Leica VT1200 User Manual

Page 1
Instructions for Use
VT1200 / VT1200 S
Vibrating-blade microtome
Leica VT1200/VT1200 S V 1.4 RevC, English - 09/2013 Order No. 14 0481 80101 RevC
Always keep this manual with the instrument. Read carefully before working with the instrument.
Page 2
Page 3
Note
The information, numerical data, notes and value judgments contained in this manual represent the current state of scientific knowledge and state-of-the-art technology as we understand it following thorough investigation in this field. We are under no obligation to update the present manual periodically and on an ongoing basis ac­cording to the latest technical developments, nor to provide our customers with additional copies, updates etc. of this manual. To the extent permitted in accordance with the national legal system as applicable in each individual case, we shall not be held liable for erroneous statements, drawings, technical illus­trations etc. contained in this manual. In particu­lar, no liability whatsoever is accepted for any financial loss or consequential damage caused by or related to compliance with statements or other information in this manual. Statements, drawings, illustrations and other information as regards contents or technical details of the present Instructions for Use are not to be considered as warranted characteristics of our products.
These are determined only by the contract provisions agreed between ourselves and our customers. Leica reserves the right to change technical specifications, as well as manufacturing pro­cesses, without prior notice. Only in this way is it possible to continuously improve the technol­ogy and manufacturing techniques used in our products. This document is protected under copyright laws. Any copyrights of this document are retained by Leica Biosystems Nussloch GmbH. Any reproduction of text and illustrations (or of any parts thereof) by means of print, photo­copy, microfiche, web cam or other methods – including any electronic systems and media – requires express prior permission in writing by Leica Biosystems Nussloch GmbH. For the instrument serial number and year of manufacture, please refer to the nameplate on the back of the instrument.
© Leica Biosystems Nussloch GmbH
Leica Biosystems Nussloch GmbH Heidelberger Str. 17-19
D-69226 Nussloch Germany Phone: +49 62 24 143-0 Fax: +49 62 24 143-268 Internet: http://www.LeicaBiosystems.com
Assembly contracted by Leica Microsystems Ltd. Shanghai
Leica VT1200 / VT1200 S
Page 4
Contents
1 Important Information ................................................................................................................4
1.1 Symbols in the text and their meanings .................................................................................... 4
1.2 Qualification of personnel ........................................................................................................... 5
1.3 Intended use/improper use ......................................................................................................... 5
2. Safety ............................................................................................................................................6
2.1 General safety notes .................................................................................................................... 6
2.2 Warnings ........................................................................................................................................ 6
2.3 Transport, unpacking and setting up......................................................................................... 7
3. Instrument characteristics ........................................................................................................8
3.1 Technical data for the VT1200 ................................................................................................... 8
3.1.1 Technical data for the VT1200 S ................................................................................................ 9
3.2 General overview – VT1200/VT1200 S ..................................................................................... 10
3.3 VT1200 Control Panel ................................................................................................................. 12
3.3.1 VT1200 S control panel .............................................................................................................. 13
4. Installation .................................................................................................................................14
4.1 Standard scope of delivery for the VT1200............................................................................. 14
4.1.1 Standard scope of delivery for the VT1200 S ......................................................................... 15
4.2 Packing and setting up the instrument ................................................................................... 16
4.3 Before commissioning the instrument .................................................................................... 17
5. Working with the Instrument ..................................................................................................18
5.1 Description of the typical application ..................................................................................... 18
5.2 Control elements on the VT1200 control panel ...................................................................... 19
5.2.1 Moving the specimen receptacle vertically........................................................................... 19
5.2.2 Fine-adjusting the blade-to-specimen proximity and feeding to the section thickness ..... 20
5.2.3 Moving the blade ........................................................................................................................ 20
5.2.4 Selecting the sectioning parameters ...................................................................................... 21
5.3 Control elements on the VT1200 S control panel .................................................................. 22
5.4 Installing the accessories ......................................................................................................... 32
5.4.1 Installing the ice tray and buffer tray ...................................................................................... 32
5.4.2 Installing the double-walled buffer tray.................................................................................. 33
5.4.3 Preparing a specimen ................................................................................................................ 33
5.4.4 Installing and removing the blade holder ............................................................................... 34
5.4.5 Inserting the blade ...................................................................................................................... 35
5.4.6 Adjusting the clearance angle ................................................................................................. 35
5.5 Routine daily maintenance and switching off the instrument – VT1200/VT1200 S .......... 36
Instructions for Use V 1.4 RevC – 09/2013
Page 5
Contents
6. Operating the VibroCheck ...................................................................................................... 37
6.1 Using the VibroCheck with the VT1200 ................................................................................... 37
6.2 Using the VibroCheck with the VT1200 S ................................................................................ 39
7. Malfunctions: Meaning and Troubleshooting .................................................................... 41
7.1 Error messages and troubleshooting ...................................................................................... 41
7.2 Replacing the main fuse ............................................................................................................ 46
8. Cleaning and Maintenance .................................................................................................... 47
8.1 Cleaning the instrument............................................................................................................. 47
9. Ordering Information for Optional Accessories, Consumables and Spare Parts .........48
9.1 Extension accessories for standard specimens .................................................................. 49
9.1.1 Buffer trays .................................................................................................................................. 49
9.1.2 Specimen plate ........................................................................................................................... 50
9.1.3 VibroCheck ................................................................................................................................... 51
9.1.4 Blades ........................................................................................................................................... 52
9.1.5 Microscope, assembly ............................................................................................................... 52
9.1.6 Magnifier, assembly ................................................................................................................... 52
9.1.7 Installing the magnifier support or microscope support .................................................... 53
9.1.8 Fiber-optic illumination, cold light source .............................................................................. 54
9.1.9 Cyanoacrylate adhesive ............................................................................................................ 55
9.1.10 Foot switch .................................................................................................................................. 55
9.1.11 Julabo FL300 – recirculating cooler/chiller ............................................................................ 55
10. Warranty and Service ..............................................................................................................56
11. Decontamination Certificate (Master) ..................................................................................58
Leica VT1200 / VT1200 S
Page 6

1 Important Information

1.1 Symbols in the text and their meanings

Warnings appear in a gray box and are marked by
a warning triangle .
Useful tips, i.e. important user information, appear in a gray box and are marked by an .
(5) (Fig. 5)
Numbers in parentheses refer to item numbers in illustrations or to the illustra­tions themselves.
Manufacturer
Date of Manufacture
The CE labeling shows that the product corresponds to one or more applicable European directives.
Observe the Instructions for Use
Order No.
Serial number
Environmental protection symbol of the China RoHS directive. The number in the symbol indicates the "Environment-friendly Use Period" of the product in years. The symbol is used if a substance restricted in China is used in excess of the maximum permitted limit.
The CSA test mark means that a product has been tested and fulfills the applicable safety and/or per­formance standards, including the relevant standards defined or ad­ministered by the American National Standards Institute (ANSI), Under­writers Laboratories (UL), the Cana­dian Standards Association (CSA), the National Sanitation Foundation International (NSF) and others.
Symbol for labeling electrical and electronic equipment in accordance with Section 7 of the German Elec­trical and Electronic Equipment Act (ElektroG). ElektroG is the law on the bringing into circulation, return and environ­mentally compatible disposal of elec­trical and electronic equipment.
Instrument model:
All information provided in these Instructions for Use applies only to the instrument type indicated on the title page. A nameplate is attached to the rear side of the instrument. The series and REF num­bers are attached to a separate label (Fig. 1a) on the right side of the instrument. Fig. 1 and 1a are provided as an example only and show a valid nameplate for this instrument.
Instructions for Use V 1.4 RevC – 09/2013
Page 7
Fig. 1a

1.2 Qualification of personnel

The Leica VT1200 and the VT1200 S may be operated by trained laboratory personnel only.
All laboratory personnel designated to operate this instrument must read these Instructions for Use carefully and must be familiar with all technical features of the instrument before attempting to operate it.
1. Important Information
Fig. 1

1.3 Intended use/improper use

The Leica VT1200 and VT1200 S are used for sectioning in the fields of medi­cine, biology and industry, and are especially designed for sectioning fixed or unfixed fresh tissue in buffer.
The instrument must be used exclusively according to the instructions con­tained in these Instructions for Use.
Any other use of the instrument is considered improper.
Leica VT1200 / VT1200 S
The VT1200/VT1200 S may be used for research purposes only. Sec­tions made using the VT1200/VT1200 S must NOT be used for diag­nostics.
Page 8

2. Safety

These Instructions for Use includes important information related to the operating safety and maintenance of the instrument.
The Operating Manual is an important part of the product, and must be read carefully prior to startup and use and must always be kept near the instrument.
If additional requirements on accident prevention and environmental pro­tection apply in the country of operation, these Instructions for Use must be supplemented by appropriate instructions to ensure compliance with such requirements.
Make sure to read all of these Instructions for Use before you work on or operate the instrument.

2.1 General safety notes

These instruments have been built and tested in accordance with the safety regulations for electrical measuring, control, regulating and laboratory devices.
In order to maintain this condition and to ensure safe operation, the user must follow the instructions and warnings contained in this operating man­ual.

2.2 Warnings

The current EC Declarations of Conformity can be found on the Internet:
www.LeicaBiosystems.com
The safety devices installed in this instrument by the manufacturer only constitute the basis for accident prevention. Operating the instrument safely is, above all, the responsibility of the owner, as well as the designated per­sonnel who operate, service or clean the instrument.
To ensure trouble-free operation of the instrument, make sure to comply with the following instructions and warnings.
Instructions for Use V 1.4 RevC – 09/2013
Page 9
• Extremely sharp blades pose risk of injury when touched!
• Fresh tissue poses risk of infection!
• Fire hazard from uncovered magnifier! Cover the magnifier during work breaks!
Proper handling
Always be exceptionally careful when handling the blades! Do not leave open blades lying around after removal. Always make sure to handle the blade in a way that will not cause you injury. All appropriate safety precautions must be met to avoid the risk of infection. Wearing safety gloves, a mask and safety goggles—in accordance with the "Working with Sub-
stances that Pose a Health Risk" guidelines—is absolutely mandatory. The instrument may be opened by authorized service personnel only. Always disconnect the power plug before opening the instrument. Always switch off the instrument using the power switch and disconnect the power plug before
replacing the fuse. The use of fuses other than those installed at the factory is not permitted.

