Lab-Line 3508, 3508-1, 3508-1CE User guide

Page 1
8/00
OPERATION MANUAL
MANUAL NO. 056-707-00
REV. F
LAB-LINE®
SHAKER
MODEL NOs. 3508 3508-1 3508-1CE
DESIGNERS AND MANUFACTURERS
1999 North 15th Ave., Melrose Park, IL 60160-1491 USA
PHONE: (319) 556-2241 or (800) 522-5463; FAX: (319) 589-0516
LAB-LINE INSTRUMENTS, INC.
A SUBSIDIARY of Barnstead|Thermolyne
Page 2
TABLE OF CONTENTS
2
Page 3
SECTION TITLE
1 Introduction 2 Safe Operation 3 Description 4 Specifications 5 Installation 6 Operation 7 Maintenance 8 Replacement Parts Warranty
3
Page 4
SECTION 1
INTRODUCTION
THANK YOU
for selecting Lab-Line Instruments for your equipment needs. For maximum value and ease of start-up,
PLEASE PROCEED AS FOLLOWS:
Inspect the carton and contents for shipping damage. Notify the carrier immediately if damage is found.
Use the Accessory Checklist when unpacking to verify that the complete unit has been received. Do not discard packing materials until all is accounted for.
Read this Operation Manual thoroughly before deciding upon an appropriate location for the unit: you will want to consider the availability of power and other unit requirements, as well as user convenience.
Insist that every operator of this unit becomes familiar with the Operation Section of this manual.
Be sure to fill out the Warranty Registration Card and mail it in to Lab-Line Instruments within seven (7) days after receiving the unit.
IF
after reading this manual you should have any difficulties with the installation or operation instructions, please call:
Lab-Line Customer Relations Department (319) 556-2241 or (800) 522-5463
ALL RIGHT RESERVED
The information contained in this manual is the exclusive property of Lab-Line Instruments, Inc., and has been pro- vided solely to enable the users of the equipment described herein to operate and maintain such equipment. Any other use of this information, or the reproduction or transmission of all or any portion of this manual without prior written consent of the manufacturer is expressly prohibited. © 2000, Lab-Line Instruments, Inc.
4
Page 5
SECTION 2
SAFE OPERATION
BE ADVISED:
YOURSELF THE CONTROLS AND ACCESSORIES. WHEN OPERATING THIS EQUIPMENT, ALWAYS WEAR PROTECTIVE
CLOTHING, GLASSES AND OTHER ACCESSORIES AS SPECIFIED BY THE SAFETY REGULATIONS OF YOUR ORGANIZATION.
DO NOT MODIFY CONSTRUCTION AND/OR ASSEMBLY OF EQUIPMENT. KEEP GUARDS IN PLACE. DO NOT REMOVE TAGS, LABELS, DECALS OR OTHER INFORMATION FROM THE UNIT.
MAKE SURE ALL VESSELS ARE SECURELY CLAMPED BEFORE TURNING ON
IF SHAKING ACTION WILL RESULT IN THE EVOLUTION OF GASES OR FUMES, CARRY OUT THE OPERATION UNDER A WELL-VENTILATED LABORATORY HOOD.
TO ELIMINATE HAZARD OF ELECTRICAL SHOCK, MAKE SURE FLOOR AROUND MACHINE IS DRY IN THE EVENT OF ACCIDENTAL SPILLING OR SPLASHING OF LIQUIDS, CLEAN UP AND/OR NEUTRALIZE THE SPILLED LIQUIDS BEFORE CONTINUING.
READ OPERATION MANUAL BEFORE USING EQUIPMENT TO FAMILIARIZE
WE RECOMMEND THAT YOU DO NOT LEAVE EQUIPMENT UNATTENDED WHILE IT IS IN OPERATION. DO NOT OPERATE EQUIPMENT WITH A DAMAGED ELECTRICAL CORD OR ALLOW CORD TO COME INTO CONTACT WITH A HOT SURFACE. DO NOT OPERATE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE. DO NOT INSERT FINGERS INTO EQUIPMENT WHEN IT IS OPERATING.
UNIT. DO NOT OPERATE THE UNIT AT SPEEDS THAT WILL CAUSE THE CONTENTS OF VESSELS TO BE THROWN OUT. WHEREVER POSSIBLE, VESSELS SHOULD BE STOPPERED TO PREVENT HAZARDOUS SUBSTANCES BEING THROWN OUT DURING THE MIXING ACTION. USE EXTRA CAUTION WITH HOT OR VOLATILE SUBSTANCES SO THAT PRESSURE BUILDUP IN CONTAINER(S) DOES NOT CAUSE THE STOPPER TO BLOW OFF. DO NOT RUN EQUIPMENT WITH AN UNBALANCED LOAD. DO NOT EXCEED RECOMMENDED LOAD.
