Leica ASP300 User Manual

Instruction manual
Leica ASP300 V1.3 English - 12/2003 Always keep this manual together with the instrument. Read this instruction manual carefully before working with the instrument.
Leica ASP300
Advanced Smart Processor
tissue processor
NOTE
The information, numerical data, notes and value judgments contained in this manual rep­resent the current state of scientific knowl­edge and state-of-the-art technology. We are under no obligation to update the present manual according to the latest tech­nical developments, nor to provide our cus­tomers with additional copies, updates etc. of this manual. For erroneous statements, drawings, techni­cal illustrations etc. contained in this manual we exclude liability as far as permissible ac­cording to the national legal system applicable in each individual case. In particular, no liabili­ty whatsoever is accepted for any financial loss or consequential damage caused by or related to compliance with statements or other information in this manual. Statements, drawings, illustrations and other information as regards contents or technical details of the present manual are not to be considered as warranted characteristics of our products.
These are determined only by the contract provisions agreed between ourselves and our customers. Leica reserves the right to change technical specifications as well as manufacturing pro­cesses without prior notice. Only in this way is it possible to continuously improve the tech­nology and manufacturing techniques used in our products. This document is protected under copyright laws. Any copyrights of this document are re­tained by Leica Microsystems Nussloch GmbH. Any reproduction of text and illustrations (or of any parts thereof) by means of print, photo­copy, microfiche, web cam or other methods – including any electronic systems and media – requires express prior permission in writing by Leica Microsystems Nussloch GmbH.
© Leica Microsystems Nussloch GmbH
Published by: Leica Microsystems Nussloch GmbH
Heidelberger Str. 17 - 19 D-69226 Nussloch Germany Telephone: +49 (0)6224 143-0 Fax: +49 (0)6224 143-200 Internet: http://www.histo-solutions.com
Leica ASP300
3
Table of contents
1. Important notes............................................................................................................................. 6
1.1 Symbols used in this manual................................................................................................... 6
1.2 Designated use .......................................................................................................................... 6
2. Safety .............................................................................................................................................. 7
2.1 Safety instructions..................................................................................................................... 7
2.2 Warnings...................................................................................................................................... 8
2.3 Integrated safety devices...................................................................................................... 10
3. Instrument components and specifications .......................................................................... 11
3.1 Specific instrument options................................................................................................... 11
3.2 Overview - instrument components .................................................................................... 12
3.3 Standard delivery - packing list............................................................................................ 14
3.4 Technical data ......................................................................................................................... 15
3.5 Compatible reagents ............................................................................................................... 18
4. Setting up the instrument .......................................................................................................... 19
4.1 Site requirements..................................................................................................................... 19
4.2 Electrical connection.............................................................................................................. 19
4.3 Installing the accessories ..................................................................................................... 21
4.4 Setting up the data connections.......................................................................................... 22
4.5 Alarm functions ........................................................................................................................ 23
4.6 Switching the instrument on ................................................................................................. 24
4.7 Touch screen functions......................................................................................................... 26
4.7.1 User interface - overview ...................................................................................................... 26
4.7.2. Access control ......................................................................................................................... 27
4.8 Check list for first operation (at supervisor access level).............................................. 28
5. Operation...................................................................................................................................... 29
5.1 Reagent handling..................................................................................................................... 29
5.1.1 Fill/drain reagents (other than wax) .................................................................................... 29
5.1.2 Changing the wax.................................................................................................................... 31
5.2 Setting up the instrument parameters................................................................................ 34
5.2.1 System setup............................................................................................................................. 34
5.2.2 Editing the reagent list ............................................................................................................ 37
5.2.3 Viewing the program list......................................................................................................... 40
5.2.4 Adding and/or modifying programs .................................................................................... 41
5.2.5 Favorites (maximum of 5 programs) .................................................................................... 44
4
Instruction manual V 1.3 – 12/2003
Table of contents
5.2.6 Editing the stations .................................................................................................................. 46
5.2.7 Reagent groups........................................................................................................................ 47
5.3 Programs.................................................................................................................................... 48
5.3.1 Editing a selected program ................................................................................................... 49
5.3.2 Starting a program................................................................................................................... 51
5.4 Reagent status ......................................................................................................................... 52
5.5 System Monitor......................................................................................................................... 53
5.6 On-line help ............................................................................................................................... 54
5.7 Configuring a printer (optional)............................................................................................. 55
6. Trouble shooting ........................................................................................................................ 56
6.1 General ....................................................................................................................................... 56
6.2 Power failure ............................................................................................................................. 56
6.3 Trouble shooting...................................................................................................................... 57
6.4 Typical fill or drain problems ................................................................................................. 58
7. Cleaning and maintenance ......................................................................................................59
7.1 Clean programs ........................................................................................................................ 59
7.1.1 Retort clean programs............................................................................................................ 59
7.1.2 Wax clean programs .............................................................................................................. 61
7.1.3 Smart clean ............................................................................................................................... 62
7.2 General cleaning steps .......................................................................................................... 63
7.3.1 Daily cleaning and maintenance ......................................................................................... 65
7.3.2 Periodic cleaning and maintenance................................................................................... 66
7.4 Preventative maintenance schedule ................................................................................. 67
8. Warranty and service ................................................................................................................ 69
Leica ASP300
5

1. Important notes

1.1 Symbols used in this manual and their meaning

Warnings appear in a grey box and are marked by a warning triangle .
Notes, i.e. important user information appear in a grey box and are marked by an information symbol .
Flammable solvents and reagents are marked with this symbol.
Instrument surfaces which become hot during operation are marked with this symbol.

1.2 Designated use

The Leica ASP300 is an automated tissue pro­cessor for the following laboratory applications:
fixation
dehydration
paraffin wax infiltration of histological tissue specimens.
The Leica ASP300 must be operated exclu­sively with the reagents listed in
Chapter 3.5 - “Compatible reagents“.
Any other use of the instrument
is considered improper!
(5)
ENTER
6
Figures in brackets refer to item nos. in drawings.
Function keys to be pressed on the instrument touch screen are written in bold-print capital letters.
Instruction Manual V 1.3 – 12/2003
Make sure to comply with the safety instructions and warnings in this chapter. Make sure to read these instructions, even if you are already familiar with the operation and use of other Leica products.