2.3 Transport, unpacking and setting up

2. Safety
The instrument must be set up so that the main power switch on its right side (item 7 in Fig. 14) is easily accessible at any time.
Because the weight of the instrument is approx. 56 kg, carrying the instrument requires 2 persons (1 carrying handle per person).
Leica VT1200 / VT1200 S
• Whenunpackingtheinstrument,comparethepartsreceivedwiththe
parts ordered. If the parts received do not match your order, contact the sales company responsible for your order immediately.
• Beforeconnectingtothepowersupplysystem,pleaseobserve"Techni­cal Data"!
• Neverconnecttheinstrumenttoapowersocketthatdoesnothavea
protective conductor terminal.
Page 10

3. Instrument characteristics

3.1 Technical data for the VT1200

General data:
Sectioning frequency (±10 %) ........................................................................................................... 85 Hz (± 10 %)
Amplitude.................................................................................................... from 0 - 3 mm, in 0.05 mm increments
Sectioning speed (± 10 %) ................................................................................................................ 0.01 - 1.5 mm/s
Return speed (± 10 %) ..................................................................................................................................2.5 mm/s
Total vertical specimen stroke .................................................................................................20 mm (motorized)
Cutting range ..............................................................................................................................45 mm (adjustable)
Maximum specimen size:
With standard blade holder ............................................................................................................. 33 x 50 mm
Specimen orientation, rotating .................................................................................................................... 360°
Specimen plate, swiveling ....................................................................................................................... 0 - 10°
Section thickness adjustment .................................................................................. manual, in 1 µm increments
Ambient conditions:
Operating temperature range: ........................................................................................... min. 10 °C - max. 35 °C
Relative humidity: ....................................................................................................................................... max. 60 %
Storage temperature: ................................................................................................................................... 5 - 55 °C
Storage humidity: ............................................................................................................................................. < 60 %
Height: ........................................................................................................................ up to 2000 m above sea level
Electrical data:
Rated voltage range (± 10 %): ....................................................................................................................... 100 V - 240 V
Nominal frequency (±10 %): .................................................................................................................................. 50/60 Hz
Power consumption: ...................................................................................................................................................35 VA
Protective class: ...................................................................................................................................................................I
Power fuse: ...................................................................................................................................................... T 1 A L 250 V
Pollution degree: ..................................................................................................................................................................2
Overvoltage category: ....................................................................................................................................................... II
Electrical overload protection: ..................................................................................................................................... Yes
Internal current limit for the electronics: ................................................................................................................... Yes
Dimensions:
L x W x H: ............................................................................................................................... 600 mm x 250 mm x 230 mm
Height with magnifier support ............................................................................................ 600 mm x 250 mm x 320 mm
Height with microscope: ..................................................................................................... 600 mm x 250 mm x 490 mm
L x W x H control unit (when the bases are folded in): ................................................... 165 mm x 120 mm x 72 mm
Weight:
Without magnifier support and control unit .............................................................................................................56 kg
VT1200 control unit .........................................................................................................................................................1 kg
Magnifier support ...........................................................................................................................................................2 kg
Microscope support with stereomicroscope ............................................................................................................4 kg
Instructions for Use V 1.4 RevC – 09/2013
Page 11
3. Instrument Characteristics

3.1.1 Technical data for the VT1200 S

General data:
Sectioning frequency (± 10 %): ...................................................................................................................85 Hz (± 10 %)
Amplitude: ....................................................................................................... from 0 - 3 mm, in increments of 0.05 mm
Sectioning speed (± 10 %): ........................................................................................................................0.01 - 1.5 mm/s
Return speed (± 10 %): .................................................................................... 1.0 - 5 mm/s, in increments of 0.5 mm/s
Total vertical specimen stroke: ......................................................................................................... 20 mm (motorized)
Sectioning range: ..................................................................................................................................................... 45 mm
Sectioning window: .................................................................................................................................. 0.5 mm - 45 mm
Specimen retraction: .............................................................................. 0 - 100 µm (adjustable; can be deactivated)
Maximum specimen size:
with standard blade holder: ..................................................................................................................... 33 x 50 mm
Specimen orientation, rotating: ........................................................................................................................... 360°
Specimen plate, swiveling: ...............................................................................................................................0 - 10°
Sectioning thickness setting: ............................................manual in 1 µm increments or automatic max. 1000 µm
Ambient conditions:
Operating temperature range: .....................................................................................................min. 10 °C - max. 35 °C
Relative humidity: ............................................................................................................................................... max. 60 %
Storage temperature: .............................................................................................................................................5 - 55 °C
Storage humidity: .......................................................................................................................................................< 60 %
Height: .................................................................................................................................up to 2000 m above sea level
Electrical data:
Rated voltage range (± 10 %): ...................................................................................................................... 100 V - 240 V
Nominal frequency (± 10 %): ................................................................................................................................ 50/60 Hz
Power consumption: ..................................................................................................................................................35 VA
Protective class: ..................................................................................................................................................................I
Power fuse: ..................................................................................................................................................... T 1 A L 250 V
Pollution degree: .................................................................................................................................................................2
Overvoltage category: ...................................................................................................................................................... II
Electrical overload protection: .................................................................................................................................... Yes
Internal current limit for the electronics: .................................................................................................................. Yes
Dimensions:
L x W x H: .............................................................................................................................. 600 mm x 250 mm x 230 mm
Height with magnifier: ......................................................................................................... 600 mm x 250 mm x 320 mm
Height with microscope: .................................................................................................... 600 mm x 250 mm x 490 mm
Control unit (when the bases are folded in): ..................................................................... 190 mm x 150 mm x 72 mm
Weight:
Without magnifier support and control unit: ...........................................................................................................56 kg
VT1200 S control unit: ...................................................................................................................................................1 kg
Magnifier support: .........................................................................................................................................................2 kg
Microscope support with stereomicroscope: ..........................................................................................................4 kg
Leica VT1200 / VT1200 S
Page 12
3. Instrument characteristics

3.2 General overview – VT1200/VT1200 S

Fig. 3, Magnifier
Fiber optic
light guide
Cold light source
Fig. 4, Microscope
Fig. 5
Cutting head
Fig. 2, Basic instrument
10
Fig. 6, Foot switch
Blade holder
Fig. 8, Control panelm VT1200 Smm
Fig. 7, VT1200 control panel
Instructions for Use V 1.4 RevC – 09/2013
Page 13
Attachments on dovetail receptacle
3. Instrument Characteristics
Ice tray
Buffer tray
Buffer tray, plastic
Buffer tray, metal
Specimen plate
for specimens 1 c m in height
for specimens 2 cm in height
Double-walled buffer tray
Julabo FL300 (recirculating cooler/chiller)
Specimen plate
for specimens 1 c m in height
for specimens 2 cm in height
VibroCheck
Fig. 9
Blades for the blade holder
Sapphire blade
Directional
Cyanoacrylate adhesive
Leica VT1200 / VT1200 S
Directional
Cyanoacrylate adhesive
11
Page 14
3. Instrument characteristics

3.3 VT1200 Control Panel

LED display for blade feed
rate and amplitude
Setting dial 1, for blade feed
rate and amplitude
LED display for section
thickness and
section thickness totalizing
Setting dial 2, for section
thickness and moving the
specimens vertically
Selection of the step size,
1 µm,10 µm or
Quickly moves the speci-
men upwards (press the
button until the desired
Quickly moves the specimen
downwards (press once to
adjust to the lowest position)
(for approx. 3 sec.) to save a
frequently used feed value
(pressing it more than once
feeds by the saved section
thickness multiple times).
100 µm steps
position is reached)
Press this button
Moves the blade to-
wards the specimen
Caution: Keep it pressed down until the desired
position is reached. When the end position is reached, the LED lights up.
LED for feed rate
LED on, for enabled ampli-
1
2
Moves the blade away from the specimen
tude adjustment
Toggles between SPEED (blade feed rate) and AMPL (amplitude)
Section thickness Sets the current specimen
position between 0 (lower) and 20,000 µm (upper) or - after zero, totals the section thicknesses
Pressing this button toggles between µm and µm (ac­tively illuminated).
Resets the totalized section thickness display µm to zero.
Immediately starts or stops the sectioning process
ON/OFF lamp
Fig. 10
12
The Leica VT1200 is a semiautomatic microtome with a vibrating blade. Before each cut, a man­ual feed to the desired section thickness must be carried out using the setting dial for section thickness. The VT1200 does not include an automatic specimen retraction; however, retraction can be performed manually.
Instructions for Use V 1.4 RevC – 09/2013
Page 15