5
Page 6
SAFE OPERATION: (Con’t)
DO NOT USE EQUIPMENT FOR OTHER THAN ITS INTENDED PURPOSE. USE ONLY THE ACCESSORIES AND ATTACHMENTS THAT ARE SHIPPED WITH THE EQUIPMENT OR ARE SPECIFIED FOR IT. OTHER ACCESSORIES CAN PRODUCE HAZARDS OR MAKE THE UNIT INOPERATIVE.
DO NOT OPERATE SHAKER FOR EXTENDED PERIODS UNDER HIGH HUMIDITY CONDITIONS. DO NOT OPERATE SHAKER IN AMBIENT TEMPERATURES EXCEEDING 40ºC. DO NOT OPERATE SHAKER IN A CO2 ENRICHED ATMOSPHERE.
PERFORM REGULAR MAINTENANCE SERVICE AS SPECIFIED IN THIS MANUAL AND KEEP UNIT IN GOOD REPAIR. DO NOT OPERATE WITH KNOWN DEFECTS. ALWAYS DISCONNECT EQUIPMENT FROM POWER SOURCE BEFORE SERVICING OR PERFORMING ANY MAINTENANCE PROCEDURES.
SECTION 3
6
Page 7
DESCRIPTION
The Lab-Line Dual-Action Orbital/Reciprocating Shaker is sturdy, heavy-duty
and designed to produce either a 3/8-inch circular shaking motion or a 3/8-inch linear shaking motion. This versatility is increased through the availability from Lab-Line of interchangeable shaking platforms and clips to accommodate a variety of laboratory vessels and glassware. Rubber feet on the unit provide stability. Switching from one type of shaking motion to the other is accomplished easily by moving a control lever on the front of the unit to either of two positions: left for a reciprocating motion; right for an orbital motion. It is not necessary to stop the Shaker to change motion. Shaking speed for either of these two shaking motions can be varied and controlled from approximately 50 to 400 rpm, depending on the load of the vessels on the platform and the laboratory procedure or experiment being conducted. A solid­state speed control is provided for this purpose.
Model 3508 is a 120-volt unit and Model 3508-1(CE) is a 240-volt unit.
SECTION 4
7
Page 8
SPECIFICATIONS
POWER REQUIREMENTS: Model 3508: 120 V, 50/60 Hz, 100W, 0.8 Amps Model 3508-1(CE): 240 V, 50/60 Hz, 100W, 0.4 Amps
DUTY CYCLE: Continuous SHAKING SPEED: 50 to 400 rpm MAXIMUM LOAD: 15 lbs. (7 kg). (Heavier loads can be
shaken at lower speeds but linear motion may become erratic.)
SHAKER MOVEMENT: 3/8-inch (0.9 cm) orbital or linear.
DIMENSIONS: Overall: 11½"W x 14"D x 7¾" (29 x 36 x 20 cm) Platforms: 13" x 11" (33 x 28 cm) 18” x 18” (46 x 46 cm)
SHIPPING WEIGHT: 36 lbs. (16 kg) UNIT’S ENVIRONMENTAL OPERATING CONDITIONS:
POLLUTION DEGREE: 2 INSTALLATION CATEGORY: II ALTITUDE: 2000 Meters MSL (Mean Sea Level) HUMIDITY: 80% maximum, non-condensing ELECTRICAL SUPPLY: 120VAC or 240VAC VOLTAGE TOLERANCE: ±10% of normal rated line TEMPERATURE: 15ºC to 40ºC PRODUCT USAGE: This product is intended for use indoors only
SECTION 5
8
Page 9
INSTALLATION
SHIPPING CARTON:
This should be inspected upon delivery. When received, carefully examine for
any shipping damage before unpacking. If damage is discovered, the delivering carrier should both specify and sign for the damage on your copy of the delivery
receipt. Open the carton carefully making certain that all parts are accounted for before packaging materials are discarded— after unpacking, if damage is found, promptly report it to the carrier and request a damage inspection promptly.
IMPORTANT: Failure to request an inspection of damage within a few days
after receipt of shipment absolves the carrier from any liability for damage: you must
call for a damage inspection promptly. LOCATION:
Place the shaker on a level surface in a convenient location and near a
properly grounded electrical power outlet that matches the requirements of the unit
indicated on its nameplate. A minimum of 1-inch of clearance is required on all sides
to allow unobstructed orbiting. PLATFORM INSTALLATION:
Select the appropriate platform for the laboratory vessels or glassware to be
shaken. Tighten the 4 thumbscrews at each corner to secure the platform to the
shaker. Make sure both sides of the platform are outside of the mounting plate. ELECTRICAL REQUIREMENTS:
Model 3508 requires a 120 VAC, 50/60 Hz power source, Model 3508-1(CE), 240 VAC, 50/60 Hz.