2.1 Safety instructions

2. Safety

This instruction manual includes important in­formation related to the operating safety and maintenance of the instrument and it is an im­portant part of the product.
If additional requirements on accident prevention and environmental protec­tion exist in the country of operation, this instruction manual must be supplemented by appropriate instruc­tions to ensure compliance with such requirements.
Instrument type
All information provided in this manual applies only to the instrument type indicated on the title page. A name plate indicating the instrument serial number is attached to the back of the instru­ment.
This instrument has been built and tested in accordance with the following safety regula­tions on electrical measuring, control, regulat­ing and laboratory devices:
DIN EN ISO 9001
EN 61010-1
EN 61010-2
EN 61326-1 and EN 61326-2
EN 61000-3-2 and EN 6100-3-3.
The protective devices on both in­strument and accessories may neither be removed nor modified. Only authorized and qualified service personnel may repair the instrument and access the instrument’s internal components.
Leica ASP300
7
2. Safety

2.2 Warnings

The safety devices installed in this instrument by the manufacturer only constitute the basis for accident prevention. Primarily responsible for accident-free operation is above all the institution which owns the instrument and, in addition, the designated personnel who operates, services or repairs the instrument. To ensure trouble-free operation of the instrument, make sure to comply with the following instructions and warnings.
Transport and installation
Warnings and instructions on transport and installation of the instrument can be found on a separate information sheet attached to the outside of the instrument’s shipping crate.
Warnings - reagent handling
To avoid damage to the instrument, only use the reagents listed in chapter 3.5!
When handling reagents, follow the instructions of the manufacturer and make sure all laboratory regulations in force in your country are complied with. Dispose of waste solvents with care according to local regulations and the waste manage­ment policy of the company or institution.
Use caution when handling hot, molten paraffin. Do not touch paraffin stations and retort walls – they can be very hot.
Do not clean the reagent bottles in an automatic dishwasher.
Material safety data sheets can be obtained from the supplier of the chemicals. They are also available on the Internet: http://www.msdsonline.com
8
Instruction manual V 1.3 – 12/2003
Warnings - cleaning and maintenance
Prior to each maintenance switch the instrument off and disconnect mains power.
Do not clean the instrument with solvents containing acetone or xylene. No liquid may be spilled into the internal components of the instrument.
Empty the condensate bottle at regular intervals.
Warnings - working with the instrument
The Leica ASP300 may only be operated by trained laboratory personnel and as per the present instruction manual.
In case of emergency switch off mains (item no. 5 in Fig. 3) at the rear of the instrument and unplug the power cord. The instrument must be installed so that the mains switch at the rear of the instrument can easily be accessed at any time.
Always press PAUSE before opening the retort lid during processing to allow the retort to vent.
2. Safety
Do not remove the wax drain hose or the remote fill/drain hose until the fill or drain has been entirely completed.
After refilling / replacing the reagent bottles ensure that the lids are tight. The bottles must be properly pushed home into the connection manifolds at the rear inner wall of the reagent module.
Fixatives containing mercuric salts, acetic or picric acid will corrode metallic components in the instrument.
After each paraffin step a retort clean cycle must be run.
Leica ASP300
9
2. Safety

2.3 Integrated safety devices

The instrument is equipped with the following protection features:
overpressure protection,
protection against excessive vacuum,
overcurrent protection,
overheating protection.
Overpressure protection
The design of the pressure system ensures that no dangerous levels of pressure can build up.
Protection against excessive vacuum
The vacuum system is not capable of generating a dangerous vacuum condition.
Overcurrent protection
Overheating protection
10
Overcurrent conditions are protected against by both the mains fuse and the separate heating power fuses.
The design of the heating system ensures that no dangerous overheat­ing conditions can occur.
Instruction manual V 1.3 – 12/2003

3.1 Specific instrument options

Reagent Management System (RMS).
The specimen cassette baskets may be used with or without cas­sette dividers and spring separators designed to separate cassettes and improve fluid penetration.
In-process fluid recirculation (‘tidal agitation’) for continuous and ef­ficient mixing of fluids.
Remote fill/drain system for reagents.
Remote fill/drain system for paraffin.
Automatic optical fluid level sensor.
Active paraffin wax solvent removal feature.
Magnetic stirrer.
Programmable program start time delay.
Programmable finish time for processing programs.
3-step retort drain feature.
Specimen processing under pressure, vacuum, alternating pressure / vacuum cycles or under ambient pressure.
User-programmable clean cycles.

3. Instrument components and specifications

Leica ASP300
11
3. Instrument components and specifications

3.2 Overview - instrument components

3
23
22
5
21
20
13
17
18
4
6
1
10 9
7
12 11 8
19
12
2
15
14
16
Fig. 1
Instruction manual V 1.3 – 12/2003
3. Instrument components and specifications
Overview - Instrument components
1 - Basic instrument - processor module 2 - Basic instrument - reagent module 3 - Wax bath lids 4 - Retort lid 5 - Touch screen 6 - Cover flap of instrument console 7 - Instrument console comprising: 8 - Printer port 9 - Local alarm connection 10 - Remote alarm connection 11 - Floppy disk drive 12 - Serial port 13 - Activated carbon filter 14 - Condensate bottle 15 - Reagent bottle (13 units) 16 - Drip tray 17 - Wax drain connection 18 - Remote drain connection 19 - Retort 20 - Wax baths 21 - Specimen basket 22 - Lid for specimen basket 23 - Splash protector
Instrument components and accessories
The processor module comprises three paraf­fin wax baths and the processing retort.
Three baskets (21) contain the cassettes. Each basket can hold up to 300 cassettes. All processing occurs in the stainless steel re-
tort under the pressure, vacuum and tempera­ture conditions selected.
The reagent bottles are stored in the reagent cabinet.
Leica ASP300
13
3. Instrument components and specifications

3.3 Standard delivery - packing list

Leica ASP300 basic standard delivery: 1 Leica ASP300 - Basic instrument 0476 34100
1 Accessory kit consisting of: 0476 35929
- 1 Basket hook
0476 34713
- 3 Specimen baskets, assy., including lid, dividers, springs and handle 0476 34193 (accessories for maximum and optimum usage of basket space)
- 1 Reagent bottle, plastic 0476 34274
- 2 Activated carbon filters, assy. 0476 34150
- 1 Remote fill/drain hose, assy. with special connecting device for remote filling / draining of processing retort 0476 34716
- 1 Paraffin drain hose 0476 34721
- 1 Paraffin wax scraper, plastic 0476 35923
- 1 Lubricant ‘Molykote 111’, for valves and O-rings, 100 g 0336 35460
- 1 Jumper cable - mains 0411 34604
- 1 Maintenance kit (2 spare lids, 9 O-rings) 0476 35921
- 1 Stirrer bar, assy. 0476 34202
- 2 Sets of labels for reagent bottles, each set containing 15 labels 0476 35792
- 1 Splash protector 0476 34770
1 Set of power cords:
- 1 Power cord D 0411 13558
- 1 Power cord UK ST-BU F-5A 0411 27822 13 Reagent bottles, plastic (included in the instrument) 0476 34274 1 Condensate bottle, plastic (included in the instrument) 0476 34278 1 Service kit for pump (included in the instrument) 0476 35922 1 Drip tray (included in the instrument) 0476 34145 1 Cover sheet for overflow (included in the instrument) 0476 34852 1 Console cover (mounted to the instrument) 0476 34162 1 3,5 Diskette, empty (taped to the back of console cover)
1 Instruction manual Leica ASP300 multilingual 0704 37102 1 Demo-Program, CD-Rom
14
Instruction manual V 1.3 – 12/2003
3. Instrument components and specifications