3.3.1 VT1200 S control panel

Sets the blade feed rate.
Select the values using set-
ting dial 1.
Setting dial 1, for blade
feed rate (SPEED),
AMPLitude or section
thickness for automatic
mode (AUTO FEED)
Sets the amplitude.
Select the values using
setting dial 1.
3. Instrument Characteristics
Possible in automatic sectioning mode only. Select the desired sec­tion thickness for the automatic feed using setting dial 1.
ON/OFF lamp
Setting dial 2, for section
thickness and moving the
specimens vertically
Selection of the step size,
1 µm, 10 µm or 100 µm steps
First cutting window edge,
LED illuminates when the
window edge is set
Moves the specimen up-
wards (as long as the button
is pressed). When the up-
permost position is reached,
the LED lights up.
Second cutting window
edge, LED illuminates when
the window edge is set
Fig. 11
Toggles between automatic (AUTO) and
semiautomatic (MAN) sectioning mode
The Leica VT1200 S is a fully automatic microtome with vibrating blade. It can be operated in automatic as well as semiautomatic sectioning mode. In semiautomatic sectioning mode, a manual feed to the desired section thickness must be carried out before each cut. There is no automatic specimen retraction in this mode; however, retraction can be performed manually. In automatic mode, an automatic feed (AUTO FEED) to the selected section thickness is carried out before each cut, and the specimen is lowered to the desired retraction value after each cut to prevent the specimen surface and the blade from coming into contact while the blade is being retracted.
Moves the speci-
men into the lowest
position, (LED is
illuminated when the
lowest position is
reached)
Moves the
blade towards
the specimen Caution: Keep it pressed down until the desired posi­tion is reached. When the end position is reached, the LED lights up.
Moves the blade away from the specimen
Toggles in and out of the menu. Memory for 8 param­eter sets.
Sets the totalized section thickness display åµm to "0".
Interrupts the sectioning process in automatic mode. Pressing the button again reactivates the sectioning process.
Starts or stops the section­ing process. In semiautomatic mode, the sectioning process stops immediately; in automatic mode, the sectioning process is finished completely.
Toggling between single stroke (SINGLE) and continu­ous stroke (CONT) in automatic mode is possible. In semiauto­matic mode, only single stroke (SINGLE) is possible.
Leica VT1200 / VT1200 S
13
Page 16

4. Installation

4.1 Standard scope of delivery for the VT1200

VT1200 basic instrument ............................................................... 14 0481 42065
1 control panel ................................................................................ 14 0481 43395
1 toolset:
- 1 Allen key, size 3.0 ................................................................. 14 0194 04764
- 1 Allen key, size 6.0 ................................................................. 14 0222 04141
- 1 cryo-manipulator .................................................................. 14 0462 28930
- 1 replacement fuse T 1 A ....................................................... 14 6943 01000
1 set of power cables:
- 1 power cable "D" .................................................................... 14 0411 13558
- 1 power cable "USA-C-J" ....................................................... 14 0411 13559
- 1 power cable "UK" ST/BU F-5A ............................................ 14 0411 27822
1 dust cover (basic instrument), small ......................................... 14 0212 43742
1 bottle of cyanoacrylate adhesive, contents 10 gr. .................. 14 0371 27414
1 Instructions for Use for Leica VT1200/VT1200 S...................... 14 0481 80101
1 language CD .................................................................................. 14 0481 80200
14
VT1200 configuration ....................................................................... 14 912000001
The above scope of delivery, plus:
Ice tray, assembly ........................................................................... 14 0481 42010
Buffer tray (metal), assembly ........................................................ 14 0481 42084
When ordering additional accessories, compare the parts received with the parts ordered. If the parts received do not match your order, contact the sales company responsible for your order immediately.
Instructions for Use V 1.4 RevC – 09/2013
Page 17

4.1.1 Standard scope of delivery for the VT1200 S

VT1200 S basic instrument ............................................................. 14 0481 42066
1 control panel ................................................................................ 14 0481 43396
1 toolset:
- 1 Allen key, size 3.0 ................................................................. 14 0194 04764
- 1 Allen key, size 6.0 ................................................................. 14 0222 04141
- 1 cryo-manipulator .................................................................. 14 0462 28930
- 1 replacement fuse T 1A ........................................................ 14 6943 01000
1 set of power cables:
- 1 power cable "D" .................................................................... 14 0411 13558
- 1 power cable "USA-C-J" ....................................................... 14 0411 13559
- 1 power cable "UK" ST/BU F-5A ............................................ 14 0411 27822
1 dust cover (basic instrument), small ......................................... 14 0212 43742
1 bottle of cyanoacrylate adhesive, contents 10 gr. .................. 14 0371 27414
1 Instructions for Use for Leica VT1200/VT1200 S...................... 14 0481 80101
1 language CD .................................................................................. 14 0481 80200
4. Installation
Leica VT1200 / VT1200 S
VT1200 S configuration ...................................................................14 91200S001
The above scope of delivery, plus:
Ice tray, assembly ........................................................................... 14 0481 42010
Buffer tray (metal), assembly ........................................................ 14 0481 42084
When ordering additional accessories, compare the parts received with the parts ordered. If the parts received do not match your order, contact the sales company responsible for your order immediately.
15
Page 18
4. Installation

4.2 Packing and setting up the instrument

Ensure that the instrument is standing on a work surface that is as free of vibrations as possible.
Before each transport, the handles must be screwed onto the instrument using the provided screws (see Fig. 13). Check to ensure that the handles are firmly in place and will hold reliably!
Packing the instrument
1. Have 2 people grab the instrument by the transport handles (1), place it
1
2
1
2
3
on the wooden pallet and screw it to the pallet using the 4 size 6 Allen screws (2).
2. Pull the transparent protective hood (3) over the instrument. Place the wooden box (4) on the baseplate. Insert the inner carton ring (5).
3. Insert the accessories box (6 – contains accessories) and fill it with packing material (7).
4. Attach the cover (8) and screw it into place using 8 Phillips screws (9).
5
16
4
6
7
8
9
Instructions for Use V 1.4 RevC – 09/2013
9
Fig. 12
Page 19

4.3 Before commissioning the instrument

1
1
Fig. 13
2
7 (AN)
4
4. Installation
Once the instrument is in its final location, un­screw the transport handles (1) from the instru­ment using the size 6 Allen key provided and store them, along with the screws, in a safe place.
1. Set the power switch (7) on the right side of the instrument to OFF ().
2. Make sure the power cable is correctly con­nected in the power socket (2) on the right side of the instrument.
3. Connect the control panel to the socket (3).
4. Install the ice tray (4).
5. Install the buffer tray (5).
6. Install any optional. accessories, such as the
magnifier, microscope, foot switch, etc. ac­cording to p. 49.
7. Use the power switch (7) to switch on the instrument.
3
Fig. 14
Leica VT1200 / VT1200 S
To make it easier to insert the speci­men, adjust the specimen receptacle to its lowest position, and adjust the blade
5
6
holder to its rearmost position during the first reference run.
17
Page 20

5. Working with the Instrument

5.1 Description of the typical application

The Leica VT1200 / VT1200 S is a microtome with a vibrating blade and is predominately used for sectioning fixed and unfixed specimens during neurological research.
• Tosimplifytheprocessofinsertingthespeci­men, you can quickly move the specimen re­ceptacle to the lowest position by pressing the DOWN button.
• To preparehigh-quality sections,especially
for unfixed tissue, Leica suggests using the optional VibroCheck measuring instrument to determine the height amplitude of the blade after each blade change, and then to minimize it using the setting screw on the blade holder.
To perform the aforementioned process, install
the VibroCheck instrument (see p. 37 for the
VT1200 or p. 39 for the VT1200 S), install the
blade and adjust the clearance angle to the desired position. Then take the measurement and make the appropriate adjustments to the placement of the blade holder.
Remove the VibroCheck according to the in-
structions, and turn the blade 90° toward the top to ensure that the ice tray and buffer tray can be installed safely.
• Insertthebuffertrayintotheicetrayandcover
it with the lid. Fill the ice tray with crushed ice.
• Removethecoverandfillthebuffertraywith
pre-cooled physiological buffer.
• Pushthe icetray and buffertray onto the
dovetail guide and clamp them down.
• Usecyanoacrylate adhesive toadhere the
specimen to the specimen plate and use a manipulator to place it in the buffer tray. Insert
the hose for gassing the buffer into the hose clamp.
• UsetheUP button to lift the specimen to the sectioning level of the blade quickly. Fine­adjusting the blade-to-specimen proximity can be performed using the desired step size 1, 10 or 100 µm.
• Advancethebladetowardthespecimenusing
the "Blade forward" key.
• Aftersectioningthespecimen,usethesetting
dial to feed to the desired section thickness using the selected step size (1, 10 or 100 µm). This raises the specimen receptacle to the desired value.
• Startthesectioningprocessbypressingthe
RUN/STOP key. The process can be stopped again by pressing this key after the cut has been completed. Use the "Blade back" key to move the blade in front of the specimen. Select the section thickness for the next cut and start the sectioning process again.
• Aftercompletingthe sectioning process,
remove the blade, dispose of the specimen receptacle by pushing the DOWN key into its lowest position, and unclamp, empty and clean out the ice tray and buffer tray.
18
Instructions for Use V 1.4 RevC – 09/2013
Page 21

5.2 Control elements on the VT1200 control panel

Caution: To practice the key functions, always make sure there are no blades installed! Do not install the blades until you are familiar with all of the key functions.
When the instrument is switched on, to make it easier to insert a specimen, the specimen receptacle automatically moves into the lowest position (the LED in the DOWN key lights up) and the blade holder moves into the rear-
Switch on the instrument. 1 = on 0 = off
most position (the LED in "Blade away" from the specimen lights up). The parameters last used before switching off the instrument: The blade
feed rate (SPEED), selected amplitude (AMPL) and section thickness saved by the MEMO key are called up again.
The LED illumination is switched on automatically. You can switch it off using the ON/OFF key (lamp symbol).

