Leica 13104810, 13104940, 13104950, 13104800 User Manual

Leica Abbe Mark II
Refractometer
Models 13104800/13104810/
13104940/13104950
Instruction Manual
Gebrauchsanleitung
Manuel d’utilisation
1.0 INTRODUCTION.................................................................. 3
2.0 UNPACKING........................................................................ 3
3.0 POWER REQUIREMENTS.................................................. 3
4.0 SET UP .................................................................................. 3
4.1 Equipment Installation........................................................ 3
5.0 CONTROLS............................................................................ 3
5.1 Variable Conditions in Taking Readings .......................... 4
5.2 Calibrating With Distilled Water ...................................... 4
6.0 OPERATING INSTRUCTIONS .......................................... 4
7.0 LEICA ABBE MARK II PLUS REFRACTOMETER............ 5
7.1 Computer Interface............................................................ 5
7.2 Setting the Baud Rate ...................................................... 5
7.3 Printer Interface ................................................................ 5
7.4 Custom Programming........................................................ 5
8.0 REFRACTOMETER USE, TECHNIQUES AND
REFRACTION PRINCIPLES ................................................ 6
8.1 High Index/High Dispersion Samples ............................ 6
9.0 CALIBRATION WITH TEST GLASS.................................. 6
10.0 ILLUMINATOR BULB REPLACEMENT ..............................7
10.1 Fuses .................................................................................... 8
11.0 CARE AND CLEANING ........................................................8
12.0 SERVICING ............................................................................8
13.0 SPECIFICATIONS..................................................................8
13.1 The following specifications refer only to the
LEICA ABBE MARK II PLUS Refractometer.................... 9
Explanattion of Warning Symbols
1.0 INTRODUCTION
NNOOTTEE::
Please read these instructions carefully before attempting
to use the instrument.
This Instruction Manual is designed specifically for the LEICA ABBE MARK II Refractometer and describes proper set up, use and routine maintenance of the instrument.
SSeeccttiioonnss 77..00 -- 77..44 aarree
aapppplliiccaabbllee oonnllyy ttoo tthhee LLEEIICCAA AABBBBEE MMAARRKK IIII PPlluuss RReeffrraaccttoommeetteerr.. The LEICA ABBE MARK II Refractometer is a durable bench-top
instrument. It provides efficient, reliable and repeatable readings in three modes:
• % Solids (
O
Brix)
• % Solids (
O
Brix) temperature compensated or (BX-TC) and
• Refractive index (nD) The LEICA ABBE MARK II Plus Refractometer can accommodate
a custom program on the % Solids (
O
Brix) scale. This can be ac-
complished with pre-programming or retrofitting at the factory.
2.0 UNPACKING
The LEICA ABBE MARK II Refractometer has been carefully pack­aged for shipping. Before setting aside the packing materials, check the box(es) to assure that all items ordered have been received. Among the accessories in the refractometer box will be:
• a refractive index test glass (Catalog No. 10451)
• a 1/16" ( No. 4) Allen wrench (Catalog No. X30600)
• a bottle of 1- Bromonaphthaline for measuring solids (Catalog No. 10452)
• a dust cover (Catalog No. 10450-140)
Check the instrument and model number against the original order before discarding any packing materials. The original packing materials may be saved for future storage or transporta­tion of the refractometer.
3.0 POWER REQUIREMENTS
Models 13104800/13104940 115~ +/-10%, 50/60 Hz Models 13104810/13104950 230~ +/-10%, 50/60 Hz
• Input power 25P max
• Indoor use only
• Altitude up to 2000m
• Temperature range 5
O
C to 40OC
• Pollution degree 2
• Installation category 2
• Max relative humidity 80% for temperatures up to 31OC
decreasing linearly to 50% relative humidity at 40OC.
4.0 SET UP
The LEICA ABBE MARK II Refractometer is easy to use. Very little training is needed to set up and operate the instrument to obtain temperature and refractive index measurements.
4.1 Equipment Installation
The LEICA ABBE MARK II Retractometer must be used on a hard, stable and level surface such as a laboratory table or work bench.
5.0 CONTROLS
THE LEICA ABBE MARK II Refractometer controls are identified in Figure 1.
NNOOTTEE::
The mode selector switch no longer functions as the on/off switch. The ON/OFF switch is located on the lower left front cover of the instrument above the Leica logo.
PPoowweerr SSwwiittcchh:: The Power Switch is located in the lower right corner of the front
panel (Figure 1). | = on 0 = off
English
Leica ABBE MARK II Refractometer Instruction Manual Contents
5X20 250mA T/250V (Type 10480/10494)
CAUTIONN:
Risk of electric shock
CAUTIONN:
(Refer to accompanying documents)
!
CAAUTION:
As a safety precaution, the power cord has been grounded to the instrument base. Never use a 2-prong adapter between the power cord and the power source; it will render the grounding feature ineffective.
CAAUTION:
Use of this equipment other than in the man­ner specified in this manual may result in reduction of safety protection.
