DEUTSCH Gebrauchsanweisung 31
ITALIANO Instruzioni per l'uso 55
ESPANÕL Instrucciones de utilización 79
NORSK Bruksanvisning 103
SVENSKA Bruksanvisning 127
FRANÇAIS Mode d’emploi 151
NEDERLANDS Handleiding 175
DANSK Brugsanvisning 199
SUOMI Käyttöohje 223
CPRmeterTM Directions for Use
Contents
1 - Description and Indications for Use 7
2 - Setup 8
3 - Use during an Emergency 10
4 - Compression Feedback 12
5 - Debriefing after Use 15
6 - Maintenance and Cleaning 17
7 - Troubleshooting Guide 20
8 - Accessories, Consumables and Spare Parts 21
9 - Glossary of Symbols 23
10 - Additional Warnings and Cautions 24
11 - Recommendations 25
12 - Specifications 26
13 - Environmental Considerations 28
English
14 - Electromagnetic Conformity 29
About this edition
The information in this Directions for Use applies to the model 801-00140 CPRmeter™.
This information is subject to change. Please contact your local Laerdal representative
for information on revisions.
Laerdal Medical AS, Tanke Svilandsgate 30, PO Box 377, 4002 Stavanger, Norway,
(+47) 51511700.
The CPRmeter™ with Q-CPR® technology is protected under U.S. Patents 6,306,107,
6,351,671, 6,390,996, 7,074,199, 7,108,665, 7,118,542, 7,122,014, 7,220,235 and other
foreign patents. Other U.S. and foreign patents pending. The design of CPRmeter™ is
protected under design registrations/design patents US609,813; US628,212; EU997416,
EU997424, NO81223, NO81219, AU323899, AU324044 and other foreign design
registrations.
CPRmeter™ and Q-CPR® are trademarks or registered trademark of Laerdal
Medical AS. Energizer® is a registered trademark of Eveready Battery Company,
Inc. Panasonic® is a registered trademark of Matsushita Electric Industrial Co., Ltd.
Duracell® is a registered trademark of The Gillette Company or its affiliates. SD™ and
microSD™ are trademarks of SD-3C LLC.
5
CPRmeter Overview
Front View Rear View
(this surface up) (apply this surface to patient)
Status Light
Display
Area
Compression
Area
Placement
Guide
Rear
Cover
Yellow Patient
Adhesive
Mount Area
Vent
Membrane
On / Off Button
Side View
Rear Cover
Screws
6
English
1 Description and Indications for Use
The CPRmeter with Q-CPR® technology is a small, lightweight device
powered by a replaceable battery. The CPRmeter is intended for use by
responders who have been trained in CPR and use of the CPRmeter.
The CPRmeter is used as a guide in administering cardiopulmonary
resuscitation (CPR) to a suspected sudden cardiac arrest (SCA) victim at
least 8 years old. If in doubt about the appropriateness for use, perform CPR
without using the CPRmeter.
When attached to the bare chest of a suspected victim of SCA, the
CPRmeter provides real-time feedback on CPR compressions in accordance
with current CPR guidelines. It displays CPR feedback indicators for depth,
release, and rate of chest compressions. It also counts the number of
compressions in a series, and provides notification of lack of expected CPR
activity.
WARNING: The CPRmeter is not intended for use on SCA victims under
8 years old.
IMPORTANT NOTE: CPR cannot assure survival, no matter how well it is
performed. In some patients, the underlying problem causing the cardiac
arrest is not survivable despite any available care.
7
2 Setup
Check the contents of the CPRmeter box. It should contain:
– the CPRmeter
– red sleeve
– set of 3 CPRmeter patient adhesives in a resealable pouch
– Directions for Use
– Laerdal Global Warranty and WEEE-statement leaflets
– microSD™ memory card with SD™ card adapter
– 3V Lithium 123 (Li/MnO2) battery
Remove the protective film from the CPRmeter display screen before first use.
Inserting a new Battery
1. Unscrew the rear cover using a flathead
screwdriver (not supplied) or a coin,
and lift out the cover.
