Ultimate Hurt (201*00001) is a full-body, lifelike vinyl
manikin that realistically simulates a trauma patient. It is
specifically designed for training professionals in the field of
emergency care and transportation of the sick and injured.
The manufacturing quality of this trainer should provide
many sessions of training when reasonable care and
maintenance is practiced.
Laerdal Recommends:
Endotracheal tube - Size 7.5 ID or smaller
10 ml syringe
Laryngoscope
Laryngoscope blade - Sizes #3 or #4, straight or curved
Items Included:
(1) Full Body Manikin
(1) Mr. Hurt Head Trauma Simulator
(1) Trauma Intubation Head
(1) Deluxe Intubation Head
(1) Advanced Trauma Modules
(1) Jacket
(1) Pants
(1) Carrying Case
(1) Laerdal Global Warranty Booklet
(1) Directions for Use
Skills Taught:
• Oral intubation
• Nasal intubation
• Digital intubation
• EOA/PTL intubation
• Combitube® intubation
• Right mainstem intubation
201-00001 Laerdal
4
• Oropharyngeal and nasopharyngeal airway
insertion
• Bag/Valve Mask
• Suctioning techniques
• Wound management
• Carotid pulse palpation
• Extrication
• Basic patient handling
• Stomach auscultation to verify proper airway
positioning
Preparing the Manikin for Use:
Head:
Head is connected to torso with a long bolt at base of neck,
secured with a lock nut.
To remove:
1. Remove chest plate.
2. Lift esophagus and trachea out of the way.
3. Detach stomach reservoir.
4. Remove lock nut.
Chest Plate:
Chest plate fits into torso by inserting ledge on plate into
corresponding groove on torso cavity.
To remove:
1. Apply pressure to center of plate.
2. Work your fingers under edge of plate.
3. Lift out.
To insert:
1. Position plate over cavity.
2. Work edges into groove on torso cavity.
Belly Plate:
Belly plate fits into lower torso by inserting ledge on plate
into corresponding groove on torso cavity.
201-00001 Laerdal
5
To Remove:
1. Pull plate out by applying pressure to center of plate.
2. Work your fingers under edge of plate.
3. Lift out.
To Insert:
1. Position plate over cavity.
2. Work edges into groove on torso cavity.
Injection Pads:
To remove:
Squeeze pad together with fingers and pull.
To replace, reverse procedure.
Waist, Elbow, Knee and Ankle Joints:
These joints are connected by a bolt and lock nut.
To remove:
1. Unscrew lock nut.
2. Pull bolt out of joint.
The segments will now easily
Laerdal Recommends:
When assembled, dust with
talcum powder to enhance
articulation.
disconnect.
Arms:
Arms are attached with a bolt, two washers, a spring and a
wing nut.
To attach:
1. Remove chest plate from manikin to locate hole in
shoulder.
2. Remove deltoid injection pad from arm.
3. Unscrew wing nut from bolt in arm and remove wing
nut, two washers and spring.
4. Insert bolt through hole in shoulder.
5. Slide one washer, spring and second washer back over
bolt.
6. Use a screwdriver to hold bolt steady while tightening
the wing nut. (See Fig. 1)
To remove arms, reverse procedure.
Fig. 1
201-00001
6
Laerdal
Legs:
Legs are attached with a bolt, two washers, a spring and
a wing nut.
To attach:
1. Remove belly plate from manikin to locate hole in
hip.
2. Remove thigh injection pad from leg.
3. Remove wing nut and washer from bolt inserted in
the leg. (Leave spring assembly in thigh of leg.)
4. Insert bolt through hole in pelvis.
5. Slide washer and wing nut back over bolt.
6. Use a screwdriver to drive bolt into wing nut (See
Fig. 2).
To remove legs, reverse procedure.
Fig. 2
Instructions for Use:
Intubation:
When using Trauma Intubation Head
or Deluxe Intubation Head, the
following equipment is recommended:
a. Endotracheal tube size 7.5 ID or smaller
b. 10 ml syringe
c. Laryngoscope blade, sizes #3 or #4 straight or curved
blade
d. Laryngoscope handle
201-00001 Laerdal
7
Laerdal Recommends
Intubation devices and airway
passages should be lubricated
with silicone or liquid soap
prior to intubation.