!
3
TThhee mmooddee sseelleeccttoorr sswwiittcchh iiss uusseedd ttoo sseelleecctt mmeeaassuurreemmeenntt mmooddeess:: TThhee mmooddeess aarree::
• % Solids-TC - Temperature Compensated % Solids as Sucrose (
O
Brix) per International Commission for Uniform Methods of
Sugar Analysis Refractive Index Tables.
•nD- Refractive Index
• % Solids - Non-Temperature Compensated % Solids as Sucrose
The reticle adjustment access hole permits vertical adjustment of the crosshair during calibration procedures.
Depress the read display button to obtain measurement values. Depressing the temperature display button will provide
a reading of the prism temperature. Readings are shown digitally in the display window.
5.1 Variable Conditions in Taking Readings
Distilled water readings above 20OC can be obtained in the refractive index % Solids-TC mode, assuming the instrument is properly calibrated. Correct readings cannot be obtained in the non-temperature compensated % Solids mode when the water sample temperature is above or below 20
O
C. Refractive index (nD) readings for distilled water can be obtained from 0OC to 80OC (see Chart 1 for values between 18-30OC).
Refractive index is inversely proportional to temperature. As temperature increases, the solution density decreases and the refractive index or % Solids reading will decrease. As tempera­ture decreases the solution density will increase and the refractive index (% Solids) will increase. The % Solids-TC mode automatically corrects the reading for temperature based on the temperature coefficient of sucrose in solution. This coeffi­cient is typical of most aqueous solutions.
5.2 Calibrating With Distilled Water
To calibrate:
1. Place a distilled water sample on the measuring prism surface and let it stand undisturbed for 3-5 minutes to obtain temperature stability. Turn the mode selector switch to the refractive index (nD) position.
2. Focus the eyepiece until the crosshair clearly appears.
3. Rotate the adjustment control clockwise until the shadow line nearly meets the crosshair. Use the dispersion correction wheel to obtain the sharpest possible color-free shadow line. Adjust the
shadow line to meet the crosshair. Press the Read button and record the n
D
value, then press the temperature button and
record the temperature.
4. If the instrument is properly calibrated, the reading will be
1.3330 less 0.0001 for each degree above 20OC or plus 0.0001 for each degree below 20
O
C.
If the instrument is not calibrated, determine the proper reading (Chart 1) and rotate the adjustment control and depress the READ display button until the display reads correctly. (The shadow line will no longer intersect the crosshair on the reticle.)
5. Insert a 1/16" Allen wrench into the reticle adjustment access hole (Figure 2) located in the dispersion correction wheel. Turn the wrench to bring the crosshair into proper position on the shadow line.
NNOOTTEE::
The final movement made with the adjustment correction control wheel should always be in a clockwise direc­tion. The shadow line will move upward.
6. Turn the mode selector to the BX-TC position. Press the READ button. An LED reading of 000.0 or 0.1 is normal. If a negative sign and a decimal point appear, no reading can be given.
Go on to the next step. If a “No Reading” condition persists after repeated READ
attempts, rotate the adjustment control until the display reads
000.0 or 0.1. Only a very slight adjustment will be necessary. Readjust the reticle crosshair to the shadow line.
The display should now read 000.0. Repeat the steps until the display reads 000.0.
6.0 OPERATING INSTRUCTIONS
To begin operation:
1. Open the prism assembly cover (Figure 4) and remove the protective lens tissue. Clean the measuring prism surface with alcohol and then distilled water. Wipe with a clean soft cloth or lens tissue, but do not wipe the measuring prism surface while it is dry. Any residue left on the prism can affect the accuracy of readings.
2. Turn the mode selector to the desired mode position.
3. Apply a sample solution to the measuring prism surface using a glass dropper. Close the prism cover and position the illumination arm so that the exposed face of the upper prism is fully illuminated.
4. Turn the dispersion correction wheel so that the crosshair adjustment access hole is at the six o’clock position.
English
Dispersion Correction Wheel
Adjustable Eyepiece
Reticle Adjustment Access Hole
Display Window
Read Display Button
Temp. Display Button
Power Switch
Mode Selector
Prism Surface
Illuminator Arm
Adjustment Control
Right Panel Cover
Reticle Adjustment Access Hole
FFiigguurree 11
FFiigguurree 22
4
5. Rotate the adjustment control counterclockwise to position the shadow line at the bottom of the field of view.
6. Rotate the eyepiece to bring the crosshair into focus.
7. Move the shadow line to the crosshair reticle with the coarse adjustment control.
8. Rotate the dispersion correction wheel to eliminate any red or green color at the edge of the shadow line.
9. Turn the adjustment control to center the shadow line to the crosshair. The shadow line must be perfectly centered to obtain an accurate reading (Figure 3a).
10. Depress the READ button. The value of the test sample will be digitally indicated in the display window. Depressing the TEMP button will activate a temperature sensing device located in the measuring prism. The display will digitally indicate the actual temperature of the measuring prism and sample.
