Read the user manual and safety instructions before starting up.
The Leica MS5, MZ6, MZ75, MZ95, MZ125, MZ16 and MZ16 A stereomicroscopes are optical instruments which use magnification for improved
visualization of objects, their details and specimens from technology and
natural sciences. The basic outfit, which includes a stand and a light
source, can be supplemented with various modules for photography, TV,
dual station viewing and other applications.
• The use of the instrument in a different manner from that described in
this user manual can lead to injury, malfunction and damage.
• The Leica MS5, MZ6, MZ7
5, MZ95, MZ125, MZ16 and MZ16 A stereo-
microscopes must not be used for eye examinations and operations.
• Do not fit different plugs. Do not dismantle optical systems or mechani-
cal parts unless instructions for doing so are given in the user manual.
• The Leica MS5, MZ6, MZ7
5, MZ95, MZ125, MZ16 and MZ16 A stereo-
microscopes are intended mainly for indoor use.
• If the stereomicroscope is used outdoors, protect it from dirt and mois-
ture. Lamps and electrically-operated stands from Leica may not be
used outdoors.
The optics carriers MS5, MZ6, MZ7
5, MZ95, MZ125, MZ16 and MZ16 A, the
binocular tube, ErgoTube™, ErgoWedge™ 5°–25°, the swinging-arm
stand ESD and the cold light sources Leica L2 and CLS consist of ESDdischarging material (surface resistance <10
11
ohms/squared, discharge
time <2 seconds, 1,000 V to 100 V).
– A connection socket for a ∅ 4 mm grounding cable is provided on the
microscope carrier.
– There are two connection sockets for ∅ 4 mm grounding cables in the
base of the swinging-arm stand.
The Leica M series can be used in clean rooms without any problems and
they can be cleaned as described on page 60. Please observe the
following rules:
– Do not clean our stereomicroscopes and accessories in a different
manner from that described in this user manual. Do not use unsuitable
cleaning agents, chemicals and techniques. Do not clean painted
surfaces and accessories with rubber parts such as the ErgoTube
®
10° – 50° with chemicals.
This can damage the surfaces. Particles or grindings can contaminate
the specimen.
– If the customer cleans our instruments with chemicals without our
written approval he will do it at his own risk.
– In most cases we can offer special solutions on request. Some
products can be modified or we could recommend other accessories
for use in clean rooms.
6
Leica M Stereomicroscopes – Safety concept
General directions
Permitted uses
Prohibited
uses
Place
of use
Use in ESD
protected rooms
Use in Clean Rooms
Safety concept