5.2.1 Moving the specimen receptacle vertically

When the DOWN key is pressed, the specimen receptacle automatically moves quickly into the lowest position. (The LED in the DOWN key lights up when the lower end position is reached.) The µm display is set to "0". If the DOWN key is pressed a second time while the specimen receptacle is moving downwards, the specimen receptacle comes to a stop. The current position appears on the µm display (lowest position = 0). The µm display does not change while the specimen receptacle is in motion.
5. Working with the Instrument
Leica VT1200 / VT1200 S
When the UP key is pressed and held, the specimen receptacle moves quickly upwards into the desired position. After the UP key has been re­leased, the current position of the specimen receptacle appears on the µm display. If the specimen receptacle reaches the upper end position, the LED in the UP key lights up (uppermost position = 20,000 µm). The µm display does not change while the specimen receptacle is in motion.
You can reset the display for the current position of the specimen holder to zero at any point using the CLEAR /µm key. The section thicknesses are then added in the µm display.
19
Page 22
5. Working with the Instrument

5.2.2 Fine-adjusting the blade-to-specimen proximity and feeding to the section thickness

The rotary knob 2 for moving the specimen receptacle vertically can be used both for fine-adjusting the blade-to-specimen proximity and for adjust­ing the desired section thicknesses. The step size: 1, 10 or 100 µm can be selected by pressing the STEP SIZE key. Turning the setting dial clockwise carries out the desired feed; turning the setting dial counterclockwise lowers the specimen (minus sign). The value appears in the lower display when "µm" is selected. After each sectioning process has been completed, the µm display is set to zero.
Save a commonly used section thickness by adjusting the setting dial for the section thickness and pressing the MEMO key for 3 seconds. An audible signal confirms that the value has been accepted. Feed to the saved value by quickly pressing the MEMO key. Pressing the
MEMO multiple times feeds to the value multiple times.
Neither negative values nor a section thickness of over 1000 µm are permitted. If an attempt is made to save impermissible values, an au­dible warning signal sounds three times and the last (permissible) value is retained.

5.2.3 Moving the blade

Blade
forward back
20
The "Blade forward" and "Blade back" keys have to be held down until the desired position is reached. The blade feed rate is 2.5 mm/s. When each end point has been reached, the corresponding LED lights up in the key.
Instructions for Use V 1.4 RevC – 09/2013
Page 23

5.2.4 Selecting the sectioning parameters

SPEED: Blade feed rate – upper display – LED mm/s. The desired blade feed
rate can be adjusted from 0.01 to 1.5 mm/s using rotary knob 1:
0.01 – 0.1 in 0. 01 mm/s increments,
0.10 – 0.5 in 0.02 mm/s increments,
0.50 – 1.5 in 0.10 mm/s increments. AMPL: upper display – LED mm: Display of the amplitude in mm: from 0 - 3 mm in 0.05 mm increments
Starts the sectioning process
Starts the sectioning process using the selected amplitude and blade feed rate. You can start the sectioning process by pressing the RUN/STOP key a second time, or you can stop the process immediately by pressing the "Blade back" or "Blade forward" key. The µm display is then reset to "0". To start a new sectioning process, use the "Blade back" key to move the blade to the beginning of the specimen, set the desired section thickness and restart the sectioning process.
5. Working with the Instrument
Leica VT1200 / VT1200 S
21
Page 24
5. Working with the Instrument

5.3 Control elements on the VT1200 S control panel

The Leica VT1200 S is a fully automatic microtome with a vibrating blade that can be operated in semiautomatic or automatic sectioning mode.
Key / setting dial Semiautomatic sectioning mode Automatic sectioning mode
Switch on the instrument. 1 = on 0 = off
When the instrument is switched on, to make it easier to insert a specimen, the specimen recep­tacle automatically moves into the lowest position (the LED in the DOWN key lights up) and the blade holder moves into the rearmost position (the LED in "Blade away" from the specimen lights up). If semiautomatic sectioning mode was selected before the instrument was last switched off, the following saved param­eters are called up again once it is switched back on:
• Bladefeedrate(SPEED),
• Selectedamplitude(AMPL)
The LED illumination is switched on automatically. You can switch it off using the ON/OFF key.
ditto
• Bladefeedrate(SPEED),
• Selectedamplitude(AMPL)
• Savedsectionthickness(AUTO
FEED)
ditto
22
The LED MAN is active. semiautomatic sectioning mode is enabled.
The LED AUTO is active. The automatic sectioning mode is enabled.
Instructions for Use V 1.4 RevC – 09/2013
Page 25
5. Working with the Instrument
Key / setting dial Semiautomatic sectioning mode Automatic sectioning mode
Toggling from AUTO to MAN
In semiautomatic sectioning mode, before each cut, a manual feed to the desired section thick­ness must be carried out using the setting dial for section thick­ness. There is no automatic specimen retraction in this mode; howev­er, retraction can be performed manually.
The following key functions are disabled in semiautomatic mode:
• Settingcuttingwindowedges
• Selecting the continuous
stroke (CONT)
• Selectingthesectionthickness
for the automatic feed (AUTO FEED)
• PAUSE
not active.
In automatic mode, the selected section thickness feed (AUTO FEED) is carried out along the 1st selected cutting window edge automatically before each cut. To prevent the specimen surface and the blade from coming into con­tact while the blade is being re­tracted, the specimen is lowered by the desired retraction value along the 2nd cutting window edge after each complete cut.
Toggling from MAN to AUTO
Leica VT1200 / VT1200 S
An audible warning signal sounds when these keys are pressed.
The following functions keys be­come enabled again in automatic mode:
• Cuttingwindow edges that
have already been set
• Sectionthickness(AUTO FEED) and continuous stroke (CONT)
23
Page 26
5. Working with the Instrument
Key / setting dial Semiautomatic sectioning mode Automatic sectioning mode
1
The blade feed rate can be ad-
ditto
justed from 0.01 to
1.5 mm/s:
0.01 - 0.1 in 0.01 mm/s increments,
0.10 - 0.5 in 0.02 mm/s increments,
0.50 - 1.5 in 0.10 mm/s increments.
1
Adjusting the amplitude from
ditto
0 to 3 mm in 0.05 mm increments
1
Not possible. Setting the section thickness for
automatic mode - max. 1000 µm.
24
Display for the current speci­men holder position (lowest posi­tion = 0 µm, uppermost position = 20,000 µm.) You can reset the display to "0" at any point by pressing the CLEAR/µm key. The section thicknesses are then added in the µm display.
ditto
Instructions for Use V 1.4 RevC – 09/2013
Page 27
5. Working with the Instrument
Key / setting dial Semiautomatic sectioning mode Automatic sectioning mode
The rotary knob for moving the
ditto
specimen receptacle vertically
2
can be used for fine-adjusting the blade-to-specimen proximity. Turning the setting dial clockwise moves the specimen up into the desired proximity; turning the set­ting dial counterclockwise lowers the specimen (minus sign). The step size: 1, 10 or 100 µm can be selected by pressing the STEP SIZE key. After the setting dial is turned clockwise or counterclockwise, the specimen receptacle position is updated in the µm display.
In semiautomatic mode, the de-
Not possible. sired section thickness is se­lected using the setting dial. The selected section thickness appears in the µm display and the current specimen receptacle po­sition appears in the µm display. After each sectioning process has been completed, the µm dis­play is set to "0".
Leica VT1200 / VT1200 S
When the DOWN key is pressed, the specimen receptacle auto­matically moves quickly into the lowest position. (The LED in the DOWN key lights up when the lower end position is reached.) The µm display is set to "0".
ditto
25
Page 28
5. Working with the Instrument
Key / setting dial Semiautomatic sectioning mode Automatic sectioning mode
If the DOWN key is pressed a second time while the specimen receptacle is moving downwards, the specimen receptacle comes to a stop and the current posi­tion appears on the µm display (lowest position = 0, uppermost position = 20,000 µm). The ∑µm display does not change while the specimen receptacle is in motion.
When the UP key is pressed and held down, the specimen recep­tacle moves quickly upwards into the desired position. After the UP key has been released, the cur­rent position of the specimen re­ceptacle appears on the µm dis­play. If the specimen receptacle reaches the upper end position, the LED in the UP key lights up (uppermost position = 20,000 µm). The µm display does not change while the specimen receptacle is in motion.
ditto
ditto
Blade forward
Blade back
26
The "Blade forward" and "Blade back" keys have to be held down
until the desired position is reached. The blade feed rate can be set in the menu: 1 – 5 mm/s, in increments of 0.5 mm/s. Each time an end point is reached, the corresponding LED lights up in the key.
ditto
Instructions for Use V 1.4 RevC – 09/2013
Page 29
5. Working with the Instrument
Key / setting dial Semiautomatic sectioning mode Automatic sectioning mode
General information about the cutting window
Activated cutting window edges can be deactivated by pressing the corresponding key for approx. three seconds.
Not possible. The horizontal cutting path can be
reduced to the specimen size. The two cutting window edges can be adjusted and changed indepen- dently. Holding the key down for a longer duration (audible warn­ing signal) sets the beginning or end (depending on the key) of the cutting window to the maximum value. Smallest possible section­ing window: 0.5 mm . If a window smaller than 0.5 mm is set or the user reverses the beginning and end, the last entered value is ac­cepted and the previous value is set to the maximum value. The cutting window is not saved when the in­strument is switched off; however, it is retained when you switch from automatic mode (AUTO) to semiau­tomatic mode (MAN).
Leica VT1200 / VT1200 S
Not possible. Advance the blade toward the
specimen using the "Blade for- ward" key. Press the "1st cutting window edge" key until the LED in the key lights up.
Not possible.
Advance the blade to the end of the specimen using the "Blade forward" key and press "2nd cutting window edge" until the LED in the key lights up.
27
Page 30
5. Working with the Instrument
Key / setting dial Semiautomatic sectioning mode Automatic sectioning mode
Only single stroke (SINGLE) is pos­sible. If an attempt is made to switch to continuous stroke (CONT), an audible warning signal sounds.
Starts the sectioning process using the selected amplitude (AMPL) and blade feed rate (SPEED). Pressing the RUN/STOP key a second time stops the sec­tioning process immediately.
The µm display is then reset to "0". The programmed section thick-
Toggles between single (SINGLE) and continuous stroke (CONT). The corresponding LED lights up to indicate the current selection.
Starts the sectioning process us­ing the selected section thickness (AUTO FEED) amplitude (AMPL) and the blade feed rate (SPEED). If single stroke (SINGLE) is select­ed, only one sectioning process is carried out. – If continuous stroke (CONT) is selected, a continuous sectioning process takes place. Pressing the RUN/STOP key a second time stops the sectioning process that is in progress. The blade moves to the 1st cutting window edge and stays there.
ness (AUTO FEED) appears in the µm display continuously.
28
Not possible. A sectioning process in progress
can be interrupted immediately by pressing the PAUSE key and restarted by pressing the PAUSE key again. If PAUSE has been pressed to interrupt a section­ing process, pressing the RUN / STOP key or the "Blade forward" or "Blade back" keys discontinues the sectioning process.
Instructions for Use V 1.4 RevC – 09/2013
Page 31
5. Working with the Instrument
Key / setting dial Semiautomatic sectioning mode Automatic sectioning mode
Press the MENU key
ditto
8 sets of user parameters can be
1
saved; Current user selection – User 1: turn rotary knob 2 clockwise,
2
then press Menu again.
The blade feed rate (SPEED) can
1
be set from 0 to 1.5 mm/s using rotary knob 1.
2
Rotary knob 2
1
The amplitude (AMPL) can be set to 0 - 3 mm using rotary knob 1.
2
1
2
The values can be selected; however, automatic feed is not possible in semiautomatic sec­tioning mode. If the AUTO FEED key is pressed, one feed motion takes place according to the value programmed in automatic mode. If the key is pressed more than once, several feed motions take place.
Leica VT1200 / VT1200 S
Rotary knob 2
The automatic section thickness feed (AUTO FEED) can be ad­justed using predefined step sizes (1, 10 or 100 µm) max. 1000 µm us­ing rotary knob 1.
29
Page 32
5. Working with the Instrument
Key / setting dial Semiautomatic sectioning mode Automatic sectioning mode
Rotary knob 2
1
Mode: Choice between AUTO and MAN with rotary knob 1; for semiautomatic sectioning mode,
2
MAN must be selected.
Rotary knob 2
1
Stroke type (CUT): Only single stroke (SINGLE) can be selected using rotary knob 1. If continuous
2
stroke (CONT) is selected, an au-
Rotary knob 2
Mode: Choice between AUTO and MAN with rotary knob 1; for the automatic sectioning mode, AUTO must be selected.
Rotary knob 2
Stroke type (CUT): Choice be­tween single stroke (SINGLE) and continuous stroke (CONT) with rotary knob 1.
dible warning signal sounds.
1
Specimen retraction (RETRACT) cannot be set.
Rotary knob 2
Rotary knob 2
The specimen retraction (RE­TRACT) can be set from 0 to
100 µm in 10 µm increments using
2
Value cannot be changed. NO automatic retraction is possible in semiautomatic section­ing mode.
rotary knob 1.
30
Rotary knob 2
The LED illumination can be ad­justed to 5 different levels of brightness using rotary knob 1.
ditto
Instructions for Use V 1.4 RevC – 09/2013
Page 33
5. Working with the Instrument
Key / setting dial Semiautomatic sectioning mode Automatic sectioning mode
Rotary knob 2
1
The feed rate (FOR/REV) for the "Blade forward" and the "Blade
ditto
back" keys can be set between 1 and 5 mm/s in 0.5 mm/s incre-
2
1
ments with rotary knob 1.
Rotary knob 2
Switches the vibration movement
ditto
(FOR/VIB) of the "Blade forward" key either on or off using rotary
2
1
knob 1.
Rotary knob 2
Key acknowledgment (BEEP) on
ditto
or off with rotary knob 1.
2
Save the parameter and leave by
ditto pressing the menu MENU key.
If you want to call up the parameters of a certain user (for example, user 3), press the Menu key and select user 3. Then confirm by pressing the Menu key 2x. The parameters stored under user3 are now enabled.
Leica VT1200 / VT1200 S
Leaving the menu and saving the parameter is possible at any point in the menu.
31
Page 34
5. Working with the Instrument