NNOOTTEE::
If the LEICA ABBE MARK II Refractometer is to be used in an environment where foreign materials may contaminate the display buttons, use the splash guard accessory (Catalog No.
13104590). It will prevent instrument damage and possible instru­ment failure.
7.0 LEICA ABBE MARK II PLUS REFRACTOMETER
NNOOTTEE::
Sections 7.0 through 7.4 are applicable only for the LEICA ABBE MARK II Plus Refractometer - Model 13104940/13104950.
The LEICA ABBE MARK II Refractometer provides computer interface, custom programming and printer interface capabilities.
The instrument becomes the LEICA ABBE MARK II Plus Refrac­tometer (Model 13104940/13104950) with the addition of the RS232 port. The LEICA ABBE MARK II Plus can be ordered as:
• an original purchase
• a factory upgrade or
• a retrofit to an existing LEICA ABBE MARK II Refractometer.
7.1 Computer Interface
Turn on the LEICA ABBE MARK II Plus Refractometer with the switch located on the instrument’s front panel. When the refrac­tometer is first turned on, the software version (
e.g.,
1.6) will appear in the read display for a few seconds. The refractometer is ready to use when this message disappears.
By interfacing the LEICA ABBE MARK II Plus to a computer, sample reading results can be transferred to IBM-PC compati­ble data-handling software programs with communications
capabilities (
e.g.,
Microsoft Works®, Framework II™ and
ProComm®)
1
. Software can be used to store, graph, analyze and print data. Leica also offers LEICA ABBE MARK II Plus Refracto­meter interface software (MARKPLUS) specifically designed for use with the instrument.
The MARKPLUS software allows the instrument user to repro­gram the % Solids scales from the default "Brix" (% Solids) scale to a user defined scale.
7.2 Setting the baud rate
The baud rate of the LEICA ABBE MARK II Plus is set to 9600. To change the baud rate remove the cap on the left side of the instrument (Figure 3b). Turn the instrument on and using the eraser end of a pencil press the post switch which is centered in the hole. The display will show the current baud rate. To change the rate press the read button on the front of the instrument panel until the appropriate baud rate is displayed. Choose the baud rate from the following options: 300, 1200, 2400, 4800, 9600. Once the appropriate baud rate is displayed press the switch again. Turn off the instrument and replace the cap. The new baud rate will be permanently retained in the instrument memory.
7.3 Printer Interface
The LEICA ABBE MARK II Plus Refractometer can be interfaced to a serial port printer for printed data results using the 9-pin RS232 output. A parallel port printer may be used with a serial to parallel converter.
7.4 Custom Programming
LEICA ABBE MARK II Plus Refractometer interface software (MARKPLUS) may be configured for customized calibration curves and temperature compensation. (The standard LEICA ABBE MARK II Refractometer corrects to 20OC only.) However, MARKPLUS must be configured for each specific application or custom programming. Leica’s MARKPLUS is designed to accept a Table of Refractive Index, Temperature and Concentration val­ues and then generate custom scale data for instrument pro­gramming. Data is entered from a text file. Contact your Leica representative for more information.
English
FFiigguurree 33aa FFiigguurree 33bb
Baud Rate Adjustment
RS232 Port
Left Panel Cover
5
8.0REFRACTOMETER USE, TECHNIQUES AND REFRACTION PRINCIPLES
Refractive Index measurements taken on the LEICA ABBE MARK II and LEICA ABBE MARK II Plus Refractometers are affected by temperature. Different values can be obtained when the same sample is measured at different temperatures. For example: as the temperature of a sugar solution sample rises, its density and refractive index decrease. This causes an apparent decrease in concentration when using the non-temperature compensated % Solids mode. The % Solids-TC (Temperature Compensated) mode automatically corrects the reading to the correct value at 20
O
C.
For every 1OC change in temperature, an aqueous solution con­taining sugar will change readings by approximately 0.08%. This is the average value for samples of 0 to 85% at 15OC to 45OC. As temperature rises the % Solids reading decreases. As tempera­ture decreases, the % Solids reading increases.
The LEICA ABBE MARK II Refractometer temperature will rise approximately 1OC every 10 minutes for 1 hour until it stabilizes. An instrument that has been turned on for more than 1 hour in a 20OC room could read 26OC. The rise in temperature is principally due to heat generated by the power transformer inside the instrument housing.
The sample placed on the measuring prism surface of a LEICA ABBE MARK II Refractometer that has been turned on for more than an hour in a 20OC environment will assume the temperature of the prism (
i.e.,
approximately 26OC). Although the sample is now warmer and would indicate a lower concentration when measured in the uncompensated Brix mode, the BX-TC mode automatically compensates for the temperature difference and displays the concentration of the solution as if its temperature were 20OC. Without compensation for the 6Orise, the readout of a 70% solution would be 0.5% less (6OC x 0.08%) or 69.5%.