3.4 Technical data

Nominal voltage: Two factory-preset voltage settings
(not user-adjustable): 100 to 120 V + 10% or 230 to 240 V + 10%
Nominal frequency: 50 to 60 Hz Main fuses: 2 fast-blowing melting fuses, 20 x 5 mm, UL-listed
Order no.
100 to 120 V; F 10 A 250 VAC 69411002
230 to 240 V; F 5 A 250 VAC 69411000 Nominal power: 1,000 VA Dimensions (W x D x H), in mm: 595 x 680 x 1,325 Dry weight, unpacked: approx. 140 kg Weight, packed: 220 kg Operating temperature range: 15 °C to 35 °C Relative humidity: 10% to 80% - non-condensing IEC 1010 classification: Protective Class 1
Pollution Degree 2 Overvoltage Installation Category II:
800 V impulse (120 V systems)
1500 V impulse (240 V systems) Maximum operating altitude: maximum 2,500 m Local / remote alarm relays: 30 V DC, maximum 2 A
2 connections: both single voltage-free, change-over contacts (can be operated as both normally-open and normally-closed connections)
Paraffin wax baths
Number of baths: 3 Volume (liters): 4.3 l per bath Melt time: approx. 10 hrs
Temperature: 40 to 65 °C Temperature accuracy:
+ 1 °C
Leica ASP300
15
3. Instrument components and specifications
Retort
Capacity: max. 300 cassettes Reagent volume (in liters): 4.3 Temperature (paraffin): 40 to 65 °C Temperature (processing reagents): ambient or 35 to 55 Temperature (cleaning reagents): 50 to 65 °C Temperature accuracy:
+ 1 K Filling time: approx. 90 sec Drain time : approx. 80, 120, 140 sec (adjustable) Impregnation vacuum: -70 kPa (g) Impregnation pressure: 35 kPa (g) Fill vacuum: -70 kPa (g) Drain pressure: 35 kPa (g)
General
Reagent bottles 10 Cleaning solution bottles: 3 Maximum bottle volume: 4.3 l
Printer: Optional accessory Pretest check: ON/OFF Fluid level sensor: ON/OFF Recirculation (pump in/out): ON/OFF (a) Time before 1st cycle: 12 min (b) Time between cycles: 20 min
o
C
System setup
Password status: Supervisor / User Type of password: alphanumeric, freely selectable
Reagent Management Status: ON/OFF Software access control (‘Lock Mode’): ON/OFF
16
Instruction manual V 1.3 – 12/2003
3.4 Technical Data (continued)
Hardware and Software:
Large color liquid crystal display (LCD) and touchscreen.
User-friendly, intelligent software.
3,5“ floppy drive and printer port.
Alarm system with two remote alarm sockets.
Password-protected instrument supervisor mode.
Multiple specimen protection system.
Capacities:
15 programs of up to 10 reagent and 3 paraffin wax steps each.
Time per program step: 0 to 99 hours , 59 minutes.
Delay time: max. 7 days.
Up to 300 cassettes can be processed simultaneously.
Four user-programmable retort clean programs.
Wax clean program.
10 Reagent bottles.
3 Wax baths.
3 Cleaning solution bottles.
1 Condensate bottle.
Reagent temperature adjustable from 35°C to 55°C or ambient.
Paraffin temperature adjustable from 40°C to 65°C.
Choice of three retort drain rates of 80, 120 and 140 seconds.
Up to 100 reagent names in memory.
3. Instrument components and specifications
Leica ASP300
17
3. Instrument components and specifications

3.5 Compatible reagents

Fixatives
1. Formalin, buffered or unbuffered
2. Formalin saline
Dehydration
1. Ethanol 4. Butanol
2. Isopropanol 5. Industrial Methylated Spirits
3. Methanol
Clearing
1. Xylen (recommended)
2. Toluol
*
3. Chloroform *
Paraffin wax
1. Paraffin
* Prior to using these reagents, contact
your Leica representative or Leica di­rectly for information on necessary pro­tective measures. Using these reagents will require an increased number of main­tenance interventions by Leica Techni­cal Service.
18
Caution: Reagents other than those listed here may damage some com­ponents of the instrument. Do not use acetone, benzene or trichlorethane in the instru­ment.
Fixatives containing mercuric salts, acetic or picric acid will corrode metallic components in the instrument and shorten in­strument life. If you choose to work with such fixatives, it is essential to per­form a clean cycle which contains multiple water rinses each time after use, to minimize damage. In addition, we recommend frequent and regular preventive maintenance by the Leica Technical Service.
Instruction manual V 1.3 – 12/2003

4.1 Site requirements

Moving the instrument

4. Setting up the instrument

The instrument needs approx. 650 x 700 mm floor space.
Practically vibration-free floor.
Room temperature constantly between +15 °C and +35 °C.
Relative humidity maximum 80%, non-condensing.
Avoid vibrations, direct sunlight and heavy variation in temperature.
The chemicals to be used in the Leica ASP300 are both flam­mable and noxious. Therefore, the instrument must be set up in a well ventilated area, free from any ignition sources. The room where the instrument is installed is not suitable for people to stay in for longer periods of time, unless it is equipped with a fume extraction system. Do not operate the instrument in rooms with explosion hazard.
The instrument must be installed such that the mains switch at the rear of the instrument (item no. 5 in Fig. 3) can easily be ac­cessed at any time.
After the instrument has been unpacked (see unpacking instructions attached to outside of transport crate) it must be touched only at the points marked with dots ‘ ‘ for transport to its final site of installation. Once the instrument has been installed there, the brakes on the instru­ment rollers must be locked.

4.2 Electrical connection

Check the voltage label (Fig. 2) on the rear of the instrument to ensure
Leica ASP300
that the instrument is set to the correct voltage range.
Fig. 2
The power supply voltage for the instrument is factory-preset and CANNOT be altered by the user.
19
Connecting the instrument to mains
The instrument MUST be connected to an earthed mains power outlet socket. It is recommended that the ASP300 be plugged into a wall socket that has Ground Fault Circuit Interruption (GFCI) protection - as an additional electrical safeguard. Make sure to use the appropriate mains cable for the local volt­age supply (wall outlet).
Connect mains cable (6) to mains power supply socket (1) (see Fig. 3).
Jumper cable (3) connects the mains power supply output (2) to the electronics module input (4).
Switch ON/OFF switch (5) ON.
Switch ON/STOP switch at the right of the instrument ON.
Electrical connections at the
1
5
rear panel of the instrument
4. Setting up the instrument
6
20
4
2
3
Cables
connected
Fig. 3
correctly
Important:! Specification for connection (2) is as follows: 100 - 120 V or 230 - 240 V, max. 200VA.
Instruction manual V 1.3 – 12/2003
4. Setting up the instrument
6

4.3 Installing the accessories

Unpack the activated carbon filter and in­sert it into the instrument (see Fig. 4). Make sure to insert the filter with the correct side facing upwards. See label (6) at the front of
7
the filter - the arrow must point upwards.
Unpack the magnetic stirrer and insert into the retort.
Fig. 4
Install the console flap (covers floppy drive and ports).
l Prior to filling the bottles with reagent for the
first time, check the connections. If neces­sary, tighten manually.
Remote fill / drain hose
Connect the remote fill / drain hose as shown in Fig. 5.
8
Important! When inserting the hose into the fill/drain
connection (item 7 in Fig. 5) the connecting device (8) of the hose must lock with an au­dible click.
Leica ASP300
7
Fig. 5
21
4. Setting up the instrument
11
12