5.4 Installing the accessories

3

5.4.1 Installing the ice tray and buffer tray

• Alever(2), which needs to be pushed forward, is located on the underside of the ice tray (1).
• Nowpushthetrayontothedovetailholder(3) from the front. Clamp it down by pushing the lever (2, Fig. 28) towards the rear.
2
Ice tray Underside
The ice tray can be placed on the stage
1
separately for preparation.
• Pushthebuffertray(5) in as far as it will go (small pins (4) guide it at the side and front).
It is held in place by three strong magnets,
which are integrated into the bottom of the
2
4
Fig. 15
buffer tray.
• Coverthebuffertray(5) with a Plexiglas lid (6).
• Nowfilltheicetraywithcrushedice.
• Coverthe buffer trayand fill itwith cooled
buffer solution.
6
To remove the buffer tray from the ice tray, pull on the rounded edges carefully (7). They help to remove the tray, as they
5
7
Fig. 16
are not magnetic.
32
Instructions for Use V 1.4 RevC – 09/2013
Page 35

5.4.2 Installing the double-walled buffer tray

11
Fig. 17
11
2
5. Working with the Instrument
Clamps for holding the gassing hose for the buf­fer in the proper position can be added to the double-walled buffer tray.
When using the double-walled buffer tray, the recirculating cooler/chiller must be installed PRIOR TO working with specimens.
Connect hoses (2, in the scope of delivery of the double-walled buffer tray) to the empty buffer tray (bottom of Fig. 17). Access is easier if you make the left connection first. To do this, pull back the closure coupling (11), fit the hose until it engages audibly, and then make the right-hand connection.

5.4.3 Preparing a specimen

6
Fig. 18
Leica VT1200 / VT1200 S
• Fixthespecimentothespecimenplateusing
the cyanoacrylate (included in the standard scope of delivery).
• Screwthemanipulator(6) onto the specimen
plate, place it in the buffer tray and adjust it to the desired position.
• Thespecimenplateisheldinplaceinthebuf-
fer tray magnetically.
33
Page 36
5. Working with the Instrument

5.4.4 Installing and removing the blade holder

7
9
8
Always remove the blade BEFORE in­stalling or removing the blade holder!
For quality and service reasons, the blade holder (7) is available as a com­plete unit only.
• Before the blade holdercan be replaced,it
must be brought into a 45° inclined position.
To do so, insert the size 3 Allen key sideways
into the blade holder through the hole (8) and rotate it 45° clockwise.
Turning the screw (9) counterclockwise
releases the blade holder, allowing it to be replaced.
• Theinstallationiscarriedoutinreverseorder.
34
Cleaning the blade holder
To clean the blade holder after removal, spray it with alcohol. It can then be wiped down with
7
a piece of cellulose and placed on a cellulose towel to dry completely.
Fig. 19
Instructions for Use V 1.4 RevC – 09/2013
Page 37
5. Working with the Instrument

5.4.5 Inserting the blade

8
10
11
9
Fig. 21

5.4.6 Adjusting the clearance angle

Fig. 20
The blade holder can be used for razor blades, injector blades and sapphire blades (universal blade holder).
• Inserta size 3Allen key sidewaysinto the
blade holder through the hole (8) and rotate it 90° clockwise.
Clamp the blade as follows:
• Inserttheprovidedsize3Allenkeyfromthetop
through the opening (10) into the blade holder (11) and open the blade holder (BH).
• Holdtheentirerazorblade(9) (not separated)
on the left and right with both hands and insert it into the blade holder. Hook in the blade over bottom pressure plate (see Fig. 21).
• ClampdowntheBHbyturningthesize3Allen
key clockwise until hand-tight.
The clamping screw (11) on the blade holder must not be tightened too much!
Now return the blade holder to the cutting posi­tion.
• Todo so, insertasize 3 Allenkeysideways
into the blade holder through the hole (8) and rotate approx. 90° counterclockwise.
For more information about adjusting the clear­ance angle, see Fig. 22.
• Inserta size 3Allen key sidewaysinto the
blade holder through the hole (8) and rotate it to the desired clearance angle mark.
12 - 15° 13 - 18° 14 - 21°
12 13 14Fig. 22
Leica VT1200 / VT1200 S
For 15°, the effective clearance angle is "0". The most commonly used setting is 18°
(13).
35
Page 38
5. Working with the Instrument

5.5 Routine daily maintenance and switching off the instrument – VT1200/VT1200 S

After all daily procedures have been finished, perform the following:
• Switchoffthemainswitchonthesideoftheinstrument.
• Placethemagnifiercoveronthemagnifier.
• Removethebladefromthebladeholderanddisposeofitsafely.
• Pulltheicetrayandbuffertrayoffofthedovetailguideandplacethem
on the stage.
• Removeandemptyoutthebuffertray.Disposeofthecontentsofthe
buffer tray properly.
• Removethespecimenplateandlayitflatonthestage.
• Removethespecimenusingasingle-sidedbladeandremoveanycya-
noacrylate adhesive residue from the specimen plate.
36
Caution! The contents of the ice tray can become contaminated if buffer solution is spilled over it.
Instructions for Use V 1.4 RevC – 09/2013
Page 39

6.1 Using the VibroCheck with the VT1200

The following instructions must be adhered to exactly. Noncompliance can cause serious damage to the instrument.
We recommend using the VibroCheck after each blade change to check the optimum position of the blade and minimize the vertical vibration.
Foot switch for VT
(2) - VibroCheck
Control panel for VT
Fig. 23
3