The measuring prism surface temperature can be lowered by running cooling fluid through the water jackets in the prism housing. However, the heat generated inside the instrument housing still affects the prism temperature. For this reason it is necessary to run fluid slightly below the desired temperature. A thermometer placed in the fluid line may measure several degrees below the actual displayed temperature of the measur­ing prism surface displayed when the temperature button is pressed. The opposite is true if the prism temperature has to be elevated above the ambient temperature.
The fluid circulation system allows the prism to be adjusted for temperatures up to 80
O
C. This may be necessary when using
samples that solidify at lower temperatures. Measuring viscous samples, such as high Brix sugar syrups, may
be more difficult than measuring products with lower total dis­solved solids.
Stir the sample on the measuring prism surface with a rubber spatula or an angled piece of clear plastic tubing. Completely cover the surface making sure all the air pockets between the sample and the measuring prism surface are removed.
For removing carbonation from carbonated soft drink samples, open and close the prism cover two or three times after loading. Doing this provides the best possible reading line and the best way to judge where the shadow line intersects the crosshair reticle.
NNOOTTEE::
Remember the accuracy of the instrument is only as good
as the clarity of the shadow line.
8.1 High Index/High Dispersion Samples
High refractive index samples with high dispersion properties may require additional use of Catalog #13K30650 589nm bandpass filter accessory. This accessory will insure that the shadowline will be sharp for these samples.
9.0 CALIBRATION WITH TEST GLASS
Refer to Chart 1, Refractive Index of Triple Distilled Water at Various Temperatures to compare the refractive index reading with the calibrated temperature reading.
To calibrate the LEICA ABBE MARK II Refractometer use:
• a refractive index test glass (Catalog No. 13104510)
• a bottle of 1- Bromonaphthaline (Catalog No. 13104520)
• a 1/16" (No. 4) Allen wrench (Catalog No. 13X30600)
• lens tissue The refractive index test glass is imprinted with the refractive
index scale. The bottom and front surfaces of the glass are pol­ished. It is important that these surfaces are very clean when beginning calibration procedures.
TToo ccaalliibbrraattee::
1. Turn the “Mode” selector switch to the refractive index position.
2. Apply a single drop of 1-Bromonaphthaline to the surface of the measuring prism surface.
3. Place the refractive index test glass on the measuring prism surface with the imprinted side facing up. Position the polished end of the test glass toward the illuminator - about 1/8" (2mm) from the edge (Figure 4).
4. Align the illuminator so that the light is directed onto the pol­ished face of the test glass. Cover the illuminator lens with lens tissue to diffuse the light.
5. Adjust the illuminator arm, dispersion correction wheel, eye­piece and adjustment control so that the shadowline is centered, in focus, free of color. This is the same procedure used with a
English
Back Plate
Test Glass
Test Glass Position 1/8” [2mm] from edge)
Base Plate
Measuring Prism
Fastening Screws
Illuminating Prism
CChhaarrtt 11:: RReeffrraaccttiivvee IInnddeexx ooff TTrriippllee DDiissttiilllleedd WWaatteerr aatt VVaarriioouuss TTeemmppeerraattuurreess
TTeemmpp OOCC RReeffrraaccttiivvee IInnddeexx
18 1.3332 19 1.3331 20 1.3330 21 1.3329 22 1.3328 23 1.3327 24 1.3326 25 1.3325 26 1.3324 27 1.3323 28 1.3322 29 1.3321 30 1.3319
Source: Molar Refraction of Organic Compounds
FFiigguurree 44
6
liquid sample.
6. Depress the READ button and record the digital reading which appears in the display window.
If the reading differs from the value printed on the test glass by more than ±0.0001, repeat the measurement several times. Clean the measuring prism and test glass surfaces with alcohol, followed by water after each test.
If an error persists, make the following adjustment:
7. Turn the shadow line adjustment control slightly and depress the READ button. Repeat this procedure until the display agrees with the test glass value.
The shadowline will now be above or below the center of the crosshair reticle.
8. Insert the Allen wrench through the reticle access hole in the dis­persion correction wheel. Adjust the reticle until the shadow line passes through the center of the crosshair (Figure 2 & Figure 3a).
The instrument is now calibrated.
10.0 ILLUMINATOR BULB REPLACEMENT
The illuminator bulb can be replaced by removing the condenser sleeve (Figure 6). Always use the suggested replacement bulb (Catalog No. 13104620, pkg. of 3) for maximum efficiency and safety. After inserting a new bulb, the socket can be adjusted to center the light on the measuring prism with the condenser sleeve in place. If necessary, push down on the spring-loaded condenser sleeve lock pin and slide back and forth until the pin snaps into place.
DAANGER
: The glass envelope of the bulb may be extremely hot. Do not attempt to change the bulb before it is completely cooled or without wearing protection.
WAARNING
: Unplug the refractometer power cord before
attempting to replace the illuminator bulb.