The shaker is supplied with a 3-wire line cord and it is necessary to
connect it to an outlet wired for 3-prong plugs. If an extension cord is used, it also
should be the 3-wire grounded type. For an outlet designed to accept 2-prong plugs (ungrounded), it is required
that a qualified electrician replace the outlet with a new grounded outlet. If a plug
must be installed, use only the 3-prong grounded type, rated for the unit load
requirements and matching the power outlet. Make sure the green ground wire is
secured to the plug ground terminal. It is best to leave the unit disconnected when not in use.
9
Page 10
SECTION 6
OPERATION
DANGER: DO NOT USE IN THE PRESENCE OF FLAMMABLE OR COMBUSTIBLE MATERIALS OR EXPLOSIVE GASES. FIRE OR EXPLOSION MAY RESULT CAUSING DEATH OR SEVERE INJURY
NOTE: DEPRESS METAL TAB PRIOR TO ROTATING KNOB—THE TAB FIXES KNOB SETTING AND PREVENTS INADVERTENT MOTION OF THE KNOB.
METAL TAB
Located on the control panel are power switch, control lever for 2 types of
shaking motion and the speed control knob. The shaker is off when the power switch
is not lit. Keep speed control knob turned to a full counterclockwise position before
starting unit. STARTING THE SHAKER:
Select shaking platform and fasten securely to shaker. Insert vessels to be
shaken. Be careful to fill vessels to levels that will not splash or spill as a result of the
shaking action. Whenever possible, use stoppers in vessels to prevent accidental
spills. Select type of shaker motion by pushing lever (with knob) either to the left for reciprocating motion or to the right for orbital motion. Press in top half of rocker switch to start shaker.
10
Page 11
OPERATION: (Con’t)
SETTING THE SPEED CONTROL:
Rotate speed control knob slowly clockwise until the desired shaking speed is
attained. The operator should use caution in setting the speed, so that the vessels
do not come out of their holders or spill their contents. The maximum safe speed
varies with the load. A heavier load will need to be shaken more slowly. To stop the shaker, turn the speed control fully counterclockwise and press in the lower half of the power switch. If the shaker will not be used again soon, disconnect the plug from its outlet.
CONTROL PANEL
POWER SPEED CONTROL
#440-375-00 (SWITCH)* # 560-223-00 (KNOB)*
*AS LISTED ON UPCOMING REPLACEMENT PARTS LIST
ON
11
Page 12
SECTION 7
MAINTENANCE
BE ADVISED:
NOTE: MAKE NO ATTEMPT TO SERVICE OR REPAIR A LAB-LINE PRODUCT UNDER
WARRANTY BEFORE CONSULTING YOUR LAB-LINE DEALER. AFTER THE WARRANTY
PERIOD, SUCH CONSULTATION IS STILL ADVISED, ESPECIALLY WHEN THE REPAIR MAY BE
TECHNICALLY SOPHISTICATED OR DIFFICULT.
IF ASSISTANCE IS NEEDED BEYOND WHAT THE DISTRIBUTOR CAN PROVIDE, PLEASE CALL
THE LAB-LINE CUSTOMER RELATIONS DEPARTMENT AT (319) 556-2241 OR (800)
522-5463. NO MERCHANDISE, HOWEVER, SHOULD BE RETURNED DIRECTLY TO LAB-LINE
WITHOUT PRIOR APPROVAL FROM LAB-LINE.
PREVENTIVE MAINTENANCE (with every 3 months of constant use): WARNING: TO AVOID RISK OF ELECTRIC SHOCK, DISCONNECT
EQUIPMENT FROM POWER SOURCE BEFORE PERFORMING ANY MAINTENANCE.
Remove the platform by loosening the 4 thumbscrews. Remove the platform mounting plate by removing the 4 flathead screws that hold it to the rubber mounts. Inspect the drive belt and pulleys for wear. Order a replacement if necessary.
Check the 4 rubber mounts for cracks, wear or looseness. If these conditions are noted, replace or tighten the mounts.
Remove 4, 1/4-inch hex bolts from the drive plate. Lift the plate out and turn it over to expose the motor—take care not to strain the wires. 2 brushes are located under plastic caps on opposite sides of the lower part of the motor. Unscrew the plastic caps and slide the brushes out. Replace the brushes when they are worn down to 3/16 of an inch or less in length.