6. Operating the VibroCheck

Prior to assembly, the dovetail guide (3) is brought down to the lowest position using the
DOWN key!
1. The user installs the VibroCheck: Push the VC
along the dovetail guide (3) to behind the mark on the baseplate of the instrument (rear stop) and clamp it down using the lever (15). Insert the blade and clamp tightly. Return the blade to the cutting position (see Fig. 20).
2. Plug the connecting plug of the VibroCheck
(VC) into the socket (2 ) on the left side panel. Short red flash of LED on VC --> control panel acknowledges VC. LED then remains red. The LED in the DOWN key flashes green.
Leica VT1200 / VT1200 S
Fig. 24
15
Fig. 25
3. The user presses the DOWN key. The VC
moves to the bottom position, after that the blade moves to the rearmost position—the LED in the RUN/STOP key flashes. Short red flash of LED on VC --> it then remains red.
4. The user presses the RUN/STOP key: First, the
blade moves forward (into the position exactly above the VibroCheck), then the VC moves into a position in which the blade partially covers the light barrier. (If the VC does not receive any signal via the light barrier, the operation is canceled and DOWN is enabled.)
LED on VC flashes green - RUN/STOP flashes
yellow.
37
Page 40
6. Operating the VibroCheck
18a
1
When LED on VC is green and LED in RUN/
STOP is yellow, the blade starts to vibrate with
the amplitude set. Speed = 0, amplitude can be changed at any time.
On the (5-digit) display, the deviation in height
amplitude is shown in µm (e.g. 0.9 µm). This value can be either positive or negative. The display can be switched using the µm/µm key. Then a number appears (e.g. 0.4). This means clockwise rotation by a turn of 0.4 ("+" see 18a) – (Neg. op-
2
Fig. 26
erational sign means counterclockwise rotation – see 18a.) If the value is "0", no improvement of the height amplitude is possible.
17
5. Press the STOP key. Using a size 3 Allen key, unscrew the clamping screw (16) only slightly, pull off the cap (17) by pulling it upwards (keep it in a safe place) and turn the adjusting screw (18) by the corresponding value (here, a turn of
0.4) clockwise (in the "+" direction, 18a) using a
18
size 3 Allen key. Tighten the clamping screw (16) clockwise.
6. Press the RUN key, check the value, repeat steps 5-7 if necessary.
38
18
16
If the display shows µm (2) "0" (opti­mal) and the value in µm (1) is unaccept­ably high, the blade must be replaced.
Fig. 27
7. If the measured value is accepted, press DOWN (flashes green). VC moves to lowest
position - blade is moved to the rear. LED on VC illuminates red again (LED on RUN/STOP is off).
8. The instrument now expects the VibroCheck to be removed. To do so, disconnect the USB
plug connection of the VC to the basic instru­ment and pull the VC off of the dovetail guide. LED in DOWN key continues to flash – wait until LED in DOWN button goes out.
The normal operating status is now restored.
Instructions for Use V 1.4 RevC – 09/2013
Page 41
6. Operating the VibroCheck

6.2 Using the VibroCheck with the VT1200 S

The following instructions must be adhered to exactly. Noncompliance can cause seri­ous damage to the instrument.
We recommend using the VibroCheck after each blade change for checking the optimum posi­tion of the blade and important sectioning parameters.
Prior to assembly, the dovetail guide (3) is
Foot switch for VT
VibroCheck
Control panel for VT
Fig. 28
3
15
Fig. 29
Fig. 30
Fig. 31
brought down to the lowest position using the
DOWN key!
1. The user installs the VibroCheck: Push the VC
along the dovetail guide (3) to behind the mark on the baseplate of the instrument (rear stop) and clamp it down using the lever (15). Insert the blade and clamp tightly. Return the blade to the cutting position (see Fig. 19).
2. Plug the connecting plug of the VibroCheck (VC) into the corresponding socket on the left side panel. Short red flash of LED on VC --> it then remains red. Control panel detects VC (see Fig. 30). The LED in the DOWN key flashes green.
3. The user presses the DOWN key. The VC moves to the bottom position, in addition to that the blade moves to the rearmost position—the LED in the RUN/STOP key flashes. Short red flash of LED on VC --> it then remains red.
4. The user presses the RUN/STOP key: First, the blade moves forward (into the position exactly above the VibroCheck), then the VC moves into a position in which the blade partially covers the light barrier. LED on VC flashes green - RUN/STOP flashes yellow. Control panel shows: "VIBRO search" (see Fig. 30). A search can last up to 1 minute. When LED on VC is green and LED in RUN/STOP is yellow, the blade starts to vibrate.
Leica VT1200 / VT1200 S
39
Page 42
6. Operating the VibroCheck
21
18a
18
Fig. 32
17
The deviation of the height amplitude is shown
on the display in µm (1). This value can be ei-
ther positive or negative. A number (e.g. –0.3)
appears in the µm display (2). This means a counterclockwise rotation (because of the "-") by a turn of 0.3 and reduces the height ampli­tude to a minimum. (If there is no operation sign, the rotation is in a clockwise direction "+".) If the value is "0", no improvement of the
height amplitude is possible.
5. Press the STOP key. Unscrew the clamping screw (16) only slightly using the size 3 Allen key, pull the cap (17) up and off (keep it in a safe place) and turn the adjusting screw (18) counterclockwise (the "-" direction, 18a) by a turn of 0.3 using the size 3 Allen key. Tighten the clamping screw (16) clockwise.
6. Press the RUN key, check the value, repeat steps 5-7 if necessary.
40
18
If the display shows µm (2) "0" (opti­mal) and the value in µm (1) is unaccept­ably high, the blade must be replaced.
7. If the measured value is accepted, press DOWN (flashes green). VC moves to lowest
position - blade is moved to the rear. Control
16
Fig. 33
panel of the VT shows: "VIBRO END". The LED on the VC is again illuminated in red.
8. The instrument now expects the VibroCheck to be removed. To do so, disconnect the USB plug connection of the VC to the basic instru­ment and pull the VC off of the dovetail guide. LED in DOWN key continues to flash – wait until LED in DOWN button goes out.
The normal operating status is now restored.
Fig. 34
Instructions for Use V 1.4 RevC – 09/2013
Page 43

7. Malfunctions: Meaning and Troubleshooting

7.1 Error messages and troubleshooting

Error No. / INF
Error Troubleshooting Comment
No.
Error 01 - Wrong control panel
(VT1200 or VT1200 S)
Error 21 - Head does not swing into
place.
Error 22 - Timeout while initializing
the x-axis. (Timeout)
Error 23 - DC motor of x-axis does
not rotate (during initial­ization or in normal op­eration)
Error 24 - "X-Start" limit switch is
not reached.
Error 25 - "X-Start" limit switch can-
not be moved away from.
Error 26 - "X-Stop" limit switch is not
reached.
- Use the correct control panel for the instrument.
1. Check to see if an am­plitude value has been selected.
2. Try to make the head swing into place by gently striking it with your hand.
3. If head does not swing into place, inform service.
- Inform service!
- Inform service!
- Inform service!
- Inform service!
- Inform service!
- Using an incorrect control panel does not harm the instrument, but no func­tions whatsoever are available.
After all error messages, the instrument MUST be switched off, then on again using the main switch.
Leica VT1200 / VT1200 S
41
Page 44
7. Malfunctions: Meaning and Troubleshooting
Error No. /
Error Troubleshooting Comment
INF No.
Error 27 - "Bottom" limit switch of
the Z-axis is not reached during initialization or operation.
Error 28 - The "top" limit switch of
the Z-axis is not reached.
Error 31 - Both X sensors activated
(during initialization or in normal operation)
Error 32 - Both Z sensors activated
(during initialization or in normal operation)
- Check whether an ob­stacle is blocking the travel path of the ice tray. Remove the obstacle. If there is no obstacle and the error message per­sists after the instrument is switched on again: Inform service.
- Check whether an ob­stacle is blocking the travel path of the ice tray. Remove the obstacle. If there is no obstacle and the error message per­sists after the instrument is switched on again: Inform service.
- Inform service!
- Inform service!
42
After all error messages, the instrument MUST be switched off, then on again using the main switch.
Instructions for Use V 1.4 RevC – 09/2013
Page 45
7. Malfunctions: Meaning and Troubleshooting
Error No. /
Error Troubleshooting Comment
INF No.
InF 41 - Control board (C1/back
panel) and keyboard (C2) have different software versions.
InF 42 - Control board (C1/back
panel) and VibroCheck (C3) have different soft­ware versions.
"InF" messages can be removed by pressing the Clear key. The message appears once each time the instrument is switched ON. "InF" messages do not cause the instrument to be locked. After all error messages, the instrument MUST be switched off, then on again using the main switch. – Exceptions are InF 41 and 42.
- It is possible that due to different software versions, some or all functions of the instru­ment may be partly or entirely unavailable. In­form service and have the software updated to the latest version.
- It is possible that due to different software ver­sions, some or all func­tions of the instrument may be partly or entirely unavailable.
Error 51 - Horizontal calibration of
VibroCheck is not pos­sible.
Leica VT1200 / VT1200 S
- Possible operating error, see the Instructions for Use.
- Great deal of damage or dirt on the blade or blade holder. Use a new blade or blade holder.
- Controllers have different software versions. Have the software updated.
- VibroCheck is defective. Have it checked by ser­vice.
- A calibration is carried out even if the blade is jagged.
A great deal of obvious
damage or dirt must be present before a blade calibration is canceled automatically.
43
Page 46
7. Malfunctions: Meaning and Troubleshooting
Error No. /
Error Troubleshooting Comment
INF No.
Error 52 - Basic calibration of light
barrier is not possible for VibroCheck.
Error 53 - VibroCheck blade search
not successful.
Communication between key­board - VT1200/S
- The knife/ blade collides with the buffer tray
- The Vibro­Check collides with the blade
- Communication error between control unit and VT
- The buffer tray was not pushed in up to the rear pin or was not clamped down during installation.
- The VibroCheck was not pushed all the way in or was not clamped down during installation.
- Switching from SINGLE to CONT is not possible.
- Emitter diode or receiver is defective.
- VibroCheck is defective.
- Inform service!
- No blade and/or blade holder swung out or not installed.
- Emitter diode or receiver dirty. Clean
- VibroCheck is defective.
- VT1200 S display lights up, but remains blank.
- VT1200: row of dots flash­es.
- Inform service!
Push in the buffer tray as
far as it will go and clamp it down with the lever.
Push in the VibroCheck as
far as it will go and clamp it down with the lever.
- In MAN mode, only single cuts are possible.
- VT1200 drives to the low­est Z position using the stepper motor after an unsuccessful search. (Allows blade installation)
44
- The cutting window can­not be defined.
- In MAN mode, only single cuts are possible.
- If multiple cuts are required, switch to AUTO mode.
Instructions for Use V 1.4 RevC – 09/2013
Page 47
7. Malfunctions: Meaning and Troubleshooting
Error No. / INF No.
The instrument is not working.
Error Troubleshooting Comment
- Not accepted:
 •Negativevalues  •Valuesofmorethan
1000 µm
- In MAN mode, pressing the AUTO FEED key feeds the last saved value.
When a key is pressed, a beep sounds.
If the 1000 hour limit is exceeded while the interment is in normal operating mode, a flashing "SEr" appears in the upper 3-digit LED for approx. three sec­onds after the sectioning process has been ended with the RUN/STOP key. This indicates that the vibrating component needs to be serviced.
The plug connections are loose or the instrument is connected to the power supply improperly or not at all. The power fuses are defec­tive.
- In MAN mode, if the
VT connection cable: Check the control unit and power plug. Replace the power fuses.
Blocked when RUN is active!
PAUSE key has not been
assigned a function, a short beep sounds.
Leica VT1200 / VT1200 S
45
Page 48
7. Malfunctions: Meaning and Troubleshooting