English
Main Control Board (November 1992)
WWiirriinngg KKeeyy CCoolloorr CCooddee W = White B = Black G = Green O = Orange R = Red y = Yellow
11331100449944--992244
Not Used
Thermistor Plug
RS232 Plug (active on 13104940, 13104950)
Not Used
Limit Switch
Baud Rate Switch
Power
Motor
FFiigguurree 55
!
7
10.1 Fuses
The LEICA ABBE MARK II Refractometer electrical circuitry is protected by two (2) fuses. To replace the fuses turn the Adjust­ment Control Knob (Figure 1) two turns clockwise. This raises the elevator assembly to provide clear access to the fuses. Remove the adjustment control knob with a 1/8” Allen wrench. Remove the two Phillips head screws which hold the right panel cover to the bottom of the instrument housing. Slide the panel cover down as far as possible to facilitate its removal. With the right panel cover removed the fuses will be visible on the bottom plate of the instrument (Figure 5b). Replace the fuses and reassemble.
11.0 CARE AND CLEANING
The LEICA ABBE MARK II Refractometer is a precision optical in­strument designed for years of trouble-free operation. The interi­or design has been carefully constructed for protection from dust or spills. The exterior is painted with a durable enamel. The stain­less steel refractive prism assembly is sealed with corrosion­resistant epoxy.
The use of strong solvents such as N, N-Dimethylformamide or N, N-Dimethylacetamide, cresols, phenols and other tar acids or strong acetic solutions should be avoided. Materials such as Tetrahydrofuran or some lacquer thinners and their components can deteriorate the prism seal over a long period of time or when applied in elevated temperatures.
An accumulation of dust, solvents or sample, particularly on the prism assembly, may eventually penetrate the refractometer and affect accuracy or accelerate wear.
When the instrument is not in use, the measuring prism assembly should be kept closed with a folded sheet of lens tissue between the prisms. The prism surfaces should be cleaned with distilled water, methanol or an appropriate solvent immediately after each sample is used.
Warm distilled water is recommended for removing sugar residues and methanol for oil-based samples. The prism may be wiped with lens cleaning tissue, but it should not be wiped with a hard, dry cloth. After wiping the sample from the prism, rinse the prism again with distilled water.
NNOOTTEE::
Do not wipe the measuring prism surface dry and do not wipe it with anything if it is dry. Do not contact the prism or the surrounding seal with metal or with sharp or pointed objects.
The glass parts of the prism assembly are relatively hard and have excellent resistance to staining. Strong acids or bases should not be measured with this instrument, nor used as clean­ing agents.
To clean the eyepiece, snap out the rubber insert at the end of the adjustable eyepiece tube. Clean the lens with a cotton swab or lens tissue lightly dampened with acetone or methanol. Replace the rubber insert.
12.0 SERVICING
The LEICA ABBE MARK II and LEICA ABBE MARK II Plus Refractometers should be serviced or repaired by factory­trained service and repair technicians. If operating difficulty is experienced contact Leica’s Refractometer Customer Service Department at 716-686-3000 or fax 716-686-3085. Outside the United States contact your local Leica representative.
13.0 SPECIFICATIONS
LLEEIICCAA AABBBBEE MMAARRKK IIII RReeffrraaccttoommeetteerr MMooddeell 1133110044880000//1133110044881100 LLEEIICCAA AABBBBEE MMAARRKK IIII PPlluuss RReeffrraaccttoommeetteerr
MMooddeell 1133110044994400//1133110044995500
DDiissppllaayy
Direct reading LED
MMooddeess ooff OOppeerraattiioonn
Refractive index (nD), %Solids (OBrix) and
temperature compensated Brix - BX-TC
MMooddeess RRaannggee nn
D
1.3200-1.7000
BBrriixx
0-95
BBXX--TTCC aatt 2200
O
0-95
TTeemmppeerraattuurree OCC
15º-50ºC PPoowweerr RReeqquuiirreemmeennttss Model 13104800 115V/60Hz
Model 13104810 230/50Hz Model 13104940 115V/50-60Hz Model 13104950 230V/50-60Hz
Illuminator Arm
Socket Adjustment Screw
Illuminator Bulb
Revolving Condenser Sleeve
FFiigguurree 55bb FFiigguurree 66
Warning:
Unplug the Refractometer Power Cord before attempt-
ing to check or replace fuses. CAAUTION:
Replacement of fuses is not user serviceable. Fuses must be replaced by qualified service personnel. The fuse shall be replaced only with the same size, type and rating.
!