Reassemble the unit carefully by reversing the preceding steps.
NOTE: IF A SINGLE BRUSH IS WORN, BOTH MUST BE REPLACED.
MAINTENANCE: (Con’t)
12
Page 13
REPLACING PARTS: SPEED CONTROL WARNING: TO AVOID RISK OF ELECTRIC SHOCK, DISCONNECT
EQUIPMENT FROM POWER SOURCE BEFORE PERFORMING ANY MAINTENANCE.
Speed control replacement requires removal of the control panel. Remove the 2 screws that hold the panel and bring the panel forward.
To replace the speed control, remove the control knob (2 setscrews) and a retaining nut to separate it from the panel. Unplug the terminal connections from the old control unit and plug them into the new one. Reverse this sequence to put the control unit and panel back in place.
REPLACING PARTS: SPEED CONTROL CALIBRATION WARNING: BE CAREFUL TO AVOID INADVERTENT CONTACT WITH LIVE
ELECTRICAL PARTS WHILE PERFORMING THESE OPERATIONS.
Turn on the power. Slowly rotate the speed control knob clockwise—verify shaker rotates counterclockwise.
Speed control adjustment: Rotate speed control knob fully counterclockwise. Adjust speed control MIN pot for “just off” shaking.
Adjust the speed control to maximum clockwise setting. Using the tachometer, read the RPM—it should read 400 RPM. If not, adjust the MAX pot so that the shaker runs at 400 RPM with the speed control at the maximum clockwise setting.
REPLACING PARTS: SWITCH
Switch replacement requires removal of the control panel. Remove the 2 screws that hold the panel and bring the panel forward. To replace the switch, disconnect the wires. The switch is a snap-in type. Squeeze the tangs together and push the switch through the front of the panel. Install the new switch and replace the wires.
MAINTENANCE: (Con’t)
REPLACING PARTS: BELT
13
Page 14
Replacement of the belt is accomplished by removing the platform mounting plate to expose the belt. Remove the old belt. Clean the pulley grooves with a solvent such as alcohol or mineral spirits. Install the new belt by stretching over the pulleys, center-distance is factory set and belt requires no further tensioning. Plug unit in and run slowly.
CAUTION: WITH PLATFORM MOUNTING PLATE REMOVED, UNIT IS OUT OF BALANCE AND WILL” WALK” IF IT IS NOT ATTENDED. BE SURE SPEED CONTROL IS AT ITS LOWEST SETTING BEFORE STARTING AND THEN INCREASE SPEED SLOWLY. DO NOT LEAVE UNIT UNATTENDED WHILE IT IS RUNNING.
With the unit unplugged, reassemble the unit carefully by following the reverse order from the above.
ABOUT LUBRICATION OF BEARINGS:
The shaking mechanism is equipped with permanently lubricated
bearings that do not require further lubrication or adjustment.
LUBRICATION OF GUIDES AND ROLLERS: (See next page for illustration)
Turn the platform mounting plate upside down. Wipe off old grease from the cam guide and cam rollers. Re-grease the guide and rollers.
MAINTENANCE: (Con’t)
REPLACING PARTS: (Con’t)
14
Page 15
VIEW OF CAM GUIDE AND ROLLERS:
CAM ROLLERS CAM GUIDE
#130-196-00*
*AS LISTED ON UPCOMING REPLACEMENT PARTS LIST
15
Page 16
SECTION 8
REPLACEMENT PARTS
DESCRIPTION PART NUMBER Bearing Plate Assembly: 014-966-00
Cam Roller Guide: 130-196-00 Cordset: 470-132-00 Drive Belt: 150-299-00 Eccentric Shaft Crank Assembly: 150-298-00 Knob, Selector: 560-015-00 Motor: 370-275-00 Motor Brushes (set of 2): 370-277-01 Mount, Flexible (4 required): 790-296-00 Platform Mat: 790-316-11 Pulley, Motor: 150-297-00 Speed Control Assembly: 227-598-00 Speed Control Knob: 560-223-00 Switch, Rocker: 440-375-00 Transformer (240V unit): 460-139-00 Mounting Feet:: 790-087-00 Wiring Schematics: 3508, 3508-1: 227-783-00 3508-1CE: 228-823-00
NEED A PART? CALL THE LAB-LINE PARTS HOTLINE. CALL: (319) 556-2241 or (800) 522-5463; FAX: (319) 589-0516.
LAB-LINE RESERVES THE RIGHT TO MAKE CHANGES TO SPECIFICATIONS WITHOUT PRIOR NOTICE.