2. Insert a new battery into the battery
well, in the orientation as directed by the
battery symbol inside the well.
3. Ensure that the rear cover’s vent
membrane is not soiled or damaged.
Refer to Chapter 6, Replacing the Rear Cover for instructions.
4. Position the rear cover on the CPRmeter.
Thread the screws through the washers
and into the cover, and tighten the screws.
The CPRmeter’s status light will flash green
every 5 seconds to indicate that it has
passed its self-test and the battery is OK
for use.
If for any reason the CPRmeter’s status indicator light is orange or does not
flash green, refer to Chapter 7, Troubleshooting Guide for instructions.
Inserting a microSD™ Memory Card
Use of a microSD™ memory card with the CPRmeter is optional. Refer to
Chapter 5, Q-CPR® Review for further details.
1. Remove the Rear Cover and Battery.
2. Insert a microSD™ card into the card
slot inside the battery well, in the
orientation as directed by the card
symbol inside the well.
3. Re-insert the battery and attach the
rear cover as described in the previous
section.
8
English
Applying a new Patient Adhesive
1. Check the label on the CPRmeter patient
adhesives resealable pouch to be sure
that they are within their expiration date.
2. Open the package of CPRmeter patient
adhesives and peel one from the white
liner strip to expose the adhesive surface
on its underside.
3. Align the bottom of the patient adhesive
with the yellow patient adhesive mount
area on the rear cover of the CPRmeter.
Make sure the channel on the adhesive is
directly over the vent membrane. Press
the patient adhesive into place.
4. Do not peel off the green liner from the
attached adhesive until you are ready to apply the CPRmeter to a patient
in an emergency. Return remaining patient adhesives to their resealable
pouch.
Storing the CPRmeter between Use
Use the CPRmeter red sleeve to shield the display screen from scratches
and to protect the patient adhesive from damage. Ensure that the On/
Off button can not be inadvertently activated during storage.
9
3 Use During an Emergency
Attach the CPRmeter to the Patient’s Chest
1. Make certain the patient is on a firm surface.
2. Remove clothing from the patient’s chest.
3. Remove the CPRmeter from its red sleeve.
4. Turn on the CPRmeter by pressing the On /
Off button once.
5. If the patient’s chest is wet, dry it before
attaching the CPRmeter.
6. Peel off the green liner from the CPRmeter
patient adhesive to expose the white adhesive
surface.
7. Position the CPRmeter so that the
compression area is on the lower half of the
sternum (breastbone), on the centerline of
the bare chest, as illustrated on the front of
the CPRmeter.
8. If the CPRmeter moves during use, re-position
it to the center of the chest, as described
above.
9. If difficulty is encountered in applying the CPRmeter,
do not delay initiation of CPR. Remove the CPRmeter
and begin compressions.
WARNING: Do not use the CPRmeter in conjunction
with any mechanical or automated compression device.
WARNING: Do not use the CPRmeter on top of
defibrillation pads, unless the manufacturer of the
defibrillator and the defibrillation pads has explicitly
stated that the CPRmeter can be used in such manner.
10
English
Perform CPR. Follow Feedback.
Using standard CPR technique, place the heel
of one hand directly onto the center of the
compression area of the attached CPRmeter.
Place your other hand on top of the first.
You should be able to see the display area
of the CPRmeter to look for feedback.
1. Provide chest compressions according to your CPR protocol.
2. When compressions are first detected by the CPRmeter, the display
changes to enlarged compression indicators, as shown:
3. Follow feedback provided by the indicators on the CPRmeter display. The
indicators are described in the following section.
WARNING: When the CPRmeter is used together with a defibrillator, make
sure to follow the defibrillator manufacturer’s instructions. Stop compressions,
remove hands from the CPRmeter and remain clear of all patient contact
during defibrillation or when otherwise required, in accordance with a proper
defibrillation protocol.
11
4 Compression Feedback
Compression Rate Target
Compression
Release Target
Compression
Depth Indicator
Compression
Depth Target
The display of the CPRmeter provides real-time feedback to the responder as
described below.