IMPORTANT: MR HURT HEAD IS NOT
AN INTUBATION TRAINER!
Care and Maintenance:
1. Clean with mild soap and water; do not submerse
manikin or parts in cleaning fluids or water.
2. Use only on clean surface. Avoid felt tipped markers, ink
pens, acetone, iodine or other staining products and
avoid placing the manikin on newsprint or inked lines of
any kind.
3. To ensure longevity, each manikin should be cleaned
after each training session and a general inspection
should be conducted regularly.
4. Modules and all other parts should be drained and air
dried thoroughly before storage and disinfected when
needed. After use of injection pads (use water only),
accumulated water should be squeezed out. Do not
store wet foam pads in the skin. To prevent mildew or
mold, pads can be soaked in a mild solution of
disinfectant and water or bleach and water. Squeeze
excess solution from pads, allow them to dry, then store
or reinsert in manikin.
5. Articulating parts will benefit from a light application of
talcum powder prior to training sessions.
6. Store properly between teaching sessions.
Cautions and Warnings
This product contains Natural Rubber latex which may
cause allergic reactions when in contact with humans.
201-00001 Laerdal
8
Replacement Parts:
Product Number: Description:
380460 Blank Genitalia, Adult
Male
380475 Pelvis Pin
380600 Adult Left Leg,
380650 Adult Right Leg
380800 Plain Adult Male Left
Arm
380801 Left Hand, Adult Male
381401 Upper Right Arm,
Adult Male
381402 Adult Injection Deltoid
Pad
381502 Adult Right Crushed
Foot
381503 Deltoid Pad, Gunshot
Large Caliber Adult
381504 Deltoid Pad, Gunshot
Small Caliber Adult
381505 Ventro-gluteal Pad,
Gunshot Small Caliber
Adult
381506 Gluteal Gunshot Pad,
Large Caliber Adult
381507 Impaled Thigh Pad
Adult
381508 Open Fracture Thigh
Pad, Adult
381509 Lower Left Leg,
Contused Foot/Ankle
Adult
381510 Lower Right Leg
with Fracture Adult
381512 Abdominal Skin
with Exposed Viscera
Adult
101-10150 Plain Abdominal Plate,
Hospital Version
Adult
200-02150 Adult Jacket
200-02250 Adult Pants
201-00150 Intubation Head
with Lungs Adult Male
201-00001 Laerdal
9
201-00250 Torso, Hospital
Version Adult Male
261-02001 Mr. Hurt Head
276-00150 Case
276-00250 Trauma Head
276-00350 Right Forearm
with Lacerated
Hand Dorsum
276-00450 Left Forearm with
Burnt Hand Adult Male
276-00550 Abdominal Plate
Exposed Viscera Adult
Hospital Version
276-00650 Right Foreman Pad,
Open Fracture
276-01150 Gunshot Wound
Chest Plate
300-00150 Thigh Injection Pad
Adult Version
300-00250 Ventro-gluteal Injection
Pad Adult Male
300-00350 Gluteal Injection Pad
Adult Male
300-00650 Lubricant (4 oz)
300-00950 Upper & Lower Adult
Denture Set
300-01350 Stomach Reservoir
Hospital Version
Adult Male
300-01450 Pelvis Hospital Version
Adult Male
300-01550 Lung Bag Assembly
300-01850 Adult Upper Teeth
300-01950 Adult Lower Teeth
300-02350 Adult Male Right
Forearm
300-02450 Right Forearm Pad,
Non-Injection Adult
Male
300-02550 Right Hand, Adult Male
300-02650 Upper Left Arm, Adult
Male
300-02750 Adult Male Left
Forearm
201-00001 Laerdal
10
300-02850 Left Forearm Pad,
Non-Injection Adult
Male
300-02950 Adult Right Thigh
300-03050 Adult Right Lower Leg
300-03150 Adult Right Foot
300-03250 Adult Left Thigh
300-03350 Adult Left Lower Leg
300-03450 Adult Left Foot
300-04250 Adult Hardware Set,
Legs & Arms to Body
300-04350 Adult Hardware Set,
Pelvis to Torso
300-05150 Plain Adult Right Arm
300-10250 Plain Chest Plate,
Hospital Version
Adult Male
Please contact Customer Service Representatives for more
information on Replacement Parts in other skin tones.