English
8
EElleeccttrriiccaall AApppprroovvaallss
UL listed, CSA certified, VDE listed, CE
Accuracy ±0.0001n
D
±0.1% Solids (OBrix)- 0.0 ­95% ±0.1 on custom scale
TTeemmppeerraattuurree CCoommppeennssaattiioonn
Default mode corrects %Solids scale readings taken at 15OC to 50OC to a datum temperature of 20OC; Mark II Plus Custom Mode corrects to user specified data
TTeemmppeerraattuurree MMeeaassuurreemmeenntt
Thermistor mounted in prism assembly
TTeemmppeerraattuurree CCoommppeennssaattiioonn
±0.1OC
AAccccuurraaccyy SSaammppllee TTyyppeess
Transparent or translucent liquids and solids
DDiimmeennssiioonnss
6"W x 10.5"L x 14"H (152mm x 262mm x 356mm)
WWeeiigghhtt
(net) 16lbs (7.3kg)
SSaammppllee TTeemmppeerraattuurree
Fittings on prism assembly
CCoonnttrrooll
for connecting temperature
controlled water circulator
PPrriissmm AAsssseemmbbllyy
Stainless steel, glass sealed with solvent-resistant epoxy
OOppttiiccaall CCoommppeennssaattiioonn
Dual amici prism compen­sator for sharp shadow-lines and 589nm (Sodium D line) illumination
All models have passed RF Interference Testing, Class A
13.1 The following specifications refer only to the LEICA ABBE MARK II Plus Refractometer
DDiissppllaayy
Direct digital readout of refractive index or concentration (% solids or custom scale) and temperature
DDeeffaauulltt MMooddeess
1. Refractive Index 1.3200 to 1.7000n
D
2. % Solids (OBrix) 0.0 to 95.0%
3. % Solids (
O
Brix) temperature compensated 0.0 to 95.0% Default % Solids (OBrix) scales (0.0-85.0%) based 1974 I.C.U.M.S.A. Tables (% Sugar vs. nDat 20OC)
CCuussttoomm MMooddeess
• 1. 0.0 to 999.9 readout of concentration. User inputs concentra­tion vs. refractive index data
• 2. 0.0 to 999.9 Temperature Compensated readout of concen­tration vs. refractive index vs. temperature data Custom Mode data are input using Leica software program (IBM PC compatible)
IInntteerrffaaccee
RS232 Serial
CCoommmmuunniiccaattiioonn
User selectable Baud rate, 300, 1200, 2400, 4800, 9600 bps - 8 data bits, one stop bit, no parity, Baud rate factory set at 9600. ASCII output with tab separators
CCoonnnneeccttiioonn
DB9 Female
PPiinn AAssssiiggnnmmeennttss
2 output, 3 input, 5
ground
English
9
LEICA ABBE MARK II Refraktometer
Gebrauchsanleitung Inhaltsverzeichnis
1.0 Einleitung ............................................................................ 10
2.0 Auspacken des GerätES....................................................10
3.0 NetZspannung...................................................................... 10
4.0 AuFSTELLen des GerätES.................................................. 10
4.1 Ausrüstung Installation...................................................... 10
5.0 Bedienungselemente.......................................................... 10
5.1 Messungsvariabeln ............................................................ 11
5.2 Kalibrieren des Instruments mit Destilliertem Wasser 11
6.0 Bedienungsanleitung ........................................................ 11
7.0 LEICA ABBE MARK II Plus Refraktometer.................... 12
7.1 Computer-Interface .......................................................... 12
7.2 Einstellung der Übertragungsgeschwindigkeit............ 12
7.3 Druckerschnittstelle .......................................................... 12
7.4 Benutzerspezifisches Programmieren .......................... 12
8.0 Der Einsatz des Refraktometers. ....................................13
Techniken und Grundsätze der Refraktion
8.1 Hoher Brechungsidex/Hohes Dispersion Muster........ 13
9.0 Kalibrieren mit Testglas .................................................... 13
10.0 Auswechseln der Lampe....................................................15
10.1 Auswechseln der Sicherung .......................................... 15
11.0 Pflege und Reinigung ..........................................................15
12.0 Wartung..................................................................................16
13.0 Technische Daten................................................................16
13.1 Die nachfolgenden Angaben beziehen sich auss-
chließlich auf das LEICA ABBE MARK II Plus Refraktometer 16
Erläuterung für Warnungs Symbole
1.0 EINLEITUNG
AACCHHTTUUNNGG!!
Vor Inbetriebnahme des Geräts, lesen Sie diese
Anleitung bitte sorgfältig durch. Diese Gebrauchsanleitung wurde eigens für das LEICA ABBE
MARK II Refraktometer geschrieben und beschreibt die richtige Aufstellung, Benutzung und routinemäßige Wartung des Geräts. Abschnitte 7.0 bis 7.4 beziehen sich ausschließlich auf das LEICA ABBE MARK II Plus Refraktometer.
Das LEICA ABBE MARK II Refraktometer ist ein robustes Tischgerät. Es liefert zuverlässig wiederholbare Meßergebnisse in drei Modi:
• % Feststoffe (ºBrix)
• % Feststoffe (ºBrix ) Temperaturausgeglichen oder (BX-TC) und
• Brechungsindex (nD)
Das LEICA ABBE MARK II Plus Refraktometer kann durch ein
anwendungsspezifisches Programm für die % Feststoffe (ºBrix)­Skala erweitert werden. Dies kann durch Vorprogrammierung oder Nachrüstung in der Fabrik geschehen.
