Introduction
The Interchangeable Enema and Catheterization Trainer (375-21001)
is a life-like, vinyl female pelvis, designed to teach enema and urinary
catheterization procedures.
The manufacturing quality of this simulator should provide many sessions of
training when reasonable care and maintenance is practiced.
Laerdal Recommends
Urethral Catheterization – 16 French
Enema Catheterization – 7mm
Belly Plates
The belly plate ts into the lower torso by a ledge on the plate and
corresponding groove on the torso cavity.
To remove:
Pull out plate out by applying pressure to center of plate then work ngers
under edge of plate and lift out.
To insert:
Position plate over cavity and work edges into groove on torso cavity.
Injection Pads
To remove:
Squeeze pad with ngers and pull.
To insert:
Squeeze pad with ngers and insert.
Hip Joints
The hip joints are attached with a bolt, a spring, two washers and a wing
nut.
To remove thighs:
a. Remove belly plate from trainer.
b. Remove thigh injection pad from thigh.
b. Remove wing nut and washer from bolt. (Leave spring assembly in thigh.)
c. Remove bolt.
English
Caution: Latex
This product contains Natural Rubber latex which may cause
allergic reactions when in contact with humans.
Items Included
(1) Interchangeable Catheterization and Enema Trainer
(1) Male genitalia assembly
(1) Female genitalia assembly
(1) Enema reservoir assembly
(3) Urinary valve assemblies
(3) Anal valve assemblies
(2) Thigh Pads
(1) Ventro-Gluteal Pad
(1) Dorsal-Gluteal Pad
(1) Carry Case
(1) Laerdal Global Warranty Booklet
(1) Directions for Use
Skills Taught
• Urinary catheterization procedures
• Enema procedures
• Intramuscular injections:
Ventro-gluteal
Dorsal-gluteal
Thighs
To attach thighs:
a. Remove belly plate from manikin to locate hole in hip.
b. Remove thigh injection pad from thigh.
c. Remove wing nut and washer from bolt already inserted in thigh. (Leave
spring assembly in thigh.)
d. Insert bolt through hole in pelvis.
e. Slide washer and wing nut back over bolt.
f. Use a screwdriver to drive bolt into wing nut. (Figure 1)
Figure 1
Interchangeable Catheterization and Enema Trainer 3 Laerdal
Page 4
Genitalia
Both male and female genitalia have been provided for urinary
catheterization and enema training procedures.
Valve tubes, with a screw cap on one end, will connect urethra and anus to
corresponding reservoirs. This cap must be connected to genitalia openings.
Soft vinyl end of connector is attached to urinary and anal reservoirs with a
white C-clamp. (Long valve connects urinary reservoir; short valve connects
anal reservoir.)
Pelvic support pin must be removed and replaced immediately when
assembling and disassembling genitalia. (Figure 2)
on connector.)
Photo 1 (Slide tube to meet ridge
Figure 2
Pelvic support pin must be replaced.
Failure to replace pin will result in pelvic spread.
If this occurs, the genitalia module will no longer t properly
in the manikin.
To disassemble:
1. Lift up on belly plate.
2. Pull upper portion of genitalia back to remove pelvic support pin.
3. Belly plate, genitalia with connectors, and colon reservoir may now be
removed simultaneously.
To assemble: (Photos 1, 2 & 3)
1.Screw both valves onto genitalia. (When attaching valves, be sure
alignment of genitalia, urinary reservoir (belly plate) and colon reservoir
are as illustrated in Figure 3).
2. Lift up on edge of belly plate.
3. Pull down upper portion of genital.
4. Insert into pelvic cavity.
5. Replace the pelvic support pin.
6. Press belly plate into position.
Photo 2 (Area where clamp is placed.)
Photo 3 (Apply clamp and squeeze to tighten.)
Urinary Catheterization and Enema Simulation
Catheters should be well lubricated with manikin lubricant or full
strength liquid soap prior to using.
If catheter hangs during removal, work it with an “in and out” motion.
Use warm water in reservoirs when possible to keep soap from clogging
the catheters. Make sure reservoirs are well drained of all water before
disconnecting valves.
A size 16 French catheter is suggested for urethral catheterization.
Size 7mm catheter is suggested for enema simulation.
Figure 3
Care and Maintenance
1. Clean with mild soap and water; do not submerse manikin, module or
Long valve connects urinary reservoir; short valve connects anal
reservoir.
Laerdal 4 Interchangeable Catheterization and Enema Trainer
parts in cleaning uids or water.
2. Use only on clean surface. Avoid felt tipped markers, ink pens, acetone,
iodine or other staining products and avoid placing the module/manikin
on newsprint or inked lines of any kind.
3. To ensure longevity, each manikin and module/manikin should be cleaned
after each training session and a general inspection should be conducted
regularly.
Page 5
4. Modules and all other parts should be drained and air-dried thoroughly
before storage and disinfected when needed. After use of injection pads
(use water only), accumulated water should be squeezed out. Do not store wet foam pads in the skin. To prevent mildew or mold,
pads can be soaked in a mild solution of disinfectant and water or bleach
and water. Squeeze excess solution from pads, allow them to dry, then
store or reinsert in manikin.
5. Articulating parts will benet from a light application of talcum powder