16
Page 17
WARRANTY
Page 1 of 2
LAB-LINE INSTRUMENTS, INC. (“Lab-Line”) warrants that the product manufactured by Lab­Line shall be free of defects in materials and workmanship for a period of time defined on the following page from the first to occur of (i) the date the product is sold by Lab-Line or (ii) the date the product is purchased by the original retail customer (the “Commencement Date”). Except as expressly stated above,
LAB-LINE MAKES NO OTHER WARRANTY, EXPRESSED OR IMPLIED, WITH RESPECT TO THE PRODUCTS AND EXPRESSLY DISCLAIMS ANY AND ALL WARRANTIES, INCLUDING BUT NOT LIMITED TO, WARRANTIES OF DESIGN, MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
An authorized representative of Lab-Line must perform all warranty inspections. In the event of a defect covered by Lab-Line’s warranty, Lab-Line shall, as its sole obligation and exclusive remedy, provide free replacement parts to remedy the defective product. In addition, for products sold by Lab-Line within the continental United States or Canada, Lab­Line shall provide free labor to repair the products with the replacement parts, but only for a period of ninety (90) days from the Commencement Date.
Lab-Line’s warranty provided hereunder shall be null and void and without further force or effect if there is any (i) repair made to the product by a party other than Lab-Line or its duly authorized service representative, (ii) misuse (including use inconsistent with written operating instructions for the product), mishandling, contamination, overheating, modification or alteration of the product by any customer or third party or (iii) use of replacement parts that are obtained from a party who is not an authorized dealer of Lab­Line.
IN NO EVENT SHALL LAB-LINE BE LIABLE TO ANY PARTY FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, OR FOR ANY DAMAGES RESULTING FROM LOSS OF USE OR PROFITS, ANTICIPATED OR OTHERWISE, ARISING OUT OF OR IN CONNECTION WITH THE SALE, USE OR PERFORMANCE OF ANY PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, TORT (INCLUDING NEGLIGENCE), ANY THEORY OF STRICT LIABILITY OR REGULATORY ACTION.
The name of your nearest authorized Lab-Line dealer may be obtained by calling 1-800­522-5463.
1999 North 15th Ave., Melrose Park, IL 60160-1491 USA
PHONE: (319) 556-2241 or (800) 522-5463; FAX: (319) 589-0516
DESIGNERS AND MANUFACTURERS
A SUBSIDIARY of Barnstead|Thermolyne
17
Page 18
WARRANTY
Page 2 of 2
12 MONTH PARTS WARRANTY:
All Environmental Chambers
Low Temperature B. O. D. Incubators
Animal Study Chamber
Controlled Environment Centers
Biological Work Station
Refrigerators, Freezers
Chromatography Refrigerators (5 year parts warranty on compressor only)
Large Capacity Refrigerators and Freezers (5 year parts warranty on compressor
only)
24 MONTH PARTS WARRANTY:
Frame Clamps, Frame Sets, Lab Jacks
Saybolt Viscosimeter
Timers, Samplers, Flasks
Saf-T-Shield, Safety Tongs
All Incubators & Ovens
Dual Action Open Air Shaker
Reciprocating Shakers (open air and water bath)
Rockers and Rotators
Low Cost Shakers
Environ Blok Shaker
Titer Plate Shaker
Multi Wrist Shaker
Water Baths (excluding Aquabaths), Ultrasonic Cleaners
Slide Warmers
Mixers, Stirrers, Hotplates
Thermal Cyclers
Blok Heaters
Aquabaths, lifetime warranty on heaters
LIFETIME PARTS WARRANTY:
All ORBITAL Shakers (not carrying a 24 month parts warranty) offer a lifetime parts warranty on the drive mechanism and a 5 year warranty on all other parts
Refrigerated Orbital Shakers carry a lifetime warranty on the drive mechanism, 1 year parts warranty on the compressor, and a 5 year warranty on all other parts.
18
Page 19
FIRST IN INSTRUMENTS SERVICING SCIENCE, INDUSTRY, RESEARCH
AND EDUCATION
SINCE 1908.
19
Page 20
20
Page 21
ACCESSORY CHECKLIST
The following loose parts and accessories are packed with this unit. Before discarding any packing materials, please be sure that nothing has been overlooked.
MODEL NOs. : 3508, 3508-1, 3508-1CE
CHECKED BY: __________ DATE __________ PACKED BY __________
CHECKED ITEM PART NUMBER QUANTITY
_________ Operation Manual 056-707-00 1 _________ Thumb Screw 562-184-10 4 _________ Warranty Card 528-020-00 1
21