2.0 AUSPACKUNG DES GERÄTES
Für den Transport wurde das LEICA ABBE MARK II Refraktometer sehr sorgfältig verpackt. Bevor Sie das Verpackungsmaterial wegwerfen, kontrollieren Sie den Karton bzw. die Kartons, um sicherzustellen, daß alle bestellten Teile vorhanden sind. Zu dem Zubehör im Refraktometer-Karton gehören:
• ein Brechzahl-Testglas (Kat.-Nr. 10451)
• ein 1/16" (Nr. 4) Sechskantstift-Schlüssel (Kat.-Nr. X30600)
• eine Flasche 1- Bromonaphthalin für Feststoffmessungen (Kat.-Nr. 10452)
• ein Staubschutz (Kat.-Nr. 10450-140)
Vergleichen Sie das Gerät und die Typ-Nr. mit der Bestellung, bevor Sie irgendwelche Verpackungsmaterialien wegwerfen. Das Verpackungsmaterial kann auch zwecks künftiger Lagerung bzw. Transport des Geräts aufgehoben werden.
3.0 NETZSPANNUNG
Netzspannung und Typ 13104800/13104940 120V ~ ±10% - 60Hz Bereich Frequenz Typ 13104810/13104950 240V ~ ±10% - 50Hz Stromaufnahme 25P Max Bereich der Umweltbedingungen
• Höhe nicht über 2000m
• nicht für den Einsatz im Freien
• Temperaturen von 5ºC bis 40ºC
• maximale relative Feuchte - 80% bei Temperaturen bis zu 31º,
linear absteigend bis 50% relative Feuchte bei 40ºC
• Installationskategorie II (Kategorie Überspannung)
• Verschmutzungsgrad 2
4.0 AUFSTELLEN DES GERÄTES
Das LEICA ABBE MARK II Refraktometer is sehr benutzerfre­undlich. Fast keine Einarbeitung ist notwendig, um das Gerät aufzustellen und zu bedienen; somit sind Temperatur- und Brechzahlmessungen sehr leicht zu erhalten.
4.1 Ausrüstung Installation
Das LEICA ABBE MARK II Refraktometer muß an einer harten und ebenen Fläche z.B eine Werkbank oder Tisch aufgestellt werden.
5.0 BEDIENUNGSELEMENTE
Die Bedienungselemente des LEICA ABBE MARK II Refraktometers sind in Abb. 1 dargestellt.
Der an der Vorderseite des Geräts angebrachte Wahlschalter für den Betriebsmodus dient nicht mehr als Ein- und Ausschalter.
EEiinn//AAuusssscchhaalltteerr:: Der Ein/Ausschalter ist an der rechten, unteren Ecke der
Vorderseite des Gerätes (Abb. 1). |=Ein 0=Aus
Deutsch
5X20 100mA T/250V (Typ 1310481/1310495)
VVOORRSSIICCHHTT::
Risiko von elektrischem Schock
VVOORRSSIICCHHTT::
Bitte verweisen sie zu begleitenden
Dokumente
!
VVOORRSSIICCHHTT::
Als Sicherheitsmaßnahme ist der Fuß des Mikroskops über ein dreipoliges Netzkabel mit einem Schutzleiter versehen. Verwenden Sie nie einen zweipoligen Adapter zwischen dem Neztkabel und der Steckdose; dies würde zum Verlust des Schutzeffektes führen.
!
10
DDeerr WWaahhllsscchhaalltteerr wwiirrdd bbeennuuttzztt uumm ddeenn BBeettrriibbssmmoodduuss eeiinnzzuusstteelllleenn::
DDiiee MMooddii ssiinndd::
• % Feststoffe-TC - TC Temperatur ausgeglichen % Feststoffe als Rohzucker (ºBrix) nach der Internationalen Kommission für gleiche Verfahren der Zucker Prüfung, Brechindex Tabellen.
• nD - Brechindex
• % Feststoff - Nicht Temperatur ausgeglichen % Feststoffe als Rohzucker.
5.1 Messungsvariabeln
Bei Temperaturen über 20° C kann destilliertes Wasser im Brechungsindex % Feststoffe-TC-Modus gemessen werden, vorausgesetzt, das Instrument wurde richtig kalibriert. Genaue Meßwerte sind im % Feststoffe-Modus ohne Temperaturausgleich nicht möglich, wenn die Wasserprobe von 20° C abweicht. Bei destilliertem Wasser kann eine Brechzahlmessung (nD) von 0° C bis 80° C vorgenommen werden (siehe Tabelle 1 auf Seite 10 für Werte zwischen 18-30° C).