7.2 Replacing the main fuse

Caution: Always unplug the power supply before replacing the fuses!
• Removethefusehousingontherightsideoftheinstrumentabovethe
power switch. To do so, push an appropriate tool (small screwdriver) into the slots on the left and right (Fig. 35a) and carefully pull it out.
• Removethedefectivefuseandreplaceitwith
the replacement fuse included in the standard scope of delivery (see Fig. 35c).
46
35a
35c
35b
• Reinsertthefusehousingintotheholderonthe
instrument as shown in Figure 35b and gently push it in until you hear it click into place.
Fig. 35
Instructions for Use V 1.4 RevC – 09/2013
Page 49

8. Cleaning and Maintenance

8.1 Cleaning the instrument

Always put the blades back into the blade case/dispenser when not in use. When using cleaners, comply with the safety instructions from the manufacturer and the labor-safety regulations at your laboratory. When cleaning the outer surfaces of the instrument, do not use xylene or solvents con­taining acetone or xylene. The finished surfaces are not resistant to xylene or acetone! Ensure that liquids do not enter the interior of the instrument during cleaning.
Before each cleaning, carry out the following preparatory steps:
• Switchoffthemainswitchonthesideoftheinstrument.
• Placethemagnifiercoveronthemagnifier.
• Removethebladefromthebladeholderanddisposeofitsafely.
• Pulltheicetrayandbuffertrayoffofthedovetailguideandplacethem
on the stage.
• Removeandemptyoutthebuffertray.Disposeofthecontentsofthe
buffer tray properly.
• Removethespecimenplateandlayitflatonthestage.
• Removethespecimenusingasingle-sidedbladeandremoveanycya-
noacrylate adhesive residue from the specimen plate.
Instrument and outside surfaces
If necessary, the varnished outside surfaces of the control panels can be cleaned with a mild commercial household cleaner or soap water and then be wiped with a cloth. The instrument must be completely dry before it can be used again.
Cleaning the blades
Always wipe the blade from the back of the blade (sapphire blade) to the cutting edge. NEVER wipe in the opposite direction - risk of injury!
Clean using an alcohol-based solution or acetone.
Leica VT1200 / VT1200 S
47
Page 50

9. Ordering Information for Optional Accessories, Consumables and Spare Parts

Ice tray ............................................................................................................ 14 0481 42010
Buffer tray
Buffer tray, assembly (plastic) .................................................................... 14 0481 42089
Buffer tray, assembly (metal) ...................................................................... 14 0481 42084
Buffer tray, double-walled, assembly ...................................................... 14 0481 44837
Lids for covering the buffer trays made of
plastic or metal .............................................................................................. 14 0481 42090
Hose clamp ................................................................................................... 14 0481 41952
Specimen plate
Specimen plate, non-directional (for 20 mm high specimens) .............. 14 0481 42086
Specimen plate, non-directional (for 10 mm high specimens) .............. 14 0481 43399
Specimen plate, directional, incl. swivel handle ..................................... 14 0481 42068
Blade
Sapphire blade, knife angle of 22° ............................................................. 14 0216 39372
Blade holder, assembly ............................................................................... 14 0481 42030
VibroCheck .................................................................................................... 14 0481 42075
Display
Microscope, assembly ................................................................................. 14 0481 42024
Cover for the LED connection ..................................................................... 14 0481 43402
Magnifier, assembly ..................................................................................... 14 0481 42035
Cold light sources
Cold light source CLS 100X 100 - 120 V/50-60 Hz ..................................... 14 0502 30214
Cold light source CLS 100X 230 - 240 V/50-60 Hz ..................................... 14 0502 30215
Cold light source CLS 100X 240 V/50-60 Hz ............................................... 14 0502 30216
Fiber-optic light guide ................................................................................. 14 0502 30028
Adhesive
Cyanoacrylate adhesive containing 10 gr. ............................................... 14 0371 27414
Foot switch .................................................................................................... 14 0481 43397
Protective cover, small ................................................................................ 14 0212 43742
Protective cover, large ................................................................................ 14 0212 43743
Fuse: cut-out T1A, 5*20 ................................................................................ 14 6943 01000
Cover for the LED connection ..................................................................... 14 0481 43402
Julabo FL300, recirculating cooler/chiller
100 V/50 /60 Hz ............................................................................................... 14 0481 48439
115 V/50 Hz ..................................................................................................... 14 0481 48437
230/50-60 Hz ................................................................................................... 14 0481 48436
230 V/60 Hz ..................................................................................................... 14 0481 48438
Antifrogen N .................................................................................................. 14 0481 45443
48
Instructions for Use V 1.4 RevC – 09/2013
Page 51
9. Ordering Information for Optional Accessories, Consumables and Spare Parts

9.1 Extension accessories for standard specimens

Ice tray
• Removable
• Drawerdesign
• Remainsstableonthestagewhenpreparing
specimen
• Integratedhand rests foruse with Leica
VT1200/VT1200 S.
Order No. ........................................... 14 0481 42010
Fig. 36

9.1.1 Buffer trays

Buffer tray, assembly (plastic)
• Buffertrayandlid
• Magneticholderforthespecimenplate
• Specimenplate,non-directional
• Clamp(2x)forholdingthehoseinpositionin
the buffer tray
• Volume:125cm3 *
• Autoclavable
Hose clamp
HC
Fig. 38
Leica VT1200 / VT1200 S
Underside
Fig. 37
Order No. ........................................... 14 0481 42089
Buffer tray, assembly (metal)
• Buffertrayandlid
• Magneticholderforthespecimenplate(see
the minimized illustration)
• Specimenplate,non-directional
• Clamp(2x)forholdingthehoseinpositionin
the buffer tray
• Volume:125cm3 *
• Non-autoclavable
Order No. ........................................... 14 0481 42084
* (Specifications without the blade holder, measured
4 mm below the upper edge of the buffer tray)
49
Page 52
9. Ordering Information for Optional Accessories, Consumables and Spare Parts
Buffer tray, double-walled, incl. integrated hand rest, assembly
• Buffertray,double-walled
• Magneticholderforthespecimenplate
• Specimenplate,non-directional
• Clampforholdingthehoseinpositioninthe
buffer tray
• Volume:400cm3 *
• Non-autoclavable
Fig. 39
* (Specifications without the blade holder, measured
4 mm below the upper edge of the buffer tray)
• Hosesetforconnectingarecirculatingcooler/
chiller (e.g. 14 0481 48436)
Order No. ........................................... 14 0481 44837
Lid for covering the buffer tray
• Thebuffertrayisavailableinplastic(140481
42089) or metal (14 0481 42084)
Order No. ........................................... 14 0481 42090

9.1.2 Specimen plate

50
Fig. 40
Specimen plate, non-directional
• Forspecimens2cminheight
• Canberotated360°
• Attachedwithmagnetsinthebuffertray.
Order No. ........................................... 14 0481 42086
Fig. 41
Instructions for Use V 1.4 RevC – 09/2013
Page 53
9. Ordering Information for Optional Accessories, Consumables and Spare Parts
Specimen plate, non-directional
• Forspecimens1cminheight
• Canberotated360°
• Attachedwithmagnetsinthebuffertray.
Order No. ........................................... 14 0481 43399
Fig. 42
10°

9.1.3 VibroCheck

Leica VT1200 / VT1200 S
Magnet on the underside
Fig. 44
Specimen plate, directional
• Inc.swivelhandle
• Inc.marksforanglingspecimen5°and10°
Order No. ........................................... 14 0481 42068
Fig. 43
Optional measuring instrument for displaying the vertical deviation of the blade (in µm) and for displaying the direction of rotation of the screws used to minimize the vertical deviation of the blade. Can be adjusted using the adjusting screw on the blade holder.
Order No. ........................................... 14 0481 42075
51
Page 54
9. Ordering Information for Optional Accessories, Consumables and Spare Parts

9.1.4 Blades

Sapphire blade, knife angle of 22°
The blade can be reground.
Order no. ............................................ 14 0216 39372
Fig. 45