4.4 Setting up the data connections

9
8
7
10
Fig. 6
Printer (optional)
Lift off the instrument console flap (12).
Connect the printer to the printer port (7) via a standard Centronics cable (refer Fig. 6).
Other connections
For additional data transfer and storage the instrument console is equipped with a stan­dard 1.44 MB floppy drive (10) for 3.5“ dis­kettes and with an RS 232 serial port (11) .
Only instruments which satisfy the re­quirements for SELV circuits may be connected to interfaces (7), (8), (9) and (11).
22
Instruction Manual V 1.3 – 12/2003

4.5 Alarm functions

4. Setting up the instrument
The Leica ASP300 is equipped with two different alarm functions:
Instrument alarm
This alarm is generated from within the Leica ASP300. The instrument alarm is generated in all alarm situations.
Remote alarm
This alarm device is installed outside of the Leica ASP300. All alarms (all types of alarm) are forwarded to the remote dialer con-
nected to the instrument, regardless of whether the local instrument alarm (type 1) is connected or disconnected.
The optional remote alarm system works via a potential-free relay. If an alarm situation occurs, the circuit is closed. The remote alarm device connected to the instrument must be rated at less than 2 amp. A maximum voltage of 30 V DC may be present.
Leica ASP300
Connecting a remote alarm device:
A remote alarm device can be connected to the Leica ASP300 via a 6.3 mm-diameter jack connector (optional accessory).
For details on how to connect a remote alarm device to the Leica ASP300, please contact your local Leica office or the manufacturer Leica Microsystems Nussloch directly.
23
4. Setting up the instrument

4.6 Switching the instrument on

Once the instrument has been switched on, the mains switch (ON/OFF) (5) should always re­main in position ON.
Fig. 7
After switching it on the instrument will take a few minutes to initialize (see opposite -
Fig. 7).
After initialization, the FAVORITES screen will be displayed (Fig. 8).
Screen saver
A screen saver will operate turning the screen off when no key has been pressed for a (user-) programmable time. Press any part of the touchscreen to restore the screen. After having been restored, the screen will be inoperational for a few seconds to avoid accidental activation of any key symbols.
.
24
Fig. 8
Instruction Manual V 1.3 – 12/2003
4. Setting up the instrument
System diagnostics menu If, during initialization, ‘Touch here
...’ is touched (refer Fig. 7) and sub-
2
1
3
4
5
6
7
8
sequently the supervisor pass­word is entered, the SYSTEM DIAGNOSTICS menu (Fig. 9) opens up, giving access to standard in­strument settings.
Attention! Serious instrument malfunctions
may occur if any settings are modified wrongly.
Fig. 9
The following functions can be selected:
1 - Access for service engineers only. 2 - Access INSTALLATION menu. 3 - Abort the current program. 4 - Abort the current program and delete the
current allocation of reagents to reagent bottles and retort.
5 - Delete all programs and reset instrument
status (all lists will be empty).
6 - Store actual instrument status on floppy
disk.
7 - Display run log. 8 - Restart instrument.
Fig. 10
To exit this menu, the instrument has to be re­started - press: RESTART APPLICATION.
Press YES to confirm ‘ARE YOU SURE YOU WANT TO ...’ (refer Fig. 10) .
Initialization starts again from the screen
shown in Fig. 7.
Leica ASP300
25
4. Setting up the instrument

4.7 Touch screen functions

4.7.1 User interface - overview

The Leica ASP300 is programmed and operated via a colored LCD touch screen.
An on-line help feature provides detailed information on all instrument functions.
HELP can be accessed from any screen by pressing the HELP button.
Button symbols
To activate any functions on the Leica ASP300 touchscreen, press the corresponding button symbols. Examples of buttons are:
26
Enabled
Buttons on the Leica ASP300 touch screen can either be enabled or dis­abled.
Disabled
A button that is disabled appears with a thinner border than a button that is enabled.
Instruction manual V 1.3 – 12/2003

4.7.2. Access control

4. Setting up the instrument
The Leica ASP300 may be configured to allow two levels of user ac­cess.
‘Operator’ symbol
‘Supervisor’ symbol
Operator access level:
Operators may run programs and view results. On this level, the OP­ERATOR symbol is displayed in the upper right corner of the touch
screen.
Supervisor access level:
Supervisors may perform all Operator functions, and additionally cre­ate programs and perform the instrument sit-up functions.
To access Supervisor level, press SUPERVISOR, enter the required password and confirm. The SUPERVISOR symbol will then be dis­played instead of the OPERATOR symbol and all enabled buttons will be surrounded by a black border instead of a blue one.
Lock feature against unauthorized use
The Leica ASP300 provides a feature called "Lock" to prevent ac­cess to any operations of the Leica ASP300 to unauthorised users.
The ‘Lock’ mode and the corresponding ‘Lock’ password are config­ured from the SYSTEM SETUP menu.
Leica ASP300
27
4. Setting up the instrument

4.8 Check list for first operation (at supervisor access level)

Screen
Press button
→→
→→
→→
or
Parameter settings
Enter instrument name and select lan­guage. Activate printer (optional).
Enter carbon filter age warning and pump warning threshold.
Set all parameters as required, especially wax bath set temperature. Check whether date and time are correct.
Add the desired reagents and edit the ‘REPLACE REAGENTS’ warning thresholds.
Allocate a reagent to each of station.
Fill the reagent stations (bottles), either via the ASP300 remote drain/ fill function (SMART SCREEN) or manually (REAGENT STATUS).
28
→→
Create the programs you need. Programs can be copied and modified.
Allocate the most popular programs to the FAVORITES menu and, if desired, set a fin­ish time for these programs and allocate a symbol.
Instruction manual V 1.3 – 12/2003

5.1 Reagent handling

5.1.1 Fill/drain reagents (other than wax)

1. Manually changing the reagents
Remove the appropriate reagent bottle from the Leica ASP300 re-
agent cabinet, remove the bottle screw cap.
Drain/refill the reagent by pouring to/from a bulk container.
On the REAGENT STATUS screen, verify correct allocation and set the reagent bottle as full.
2. Use the ‘Remote Fill / Drain’ features of the Leica ASP300

5. Operation

Connect the supplied remote fill/drain hose to
the remote drain connection (refer Fig. 11).
Important! Ensure that the hose is pushed all of the way onto the connection, until a clear click is heard
Leica ASP300
Put the loose end of the remote fill/drain hose into the desired container for the drain/fill operation as desired.
Fig. 11
29
5. Operation
Fill / drain reagents
In FAVORITES, press MENU. In MENU FUNC­TIONS, press SMART SCREEN.
The SMART SCREEN (Fig. 12) pops up.
Fig. 12
Select station Press key symbol
30
Remote drain
On the screen, select reagent bottle no. 3 and the bulk container (Remote). The press STATION DRAIN.
Check for leaks.
Remote fill
On the screen, select reagent bottle no. 3 and the bulk container (Remote).
The press STATION REFILL.
The remote fill procedure takes less than 3 minutes.
Check for leaks.
Instruction Manual V 1.3 – 12/2003

