots
eciffoecivresyfiton,srucer
egassemnehw;egrahcsidtsetesaeler
elbissopnoitallirbifed;ygrene
egassemnehw;egrahcsidtsetesaeler
elbissopnoitallirbifed
eciffoecivresyfiton,srucer
ecived(niaganodnaffoecivedhctiws
ecivedhtiwselddapfoegnahcxe,tcefed
elbissoptonnoitallirbifed,nodenrut
elddaprofffodehctiwsebtsum
yfiton,sruceregassemnehw;)egnahcxe
eciffoecivres
yltcerrocnoitcnuftonseodeciveDecnabrutsiderawtfosniaganodnaffoecivedhctiws
sdapecapkcehCecnerefretnienilCAevissecxe,retlifenilCAehtelbane
noitacilppaedortcelekcehc
gnirudetanimullitonseodDELneerG
tinuylppusrewopevitcefedecivresyfiton
noitareporewopenil
gnirudetanimullitonseodDELwolleY
tinuylppusrewoproyrettabevitcefedecivresyfiton
noitareporewopenil
227 446 32-D CardioServ V4.1 93
For your notes
Error Indications and Messages
94 CardioServ V4.1 227 446 32-D
Cleaning and Maintenance
13. Cleaning and Maintenance
This section of the manual describes
how to clean and disinfect the device and the electrodes
how to sterilize the internal electrodes
how to clean and disinfect the suction electrodes
how to perform routine maintenance on the device
227 446 32-D CardioServ V4.1 95
Cleaning and Maintenance
13.1 Cleaning, Disinfection and
Sterilization
Before cleaning the device, disconnect it from the power line. Turn off the device before touching
Danger
the contact susfaces of the paddles, and be careful not to switch it on again while cleaning it. Dan­ger to life! As a safety precaution, remove the battery and disconnect the paddles from the defibrillator.
Do not use hot air to sterilize the internal defi­brillation electrodes.
Caution
The defibrillation electrodes and handles must be disconnected from the cable before autoclaving
Caution
(arrow, Figure 13-2).
Device and Paddles
* Discard all disposable electrodes immediately after use to
prevent inadvertent reuse.
* The paddles and their leads can be cleaned and disin-
fected by wiping them down with a gauze pad moistened in a cleaning solution. Before applying the paddles again, check that they have thoroughly dried.
* Clean the instrument surface with a cloth moistened in a
cleaning solution. Take care that the solution does not enter the defibrillator enclosure.
Any hospital-grade cleaning solution and disinfectant contain­ing up to 70% alcohol is suitable.
* The internal defibrillation electrodes are initially cleaned
in the same way as external paddles. For electrodes and connection cables, low-temperature plasma sterilization is the recommended method. Alternative methods are ETO sterilization, water vapor (autoclave at 134 °C) or ionizing radiation. Internal electrodes must be sterilized after each resuscitation code.
96 CardioServ V4.1 227 446 32-D
Cleaning and Maintenance
1 2 3
Figure 13-1. Inserting the contact paddle
Having loosened the counter nut, you can easily alter
the position of the contact paddle.
To insert the contact paddles, proceed as follows:
* Screw the counter nut (2, Figure 13-1) onto the elec-
trode as far as it will go.
* Screw the contact paddle (1) into the handle as far as it
will go, then bring it into the appropriate position.
* Now fix the contact paddle by screwing the counter nut
(2) tight against the handle (3).
External Counter Electrode for Internal Defibrillation
* Disconnect the electrode from its lead before cleaning or
sterilizing it (Figure 13-2).
* Clean the electrode by rubbing it down with a cloth
moistened in soap water. Use a disinfectant for disinfec­tion. Do not immerse the electrode in the liquid.
* Low-temperature plasma sterilization is the recom-
mended sterilization method. Alternative methods are ETO sterilization and ionizing radiation. (Please note:
Figure 13-2. Counter electrode for internal defibrillation
227 446 32-D CardioServ V4.1 97
Frequently sterilizing the electrodes with ethylene oxide reduces the life of the plastic material!) Do not autoclave the electrodes!
Cleaning and Maintenance
Exchanging the Defibrillation Electrode Cable
* Switch off CardioServ.
* Grasp the plug and remove it from the socket (do not
pull on the cable).