Compression
Rate Indicator
Good Compressions
Each compression performed is represented on
the CPRmeter display by a moving white
compression depth indicator bar. When the
CPRmeter detects that a compression meets
the targets for depth or release, the target briefly
illuminates.
If the CPRmeter detects that the rate of
compressions is within the target area, the
“speedometer” needle on the compression rate
indicator points to the green target area and the
green target briefly illuminates.
Compression Depth
If the CPRmeter detects a compression that fails
to meet the proper minimum compression depth
target (50 mm / 2 inches – for a patient on a hard
surface), the compression depth target does not
illuminate.
If 4 consecutive compressions fail to meet the
compression depth target, the display shows a
yellow arrow pointing to the compression depth
target.
If the CPR meter detects a compression that
exceeds 70 mm (2.75"), the CPRmeter shows the depth indicator below
the target area. If a specific CPR event requires CPR to be performed
on a patient lying on a mattress, slide a backboard under the patient and
compensate for the mattress softness by ensuring that for each chest
compression the area below the compression depth target lights up.
WARNING: When performing CPR on a patient lying on a mattress, a
backboard must be used to limit the amount of compressed depth which is
absorbed by the mattress. Depending on characteristics of the mattress,
backboard and patient, the depth compensation does not guarantee that
the patient chest is compressed by 50 mm / 2 inches.
12
English
Incomplete Release
If the CPRmeter detects that pressure is not fully
released between compressions, the release target
does not illuminate. If 4 consecutive compressions
fail to meet the release target, the display shows a
yellow arrow pointing to the compression release
target. Release pressure completely between
compressions.
IMPORTANT NOTE: Release pressure completely between compressions.
Compression Rate
If the CPRmeter detects that the compression
rate is greater than the target rate, the needle on
the compression rate indicator points to the right
of the green area. If the CPRmeter detects that the
compression rate is less than the target rate, the
needle on the compression rate indicator points to
the left of the green area.
Compression Counter
When compressions start, the CPRmeter display
shows a numerical compression counter. The
counter digits turn solid white between 25 and 30
compressions, as a guide to the responder when
performing a cycle of 30 compressions. Beyond 30
compressions, the counter digits flash solid white for
every tenth compression*.
The compression counter is reset after 3 seconds
without a compression.
* When performing continuous compression CPR on an intubated patient,
the compression counter may be used to time delivery of ventilations. At a
compression rate of 100 per minute: If a ventilation is delivered for every 10
continuous compressions, ventilation rate should be 10 per minute.
25
Compression
Counter
13
Inactivity
If compressions are stopped during a CPR event,
after 3 seconds the CPRmeter displays an inactivity
timer which counts the seconds since the last
compression.
After 20 seconds since the last compression, the
inactivity timer starts flashing.
After 1 minute, the CPRmeter display fades down
to conserve battery power. The display is restored
when a new compression is delivered.
After 10 minutes of inactivity, the CPRmeter is automatically turned off. To
restart the CPRmeter, press the On/Off button.
Inactivity Timer
0:23
Turning Off the CPRmeter
To turn the CPRmeter off, press and hold the On / Off button for at least 1
second.
14
English
5 Debrieng After Use
Q-CPR® Quick Review
The CPRmeter can display CPR performance statistics for the last CPR
event. When the CPRmeter is turned on, press the On / Off button once to
activate Q-CPR Quick Review. The statistics are shown over two displays.
–
Percentage of compressions with release force
within the Compression Release Target.
– Percentage of compressions with compression
depth which have reached the Compression
Depth Tar get.
– Percentage of compressions with compression
rate within the Compression Rate Target.
Press the On / Off button once to cycle to the next
display.
–
Duration of the CPR event (minutes: seconds).
– Percentage of the CPR event duration where
chest compressions have been delivered.
Press the On / Off button once to cycle between
the displays.
98%
96%
70%
24:35
73%
To turn the CPRmeter off, press and hold the On / Off button for at least 1
second.