5.1.2 Changing the wax

1
Draining the wax
5. Operation
2
Fig. 13
Connect the paraffin drain hose (2) as shown in Fig. 13 and insert the loose end of the hose into a remote collecting container.
+
Make sure the hose end is securely kept in the remote container until both the drain has been completed and the hose has been purged with pressurized air.
From the FAVORITES screen, press SMART SCREEN (see previous chapter).
Select the wax bath to be drained and the remote container (‘Remote’). Then press STATION DRAIN.
Several queries will be displayed each of which has to be confirmed individually.
Leica ASP300
31
5. Operation
Filling with wax pellets
1
2
Fig. 14
Remove the splash protector.
Fill the wax bath (Fig. 15) with wax pellets up to the top.
Press BACK and REAGENT STATUS to ac­cess the reagent status table (Fig. 14). High­light the wax bath you just filled and press SET AS FULL.
After about 90 minutes, add another 0.5 kg of wax pellets.
The pellets will take a total of approx. 10 hours to melt.
Important! During the melting process, the paraf­fin level MAY NOT drop below the MIN level (2, Fig. 15).
Once the paraffin is completely molten, re­insert the splash protector.
Software control
The software calculates when the wax will be ready for use and selects program start­ing times accordingly, i.e. the paraffin will definitely be molten when it is required for the first paraffin step.
32
Fig. 15
Instruction Manual V 1.3 – 12/2003
Adding molten paraffin
5. Operation
Molten wax may have a maximum temperature of 70°C.
If filling with molten wax, do not fill above the MAX level (see Fig. 15).
After a wax bath has been filled, it must be marked as ‘full’ in the
REAGENT STATUS screen.
Software control
If the wax baths are filled with molten wax, it is possible to override the Leica ASP300 melting time calculation.
For that purpose, in supervisor mode, access the SERVICE FUNCTIONS screen (Fig. 16).
Press TIME UNTIL WAX MELTED.
Enter the required delay time and press OK to confirm.
Fig. 16
If the instrument attempts a paraffin step with the paraffin only partly molten, severe instru­ment malfunctions may occur.
Leica ASP300
33
5. Operation

5.2 Setting up the instrument parameters

5.2.1 System setup

From the start screen, press MENU to go to the MENU FUNCTIONS screen, then press MORE.
In the MORE MENU FUNCTIONS screen, press SYSTEM SETUP.
SYSTEM SETUP screen (Fig. 17).
34
Fig. 17
The DISPLAY/BUZZER field is practically self-explanatory:
The number next to DISPLAY OFF indicates the remaining time (in min) until the screen saver (after the last user action) is activated.
LCD CONTRAST can be set from 0 to 14.
The BUZZER VOLUME can be set between 1 and 10.
Instruction manual V 1.3 – 12/2003
5. Operation
Program Options
Selection of the options on how programs are carried out.
STATION ALLOCATION: By age or Sequential
By age - RMS is activated, reagents will automatically be
used in order of increasing cleanliness.
Sequential - reagents will be used in sequential order of stations.
WAX BATH ORDER: Auto or 1.; 2. ; 3.
Wax bath order can be activated only when STATION ALLOCATION is set to Sequential.
PROMPT FOR NUM BLOCKS : Enabled or Disabled
Enabled- When starting a program, the actual number of
blocks must be entered. This is recommended with the RMS being enabled.
Disabled - Number of blocks can be entered optionally (not
mandatory).
Leica ASP300
Date / Time
Be sure to verify that the date and time entered do actually correspond to the local time/date.
Press the buttons to access the entry windows where time/date can be corrected, if necessary.
Instrument
Press WAX BATH SET TEMP (°C) to access the wax bath temperature entry window. Select the highest temperature compatible with the type of paraffin used as paraffin temperature will drop slightly once the paraf­fin is in the retort.
35
5. Operation
Security
To determine the level of access to the instrument.
SUPERVISOR PASSWORD: Enabled or Disabled Enabled - A password is required for access to the supervisor
level.
Disabled - All instrument and software functions are fully
accessible to all staff.
LOCK MODE: Enabled or disabled. Enabled - No screen functions can be accessed.
To unlock, a password has to be entered.
Disabled- Screen can be accessed at any time until the screen
saver turns off the screen.
Entry keyboard
The keyboard is displayed any time text needs to be entered. The key­board headline (1) tells you what kind of text to enter. 30 characters can be entered into each entry field, though sometimes not all characters entered can be displayed.
36
1
Important keys Shift:
To shift to upper case symbols. AltGr :
For access to a number of special characters.
<-- : Deletes last character entered.
Clear: Deletes entire line.
Fig. 18
Instruction manual V 1.3 – 12/2003

5.2.2 Editing the reagent list

Adding new reagents
1. Enter the reagent name and allocate the new reagent to the reagent
2. Define the stations (bottles) that will be filled with the new reagent.
3. Fill the stations (bottles) according to the reagent list.
Enter reagent names
Enter reagent names via the REAGENTS screen. You must be logged on at supervisor access level to proceed.
From the start screen, press MENU to go to the MENU FUNCTIONS
Press REAGENTS.
The SET UP REAGENTS AND WARNING THRESHOLDS screen opens.
To add a reagent:
5. Operation
group it belongs to.
screen, then press MORE to go to MORE MENU FUNCTIONS.
Press INSERT.
Enter the new reagent name.
Press OK to confirm.
You will then automatically be prompted to select the reagent group:
Leica ASP300
Select the reagent group
Allocate the new reagent to the desired group and press OK to confirm.
Fig. 19
37
5. Operation
Allocation of reagents to the correct reagent group is the basis for compatibility monitoring. Allocation to the wrong reagent group can lead to reagent cross contamination.
Changeable parameters
Entering / modifying reagent thresholds
Highlight the reagent to be modified.
Press the headline key symbol of the parameter to be changed.
Enter a new threshold or - if no warning is desired - press CLEAR to remove the threshold altogether.
Press OK to confirm.
The same threshold value applies for all reagent stations con­taining the same reagent.
Changing reagent names or reagent groups
Highlight the reagent.
Press the corresponding headline key symbol.
In the entry window (or via the keyboard), enter the new reagent group allocation.
Press OK to confirm and save.
Reedit all station and programs allocated to the reagent changed.
38
Instruction manual V 1.3 – 12/2003
5. Operation
Deleting reagents
Highlight the reagent to be deleted in the SET UP REAGENTS AND
WARNING THRESHOLDS screen.
Press DELETE.
Press OK in the screen prompt.
Please remember that a reagent which is already used in a pro­gram cannot be deleted.
Adding new reagents to stations
Go to SMART FUNCTIONS to remote-fill the reagent bottle.
If a bottle is filled manually, it must be defined as full afterwards. For that purpose:
Press MENU to access the MENU FUNCTIONS screen.
Press REAGENT STATUS.
In the reagent list, highlight the station / reagent that has been filled
manually.
Press SET AS FULL to mark the station as ‘full’.
Leica ASP300
Upon activation of the RMS all warning thresholds for the se­lected reagent are automatically reset to ‘0’.
39
5. Operation

5.2.3 Viewing the program list

This list (Fig. 20) displays all programs currently existing in the ASP300.
You can create:
up to 15 processing programs,
3 retort cleaning programs,
1 wax cleaning program.
At supervisor access level:
the processing program names can be edited,
new processing programs can be created and existing ones can be deleted.
Fig. 20
Important! New processing programs are created by copying existing ones. Therefore, the list must contain at least one program at all times.
40
Program duration cannot be edited directly.
To alter the duration of a program, the duration of one or more indi­vidual program steps has to be modified.
Retort and wax cleaning programs are predefined and cannot be altered.
Instruction manual V 1.3 – 12/2003