Der Brechungsindex steht im umgekehrten Verhältnis zur Temperatur. Bei steigenden Temperaturen sinkt die Lösungsdichte und damit auch der Brechungsindex oder % Feststoffe-Meßwert. Bei sinkender Temperatur steigen dagegen die Lösungsdichte und der Brechungsindex (% Feststoffe). Der % Feststoffe-TC- Modus gleicht den Meßwert auf Grund des Temperatur-Koeffizients für Zuckerlösungen aus. Dieser Koeffizient gilt für die meisten wasserhaltigen Lösungen.
5.2 Kalibrieren des Instruments mit destilliertem Wasser
Um das Refraktometer zu kalibrieren:
1. Geben Sie etwas destilliertes Wasser als Probe auf die
Oberfläche des Meßprismas und warten Sie 3 - 5 Minuten ab, um eine stabile Temperatur zu erreichen. Wählen Sie mit dem Wahlschalter den Brechzahlmodus (nD).
2. Schärfen Sie das Okular solange ein, bis das Fadenkreuz deut-
lich erscheint.
3. Drehen Sie das Einstellrad im Uhrzeigersinn so lange, bis die
Schattenlinie das Fadenkreuz fast berührt. Rotieren Sie das Dis­persions-Korrekturrad bis eine möglichst scharfe und farbenfreie Schattenlinie erreicht ist. Dann stellen Sie die Schattenlinie so ein, daß sie mit dem Fadenkreuz in Berührung kommt. Drücken Sie zuerst die Meßtaste und verzeichnen Sie den n
D
Wert, dann die Temperaturtaste drücken und den Temperatur Wert verze­ichnen.
4. Wenn das Gerät richtig kalibriert wurde, beträgt der Meßwert 1,3330 minus 0,0001 pro Grad über 20ºC bzw. plus 0,0001 pro Grad
unter 20ºC. Wenn das Gerät nicht kalibriert wurde, ermitteln Sie nach
Tabelle 1 den richtigen Meßwert. Rotieren Sie das Einstellrad und drücken Sie die Meßtaste solange, bis die Anzeige stimmt. (Die Schattenlinie entfernt sich dabei vom Fadenkreuz auf der Meßplatte.)
5. Führen Sie einen 1/16" Sechskant-Stiftschlüssel in das Fadenkreuz-Einstell-Loch im Dispersions-Korrekturrad (Abb. 2). Drehen Sie den Schlüssel, bis das Fadenkreuz an der Schattenlinie liegt.
AACCHHTTUUNNGG!!
Die letzte Drehung des Einstellrads soll immer im Uhrzeigersinnn erfolgen. Die Schattenlinie bewegt sich dabei nach oben.
6. Drehen Sie den Modus-Umschalter auf die Stellung BX-TC. Drücken Sie die Meß-Taste. Ein angezeigter Wert von 0000.0 oder 0.1 ist normal. Beim Erscheinen eines Minuszeichens und eines Dezimalpunkts ist keine Meßwertanzeige möglich.
Gehen Sie zum nächsten Schritt über. Wenn die Fehleranzeige auch nach wiederholten Meßversuchen
andauert, drehen Sie das Einstellrad solange, bis 000.0 oder 0.1 erscheint. Dazu bedarf es nur einer geringfügigen Verstellung. Bringen Sie das Fadenkreuz erneut auf die Schattenlinie.
Das Anzeigefeld soll jetzt 000.0 zeigen. Wiederholen Sie diese Schritte solange, bis die Anzeige 000.0
beträgt.
6.0 BEDIENUNGSANLEITUNG
Inbetriebnahme des Gerätes:
1. Öffnen Sie den Prismadeckel (Abb. 4) und entfernen Sie das Linsen-Schutztuch. Reinigen Sie das Meßprisma mit Alkohol und dann mit destilliertem Wasser. Wischen Sie es dann mit einem weichen fuselfreien Tuch oder einem Linsenreinigungstuch ab, aber ohne dabei die Oberfläche des Prismas ganz trocken zu wischen, da eventuell auf dem Meßprisma verbleibende Reststoffe die Meß-Leistung beeinträchtigen können.
2. Wählen Sie den erwünschten Modus mit dem Modus­Umschalter.
3. Geben Sie eine Flüssigkeitsprobe auf die Oberfläche des Meßprismas. Benutzen Sie dazu eine Glaspipette. Schließen Sie den Prismadeckel und richten Sie den Beleuchter-Schwenkarm so aus, daß die freie Fläche des oberen Prismas voll beleuchtet ist.
4. Drehen Sie das Dispersions-Korrekturrad bis das Fadenkreuzeinstellloch direkt nach unten zeigt.
Dispersions-Korrekturrad
Verstellbares Okular
Fadenkreuz-Einstell­Loch
Meßwert-und Temperaturanzeige
Temperatur-Taste
Temperatur-Taste
Taste
Wahlschalter
Prisma-Oberfläche
Beleuchter-
Schwenkarm
Einstrellrad
Rechte Außeuseite
AAbbbb.. 11
Fadenkreuz-Einstell Loc
AAbbbb.. 22
Deutsch
11
Loading...
+ 23 hidden pages