The CPRmeter reverts to Compression Feedback mode if a compression is
delivered.
The CPR event statistics are stored when the CPRmeter is turned off. When
turned on again, the statistics from the stored CPR event can be reviewed,
as described above.
When the CPRmeter is used in a new CPR event, the preceding event’s
statistics are deleted and the new event’s statistics are stored in Q-CPR
Quick Review.
CPR performance statistics are only calculated if at least 10 compressions
have been delivered.
15
Q-CPR® Review
When a microSD™ memory card has been
inserted into the CPRmeter prior to use, data
from CPR events are stored on the card. The
data can only be viewed on a PC with the
Laerdal Q-CPR Review software program,
version 3.1 or newer, installed.
Q-CPR Review software is available for download
by visiting: www.laerdal.com/downloads
To transfer CPRmeter data to a PC with a SD™ memory card reader:
1. Remove the microSD™ card from the CPRmeter and insert it into the
provided SD™ card adapter.
2. Inser t the SD™ card adapter into the SD™ card reader.
3. Star t the Q-CPR Review software on the PC and follow instructions.
Remember to re-insert a microSD™ card into the
CPRmeter before further CPR event data recording.
If the ’remove memory card’ symbol is shown
on the CPRmeter display, the memory card is
full and should be replaced to prevent data from
subsequent CPR events from being lost.
The CPRmeter’s compression feedback and Q-CPR Quick Review features
will function as normal even if the memory card is full.
16
English
6 Maintenance and Cleaning
Routine Maintenance
1. On a routine basis, check the CPRmeter’s status light. If the status light is
not flashing green every 5 seconds: See Chapter 7, Troubleshooting Guide.
2. On a routine basis, check that the CPRmeter has a patient adhesive in place
and that the liner remains on it. Replace the patient adhesive at least every
2 years if it is not used.
3. Replace the battery at least every 2 years. Replace the patient adhesive
every time the rear cover is opened.
Battery Monitoring
The CPRmeter continuously monitors the power of its battery. If the
remaining power is less than that required for an entire CPR event, the
following visual indicators signal that the battery should be replaced before
the next use:
– The green status light does NOT flash green
every 5 seconds when the CPRmeter is turned
off.
– A small battery-low icon is shown in the upper
right corner of the display when the CPRmeter
is being turned on.
– A large battery-low icon is shown on the display
when the CPRmeter is being turned off.
IMPORTANT NOTE: If the remaining battery
power during use becomes too low to sustain further operation, the batterylow icon is shown for 10 seconds and then the CPRmeter turns itself off.
WARNING: Do not interrupt CPR to replace the battery. Continue CPR
without feedback from the CPRmeter.
Battery Replacement
1. Remove the CPRmeter patient adhesive from the back of the CPRmeter.
2. Unscrew the rear cover using a flathead screwdriver (not supplied) or a
coin, and lift out the cover.
3. Remove the old battery and discard.
4. Insert a new battery into the battery well, in the orientation as directed
by the battery symbol inside the well. See Chapter 12, Specifications, for
battery specifications.
5. Ensure that the rear cover’s vent membrane is not soiled or broken.
6. Position the rear cover on the CPRmeter. Thread the screws through the
washers and into the cover, and tighten the screws.
7. Apply a new CPRmeter patient adhesive to the CPRmeter as described in
Chapter 2, Setup.
17
After each Use
After use on a patient, the CPRmeter may be contaminated and should be
handled appropriately.
1. Place the contaminated CPRmeter in a plastic bag until it can be cleaned.
Do not insert a contaminated CPRmeter into the red sleeve.
2. If it is visibly soiled, wipe the CPRmeter with a soft cloth or paper towel to
remove as much contamination as possible.
3. Remove the CPRmeter patient adhesive from the back of the CPRmeter.
4. Clean the CPRmeter as described under Cleaning.
5. Examine the vent membrane in the center of the yellow rear cover. If the
vent membrane is soiled or damaged, see Replacing the Rear Cover.
6. Check the exterior of the CPRmeter for signs of damage. Contact Laerdal
to arrange for replacement if needed.