5.2.4 Adding and/or modifying programs

Creating a new program
You must be logged on at supervisor level.
In the VIEW/EDIT PROGRAMS screen (Fig. 20) highlight a program as
similar as possible to the program you wish to create.
Press COPY to copy the selected program.
The new program will have the same name as the program copied, however, the figure “(2)“ will be added.
Highlight the line containing the new program.
Press PROGRAM NAME.
Enter the new program name.
Editing program steps
Press EDIT (refer Fig. 20) to access the PROGRAM STEPS screen.
In the headline you will find the PROGRAM NAME.
The colors on the left border of the table indicate the reagent groups
to which the reagent of the step belongs.
The program steps (max. 13 steps per program) are displayed in the
order in which they are carried out.
The following characteristics can be edited:
reagent name
duration of step
(with the exception of fill and drain steps)
retort temperature
(if “Ambient“ is selected, the retort temperature indication remains empty)
type of pressure and/or vacuum cycle
retort drain time
delay step.
5. Operation
Leica ASP300
41
5. Operation
Editing program steps
To edit a step, highlight the corresponding line and press the re­spective headline.
In the entry windows that pop up enter / select the program step set­tings.
42
↓↓ ↓
Fig. 21
Instruction manual V 1.3 – 12/2003
Editing program steps (continued)
Setting a delay step
A so-called ‘delay step’ ensures that a program is finished at a specified time.
Highlight the program step you wish to define as a delay step.
Press DELAY.
The delay symbol is moved to the selected step, thus defining the step as delay step.
Copying steps
Highlight the step to be copied.
Press COPY.
Use the MOVE UP / MOVE DOWN symbols to move program steps up
or down within an existing program without having to recreate those steps.
5. Operation
Leica ASP300
DELETING PROGRAM STEPS
Highlight the step to be deleted.
Press DELETE.
43
5. Operation

5.2.5 Favorites (maximum of 5 programs)

Favorites can be programmed to end either:
“as soon as possible“ (ASAP) or
at a predetermined finish time.
Adding and/or modifying Favorites
Highlight the program line to be altered.
Press PROGRAMS.
Select the desired program and press OK to confirm.
Fig. 22
44
Setting a finish time
Press FINISH TIME.
Enter the finish time (24-hour clock). Select ASAP, if the program finish time is not to be delayed.
Press OK to confirm.
Instruction manual V 1.3 – 12/2003
Assigning a finish day
5. Operation
Press DAY.
Select the desired week day in the list and press OK to confirm.
Assigning an icon to a “Favorite“
Press SYMBOL in the table header.
Select the desired icon and press OK to confirm.
The first icon on the list is empty - select this option, if you do not want to assign an icon to a specific program.
Leica ASP300
Deleting a ‘Favorite’
Highlight the Favorite to be deleted.
Press DELETE.
The ‘Favorite“ is deleted from the FAVORITES list only - the program itself is not deleted.
45
5. Operation

5.2.6 Editing the stations

Fig. 23 shows all stations / allocated reagents cur-
rently defined in the instrument.
Adding / changing reagents
Highlight the desired station.
Press REAGENT NAME.
The SELECT REAGENT entry window pops up.
Only those reagents that are compat­ible with the respective station will be available for selection.
Highlight the desired reagent and press OK to confirm.
46
Fig. 23
Allocating reagents
Only processing reagents (except paraffin) can be allocated to stations 1 – 10.
Only a cleaning reagent can be allocated to station 11.
Only a cleaning alcohol can be allocated to station 12.
Only cleaning water/detergent can be allo­cated to stations 13 and 13-ext.
Only paraffin wax can be allocated to the wax baths.
Instruction Manual V 1.3 – 12/2003

5.2.7 Reagent groups

5. Operation
In this screen (Fig. 24) you find the colors used for each reagent group in the graphic program display.
Leica ASP300
Fig. 24
Selecting a new color for a reagent group
Highlight the line containing the reagent color you wish to change.
Press COLOR in the table header: the entry window SELECT COLOR pops up.
Select the desired color and press OK to confirm.
If you want to close the entry field without changing the color, press CANCEL.
47
5. Operation

5.3 Programs

Programs can be started from two different screens:
FAVORITES (max. 5 programs)
Fig. 25 Fig. 26
A ”Favorite” is a processing program that is frequently used. All settings are already programmed. Only the number of blocks has to be entered (if this function is enabled).
Starting a “Favorite“
To start a Favorite, press the corresponding symbol on the Favorites screen - the program will be started immediately.
ALL PROGRAMS
In the Favorites screen, press ALL PROGRAMS.
Press the symbol corresponding to the pro­gram you wish to start.
A graphic representation of the stations / retort will be displayed with the stations re­quired for the program you are about to start colored in the corresponding reagent group colors (Fig. 27).
Prior to starting it, modify the program as required. All processing step modifications made will only be applied to the current program.
48
Instruction Manual V 1.3 – 12/2003

5.3.1 Editing a selected program

Fig. 27
To modify a program prior to starting it, press EDIT to access the Edit Program screen.
5. Operation
Redefine the program finish time:
Press FINISH TIME. In the entry window, set
a new finish time.
Leica ASP300
Fig. 28
Enter the day when you want the program
finished.
Enter the finish time (24-hour clock format).
Press OK to confirm.
Fig. 29
49
5. Operation
Access the RUN OPTIONS function to display the options selected for the current program.
LEVEL CHECK enabled: The lower level sensor in the retort is acti­vated.
RECIRCULATION enabled: During each processing step, the reagent is pumped periodically from the bottle to the retort and back for continuous and efficient mixing of fluid. The first pump step is carried out after 16 minutes, all subsequent pump steps in 12-minute intervals.
PRETEST enabled: After starting the program a test run is per­formed: the first reagent is pumped into the retort and back into the bottle.
STIRRER enabled: Magnetic stirrer is on.
50
PRINTER enabled: At the end of the program run, a full run log will be printed out.
Fig. 30
Important! The options selected will be applied to ALL programs! All run options can be enabled / disabled during the program run if necessary.
The printer option will be displayed only if a printer has been configured for the instrument.
Instruction Manual V 1.3 – 12/2003

5.3.2 Starting a program

5. Operation
Fig. 31
Press START to start the program. Press OK to confirm the query that follows. Enter the number of blocks, if the NUMBER OF BLOCKS op­tion has been enabled.
Leica ASP300
The FINISH TIME button shows the time when the current program will be finished.
Before a program in progress can be modified, the PAUSE symbol must be pressed. To modify or delete program steps, press EDIT and, in the entry win­dow, carry out any changes you wish to make. Press CONTINUE to re­start the program.
The finish time of a program in progress can only be changed as long as the delay step has not yet been completed.
Press TABLE to change from graphic to tabular program display, press GRAPHICAL to go back to graphic display.
51
5. Operation