* When inserting the plug, observe the orientation
(beveled edges) and click it into place.
Cleaning and Disinfecting Suction Electrodes
Do not use pointed metal objects to remove solid particles of dirt from the electrodes as this would
Caution
destroy the silver/silver-chloride layer. Use a commercially available fiberglass eraser instead.
* Clean the electrodes with water and a detergent. Use a
small brush to remove grime.
* All commercially available cleaning agents used for
surgical instruments are suitable for cleaning of the electrodes. Follow the manufacturers instructions to mix the preparation. Do not use metal dishes and take care not to immerse plugs and metal sockets in the solution.
* To disinfect the electrodes, wipe them down with a cloth
moistened in 70% alcohol.
98 CardioServ V4.1 227 446 32-D
13.2 Maintenance
Cleaning and Maintenance
Before each application
* visually check the device and all accessories (leads, elec-
trodes, etc.) for signs of damage
* test the device performance (refer to section 3).
If you detect damages or impaired functions so that the safety of the patient and user are no longer guaranteed, CardioServ must be repaired before it can be applied again.
Checks at regular intervals
CardioServ defibrillators are emergency medical devices
designed to save and preserve life; they must be ready for use at all times. Operational readiness must also be ensured for battery power operation. Therefore, the devices must be subjected to the following checks at regular intervals:
Every month
If batteries are repeatedly partially discharged, the resulting memory effect may dramatically reduce the battery capacity. This effect can be efficiently minimized by regular condition­ing. If the capacity of a relatively new battery is drastically reduced, the battery may be reconditioned by repeated charg­ing and discharging. Proper maintenance of NiCd batteries is essential and considerably promotes their proper performance. Routine preventive maintenance should be carried out by qualified service technicians on a regular basis (recommended interval: 30 days).
It is the users decision whether or not to recondition the battery at regular intervals. Batteries which are not recondi­tioned have a shorter service life and will have to be replaced more frequently.
We recommend our Accu Service Unit (Figure 3-5)
for optimal care of the batteries. It prolongs the batteries service life and guarantees their opera­tional readiness at all times.
* visually inspect the device and accessories
* test the device performance as described in section 3.
Battery Maintenance
Rechargeable batteries require special maintenance and con­tinued checks to assure they function in emergency situations. It is normal for batteries of this type to self-discharge when not in use.
227 446 32-D CardioServ V4.1 99
If you decide not to use the Accu Service Unit, the battery can be checked and reconditioned as described below. With this method, however, the continued operational readiness of CardioServ on battery power is not guaranteed.
1. Disconnect CardioServ from the power line and dis-
charge fully charged battery in the monitoring mode. To do so, set energy selector switch to (SpO2 sensor not connected) and wait until device switches off.
Cleaning and Maintenance
2. Check how long it takes before battery is depleted. If the time is less than 1.8 hours, the battery is too old or improperly maintained and should be replaced.
3. Recharge the battery. This will take 16 hours.
Technical Inspections
For safety, the devices require regular maintenance. To ensure functional and operational safety of the CardioServ Technical Inspections should be carried out annually.
The following checks can be carried out within the framework of a service contract. Otherwise it should be assured that the person inspecting the device is adequately trained and experi­enced.
* The device and accessories should be visually inspected
for signs of mechanical damage which may impair the device functions.
* The electrode leads of defibrillators and pacemakers
must be carefully checked for signs of mechanical dam­age, short-circuits and breaks.
* All batteries that are 3 years or older must be replaced.
* The warning system of the defibrillator must be checked.
Testing the Pacemaker Performance
The performance of the pacemaker can be tested with a commercially available pacemaker tester (e.g. CS300 Simula­tor from Marquette Hellige, part no. 417 983-001).
* Devices which are not in perfect working order or the
use of which is unsafe must be immediately repaired or labelled accordingly to prevent their use.
The device does not require any other maintenace.
* All safety-related labels and instructions printed on the
device must be inspected for legibility.
* A performance test as described in section 3 is to be
performed. The hardware and software functions are to be tested by means of the power-on self-test. All seg­ments of the LCD must be visible.
* On the defibrillator the energy delivered into a 50-ohm
resistance must be measured.
100 CardioServ V4.1 227 446 32-D
Loading...