7. Apply a new CPRmeter patient adhesive to the CPRmeter as described in
Chapter 2, Setup.
Cleaning
If the CPRmeter has been used in a training situation, it may be wiped using
an alcohol wipe (70% ethanol solution).
If the CPRmeter has been used in a clinical situation, clean it as follows:
1. Clean and scrub the exterior using a mild detergent and either a soft cloth
or toothbrush until the surfaces are visually clean.
2. Wipe the exterior with a soft cloth dampened in lukewarm water.
3. Clean the exterior using a 0.55% solution of or tho-phthalaldehyde. Spray the
solution on to cover all exterior surfaces, and allow to sit for a minimum of
10 minutes. An alternative cleaning agent is isopropyl alcohol (70% solution).
4. Wipe the exterior with a clean soft cloth dipped in water. Allow to dry
completely.
WARNING: The above procedure does not ensure disinfection. Cleaning
the device with 0.55% solution of ortho-phthalaldehyde ensures that the
CPRmeter is thoroughly cleaned.
CAUTION: Do not immerse the CPRmeter in water, hold it under running
water, or allow moisture to penetrate it. Do not sterilize the CPRmeter.
18
English
Replacing the rear cover
The vent membrane allows the CPRmeter to adapt to minute air pressure
changes while protecting it from fluid entry.
If the vent membrane in the center of the yellow rear cover is soiled,
contaminated, scratched or damaged, the rear cover must be replaced. Do
not attempt to clean the vent membrane.
WARNING: A soiled or damaged vent membrane may lead to inaccurate
feedback or fluid entry into the CPRmeter. If there are signs of fluid entry,
remove the CPRmeter from service and contact your Laerdal representative
for advice.
If the rear cover is damaged or does not fit flush with the CPRmeter housing,
or the rear cover screws have sharp edges or the washers under each of the
rear cover screws are missing, the rear cover must be replaced. See Chapter
8, Accessory, Consumables and Spare Parts.
1. Unscrew the rear cover using a flathead screwdriver (not supplied) or a
coin, and lift out the cover.
2. Remove the replacement rear cover from its packaging and check to be
sure it is not damaged.
3. Position the rear cover on the CPRmeter. Thread the screws through the
washers and into the cover, and tighten the screws.
Customer Service Indicator
The Customer Service Indicator (right) appears
on the CPRmeter only at shutdown, after 500 000
chest compressions. Contact your local Laerdal
representative for further instructions.
19
7 Troubleshooting Guide
ProblemSuggested Action
The CPRmeter display is dark. Make sure the CPRmeter is
turned on.
CPRmeter status light does not
flash green.
CPRmeter status light is orange
(solid or flashing) and the
CPRmeter display is dark.
CPRmeter does not adhere to
patient's chest.
User is unfamiliar with the
CPRmeter, or thinks there is a
problem with the CPRmeter.
Replace the battery. Ensure that
the battery is inser ted in the
correct orientation.
Remove the CPRmeter from
service. Contact Laerdal for
technical support.
If there is no other responder
present, provide CPR using the
CPRmeter even if it does not
adhere to the patient's chest.
Be sure to keep the CPRmeter
properly positioned.
If there is another responder
present to provide CPR, remove
the CPRmeter and replace the
CPRmeter patient adhesive.
Dry the patient's chest if wet and
re-apply the CPRmeter with a
minimum interruption to CPR.
Ignore feedback from the
CPRmeter and continue with
CPR.
At an appropriate time after
the CPR event, inspect the vent
membrane for soiling or damage.
Apply a new patient adhesive
and let a trained user perform
chest compressions with the
CPRmeter on a CPR manikin. If
the CPRmeter does not appear
to function as expected, remove
it from service and contact
Laerdal or approved Laerdal
representative for technical
support.