5.4 Reagent status

In the reagent status screen (Fig. 32) you can display and update the actual reagent status of each station and you can set the stations as full or empty.
Warning thresholds that trigger warning messages appear on a red background.
To change a reagent status, highlight the corresponding line and press the corresponding key symbol .
Reagent warnings
If the RMS is enabled, the instrument will give you warning messages when reagents have been used for too long.
Warning messages for normal reagents are given based on up to 3 criteria; warning mes­sages for paraffin wax on up to 6 criteria.
The warning messages are displayed at the end of a cleaning cycle. From there, you can go directly to the REAGENT STATUS screen.
In addition, warning messages (on a yellow background) are displayed in the program start screens.
52
Fig. 32
Criteria for warning messages
For normal reagents, the following warning cri­teria apply:
number of blocks processed
number of programs carried out
number of day lapsed since the reagent has last been changed.
For paraffin wax, three additional criteria can be applied:
number of blocks processed
number of programs carried out
number of days lapsed since the wax has last been cleaned.
Instruction Manual V 1.3 – 12/2003

5.5 System Monitor

5. Operation
After replacing the filter press RESET.
ROTARY VALVE RETRIES:
If a figure higher than 50 is indicated, a preventive maintenance is recommended.
The SYSTEM MONITOR screen supplies important infor­mation on the Leica ASP300 system status.
PUMP AGE:
Indication of hours of service of the pressure/ vacuum pump. After approx. 1,000 hours of use, have Leica Technical Service carry out a preventive maintenance of the pump.
CARBON FILTER AGE:
Indication of hours of service of the activated carbon filter. The carbon filter must be ex­changed after approx. 4 months of use.
Fig. 33
Leica ASP300
FILL/DRAIN:
This screen displays the average fill / drain times of all stations (in seconds).
Short term = average duration of last 5 fills / drains. Long term = average duration of last 20 fills / drains.
If one of the figures displayed exceeds 250, call Leica Technical Service.
53
5. Operation

5.6 On-line help

The Leica ASP300 features an extensive on-line Help system. For information on basic problems refer to the Help Index (to access
press INDEX on the Help screen).
Use the UP/ DOWN arrows to scroll.
Press PREVIOUS TOPIC (1) to return to the index.
Press BACK (2) to go back to the screen from where you accessed Help.
CUSTOM HELP (3) is only displayed, if user­specific help files have been installed.
23
1
Fig. 34
Detailed information on how to create and install customer-specific help files can be found in the “What is Custom Help?“ help text.
Accessing ‘Help’ in the event of error messages
Press DETAILS, or
Press the HELP button on the error message.
54
Instruction Manual V 1.3 – 12/2003

5.7 Configuring a printer (optional)

5. Operation
With a Centronics standard cable, connect the printer to
Printing run logs
Printing is possible when the PRINT symbol is displayed.
To select and configure a printer, proceed as follows:
the printer port (refer chapter 4.4, Fig. 6).
Leica ASP300
For the PRINT symbol to be displayed, a printer must have been configured for the instrument.
The following lists and protocols can be printed:
reagent list,
stations list,
reagent status list,
all programs,
run log,
error log.
55

6. Trouble shooting

6.1 General

The Leica ASP300 is equipped with a trouble shooting system. If an error is detected, detailed error messages and corresponding user
instructions on what to do are displayed on the touch screen.
For all correctable error conditions the instrument will take corrective actions to protect the specimens.

6.2 Power failure

Details of all events are recorded in “Run Log”. Run log can be viewed by pressing MENU and then RUN LOG.
Power failure during a processing program
In the event of a power failure, the instrument automatically compen­sates any impact the power failure might have on the specimens.
Power failure during a paraffin or retort clean cycle
If a power failure occurs during a clean cycle, the total duration of the cycle will be extended by the duration of the power failure.
56
Instruction manual V 1.3 – 12/2003

6.3 Trouble shooting

6. Trouble shooting
Check for error messages indicating the nature of the problem.
Press the Help button on the message for any information relating to
the cause.
Check the Run Log for any information on the nature of the failure.
Use the SMART SCREEN functions to individually test opera­tions such as fill, drain, pressurizing and evacuation operations.
Power failure
Verify that the mains lead is plugged in to the wall socket, and that, if
applicable, the wall socket is switched ON.
Verify that the jumper lead or UPS lead are fully inserted into their re-
spective socket.
Verify that the power switches on the rear (adjacent to the power
lead) and on the side of the instrument are both switched ON.
Leica ASP300
If, by carrying out the above listed steps, you cannot identify the problem, call Leica Technical Service.
57
6. Trouble shooting

6.4 Typical fill or drain problems

Fill and drain problems may be due to several causes:
1. There is insufficient reagent
Verify that both the reagent bottles and the wax stations have
sufficient reagent.
2. Insufficient pressure or vacuum
The Leica ASP300 is unable to create sufficient vacuum (for a fill) or pressure (for a drain) in the retort.
Verify that the reagent bottle in question is correctly inserted in
the reagent cabinet.
Verify that all reagent bottles are fully pushed home into their
sockets.
Check whether there is some debris under the retort lid seal
causing a leak.
3. There is a blockage in the air or reagent lines
If there is reagent in the retort that cannot be drained, try to remove the blockage as follows:
Heat the retort to the maximum possible temperature (consistent
with the reagent currently in the retort, if any) and leave it at this temperature for at least 15 minutes.
Whilst the retort is at maximum temperature, attempt to fill or
drain.
If the retort is empty:
Run the extended retort clean program. Use the maximum set­tings for number of clean cycles and temperature (consistent with the type of reagents/solvents used in the clean program).
58
If you cannot eliminate the problem by carrying out the above listed steps, call Leica Technical Service
Instruction manual V 1.3 – 12/2003

7.1 Clean programs

7.1.1 Retort clean programs

7. Cleaning and maintenance

You find the clean programs at the end of the programs list (Fig. 35). To display a clean program, highlight the re­spective line and press EDIT.
The ASP300 offers three retort clean programs and one paraffin clean program. Clean programs cannot be copied or deleted.
A clean program comprises a minimum of three and a maxi­mum of four steps:
1. Cleaning Solvent 1 step from station 11
2. Cleaning Alcohol 1 step from station 12
3. Cleaning Water 1 step from station 13
Leica ASP300
Optional step between steps 2 and 3: Cleaning Water / Detergent 1 step from station 13-ext.
Fig. 35
The steps of retort clean programs are listed in the order in which they are carried out.
The clean programs skip any steps not required to complete the re­tort clean.
59
7. Cleaning and maintenance
Editing steps
To edit a program step highlight the corresponding line and press the respective table header.
To disable a step, set the number of cycles to “0“. Only cleaning wa­ter steps can be disabled.
The following characteristics of each step can be edited:
Reagent name - only those reagents that are compatible with the re­agent station being edited will be displayed.
Number of cycles - the duration displayed is calculated based on the number of cycles.
Retort temperature - temperatures from 50 to 65 °C can be set.
Station 13-ext
The term “13-ext“ stands for an additional station which can be used instead of station 13.
Station 13 always contains cleaning water.
If a program contains a “13-ext“ step, during the clean cycle the op­erator is prompted to insert bottle “13-ext“ when needed, and to re­move the bottle as soon as the step has been completed.
60
Only reagents belonging to the cleaning water group can be al­located to station 13-ext. Station 13-ext may be used for clean programs only.
Instruction manual V 1.3 – 12/2003
7. Cleaning and maintenance