9.1.5 Microscope, assembly

9.1.6 Magnifier, assembly

- Microscope support
- S6 Stereozoom microscope
- 2 eyepieces, 10x23 B, adjustable
- Protective objective glass
- Bearing cylinder with 3 screws
- Dust cover, large
- Adapter for the fiber-optic light guide slot
- Cover for the LED connection
Fig. 46
Order No. ........................................... 14 0481 42024
- Magnifier support
a
- Lens (2x magnification) with lens cover (a)
- Bearing cylinder with 3 screws
- Dust cover, large
- Adapter for the fiber-optic light guide slot
52
Fig. 47
Order No. ........................................... 14 0481 42035
Instructions for Use V 1.4 RevC – 09/2013
Page 55
9. Ordering Information for Optional Accessories, Consumables and Spare Parts

9.1.7 Installing the magnifier support or microscope support

Fig. 48
3
• First,remove thecover (3) from the basic instrument and store it in a safe place.
• InserttheprovidedAllenscrews(a) into the hole (4) in the bearing bush. Use the size 3 Allen key to screw the screws into the open slot on the top side of the basic instrument.
• Pushthemagnifiersupportormicroscopesup-
a
port into the bearing bush as far as it will go.
• Adjustthemagnifiertothedesiredinclination
angle and screw in the screw (2, Fig. 49) using the size 3 Allen key, or
4
• InserttheS6Stereozoommicroscopeintothe
ring holder and attach it by screwing it down (5, Fig. 50 b).
• Determinethedesiredinclinationangleofthe
microscope and fix it in place by tightening the screw (6, Fig. 50a) clockwise using a size
2
3 Allen key.
• The microscopeheight canbe adjustedusing
the setting dials for height adjustment (7, Fig. 50a) and adapted to the respective specimen.
6
7
Leica VT1200 / VT1200 S
Fig. 49
7
Adapter for installa­tion of a fiber-optic light guide
5
5
Fig. 50a
Fig. 50b
53
Page 56
9. Ordering Information for Optional Accessories, Consumables and Spare Parts

9.1.8 Fiber-optic illumination, cold light source

Fiber-optic illumination
• Thefiber-opticilluminationisinstalledafterthe
magnifier has been installed in the magnifier holder and then connected to the cold light source.
Fig. 52
Order no. 14 0502 30028
Leica CLS 100 cold light source
• Servesas a lightsource for thefiber-optic
illumination.
Fig. 51
54
Before use, carefully read the separate­ly provided Instructions for Use!
100-120 V, 50/60 Hz Order No. 14 0502 30214 230-240 V, 50/60 Hz Order No. 14 0502 30215 240 V, 50/60 Hz Order No. 14 0502 30216
Fig. 53
The cold light source must be operated at the power supply voltage specified on the nameplate (on the underside of the instrument).
Instructions for Use V 1.4 RevC – 09/2013
Page 57
9. Ordering Information for Optional Accessories, Consumables and Spare Parts

9.1.9 Cyanoacrylate adhesive

Single-component adhesive for adhering speci­mens to the specimen plate – Contains 10 gr.
Order no. ............................................ 14 0371 27414
Fig. 54

9.1.10 Foot switch

Insert the foot switch into the uppermost socket
5
marked "FS" (5) on the left side of the basic instru­ment.
• Thefoot switch performsthe START/STOP function.
Order no. ............................................ 14 0481 43397
Fig. 56

9.1.11 Julabo FL300 – recirculating cooler/chiller

Recirculating cooler/chiller for connecting to the double-walled buffer tray in the Leica VT1000 S and VT1200/VT1200 S.
Selectable temperature range: –20 °C to +40 °C. Recommended cooling medium: Antifrogen N (14 0481 45443) Mixture with water (50 %/50 %)
Application example:
If (at an ambient temperature of 20 to 22 °C) a temperature of 4 °C is to be reached in the buffer tray, a setting value of 0.5 to 2 °C must be selected.
For additional information, refer to the Instructions for Use provided
Fig. 57
Leica VT1200 / VT1200 S
with this instrument.
55
Page 58

10. Warranty and Service

Warranty
Leica Biosystems Nussloch GmbH guarantees that the contractual product delivered has been subjected to a comprehensive quality control proce­dure based on the Leica in-house testing standards, and that the product is faultless and complies with all technical specifications and/or agreed characteristics warranted.
The scope of the warranty is based on the content of the concluded agree­ment. The warranty terms of your Leica sales organization or the organiza­tion from which you have purchased the contractual product shall apply exclusively.
Service information
If you are in need of technical customer support or spare parts, please contact your Leica representative or the Leica dealer where you purchased the instrument.
Please provide the following information:
• Modelnameandserialnumberoftheinstrument
• Locationoftheinstrumentandnameofthepersontocontact
• Reasonfortheservicecall
• Deliverydate
Decommissioning and disposal
The instrument or parts of the instrument must be disposed of according to existing applicable, local regulations.
56
Instructions for Use V 1.4 RevC – 09/2013
Page 59
Leica VT1200 / VT1200 S
57
Page 60

11. Decontamination Certificate (Master)

Decontamination Certificate
Dear Customer,
Any product that is to be returned to Leica Biosystems or serviced on site, must be cleaned and decontaminated in the appropriate manner. Since it is not possible to decontaminate for prion diseases, such as CJD, BSE, CWD etc., equipment exposed to specimens containing prion diseases cannot be returned to Leica Biosystems for repair. On-site repair of prion contaminated equipment will only be conducted after the Field Service Engineer has been educated in the risks, instructed in the policies and procedures of the institution, and provided with personal protective equipment. Please fill out this confirmation carefully and enclose a copy with the instrument. Attach the confirmation to the outside of the flight case or hand it directly to the service technician. Packages will not be opened, nor servicing commenced until the Company or service engineer have received a satisfactory certificate. Should returned goods be considered a hazard by the Company, they will be returned immediately to the customer at his/her expense. Note: Microtome knives must be in boxes.
Mandatory information: Fields marked with * are mandatory. Depending on whether the instrument is contaminated, please also complete either section A or section B.
Nameplate information Model (see nameplate)*
REF (see nameplate)*
SN (see nameplate)*
A B
Yes
No
No
Yes
Yes No
Yes
Tick Box A if applicable. Otherwise please complete all parts of B, providing further information as requested or appropriate.
This equipment has not been in contact with unfixed biological samples.
This equipment has been exposed internally or externally to hazardous materials as indicated
1
below:
Blood, body fluids, pathological samples
Other biohazards
Chemicals/substances hazardous to health
Other hazards
Radioactivity
This equipment has been cleaned and decontaminated:
2
If yes, give details of the method:
If no**, please indicate why not:
** Such equipment must not be returned without the written agreement of Leica Biosystems.
The equipment has been prepared to ensure safe handling/transportation.
3
Whenever possible, please use the original transportation case/box.
Please provide further detail here:
Please provide further detail here:
58
Instructions for Use V 1.4 RevC – 09/2013
Page 61
Tick Box A if applicable. Otherwise please complete all parts of B, providing further information as requested or appropriate.
Nameplate information Model (see nameplate)*
SN (see nameplate)*
REF (see nameplate)*
This equipment has not been in contact with unfixed biological samples.
A B
This equipment has been exposed internally or externally to hazardous materials as indicated below:
Yes
No
Blood, body fluids, pathological samples
Other biohazards
Chemicals/substances hazardous to health
Other hazards
Yes
This equipment has been cleaned and decontaminated:
Yes
No
If yes, give details of the method:
If no**, please indicate why not:
Please provide further detail here:
Please provide further detail here:
** Such equipment must not be returned without the written agreement of Leica Biosystems.
The equipment has been prepared to ensure safe handling/transportation. Whenever possible, please use the original transportation case/box.
Yes No
1
3
2
Important - to avoid refusal of shipment:
Version 3.5 - 01/2013, Order-No. 14 0700 83101
Dear Customer, Any product that is to be returned to Leica Biosystems or serviced on site, must be cleaned and decontaminated in the appropriate manner. Since it is not possible to decontaminate for prion diseases, such as CJD, BSE, CWD etc., equipment exposed to specimens containing prion diseases cannot be returned to Leica Biosystems for repair. On-site repair of prion contaminated equipment will only be conducted after the Field Service Engineer has been educated in the risks, instructed in the policies and procedures of the institution, and provided with personal protective equipment. Please fill out this confirmation carefully and enclose a copy with the instrument. Attach the confirmation to the outside of the flight case or hand it directly to the service technician. Packages will not be opened, nor servicing commenced until the Company or service engineer have received a satisfactory certificate. Should returned goods be considered a hazard by the Company, they will be returned immediately to the customer at his/her expense. Note: Microtome knives must be in boxes.
Mandatory information: Fields marked with * are mandatory. Depending on whether the instrument is contaminated, please also complete either section A or section B.
Radioactivity
Place one copy in the unit prior to packaging, or hand it over to the service engineer. Customer assumes all responsibility for the immediate return shipment of articles sent to Leica without proper decontamination documentation. If you have any further questions, please call your local Leica organization. Leica Internal Use: If applicable, note corresponding Job and RAN-/RGA-Number:
Job Sheet No.: _______________ BU Return Authorization Number: ______________ SU Return Goods Authorization: ______________
Signature/Date*
Name*
Position*
eMail
Leica Biosystems Nussloch GmbH Heidelberger Str. 17-19 69226 Nussloch, Germany
Phone: ++49 (0) 6224 143 0 Fax: ++49 (0) 6224 143 268 www.LeicaBiosystems.com
11. Decontamination Certificate (Master)
Institute*
Department*
Address*
Phone*
Fax
Leica VT1200 / VT1200 S
59
Page 62
www.LeicaBiosystems.com
Leica Biosystems Nussloch GmbH Heidelberger Straße 17-19
D- 69226 Nussloch Phone: +49 - (0) 6224 - 143 0
Fax: +49 - (0) 6224 - 143 268 Web: www.LeicaBiosystems.com
Loading...