20
English
8 Accessories, Consumables and Spare Parts
Contact Laerdal to order the following replacements for the CPRmeter:
– 10 replacement packs of CPRmeter patient
adhesives (each pack holds 3)
[REF: 801-10850]
– Replacement rear cover with screws and rubber washers
[REF: 801-10750] (spare recommended)
– CPRmeter Red Sleeve
[REF: 801-10150]
– CPRmeter Hard Case
[REF: 801-10550]
The accessory CPRmeter Hard Case for holding the
CPRmeter can be attached to a belt or a clip hook
– Reusable Silicone Cover kit,
includes one pack of 3 large patient adhesives
[REF: 801-10650]
– 10 replacements packs of CPRmeter large patient
adhesives (each pack holds 3)
[REF: 801-10950]
– CPRmeter battery (each pack holds 5)
[REF: 801-10350]
– microSD card kit (each pack holds 5)
[REF: 801-10450]
21
Reusable Silicone Cover Kit
The CPRmeter can be fitted with the accessory Reusable Silicone Cover and
the large patient adhesive (single-use). This increases the contact area to the
patient.
Applying the Reusable Silicone Cover and large patient adhesive
1. Check that the silicone cover has been cleaned
and disinfected after last use.
2. Inser t the CPRmeter
into the sleeve.
3. Check the label
on the CPRmeter large
patient adhesives resealable
pouch to be sure that they are within their expiration date.
4. Open the package of CPRmeter large patient adhesives and peel one
from the white liner strip to expose the adhesive surface on its underside.
5. Align the bottom of the patient adhesive with the yellow patient adhesive
mount area on the rear cover of the CPRmeter and the adjacent silicone
cover’s bottom surface. Make sure the channel on the adhesive is directly
over the vent membrane. Press the large patient adhesive into place.
6. Do not peel off the green liner from the attached large patient adhesive
until you are ready to apply the CPRmeter to a patient in an emergency.
Return remaining patient adhesives to their resealable pouch.
IMPORTANT NOTE: Do not delay CPR. If the Reusable Silicone Cover is
generally used with the CPRmeter, the cover and the Large Patient
Adhesives should be applied to the CPRmeter in advance.
Use and maintenance
The CPRmeter use or maintenance is not changed by use of the silicone
cover. Clean the silicone cover separately from the CPRmeter. Disinfect the
silicone cover using the following method:
1. Clean and scrub the cover with a small soft brush, using a mild detergent
until the cover is visibly clean.
2. Rinse in lukewarm water.
3. Submerge the cover in a 0.55% solution of ortho-phthalaldehyde in
accordance with manufacturer’s specifications.
4. Rinse again and dry.
The silicone cover can also be autoclaved in distilled water at 136 ºC
(210 ºF) and 2.0 kg/cm² for 10-20 minutes.
WARNING: The CPRmeter with the silicone cover can not be
considered as disinfected or sterilized.
22
English
9 Glossary of Symbols
IP55
LATEX
The following symbols appear on the CPRmeter, accessories and packaging:
SymbolDenition
The product is in compliance with the essential requirements of Council
Directive 93/42/EEC as amended by Council Directive 2007/47/EC.
Compliance with applicable U.S and Canadian safety standards has
been certified by Canadian Standards Association.
These CPRmeter patient adhesives are disposable and are for single
patient use only. Do not re-use. Re-use will lead to increased risk of
cross contamination, and/or degradation of adhesive performance.
Defibrillation protection. The CPRmeter is defibrillation protected,
type BF patient connection.
Manufacturer.
Dispose of in accordance with your country's requirements.
Reference order number.
The CPRmeter meets IEC 60529 class IP55.
Serial number.
Expiration date for patient adhesives, formatted YYYY-MM
(year-month).
The patient adhesives, CPRmeter or other parts do not contain
natural rubber latex.
Warning / Caution symbol
Store the CPRmeter patient adhesives between the temperatures
shown. See Chapter 12, Specifications.
= #
Contains number of CPRmeter patient adhesives shown as “#.”
Refer to Directions for use.
Consult Directions for Use
Lift here to peel off the patient adhesive liner and apply to patient’s
bare chest.