7.1.2 Wax clean programs The Leica ASP300 contains a wax clean fea-

ture that removes solvents (xylene) from used wax.
Follow the instructions on the left (Fig. 36) to run a wax clean program.
Leica ASP300
Fig. 36
The duration of the wax clean program is user-programmable.
Follow the instructions in Fig. 37.
Fig. 37
61
7. Cleaning and maintenance

7.1.3 Smart clean

The Smart Clean prepares the ASP300 for transport. Prior to starting a Smart Clean, a complete retort clean program must be
run. To start a Smart Clean, press SMART CLEAN on the MORE MENU FUNC-
TIONS screen and press OK to confirm. The Smart Clean takes approx. 15 minutes and consists of 3 phases:
Phase 1:
Make sure all reagent containers are inserted correctly. All internal tubing is cleaned with pressurized air.
Phase 2:
Before phase 2 starts, the operator is prompted to remove all reagent bottles from the instrument. All internal tubing is again cleaned with pressurized air.
Phase 3
Before phase 3 starts, the operator is prompted to insert an empty bottle into station 11. The tubing is cleaned with pressurized air for the third time - remaining rests of reagents, if any, are forced back into station 11.
62
Empty and clean all reagent bottles thoroughly before reinsert­ing them into the instrument to make sure there will be no re­agent spills from the bottles that could cause damage during transport.
The SMART clean is ideal for removing residual liquid from the internal pipes of the instrument, as it thoroughly purges them with pressurized air. A SMART clean should be carried out once every month.
Instruction manual V 1.3 – 12/2003

7.2 General cleaning steps

Paraffin wax baths
Wipe the wax baths and lids clean.
The lid can be removed for cleaning purposes.
If they are dirty, remove the wax strainers from the wax baths. Clean,
dry and reinsert them.
Ensure that the air vent hole at the top rear right hand corner is un-
obstructed.
Instrument exterior
Clean the instrument exterior as necessary. Wipe with a damp cloth moistened with mild detergent and dry.
7. Cleaning and maintenance
Warning! Be careful as the walls of the wax baths are very hot and may cause burns!
Do not use solvents on painted surfaces and/or the touch screen!
Checking the drip tray
Leica ASP300
Regularly check the drip tray for spilled reagents (refer Fig. 38).
Fig. 38
63
7. Cleaning and maintenance
Draining the condensate bottle
Exchanging the activated carbon filter
Remove and empty the condensate bottle and reinsert (Fig. 39).
Fig.39
Exchange the filter as shown in Fig. 40.
Replace the filter at least every four months – make sure to insert correctly – (see arrow on label)!
Perform a Smart Clean prior to changing the filter.
64
Fig. 40
Filter label
Instruction Manual V 1.3 – 12/2003

7.3.1 Daily cleaning and maintenance

Cleaning the retort lid
Remove wax from the inside of the retort lid with the plastic scraper provided. Make sure to remove all wax from around the retort lid seal.
For convenience, the lid may be removed during cleaning by lifting it vertically and sliding it towards the console.
Cleaning the retort
The retort may be wiped clean using a cloth moistened with solvent (xylene or alcohol) or mild detergent. Ensure that the air holes at the top front of the retort are open - remove all dirt!
Cleaning the retort strainer
Use alcohol or xylene to clean the strainer located at the bottom of the retort. For convenience, the strainer may be taken out to remove all solid dirt.
Cleaning the touch screen
On the MENU FUNCTIONS screen press CLEAN TOUCH SCREEN.
Clean the touchscreen.
Press ENABLE EXIT and then BACK to return to MENU FUNCTIONS.
7. Cleaning and maintenance
Leica ASP300
Cleaning the processor module top surface
Remove both lids for cleaning.
First scrape the stainless steel surface with the plastic scraper and then wipe down.
Reagent and condensate bottle seal lubrication
To ensure easy removal of the reagent and condensate bottles, lu­bricate the O-ring seals on the plug-in nozzles with the O-ring lubri­cant supplied.
Check regularly whether the reagent bottles are still properly con­nected to the connection manifolds in the instrument. Retighten, if necessary.
65
7. Cleaning and maintenance
This procedure is particularly important with bottles containing chloroform. Chloroform tends to swell the bottle seals, making them difficult to remove. This problem can be prevented by regular lubrication.
Checking the retort and wax bath lid seals
Regularly check the retort and wax bath lid seals for damage. If a seal is damaged, it must replaced without delay.

7.3.2 Periodic cleaning and maintenance

Cleaning the reagent bottles
Remove the reagent bottles from the reagent cabinet. Empty and clean them using a bottle brush and a laboratory deter-
gent in warm water.
Never clean the reagent bottles in an automatic dishwasher.
66
Wipe the stainless steel inner walls of the reagent cabinet with a damp cloth moistened with a mild detergent.
Refill and reconnect the bottles once cleaned. Make sure the bottle lids are tight.
Make sure the bottles are properly pushed home into the connection manifolds at the rear inner wall of the reagent module.
Instruction manual V 1.3 – 12/2003
7. Cleaning and maintenance

7.4 Preventative maintenance schedule

Maintenance activity daily weekly monthly every
4 months
Lubricate reagent bottle O-rings √√ and check for damage.
Wipe dry retort and lid after completing retort clean cycle.
Clean exterior of instrument with soft cloth and minimal amount of xylene.
Check retort strainer for any tissue or wax debris.
Remove residual paraffin from inner surface of wax bath lid.
Inspect and, if necessary, clean retort lid seal.
Inspect and, if necessary, clean wax bath lid seal.
Inspect and, if necessary, adjust paraffin levels.
Inspect reagent container fluid levels.
* if the automatic fill and drain function is actually being used
Leica ASP300
67
7. Cleaning and maintenance
Maintenance activity daily weekly monthly every
4 months
Inspect and, if necessary, clean wax bath strainers .
Inspect wax bath air vent and,
if necessary, clear.
Inspect and empty condensate bottle.
Clear inlet holes.
Examine activated carbon filter.
Replace carbon filter.
emove fluid level sensor (float),
R
inspect and clean.
Inspect electrical connectors at the rear panel of the instrument.
Inspect and, if necessary,
clean inside reagent bottles.
Check that the bottle caps and seals (O-rings) of the reagent bottles are in place and tight.
Check that the reagent bottles are fully inserted in the connection manifolds.
68
Instruction manual V 1.3 – 12/2003
Warranty
Leica Microsystems Nussloch GmbH guarantees that the contractual product delivered has been subjected to a comprehensive quality con­trol procedure based on the Leica in-house testing standards, and that the product is faultless and complies with all technical specifications and/or agreed characteristics warranted.
The scope of the warranty is based on the content of the concluded agreement. The warranty terms of your Leica sales organization or the organization from which you have purchased the contractual product shall apply exclusively.
Technical service information
If you require technical service or replacement parts, please contact your Leica sales representative or dealer who sold the product.
Please provide the following information:
Model name and serial number of the instrument.
Location of the instrument and name of the person to contact.
Reason for the service call.
Date of delivery.

8. Warranty and service

Decommissioning and disposal
The instrument or parts of the instrument must be disposed of in compli­ance with the local laws.
Leica ASP300
69
Notice
70
Instruction manual V 1.3 – 12/2003
Loading...