Not for use on children under 8 years old.
microSD memory card
Battery – 3V Lithium 123 type
23
10 Additional Warnings and Cautions
A Warning identifies conditions, hazards, or unsafe practices that can result
in serious personal injury or death. A Caution identifies conditions, hazards,
or unsafe practices that can result in minor personal injury or damage to the
CPRmeter.
– WARNING: The CPRmeter is not intended for use in a moving
environment, such as an ambulance. If used during patient transport,
the CPRmeter may provide inaccurate feedback. If CPR is indicated in
a moving environment, do not rely on the CPRmeter depth feedback
during such conditions. It is not necessary to remove the device from the
patient.
– WARNING: Do not practice by using the CPRmeter on a person. The
CPRmeter may be used with a training manikin or simply on a compliant
surface for practice.
– WARNING: Properly performed CPR may result in fracturing of the
patient’s ribs.* If rib integrity has been compromised, continue to provide
CPR in accordance with your local protocol.
– WARNING: Properly performed CPR may result in chest injuries* e.g.
external chest wall bruising or abrasion.
– WARNING: Do not rely on CPRmeter feedback during aircraft ascent
and descent, as its accuracy is reduced in such conditions.
– CAUTION: Do not apply the CPRmeter to an open wound or recent
incision site.
– CAUTION: The CPRmeter is designed to be used only with Laerdal-
approved accessories. The CPRmeter may perform improperly if nonapproved accessories are used. Do not attempt to modify the CPRmeter
in any way.
– CAUTION: Use only model 801-10850 Patient Adhesives with the
CPRmeter. If the Reusable Silicone Cover is applied to the CPRmeter, use
only model 801-10950 Large Patient Adhesives.
* Black CJ, Busuttil A, Robertson C. Chest wall injuries following
To prolong the life of the display, avoid storing the CPRmeter where it is
exposed to direct sunlight when not in use.
Responders should receive training, including regular refresher training, in
use of the CPRmeter. When training with the CPRmeter on a CPR manikin,
disable or ignore feedback from the manikin.
25
12 Specications
CPR Targets
CategorySpecication
Compression Depth Tar get> 50 mm
( 2”)
Depth accuracy: ±10%
Compression Release Tar get
Force
Compression Rate Tar get100 to 120/min ± 3/min
< 2,5 kg (5,5 lbs).
Force accuracy: +1,5 kg, -2,0 kg
(+3,3 lbs, - 4,4 lbs)
CPRmeter [REF 801-00140]
The CPRmeter meets the performance requirements of IEC 60601-1, 2nd
and 3rd edition.
CategorySpecication
Dimensions154 mm x 64 mm x 28 mm
(6.1” x 2.5” x 1.1”)
Weight227g (7,3 oz)
Battery3V Lithium 123 (Li/MnO2)
The CPRmeter has been tested
with Energizer® batteries. Use
only Energizer®, Panasonic® or
Duracell® batteries.
microSD™ Memory CardThe CPRmeter has been tested
with 2Gb microSD™ memory
cards. Always test a new card
before using it in data collection.
Temperature and Relative
Humidity
Electromagnetic CompatiblityMeets IEC 60601-1-2 and RTCA/
Recovery timeRecovery time after defibrillation: 0 s
Storage temperature:
-20° to 70° C (-4° to 158° F).
Relative humidity: 0% to 75%.
Operating temperature:
0° to 40° C (32° to 104° F).
Operating atmospheric pressure :
1000 to 540 mbar (100 to 54 kPa).
Relative humidity 0% to 95%.
Sealing: Meets ISO/IEC 60529 class
IP55.
Product temperature can reach
60° C.
DO-160E.
26
English
CPRmeter Patient Adhesives [REF 801-10850]
CategorySpecication
Dimensions39 mm x 90 mm (1.5” x 3.5”).
Temperature and Relative
Humidity
MaterialFoam pad with biocompatible
Shelf Life2 years when applied to
Storage temperature:
-20° to 70° C (-4° to 158° F )
Relative humidity 0% to 75%
Operating temperature:
0° to 50° C (32° to 122° F)
Relative humidity 0% to 95%.
adhesive on each side.
the CPRmeter or 4 years in
unopened package. Do not
exceed the expiration date on
the packaging.
CPRmeter Large Patient Adhesives [REF 801-10950]
CategorySpecication
Dimensions64 mm x 128 mm (2.5” x 5”).
Reusable Silicone Cover [REF 801-10650]
CategorySpecication
Dimensions66 mm x 156 mm x 31 mm (2.6”
x 6.1” x 1.2”)
Weight36.5 g (1.3 oz)
MaterialSilicone
Warranty
The Laerdal CPRmeter has a one-year limited Warranty. Refer to the Laerdal
Global Warranty for terms and conditions.
27
13 Environmental Considerations
ProductInformation
CPRmeter The CPRmeter contains electronic
components. Dispose of it at an
appropriate recycling facility in
accordance with local regulations.
CPRmeter Patient AdhesiveThe used patient adhesive may be
contaminated with body tissue,
fluid, or blood. Dispose of it as
infectious waste.
28
English
14 Electromagnetic Conformity
Guidance and manufacturer’s declaration: The CPRmeter is intended for use in the
electromagnetic environment specified in the tables below. The user of the CPRmeter
should assure that it is used in such an environment.
Electromagnetic Emmissions
Emissions TestCompliance
RF CISPR 11Group 1 ClassThe CPRmeter uses RF energy only for its
Electromagnetic environment - Guidance
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The CPRmeter is suitable for use in
all establishments, including domestic
establishments and those directly connected
to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Electromagnetic Immunity
Immunity TestIEC 60601
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
Power
Frequency
(50/60/400 Hz)
Magnetic Field
IEC 61000-4-8
Radiated RF
IEC 61000-4-3
Test Level
±6 kV contact
±8 kV air
3 A/m3 A/mPower frequency
10 V/m
80 MHz to
2.5 GHz
Compliance
Level
±6 kV contact
±8 kV air
10 V/mPortable and mobile
Electromagnetic
Environment -
Guidance
There are no special
requirements with
respect to electrostatic
discharge.
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial/
hospital environment.
There are no special
requirements for
non-commercial/
non-hospital
environments.
RF communications
equipment should be
used no closer to any
part of the CPRmeter,
than is absolutely
necessary.†,‡ The
recommended
separation distances for
various transmitters and
the CPRmeter are
shown in the following
table.
Interference may
occur in the vicinity of
equipment marked
with the following
symbol:
29
† The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are
6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz;
and 40,660 MHz to 40,700 MHz.
‡ Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
CPRmeter is used exceeds the applicable RF compliance level above, the CPRmeter
should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the CPRmeter.
Recommended separation distances between portable and mobile
RF communications equipment and the CPRmeter
The CPRmeter is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the CPRmeter
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
CPRmeter as recommended below, according to the maximum output power of the
communications equipment.
Electromagnetic Emmissions
Rated maximum output Separation distance according to frequency of
power of transmitter (W) transmitter (M)
80 MHz to 800 MHz 800 MHz to 2.5 GHz
d= 1.2 √P d= 2.3 √P
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be determined using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE: 1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2. The ISM (industrial, scientific and medial) bands between 150 kHz and 80 MHz
are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13, 567 MHz; 26,957 MHz to
27,283 MHz; and 40,660 MHz to 40,700 MHz.
NOTE 3. An additional factor of 10/3 is used in calculating the recommended separation
distance for transmitters in the ISM frequency bands between 150 kHz and 80 MHz
and in the frequency range 80 MHz to 2.5 GHz to
decrease the likelihood that mobile/portable communications equipment could cause
interference if it is inadvertently brought into patient areas.
NOTE 4. These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people
NOTE 5. Transmitters/antennas of this power-level are most likely mounted on an
emergency vehicle chassis. The distances cited here are for open field. For an external
antenna, the separation distance is most likely shorter.
0,12
0,38
1,2
3,8
12
0,23
0,72
2,3
7,28
23
30
30
Loading...
+ 218 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.